Green Lion GL-MN15 Portable Nebulizer
SKU: GNMININEBWH
Product Overview
Thank you for choosing our small nebulizer. Please read this manual carefully to ensure safe and effective use. Keep it in a safe place for future reference.
Scope of Use: For evaporating medications for therapeutic use.
Target Audience: Suitable for adults and children.
This nebulizer works using rapid circuit oscillation, which causes resonance in the piezoelectric ceramic transducer. This vibrates the fine mesh network, allowing the medication to pass through tiny pores, forming mist particles. These particles are directed to the patient's respiratory system via a face mask or mouthpiece for inhalation therapy.
The vaporized medication is absorbed through the mucous membranes and tissues of the oral cavity, throat, trachea, bronchi, and alveoli, ensuring the desired therapeutic effect.
Product Components
The nebulizer consists of the following components:
- Main Unit
- Medicine Cup
- Mouthpiece
- Adult Mask
- Child Mask
- USB Cable
- User Manual
Diagram Description: A diagram illustrates the main unit with labels pointing to the Medicine Cup, Power Button, and USB Type-C Charging Port, and Mist Outlet.
Technical Specifications
Parameter | Specification |
---|---|
Operating Frequency | 120 kHz |
Nebulization Rate | ≥ 0.3 ml/minute |
Medicine Temperature | ≤ 45°C |
Operating Noise | ≤ 22 dB |
Median Particle Size | 3 micrometers |
The average mass median particle size of the nebulizer was measured at 25°C and 46% humidity, using a 0.9% sodium chloride solution as a test solution. The figure below shows the particle size distribution curve measured under these conditions.
Diagram Description: A graph showing particle size distribution, with particle diameter on the x-axis (ranging from 0.1 to 100 micrometers) and cumulative distribution percentage on the y-axis (ranging from 0% to 100%).
Installation Instructions
- Remove all packaging materials and set them aside. Take out the main unit and accessories.
- Attach the assembled medicine cup to the main unit. You should hear a clear "click" when it is correctly installed.
- Attach accessories as shown in the medicine cup installation diagram.
- To add medication, invert the medicine cup, insert the cup opening into the main unit, and turn to secure it.
- Power Source Setup: This product does not come with a power adapter. Please use a reliable manufacturer's adapter that meets requirements IEC 60601-1 or GB/T 2010-2009. The power adapter specifications must comply with a 5.0V DC, 1A output.
Note: This device uses a rechargeable lithium battery, which is not replaceable and has a lifespan of 5 years.
- Connect one end of the USB data cable to the USB charging port and the other end to a USB power source to charge the lithium battery. Once the battery reaches operating voltage, the device will function normally.
Note: Before charging, ensure the power plug is connected. Do not charge the device with other electronic devices simultaneously.
6. Charging the Battery: When the battery charge level is low, use the power adapter (5V DC, 1A) to charge the lithium battery. Charging takes approximately 40 minutes. During charging, if the battery is not fully charged, the blue light will flash. After charging is complete, the blue light will remain solid.
Note: The lithium battery is pre-installed; do not attempt to disassemble or replace it yourself. The device cannot be used while charging.
Additional Notes
- Before using the device for the first time, ensure it is charged for at least 30 minutes.
- If the mist rate significantly decreases or stops, recharge the device immediately.
- The fully charged lithium battery provides up to 60 minutes of continuous operation.
- If the device is not used for more than a month, recharge the battery at least once a month during storage.
- When taking the device out of storage at approximately 20°C (whether from a higher or lower storage temperature), let it rest for 5 minutes before use.
Warning: Dispose of used batteries according to local environmental regulations. Do not dispose of them with household waste to avoid environmental pollution. Never attempt to disassemble or repair the device or its components. Do not attempt to replace the battery yourself.
Indicator Light Guide
Indicator Light | Status |
---|---|
Solid Green | Device is currently operating. |
Blinking Green (for 10 seconds) | The device will automatically stop after 10 minutes. |
Blinking Green (continuously) | Device voltage is below 3.2V and requires charging. |
Blinking Blue | Device is currently charging. |
Solid Blue | Device is fully charged. |
Usage Instructions
- Press the power button once to turn on the device and start the inhalation process.
- Before starting, gently shake the device horizontally to ensure the medication fully contacts the mesh.
- Inhale the medication slowly and deeply for optimal absorption.
- After completing the inhalation, press the power button again to turn off the device. Dispose of any remaining medication from the cup. Do not reuse leftover liquid.
- Rinse the medicine cup, lid, and accessories with tap water, then disinfect them according to the recommended guidelines.
Notes:
- Ensure the medication fully contacts the mesh during inhalation. Gently shaking the device is sufficient, but avoid tilting the device backward during use.
- Always follow your doctor's treatment instructions and maintain a calm, comfortable posture during the session for effective results.
- Over time, medication may accumulate around the mesh, potentially affecting mist performance. In such cases, turn off the device, remove accessories, and gently wipe away residue with a sterile gauze. Avoid touching the central mesh area to prevent potential damage.
Cleaning and Disinfection
To maintain optimal performance and hygiene, it is essential to clean and disinfect the medicine cup assembly (including the mist outlet and cup lid) and accessories after each use. Follow the recommended cleaning and disinfection procedures:
1. Product Cleaning
- Component Disassembly: Carefully detach the medicine cup assembly, including the mist outlet, cup lid, and accessories, from the main unit. Soak them in clean tap water at a temperature not exceeding 40°C for approximately 5 minutes to remove any residual medication.
- Drying Parts: After cleaning, use clean, sterile gauze to remove excess water. Allow all parts to air dry completely.
- Cleaning the Main Unit: Wipe the exterior surface of the main unit. If there is any medication residue on the electrode contacts, use clean, sterile gauze dampened with water (and wrung out) to gently clean the area. Then, allow the main unit to air dry completely.
- Storage: Store the device and its components in a clean, dry place to prevent contamination.
Notes:
- Do not wash the main unit directly to prevent water from entering the device.
- Ensure the main unit and its components are completely dry using clean, sterile gauze before storing.
- The inhalation mask and mouthpiece are intended for use by one person only to prevent cross-contamination. Refer to the "Cleaning and Disinfection" section for instructions.
- Always dispose of any leftover medication after use. Failure to do so may clog the nebulizer mesh, affecting its performance.
2. Product Disinfection
To maintain a high level of hygiene and safety, the medicine cup assembly (including the mist outlet and cup lid) and accessories must be disinfected after each use. The following disinfection methods are recommended:
- Chemical Disinfection (Recommended for medical institutions): Soak the medicine cup assembly (including the mist outlet, cup lid) and accessories in a solution of Benzalkonium Bromide (a new disinfectant) or 3% Hydrogen Peroxide for 10 minutes for effective disinfection. Alternatively, you can wipe the components with the disinfectant.
- After disinfection, rinse thoroughly with tap water and dry with clean, sterile gauze, or allow the parts to air dry. Note: Follow the instructions for Benzalkonium Bromide and Hydrogen Peroxide. Do not soak components for extended periods. Avoid using strong oxidizers like perchlorates or disinfectants that may damage metals, polymers, or plastics.
- Ethanol Disinfection: Soak the medicine cup assembly (including the mist outlet, cup lid) and accessories in 75% medical-grade ethanol for 5-10 minutes for disinfection. You can also disinfect them by wiping with ethanol.
- After disinfection, rinse thoroughly with tap water and dry with clean, sterile gauze, or allow the parts to air dry. Notes:
- Wipe any disinfectant residue from the components using sterile gauze before storage to ensure safe use.
- When rinsing or wiping the mist outlet, avoid touching the central mesh area to prevent damage and maintain the nebulizer's function.
Maintenance and Storage
1. Nebulizer Storage
- Store the nebulizer in an environment with a temperature between 10°C and 40°C.
- Relative humidity should be ≤ 80%, without condensation.
- The ideal air pressure for storage ranges between 860 kPa and 1060 kPa.
- Ensure the nebulizer is stored away from direct sunlight and in a well-ventilated area.
- Store the device in a location free from corrosive gases and away from heating devices or exposed heaters.
- After each use, clean and disinfect the nebulizer immediately.
- Once the device and its accessories are completely dry, store them together in the original packaging.
- Always ensure that the storage area meets the recommended environmental conditions, and avoid exposing the nebulizer to any shocks or rough handling.
Contraindications and Precautions
Warning:
- This product is not permitted for use with Pentamidine medications.
- It is prohibited for use in patients with pulmonary edema.
- Do not use the device during acute asthma attacks or severe obstructive pulmonary disease.
- This nebulizer is a medical device suitable for adults and children. Please follow the instructions in the manual precisely or use it under medical supervision.
- Children and individuals requiring special care should use it under the supervision of a guardian.
- In case of malfunction, please refer to the "Troubleshooting" section. Do not attempt to repair the device yourself. Contact an authorized service center for assistance.
- Always clean and disinfect the nebulizer before each use. Refer to the "Cleaning and Disinfection" section for appropriate instructions.
- Do not carry or store the device with medication inside the cup.
- Before each use, clean and disinfect the inhalation mask or mouthpiece. These tools should not be shared to prevent cross-infection. Refer to the "Cleaning and Disinfection" section for instructions.
- Do not use the nebulizer while bathing.
- Do not use the nebulizer in environments containing flammable gases or near oxygen/anesthetic gas mixtures.
- The device is designed for use in a controlled electromagnetic environment. Keep it away from potential sources of interference.
- Do not operate the device in high-temperature environments, as this may cause malfunction or fire.
- Do not expose the main unit or any other components to strong vibrations or shocks (such as drops or impacts).
- Do not use medications containing esters, oils, or suspended particles (including herbal extracts). Use only standard prescribed nebulizer solutions to avoid mesh damage.
- During cleaning, do not immerse the device in water or rinse it directly under the tap, especially near the USB port. Do not direct running water towards the buttons.
- Do not dry or disinfect any accessories in a microwave oven, as this may cause damage.
- When using the mask, cover the nose and mouth gently. Do not press the mask firmly against the face. Children are prohibited from using the nebulizer without supervision.
- Do not touch the mesh area with your fingers or sharp objects, as this may damage it or stop mist production.
- The nebulizer is rated IP22 for protection against dust and moisture. Do not store the product in humid, dusty, or sunny environments. Keep it away from insects.
- Avoid dropping or forcefully hitting the main unit or medicine cup, as this may cause damage and render the product unusable.
- Upon expiration of its service life, dispose of the device and accessories as medical waste. Follow local regulations for proper disposal. Illegal disposal may cause environmental pollution. Users should thoroughly disinfect the medical device before disposing of it as general waste to minimize environmental impact.
- Individuals who are not experienced should contact local authorities for the proper disposal of components that may pose biological hazards.
- Do not allow children to play with the USB power cable. Avoid charging near the patient's bed to prevent potential suffocation hazards.
- The mouthpiece is a small part. Keep it out of reach of children to prevent accidental ingestion.
Troubleshooting
Problem: Device not working
- Possible Cause 1: Power button is stuck or damaged. Solution 1: Please contact an authorized service center for assistance.
- Possible Cause 2: Battery power is low. Solution 2: Recharge the battery and restart the device.
- Possible Cause 3: Battery is damaged. Solution 3: Consult a qualified technician for battery replacement.
- Possible Cause 4: Printed Circuit Board (PCBA) is damaged. Solution 4: Please contact an authorized service center for assistance.
Problem: Cannot charge
- Possible Cause 1: USB cable is not connected properly or is faulty. Solution 1: Ensure the USB cable is properly connected to both the nebulizer and the power adapter.
- Possible Cause 2: Micro USB port is damaged. Solution 2: Contact an authorized service center for repairs.
- Possible Cause 3: Battery is damaged. Solution 3: Please contact an authorized service center for assistance.
Problem: No medicine output
- Possible Cause 1: Incorrect grip position, and medication is not contacting the mesh. Solution 1: Ensure the correct grip position for the nebulizer.
- Possible Cause 2: Medicine cup is not installed correctly. Solution 2: Reinstall the medicine cup and all its components correctly to ensure proper operation.
- Possible Cause 3: Medication is not suitable for use with the nebulizer. Solution 3: Ensure the medication is sufficient and suitable for use with the device. Consult your doctor for appropriate medication recommendations.
- Possible Cause 4: Excess bubbles in the medication. Solution 4: Adjust the nebulizer's angle and gently shake it to start the misting process.
- Possible Cause 5: Mesh is damaged. Solution 5: Inspect the mesh for any wear or cracks. If any damage is found, please contact an authorized service center for repair or replacement.
- Possible Cause 6: Poor connection or damage to the mesh sensor probe, connection points, or wires. Solution 6: Check for any wear or cracks on the mesh and sensor probe. Contact an authorized service center if necessary.
- Possible Cause 7: Leakage causing internal corrosion of the device. Solution 7: Inspect the device for any signs of leakage or internal corrosion. If any damage is found, please contact an authorized service center for assistance.
Problem: Low mist output
- Possible Cause 1: Insufficient battery power. Solution 1: Charge the battery.
- Possible Cause 2: Medication is not fully contacting the mesh. Solution 2: Adjust the nebulizer's angle so that the medication fully contacts the mesh.
- Possible Cause 3: Insufficient medication. Solution 3: Add more medication.
- Possible Cause 4: Medication is unsuitable for nebulization. Solution 4: Ensure the medication is suitable for nebulization. Consult your doctor if necessary.
- Possible Cause 5: Mesh openings are blocked. Solution 5: Clean the mesh, remove any blockages, and restart the device.
- Possible Cause 6: Mesh wear. Solution 6: Contact an authorized service center for further assistance.
Problem: Medicine cup leakage
- Possible Cause 1: Medicine cup latch is damaged. Solution 1: Replace the medicine cup.
- Possible Cause 2: Medicine cup lid sealing ring is missing or damaged. Solution 2: Contact an authorized service center.
- Possible Cause 3: Cracks in the medicine cup. Solution 3: Contact an authorized service center for replacement.
Problem: Medicine residue in the device
- Possible Cause: Nebulization stops, and the device shuts off automatically. Solution: This is normal. Clean the medicine residue after nebulization.
Problem: Liquid accumulation
- Possible Cause 1: Due to temperature differences, mist particles condense at the mist outlet.
Statement
The device is designed for use in the electromagnetic environment specified below. The buyer or user must ensure that the device is used in accordance with this electromagnetic environment.
Specifications
Model Number | GL-MN15 |
---|---|
Material | ABS + PC |
Battery | 250 mAh |
Power | 1 Watt |
Medicine Capacity | 0.5 ml - 10 ml |
Nebulization Rate | ≥ 0.3 ml/minute |
Particle Size | 3 micrometers |
Vibration | 120 kHz |
Noise Level | ≤ 22 dB |
Water Resistance | IP22 |
Charging Port | USB-C |
Charging Time | 1 hour |
Working Time | 60 minutes |
Product Weight | 39 grams |
Product Dimensions | 41*63*35 mm |
Electromagnetic Compatibility (EMC) Information:
Note: UT refers to the alternating current (AC) voltage measured before applying the test level.
Immunity Test: Electrostatic Discharge (GB/T 26262-2019)
- Test Level per IEC 60601: Requires contact discharge of 6 kV and air discharge of 8 kV.
- Compliance Level: The device complies with 6 kV contact discharge and 8 kV air discharge.
- Electromagnetic Environment Guidance: The floor should be made of wood, concrete, or tile. If the floor is covered with synthetic materials, the relative humidity should be at least 30%.
Immunity Test: Electrical Fast Transient/Burst (GB/T 26262-2019)
- Test Level per IEC 60601: Requires ±2 kV for power lines and ±1 kV for input/output lines.
- Compliance Level: This test is not applicable to the device.
- Electromagnetic Environment Guidance: The device should be used with a non-grid power supply.
Immunity Test: Surges (GB/T 26262-2019)
- Test Level per IEC 60601: Requires ±1 kV for differential mode and ±2 kV for common mode.
- Compliance Level: This test is not applicable to the device.
- Electromagnetic Environment Guidance: The device should be used with a non-grid power supply.
Immunity Test: Voltage Dips, Short Interruptions, and Voltage Variations on Input Power Supply Lines (GB/T 26262-2019)
- Test Level per IEC 60601: The device can withstand less than 0% voltage for 0.5 seconds (95% voltage dip), 40% voltage for 5 cycles (60% voltage dip), and 70% voltage for 25 cycles (30% voltage dip). The device can also withstand 0 voltage for 5 seconds (>95% voltage dip).
- Compliance Level: This test is not applicable to the device.
- Electromagnetic Environment Guidance: The device should be used with a non-grid power supply.
Immunity Test: Power Frequency Magnetic Fields (50 Hz / 60 Hz) (GB/T 26262-2019)
- Test Level per IEC 60601: Requires 3 A/m.
- Compliance Level: The device complies with 3 A/m.
- Electromagnetic Environment Guidance: The power frequency magnetic field should match those typically found in a commercial or hospital environment.
Immunity Test: Radiated RF Communications (GB/T 26262-2019)
- Test Level per IEC 60601: 3 V/m from 150 kHz to 80 MHz.
- Compliance Level: 3 V/m.
- Electromagnetic Environment Guidance: Portable and mobile RF communication equipment should not be used closer to any part of the device, including cables, than the recommended isolation distance. This distance is calculated using the formula related to the transmitter's frequency.
Recommended Isolation Distance:
- d = 1.2 √P for 150 kHz to 80 MHz
- d = 2.3 √P for 80 MHz to 800 MHz
- d = 2.3 √P for 800 MHz to 2.5 GHz
Note: P refers to the maximum output power of the transmitter as stated by the manufacturer, in Watts (W). The field strength for a fixed transmitter is determined by an electromagnetic field survey and should be less than the compliance level in each frequency range.
Note: If the device's field strength exceeds the above-specified RF compliance level, the device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the device.
Note: For the entire frequency range from 150 kHz to 80 MHz, the field strength should be maintained at a maximum of 3 V/m.
Note: These guidelines may not apply in all situations. Electromagnetic wave propagation is affected by absorption and reflection from buildings, objects, and the human body.
Note: Interference may occur near equipment bearing the following symbols: [Symbol indicating RF interference]
Disposal
This product should not be disposed of as unsorted municipal waste. It is important to separate this waste for treatment and recycling in accordance with local waste management regulations.
Warranty
Products purchased directly from the Green Lion website or from our store come with a 24-month warranty.
When purchasing Green Lion products from any of our authorized distributors, you will receive a 12-month warranty. If you wish to extend this warranty, please visit our website at https://www.greenlion.net/warranty and fill out the form with your information. Don't forget to upload a photo of the product as well. After we review and accept your request, we will send you an email to confirm the extension of your product's warranty.
For more information, please visit: https://www.greenlion.net/warranty