Protekt Aire 8900 - USER MANUAL

Low Air Loss Alternating Pressure Mattress with Side Air Bolsters and Dual Compressor - Protekt Aire 8900


File Info : application/pdf, 24 Pages, 1.75MB

PDF preview unavailable. Download the PDF instead.

protekt-8900
Protekt® Aire 8900 Low Air Loss/Alternating Pressure/Pulsation Mattress System w/on Demand Side Air Bolsters
Ver. C

IMPORTANT SAFEGUARDS READ ALL INSTRUCTIONS BEFORE OPERATING THIS DEVICE
NOTE, CAUTION AND WARNING STATEMENTS: NOTE ­ Indicate some tips. CAUTION ­ Indicate correct operating or maintenance procedures in order to
prevent damage to or destruction of the equipment or other property. WARNING ­ Call attention to a potential danger that requires correct procedures or
practices in order to prevent personal injury.
WARNING­ To reduce the risk of electrocution: 1. Patients are not allowed to operate the product. Always unplug this product immediately while it's not in use. 2. Do not disassemble the Pump to avoid electrocution. 3. Do not place or store product where it can fall or be pulled into a tub or sink. 4. Do not place in or drop into water or other liquid. Do not use while bathing. 5. Do not reach for a product that has fallen into water. Unplug immediately.
WARNING­To reduce the risk of burns, electrocution, fire or injury to person: 1. The system must be operated with the mattress connected to the pump.
Please do not power-off or unplug the Pump while in use. 2. Always use the same voltage as outlet label. Do not use the other power cords
on the pump. Keep the plastic packing away from children and pets to prevent suffocation hazards. 3. Equipment is not suitable to use in the presence of a flammable anesthetic mixture with air, with oxygen or nitrous oxide. 4. Keep away from sharp objects. 5. Close supervision is necessary when this product is used by, on, or near children or people with disabilities. 6. Use this product only for its intended use as described in this manual. Do not use attachments not recommended by the manufacturer. 7. Never operate this product if it has a damaged cord or plug, if it is not working properly, if it has been dropped or damaged, or dropped into water. Return the product to a service center or to the distributor for examination and repair. 8. Keep the power cord away from heated surfaces. 9. Never block the air openings of this product or place it on a soft surface, such as a bed or couch, where the openings may be blocked. Keep the air opening free of lint, hair, and other similar particles. 10.Never drop or insert an object into any opening or hose.
2

11.Connect this product to a properly grounded outlet only. See Grounding Instruction.
12.Avoid dropping or putting any heavy object on the pump. 13.Put power cord or hose tube at patient's foot area to avoid winding around
patient's neck. 14.To avoid electromagnetic interference, the patient environment should not
have strong electro-magnetic or RF generated equipment nearby. 15.The PUMP will have minor heat generated in operation, please do not directly
contact the surface continuously for more than 1 minute. 16.Avoid the risk of electrocution by grounding the ground pin with grounding
power outlet. 17.When the main supply is lost or has temporarily failed, the pump will stop and
the power failure alarm will alarm up to 20 minutes. This is normal and the product will return to normal operation once the main supply is resumed. 18.The connection of lift-up detection to the PUMP shall be performed by the operator. It's not designed for patient to do the connection.
3

SYMBOLS I O
IP21

Product Symbol Description DESCRIPTION
POWER ON POWER OFF ATTENTION
DOUBLE ISOLATION
"BF" SYMBOL, INDICATE THIS PRODUCT IS ACCORDING TO THE DEGREE OF PROTECTION AGAINST ELECTRIC SHOCK FOR TYPE BF EQUIPMENT, APPLIED PART:MATTRESS CAUTION, READ THE INSTRUTION BEFORE USE
AWAY FROM THE FLAME
WATER AND DUST PROTECTION CLASSIFICAITON FUSE SPECIFICATION
DISPOSAL OF ELECTRICAL & ELECTRONIC EQUIPMENT (WEEE): THIS PRODUCT SHOULD BE HANDED OVER TO AN APPLICABLE COLLECTION POINT FOR THE RECYCLING OF ELECTRICAL AND ELECTRONIC EQUIPMENT. UL CERTIFICATION LOGO (COMPLIACE WITH IEC60601-1) WITH RESPECT TO ELECTRICAL SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH IEC60601-1. CB CERTIFICATION LOGO
CE CERTIFICATION LOGO

4

Contents

1.

INTRODUCTION ....................................................................................................................... 6

2.

INTENDED USE ......................................................................................................................... 6

3.

PRODUCT DESCRIPTION .......................................................................................................... 6

4.

PRODUCT INSTALLATION GUIDE .............................................................................................. 7

5.

PANEL DISPLAY AND OPERATION GUIDE ............................................................................... 10

6.

CLEANING .............................................................................................................................. 15

7.

STORAGE ................................................................................................................................ 17

8.

MAINTENANCE ...................................................................................................................... 17

9.

THE DISPOSAL OF AIR MATTRESS .......................................................................................... 17

10. TROUBLESHOOTING .............................................................................................................. 18

11. TECHNICAL DATA.................................................................................................................... 19

5

1. INTRODUCTION This manual provides the information required for the initial set up and for the normal operation of the Proactive Medical Product's Protekt® Aire 8900 Mattress System. Before operating the Protekt® Aire 8900 Mattress System, be sure the operator had read and understood in detail the content of this manual.
2. INTENDED USE The Protekt® Aire 8900 Mattress System is intended to help and reduce the incidence of pressure ulcers while optimizing patient's comfort. The Protekt® Aire 8900 Mattress System may be used in a variety of settings including, but not limited to individual home care setting and long-term care of whom suffering from pressure ulcer, or pain management as prescribed by physician.
The connection of Lift-Up Detection to the Pump shall be performed by the operator. It's not designed for the patient to perform the connection.
NOTE: Equipment is not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
3. PRODUCT DESCRIPTION The Protekt® Aire 8900 Mattress System is an alternating mattress replacement system used in the prevention and relief for patients with, or vulnerable to, pressure ulcers. The Protekt® Aire 8900 Mattress System offers patients a comfortable and relaxing support surface by using the established principles of alternating therapy, which can both prevent skin breakdown and enhance healing.
The Control Unit of the Protekt® Aire 8900 Mattress System is a Pump featuring a digital pressure adjustment function, mode selections, and audiovisual alarms. The 18 air cells mattress unit provides a unique design which keeps the lower sections of air cells constantly inflated while alternating and deflating the upper sections. The 3 static head cells remain static and provide a "pillow" support for optimum comfort. The mattress has a heavy-duty polyester base sheet with a vapor permeable nylon PU coated cover.
The system includes a rapid release twist CPR valve by the head section of the mattress for the event of cardiac arrest.
6

4. PRODUCT INSTALLATION GUIDE 1. Unpack the box to inspect all items for any damage that may have occurred during shipping. If there is any damage, please contact your dealer immediately for assistance. 2. Place the mattress on top of the bed frame. The feet symbol on both side of the mattress indicates location of the foot end.

Patient Head

Air Mattress
Model No. (Inside the cover)

CPR
Side Bolster (Optional)
3. Secure the mattress onto the bed frame by using the securing side straps.

Pump Unit Hose Connector

4. Ensure the CPR valve is at CLOSE position before turning on the power.

Securing Side Strap CPR Position Indicator

5. Use the yellow Bolster Inflator valve to inflate (CLOSE position) or deflate (OPEN position) the side bolsters.

7

6. Position the Pump by its elastic

hanger brackets over footboard of

the bed. The elastic hanger brackets

will self-adjust onto the footboard

6

tightly.

7. Remove the Transport Cap of the

8

hose connector and connect the

hose connector to the Pump unit.

Firmly push the hose connector into 7 position and a "click" sound will

secure the connection.

8. Connect the Lift-Up Detection cable

to the Pump unit by first align and

push-in the 3 pins connector then

gently turn the screw clockwise to

secure the connection.

80089 Pump

Follow the direction for connection.

9. Connect the power cord to the

Pump. The power switch should

remain off.

10. Press the red power cord protector

downward to secure the power

9

cord.

11. Plug the power cord into the

electrical outlet.

8

NOTE: Check and ensure the Pump

unit is suitable for the local power

voltage.

1

CAUTION: The Pump can only be

applied to the mattress

recommended by the

manufacturer. Do not use the

8

Pump for any other purpose. WARNING: Do not place the Pump unit in area where power cord can come off easily.

12. For patient's transportation, press "Auto Firm" button and wait for 5 minutes for the mattress to be inflated. Disconnect the hose from the Pump unit and put on the hose connector Transport Cap to keep the mattress inflated.

Bi-directional Transport Cap

9

5. PANEL DISPLAY AND OPERATION GUIDE
1 2

7

6 543

5.1 PANEL DISPLAY  Alarm Mute and Alarm Indicator  Low Pressure Alarm Indicator  Power Failure Alarm Indicator  Service (Malfunction) Alarm Indicator  Alternate/Pulsate Cycle Time or Warning code Display  Operating or Standby  Auto-Firm  Function Mode Selection (Alternate/Pulsate/Static)  Panel Lock-out  Comfort Control

5.1.1 Alarm Mute
Press the alarm mute button to suspend temporarily alarms. Should the situation not resolved and fault conditions continue, the alarm will resume notifying the patient and caregiver.

5.1.2 Alternate Time Display Alternate cycle time can be selected from 10-30 minutes by pressing the Cycle button.

5.1.3 Operate or Standby
10

Press this button to start operating or go into standby.
5.1.4 Auto-Firm The Pump will go into the inflation mode (LED lights flashing) every time the operate mode is triggered. This insures the mattress to be able to reach its maximum operating pressure. Once the max pressure level is reached, the Pump will automatically switch into the previous selected mode and comfort level. User can also use this function as full mattress inflation during patient sit-up or ingress/egress for better support.
5.1.5 Function Mode Switch  Alternate ­ the air cells of the mattress will be proportionally deflated to reduce the surface pressure. The alternating cycle will continue at the selected cycle time until another mode is selected.  Static ­ the mattress maintains at the selected pressure. The Pump will automatically fallback to alternate mode after 20 minutes.  Pulsate ­the mattress maintains at the selected pressure in static mode and changes from 100% to 60% of the selected pressure according to the selected Pulsate cycle time. The Pulsate cycle will continue until another mode is selected.
5.1.6 Panel Lock-Out Press the Lock-Out button to lock the panel. Should the panel remain untouched for 30 seconds, the Lock-Out feature will lock the panel to prevent accident from changing setting without notice. To unlock, press the Lock-Out button for 3 seconds.
5.1.7 Comfort Level Comfort level controls the air pressure output level. Press Firm button and the output pressure will increase and higher pressure output will support heavier weight patient, for decreasing air pressure, vice versa. Check to see if the suitable pressure is selected by sliding one hand between a deflated air cell and the patient's buttocks areas and there should be minimum contact. Always leave at
11

least 1 inch space between a deflated air cell and patient's buttocks areas to prevent "bottoming-out". Refer to Table 1 Weight and Comfort Level Reference for weight and comfort level suggestion.

5.2 OPERATION GUIDE 5.2.1 General Operation:

NOTE:

The power switch is located on the side of the Pump.

 Press to turn on the unit, all LED indicators on the control panel will light

up accompanied with a beep for 2 seconds (check for indicator failure if any),

and the indicator of Standby on the control panel will light up. If the Pump was

previously shut off in operate mode then the Pump will enter operate mode

directly.

P.S. To test the battery, press

to turn off the power and the power

failure alarm should be triggered. Refer to 5.2.3 Audiovisual Alarm if the

alarm is not triggered.

 Press the Operate button

and the system will begin to inflate and the

"Auto-Firm" indicator will be flashing.  The mattress should be fully inflated within 60 minutes, and automatically
enter the previous operating mode, otherwise the low pressure alarm with warning code " " will be triggered.

Note: Do not proceed to other settings before inflation is completed.

 After initial inflation is completed. Press Auto-Firm button

for moving

the patient onto the mattress. The mattress will turn into a steady condition in

around 5 minutes. Move the patient onto the mattress and press Auto-Firm

button again to cancel Auto-Firm mode and select the appropriate mode.  According to the weight of the patient, adjust the pressure setting to the most

suitable level without "bottoming-out". User can determine an appropriate

pressure by adjusting the Comfort Level. Please consult with your physician for

a proper setting.

Warning: the Pump unit should always be operating to prevent bedsore

occurring.  In operate mode, press operate/standby button

for the system to enter

standby mode. The system should be in standby mode before shut down.

12

Switch the power switch

to off and the warning code" "will appear

on the display to shut off the system.

Note: For reminding purpose, power failure alarm will be triggered if the power is switched off in operating mode (refer to 5.2.3 Audiovisual Alarm). Press power switch to restart the system, or press Alarm Mute to turn off the system (refer 5.2.4 Alarm Mute).

Table 1 Weight and Comfort Level Reference Protekt® Aire 8900

5.2.2 CPR  When CPR needs to be performed, quickly rotate the CPR valve to "OPEN" position, at the same time, disconnect the hose connector from the Pump to speed up the air release.

5.2.3 Audiovisual Alarm  Power Failure ­ When electrical shortage occurred or power cord is unplug

without turning off the Pump or

is pressed (intentionally or

unintentionally), the "Power Failure" indicator will light up along with buzzer

and will last 20 minutes.

NOTE: When the Pump has not been used for more than 3 months, the Pump may need 6 hours or more of operating time for the alarms to function properly.

 Low Pressure ­When an abnormal low pressure occurred in body section, the "Low Pressure" indicator will flash and beep. Should the situation not resolved and fault conditions continue, the alarm will resume.

 Service (Malfunction) ­When fault conditions occur, the "Service" indicator will light up along with buzzer.

13

Note: Refer to Table 2 for Warning Code Reference Table if error code appears on the display or refer to 10. TROUBLESHOOTING. 5.2.4 Alarm Mute  When alarms are triggered, both LED light and buzzer will turn on to warn the
patient and caregiver. By pressing the button, it will temporarily mute the buzzer so the caregiver may check for possible causes. Should the situation not resolved and fault conditions continue, the alarm will resume. Refer to 10. TROUBLESHOOTING for diagnosis.  During "power failure", pressing "alarm mute" will cease all buzzers and indicators and turn off the system.  During "low pressure alarm" if the pressure resumes back to normal then the low pressure alarm will stop.  When more than one alarm is triggered, the alarm will be performed according to priority level. Refer to Table 2 Warning Code Reference Table for priority level.
14

Table 2 Warning Code Reference

PRIORITY

HIHG WARNING INDICATOR

 CODE

LED

LOW

0

N/A

N/A

AUDIBLE OUTPUT MODE
ONCE

CONDITION OF OUTPUT

WARNING DESCRIPTION

REMARKS

Not in System Shutdown

Key Tone

Key Tone from Functional Button

1

Power Failure ONCE POWER-OFF System Shutdown

Shutdown

2

ALL LED

ONCE

OPERATE OR STANDBY

Power-On

All Indicators Light On

3

N/A

N/A

ONCE

OPERATE OR STANDBY

State/Mode Switching

No Display

4

Auto Firm ONCE

OPERATE

Mattress Inflation Completion

Inflation Ended

5

Auto Firm ONCE OPERATE Auto-Firm Completion Auto-Firm Ended

6

Static

ONCE OPERATE Static Completion

Static Ended

7

N/A

Power

Failure

REPEAT (Cycle 4 sec.)

POWER-OFF

Power Failure Alarm

No Display

8

Low Pressure

REPEAT (Cycle 4 sec.)

OPERATE OR STANDBY

Power-On Alarm

Inflation

Failure

Inflate

Failure

9

Low Pressure

REPEAT (Cycle 4 sec.)

OPERATE OR STANDBY

Auto-Firm Failure Alarm

Auto-Firm Failure

10

Low Pressure

REPEAT (Cycle 4 sec.)

OPERATE OR STANDBY

Low Pressure Overtime Alarm

Low Pressure

11

Service

REPEAT (Cycle 4.5
sec.)

OPERATE OR High Pressure Overtime STANDBY Alarm

High Pressure

12

Service

REPEAT (Cycle 4.5
sec.)

OPERATE OR High Ambient STANDBY Temperature Alarm

High Temperature

13

Service

REPEAT (Cycle 4.5
sec.)

OPERATE OR Air Valve 1 Positioning STANDBY Failure Alarm

Air Valve 1 failure

14

Service

REPEAT (Cycle 15
sec.)

OPERATE OR STANDBY

Battery Low Alarm

Battery would need to be replaced

15

NONE

NONE

FACTORY CALIBRATION
MODE

Calibration Not Completed

Calibration Unfinished

FACTORY

16

NONE

NONE CALIBRATION Calibration Completed Calibration Completed

MODE

5.2.5 Lift-Up Detection  Fowler Mode will be triggered and increase the pressure to Auto-Firm mode when the upper half of the mattress is elevated and exceeded an angle of 45  (±15) or more. The Pump unit will resume to previous setting when it's lowered than an angle of 45 (±15).

6. CLEANING Wipe the Pump unit with a damp cloth pre-soaked with a mild detergent, and keep the Pump unit away from dust. If other detergent is used, choose one that will have no chemical effects on the surface of the plastics case of the Pump unit.

15

CAUTION: Do not immerse or soak the Pump unit.
Clean the mattress cover by using single use wipes with a solution of neutral detergent and hand hot water. Rinse thoroughly with clean water and damp dry the mattress using single use wipes.
Disinfecting the cover If the cover is heavily soiled or has been exposed to bodily fluids such as blood, it will require a more thorough cleaning procedure.
Use single use wipes with a 0.1% chlorine solution (1,000ppm) and cold water to wipe the cover. Rinse thoroughly with clean water and damp dry the mattress using single use wipes. Ensure the cover is completely dried before refitting to the mattress.
Frequent or prolonged exposure to higher concentration disinfectant solutions may prematurely age the fabric cover of mattresses. Cover surfaces should be protected during use and rinsed and dried thoroughly after disinfectant.
Laundering  Before laundering, mattress cover should be completely removed.  Remove Lift-Up Detection Sensor located on the bottom base before laundering.  Mattress covers can be laundered as following:
 Prewash 60 +15 minutes
 Main wash 60+15 minutes  This should be followed by a cold rinse and extraction.  Lift-Up Detection Sensor is not washable. It can be wiped as described, and
install back after the mattress is completely dry.
Drying Mattress covers should be hung from a line or bar and drip dried in a clean indoor environment. Covers must be completely dried before refitting to the mattress.
Mattress covers can be tumble dried on a low heat setting for 90 minutes. Drying temperature must not exceed 40°C. Exceeding the temperature can cause significant damage to the mattress cover.
CAUTION: Do not use phenolic-based product for cleaning.
16

CAUTION: After cleaning, dry the mattress without direct exposure of sunlight.
7. STORAGE  To quickly vacuum air out from mattress for storage, rotate the CPR valve to OPEN position and disconnect the hose connector to release the air.  Lay the mattress out flat and upsides down.  Roll from the head end towards the foot end.  Packing strap can then be stretched around the rolled mattress to prevent unrolling.  The power cord could be wrapped around the Pump bumper or disconnected for storage.
8. MAINTENANCE General  Check main power cord and plug if there are abrasions or excessive wears.  Check mattress cover for signs of wear or damage. Ensure mattress cover and tubes are stubbed together correctly.  Check the air hoses for any kink or break. For replacement, please contact your local dealers.
Fuse replacement  Disconnect the plug from mains power when a blown fuse is suspected.  Remove the cover of the fuse holder by means of a small screwdriver.  Insert a new fuse of the correct rating in, and replace the cover of the fuse holder
back. The fuse rating should comply with the requested specification.
Air Filter Replacement After checking 10. TROUBLESHOOTING, if the air filter needs to be replaced:  Replace the air filter located at the back of the Pump.  The filter is reusable and can be washed gently with a mild detergent and water.
Dry the filter before use.  Check and replace air filter regularly if environment is dirty.
9. THE DISPOSAL OF AIR MATTRESS When the air mattress is broken or no longer be useable, the mattress and the Pump may be discarded for recycle.
17

10. TROUBLESHOOTING

PROBLEM

SOLUTION

The mattress is not able to  Check if the mattress model (model no. located inside the cover by

connect with the Pump

the foot end) Protekt Aire 8900 matches with the Pump model

Protekt Aire 8900. The AAA should be the same as BBB. If not,

please contact with the agent or distributor.  Check if the Transport Cap is removed and make sure the

connector is not broken.

The Pump is showing no indications of working

 Check if the plug is connected to the mains supply.  Check if the power switch is switched to ON position (press ).  Check if there is any blown fuse.

Power failure alarm failed  If the Pump is in operation but failed to trigger the power failure

alarm during power off. Charge the Pump for 6 hours or more of

operating time and if the power failure still not works then please

contact the dealer or agent for further investigation.

The low pressure light is  Check if the CPR is at CLOSE position constantly flashing and the  Check if the connection between air tubes to Pump unit is tightly

alarm is sounded

secured.  Check if all coupling connections along mattress are secured.  If the mains supply is normal but there is no sound of the Pump,

please remove the connector from the Pump to check if there's air

comes out. If not, please turn off the machine and contact the

dealer or agent for further investigation.  If all of above steps have been checked. Press "Alarm Mute" for

system to be verified again.

The Pump is on but the mattress is not alternated

 Ensure the mattress inflation is completed.  Check the Pump control panel the indicator of "Alternate" should

be lighted on, if not, switch it to "Alternate."  Check if "Service" alarm indicator is on with buzzer, if yes, contact

the dealer or agent for further investigation.

Service (Malfunction)

 Press "alarm mute" for system to be verified again. If the alarm is

Alarm is on

still on, please contact dealer or agent.

The Pump is operating noisily

 Make sure the Pump is resting against a solid surface.  If the noise is getting louder, contact the dealer or agent for further

investigation.

Patient is bottoming out  Pressure setting might be inadequate for the patient, adjust

(without alarm triggered)

comfort level to FIRM (refer to Table 1 Weight and Comfort Level

Reference Table) and wait for a few minutes for better comfort.  Follow the procedures "The low pressure light is constantly flashing

and the alarm is sounded" for inspection.

If the above information does not solve the problem, please contact your local dealer or agent for further support.

18

11. TECHNICAL DATA

11.1 Product Specification

PUMP UNIT

Model

80089

Dimension (cm)

42 (W) x 23 (D) x13 (H)

Weight (kg) Alternate Cycle Time (minutes) Static Time (minutes)

4.6 10/15/20/25/30 20

Pulsate Time (minutes)

10-30

Auto Firm Time (minutes) Pump Output Flow Range (Liter)

20
9 + 9
Note: The flow rate may be varied

because of the fluctuation of input

voltage

Pump Output Pressure Range (mmHg)

25 to 60 (±5)

Power

AC120V 60Hz

Current

0.43AMAX (@132V~)

Fuse Rating Frequency Classification

T1A 250VAC 60Hz (120V) Class II

Model

AIR MATTRESS 8"Mattress

Dimension (cm)
Weight (kg) Cell Material

200(L) x 90(W) x 20(H) 200(L) x 90(W) x 30(H) with bolsters 7.0

Nylon-TPU

No. of Air Cell

18 Cells

Cover Material

Nylon-PU with Quilted

Bottom Material Standard Max. Weight

Polyester-PU 660 lbs.

Warranty

Type BF 2 years

Operation Environment Storage Environment Environment Pressure

Warranty

2 year

ENVIRONMENTAL CONDITIONS

5 ~40

15%RH ~ 93%RH(no condensation)

-25~70 93%RH(no condensation)

70 kPa-101.3kPa

Environment Level

3000m

19

Water and Dust Protection Classification IP21 20

11.2 EMC Information (120V)

Guidance and manufacturer's declaration-electromagnetic emissions

The system is intended for use in the electromagnetic environment specified below. The customer or the user of the system should assure that it is used in such an environment.

Emission test

Compliance Electromagnetic environment-guidance

RF emissions

Group 1

The system uses RF energy only for its

CISPR 11

internal function. Therefore, its RF emissions

are very low and are not likely to cause any

RF emissions CISPR 11

Class B

interference in nearby electronic equipment. The system is suitable for use in all establishments, including domestic

Harmonic emissions Class A IEC 61000-3-2 Voltage fluctuations Compliance /flicker emissions IEC 61000-3-3

establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.

21

Guidance and manufacturer's declaration-electromagnetic immunity The system is intended for use in the electromagnetic environment specified below. The customer or the user of the system should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment-guidance

Electrostatic discharge(ESD) IEC 61000-4-2

+ 6 kV contact + 8 kV air

+ 6 kV contact + 8 kV air

Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%

Electrical fast transient/burst IEC 61000-4-4

+ 2kV for power supply + 2kV for power supply Mains power quality should be that

lines

lines

of a typical commercial or hospital

+ 1kV for input/output Not applicable

environment.

lines

Surge IEC 61000-4-5

+ 1kV line(s) to line(s) + 2kV line(s) to earth

+ 1kV differential mode Mains power quality should be that

Not applicable

of a typical commercial or hospital

environment.

Voltage Dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11

<5% UT(>95% dip in UT) for 0,5 cycle 40% UT(60% dip in UT) for 5 cycles 70% UT(30% dip in UT) for 25 cycles <5% UT(>95% dip in UT) for 5 s

<5% UT(>95% dip in UT) for 0,5 cycle 40% UT(60% dip in UT) for 5 cycles 70% UT(30% dip in UT) for 25 cycles <5% UT(>95% dip in UT) for 5 s

Mains power quality should be that of a typical commercial or hospital environment. If the user of the system requires continued operation during power mains interruptions, it is recommended that the system be powered from an uninterruptible power supply or a battery.

Power frequency(50/60 Hz) 3 A/m magnetic field IEC 61000-4-8

3 A/m

The system power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.

NOTE UT is the a.c. mains voltage prior to application of the test level.

22

Guidance and manufacturer's declaration-electromagnetic immunity

The system is intended for use in the electromagnetic environment specified below. The customer or the user of the system should assure that is used in such and environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment-guidance Portable and mobile RF communications equipment should be used no closer to any part of the system including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.

Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3

3 Vrms 150 KHz to 80 MHz

3 Vrms

3 V/m 80MHz to 2,5 GHz

3 V/m

Recommended separation distance: d = 1,2  d = 1,2  80MHz to 800 MHz d = 2,3  800MHz to 2,5 GHz Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range.b
Interference may occur in the vicinity of equipment marked with the following symbol:

NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the system is used exceeds the applicable RF compliance level above, the system should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the system.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
23

Recommended separation distance between portable and mobile RF communications equipment and the system

The system is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the system can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the system as recommended below, according to the maximum output power of the communications equipment.

Rated maximum output power of transmitter W

Separation distance according to frequency of transmitter m

150 kHz to 80 MHz d =1,2

80 MHz to 800 MHz d =1,2

800 MHz to 2,5 GHz d =2,3

0,01

0,12

0,12

0,23

0,1

0,38

0,38

0,73

1

1,2

1,2

2,3

10

3,8

3,8

7,3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can

be estimated using the equation applicable to the frequency of the transmitter, where p is the maximum output power rating of

the transmitter in watts (W) according to the transmitter manufacturer.

NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

24

