VPFlowScope M Firmware Update and VPStudio 3.2 Installation
Information regarding the latest firmware update for the VPFlowScope M and the required VPStudio software.
Introduction
VP Instruments informs you about a new firmware version for the VPFlowScope M. To configure your products with the correct pipe diameter settings and communication outputs, you need to install VPStudio 3.2.
Download VPStudio 3.2 from: VPInstruments Website. This software is continuously improved, so check for updates regularly.
Compatibility Note: VPFlowScope M with firmware 2.2.0 or higher is not compatible with older versions of VPStudio (1.0 and 2.0). VPStudio 3.2 is compatible with VPFlowScope M firmware 2.2.0 or higher.
Warning: Read this entire instruction first before updating your VPFlowScope M Transmitter. Pay especial attention to compatibility with versions and data logger functionality.
Upgrade Your VPFlowScope M
Before Upgrade: Settings Migration
The updater migrates the following settings from your device to the new firmware:
- Serial number
- Kind of product (VPFlowScope M without display: D000, with display: D010, and with display and datalogger: D011)
- Minimum and maximum 4..20 mA calibration value
- MAC address
- Production date
Any other settings (including e.g. Static IP) not mentioned here are not being migrated in a firmware update!
Upgrade Steps
- Download firmware version 2.3.2 from the VPInstruments website.
- Download VPStudio 3.2 from the VPInstruments website.
- Unzip the VPStudio 3.2 file and install VPStudio 3.2 using the Setup file.
- During VPStudio 3.2 installation, the VPFlowScope M Firmware Updater will also be installed on your PC. You can find it by searching in your start menu.
- Connect your VPFlowScope M via a mini USB cable to the PC. Open the VPFlowScope Firmware Updater and follow its instructions to update your VPFlowScope M.
- Be aware that the firmware file needs to be on your PC to be selected.
- Once ready, your VPFlowScope M is updated and ready to be configured via VPStudio 3.2.
- To update another VPFlowScope, restart the Firmware Updater first.
Visual Description: The document includes screenshots illustrating the process. One shows how to find the "VPFlowScope M Firmware updater" application in a Windows start menu search. Another displays the "VPFlowScope M Firmware Updater" interface, detailing fields for firmware information, update progress, and options to browse for firmware files and initiate the update.
VPFlowScope M Firmware 2.3.2 Features
The new firmware version 2.3.2 introduces the following improvements:
- Enjoy faster screen refreshes: The interface is removed from the data download process, and the "Download in progress" screen is eliminated, restoring access to the local interface during data transfers.
- Improved disconnection handling: The transmitter smartly works through brief probe disconnections, ensuring continuous process.
- Enhanced safety and functionality with 24V connection protocol: This protocol automatically activates upon probe connection and deactivates upon disconnection. It optimizes the cartridge's electrical performance and safeguards against accidental short circuits when probes are not connected.
- The pulse width of the pulse output is now set at 1000ms, aligning with manual specifications.
- VPFlowScope M Transmitter firmware 2.2.0 or higher is compatible with VPSensorCartridges with serial number 6100658 and higher.