LEPU MEDICAL BP2/BP2A Automatic Upper Arm Blood Pressure Monitor

User Manual

1. The Basics

This manual contains the instructions necessary to operate the product safely and in accordance with its function and intended use. Observing this manual is a prerequisite for proper product performance and correct operations, which ensures patient and operator safety.

1.1 Safety

⚠️ Warnings and Tips

2. Introduction

2.1 Intended Use

The product is intended to be used for measuring, displaying, reviewing, and storing ECG and blood pressure variations in the home or healthcare facilities. ECG and Blood pressure variation is intended for use with adult. The data and results provided by this device are for precheck screening purposes only and cannot be directly used for diagnosis or treatments.

Product Features

Model Blood Pressure Measurement ECG Measurement
BP2
BP2A

2.2 Contraindications

2.3 About the Product

Device Components:
  • 1. Start/Stop button: Power On/Off; Press to Start/Stop measuring blood pressure.
  • 2. Function button: Press to start the ECG measurement; Press and hold for 2 seconds to review historical data.
  • 3. Display screen
  • 4. LED indicator: Blue light is on: the battery is charging; Blue light is off: the battery is fully charged.
  • 5. ECG electrodes: Press and hold them while taking ECG measurements.
  • 6. Cable connector: Connect with the charging cable.

2.4 Unpacking

Main Unit; Charging Cable; User Manual; Quick Guide

2.5 Symbols

Symbol Description
[factory] Manufacturer
[calendar] Date of manufacture
SN Serial number
[recycle-bin] Indicates a medical device that is not to be disposed of as unsorted municipal waste.
[book] Follow Instructions for Use.
[medical-device] Type BF Applied Part
[cross] No alarm system
[mr-unsafe] MRI unsafe. Presents hazards in all MR environments as device contains strongly ferromagnetic materials.
IP22 Resistant to liquid ingress
CE 0197 CE marking
EC REP Authorized representative in the European community
UK CA, UK REP UKCA marking
FC This product complies with the rules and regulations of the Federal Communication Commission.
[antenna] Non-ionizing radiation
[recycle] Our products and packaging can be recycled, don't throw them away! Find where to drop them off on the www.quefairedemesdechets.fr site (Only applicable for French market).
[heart] Heartbeat symbol
[battery] Battery symbol
[bluetooth] Bluetooth symbol
[upload] Data transmission symbol
[waveform] ECG waveform
[progress-bar] ECG measurement progress bar
NEXT Press to display next record
HOME Press to return to home screen
13/50 Current record / Total records

3. Using the Monitor

3.1 Charging the Battery

Use the USB cable to charge the monitor. Connect the USB cable to a USB charger or to the PC. It takes 2 hours to fully charge. When the battery is fully charged, the indicator will be off. The monitor works on very low power consumption, and a full charge usually lasts for months. A battery symbol, which indicates the battery status is displayed on the screen.

Note: The device cannot be used while charging.

3.2 Blood Pressure Measurement

3.2.1 Before Taking Measurements

To help ensure accurate measurements, follow these directions:

3.2.2 Applying the Arm Cuff

  1. Wrap the cuff around the upper arm, about 1 to 2 cm above the elbow joint, as shown.
  2. Place the cuff directly against the skin, as clothing may cause a faint pulse and result in measurement errors.
  3. Constriction of the upper arm, caused by rolling up a shirtsleeve, may prevent accurate readings.
  4. Confirm that the artery position mark is lined up with the artery.

Diagram Description: Shows the arm cuff placement, indicating the artery area and a measurement of 1-2cm (1/2in) above the elbow joint.

Note: Keep the host wear position aligned with the middle finger.

3.2.3 Sitting Correctly

To take measurements, you need to be relaxed and comfortably seated. Sit in a chair with your legs uncrossed and your feet flat on the floor. Place your arm on a table, so that the cuff is level with your heart.

3.2.4 Measuring Blood Pressure

  1. Power on the blood pressure monitor.
  2. Press the Start/Stop button to start taking blood pressure measurements.
  3. The monitor will automatically deflate the cuff slowly while taking measurements. A typical measurement takes about 30s.
  4. The readings will be displayed when the measurement is finished.

You can press Start/Stop button again to stop the blood pressure measurement.

Note: While taking measurements, you should remain still and not squeeze the cuff.

3.2.5 After Taking Measurements

The monitor will automatically release the gas in the cuff once finished taking measurements. Press the button to turn the power off after taking measurements. Remove the cuff.

Note: The device has an automatic power shut-off function, which turns off the power automatically 2 minutes after taking measurements.

3.3 Recording ECG

3.3.1 Before Taking Measurements

To help ensure accurate measurements, follow these directions:

3.3.2. Recording ECG without the Cable

There are four methods to record ECG without the cable.

ECG Recording Methods:
  • A. Right hand to left Leg (Lead II)
  • B. Hand to hand (Lead I)
  • C. Right hand to abdomen (Lead II)
  • D. Right hand to chest (Chest Lead)

To start an ECG Recording:

  1. Put your right palm on the right side electrodes of the monitor.
  2. Place the left side electrodes to the body position that you desired to measure.
  3. Once the body parts are placed on the electrodes, press the Function button to start EKG recording.
  4. Wait for 30 seconds, the result readings will be displayed.

To start recording ECG again, press the Start/Stop button to return to the Home screen, then press the Function button.

Note:

3.3.3. Turn on/off Heartbeat Sound

The buzzer beeps when a heartbeat is detected while recording ECG. You can turn on/off the heartbeat sound on the App.

3.4 Reviewing History Records

You can review the history results and replay the recorded ECG waveform on the History screen. Hold the Function button for 2 seconds to enter the History screen. The last measurement results will be displayed by default.

Example Record: 3:15 pm Mar 18, 2019, 100/120, ♥ 55 (Record 3/10)

For ECG recordings, the Heart Rate will be displayed first and then replayed over the 30-second ECG waveform.

Example ECG Record: 3:18 pm Mar 18, 2019, Regular ECG, ♥ 55 (Record 1/10)

To view the next records, press the NEXT button. To exit the History screen, press the HOME button.

3.5 Connecting with the App

App name: ViHealth

iOS: App Store

Android: Google Play

[App Store icon] [Google Play icon]

The monitor has built-in Bluetooth wireless connectivity, which enables exporting measurement records to mobile phones and tablets running iOS or Android. Hold the function button and Start/Stop button simultaneously to enable Bluetooth and start pairing it to your smart device.

4. Troubleshooting

Problem Possible Cause Solution
No power.
No display appears on the device.
Battery is depleted Recharge the battery
Blood pressure readings appear too high or too low Blood pressure varies constantly. Many factors including stress, time of day, and/or how you apply the arm cuff, may affect your blood pressure. Hold still for a moment and try again
Error 1: Apply the arm cuff tighter. Arm cuff is applied too loosely Tighten the arm cuff and try again.
Error 2: Do not move or talk, remain still. Moving or talking while taking measurements, and the pressure of the cuff is interfered with. Remain still and do not talk while taking measurements
Error 3: Remove any clothing interfering with the cuff. The signal is poor while measuring blood pressure. Take blood pressure on a bare arm.
Error x (x>4): Contact customer service. The device has malfunctioned. Contact customer service.
Check the connection The ECG cable is off, or the body parts are not placed on the electrodes closely enough. Check the ECG cable, or place the electrodes closer to your body part.
ECG waveform drifts The pressure exerted on the electrode is not stable or too much. Hand or body may be moving. Hold the device stably and gently. Try to keep perfectly still and test again.
The ECG waveform amplitude is small The measurement method you chose is not suitable for you. Change another lead and try again.

5. Maintenance

5.1 Maintenance

To protect your monitor from damage, store the monitor and the components in a clean, safe location.

Caution: DO NOT disassemble or attempt to repair this monitor or other components. This may cause inaccurate blood pressure readings and/or EKG recordings.

5.2 Cleaning

5.3 Storage

Keep your monitor and other components in the storage case when not in use.

6. Specifications

Classifications

EC Directive MDD, 93/42/EEC
RED, 2014/53/EU
ROHS 2.0, 2011/65/EU
Degree of protection against electrical shock Type BF

Environmental

Item Operating Storage
Temperature 5 to 40°C -25 to 70°C
Relative humidity (non-condensing) 10% to 95% 10% to 95%
Barometric 700 to 1060 hPa 700 to 1060 hPa
Degree of dust & water resistance IP22

Physical

Drop test 1.0 m
Size (main unit) 135mm(L)×45mm(W)×20mm(H)
Weight (main unit) 240 g
Cuff size Adult cuff: 22-42cm
Small adult cuff (optional): 17-22cm
Adult cuff (optional): 22-32cm
Wireless connectivity Built-in Bluetooth 4.0 BLE

Power Supply

Charge input USB Type-C, DC 5V
Battery type Rechargeable lithium-polymer battery
Battery run time 500 measurements
Charge time 2 hours

Blood Pressure Measurements

Technology Oscillometric Method
Pressure measurement range 0 – 300mmHg
Pressure measurement accuracy ±3mmHg
Pulse rate range 40 to 200 /min
Pulse rate accuracy ±2/min
Clinical accuracy Meet IEC80601-2-30

ECG Recording

Lead type Integrated ECG electrodes
Lead set Lead I, Lead II, Chest Lead
ECG length 30s

Storage

Blood pressure records 50
ECG records 10

Bluetooth RF

Frequency range 2.402 – 2.480 GHz
Max RF power -10 dBm

7. Electromagnetic Compatibility

The device meets the requirements of EN 60601-1-2.

⚠️ Warnings and Tips

Guidance and Declaration - Electromagnetic Emissions

The Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the Blood Pressure Monitor should assure that it is used in such an environment.

Emission tests Compliance Electromagnetic environment - guidance
RF emissions CISPR 11 Group 1 The Blood Pressure Monitor uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11 Class B The Blood Pressure Monitor is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions IEC61000-3-2 N/A
Voltage Fluctuations / Flicker Emissions IEC 61000-3-3 N/A

Guidance and Declaration - Electromagnetic Immunity

The Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the Blood Pressure Monitor should assure that it is used in such an environment.

Immunity test IEC60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic discharge (ESD) IEC 61000-4-2 ± 8kV contact
±2kV, ±4kV, ±8kV, ±15kV air
± 8kV contact
±2kV, ±4kV, ±8kV, ±15kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/ burst IEC 61000-4-4 ± 2kV for power supply lines
100kHz repetition frequency
± 1kV for input/output lines
N/A --
Surge IEC 61000-4-5 ± 0.5kV, ±1kV differential mode line-line N/A --
Voltage dips, short Interruptions and Voltage variations on power supply input lines IEC 61000-4-11 0% UT (100% dip in UT) for 0.5 cycle at 0°, 45°, 90°, 135°,180°, 225°, 270°, and 315°
0% UT (100% dip in UT) for 1 cycle at 0°
70% UT (30% dip in UT) for 25/30 cycles at 0°
0% UT (100% dip in UT) for 250/300 cycles at 0°
N/A --
Power frequency (50/60 HZ) magnetic field IEC 61000-4-8 30 A/m, 50/60Hz 30 A/m, 50/60Hz Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.

Note: UT is the AC mains voltage prior to application of the test level.

Guidance and Declaration - Electromagnetic Immunity

The Blood Pressure Monitor is intended for use in the specified electromagnetic environment. The customer or the user of the Blood Pressure Monitor should assure that it is used in such an environment as described below.

Immunity test IEC60601 test level Compliance level Electromagnetic environment - guidance
Conducted RF IEC61000 -4-6 3Vrms 150kHz to 80 MHz
6Vrms 150 kHz to 80 MHz outside ISM bands
N/A Portable and mobile RF communications equipment should be used no closer to any part of the system, including cables, than the recommended separation distance calculated from the equation appropriate for the frequency of the transmitter.
Radiated RF IEC61000 -4-3 10V/m 80MHz to 2.7 GHz 10 V/m Recommended separation distances:
d = [3.5 / √P] m for 80MHz to 800MHz
d = [3.5 / E1 √P] m for 80MHz to 800MHz
d = [7 / E1 √P] m for 800MHz to 2.7GHz
Where, P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey ª, should be less than the compliance level in each frequency range b. Interference may occur in the vicinity of equipment marked with the following symbol: [antenna-interference]

Note 1: At 80 MHz to 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

a The ISM (industrial, scientific and medical) bands between 0,15 MHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0 MHz.

b The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2,7 GHz are intended to decrease the likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas. For this reason, an additional factor of 10/3 has been incorporated into the formulae used in calculating the recommended separation distance for transmitters in these frequency ranges.

c Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Blood Pressure Monitor is used exceeds the applicable RF compliance level above, the Blood Pressure Monitor should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Blood Pressure Monitor.

d Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Recommended separation distances between portable and mobile RF communications equipment and the Blood Pressure Monitor

The Blood Pressure Monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Blood Pressure Monitor can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Blood Pressure Monitor as recommended below, according to the maximum output power of the communications equipment.

Rated max. output power of transmitter (W) Separation distance according to frequency of the transmitter (m)
150kHz to 80MHz 80MHz to 800MHz 800MHz to 2.7GHz
0.01 0.12 0.04 0.07
0.1 0.37 0.12 0.23
1 1.17 0.35 0.70
10 3.70 1.11 2.22
100 11.70 3.50 7.00

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

Models: BP2, BP2 Automatic Upper Arm Blood Pressure Monitor, Automatic Upper Arm Blood Pressure Monitor, Upper Arm Blood Pressure Monitor, Arm Blood Pressure Monitor, Blood Pressure Monitor, Pressure Monitor, Monitor

File Info : application/pdf, 27 Pages, 1.29MB

PDF preview unavailable. Download the PDF instead.

148392-bp2-upper-arm-blood-pressure-monitor-ecg-lepu-medical-user-manual

References

Related Documents

Preview Lepu Medical ER2-S ECG Recorder User Manual
Comprehensive user manual for the Lepu Medical ER2-S ECG Recorder, detailing its features, usage instructions, safety precautions, maintenance, and specifications. Learn how to accurately record and transfer ECG data for health monitoring.
Preview Shenzhen Lepu Intelligent Medical Equipment Co., Ltd. Fingertip Pulse Oximeter LOX100A/B/C/D User's Manual
This user manual provides comprehensive instructions for the Shenzhen Lepu Intelligent Medical Equipment Co., Ltd. Fingertip Pulse Oximeter, models LOX100A, LOX100B, LOX100C, and LOX100D. It covers product overview, intended use, principles of measurement, warnings, symbol descriptions, battery installation, operating instructions, settings, lanyard installation, product accessories, maintenance, storage, transportation, technical specifications, and electromagnetic compatibility.
Preview SARS-CoV-2 Antigen Rapid Test Kit (Colloidal Gold Immunochromatography) - Lepu Medical
Instructions for use and technical specifications for the Lepu Medical SARS-CoV-2 Antigen Rapid Test Kit, a qualitative immunochromatographic assay for detecting SARS-CoV-2 antigens in nasal swabs. Includes product details, test method, results interpretation, performance data, and precautions.
Preview LeECG & NeoECG Series Electrocardiograph Operator's Manual
This operator's manual provides comprehensive instructions for the LeECG & NeoECG Series Electrocardiograph by Lepu Medical. It covers safety guidance, product introduction, operation preparations, screen navigation, patient information entry, acquisition, analysis, printing, file management, system settings, troubleshooting, and maintenance.