Ultrasonic Cleaner Instruction Manual
Product Overview
This manual provides instructions for the VEVOR Ultrasonic Cleaner, covering both Mechanical and Digital models.
Applications: Suitable for cleaning a wide range of items including PCB, jewelry, electronic components, car parts, hardware, metal tools, computer parts, laboratory instruments, clock & watches, eyeglasses, contact lens, dentures, dental tools, DVD & VCD, golf, table-ware, printer inkjet, seals, combs, toothbrushes, and coins.
Industries Served: Semiconductor, Electrical & Electronic Components, Optical Device, Hardware & Machinery Parts, Medical Instrument, Electroplate, Car Parts, Chemical Fiber.
Models and Specifications
Mechanical Models
Model | Frequency (KHz) | Transducer (PCS) | Capacity (L) | Drain Valve | Tank Size (mm) | Unit Size (mm) | Heater Power (W) | Ultrasonic Timer (MIN) | Heater (°C) | |
---|---|---|---|---|---|---|---|---|---|---|
C10 | 40 | 1 | 2 | NO | 150x135x100 | 175x160x210 | 60 | 100 | 0~30 | 20~80 |
C20 | 40 | 2 | 3.2 | NO | 240x135x100 | 265x165x220 | 100 | 120 | 0~30 | 20~80 |
C30 | 40 | 3 | 4.5 | NO | 300x150x100 | 325x180x225 | 200 | 180 | 0~30 | 20~80 |
C31 | 40 | 3 | 6.5 | YES | 300x150x150 | 325x180x280 | 200 | 180 | 0~30 | 20~80 |
C40 | 40 | 4 | 10 | YES | 300x240x150 | 325x265x280 | 200 | 240 | 0~30 | 20~80 |
C60 | 40 | 6 | 15 | YES | 330x300x150 | 360x325x285 | 300 | 360 | 0~30 | 20~80 |
C80 | 40 | 8 | 22 | YES | 500x300x150 | 530x325x285 | 500 | 480 | 0~30 | 20~80 |
C100 | 40 | 10 | 30 | YES | 500x300x200 | 530x325x325 | 500 | 600 | 0~30 | 20~80 |
Digital Models (Product Specifications)
Model | Frequency (KHz) | Transducer (PCS) | Capacity (L) | Drain Valve | Tank Size (mm) | Unit Size (mm) | Heating Power (W) | Ultrasonic Power (W) | Time (M) | Temperature (°C) |
---|---|---|---|---|---|---|---|---|---|---|
JPS-08A | 40 | 1 | 1.3 | NO | 150x135x65 | 175x160x175 | 100 | 60 | 0~30 | room temp~80 |
JPS-10A | 40 | 1 | 2 | NO | 150x135x100 | 175x160x210 | 100 | 60 | 0~30 | room temp~80 |
JPS-20A | 40 | 2 | 3.2 | NO | 240x135x100 | 265x165x220 | 1 | 120 | 0~30 | room temp~80 |
JPS-30A | 40 | 3 | 6.5 | YES | 300x150x150 | 325x180x280 | 200 | 180 | 0~30 | room temp~80 |
JPS-40A | 40 | 4 | 10 | YES | 300x240x150 | 325x265x280 | 200 | 240 | 0~30 | room temp~80 |
JPS-60A | 40 | 6 | 15 | YES | 330x300x150 | 360x325x285 | 300 | 360 | 0~30 | room temp~80 |
JPS-80A | 40 | 8 | 22 | YES | 500x300x150 | 530x325x285 | 500 | 480 | 0~30 | room temp~80 |
JPS-100A | 40 | 10 | 33 | YES | 500x300x200 | 530x325x325 | 500 | 600 | 0~30 | room temp~80 |
Cautions and Safety
- Read the manual carefully before operation to avoid damage or personal danger. Keep for reference.
- Ensure the power supply is within the rated range before connecting the power cable. Refitting is strictly prohibited. The control panel may be eroded by organic solution, strong acid, and strong alkali.
- Ensure the earth wiring is well connected before starting.
- Do not operate if the tank is empty, or the ultrasonic generator will be damaged. If heating is needed, the water level should not be less than 2/3.
- Close the lid to reduce noise and pay attention to water and steam in case of burning skin when opening the lid.
- Do not relocate the machine when fluid is in the tank to prevent overflow.
- Suggest using water-soluble liquid for ultrasonic cleaners. Strong acid or flammability cleaner is forbidden.
- Do not use the machine in severe environments: places where temperature changes fiercely, humidity is high, vibration or impact is strong, corrosive gas or dust is present, water/oil/chemicals splash, or explosive/flammable gas is present.
- Shorten the daily working time. Suggestion is to stop for a few minutes for heat dissipation after working over 30 minutes.
- Heater is forbidden when using flammable liquid (such as alcohol, solvent cleaner etc.) to clean objects. Cleaning effect will be much better if using the machine together with proper solvent.
Operation Procedure
Mechanical Models
- Start: Rotate the 'ULTRASONIC' knob clockwise to set the time required (0-30 min). When the indicator light is on and makes a “ZIZI” sound, ultrasonic operation is working.
- Heating: If heating is needed, start 'HEATING' to adjust the required temperature, normally 40~60°C is recommended.
- Stop Cleaning: Press 'OFF' once to stop ultrasonic operation (indicator light will turn off). Rotate the heating control knob to 'OFF' (indicator light will turn off). Then switch off the unit and disconnect the power supply. Empty out the liquid and clean the tank and the unit with a clean cloth for next use.
Digital Models
- Timer Setting: When power is connected, the default setting is "05:00". Press 'TIME +' once to increase time by 1 min; Press 'TIME -' once to decrease time by 1 min. (Free choice and digital countdown control).
- Temperature Setting: (Heating is optional). When power is connected, the default setting is "50°C". Press 'TEMP +' once to increase temperature by 1°C; Press 'TEMP -' once to decrease temperature by 1°C. If the setting temperature is lower than the actual tank temperature, the operation will stop automatically. When the temperature rises to the setting temperature, the indicating light goes off. During ultrasonic work, two temperature screens show the set temperature and the actual temperature reached.
- Start/Stop: After setting the time and temperature, press 'ON/OFF' once; the equipment will work automatically. Press 'ON/OFF' again to stop the work process.
- Adjusting Power: The function is only available for power adjustable models. Rotate the power knob SLOWLY clockwise to increase the power from 40% to 100%, and SLOWLY counter-clockwise to decrease sonic power.
- Post-Operation: Switch off the unit and disconnect the power supply, empty out the liquid and clean the tank and the unit with a clean cloth for next use.
Maintenance
- The unit must be opened by an authorized specialized person only for maintenance and care.
- Clean the contaminant in the tank frequently.
Troubleshooting
No. | Problem | Possible Causes | Solutions | Remarks |
---|---|---|---|---|
1 | No ultrasonic | A、Power supply not connect B、Fuse broken C、Cable short circuit D、Transducer short circuit E、PCB Board broken F、Other reasons | A、Check and plug power switch. B、Check the fitted power supply and change the same specification fuse. C、Connect fitted cable or replace a new. D、Inquiry our after service engineer. E、Check the broken part and change it. F、Inquiry our after service engineer. | |
2 | Time control failure | A、Timer knob out of control B、Timer failure C、Other reasons | A、Loosen or tighten the screw. B、Replace timer or digital panel. C、Inquiry our after service engineer. | |
3 | No heating | A、Heater power switch bad linkage B、Fuse burn out C、Heating pad burn out D、Digital display board out of control E、Other reasons | A、Check heating plug and connect well. Check outlet line with multimeter: if OK and resistance value is few hundred OHMs. B、Change the same specification fuse. C、Replace the bad heating pad if open circuit. D、Check the broken part and change it. E、Inquiry our after service engineer. | 50-60°C |
4 | Temperature control failure | A、Thermostat loosen B、Thermostat tube broken C、Digital display out of control D、Other reasons | A、Fasten the thermostat header. B、Replace thermostat. C、Check the broken part and change it. D、Inquiry our after service engineer. | 0-60°C |
5 | Not well cleaning | A、Not strong ultrasonic cleaning B、Too high too low liquid surface C、Too high too low temperature D、Not suitable cleaning liquid E、Other reasons | A、Connect ultrasonic button and adjust. B、Adjust liquid into the best surface. C、Adjust temperature into the most fitted. D、Stop and switch off power supply, replace suitable liquid after the previous liquid cool down. E、Inquiry our after service engineer. | |
6 | Electric leakage | A、Customer side not grounded B、Machine not grounded | A、To ensure grounded. B、Check if machine earth wire loosen. | |
7 | Other problems |
Product Drawing & Indication
Mechanical Model Indication
The mechanical model unit features a Lid, Tank, Handle, Timer, Heater, Drainage Valve, Air Vents, Heater Light, and Ultrasonic Light.
Digital Model Indication
The digital model unit features a Lid, Tank, Handle, Drainage Valve, Air Vents, Heater indicator, Ultrasonic indicator, Digital display screen, Temp, Temp +, Heater ON/OFF, Time +, Time, and Ultrasonic ON/OFF controls.