AeroPro® Cordless Prophy Handpiece
Quick Start Guide
Please read carefully. Refer to the Instructions for Use manual for complete product details. A French guide is available online at premierdentalco.com.
Get Started!
Initial Use: Follow steps 1-3.
- [Lock] Unlock
- [Steam] Autoclave & Disinfect
- [Battery] Charge
Routine Use: Follow steps 4-7.
Components:
- A Motor Component
- B Autoclavable Outer Sheath
- C Stand
- D Charging Station
- E AC Adapter
- F AeroPro® Disposable Barriers
- G Replacement O-rings and Tool for Maintenance
Step 1: Unlock
AeroPro® is shipped in a "Locked" position and must be “Unlocked" prior to initial use.
Remove green "Unit is Locked" label from Motor Component (A).
Place the Outer Sheath (B) onto the Motor Component (A).
To Unlock: Press and hold the ON/OFF/MODE button for 3 seconds.
Confirmation: AeroPro® will beep 3 times ??? and all indicator lights will flash simultaneously 3 times to indicate the system is unlocked.
Visit premierdentalco.com for an online library of helpful tutorial videos.
Step 2: Autoclave & Disinfect
DO NOT USE DISINFECTANT SOLUTION ON THE OUTER SHEATH
Place bagged Outer Sheath (B) & Stand (C) into a steam autoclave, per the autoclave manufacturer's instructions.
DO NOT autoclave the Motor Component (A).
Sterilization Cycles:
- Gravity Steam Sterilization (Outer Sheath/Stand): Full Cycle: 135°C (275°F) for 10 minutes.
- Pre-vacuum Steam Sterilization (Outer Sheath/Stand): Full Cycle: 132°C (270°F) for 4 minutes.
Drying Cycle: Use the drying cycle of the steam autoclave. Set cycle for 20 to 30 minutes. Do not exceed 137°C (278°F).
DO NOT EXCEED 137°C (278°F)
For non-autoclavable components (Motor Component (A), Charging Station (D), AC Adapter (E)): Use only water-based non-immersion disinfectant with a CDC intermediate level rating (e.g., chlorine disinfectant wipes with a tuberculocidal claim).
Step 3: Charge
Charging AeroPro® for the first time:
Plug AC Adapter (E) into surge protector (100-240V power only) ?.
Insert the connector into the back of the Charging Station (D). Seat the Motor Component (A) on the Charging Station.
Indicator lights will illuminate in succession to confirm charging is initiated. ?➡️?➡️?
Full charge is achieved within 2 hours. ⏳ 2h
AeroPro® features ChargeSMART™ Technology, engineered to optimize battery life. See Section 4.6 in Instructions for Use or learn more at premierdentalco.com.
Step 4: Routine Use - Pre-Operational Checks and Handpiece Assembly
Insert the charged Motor Component (A) into a custom-fitted AeroPro® Disposable Barrier (F). Ensure that the nose of the Motor Component (A) is fully aligned with the opening of the disposable barrier.*
✓ Correct
✗ Incorrect
Place autoclaved Outer Sheath (B) onto Motor Component (A) after covering with the Disposable Barrier (F).
Attach a 2pro® or equivalent disposable prophy angle (DPA) into the Outer Sheath (B) nose cone.
Shown - fully assembled CDC-compliant handpiece.
*Use only AeroPro® Disposable Barriers (F) for CDC-compliance.
Step 5: Operating the AeroPro® Handpiece
To turn unit on: Once assembled, press the ON/OFF/MODE button once quickly. The DPA tip will begin to rotate at low speed of 500 rpm.
[Power Button] ON: Press & release once quickly
Use the low speed for dipping the DPA into the prophy paste, applying downward force and light pressure. The low speed is also suitable for very light stains and pedodontic patients.
To change to high speed (2800 rpm): Insert AeroPro® into the patient's mouth and, if desired, press and release ON/OFF/MODE button once quickly to start polishing.
To turn handpiece off: Press and hold ON/OFF/MODE button for 1 second.
Note: The handpiece always returns to low speed upon powering on from an off state.
[Power Button] Off: Press & Hold 1 second
Sleep mode: After 20 continuous seconds of non-use, AeroPro® will automatically enter sleep mode and indicator lights will turn off. Lights will illuminate again once the ON/OFF/MODE button is pressed.
Automatic turn off: AeroPro® will automatically turn off after 10 minutes of continuous use. Press ON/OFF/MODE button to restart.
Step 6: After Treatment / Disinfecting and Sterilization
Pull DPA straight off to remove and discard. ?️
Disassemble the handpiece and follow Infection Prevention Procedures (See Section 4.8 in Instructions for Use).
Remove Outer Sheath (B) for sterilization by steam autoclaving (See Step 2). Autoclave AeroPro® Stand (C) if used.
DO NOT USE DISINFECTANT SOLUTION ON THE OUTER SHEATH
Remove the AeroPro® Disposable Barrier (F) from the Motor Component (A) and discard the barrier. ?️
Clean and disinfect the Motor Component (A), Charging Station (D), and AC Adapter (E) using a CDC intermediate level disinfectant (See Step 2, Section 4.8.2 in Instructions for Use for disinfection instructions).
When cleaning and disinfecting the Motor Component, avoid getting excess moisture onto or around the green band and inside the working end of the motor component.
Place disinfected Motor Component (A) back into the Charging Station (D) (See Step 3 Charge).
Step 7: Changing Speed Modes
Unit must be off to change speed modes.
Factory setting: 2-speed mode (low to high speed) (L → H).
If a greater range of speeds is preferred, the 3-speed mode can be selected:
3-speed Mode:
To change from 2-speed to 3-speed, press and hold ON/OFF/MODE button for 3 seconds. Three green indicator lights will flash and beep 3 times ???. AeroPro® is switched to 3-speed mode.
2-speed Mode:
To change from 3-speed to 2-speed, press and hold ON/OFF/MODE button for 3 seconds. Two top green indicator lights flash and beep 2 times ??. AeroPro® is switched to 2-speed mode.
Speeds and Speed Modes Table:
Operating Speeds | 2-Speed Mode (factory setting) | 3-Speed Mode |
---|---|---|
Low | 500 rpm | 500 rpm |
Medium | — | 1500 rpm |
High | 2800 rpm | 2800 rpm |
When operating in 3-speed mode, switch speeds by quickly pressing and releasing the ON/OFF/MODE button (L → M → H → L).
The selected speed mode will be saved until a new speed mode selection is made (See Section 4.5 in Instructions for Use).
Kits and Accessories
Basic Kit (#5500500)
- 1 Motor Component
- 1 Autoclavable Outer Sheath
- 1 Charging Station
- 1 AC Adapter
- 1 Stand
- 1 Box of 500 Disposable Barriers
- 3 Extra Set of O-rings and Tool for Maintenance (6 Small O-rings for the Outer Sheath and 3 Large O-rings for Motor Component)
Complete Package (#5500510)
- 1 Motor Component
- 3 Autoclavable Outer Sheaths
- 1 Charging Station
- 1 AC Adapter
- 1 Stand
- 1 Box of 500 Disposable Barriers
- 3 Extra Set of O-rings and Tool for Maintenance (6 Small O-rings for the Outer Sheath and 3 Large O-rings for Motor Component)
Accessories
- 5500520: AeroPro® Autoclavable Outer Sheath - Single
- 5500530: AeroPro® Disposable Barriers - Box of 500
- 5500550: AeroPro® Motor Component
- 5500540: AeroPro® Charging Station
- 5500541: AeroPro® AC Adapter
- 5500560: AeroPro® Stand
- 5500561: AeroPro® O-ring Replacements
*Order through your authorized dealer.
Warranty Registration
Please register your AeroPro® Cordless Prophy System and retain your dealer invoice. This will ensure prompt and accurate service should your AeroPro® Cordless Prophy System require evaluation, repair or replacement during the applicable warranty period, and will enable Premier to send you important product information.