Instruction Manual for kaise models including: KF-20, KF-20 Analog Multitester, Analog Multitester, Multitester

KF-20英文取説【第3版】表

ANALOG MULTITESTER

KF-20 Analog Multimeter

Instruction Manual|KAISE CORPORATION|Measuring Instrument for Automotive Maintenance|Online Shopping


File Info : application/pdf, 2 Pages, 337.20KB

PDF preview unavailable. Download the PDF instead.

e instruction kf20 3
ANALOG MULTITESTER INSTRUCTION MANUAL
KF-20

FOR SAFETY MEASUREMENTS!!
To prevent an electrical shock hazard to the operator and/or damage to the instruments, read this instruction manual carefully before using the Multitester. WARNINGS with the symbol on the Multitester and this instruction manual are highly important. Important Symbols
The symbol listed in IEC 61010-1 and ISO 3864 means "Caution (refer to instruction manual)". WARNING : The symbol in this manual advises the user of an electrical shock hazard that could result in serious injury or even death. CAUTION : The symbol in this manual advises the user of an electrical shock hazard that could cause injury or material damages.

WARNING
Do not measure High Power Line that might exceed 2kVA. Measurement on High Power Line is very dangerous. It sometimes includes High Surge Voltage that could cause dangerous arcs of explosive short in the instrument and could result in serious injury to the operator. When measuring high voltage, even for Low Power Line (Low Energy Circuit), take extra care to avoid electrical shock hazard and/or damage to the instrument.

INTRODUCTION
Thank you for purchasing KAISE "MODEL KF-20 ANALOG MULTITESTER". To obtain the maximum performance of this instrument, read this Instruction Manual carefully, and take safe measurement.

1. UNPACKING AND INSPECTIONS

Inspect the instrument and acessories for transport damage. If there is any damage or missing items, ask your local dealer for replacement. Confirm that the following items are contained in the package.

1. Analog Multitester 2. Test Lead (100-51) 3. Holster 4. Spare Battery (1.5V R6P) 5. Instruction Manual

1 pce. 1 set 1 pce. 1 pce. 1 pce.

2. SPECIFICATIONS

2-1. MEASUREMENT SPECIFICATIONS

DC Voltage ( DC-V ) AC Voltage ( AC-V ) DC Current ( DC-mA ) Resistance (  )
1.5V Battery Test Hi Range Lo Range

0.3 / 3 / 12 / 30 / 120 / 300 / 1200V 12 / 30 / 120 / 300 / 1200V
60A / 3mA / 30mA / 300mA 5K / 500K / 5M
1.5V R20P / R14P / R6P 1.5V 250mA 1.5V 50mA

- 1 -

2k 1k
2k 1k 2k 1k 2k 1k
2k 1k

2-2. GENERAL SPECIFICATIONS

Accuracy Overload Protection

DC : ±3 f.s., AC : ±4 f.s.  : ±3 f.s.length
×1 and 300mA range : protected by internal diode and 0.3A / 250V fuse

Power Supply

1.5V R6P (AA) battery ×1

Dimensions & Weight 136(H)×90(W)×30 (D)mm, 215g

Accessories

100-51 Test Lead ×1 set Holster ×1 1.5V R6P (AA) Battery (built-in and spare) ×1 each Spare Fuse (0.3A/250V)×1 (inside the rear case) Instruction Manual

Optional Accessories

100-41 Test Lead Kit, 100-62 Test Lead Set 940 Alligator Clips, 948 Alligator Clips 793 Coil-Type Contact Pin

3. SAFETY PRECAUTIONS
Correct knowledge of electric measurements is essential to avoid unexpected danger such as operator's injury or damage to the instrument. Read carefully and observe the following precautions for safety measurements.
3-1. WARNINGS WARNING 1. Checks of Body and Test Lead
Before measurement, confirm the body of this instrument and handle insulators of the Test Lead have no cracks or any other damages. Dust, grease and moisture must be removed.
WARNING 2. High Power Line Measurements Prohibited This instrument is applicable for Low Energy Circuit measurements only. Do not use for High Power Line (High Energy Circuit) measurements such as Distribution Transformers, Bus Bars and Large Motors.
WARNING 3. Warning for High Voltage Measurements Even for Low Energy Circuits of electric/electronic appliances, such as heating elements, small motors, line cords and plugs, High Voltage Measurements are very dangerous. Do not touch Multitester, Test Leads, and any part of the circuit. Generally, shock hazard could occur when the current between the circuit, that involves more than 33V rms or 46.7V DC or peak, and ground goes up to 0.5mA or more.
WARNING 4. Dangerous Voltage Measurement Procedures For dangerous voltage measurement, strictly observe the warnings below.
(refer to fig.1) 1. Turn off the circuit to be measured. 2. Plug black test lead to "- (COM)" terminal and red test lead into "+" terminal. 3. Attach black and red alligator clips to test lead pins. 4. Set RANGE Switch to desired DC or AC voltage range. 5. Connect black alligator clip to - (earth) side and red alligator clip to + (positive)
side of the measuring circuit. 6. Do not hold Multitester and test leads in your hands. Keep safety distance
from power source or circuit to be measured not to touch the dangerous voltage. 7. Turn on the circuit to be measured, and read the meter. 8. Turn off the circuit again. Detach alligator clips after confirming the meter returns into "0".

5k 500 200

DC 0

AC

0

0

BATT

100

50 30

5 2

10 4 2

15
6 3

1 R

E

P

L

A C 0.5

E

0

20

10

20

8 4

1025

5

1.0

G O O D

1.5

KF-20

OHMS 5

0

6 1230

DC AC

BATT 1.5 VOLT

20,000/ V DC 10,000/ V DC PROTECTION

DROP PROOF

Kaise Japan

300-12

NO.

1200 300 120

 1 100 1k

1200 300 AC-V 120

30

30

12

12

3

Hi

0.3

603 DC30mA300

Lo BAT.

C O M

Black Test Lead

Allligator Clip

Red Test Lead

fig. 1

In case of live-line measurement, strictly observe the warnings below (refer to fig 2.)
1. Set RANGE Switch to desired DC or AC voltage range. 2. Attach alligator clip to black test lead pin, and connect to - (earth) side of the
measuring circuit. 3. Do not hold Multitester in your hands. Keep safety distance from power source
or circuit to be measured not to touch the dangerous voltage. 4. Connect red test lead to + (positive) side, and read the meter. 5. Detach red test lead from the circuit first, then detach black alligator clip.
- 2 -

5k 500 200

DC 0

AC

0

0

BATT

100

50 30

5 2

10 4 2

15
6 3

1 R

E

P

L

A C 0.5

E

0

20

10

20

8 4

1025

5

1.0

G O O D

1.5

KF-20

OHMS 5

0

6 1230

DC AC

BATT 1.5 VOLT

20,000/ V DC 10,000/ V DC PROTECTION

DROP PROOF

Kaise Japan

300-12

NO.

1200 300 120

 1 100 1k

1200 300 AC-V 120

30

30

12

12

3

Hi

0.3

603

30 300 DC mA

Lo BAT.

C O M

Black Test Lead

Allligator Clip

Red Test Lead
fig. 2

WARNING 5. Correct Selection of RANGE Switch Always confirm that RANGE Switch is set to the correct position. Do not measure voltage in DC-mA,  and BAT. ranges.
outlet

5k 500 200

DC 0

AC

0

0

BATT

100

50 30

5 2

10 4 2

15
6 3

1 R

E

P

L

A C 0.5

E

0

20

10

20

8 4

1025

5

1.0

G O O D

1.5

KF-20

OHMS 5

0

6 1230

DC AC

BATT 1.5 VOLT

20,000/ V DC 10,000/ V DC PROTECTION

DROP PROOF

Kaise Japan

300-12

NO.

1200 300 120

 1 100 1k

1200 300 AC-V 120

30

30

12

12

3

Hi

Lo

0.3 603 DC30mA300

BAT. C O M

when measuring DC. mA /range

×
fig. 3

WARNING 6. Maximum Measurement Range Observance Measurement exceeding the specified maximum values of each measurement range is prohibited.
WARNING 7. Test Lead Detachment Detach test leads from the measuring circuit before rotating RANGE Switch or opening rear case for battery or fuse replacement.

3-2. GENERAL WARNINGS AND CAUTIONS
WARNING 1. Children and the persons who do not have enough knowledge about electric measurements must not use this instrument.
WARNING 2. Do not measure the electricity naked or barefooted to protect yourself from electrical shock hazard.
WARNING 3. Be careful not to get hurt with the sharp test lead pins. CAUTION 1. Do not polish the case or attempt to clean it with any cleansing
fluid like gasoline or benzine. If necessary, use silicon oil or antistatic fluid. CAUTION 2. Avoid the multitester from hard mechanical shock or vibration, high temperature and strong magnetic field. CAUTION 3. Remove the batteries when the multitester is out of use for a long time. The exhausted batteries might leak electrolyte and corrode the inside.

4. NAME ILLUSTRATION

Meter Case Pointer
Zero Adjustment Screw
RANGE Switch

5k 500 200

DC 0

AC

0

0

BATT

100

50 30

5 2

10 4 2

15
6 3

1 R

E

P

L

A C 0.5

E

0

20

10

20

8 4

1025

5

1.0

G O O D

1.5

KF-20

OHMS 5

0

6 1230

DC AC

BATT 1.5 VOLT

20,000/ V DC 10,000/ V DC PROTECTION

DROP PROOF

Kaise Japan

300-12

NO.

1200 300 120

 1 100 1k

1200 300 AC-V 120

30

30

12

12

3

Hi

0.3

603

30 300 DC mA

Lo BAT.

C O M

Rear Case Meter Scale
0ADJ. Knob
 Terminal ( COM) Terminal

100-51 Test Lead

fig. 4 - 3 -

5. MEASUREMENT PROCEDURES
5-1. PREPARATION FOR USE
1. INSTRUCTION MANUAL Read INSTRUCTION MANUAL carefully to understand the specification and functions correctly. 3. SAFETY PRECAUTIONSis higly important for safety measurement.
2. BATTERY KF-20 contains 1 x 1.5V R6P (AA) battery. Replace it referring to 6-1. BATTERY REPLACEMENTwhen exhausted.
3. FUSE KF-20 contains 1 x fast-acting 0.3A/250V (5x20mm) fuse to protect Resistance and Current measurement ranges. For replacement, refer to 6-2. FUSE REPLACEMENT.
4. ZERO ADJUSTMENT Before measurement, confirm that meter points to "0". If not, adjust the pointer by Zero Adjustment Screw to take accurate measurement.
5. TEST LEAD Insert black test lead into - (COM) terminal, and red test lead into + terminal.

6. RANGE SWITCH
(1) Range Selection a. Set the switch to a desired measurement range.
To measure uncertain voltage or current, firstly measure the approximate value at the highest range, then change the switch into the suitable range. b. For correct voltage and current measurements, select the range to point the right side of the meter scale. (between center scale and maximum scale) c. For correct resistance measurement, select the range to point around the center of OHMS scale.
(2) Confirmation of the Range Always confirm that RANGE Switch is set to the proper measurement range. To avoid serious damage to the instrument, do not measure voltage in DC-mA,  and BAT ranges.
WARNING Detach test leads from the circuit before rotating RANGE Switch to avoid electric shock hazard and serious damage to the instrument.

7. Meter Scale Reading

5k 500 200

DC 0

AC

0

0

BATT

100

50 30

5 2

10 4 2

15
6 3

1 R

E

P

L

A C 0.5

E

0

20

10

20

8 4

1025

5

1.0

G O O D

1.5

KF-20

OHMS
5

0

6 1230

DC AC

BATT 1.5 VOLT

20,000/ V DC 10,000/ V DC PROTECTION

DROP PROOF

OHMS scale DC/AC scale
BATT scale

Kaise Japan

300-12

NO.

fig. 5

DC/AC scale (for AC/DC Voltage, DC Current measurements : V, A, mA)
Select suitable scale from "0 - 6", "0 - 12", or "0 - 30" applying certain multiple. Reading examples : DC 0.3V range : Read "0 - 30" scale multiplying by 1/100. DC 120V range : Read "0 - 12" scale multiplying by 10. DC 60uA range : Read "0 - 6" scale multiplying by 10. DC 300mA range : Read "0 - 30" scale multiplying by 10.

OHM scale (for Resistance measurement : ) Multiply the scale value depending on the measurement range. Reading examples : "x 1" range : Read the scale directly. "x 100" range : Multiply the scale value by 100. "x 1k" range : Multiply the scale value by 1,000.
8. Notes for the Measurement For accurate measurement, read following notes carefully. 1. Before measurement, confirm that meter points to "0". (zero-adjusted) 2. Select proper range referring to 5-1. 6-(1) Range Selection. 3. Read the meter from directly over the pointer. 4. Measurement in strong magnetic field or on iron plate causes reading error or
meter sensitivity wrong. 5. Once-a-year periodical calibration is necessary to maintain tolerance accuracy
and to take safety measurements.
9. Protection Circuit The internal protection circuit against over current for meter movement and the internal circuit is installed. But, the instrument should be defective by such as over input measurement, or applying voltage to resistance or current range.
- 4 -

2k 1k 2k 1k

2k 1k 2k 1k

9. Protection Circuit The internal protection circuit against over current for meter movement and the internal circuit is installed. But, the instrument should be defective by such as over input measurement, or applying voltage to resistance or current range.
5-2. DC VOLTAGE MEASUREMENTS (DC-V)
WARNINGS
 Do not measure high power line that might exceed 2kVA.  Do not exceed the maximum input value of each range.  Confirm that RANGE Switch is set to the correct position.  Read 3. SAFETY PRECAUTIONScarefully for safety measurement.
1. Plug Black Test Lead into "- (COM)" terminal and Red Test Lead into "+" terminal.
2. Set RANGE Switch to a desired DC-V range. NOTE : To measure uncertain voltage, firstly measure the approximate value at 1200V range, then change the switch into the suitable range. Detach test leads from the circuit before rotating RANGE Switch.
3. Connect black test lead to  (earth) side and red test lead to  (positive) side of the measuring circuit. NOTE : Connect the multitester IN PARALLEL to the circuit. NOTE : Use alligator clips for dangerous voltage measurement.
4. Read the meter on DC/AC scale referring to 5-1. 7. Meter Scale Reading. 5. Detach test leads from the circuit.

5k 500 200

DC 0

AC

0

0

BATT

100

50 30

5 2

10 4 2

15
6 3

1 R

E

P

L

A C 0.5

E

0

20

10

20

8 4

1025

5

1.0

G O O D

1.5

KF-20

OHMS 5

0

6 1230

DC AC

BATT 1.5 VOLT

20,000/ V DC 10,000/ V DC PROTECTION

DROP PROOF

Kaise Japan

300-12

NO.

1200 300 120

 1 100 1k

1200 300 AC-V 120

30

30

12

12

3

Hi

0.3

603

30 300 DC mA

Lo BAT.

C O M

Black Test Lead

Red Test Lead

battery fig. 6

5-3. AC VOLTAGE MEASUREMENTS (AC-V)
WARNINGS
 Do not measure high power line that might exceed 2kVA.  Do not exceed the maximum input value of each range.  Confirm that RANGE Switch is set to the correct position.  Read 3. SAFETY PRECAUTIONScarefully for safety measurement.
1. Plug Black Test Lead into "- (COM)" terminal and Red Test Lead into "+" terminal.
2. Set RANGE Switch to a desired AC-V range. NOTE : To measure uncertain voltage, firstly measure the approximate value at 1200V range, then change the switch into the suitable range. Detach test leads from the circuit before rotating RANGE Switch.
3. Connect black test lead to  (earth) side and red test lead to  (positive) side of the measuring circuit. NOTE : Connect the multitester IN PARALLEL to the circuit. NOTE : Use alligator clips for dangerous voltage measurement.
4. Read the meter on DC/AC scale referring to 5-1. 7. Meter Scale Reading. 5. Detach test leads from the circuit.
outlet

5k 500 200

DC 0

AC

0

0

BATT

100

50 30

5 2

10 4 2

15
6 3

1 R

E

P

L

A C 0.5

E

0

20

10

20

8 4

1025

5

1.0

G O O D

1.5

KF-20

OHMS 5

0

6 1230

DC AC

BATT 1.5 VOLT

20,000/ V DC 10,000/ V DC PROTECTION

DROP PROOF

Kaise Japan

300-12

NO.

1200 300 120

 1 100 1k

1200 300 AC-V 120

30

30

12

12

3

Hi

0.3

603

30 300 DC mA

Lo BAT.

C O M

Black Test Lead
C O M

Red Test Lead
fig. 7

- 1 -

5-4. DC CURRENT MEASUREMENTS (DC-mA)
WARNINGS
 Do not measure over 300mA DC.  Do not exceed the maximum input value of each range.  Confirm that RANGE Switch is set to the correct position.  Read 3. SAFETY PRECAUTIONScarefully for safety measurement.

1. Plug Black Test Lead into "- (COM)" terminal and Red Test Lead into "+" terminal.
2. Set RANGE Switch to a desired DC-mA range. WARNINGS
 300mA range is protected by 0.3A/250V fuse, but 60uA, 3mA, and 30mA ranges are not.
 Do not measure voltage in DC-mA range.  Do not measure wall outlet or car battery by mistake.
3. Turn off the circuit to be measured and discharge the capacitors. 4. Connect black test lead to  (earth) side and red test lead to  (positive) side
of the measuring circuit. NOTE : Connect the multitester IN SERIES to the circuit. NOTE : Using alligator clips is recommended for safety measurement. 5. Turn on the circuit to be measured. 6. Read the meter on DC/AC scale referring to 5-1. 7. Meter Scale Reading. 7. Turn off the measured circuit again and discharge the capacitors. 8. Detach test leads from the circuit.

5k 500 200

DC 0

AC

0

0

BATT

100

50 30

5 2

10 4 2

15
6 3

1 R

E

P

L

A C 0.5

E

0

20

10

20

8 4

1025

5

1.0

G O O D

1.5

KF-20

OHMS 5

0

6 1230

DC AC

BATT 1.5 VOLT

20,000/ V DC 10,000/ V DC PROTECTION

DROP PROOF

Kaise Japan

300-12

NO.

1200 300 120

 1 100 1k

1200 300 AC-V 120

30

30

12

12

3

Hi

Lo

0.3

603

30 300 DC mA

BAT. C O M

Red Test Lead
Black Test Lead

 

Switch fig. 8

5-5. RESISTANCE MEASUREMENTS ()
WARNINGS
 Do not measure voltage in  range.  When measuring in-circuit resistor, turn off the circuit to be measured and
discharge the capacitors.
1. Plug Black Test Lead into "- (COM)" terminal and Red Test Lead into "+" terminal.
2. Set RANGE Switch to a desired  range. 3. When measuring in-circuit resistor, turn off the circuit to be measured and
discharge the capacitors. 4. 0 Adjustment
Short-circuit metal pins of the test leads. Turn 0 ADJ. Knob until the meter points to "0". NOTE : 0 Adjustment is necessary for each time when  range is selected. NOTE : 0 Adjustment does not work if the battery is exhausted. Replace it
referring to6-1. BATTERY REPLACEMENT. 5. Connect test leads to resistor to be measured detaching its one side from the
circuit. 6. Read the meter on OHMS scale referring to 5-1. 7. Meter Scale Reading.
NOTE : Read the scale directly when measuring in "x 1" range. In "x 100" range, multiply the scale by 100, and in "x 1k" range, multiply by 1,000.
NOTE : For correct resistance measurement, select the range to point around the center of OHMS scale.
7. Detach test leads from the resistor.

5k 500 200

DC 0

AC

0

0

BATT

100

50 30

5 2

10 4 2

15
6 3

1 R

E

P

L

A C 0.5

E

0

20

10

20

8 4

1025

5

1.0

G O O D

1.5

KF-20

OHMS 5

0

6 1230

DC AC

BATT 1.5 VOLT

20,000/ V DC 10,000/ V DC PROTECTION

DROP PROOF

Kaise Japan

300-12

NO.

1200 300 120

 1 100 1k

1200 300 AC-V 120

30

30

12

12

3

Hi

Lo

0.3

603

30 300 DC mA

BAT. C O M

Black Test Lead

- 2 -

Resistor

Red Test Lead
fig. 9

5-6. 1.5V BATTERY TEST (BATT)
Battery exhausted level can be tested in Hi (1.5V 250mA), or Lo (1.5V 50mA) ranges. Testable batteries : 1.5V R20P (D), 1.5V R14P (C), 1.5V R6P (AA), 1.5V R03 (AAA)
WARNINGS
 Do not test non-specified batteries to avoid electric shock hazard and serious damage to the instrument.
 Do not test car battery.  Do not measure voltage or current in BAT range.
1. Plug Black Test Lead into "- (COM)" terminal and Red Test Lead into "+" terminal.
2. Set RANGE Switch to "Hi" or "Lo" of BAT range. NOTE : "Lo" range applies 50mA, and "Hi" range applies 250mA to the battery. Confirm consumption current of the instruments that the battery will be used to select the proper testing range. Popular range selection : 1.5V R20P (D), 1.5V R14P (C) : "Hi" range 1.5V R6P (AA), 1.5V R03 (AAA) : "Lo" range
3. Connect black test lead to  (earth) side and red test lead to  (positive) side. 4. Read the meter on BATT scale.
Meter points "GOOD (blue)" : The tested battery is good. Meter points "REPLACE (red)" : The tested battery is exhausted and needs to be replaced. NOTE : Even if meter points "REPLACE (red)" for 1.5V R6P (AA) battery tested in "Lo" range, it may be used with low consumption current instruments.

6MAINTENANCE
6-1. BATTERY REPLACEMENT
WARNINGS
 To avoid electrical shock, detach test leads from circuit when when opening rear case for battery or fuse replacement.
 Be sure to tighten the screw of rear case after the battery replacement
Replace the batteries when it is exhausted and "0 Adjustment" does not work. 1. Detach test leads from the circuit. 2. Unscrew rear case. 3. Remove rear case opening from meter side and slide to the upward. 4. Replace an exhausted battery into a new 1.5V R6P (AA) battery in correct
polarity. 5. Close rear case fitting from bottom to up, and tighten the screw.

Screw

1.5V R6P Battery

Rear Case

fig. 10

NOTE : Installed battery is for inspection purpose and might be exhausted earlier than the new battery.
6-2. FUSE REPLACEMENT
WARNINGS
 Always use the specified fuse. Do not use the instrument shorting fuse holder or without using the fuse.
 Be sure to tighten the screw of rear case after the fuse replacement.
DC Current and Resistance measurement cannot be made if fuse blows out. This instrument contains one spare fuse inside the rear case. 1. Open rear case referring to 6-1. BATTERY REPLACEMENT. 2. Replace a blowout fuse into a new one.
Fuse type : Fast-acting 0.3A/250V (5x20mm) 3. Close rear case fitting from bottom to up, and tighten the screw.

- 3 -

6-3. PERIODICAL CHECK AND CALIBRATION
Periodical check and calibration is necessary to make safety measurements and to maintain the specified accuracy. The recommended check and calibration term is once a year and after the repair service. This service is available at KAISE AUTHORIZED SERVICE AGENCY through your local dealer.
6-4. REPAIR
Repair service is available at KAISE AUTHORIZED SERVICE AGENCY through your local dealer. Pack the instrument securely with your name, address, telephone number and problem details, and ship prepaid to your local dealer.
Check the following items before asking repair service. 1. Check the battery connection, polarity, and capacity. 2. Confirm that RANGE Switch is set to the correct position. 3. Confirm that the body of this instrument and handle insulators of the Test
Leads have no cracks or any other damages. 4. Check if any noise affects the instrument. This instrument is fully shielded
against noise, but possibly to be affected by very strong noise.
WARRANTY
KF-20 is warranted in its entirety against any defects of material or workmanship under normal use and service within a period of one year from the date of purchase of the original purchaser. Warranty service is available at KAISE AUTHORIZED SERVICE AGENCY through your local dealer. Their obligation under this warranty is limited to repairing or replacing KF-20 returned intact or in warrantable defect with proof of purchase and transport charges prepaid. KAISE AUTHORIZED DEALER and the manufacturer, KAISE CORPORATION, shall not be liable for any consequential damages, loss or otherwise. The foregoing warranty is exclusive and in lieu of all other warranties including any warranty of merchantability, whether expressed or implied. This warranty shall not apply to any instrument or other article of equipment which shall have been repaired or altered outside of KAISE AUTHORIZED SERVICE AGENCY, nor which have been subject to misuse, negligence, accident, incorrect repair by users, or any installation or use not in accordance with instructions provided by the manufacturer.
KAISE AUTHORIZED DEALER
422 Hayashinogo, Ueda City, Nagano Pref., 386-0156 Japan TEL : +81-268-35-1600 (REP.) / FAX : +81-268-35-1603 E-mail : sales@kaise.com http://www.kaise.com
Product specifications and appearance are subject to change without notice due to continual improvements.
70-1201-KF20-3 1610 - 4 -



References

iTextSharp™ 5.4.5 ©2000-2013 1T3XT BVBA (AGPL-version)