OperInstr WaterAnalysisDevice PCE-WO2 10 en
Englis h

User Manual
PCE-WO2 10 Oxygen Meter
User manuals in various languages (français, italiano, español, português, nederlands, türk, polski,
, ) can be found by using our product search on: www.pce-instruments.com
Last change: 8 August 2019 v1.0
© PCE Instruments

Contents 1 Safety notes ........................................................................................... 1 2 Technical specifications ...................................................................... 2 3 Delivery Scope ...................................................................................... 2 4 Installing the Batteries ......................................................................... 3 5 Display ................................................................................................... 3 6 Keypad ................................................................................................... 4 7 Filling the Electrolyte Solution ............................................................ 5 8 Connecting the Dissolved Oxygen Probe .......................................... 5 9 Switching the Tester On and Off ......................................................... 6 10 Prior to Use ............................................................................................ 6 11 Setup Menu............................................................................................ 6
11.1 Setting the default option.................................................................................................. 7
12 Setting the salinity coefficient ............................................................. 8 13 Setting the ambient pressure .............................................................. 8 14 DO Calibration in % Saturation Mode ................................................. 9
14.1 Single point calibration - 100% saturation......................................................................... 9 14.2 2 points calibration ........................................................................................................... 9
15 Temperature Calibration .................................................................... 10 16 Dissolved Oxygen Measurement....................................................... 11 17 Auto-Hold ............................................................................................. 11 18 Electrode Care and Maintenance ...................................................... 12 19 Troubleshooting.................................................................................. 12 20 Addendum: Preparation of the Zero Oxygen Solution.................... 12 21 Contact............................................................................................... 13 22 Disposal ............................................................................................... 13
© PCE Instruments

1 Safety notes
Please read this manual carefully and completely before you use the device for the first time. The device may only be used by qualified personnel and repaired by PCE Instruments personnel. Damage or injuries caused by non-observance of the manual are excluded from our liability and not covered by our warranty.
· The device must only be used as described in this instruction manual. If used otherwise, this can cause dangerous situations for the user and damage to the meter.
· The instrument may only be used if the environmental conditions (temperature, relative humidity, ...) are within the ranges stated in the technical specifications. Do not expose the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
· Do not expose the device to shocks or strong vibrations. · The case should only be opened by qualified PCE Instruments personnel. · Never use the instrument when your hands are wet. · You must not make any technical changes to the device. · The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner,
no abrasives or solvents. · The device must only be used with accessories from PCE Instruments or equivalent. · Before each use, inspect the case for visible damage. If any damage is visible, do not
use the device. · Do not use the instrument in explosive atmospheres. · The measurement range as stated in the specifications must not be exceeded under
any circumstances. · Non-observance of the safety notes can cause damage to the device and injuries to
the user. We do not assume liability for printing errors or any other mistakes in this manual. We expressly point to our general guarantee terms which can be found in our general terms of business. If you have any questions please contact PCE Instruments. The contact details can be found at the end of this manual.
© PCE Instruments 1

2 Technical specifications
Measuring range oxygen content Resolution oxygen content Measuring accuracy oxygen content Measuring range oxygen saturation Resolution oxygen saturation Measuring accuracy of oxygen saturation Temperature compensation Air pressure correction Salinity correction Sensor type Environmental conditions Power supply Dimension (length x radius) Weight
3 Delivery Scope
1 x Dissolved Oxygen Meter PCE-WO2 10 1 x Oxygen probe with measuring cable 2 x Membrane 1 x Electrolyte solution (30ml) 2 x 1.5 V AAA batteries 1 x User manual 1 x Carrying case

0.0 ... 20.0 mg/l 0.1 mg/l ±0.5 mg/l 0.0 ... 200.0 % 0.1 % ±2.0 % 0 ... +40 °C / 32 ... 104 °F, automatic 60.0 ... 112.5 kPa, 450 ...850 mmHg 0 ... 35 g/l polarography 0 ... +60 °C / 32 ... 140 °F, max. 80 % r.H. 2 x 1.5 AAA batteries 175 x 40 mm / 6.8 x 1.5" 100 g / 3.5 oz

© PCE Instruments 2

4 Installing the Batteries

·

Twist the electrode collar counter clockwise, pull the 6-pin connector away from the

tester.

·

Insert two AAA batteries into the battery compartment, note polarity.

·

Align the slot on connector, gently push the connector into the tester.

·

Twist the electrode collar clockwise until it is tight.

5 Display
PCE-WO2 10 pocket dissolved oxygen tester is equipped with an easy-to-read LCD display that used to show the measured values and mode icons. The following table describes the function of each icon.

INDEX:
ICON MEAS CAL SETUP HOLD ATC

DESCRIPTION Indicates the tester is in the measurement mode. Indicates the tester is in the calibration mode. Indicates the tester is in the setting mode. Indicates the measuring value has been locked. Indicates the temperature compensation is enabled. When the battery is depleted, the icon will disappear.
© PCE Instruments 3

6 Keypad
The tester has a succinct membrane keypad, names and symbols describe each function key controls.

INDEX: KEY MEAS/HOLD
CAL ENTER

FUNCTION
· Switches the tester ON/OFF. · Locks the measured value, press the key again to resume measuring. · Exits the calibration or setting and returns to the measurement mode.
· Press the key to start the calibration. · Press and hold the key to enter the setup menu. · In the setting mode, press the key to select the default option.
· Confirms the calibration or selected option.

© PCE Instruments 4

7 Filling the Electrolyte Solution
· Take out the dissolved oxygen probe and electrolyte solution from the packaging. Unscrew the membrane cap.
· Fill the membrane cap halfway with electrolyte solution. · Screw the membrane cap onto the probe, excess electrolyte solution will drain out. · Be sure the cathode of probe makes contact with membrane cap, the electrolyte solution in
membrane cap should be without an air bubble.
Cathode
8 Connecting the Dissolved Oxygen Probe
Insert the 6-pin connector of probe into the connector socket on the tester. After the connection is completed, DO NOT pull on the cable. Always make sure that the connector is clean and dry.
© PCE Instruments 5

9 Switching the Tester On and Off

·

Press the Meas key to switch on the tester, the display shows measured values.

·

Press and hold the Meas key for 5 seconds, the tester will switch off.

·

To enable the Auto-Power Off function, please refer to chapter SETUP MENU.

Power On

5 sec. Power Off

10 Prior to Use
Switch on the tester 10 to 15 minutes and wait for the probe to polarize.

11 Setup Menu
PCE-WO2 10 pocket dissolved oxygen tester contains an integrated setup menu that is used to customize the displayed option to meet measurement requirements.

MENU

DESCRIPTION
Set the salinity coefficient of sample. Set the barometric pressure coefficient.
Set the number of calibration points.

OPTIONS

Range: 0~35ppt Range: 450~850mmHg 1 point
2 points

mg/L
ppm Set the default measurement % unit.
Calibrate the temperature. When the option is enabled, the tester will automatically sense a stable reading and lock the measurements. When the option is enabled, the tester will automatically switch off if no key is pressed within 8 minutes.
Reset the tester to factory default settings.

Concentration Unit
% saturation Degrees Celsius Degrees Fahrenheit --Enable Disable Enable
Disable Enable Disable

DEFAULT 0.0ppt 760mmHg 1 point
mg/L, °C
--Disable Disable Disable

© PCE Instruments 6

11.1 Setting the default option · Press and hold the Cal key for 3 seconds to enter the setup menu, the tester shows a menu item and default option. · If necessary, press the Cal key again to select the desired option. · Press the Enter key to confirm, the tester moves to the next menu item. · Repeat the steps above until the tester returns to the measurement mode. Setting is completed.
Option


·
·
·

During the setting process, press the Meas key, the tester will exit the setup menu and return to the measurement mode. If you do not want to calibrate the temperature, press the Enter key to skip the °C/CAL or °F/CAL option. The Reset function will restore the tester back to factory default settings. If enabled, all of the calibration data and selected parameters will be lost or reset, the tester must be recalibrated.

© PCE Instruments 7

12 Setting the salinity coefficient

In order to get an accurate reading, the salinity of the sample should be specified which is done

with the help of a coefficient.

Salt dissolved in water will influences oxygen content of water. If your sample contains high levels

of salinity, ensure that setting an applicable salinity

coefficient prior to measurement.

· In the measurement mode, press and hold the Cal key for 3 seconds to enter the

setup menu, the display shows

(salinity coefficient) option.

· Press the Cal key again to set the value.

· Press the Enter key to confirm, the display shows the next menu item.

· Press the Meas key to exit the setting and return to the measurement mode.

13 Setting the ambient pressure
To ensure an accurate reading, the ambient pressure must be set in the unit mmHg. The following table will help to convert your ambient pressure into the required unit mmHg on the basis of your current height.

ALTITUDE (m)

kPa

mmHg

ALTITUDE (m)

kPa

mmHg

0

101.3

760

1600

82.9 622

100

100.1

750

1700

81.9 614

200

98.8

741

1800

80.9 607

300

97.6

732

1900

79.9 599

400

96.4

723

2000

78.9 592

500

95.2

714

2100

77.9 584

600

94.0

705

2200

76.9 577

700

92.8

696

2300

76.0 570

800

91.7

688

2400

75.0 563

900

90.5

679

2500

74.1 556

1000

89.4

671

2600

73.2 549

1100

88.3

662

2700

72.3 542

1200

87.2

654

2800

71.4 536

1300

86.1

646

2900

70.5 529

1400

85.0

638

3000

69.6 522

1500 ·
·
· · ·

84.0

630

3100

68.7 515

In the measurement mode, press and hold the Cal key for 3 seconds to enter the

setup menu, the display shows

.

Press the Enter key, the display shows

(Barometric Pressure Coefficient)

option.

Press the Cal key to set the value.

Press the Enter key to confirm, the display shows the next menu item.

Press the Meas key to exit the setting and return to the measurement mode.

© PCE Instruments 8

14 DO Calibration in % Saturation Mode
PCE-WO2 10 pocket dissolved oxygen tester can be calibrated quickly in air. In the percentage saturation mode, the tester is able to perform either 1 or 2 points calibration. For single point calibration, we recommend that you perform 100% saturation calibration in the air-saturated water. If the 2 points calibration is selected, the zero oxygen solution needs to be used.

14.1 Single point calibration - 100% saturation
· Make sure that you have selected 1 point calibration in the setup menu.
· Press the Cal key, the tester shows "100% / CAL1".
· Hold the dissolved oxygen probe in the air at 100% relative humidity or place the probe into the air-saturated water for 15 minutes.
· Press the Enter key. Wait for the reading to stabilize, the tester automatically shows END. Single point calibration is completed.

14.2 2 points calibration

· Make sure that you have selected 2 points calibration in the setup menu.
· Repeat the steps 1.2 to 1.4 above. When the first calibration point is completed, the display will show "0% / CAL2". The tester prompts you to continue with second point calibration.
· Immerse the dissolved oxygen probe into the zero oxygen solution, stir the probe gently.
· Press the Enter key to confirm. Wait for the reading to stabilize (this can often take more than 10 minutes), the display automatically shows END. The tester returns to the measurement mode. Calibration is completed.
 Performing a percentage saturation
calibration will simultaneously calibrate the corresponding mg/L (or ppm) concentration value.

© PCE Instruments 9

15 Temperature Calibration
During the measurement process, if the temperature reading displayed differs from that of an accurate thermometer, the tester needs to be calibrated.
· Press and hold the Cal key for 3 seconds to enter the setup menu.
· Press the Enter key until the tester shows °C/CAL or °F/CAL.
· Press the Cal key to enter the temperature calibration mode.
· Press the Cal key again to modify the temperature value (Resolution: 0.5°C).
· Press the Enter key to confirm, the tester shows the next menu item.
· Press the Meas key to return to the measurement mode. Calibration is completed.
© PCE Instruments 10

16 Dissolved Oxygen Measurement
PCE-WO2 10 pocket dissolved oxygen tester is suitable for measuring the water, wastewater, brine and other liquids. If the sample is belong to the seawater or other water containing large amounts of salt, please setting the salinity coefficient before measurement. Some gas and steam such as chloride, sulfur dioxide, sulfureted hydrogen, ammonium, carbon dioxide and iodin can permeate the membrane via diffusion. So their existence will influence the measurement of dissolved oxygen. If the sample contains the solvent, grease, sulfide and alga, the membrane on the probe will be blocked, damaged or eroded.
· Connect the dissolved oxygen probe to tester and wait for 15 minutes to polarize the probe.
· If necessary, to set the barometric pressure and salinity coefficient in the setup menu (Refer to page 5).
· Immerse the probe in the sample solution, make sure the temperature sensor on the probe is fully immersed.
· Stir the probe gently. Record the measured value when the reading is stable.
Temperature sensor

17 Auto-Hold

PCE-WO2 10 pocket dissolved oxygen tester contains an Auto-Hold function. If enabled, the

tester will automatically sense a stable reading and lock the

measurements. The

icon appears on the display. If disabled, press the Hold key, the tester

will immediately lock the displayed value. Press the Meas key to resume measuring.

© PCE Instruments 11

18 Electrode Care and Maintenance

·

Always keep the membrane of the dissolved oxygen probe is wet or moist.

·

If you do not use the probe for long periods, please screw off membrane cap and rinse

the cathode, anode and membrane with deionized water, then soak up residual water

on them with filter paper. Install the probe again.

Membrane cap Cathode and Anode

19 Troubleshooting

LCD DISPLAY

CAUSE
DO probe does not connect to tester Measured value is out of range Electrolyte solution is depleted Zero oxygen solution is contaminated

CORRECTIVE ACTION
Check the connector of probe. Check the DO membrane whether clogged, dirty or broken. Refilling electrolyte solution.
Replace the calibration solution.

20 Addendum: Preparation of the Zero Oxygen Solution
Dissolve 500mg of sodium sulfate (Na2SO3) reagent and a small amount of cobalt (II) chloride hexahydrate (CoCl2 · 6H2O) in the 250mL distilled water, mix the solution until the reagent is completely dissolved.

© PCE Instruments 12

21 Contact
If you have any questions, suggestions or technical problems, please do not hesitate to contact us. You will find the relevant contact information at the end of this user manual.
22 Disposal
For the disposal of batteries in the EU, the 2006/66/EC directive of the European Parliament applies. Due to the contained pollutants, batteries must not be disposed of as household waste. They must be given to collection points designed for that purpose. In order to comply with the EU directive 2012/19/EU we take our devices back. We either reuse them or give them to a recycling company which disposes of the devices in line with law. For countries outside the EU, batteries and devices should be disposed of in accordance with your local waste regulations. If you have any questions, please contact PCE Instruments.
© PCE Instruments 13

PCE Instruments contact information

Germany
PCE Deutschland GmbH Im Langel 4 D-59872 Meschede Deutschland Tel.: +49 (0) 2903 976 99 0 Fax: +49 (0) 2903 976 99 29 info@pce-instruments.com www.pce-instruments.com/deutsch
Germany
PCE Produktions- und Entwicklungsgesellschaft mbH Im Langel 26 D-59872 Meschede Deutschland Tel.: +49 (0) 2903 976 99 471 Fax: +49 (0) 2903 976 99 9971 info@pce-instruments.com www.pce-instruments.com/deutsch
The Netherlands
PCE Brookhuis B.V. Institutenweg 15 7521 PH Enschede Nederland Telefoon: +31 (0)53 737 01 92 info@pcebenelux.nl www.pce-instruments.com/dutch
United States of America
PCE Americas Inc. 711 Commerce Way suite 8 Jupiter / Palm Beach 33458 FL USA Tel: +1 (561) 320-9162 Fax: +1 (561) 320-9176 info@pce-americas.com www.pce-instruments.com/us

France
PCE Instruments France EURL 23, rue de Strasbourg 67250 Soultz-Sous-Forets France Téléphone: +33 (0) 972 3537 17 Numéro de fax: +33 (0) 972 3537 18 info@pce-france.fr www.pce-instruments.com/french
United Kingdom
PCE Instruments UK Ltd Unit 11 Southpoint Business Park Ensign Way, Southampton Hampshire United Kingdom, SO31 4RF Tel: +44 (0) 2380 98703 0 Fax: +44 (0) 2380 98703 9 info@industrial-needs.com www.pce-instruments.com/english
China
PCE (Beijing) Technology Co., Limited 1519 Room, 6 Building Zhong Ang Times Plaza No. 9 Mentougou Road, Tou Gou District 102300 Beijing, China Tel: +86 (10) 8893 9660 info@pce-instruments.cn www.pce-instruments.cn
Turkey
PCE Teknik Cihazlari Ltd.ti. Halkali Merkez Mah. Pehlivan Sok. No.6/C 34303 Küçükçekmece - stanbul Türkiye Tel: 0212 471 11 47 Faks: 0212 705 53 93 info@pce-cihazlari.com.tr www.pce-instruments.com/turkish

Spain
PCE Ibérica S.L. Calle Mayor, 53 02500 Tobarra (Albacete) España Tel. : +34 967 543 548 Fax: +34 967 543 542 info@pce-iberica.es www.pce-instruments.com/espanol
Italy
PCE Italia s.r.l. Via Pesciatina 878 / B-Interno 6 55010 Loc. Gragnano Capannori (Lucca) Italia Telefono: +39 0583 975 114 Fax: +39 0583 974 824 info@pce-italia.it www.pce-instruments.com/italiano
Hong Kong
PCE Instruments HK Ltd. Unit J, 21/F., COS Centre 56 Tsun Yip Street Kwun Tong Kowloon, Hong Kong Tel: +852-301-84912 jyi@pce-instruments.com www.pce-instruments.cn

© PCE Instruments 14



References

Microsoft Word 2016