R&S®ZN-Z5x, R&S®ZN-Z15x, R&S®ZN-ZE1xx Calibration Units

Instrument Security Procedures

Document Number: 1179039702 | Version: 03

Brand: Rohde & Schwarz

Tagline: Make ideas real

Introduction

This document describes the types of memory and their use in the calibration units R&S®ZN-Z5x, R&S®ZN-Z150, R&S®ZN-Z151, R&S®ZN-Z152, R&S®ZN-Z153, R&S®ZN-Z156, and R&S®ZN-ZE1xx. While every effort has been made to ensure the accuracy of the information herein, it is provided without warranty. Design iteration and revisions may result in minor differences between the information provided here and your product.

© 2022 Rohde & Schwarz GmbH & Co. KG. All rights reserved.

Note: Throughout this document, products from Rohde & Schwarz are indicated without the ® symbol, e.g., R&S®ZN-Z151 is indicated as R&S ZN-Z151.

Overview

Securing important information is crucial in many applications. Generally, highly secured environments do not allow any test equipment to leave the area unless it can be proven that no user information leaves with the test equipment, e.g., to be calibrated.

Regarding sanitization, the principal concern is ensuring that data is not unintentionally released [1]. This document provides a statement regarding the volatility of the memory types used and specifies the steps required to sanitize an instrument. The procedures in this document follow "NIST Special Publication 800-88: Guidelines for Media Sanitization" [1]. In addition, recommendations are provided to safeguard information on the product.

References: See the following literature for further information.

  • [1] Kissel Richard L. [et al.] Guidelines for Media Sanitization = Special Publication (NIST SP) = NIST SP - 800-88 Rev 1. - Gaithersburg: [s.n.], December 17, 2014.
  • [2] National Industrial Security Program Authorization Office Defense Security Service (DSS) Assessment and Authorization Process Manual (DAAPM). - May 6, 2019.
  • [3] ACSC Australian Cyber Security Centre Australian Government Information Security Manual, January 2020.

Instrument Models Covered

This section lists the various calibration unit models covered by this document.

Table 2-1: R&S ZN-Z5x, R&S ZN-Z15x, R&S ZN-ZE1xx Models

ProductPortsFrequency rangeOrder number
R&S ZN-Z5x
R&S®ZN-Z50 calibration unit2 ports, 3.5 mm (f)9 kHz to 9 GHz1335.6904.30
4 ports, 3.5 mm (f)9 kHz to 26.5 GHz1335.6904.32
R&S®ZN-Z51 calibration unit2 ports, 3.5 mm (f)100 kHz to 8.5 GHz1319.5507.34
4 ports, N (f)1319.5507.32
2 ports, N (f)1319.5507.74
R&S®ZN-Z52 calibration unit4 ports, 3.5 mm (f)100 kHz to 26.5 GHz1319.5507.72
R&S®ZN-Z53 calibration unit2 ports, 3.5 mm (f)100 kHz to 26.5 GHz1335.6991.30
2 ports, type N (f)100 kHz to 18 GHz1335.7046.32
R&S®ZN-Z54 calibration unit2 ports, 2.92 mm (f)9 kHz to 40 GHz1)1335.7046.72
R&S®ZN-Z55 calibration unit2 ports, 2.4 mm (f)9 kHz to 50 GHz1335.7117.92
R&S ZN-Z15x
R&S®ZN-Z150 calibration unit2 ports, N (f)5 kHz to 6 GHz1335.6710.72
R&S®ZN-Z151 calibration unit2 ports, SMA (f)100 kHz to 8.5 GHz1317.9134.32
2 ports, N (f)1317.9134.72
R&S®ZN-Z152 calibration unit6 ports, SMA (f)100 kHz to 8.5 GHz1319.6003.36
R&S®ZN-Z153 calibration unit4 ports, SMA (f)100 kHz to 8.5 GHz1319.6178.34
R&S®ZN-Z156 calibration unit2 ports, 1.85 mm (f)5 GHz to 67 GHz1332.7239.02
10 MHz to 67 GHz1332.7239.03
R&S ZN-ZE1xx
R&S®ZN-ZE104 calibration unit2 ports2)5 kHz to 4.5 GHz1350.8040K04
R&S®ZN-ZE109 calibration unit2 ports2)5 kHz to 9 GHz1350.8040K09
R&S®ZN-ZE118 calibration unit2 ports2)5 kHz to 18 GHz1350.8040K18
R&S®ZN-ZE126 calibration unit2 ports2)5 kHz to 26.5 GHz1350.8040K26

1) characterized up to 43.5 GHz

2) The R&S ZN-ZE1xx models are equipped with two of the following port options: R&S®ZN-ZE1-B170, N (f), port 1; R&S®ZN-ZE1-B171, N (m), port 1; R&S®ZN-ZE1-B270, N (f), port 2; R&S®ZN-ZE1-B271, N (m), port 2; R&S®ZN-ZE1-B130, 3.5 mm (f), port 1; R&S®ZN-ZE1-B131, 3.5 mm (m), port 1; R&S®ZN-ZE1-B230, 3.5 mm (f), port 2; R&S®ZN-ZE1-B231, 3.5 mm (m), port 2.

Security Terms and Definitions

Terms defined in Guidelines for Media Sanitization

According to NIST Special Publication 800-88 [1]: "Sanitization is a process to render access to target data (the data subject to the sanitization technique) on the media infeasible for a given level of recovery effort." It defines the following categories of sanitization:

  • Sanitization: Media sanitization refers to a process that renders access to target data on the media infeasible for a given level of effort.
  • Clear: Clear applies logical techniques to sanitize data in all user-addressable storage locations for protection against simple non-invasive data recovery techniques; typically applied through the standard Read and Write commands to the storage device, such as by rewriting with a new value or using a menu option to reset the device to the factory state (where rewriting is not supported).
  • Purge: Purge applies physical or logical techniques that render Target Data recovery infeasible using state of the art laboratory techniques.
  • Destroy: Destroy renders Target Data recovery infeasible using state of the art laboratory techniques and results in the subsequent inability to use the media for storage of data.

Control of Media

Another option to secure sensitive information is to keep physical media within the classified area, see [1], paragraph 4.4.

Statement of Volatility

Volatile Memory

"Memory components that do not retain data after removal of all electrical power sources, and when reinserted into a similarly configured system, are considered volatile memory components." [2]

Typical examples are RAM, e.g. SDRAM.

Non-volatile Memory

"Components that retain data when all power sources are discontinued are non-volatile memory components." [2]. In the context of this document, non-volatile memory components are non-user accessible internal memory types, e.g. EEPROM, Flash, etc.

Media

Media are types of non-volatile memory components. In the context of this document, media are user-accessible and retain data when you turn off power. Media types are Hard Disk Drives (HDD), Solid State Drives (SSD), Memory Cards, e.g. SD, microSD, CFast, etc., USB removable media, e.g. Pen Drives, Memory Sticks, Thumb Drives, etc. and similar technologies.

Notes on Memory Sizes

Due to the continuous development of memory components, the listed values of memory sizes may not represent the current, but the minimal configuration. This document uses the common notation kbyte, Mbyte and Gbyte for memory sizes, although the prefix multiplication factor is 1024.

Memory Types

Volatile Memory

Volatile memory modules refer to non-accessible internal storage devices, as described in Security terms and definitions > Volatile memory.

Table 4-1: Types of Volatile Memory

Memory typeLocationSizeContent / FunctionUser modifiable
SRAMµController52 kbyteTemporary information storage for instrument firmwareYes

Non-volatile Memory

Non-volatile memory modules refer to non-accessible internal storage devices, as described in Security terms and definitions > Non-volatile memory.

Table 4-2: Types of Non-volatile Memory

Memory typeLocationSizeContent / FunctionUser modifiable
FlashµController2x128 kbyte
  • Firmware
No
FlashOn-board flash memory128 Mbyte
  • Product identification data (serial number, hardware info)
  • Calibration data
  • Factory characterization data
  • User characterization data generated by the analyzer firmware
Yes

Media

Media memory modules refer to non-volatile storage devices, as described in Security terms and definitions > Media.

Note: R&S ZN-Z150, R&S ZN-Z151 and R&S ZN-Z156 are not equipped with a microSD card slot. Hence, they do not have a media memory.

Table 4-3: Types of Media Memory Modules

Memory typeLocationSizeContent / FunctionUser modifiable
microSD card (removable)see Location of the microSD card slot1 Gbyte (R&S ZN-Z5x, R&S ZN-Z15x)
8 Gbyte (R&S ZN-ZE1xx)
  • User data
  • User characterization data generated by the analyzer firmware
Yes

Instrument Sanitization Procedure

Volatile Memory

The volatile memory in the instrument does not have battery backup. It loses its contents when power is removed from the instrument.

To Remove Power

  • Disconnect the USB connection cable.

Provided the instrument remains without power for at least 10 minutes, all volatile memory modules lose their contents, see [3].

Non-volatile Memory

You can clear/purge the non-volatile memory by following the procedure below. The sanitization procedure complies with the definition of NIST [1], see "Terms defined in Guidelines for Media Sanitization" on page 5.

To Sanitize the Non-volatile Memory

Risk of losing data: The sanitization procedure clears/purges all user characterization data.

  1. Connect the calibration unit to the network analyzer.
  2. Access the characterization wizard of the network analyzer.
  3. Select the calibration unit from the list.
  4. Use one of the following functions to delete user characterization data from the selected calibration unit:
    • Use the "Delete" function to clear the user characterizations from the non-volatile memory.
    • Use the "Sanitize" function to purge all user characterizations from the non-volatile memory.

Note: The "Sanitize" function is available on R&S ZVA, R&S ZVB and R&S ZVT only. These network analyzers do not support calibration units R&S ZN-ZE1xx.

Media Sanitization

R&S ZN-Z150, R&S ZN-Z151 and R&S ZN-Z156 are not equipped with a microSD card slot. Hence, they do not require this kind of instrument declassification.

To Remove the microSD Card:

NOTICE! Do not remove the microSD card during operation. It can lose data.

  1. Power off the calibration unit by disconnecting the USB.
  2. Remove the microSD card at the microSD card slot of the instrument.

Table 5-1: Location of the microSD Card Slot

This table illustrates the physical location of the microSD card slot on different calibration unit models. The images show the R&S ZN-Z5x, R&S ZN-ZE1xx, and ZN-Z152 and ZN-Z153 models, with the microSD card slot highlighted by an orange box in each image.

Note: The images are descriptive and show the general placement of the microSD card slot.

3. Keep the microSD card under organizational control.

Operability Outside Secured Area

The sanitization does not affect the functionality of the R&S ZN-Z5x, R&S ZN-Z15x, R&S ZN-ZE1xx calibration unit. The instrument works properly after sanitization. When the calibration unit is back within the secured area, the original classified removable microSD card can be reinstalled.

Validity of Instrument Calibration

The Flash is the only memory type used to hold permanent adjustment values required to maintain the validity of the calibration unit's calibration. Therefore, the sanitization procedure does not affect the validity of the instrument's calibration.

Special Security Features

This section leads you to the information on how to use the security features of the network analyzer to protect the R&S ZN-Z5x, R&S ZN-Z15x, R&S ZN-ZE1xx from unauthorized access of classified information saved in the instrument.

Considerations for Saving User Characterization Data

Current Rohde&Schwarz desktop network analyzers allow you to create user characterizations of your calibration unit. For calibration units with microSD card slot, the characterization data can be written to the inserted microSD card instead of the internal flash memory.

See Recommended security settings > Saving user characterization data.

Recommended Security Settings

Basically, see the user manuals of your network analyzer models for the security concept of the calibration unit, including instructions on how to select where to save user characterization data. The user manuals of the network analyzers are provided for download on their product pages at www.rohde-schwarz.com.

The following section describes measures that protect user characterization data before issuing the instrument outside the secure environment.

Saving User Characterization Data

The R&S ZN-Z150, R&S ZN-Z151 and R&S ZN-Z156 are not equipped with a microSD card slot. Hence, the user characterization data is always saved on the internal flash memory.

To Save User Characterization Data on a microSD Card

You can save user characterization data on a microSD card instead of the calibration unit's internal flash memory.

  1. Insert a microSD card into the microSD card slot of the instrument.
  2. Connect the calibration unit to the network analyzer.
  3. Access the characterization wizard of the network analyzer.
  4. Activate the checkbox when saving the characterization data.

Tip: If the characterized calibration unit does not have an SD card slot, the checkbox is hidden. If the calibration unit has an SD card slot but the SD card is not accessible, the checkbox is grayed out.

The characterization data is saved on the microSD card.

Table 9-1: Location of the microSD Card Slot

This table illustrates the physical location of the microSD card slot on different calibration unit models. The images show the R&S ZN-Z5x, R&S ZN-ZE1xx, and ZN-Z152 and ZN-Z153 models, with the microSD card slot highlighted by an orange box in each image.

Note: The images are descriptive and show the general placement of the microSD card slot.

Models: ZN-Z5x Automatic Calibration Units, ZN-Z5x, Automatic Calibration Units, Calibration Units, Units

File Info : application/pdf, 11 Pages, 331.72KB

PDF preview unavailable. Download the PDF instead.

zn-z5x-zn-z15x-zn-ze1xx-instrumentsecurity-en-03

References

AH XSL Formatter V6.6 MR2 for Windows (x64) : 6.6.4.36256 (2018/11/29 19:37JST) Antenna House PDF Output Library 6.6.1359 (Windows (x64))

Related Documents

Preview Rohde & Schwarz R&S FPL1000 Spectrum Analyzer Instrument Security Procedures
Comprehensive guide to securing and sanitizing the Rohde & Schwarz R&S FPL1000 Spectrum Analyzer. Covers memory types, sanitization methods (Clear, Purge, Destroy), model specifics, and USB interface security, adhering to NIST SP 800-88 guidelines.
Preview R&S®QPS Walk2000 Getting Started Guide
This guide provides essential information for getting started with the R&S®QPS Walk2000 walk-through security scanner. Learn about its features, setup, operation, maintenance, and troubleshooting.
Preview R&S QPS201 Quick Personnel Security Scanner Getting Started Guide
A comprehensive guide for the R&S QPS201 Quick Personnel Security Scanner, covering essential safety instructions, regulatory compliance for Canada and the USA, key features, EMI suppression, system components, and detailed preparation for use, including installation and power connection procedures.
Preview R&S ARDRONIS: Drone Monitoring and Detection System
Discover the R&S ARDRONIS drone monitoring system from Rohde & Schwarz. This comprehensive solution offers early detection, identification, and disruption of unauthorized drones, enhancing security for critical infrastructure, public events, and sensitive areas. Learn about its advanced radio monitoring capabilities, various configurations, and integration potential.
Preview R&S®FPC Spectrum Analyzer Getting Started Guide
This guide provides essential information for setting up and operating the R&S®FPC Spectrum Analyzer, covering safety instructions, product overview, and basic operations.
Preview R&S RTB2000 Digital Oscilloscope User Manual
Comprehensive user manual for the Rohde & Schwarz R&S RTB2000 series digital oscilloscopes, detailing safety, operation, setup, analysis, measurements, and serial bus protocols.
Preview R&S®FSW-K30 Noise Figure Measurements User Manual
Comprehensive user manual for the Rohde & Schwarz R&S®FSW-K30 firmware application, detailing noise figure measurement setup, configuration, analysis, and remote control for R&S®FSW signal analyzers.
Preview R&S FSVA3000/FSV3000 Signal and Spectrum Analyzer User Manual
Comprehensive user manual for the Rohde & Schwarz FSVA3000 and FSV3000 Signal and Spectrum Analyzers, detailing features, operation, and measurement applications.