Addendum to: Cardiosave IABP Operating Instructions – Clinical Considerations

IAB IFU Balloon Membrane Perforation

For use of Cardiosave Hybrid IABP and Cardiosave Rescue IABP with Linear, Mega, Sensation, Sensation Plus, Yamato Plus-R, Trans-Ray and Trans-Ray Plus IABs.

A. Balloon Membrane Perforation

Balloon membrane perforation may be caused by:

If perforation occurs, blood may be visible in the IAB catheter. If balloon membrane perforation occurs or is suspected it may also be evidenced by: 1) bright red blood, dried blood particles or serosanguineous fluid seen in the extracorporeal tubing or catheter extender; 2) a sudden change in the diastolic augmentation pressure waveform and/or 3) certain IAB pump alarms. Do not bypass these alarms, and pay close attention to the alarm notifications. The following Cardiosave alarms are designed to assist the user to identify and address a perforated balloon earlier, which may prevent blood from traveling into the IABP:

Periodically check the IAB catheter tubing for blood both throughout therapy and when the above alarms occur. If any blood is noted or perforation is suspected, the following procedure must be performed immediately:

  1. Stop pumping by placing IABP console in Standby.
  2. Disconnect the catheter extender tubing from the IABP console to allow the balloon to deflate.
  3. Clamp extracorporeal tubing between white y-fitting and male connector.
  4. Notify physician, and prepare for IAB catheter removal.
  5. Consider IAB catheter replacement, if the patient's condition warrants.
  6. If blood is suspected of having entered the pump, take pump out of service. It should be evaluated before use in another patient by Biomed/Technical Service to determine if replacement of contaminated components is necessary.

WARNINGS

  • If you continue to pump an IAB catheter with a leak, gaseous embolic injury of organs may result or a large blood clot may form within the balloon membrane requiring surgical removal of the IAB catheter.
  • Do not inflate the IAB using a syringe or any other means if a balloon membrane leak is suspected.
  • Should an IAB membrane perforation occur, blood may be permitted to travel into the IABP console. To prevent blood loss and damage to the IABP console, stop therapy, disconnect catheter extender tubing from the IABP console, and clamp the extracorporeal tubing.
  • Perforation of an IAB membrane may indicate that the patient's vascular condition may induce abrasion or perforation in subsequent IAB membranes.
  • An unexpected shutdown due to a blood back event may threaten the hemodynamic stability of the supported patient as the User is left unaware to the status of the IABP. Additionally, any subsequent attempts to use a console that experienced a blood back event without reconditioning may delay future therapy delivery.
  • The user and subsequent maintenance or service personnel can be exposed to an unexpected biohazard should proper containment precautions not be taken.
  • Subsequent patients may be exposed to an unexpected biohazard should an impacted console not be appropriately serviced prior to use.
For the Cardiosave Intra Aortic Balloon Pump:
  • During hospital use, it is advised another Cardiosave be available to provide therapy should the Cardiosave in use be compromised.
  • It is advised not to transport a patient receiving counterpulsation therapy via Cardiosave unless the clinician deems the benefit of transport outweighs the risk of unexpected shutdown.

Note: The length of time a balloon membrane can survive such contact with plaque or unusual folding is unpredictable. A leak in an IAB catheter within the bloodstream may allow gas to enter the patient's bloodstream which may result in patient injury. Large perforations are rare, therefore the small quantity of gas released is usually asymptomatic. The rate of incidence at each individual hospital may be influenced by the degree of vascular disease in that patient population, by the location of the IAB catheter in the aorta, or by using a balloon membrane size inappropriate for the specific patient.

Models: Cardiosave Hybrid IABP Intra Aortic Balloon Pumps, Cardiosave, Hybrid IABP Intra Aortic Balloon Pumps, IABP Intra Aortic Balloon Pumps, Intra Aortic Balloon Pumps, Aortic Balloon Pumps, Balloon Pumps, Pumps

File Info : application/pdf, 2 Pages, 58.04KB

mca000000556 us addendum-en-us canada

References

Adobe PDF Library 17.0 Adobe InDesign 18.5 (Windows)

Related Documents

Preview Cardiosave IABP Balloon Membrane Perforation: Causes, Signs, and Procedures
This addendum to the Cardiosave IABP Operating Instructions provides clinical considerations regarding balloon membrane perforation. It details causes, signs of perforation, immediate procedures, and critical warnings for healthcare professionals.
Preview IAB Insertion / Cardiosave IABP Operation: Quick Reference Guide for Software B.17
A quick reference guide for the insertion and operation of the Cardiosave IABP with software version B.17, detailing procedures for both Sensation Plus, Sensation, Mega, and Linear IAB catheters, including sizing, preparation, insertion (sheathed and sheathless), pressure monitoring, and initial setup.
Preview URGENT MEDICAL DEVICE CORRECTION: Datascope IABP Battery Safety and Maintenance
Getinge issues an urgent medical device correction for Datascope Intra-Aortic Balloon Pumps (IABP) models Cardiosave Hybrid, Rescue, CS300, and CS100 concerning battery usage, charging, maintenance, and storage. Learn about potential risks, interim user actions, and Getinge's corrective measures to ensure patient safety.
Preview Cardiosave IABP Balloon Membrane Perforation: Clinical Considerations and Operating Procedures Addendum
Detailed addendum to Cardiosave IABP operating instructions covering clinical considerations for balloon membrane perforation, including causes, signs, procedures, and warnings.
Preview Getinge Datascope Cardiosave IABP Cybersecurity Vulnerabilities (Ripple20) - Field Safety Notice
Urgent field safety notice regarding cybersecurity vulnerabilities (Ripple20) affecting Getinge Datascope Cardiosave Hybrid and Rescue Intra-Aortic Balloon Pumps (IABP). Details interim actions and corrective actions.
Preview Cardiosave Enhanced Blood Detection D.03 Software: Features and Safety Updates
This document details the Cardiosave Enhanced Blood Detection D.03 software update from Getinge, covering new safety features like the Blood Surveillance Algorithm and Blood Containment Filter, and providing guidance on managing IAB catheter perforations and related alarms.
Preview Cardiosave Battery Operation Quick Reference Guide
A quick reference guide detailing the operation, functionality, alarms, and maintenance of the lithium-ion batteries for the Cardiosave Intra-aortic Balloon Pump. It covers battery location, status monitoring, charging procedures, in-use indicators, and important safety warnings.
Preview Getinge Clean Cleaning Chemistries: Solutions for Healthcare and Life Science Reprocessing
Explore Getinge Clean's comprehensive range of high-performance cleaning chemistries designed for healthcare, pharmaceutical production, and life science laboratory applications. Ensure safety, efficiency, and optimal results in instrument reprocessing.