Operation Manual eyc-tech THS88MAX
Industrial Dew Point Transmitter
eyc-tech THS88MAX

Dew Point Industrial Dew Point Transmitter
Contents
1. Security considerations ...................................................... P.02 2. Dimension ........................................................................... P.03 3. Diagram ............................................................................... P.03 4. Accessories for measuring atmospheric dew point........... P.04 5. Software and calibration operation step ........................... P.05
5.1 Application Program statement ................................ P.05 5.2 Setting RS-485 connection ........................................ P.05 5.3 Scan RS-485 connection............................................ P.07 5.4 Setting RS-485 ModBus Protocol .............................. P.09 5.5 Display and save data ................................................ P.11 5.6 Choose parameter of Output .................................... P.14 5.7 Convert 4-20mA to 0-10V ........................................... P.16 6. Inspection and maintenance ............................................. P.17
www.eyc-tech.com
1

1. Security considerations

Dew Point Industrial Dew Point Transmitter

Please read this Specification carefully, prior to use of this, and keep the manual properly, for timely reference. Solemn Statement This product can not be used for any explosion-proof area. Do not use this product in a situation where human life may be affected. eyc-tech will not bear any responsibility for the results produced by the operators
Warning!
 Installation and wiring must be performed by qualified personnel in accordance with all applicable safety standards.
 This product must be operated under the operating conditions specified in manual to prevent equipment damages.
 Please using the product under the ordinary pressure, or it will influence safe problem.  This product must be operated under the operating condition specified in this manual
to prevent equipment damages.  This product must be operated under the normally atmospheric condition to prevent
equipment damages.  To prevent products damage, always disconnect the power supply from the product
before performing any wiring and installation.  All wiring must comply with local codes of indoor wiring and electrical installation
rules.  Please use crimp type terminal.  To prevent personal injury, do not touch the moving part of product in operation.  It may cause high humidity atmosphere during the product was breakdown. Please
take safety strategy.

www.eyc-tech.com
2

2. Dimension

Dew Point Industrial Dew Point Transmitter

 Standard  M12-5PIN 2M waterproof cable
3. Diagram
M12 Connector
*Please make sure the product and the device which connect with RS-485 are on common ground, avoid damaged product.
www.eyc-tech.com
3

Dew Point Industrial Dew Point Transmitter
4. Accessories for measuring atmospheric dew point

· Environment pressure affects the value of dew point. When environment pressure is higher, water vapor condensate easier. Thus, dew point value in high pressure environment is higher than atmospheric dew point.
· In process line, pressure could change and fluctuate easily, and leads to inconstant reading value. To avoid this problem, eyc-tech suggest user to measure dew point at atmospheric pressure (atmospheric dew point).
· To help user measuring atmospheric dew point easily, eyc-tech provide related accessories, which including installation base for dew point transmitter, pressure regulator and flowmeter ...etc. User can select accessories depends on installation. Please ask our sales personnel for further information.

Part number

Description

BASE-THS-001 Installation base(1/2"PT for transmitter), SUS304, Connection for air inlet and outlet: 1/4"PT

BASE-THS-002 Installation base(1/2"PF for transmitter), SUS304, Connection for air inlet and outlet: 1/4"PT

BASE-THS-003 Installation base(5/8"UNF for transmitter), SUS304, Connection for air inlet and outlet: 1/4"PT

BASE-THS-002-1 Installation base(1/2"PF for transmitter), SUS304, Connection for air inlet : 1/4"PF, regulator included.

· A schematic of installation is shown below. Using pressure regulator to regulate pressure to 1 atm, and adjust airflow to 1~5 LPM to have stable reading of atmospheric pressure.

www.eyc-tech.com
4

Dew Point Industrial Dew Point Transmitter
5. Software and calibration operation step
5.1 Application Program statement
User may download the configuration software on eyc-tech web site. Please decompress the application prior to execute it. Operating System requirements above Windows XP. Other application program requirementsabove Microsoft Office 2003
5.2 Setting RS-485 connection
1. Connect product to PC via RS-485 cable 2. Execute "THS UI" 3. Click "Interface > Config"
www.eyc-tech.com
5

Dew Point

Industrial Dew Point Transmitter

4. Select the corresponding values of com port as

fallowing :

a.

a. PortCheck Come Port

b.

b. Baud Rate

c.

c. Data Frame

d.

d. Timeout

e.

e. Retry

f.

f. Station ID(Default 1)

5. Click "Apply" 6. Connect successfully a. Show value and trend chart of Dew point temperature b. Show "Open Port, Read successful"
a.

a. b.
www.eyc-tech.com
6

Dew Point Industrial Dew Point Transmitter
5.3 Scan RS-485 connection
Use scan function to connect when forgetting the connection information or having more facilities . 1. Connect the product to PC via RS-485 cable 2. Execute "THS UI" 3. Click "Interface > Config"
4. Select the corresponding values of com port as fallowing: a. Port b. RS-485
5. Click "Scan" to execute connection facilities
www.eyc-tech.com
7

6. Scan connection facilities and set up a. Select Station ID b. Click "CLOSE AND EXPORT"

Dew Point Industrial Dew Point Transmitter

a.
b.
7. Click "Apply"

8. Connect successfully

www.eyc-tech.com
8

Dew Point Industrial Dew Point Transmitter
a. Show values and trend chat Dew point Temperature b. Show "Open port, Read successful"
a.
a. b.
5.4 Setting RS-485 ModBus Protocol
1. Setting RS-485 connection step as step 5.1 2. Click "Setting"
3. Select Modbus Protocol parameter a. Station ID1~247
www.eyc-tech.com
9

Dew Point Industrial Dew Point Transmitter
b. Baud Rate9600, 19200, 38400, 57600, 115200 c. Data FrameNone-8Bit-1Stop, None-8Bit-2Stop, Even-8Bit-1Stop, Even-8Bit-2Stop,
Odd-8Bit-1Stop, Odd-8Bit-1Stop
a. b. c.
4. Click "Apply" 5. Execute connection as step 5.2 or 5.3 again
www.eyc-tech.com
10

5.5 Display and save data
1. Show DataClick "Display"

Dew Point Industrial Dew Point Transmitter

2. Icon function statements
Clear the chart records
Change the chart style Select the OUTPUT channel Set line color chosen from OUTPUT Snap chart Show/Not show measuring data
Adjust time range of X axis
Adjust time range of Y axis
3. Setting time interval of record a. File > Log Interval b. Select time interval of record

Save the data measuring when the system start connecting before clinking the Export icon Show/Not show the statement of Y axis Show/Not show the statement of X axis Show/Not show the statement of Y secondary axis Show/ Not show chart

www.eyc-tech.com
11

Dew Point Industrial Dew Point Transmitter
4. Save/Log measuring data a. Log measuring rangeSave the data measuring when the system start connecting before clinking the Export icon a-1. Click Display > Export
a-2. Appoint path and Key in file name > save 1. If file name is some as the path name, the original file will be over write.
www.eyc-tech.com
12

Dew Point Industrial Dew Point Transmitter
b. Log measuring dataLog the data which is on from start or off b-1. Display > Log(OFF) b-2. Appoint path and Key in file name > save > Log (ON) 1. If file name is some as the path name, the original file will be over write.
www.eyc-tech.com
13

Dew Point Industrial Dew Point Transmitter
5.6 Choose parameter of Output
1.Click "Output"

2. Select relative parameters of Output1 a. Measures b. Responding Rate c. Voltage or current analog range d. Upper and Lower point of Output
a. b. c. d.

3. Click "Apply"

www.eyc-tech.com
14

4. Select relative parameters of Output2 a. Measures b. Responding Rate c. Upper and Lower point of Output

Dew Point Industrial Dew Point Transmitter

a. b.
c.

5. Click "Apply"

www.eyc-tech.com
15

Dew Point Industrial Dew Point Transmitter
5.7 Convert 4-20mA to 0-10V
1. Select Current to 0-20mA in UI.
2. Please Connect 500 ohms in parallel at the output.
500 ohms
M12 Connector
www.eyc-tech.com
16

Dew Point Industrial Dew Point Transmitter
6. Inspection and maintenance

1. Maintenance

Since this product is inspected and calibrated for high accuracy at the factory

before shipment, no calibration on the installation site is necessary when this product is

installed. For inspection and maintenance follow the instructions below

 Periodic inspection

Periodically inspect this product for its sensing accuracy, and clean the cover. Set

the period between inspections based on atmospheric dust and other contaminants in

the installation environment.

2. Troubleshooting

 Sensor maintenance

Do not damage sensor surface during the maintenance process.

 Troubleshooting

If any problem occurs during operation, refer to the table below for appropriate

solutions.

Problem  No output  Unstable output
 Slow response to output
 Errow in output

Cleck items
 Disconnected wiring  Loose wiring  Power supply voltage  Sensor damages  Moisture /condensation on
the product  Check installed location  Check installed angle  Check dust and
contamination on the sensor

Soluations
 Re-perform wiring  Crew on terminal tightly or replace
wires  Replace the sensor  Remove the sensor and filter. Dry
power-off state sensor in clean air
seasoning  Refer to the section  Align measurement head with flow
direction  Cleaning the filter  Changing the filter  Calibrate  Replace the sensor

www.eyc-tech.com
17