Dräger Babylog® VN600
Neonatal Intensive Care Ventilation
The Babylog® VN600 offers a new user interface and design, enhancing operation ease and efficiency. This neonatal ventilator supports lung and brain protective ventilation modes throughout the respiratory cycle and integrates seamlessly into a developmental care-friendly workplace.
Key features include:
- Operation Principle and User Interface: A brilliant user interface with glass touch technology for intuitive operation and reduced training time. Features include quick and safe access to settings and clinical data, full patient data recording and export via USB, multiple view configurations, step-by-step guidance, and a 360° alarm light for clear visibility of alarm priorities.
- Lung and Brain Protective Ventilation: Supports various ventilation strategies to prevent lung injury and neurological impairment, including High-Flow Oxygen Therapy, Volume Guarantee (HFO-VG), PC-MMV/VG+PS, and original Dräger leak adaptation technology.
- Care-Centered Workplaces: Designed to support the complex needs of developing lungs, brains, and other organs in NICU environments. Products are compatible, offer flexible workplace integration, low operational noise levels, effective infection prevention through easy cleaning, and longer circuits for kangaroo time.
- Connectivity: Future-proof connectivity through CC300, enabling interoperability with other medical devices via the SDC network protocol. Features full HL7 data export to HIS and secure communication.
- Comprehensive Services: Includes product service, maintenance, IT consulting, system integration, training, multivendor service, digital services, and access to the BabyFirst community.
Physical Specifications
Dimensions and weights of the Babylog VN600:
Designation | Weight |
---|---|
Ventilation unit | approx. 16 kg (34.9 lbs) |
Display unit | 5 kg (11 lbs) |
Ventilation unit and display unit | approx. 21 kg (46.3 lbs) |
Trolley | approx. 33 kg (72.8 lbs) |
PS500 | approx. 27 kg (59.5 lbs) |
GS500 | approx. 10.5 kg (23 lbs) |
Nominal weight (weight of the ventilation unit and display unit on trolley) | 58 kg (128 lbs) |
Maximum weight (permitted maximum total weight) | 133 kg (293 lbs) |
Neonatal Ventilation Accessories
Dräger original neonatal ventilation accessories are suitable for various ventilation strategies and offer optimized length, pressure transmission, and humidification performance. They are fully compatible with the Babylog neonatal ventilator. A neonatal accessories catalogue is available.
Related Products
- Dräger Babyleo® TN500: An IncuWarmer providing optimal thermoregulation for neonates in open, closed, and transition care.
- Isolette® 8000 plus: Offers advanced thermoregulation for a stable, cocoon-like environment, with continuous monitoring of central and peripheral body temperature.
- Dräger Jaundice Meter JM-105: Provides consistent quality screening for jaundice management, optimizing efficiency and saving time and money.
- BiliLux: A compact LED phototherapy light system for treating neonatal unconjugated hyperbilirubinemia, offering superior performance and electronic documentation.
- Seattle PAP plus – Bubble CPAP System: Helps infants with respiratory distress breathe easier by utilizing Bubble CPAP therapy with oscillatory effects.
Technical Data
Ventilation Settings
- Ventilation mode: Includes Pressure controlled ventilation (PC-CMV, PC-BIPAP/SIMV+, PC-SIMV, PC-AC, PC-APRV, PC-PSV, PC-HFO, PC-MMV) and Support of spontaneous breathing (SPN-CPAP/PS, SPN-CPAP/VS, SPN-CPAP, SPN-PPS).
- Enhancements: Volume Guarantee/HF-Volume Guarantee, Smart Pulmonary View, Automatic Tube Compensation (ATC®), APRV-AutoRelease®, Apnoea ventilation, Automatic flow adjustment.
- Special procedures: Suction manoeuvre, Manual inspiration/hold, Medication nebulisation.
- Therapy types: Invasive ventilation (Tube tracheostomy), Non-invasive ventilation (NIV), O2-therapy.
Performance Data
- Control principle: Time-cycled, pressure limited, continuous flow.
- Length of intermittent PEEP: 1 to 20 expiratory cycles.
- Medication nebulisation: For 5, 10, 15, 30 minutes, continuously (∞).
- Inspiratory flow: Paediatrics Max. 60 L/min, BTPS; Neonates Max. 30 L/min, BTPS.
- Base flow: 3 L/min (paediatric), 6 L/min (neonates).
- Inspiratory safety valve: Opens if compressed gas supply fails, enabling spontaneous breathing with ambient air.
Operating Data
- Mains power supply: 100 V to 240 V, 50/60 Hz.
- Current consumption: Max. 1.3 A (at 230 V), Max. 3.0 A (at 100 V).
- Power consumption: Maximum 300 W; Ventilation unit with display unit approx. 100 W; with GS500 approx. 180 W.
Gas Supply
- O2 positive operating pressure: 2.7 to 6.0 bar.
- Air operating pressure: 2.7 to 6.0 bar.
Battery Details
- Internal battery: Without GS500: 30 minutes; With GS500: 15 minutes.
- PS500 batteries: Without GS500: 240 minutes; With GS500: 120 minutes.
Screen Values
- Babylog VN600 diagonal screen size: 15.6 inches.
- Input / Output ports: 3 external RS232, 4 USB, 1 LAN port.
- Touchscreen technology: Capacitive touchscreen with glass front.
- Aspect ratio: 16:9.
- Resolution: 1366 x 768 pixels.
- Digital machine output: Via RS232 C interface (Dräger MEDIBUS®).
Technical Data Summary
Parameter | Value |
---|---|
Respiratory rate (RR) | 0.5 to 150/min |
Inspiratory time (Ti) | 0.1 to 3 s |
Tidal volume (VT) | 2 to 300 mL |
Inspiratory pressure (Pinsp) | 1 to 80 mbar |
PEEP | 0 to 35 mbar |
O2 concentration (FiO2) | 21 to 100 Vol.% |
Trigger threshold | 0.2 to 5 L/min |
Automatic Tube Compensation (ATC®) | 0 to 100 % tube compensation |
High Frequency Oscillation (PC-HFO) | Frequency: 5 to 20 Hz; Pressure amplitude: 5 to 90 mbar |
Leakage compensation | On, off |
O2-therapy | Continuous Flow 2 to 50 L/min |
Airway pressure measurement | Range -60 to 120 mbar |
Flow Measurement (proximal) | Range -40 to 40 L/min |
Minute volume measurement | Range 0 to 30 L/min, BTPS |
Respiratory rate measurement | 0 to 300/min |
O2 measurement (inspiratory side) | 18 to 100 Vol% |
CO2 measurement in main flow | 0 to 100 mmHg |
Dynamic compliance (Cdyn) | 0 to 100 mL/mbar |
Resistance (R) | 0 to 1000 mbar/L/s |
Airway resistance of the patient (Rpat) | 0 to 1000 mbar/L/s |
Leakage minute volume (MVleak) | 0 to 30 L/min, BTPS |
Rapid shallow breathing index (RSBI) | 0 to 9999 (/min/L) |
Airway pressure (Paw) | -30 to 100 mbar |
Alarms / Monitoring | Various parameters with High/Low settings, Apnoea alarm time: 5 to 60 seconds, Disconnection alarm time: 0 to 60 seconds. |
Notes
Not all products, features, or services are available in all countries. Mentioned trademarks are registered in certain countries only. Visit www.draeger.com/trademarks for current status.
Manufacturer: Drägerwerk AG & Co. KGaA, Moislinger Allee 53-55, 23558 Lübeck, Germany.
Contact information for various regional offices (DACH, Europe, Middle East, Africa, Asia Pacific, Central and South America) is provided, along with a link to locate regional sales representatives.