Owner's Manual for MTTS models including: 4METER Lightmeter, 4METER, Lightmeter

LM-MKT-01-4meter-4.01EN

User Manual

Support | LifeKit by MTTS

Brochure

Lightmeter / 4meter | LifeKit by MTTS


File Info : application/pdf, 2 Pages, 1.56MB

PDF preview unavailable. Download the PDF instead.

LM-MKT-01-4meter-4.01EN
meter

by

4METER
An instrument for determination of the actual level of sources required for an effective medical therapy.

Key Functions
1. Blue light irradiance measurement: - verify irradiance of photo therapy devices during preventive maintenance and normal use 2. O2 measurement: - verify purity of O2 sources (tank, generator, wall) during preventive maintenance - verify mechanical blender and other blender during preventive maintenance - leak search inside device and along external tubes during repair and equipment inspection 3. Pressure measurement: - verify pressure at Y connector during use 4. Flow measurement: - verify flow output of CPAP, Hi-flow and ventilators during preventive maintenance 5. High pressure measurement: - verify hospital gas sources and check them for fluctuations 6. Temperature and humidity measurement: - verify the temperature uniformity of the radiant warmers - check temperature and humidity inside the incubators
Light Irradiance Sensor
- spectral response chosen to match wavelength used for jaundice treatment - two sided measurement for convenient measurement of double sided
phototherapy devices such as the MTTS Firefly
Displays/Indicators
- touch screen liquid-crystal display (LCD) - alarms: low battery, system failure, out of measurement - results `freeze' - bandwidth estimator - absolute and relative irradiance measurement - multi-point mode

Light Sensor Spectral Response

100
75
%
50
25

390 400 410 420 430 440 440 450 460 470 480 490 500 510 520 530
Wavelength

Dimensions and Weight

Ordering Information and Servicing

Height Width Depth Weight

148 mm 73 mm 17 mm 200 g

Performance Specifications

Application
LIGHT SENSOR Spectral response Measurement range
Resolution Cosine characteristics
Accuracy OXYGEN SENSOR FLOW SENSOR MANOMETER HIGH PRESSURE SEN. TEMP/HUMIDITY SEN.

- LED and blue fluorescent phototherapy light sources; using the lightmeter for halogen phototherapy results in a minor but not considerable error - Various gas sources used for respiratory therapy

400-520nm

single side 0.1 ­ 150.0 µW/cm²/nm

double side 0.1 ­ 300.0 µW/cm²/nm

0.1 µW/cm²/nm

±2% @ 30 degree angle

±7% @ 60 degree angle

±25% @ 80 degree angle

+/- 3% of reading

FiO2

21-100%

flow

0-120 lpm, bi-directional

pressure pressure

0-100 cm H2O ±100 psi

temperature 5-60oC

humidity 0-100% RH

Battery Specification

ORDERING INFO
Lightmeter Oxygen meter Flow meter Manometer High pressure meter Temperature/Humidity meter Rechargeable batteries and charger Storage case

ITEM ORDER NO
LM-800 LM-800-020 LM-800-030 LM-800-040 LM-800-050 LM-800-060 LM-800-070

Main Unit

LM-800-080

Calibration Parts

Required

Every 2 years after delivery date; return to MTTS for calibration
No customer serviceable parts; return to MTTS for service

MEDICAL TECHNOLOGY TRANSFER AND SERVICES Co., LTD ISO 13485: 2016 Certified Company

For more information or to place an order please contact us at:

info@mtts-asia.com

https://mtts-asia.com

Type Lifetime

2xAAA 1.5V Alkaline, Non-rechargeable Up to 120h operating time

Environmental Specification

+84 24 37 666 521

No 26, Lane 41 An Duong Vuong Street Phu Thuong Ward, Tay Ho District Hanoi, Vietnam

Operation Temperature Storage Temperature Operation Atmospheric Pressure Storage Atmospheric Pressure Operation Humidity Storage Humidity Sealing Drop/Shock
Standards

10°C to 40°C 0°C to 50°C 70 ­ 106 kPA 70 ­ 106 kPA 30 ­ 90% RH non condensing 5 ­ 90% RH non condensing IP-54 1m onto hard surface, meets or exceeds IEC 60068-2-32 (only with the protective case)

Distributor:

ISO 13485:2016 IEC 61010-1:2010 IEC 61326-1:2012

LM-MKT-01-Rev-4.01EN



References

Adobe PDF library 9.90 Adobe Illustrator(R) 15.0