User Manual for ETS-LINDGREN models including: 8000-014 RF Power Amplifier, 8000-014, RF Power Amplifier, Power Amplifier, Amplifier

8000-014 1791479

tdMcAda

RF Power Amplifier

29 mag 2024 — • User Manual (This document). • Remote Control Interface Manual. • Remote ... material for the carrier's inspection. Check that the equipment is ...20 pagine


File Info : application/pdf, 20 Pages, 377.94KB

PDF preview unavailable. Download the PDF instead.

8000-014 1791479
Model 8000-014
RF Power Amplifier
User Manual

ETS-Lindgren Inc. reserves the right to make changes to any products herein to improve functioning or design. Although the information in this document has been carefully reviewed and is believed to be reliable, ETS-Lindgren does not assume any liability arising out of the application or use of any product or circuit described herein; nor does it convey any license under its patent rights nor the rights of others. All trademarks are the property of their respective owners.
© Copyright 2024 by ETS-Lindgren Inc. All Rights Reserved. No part of this document may be copied by any means without written permission from ETS-Lindgren Inc.
Trademarks used in this document: The ETS-Lindgren logo is a registered trademark and Calibration Services Plus is a trademark of ETS-Lindgren Inc.

Revision A

Revision Record MANUAL, 8000-014 | Part #1791479, Rev. A

Description

Date

Initial Release

May, 2024

ii

www.ets-lindgren.com

Table of Contents

Notes, Cautions, and Warnings ........................................................................iv Safety Information .............................................................................................iv 1.0 Introduction ...................................................................................................6
Standard Configuration....................................................................................................... 6
2.0 Maintenance ..................................................................................................8
Service Procedures ............................................................................................................ 8 Contacting ETS-Lindgren............................................................................................. 8 Sending a Component for Service ............................................................................... 8
3.0 Specifications..............................................................................................10
Electrical Specifications .................................................................................................... 10 Physical Specifications ..................................................................................................... 10 Environmental................................................................................................................... 11
4.0 Before You Begin ........................................................................................ 12
Unpacking......................................................................................................................... 12 Safety Information ............................................................................................................ 12 Warnings........................................................................................................................... 13
5.0 Installation ................................................................................................... 14
Mains Input ....................................................................................................................... 14 Amplifier Protection .......................................................................................................... 14
6.0 Operation ..................................................................................................... 16
Equipment Use ................................................................................................................. 16 Mains Switch..................................................................................................................... 16 Standby/Operate Switch (Centre Unit) ............................................................................. 16 Fault Indicator (Red LED)................................................................................................. 17 Interlock Indicator (Orange LED)...................................................................................... 17 Local Lockout Indicator..................................................................................................... 17 RF Input (Rear Panel Mounted) ....................................................................................... 17 RF Output (Rear Panel Mounted) .................................................................................... 17 Air Inlet.............................................................................................................................. 18 Safety Interlock ................................................................................................................. 18 Fan Outlet ......................................................................................................................... 18
Appendix A: EC (UKCA) Declaration of Conformity ...................................... 20

www.ets-lindgren.com

iii

Notes, Cautions, and Warnings
Note: Denotes helpful information intended to provide tips for better use of the product.
CAUTION: Denotes a hazard. Failure to follow instructions could result in minor personal injury and/or property damage. Included text gives proper procedures.
WARNING: Denotes a hazard. Failure to follow instructions could result in SEVERE personal injury and/or property damage. Included text gives proper procedures.
Safety Information
Warning: Avoid exceeding the stated maximum input level as this can cause a risk of damaging the equipment.
Warning: There is a significant risk of an electrical or burn hazard, to avoid this always ensure the output is terminated prior to switching on the equipment.
Protective Earth Ground (Safety Ground): Indicates protective earth terminal. You should provide uninterruptible safety earth ground from the main power source to the product input wiring terminals, power cord, or supplied power cord set.
Waste Electrical and Electronic Equipment (WEEE) Directive: (European Union) At end of useful life, this product should be deposited at an appropriate waste disposal facility for recycling and disposal. Do not dispose of with household waste.

iv

www.ets-lindgren.com

This page intentionally left blank.

www.ets-lindgren.com

v

1.0 Introduction
The ETS-Lindgren Model 8000-014 RF Power Amplifier is capable of supplying a minimum of 200 watts into a 50 ohm load over the frequency range 10kHz to 250MHz. The amplifier is designed with sufficient gain so that it may be used with normal output levels of signal sources.
 A safety interlock on the rear panel is also provided, which will mute the amplifier when grounded.
 The unit is powered by a switched mode power supply for high efficiency, high power factor, and a wide voltage range operation. The unit is air cooled with integral fans and is protected against faulty cooling by excess temperature sensing.
 A front panel fault indicator is provided to indicate over-temperature.
The amplifier is designed for rugged operation into a variety of loads and is primarily intended for use as an RF power source for EMC susceptibility testing, but is also applicable to other systems requiring a wide-band linear amplifier.
This amplifier is designated as Professional Equipment and should not be operated by untrained staff. As this product has the capability to generate high levels of RF energy it is not intended for use in a residential environment.
Standard Configuration
 The 8000-014 amplifier is supplied for installation into a19-inch rack.
 The RF connectors are N type female which are located on the rear panel of the amplifier.

6

Introduction

www.ets-lindgren.com

This page intentionally left blank.

www.ets-lindgren.com

Introduction

7

2.0 Maintenance
If you have any questions concerning maintenance or if the amplifier system is misoperating, contact ETS-Lindgren Customer Service. See Service Procedures for contact information.
CAUTION: Before performing any maintenance, follow the information provided in Before You Begin on page 13.
WARNING: There are no user serviceable parts within the RF Power Amplifier system. To prevent electrical shock, DO NOT remove covers. Warranty will be void if the covers are opened.
ALWAYS UNPLUG THE UNIT BEFORE CLEANING. Clean the exterior of
WARRANTY
the cabinet using a damp cloth.

Service Procedures
CONTACTING ETS-LINDGREN
Note: Please see ets-lindgren.com for a list of ETS-Lindgren offices, including phone and email contact information.

SENDING A COMPONENT FOR SERVICE
1. Contact ETS-Lindgren Customer Service to obtain a Service Request Order (SRO).
2. Briefly describe the problem in writing. Give details regarding the observed symptom(s) or error codes, and whether the problem is constant or intermittent in nature. Please include the date(s), the service representative you spoke with, and the nature of the conversation. Include the serial number of the item being returned.
3. Package the system or component carefully. If possible, use the original packing materials or carrying case to return a system or system component to ETS-Lindgren.

8

Maintenance

www.ets-lindgren.com

This page intentionally left blank.

www.ets-lindgren.com

Maintenance

9

3.0 Specifications
Electrical Specifications
Note: The third order intercept point is a nominal value as its calculation depends upon the power level at which distortion measurements are made.
Note: Output VSWR tolerance is specified for excitation within the permitted levels and frequency range.

Frequency Range (Instantaneous)

10kHz ­ 250MHz

Rated Output Power

250W Min (10kHz ­ 80MHz) 200W Min (80MHz ­ 250MHz)

200W Min (10kHz ­ 80MHz) Output Power at 1dB Gain Compression
150W Min (80MHz ­ 225MHz)

Gain

54dB Minimum

Third Order Intercept Point

64dBm

Gain Variation with Frequency

+/- 2dB

Harmonics at 550W Output Power

Better than -20dBc

Output Impedance

50 Ohms

Stability

Unconditional

Output VSWR Tolerance

Infinity:1

Input VSWR

2:1 (Max)

RF Connector Style

Type N Female

Safety Interlock

2 x BNC Female, s/c and o/c to mute

USB/GPIB/Ethernet Interface

Standard

Physical Specifications

Case Dimensions Weight

10

Specifications

19 inch, 4U , 500mm deep 18 kg (39.6 lb)
www.ets-lindgren.com

Environmental
Operating Temperature Range Altitude Humidity

0-40°C Up to 2000m above sea level 95% (Non-condensing)

www.ets-lindgren.com

Specifications

11

4.0 Before You Begin
CAUTION: Before connecting any components, follow the information provided in Safety Symbols on page v.
Unpacking
The 8000-014 amplifier is supplied with the following contents:
 Power Cable (10A)
 Spare Fuse (T 10A H 250V 20mm)
 User Manual (This document)
 Remote Control Interface Manual
 Remote Control Interface CD
 USB Cable
 Calibration Test Report
If any signs of damage are found no attempt should be made to install or operate the amplifier system. Contact ETS-Lindgren Customer Service immediately to report the problem.
If the shipping carton has been damaged, retain the shipping carton and packing material for the carrier's inspection. Check that the equipment is complete and is in an undamaged condition, and then contact ETS-Lindgren Customer Service for further advice.
Safety Information
 This RF amplifier has been designed and tested in accordance with BS EN61010-1 and has been supplied in a safe condition. This manual contains warning information which must be followed to ensure safe operation and to retain the apparatus in a safe condition.
 This apparatus does not incorporate components liable to explode or implode during normal operating conditions.
 In normal operating conditions this apparatus does not liberate injurious or poisonous gases.

12

Before You Begin

www.ets-lindgren.com

 Sound levels of this apparatus in a rack are below 85dBA as required by EN61010-1.
 If local regulations have a lower limit, appropriate action should be taken in line with the local regulations so as to comply.
 This apparatus is of Installation Category 2.
Warnings
WARNING: This apparatus is capable of delivering harmful levels of radio frequency power. Ensure at all times during operation that the RF output is properly terminated with an adequately rated termination or transducer and that the cables and connectors attached to the apparatus are in good condition.
The use of suitably rated mains cords must only be used as specified within this document, see Rear Panel, Mains Input on Page 15.
The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. The protective action must not be negated by the use of an extension cord without a protective conductor.
The opening of covers or removal of parts is likely to expose live parts.
This apparatus must only be serviced by qualified service personnel and must be disconnected from all voltage sources before it is opened for adjustment, replacement, maintenance, or repair.
Make sure that only fuses of the required rated current and of the specified type are used for replacement. The use of makeshift fuses and the short-circuiting of fuse holders are prohibited.

www.ets-lindgren.com

Before You Begin

13

5.0 Installation
CAUTION: Before connecting any components, follow the information provided in Before You Begin on page 13.
Mains Input  Mains input is via a 10A power cable (supplied).  Access for mains disconnection must be maintained at all times.
Amplifier Protection  The amplifier is protected by a fuse of the following rating, T 10A H 250V 20mm.  This instrument must be earthed. WARNING: If the rack mounting option has been selected, the unit must be installed in the rack using a suitable tray or runners. The amplifier must not be supported only by the front panels.

14

Installation

www.ets-lindgren.com

This page intentionally left blank.

www.ets-lindgren.com

Installation

15

6.0 Operation
CAUTION: Before placing into operation, follow the information provided in Before You Begin on page 13.
Note: Make sure you are satisfied with the contents and condition of your amplifier prior to placing it into operation.
Warning: Avoid exceeding the stated maximum input level as this can cause a risk of damaging the equipment.
Warning: There is a significant risk of an electrical or burn hazard; to avoid this, always ensure the output is terminated into a suitable load prior to switching on the equipment.
Equipment Use  The equipment must only be used for its intended purpose.  All parts of the equipment must remain connected to ensure that protective circuits operate. Disconnecting parts of the equipment may impair operation and safety.
Mains Switch  The amplifier is switched on or off by pressing the mains switch on the rear panel of the amplifier.
Standby/Operate Switch (Centre Unit) Pressing the standby/operate switch toggles the state of the amplifier between Standby and Operate modes. The blue halo indicator flashes in Standby mode, whereas steady blue indicates Operate mode.

16

Operation

www.ets-lindgren.com

Fault Indicator (Red LED)
Should a fault be detected, the red LED on the amplifier front panel will be illuminated. The amplifier will be forced into standby mode and the unit will cease to function. This action is latching, therefore only when the fault has been cleared can the amplifier be reset; this is done by pressing the standby/operate switch.
The reasons for the system fault light to illuminate are as follows:
 Excess module temperature
 Air cooling failure (fans failing to operate)
Interlock Indicator (Orange LED)
When the rear panel interlock is activated the interlock indicator will illuminate and the amplifier gain stages will all be turned off. This action is latching; therefore, the amplifier will return to Standby mode when the interlock is released.
Local Lockout Indicator
Refer to the separate Remote Control Interface manual supplied.
RF Input (Rear Panel Mounted)
The RF input will accept a signal from an RF generator. This input must be within the operating frequency range of the 8000-014 amplifier. An amplitude of up to 0dBm will be sufficient to saturate the amplifier. Operation outside the specified frequency range should not be attempted as it may subject the amplifier to undue internal stress.
RF Output (Rear Panel Mounted)
WARNING: The centre conductor of the RF output represents a severe burn hazard to personnel.
Never operate the amplifier without a proper termination or with defective cables or connectors.
This connector must be suitably terminated at all times during operation. Ensure that the cable is capable of handling the power available, which may be as much as 350 watts.

www.ets-lindgren.com

Operation

17

Air Inlet
This amplifier requires a free air supply for cooling; the front air inlet must not be restricted.
Safety Interlock
 A dual interlock system can be operated by applying either short circuit (S/C) or open circuit (O/C) to mute the amplifier.
 In order for the amplifier to operate, the Interlock BNC connector must be S/C and the Interlock must be O/C. If the Interlock is O/C or the Interlock is S/C the amplifier will be muted.
 The amplifier is delivered without the interlock applied and will be ready to operate.
 The interlock function is latching so when the interlock is removed the amplifier will be in standby mode. Press the standby/operate switch to return to normal operation.
 Current drawn from the interlock BNC connectors under short circuit conditions will typically be 2.8mA.
Fan Outlet
This amplifier requires a free air supply for cooling; the rear air outlets must not be restricted.

18

Operation

www.ets-lindgren.com

This page intentionally left blank.

www.ets-lindgren.com

Operation

19

Appendix A: EC (UKCA) Declaration of Conformity

ETS-Lindgren Inc. declares these products to be in conformity with the following standards, following the provisions of EMC Directive 2014/30/EU (Electromagnetic Compatibility Regulations 2016: Great Britain), Low Voltage Directive 2014/35/EU (Electrical Equipment (Safety) Regulations 2016: Great Britain) and the ROHS Directive 2011/65/EU (The Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment Regulations 2012: Great Britain):
Model 8000-014, RF Power Amplifier

Emission:

EN 61326-1:2013, Class A Electrical equipment for measurement, control, and laboratory use.

Immunity:

EN 61326-1:2013, Industrial level, performance criteria A Electrical equipment for measurement, control, and laboratory use.

Safety:

EN 61010-1:2010, AMD1:2016
Safety requirements for electrical equipment for measurement, control, and laboratory use.

Technical Construction Files are available upon request.

20

EC (UKCA) Declaration of Conformity

ets-lindgren.com



References

Acrobat Distiller 24.0 (Windows)