Labmate LMMA-103 Melting Point Apparatus: Operational Manual

Index

Sr.noTitlePage no
1.Safety Measures2
2.Introduction3
3.Features3
4.Specifications4
5.Applications4
6.Instrument Introduction5
7.Operations7
8.Troubleshooting26

1. Safety Measures

The instrument was designed using the newest technologies while adhering to recognised safety standards. However, if taken in the following ways, it can still be dangerous:

Prioritizing Safety:

Any disregard for safety measures will result in normal work, property losses, instrument failure, and personal injury to the user.

Risk Prevention Guidelines:

  1. A technical malfunction could endanger life; therefore, be sure the power cord is plugged into a grounded outlet.
  2. Never handle the sample right out of the body, the cover, or the stove! It might get as high as 400°C.
  3. Put the device in ventilation if the dangerous gas breaks down after the sample.
  4. The paper should not be pushed against the instrument to avoid heat dissipation and instrument damage.
  5. Use it in spaces with adequate ventilation.
  6. Do not include the following effects on the environment:
    • When exposed to strong light
    • Excessive vibration
    • There is more than 80% humidity
    • High magnetic or electric field intensity
    • The outside temperature ranges from 5°C to 40°C
    • Volatile gases

2. Introduction

Melting Point Apparatus LMMA-103 utilizes a 400°C to room temperature automatic visual inspection detection method. The data can be sent using a network interface, USB port, RS-232 port, or SD port. With 20 map saves, around 600 bytes can be stored. It features a heating rate of 0.1 to 20°C per minute and supports 16 GB of storage capacity.

3. Features

4. Specifications

Model No.LMMA-103
Temperature RangeRoom Temperature - 400°C
Detection MethodAutomatic/ HD Visual Inspection
Accuracy± 0.3°C (< 250°C); ± 0.5°C (> 250°C)
Repeatability0.1°C/ min ± 0.1°C
Storage Capacity16 GB
Heating Rate0.1 – 20 °C/min
Display ModeTFT high-definition true color screen
InterfaceUSB/RS232/SD/ Network Port
Capillary SizeOD: 1.4 mm; ID: 1.0 mm
Data Save600
Number of User Management10
Map Save20
Power120 W, 220 V 50 Hz

5. Applications

Melting point apparatus are used in food, pharmaceutical, biochemical, medical, mining, and water management industries to figure out the melting point of oil and fat.

6. Instrument Introduction

Instrument Structure:

The figure below depicts the instrument's structural composition:

Figure-1 Instrument appearance: Shows the front view of the melting point apparatus with numbered components.
Figure-2 Heating furnace: A close-up view of the heating furnace component.

The capillary's dimensions are 120 mm in length, 0.9-1.1 mm in inner diameter, and 0.10-15 mm in wall thickness. The user can then put the capillary into two to three heating holes in the heating furnace. For a single test, there could be one to three samples.

IconDescription
[Icon for photo-taking]Start the photo-taking function
[Icon for camera]Start/stop the camera function
[Power switch icon]System power switch
Figure-3 Instrument rear view: Shows the rear panel of the instrument with numbered ports and interfaces.

7. Operations

7.1 Function Introduction

7.1.1 Opening Interface

Once activated, Figure 4 displays the UI. "Labmate LOGO" is in the centre, and "Melting point apparatus" is below.

Figure-4: Displays the initial Labmate logo screen for the melting point apparatus.

The login interface, depicted in Figure 5, will be brushed with the starting progress bar. Click the "Login" icon in the lower right corner to access the main interface after providing your username and password in the login interface. All users' initial passwords are 888888.

Figure-5: Shows the login screen with fields for username and password, and a login button.

7.1.2 Main interface

There are two components to the main interface:

Figure-6: Illustrates the main interface of the apparatus, showing function keys and parameter display areas.

7.1.2.1 Test interface

To access the test interface, which is displayed in Figure 7, click [Test] at the main interface. Containing the [sample name], [preset temperature], [heating rate], [stop temperature], [initial], [range], and [final] parameter settings.

Figure-7: Displays the test interface with input fields for various parameters and a numeric keypad.
ParameterDescription
[Sample name]It can contain up to 20 characters and be entered in either English or symbols.
[Start temperature]The range of the input value is 0°C to 100°C.
[Heat-up rate]The input value range is between 0.1°C and 20°C.
[Stop temperature]To await the user's next action, the instrument will automatically cool down to the start temperature once the input value range of 5 to 100 hits the stop temperature.
[τ end]The range of the optical signal intensity at the sample's ultimate melting, which also shows the melting degree, is 0–1200.
[Range]The sample final melting time signal value's permissible deviation range is 0–1200.
[β]Calculate the light signal value change after final melting, with a range of 1 to 10.
[τ initial]The time difference between the automated and actual start melting is between 1 and 120 seconds.

Function Key Descriptions:

An illustration of the significance of the parameter is as follows: The [final] value can be set to 1000, as Figure 8 displays the sample's melting curve.

Figure-8: Shows a sample melting curve with temperature plotted against time, indicating TA1, TA2, TA3, TA4, TC1, TC2, TC3, TC4.

The lowest value, 800, and the highest value, 1150, for several samples are displayed in Figure 9. Then, by setting the [end] value to 1000 and the [range] value to ± 200, the parameter setting may account for all possible ranges that might arise during the sample's final melting, improving the accuracy of the instrument's detection. Both [end] and [range] can generally be set to their default values, meaning that the entire range can decide the final melting outcome, and they typically do not need to be changed.

Figure-9: Displays multiple sample melting curves, illustrating variations in melting points and ranges.

Figure 10 illustrates the sample melting curve, which is smooth after the final melting and allows the [] value to be changed to smaller values like 1 or 2. Figure 11 shows a significant change in the melting curve trend, allowing the [] value to be set to bigger values like 4 or 5. For the real measurements, all values can be changed. To improve the accuracy of the melting point test results, choose the right value.

Figure-10: Shows a smooth melting curve.
Figure-11: Shows a melting curve with a significant change in trend.

Note:

Automatic detection interface: Click [automatic detection] to access the automated detection interface at the test interface after setting the parameters, such as the heating rate and beginning temperature. Figure 12 shows the interface.

Figure-12: Displays the automatic detection interface, showing parameters like TA, TC, RH, TP, and axes for temperature and signal strength.
ParameterDescription
TAThe average of the investigated samples' initial melting temperatures
TCAverage of the examined samples' final melting temperatures
RHRate heating
TPStarting temperature
TA1 to TA4From right to left, indicate the appropriate samples' initial melting values.
TC1 to TC4From right to left, indicate the appropriate sample's final melting value.
Horizontal axisThe value of the light passing's optical signal strength
Vertical axis °CThe current temperature of the sample
[Print]The test results will be printed by the device (if a printer is attached).
[Heating up]Start the test by heating up at the designated rate.
[Save]The instrument will save the query and test results (i.e., TA TC value) in the data interface; the user must select the appropriate option in [System settings] to save the map displayed during the test procedure.
[Return]To return to the parameter setting interface, exit the automatic interface.

Note:

7.1.2.2 Manual detection interface

Once the initial temperature and heating rate have been established, select [Manual Detection] in the test interface to access the manual detection interface (Figure 13).

Figure-13: Shows the manual detection interface with fields for sample name, temperatures, heating rate, and buttons for test control.
ParameterDescription
TAAverage initial melting temperature of the examined samples
TCFinal melting temperature average of the tested samples
[Heating up]Start the test by heating up at the designated rate.
[Save]The instrument will save the test results (i.e., TA TC value), and if you need to check the test results, you can query them in the data interface
[Print]The test results will be printed by the device (a printer is optional).
[Return]Close the test interface and then re-enter it.

After the actual temperature reaches the pre-set starting temperature (prompted by the buzzer), put in the sample to be tested, and press [Heating up] to detect the sample's melting point. At this point, it can be viewed in the video above the instrument. During the melting process of the measured sample, when the measured sample reaches the initial melting degree, click on the corresponding position in Figure 11 [Initial melting] key, that is, to record the initial melting value. When the measured sample reaches the final melting degree, click the [final melting] key in the corresponding position in Figure 11, that is, to record the final melting value.

7.2 Data interface

To access the data interface from the main interface, click [Data]. Figure 14 shows the interface. This interface mostly shows the user-saved data. The stored data group's serial number is 001~, which is the total number of saved data.

Note: The most recent data saved is at the top, so the more recent the data presented, the greater the serial number value.

Figure-14: Displays the data interface showing saved test results, with options to look back, print, and navigate pages.

To print or see the data, click on any place in the line of data that has a "" at the head of the journey. To print the user-selected data, click the [Print] key; to view the user-selected map and movie, click [Look Back]. Only when both the "map" and the "video" are shown after the peer data can the user view the pertinent information in the past.

Look back at the interface: At the data interface, click [Back] to enter the lookback interface as shown in Figure 15.

Figure-15: Shows the lookback interface with navigation buttons for playback control.

Operation declaration:

Figure-16: Displays a synchronized video and melting curve.

To switch the image display to the preview mode after finishing the callback action, press [Return], navigate to the callback interface shown in Figure 15, and then press [mode].

7.3 Set up the interface

To access the Settings interface from the main interface, click [Settings], as illustrated in Figure 15. The [system setting], [time setting], [experiment scheme], [printing setting], [user password setting], and [correction] settings are all part of the setting interface.

Figure-17: Shows the main settings menu with options for System settings, Time setting, Experimental plan, Print settings, Password settings, and Correction.

7.3.1 System setup interface

As illustrated in Figure 18, click [System Settings] to access the system settings screen.

Figure-18: Displays the system settings screen with various toggle options and a numeric keypad.
SettingDescription
[Preset prompt]A sound prompt is necessary when the instrument's real temperature exceeds the default beginning temperature.
[Heating prompt]The heating indicator will appear after the real temperature (by default) once you click the heating key.
[Heatup camera]Click the heating device during the test to simultaneously initiate the camera feature (the default is no).
[Save the map]The device will automatically save the map once the test is over (by default, it doesn't).
[Exploration mode]Unknown chemicals can be explored because, after the exploration option is chosen, the automatic detection duration can be much longer than seven minutes. [Note: 5°C per minute is the suggested heating rate.] (No is the default.)
[Overheat protection/alarm temperature]The input ranges from 350 to 400. To ensure human-machine safety, the user can preset a temperature in advance, and the device will automatically turn off the continuous heating feature once it reaches that degree (the default is 350°C).
[Return]Return to the setup interface after leaving the test interface.

7.3.2 Time setting interface

To access the time setting interface, click [Time Settings] at the settings interface, as seen in Figure 19. Press year-month-day-time-minute-second after clicking anywhere in the box. Then, click on the next digital interface to enter data, and hit Enter.

[Return]: Close the time setting window and go back to the settings window.

Figure-19: Shows the time setting interface with input fields for year, month, day, hour, minute, and second, along with a numeric keypad.

7.3.3 Experimental protocol interface

To access the experimental scheme interface, click [Experimental Scheme] at the configuration interface, as seen in Figure 20.

Figure-20: Displays the experimental scheme interface with columns for sample name, parameters like TA, TC, TP, and buttons for call, page navigation, and return.

7.3.4 Print the setup interface

As illustrated in Figure 21, click [Print Settings] to open the Print Settings screen.

Figure-21: Shows the print settings interface with fields for Title content and Operator, and buttons for confirm and return.
SettingDescription
[Title content]The printed report's title (by default, Melting Point Apparatus Report)
[Operator]Printout Report Operator Name
[Return]Return to the settings interface after leaving the print settings interface.

7.3.5 User password setting interface

To access the password setting interface, click [Password Settings] at the settings interface, as seen in Figure 22. Click the box next to the new password, type it in using the digital keyboard below, press Enter, then [confirm], and restart the computer so that the new password can be used to log in.

Figure-22: Displays the password setting interface with fields for original password and new password, and a numeric keypad.

[Return]: Return to the setup interface after leaving the test interface. Note: Just alter the user's current password.

7.3.6 Correct the interface

To access the correction parameter input interface, click [Correction] at the configuration interface, as seen in Figure 23.

Figure-23: Shows the correction interface with fields for sample melting point, preset temperature, heating rate, stop temperature, and buttons for default scheme, factory reset, automatic correction, and return.
ParameterDescription
[Sample melting point]Enter the standard sample's final melting value.
[Preset temperature]The input values range from 5-100
[Heat-up rate]The input values range from 0.1-20
[Stop temperature]The instrument will automatically cool to the start temperature once the input value range of 0-400 reaches the stop temperature. Set the temperature while you wait for the user to do something else. Consider there to be no stop temperature when the stop temperature input is 0.
[Default scheme]Return the four optical signal setting parameters [final], [range], and [initial] to the factory defaults.
[Restore factory]Restore factory raw data
[Automatic detection]Open the relevant rectification interface. It is displayed in Figure 24.
[Manual detection]Open the relevant rectification interface. The user interface is displayed in Figure 25.
[Return]Exit the correction interface and go back to the settings interface
[Correction]The corrected test result can be saved by pressing the correction key if the test result is not abnormal.
[Heating up]The detection starts when the temperature rises at a rate of heating.
[Return]Return to the correction interface after leaving the correction test interface.
Figure-24: Displays a graph related to correction, showing temperature and signal values.
Figure-25: Shows the correction interface with input fields and control buttons.

Note: To guarantee the correctness of the instrument test, it is advised to apply automatic detection correction for the functional oil melting point.

7.4 Camera interface

To access the camera interface, click [Camera] at the main interface, as illustrated in Figure 26.

Figure-26: Shows the camera interface with navigation buttons (UP, DN, OK, Menu) and mode selection.

Note:

7.4.1 Operation instructions

7.4.1.1 Preview mode
  1. To switch between preview and playback mode, press [Mode].
  2. To capture pictures, press the [Photo] key.
  3. To begin recording, press [Video]; to save and end the recording, press [Video].

    Note: The system will automatically save and stop the video recording when the continuous video file size reaches 4 GB, when the continuous video recording time reaches 60 minutes, or when the SD card is used for video recording.

  4. After that, the complete screen will display the Memory complete Memory Full.
  5. To configure the preview image to freeze or unfreeze and capture pictures while frozen, press OK.
  6. Enter the digital zoom state by pressing [DN], zoom in on the image by pressing [Video], then shrink the image by pressing [Photo]. Then, without performing the operation, press [DN] or hold for two seconds.
  7. To turn the OSD on or off, press the [UP] key. Note: OSD is the term for creating a unique glyph or graphics on the screen that gives the user access to information.
7.4.1.2 Menu

To access the View / Audio Settings menu, press [Menu], as illustrated in Figure 27. The red writing on the screen indicates the current setting, the [UP] / [DN] key chooses the setting item, and the [Photo] / [Video] key changes the option value. To save and close the View / Audio Settings menu, press [Menu] once more.

Figure-27: Shows the Video Setup menu with options for Photo Size, Photo Quality, Video Size, Video Quality, and Brightness.

(1) The video settings menu

The Manual color sub-menu, as shown in Figure 28.

Color SettingDescription
Red: Level 0-255The state is set to 128 by default
Green: Level 0-255The state is set to 128 by default
Blue: Level 0-255The state is set to 128 by default
Figure-28: Displays the Manual Color settings for Red, Green, and Blue, with their default values.

Auto Calibrate White balance automatic calibration sub-menu: When the screen says, "Put a blank sheet of paper in front of the camera, press [Photo] button to confirm and start automatic calibration, press [Video] button to exit," press [Photo] button to confirm and exit, press [Photo] button to recalibrate, press [Video] button to cancel and exit," and the RGB scale will automatically calibrate. Press OK to enter the automatic calibration submenu.

Time Lapse sub-menu, Figure 29.

Time Lapse Settings:

Figure-29: Shows the Time Lapse settings for Hour, Minute, and Second.

(2) Audio setup menu (which is not supported by this instrument)

7.4.1.3 Playback mode

  1. To switch between preview and playback mode, press [Mode].
  2. As seen in Figure 30, the screen will show up to six thumbnails at once. To get to the last or next file, press [Photo] / [Video]. To go to the first or last three files, press [UP] / [DOWN]. To play the current file in full screen, press OK.
  3. Press [UP] to turn on or off the OSD when the photo file is selected, and then [OK] or [mode] to enlarge or reduce the image. When the image scales, [UP] key up, [DN] down, LT key left, and [Video] right.
  4. You can also make the picture smaller by using the keys [UP] up, [DN] down, LT left, and [Video] right when the image scales.
  5. After selecting the video file, you can press [UP] to turn off OSD, [OK] to start or stop the video, [Photo] or [Video] to move forward or backward, [OK] to stop the video, and [DN] to pause or continue.
Figure-30: Displays a grid of captured images (thumbnails) with file names and dates.

Note: All the files will be stored in a subfolder named after the xxxIMAGE.example. For example: 100IMAGE, 101IMAGE, 102IMAGE, these subfolders are stored in the DCIM folder, each can save 1000 files, such as: PICT0001.JPG, PICT0002.AVI.

7.4.1.4 Key Operation

KeyOperation
Preview modeReplay back (thumbnail) modeReplay (individual speech piece) mode
[pattern]Go to playback modeGo to preview modeEnter the Play (thumbnail) mode
[menu]Enter the system settingsThe first three filesGo to the delete menu
UPOSD open/closeOSD open/closeOSD open/close
DNEnter the digital zoom stateThe last three filesThe previous file
LFTake a pictureThe previous fileNext file
RTStart/stop the video recordingNext fileVideo file: Start playing the video recording Photo file: enlarge the photos
OKFreeze / thaw the imagesSelect and confirm the fileSelect and confirm the file
KeyBack to the VideoMenu SettingsTime Date Settings
[pattern]Go to playback modeGo to playback modeGo to playback mode
[menu]Save the current settings and exit the menu modeExit menu modeExit menu mode
UPOSD open / closeSelect Setup ProjectIncrease the number
DNPause the play/ continue playingSelect Setup ProjectReduce the number
LFBack playSelect Set ParametersMove to the previous settings option
RTForward Play (2)Select Set ParametersMove to the next setting option
OKStart (1) / Stop playingaffirmaffirm

7.5 Help interface

To access the help interface from the main interface, click [Help]. Figure 31 shows the interface.

Figure-31: Shows the help interface with a question mark icon and a 'return' button.

[Note]: To access the precautions interface, press the precautions key. [Return]: To return to the main interface, exit the test interface.

7.6 Operations steps and use methods

7.6.1 Capillary test sample preparation

1) Large capillary preparation

First, fill the capillary's bottom with 0.2 mL of water. After cleaning the outside of the glass tube using paper towels, a large capillary is placed within the furnace body.

2) Fine capillary mounting sample preparation

Take out the double-opening capillary, submerge it in the completely melted sample until the oil is about 10 mm high, then store and harden it in accordance with national regulations.

7.6.2 Oil melting point test

1) Automatic test paradigm palm oil: sample melting point 44.8-53.4°C
  1. Turn on the power switch and preheat the instrument for 20 minutes.
  2. Enter the main interface, choose the test, enter the sample name (palm oil), as well as the end, range, and first light optical signal setting parameters, starting temperature (45°C), heating rate (1°C/min), stop temperature (55°C), and default scheme. Then, enter the appropriate interface based on [automatic detection].
  3. Once the actual temperature reaches 45°C (as indicated by the buzzer), set the sample's capillary to that temperature. The temperature in the boiler drops a little after the sample is placed, therefore, press the temperature button. When it achieves the final melting degree, the device will automatically show the initial melting value that was reached when the sample reached the initial melting degree.
  4. After all the findings are ready, you can either save the data or print it by pressing [print]. You only must do the third step if the sample is tested again. To modify the tested sample, press [Return] and go back to the test interface to adjust the relevant parameter.
2) Manual test paradigm palm oil: sample melting point 44.8-53.4°C
  1. Turn on the power switch and preheat the instrument for 20 minutes.
  2. Press [manual detection] to enter the appropriate interface after entering the main interface, choosing the test, entering the sample name (secondary acid), starting at 45°C, heating at a rate of 1°C per minute, stopping at 55°C, and choosing the test.
  3. Insert the test sample, press [heating up], and see the melting process of the test sample from the video above the instrument once the real temperature reaches 45°C (with buzzer prompts). Click the [primary melting] key in the appropriate location shown in Figure 13 once the test sample begins to glide up. Click the [final melting] key in the appropriate location shown in Figure 13 once the test sample has completely slipped out of view.
  4. You can either save the data or click [print] to print the data once all of the findings are ready. Only the third step needs to be repeated if the sample is tested again. Press [Return] to go back to the test interface for the relevant parameter setting if you wish to modify the tested sample.

7.6.3 Oil melting point correction

Correction paradigm: The sample final melting point of 50.2°C standard oil correction:

  1. Press [Automatic detection] or [Manual detection] to access the appropriate interface after entering the calibration interface in the settings, setting the test parameters for the standard sample naphthalene, entering the sample melting point (50.2°C), start temperature (45.2°C), heating rate (1°C/min), and stop temperature (55.2°C).

The final, range, and beginning light signal setting parameters can be changed based on the samples if the correction technique is carried out by automatic detection (the correction samples are three standard samples, and the parameters can be chosen from the preset scheme).

  1. Using the test sample's third step.
  2. Click [Correction] to save the updated test results if there are no abnormalities. Then, click the third step to test again. The [Return] key can be used to close the correcting interface.

8. Troubleshooting

8.1 Common faults and treatment methods

Fault phenomenonCauseSolution
There is fog on the small screen, making it blurry.Due of its high moisture content, the furnace body heats up rapidly and releases water vapour.After some time of heating, the water vapour evaporates and immediately returns to its ideal state.
After turning on, the display screen does not reflectThe power supply for the instrument is not connected.Turn on the power switch or insert the power plug.
Every operation function is unsuccessful.The heating system's software is broken.Examine and fix the heating system.
Atlas prevention failedThe save the profile option is notThe map save option is Yes
Press the [Heating Up] key without the displayPress [Save] to stay for less than two secondsPress [Save] and stay for at least two seconds
Lower the temperature setting below the initial value.Reset the stop temperature
The heating prompt option is NoThe heating prompt option is Yes
Automatic detection of anomaliesSundries have been present.Clean up the furnace core debris
The [automatic detection] button is already depressed on the front furnace capillary.Remove the capillary and restart
The parameters for automatic detection are not properly configured.Adjust the automatic detection parameters based on the sample's actual circumstances or the default scheme.
The melting point result is not permitted.No calibration for a long timeCalibrate the standard samples
There is no preview mode on the image display.The position of the platinum resistance is loose.Reconnect and tighten it
To avoid accessing the main interface, press the Login key.There has sundriesPress [mode] to change the image display mode to the preview mode after entering [Camera].
Poor reproducibilityNo mode switch was made after reviewing the modeRe-enter the correct username or password
Wrong username or password inputStrict guidelines for installing the sample
InconsistentStrictly pick up the wool tube
The capillary size is inconsistentThe instrument's regular operation is confirmed by high-purity samples.
During the test procedure, it was not possible to monitor the melting process of the sample.Low sample purityFor observation, rotate or raise the capillary to a height of less than 2 mm.
The bottom of the furnace body cannot be directly illuminated by the light source.For observation, rotate or raise the capillary to a height of less than 2 mm.
Samples were darker during meltingFor observation, rotate or raise the capillary to a height of less than 2 mm.

8.2 Treatment method of capillary fracture in the furnace body

The following figure illustrates what happens if a capillary fracture happens in the furnace during the test:

Step 1: Hold the cover of the furnace.

Figure-32: Shows a hand holding the cover of the furnace.

Step 2: Rotate 90 degrees anticlockwise.

Figure-33: Shows a hand rotating the furnace cover anticlockwise.

Step 3: Lift the cover.

Figure-34: Shows a hand lifting the furnace cover.

Step 4: On the furnace core's support, release the snail.

Figure-35: Shows a hand releasing a component on the furnace core's support.

Step 5: Remove the support for the furnace core.

Figure-36: Shows a hand removing the support for the furnace core.

Step 6: From the damaged capillary, extract the oven core.

Figure-37: Shows a hand extracting an oven core from a damaged capillary.

Labmate Scientific Inc

Email: info@labmate.com | Website: www.labmate.com

Models: LMMA-103 Melting Point Apparatus, LMMA-103, Melting Point Apparatus, Point Apparatus, Apparatus

File Info : application/pdf, 29 Pages, 1.29MB

PDF preview unavailable. Download the PDF instead.

v1

References

iLovePDF

Related Documents

Preview Labmate LMMA-103 Melting Point Apparatus Operational Manual
This operational manual provides detailed instructions for the Labmate LMMA-103 Melting Point Apparatus, covering safety measures, instrument introduction, features, specifications, operations, and troubleshooting.
Preview Labmate LMMA-104 Melting Point Apparatus Operational Manual
This operational manual provides comprehensive instructions for the Labmate LMMA-104 Melting Point Apparatus, covering safety measures, instrument introduction, features, specifications, applications, operations, and troubleshooting.
Preview Labmate LMMA-102 Melting Point Apparatus Operational Manual
This operational manual provides detailed instructions for the Labmate LMMA-102 Melting Point Apparatus, covering safety measures, features, specifications, operations, and troubleshooting.
Preview Labmate Drying Oven LMOD-A101 Operational Manual
Operational manual for the Labmate Drying Oven LMOD-A101, detailing safety measures, features, specifications, applications, operations, troubleshooting, and circuit diagram.
Preview Labmate LMPCR-A102 PCR Cabinet User Manual
Comprehensive user manual for the Labmate LMPCR-A102 PCR Cabinet, detailing its features, specifications, installation, operation, maintenance, and troubleshooting.
Preview Labmate Drying Oven LMOD-A109 Operational Manual
Operational manual for the Labmate Drying Oven LMOD-A109, detailing safety measures, introduction, features, specifications, applications, structure, operations, troubleshooting, and circuit diagram.
Preview Labmate LMPHM-A100 Portable Water Hardness Meter User Manual
Comprehensive user manual for the Labmate LMPHM-A100 Portable Water Hardness Meter, covering specifications, applications, installation, operation, maintenance, and troubleshooting.
Preview Labmate LMSAA-101 Automatic Surface Area Analyzer User Manual
This user manual provides detailed instructions for operating, installing, maintaining, and troubleshooting the Labmate LMSAA-101 Automatic Surface Area Analyzer. It covers instrument features, software operations, sample analysis, data processing, and reporting.