The Extech PH90 Waterproof pH Meter is a precision instrument designed for accurate pH measurement across various applications, including water quality testing, hydroponics, aquaculture, swimming pools, beverage production, and laboratory use. This device features a robust waterproof housing, a replaceable electrode, 3-point automatic calibration, and auto buffer recognition for ease of use and reliable results. Its compact design makes it suitable for both field and professional environments.

Extech PH90 Waterproof pH Meter

Figure 1: Extech PH90 Waterproof pH Meter. This image shows the compact, handheld pH meter with its digital display and control buttons, ready for use.

2. Pagrindinės savybės

3. Sąranka ir paruošimas

3.1. Akumuliatoriaus įdėjimas

The PH90 meter requires two CR2032 Lithium Metal batteries (included). To install or replace batteries:

  1. Unscrew the battery compartment cap located at the top of the meter.
  2. Įdėkite dvi CR2032 baterijas, užtikrindami teisingą poliškumą (+/-).
  3. Tvirtai užsukite baterijų skyriaus dangtelį atgal į vietą.

3.2. Electrode Preparation

Before first use, and after storage, prepare the electrode:

  1. Remove the protective cap from the electrode tip. The cap contains a moist sponge to keep the electrode hydrated.
  2. Rinse the electrode tip thoroughly with distilled or deionized water. Do not wipe the electrode, as this can create static charges.
  3. Allow any excess water to drip off.

3.3. Pradinis kalibravimas

For optimal accuracy, perform a 3-point calibration before initial use and periodically thereafter. Calibration buffers (pH 4, 7, and 10) are sold separately.

  1. Įjunkite matuoklį paspausdami ĮJUNGTA/IŠJUNGTA mygtuką.
  2. Immerse the electrode into a pH 7.00 buffer solution.
  3. Paspauskite ir palaikykite CAL button until "CAL" appears on the display. The meter will automatically recognize the buffer and calibrate.
  4. Nuplaukite elektrodą distiliuotu vandeniu.
  5. Repeat the process with pH 4.00 and pH 10.00 buffer solutions. The meter will automatically detect and calibrate to these points.
  6. Once calibration is complete, the meter will return to measurement mode.
Extech PH90 Meter Display and Buttons

Figure 2: Close-up of the Extech PH90 display and control buttons, including ON/OFF, MODE, and CAL buttons.

4. Naudojimo instrukcijos

4.1. Matavimas

  1. Ensure the meter is calibrated.
  2. Rinse the electrode with distilled water and gently shake off excess.
  3. Panardinkite elektrodo galiuką įample solution, ensuring the sensor is fully submerged.
  4. Gently stir the solution to ensure homogeneity and allow the reading to stabilize. This may take up to a minute, especially if the sample temperature differs significantly from the meter's previous environment.
  5. Read the pH value displayed on the LCD screen. The temperature will also be displayed.
  6. Po matavimo elektrodą kruopščiai nuplaukite distiliuotu vandeniu.

4.2. Laikymo funkcija

Norėdami užfiksuoti dabartinį rodmenį ekrane, paspauskite LAIKYTI button (if available, typically integrated with MODE or a separate button). Press it again to release the hold and return to live measurement.

4.3. Automatic Temperature Compensation (ATC)

The Extech PH90 features Automatic Temperature Compensation, which adjusts pH readings based on the sample's temperature, ensuring accuracy across varying temperatures. The temperature is displayed alongside the pH value.

5. Priežiūra

5.1. Elektrodų priežiūra ir laikymas

Proper electrode care is crucial for maintaining accuracy and extending the life of your meter:

Extech PH90 Electrode Tip

Figure 3: Close-up of the Extech PH90's replaceable electrode tip, which requires proper care and storage.

5.2. Elektrodo keitimas

The PH90 features a replaceable electrode. If the electrode performance degrades significantly despite proper cleaning and calibration, it may need replacement. Refer to the Extech websvetainėje arba susisiekite su klientų aptarnavimo tarnyba, kad gautumėte informacijos apie pirkimąasing replacement electrodes (Part PH95).

5.3. Baterijos keitimas

When the battery indicator appears on the display or the meter fails to power on, replace the two CR2032 Lithium Metal batteries as described in Section 3.1.

6. Problemų sprendimas

ProblemaGalima priežastisSprendimas
Netikslūs arba nestabilūs rodmenys
  • Electrode not calibrated
  • Dirty or dry electrode
  • Temperature not equilibrated
  • Pasibaigusio galiojimo buferiniai tirpalai
  • Pažeistas elektrodas
  • Atlikite 3 taškų kalibravimą.
  • Clean the electrode and ensure it is properly stored.
  • Allow sufficient time for the meter's temperature sensor to equilibrate with the sample.
  • Use fresh calibration buffer solutions.
  • Pakeiskite elektrodą.
Matuoklis neįsijungia
  • Išsikrovusios baterijos
  • Neteisingas akumuliatoriaus įdėjimas
  • Replace batteries with new CR2032 cells.
  • Patikrinkite akumuliatoriaus poliškumą.
"CAL" error during calibration
  • Naudotas netinkamas buferinis tirpalas
  • Contaminated buffer solution
  • Electrode issue
  • Ensure correct pH buffer (4, 7, or 10) is used.
  • Use fresh, uncontaminated buffer solutions.
  • Clean or replace the electrode.

7. Specifikacijos

ParametrasVertė
Modelio numerisPH90
pH diapazonaspH nuo 0 iki 14
pH tikslumas±0.01 pH
Temperatūros diapazonas(Not explicitly stated, but integrated for ATC)
Maitinimo šaltinis2 x CR2032 Lithium Metal batteries
Prekės svoris2.88 uncijos (apie 81.6 gramų)
Gaminio matmenys6.7 x 1.4 x 1.4 colio (apytiksliai 17 x 3.6 x 3.6 cm)
Įtraukti komponentaiFlat surface pH electrode, cap, batteries (pH buffers sold separately)
GamintojasExtech

8. Garantija ir palaikymas

For warranty information, technical support, or to purchase replacement parts and accessories, please visit the official Extech website or contact Extech customer service. You can also find more information and products at the „Extech“ parduotuvė „Amazon“ platformoje.

Susiję dokumentai - PH90

Preview „Extech ExStik™“ vandeniui atsparūs pH matuokliai PH100 ​​ir PH110, naudotojo vadovas
Išsamus „Extech ExStik™“ vandeniui atsparių pH matuoklių, modelių PH100 ​​ir PH110, naudotojo vadovas. Sužinokite apie veikimą, kalibravimą, priežiūrą ir specifikacijas.
Preview „Extech PH220“ vandeniui atsparus delninis pH matuoklis, naudotojo vadovas
„Extech PH220“ vandeniui atsparaus delninio pH matuoklio naudotojo vadovas. Apima įvadą, matuoklio aprašymą, veikimą, kalibravimą, matavimą, laikymą, baterijų keitimą, priežiūrą, trikčių šalinimą ir specifikacijas.
Preview „Extech ET40“ didelio našumo tęstinumo testerio naudotojo vadovas
„Extech ET40“ didelio našumo elektros srovės testerio naudotojo vadovas. Sužinokite, kaip saugiai ir efektyviai patikrinti neįjungtų komponentų, saugiklių, jungiklių, relių, laidų ir grandinių plokščių elektros srovės tęstinumą.
Preview „Extech“ elektromagnetinio lauko matuoklio modelis 480823, naudotojo vadovas
„Extech“ elektromagnetinio lauko matuoklio, modelis 480823, naudotojo vadovas. Šiame dokumente išsamiai aprašomos matuoklio specifikacijos, veikimas, baterijos keitimas, EML poveikio informacija, garantija ir kalibravimo paslaugos. Jis matuoja EML Gauso ir Teslos vienetais, o dažnių juostos plotis yra nuo 30 iki 300 Hz.
Preview „Extech ExStik“ vandeniui atsparūs pH matuokliai PH100 ​​ir PH110, naudotojo vadovas
Išsamus „Extech ExStik“ vandeniui atsparių pH matuoklių, modelių PH100 ​​ir PH110, naudotojo vadovas. Pateikiamos išsamios instrukcijos apie šių didelio tikslumo pH matavimo prietaisų naudojimą, kalibravimą, elektrodų priežiūrą, baterijų keitimą, saugos priemones, specifikacijas ir trikčių šalinimą.
Preview EXTECH 45170 4 viename drėgmės, temperatūros, oro srauto ir šviesos matuoklio naudotojo vadovas
EXTECH 45170, 4 viename matuoklio, matuojančio drėgmę, temperatūrą, oro srautą ir šviesą, naudotojo vadovas. Apima veikimo, specifikacijų ir garantijos informaciją.