Linshang logo Shenzhen Linshang Technology Co., Ltd.
LS155
Color Haze Meter
User Manual V1.06

Please read this manual carefully before using and reserve it for reference.

Product Introduction

The instrument is a multifunctional color haze meter. It adopts full spectrum LED light source and spectral sensor design, which is used for the measurement of haze, transmittance and color of various diffuse transmittance materials such as milky white, matte, frosted surface and various transparent and semi-transparent regular transmittance materials, and it can also display the spectral transmittance curve. The instrument is equipped with 7-inch color screen and capacitive touch screen, which provides excellent operation experience.
Standards for the product
JJF 1303-2011 Calibration Specification for Hazemeter
GBT 2410-2008 Determination of the luminous transmittance and haze of transparent plastics
GBT 36142-2018 Test methods of color and color difference for architectural glass
JJG 595-2002 Colorimeters and Color Difference Meters
GB/T 3978-2008 Standard illuminants and geometric conditions
GBT 7921-2008 Uniform color space and color difference formula
GB 3143-1982 Colour determination method of liquid chemicals (Hazen unit–platinum-cobalt scale)
JJF 1947-2021 Calibration Specification for Platinum-Cobalt Colorimeter
JJG 880-2006 Verification Regulation of Turbidimeters

Parameters

Illumination geometry D/0°
Illumination light source Full spectrum LED light source
Spectral Range 400-700nm
Spectral Interval 10nm
Measuring aperture 5mm/14mm
Measurement conditions Light source D65, field of view 10°; Light source A, field of view 2°;
Light source C, field of view 2°
Measuring time About 3s
Measuring parameter Haze, Transmittance, CIE Lab, LCh, Yxy, Spectral
transmittance,Clarity,Turbidity,Pt-Co(Hazen)
Transmittance resolution 0.01%
Transmittance accuracy Better than ±1%
Transmittance repeatability 10.03 (Standard deviation of 30 measurements at 3-second intervals on a standard haze plate with transmittance -80% and haze -30, after preheating calibration)
Haze/Clarity resolution 0.01%
Haze/Clarity accuracy Better than ±2%
Haze/Clarity repeatability Linshang LS155 Color Haze Meter - icon03 (Standard deviation of 30 measurements at 3-second intervals on a standard haze plate with transmittance -80% and haze -30, after preheating calibration)
Color resolution 0.01
Color inter- instrument agreement Linshang LS155 Color Haze Meter - iconE•ab<=0.4 (Standard haze plate with transmittance -80% and haze -30)
Color repeatability Standard deviation AEtab is within 0.03 (Standard deviation of 30 measurements at 3-second intervals on a standard haze plate with transmittance -80% and haze -30, after preheating calibration)
Turbidity resolution 0.1NTU
Turbidity range 0-1000NTU
Turbidity accuracy Linshang LS155 Color Haze Meter - icon 1(5%H11NTU) H is standard
Pt-Co resolution 0.1
Pt-Co range 0-1000
Pt-Co accuracy Linshang LS155 Color Haze Meter - icon 1(5%H11) H is standard
Display 7-inch 1024.600 dot matrix IPS color screen
Language Simplified Chinese, English
Data transmission USB (Type-C)
Operating temperature range 0-45°C, 0-85%RH (no condensation)
Storage temperature range -25-55°C, 0-85%RH (no condensation)
Size 21.1cmx26.4cmx 36.0cm(LWH)
Weight 5.3kg
Supply Voltage AC100~277V 50/60HZ
Operating Current 0.4A
Operating Power Consumption 80W

Features

  1. The instrument is fashioned with a full-spectrum LED light source and spectral sensor, and it Is used for haze, transmittance, color difference multiple parameters measurement, and can display the spectral transmittance curve.
  2. Adopt a 7-inch capacity touch panel with human-machine interaction interface for convenient operation.
  3. The instrument has a built-in compensation optical path with stable values, allowing long-period calibration.
  4. QC testing function: the threshold can be set.
  5. Can store 1000 measurement records with large storage space.
  6. A powerful function of color difference analysis as well as data import and export with PC software

Operation

  1. Power on/off
    Power on: Plug in the power supply, short press the right side button to turn on the instrument.
    Power off: Long press the right side button to turn off or press “Power off” in the menu bar.
  2. Calibration
    After entering “calibration” interface, you can perform calibration operation as the calibration animation, or skip the calibration. It is recommended to calibrate after a long time of no usage.Linshang LS155 Color Haze Meter - possibleFor failed calibration, the possible reasons are as below:
    ◆ The dust cover has not been removed or the test holes are covered with samples;
    ◆ Tested in a bright light environment;
    ◆ The attenuation of the light source leads to failure of normal use and it needs to be returned to the factory for inspection and repair.
  3. Measurement
    The default measurement parameter of the instrument is “Haze/Transmittance”. Users can click the “Param” at the top of the screen or the “Param Setting” in menu bar to enter the setting interface to select the parameters to be measured (H/T, H, T, CIE_Lab, CIE_LCh, Yxy, Spectra, Clarity,Turbidity,PtCo).Linshang LS155 Color Haze Meter - MeasurementEntering the measurement interface, you can click the ” Compare” on the top of the screen to switch the compare measurement mode.

Linshang LS155 Color Haze Meter - Measurement 1

3.1 Compare off
In the measurement interface, click on “Scan” or short press the button on the right side of the instrument to measure.

Linshang LS155 Color Haze Meter - Compare

The following steps can be performed after completing measurement:
⚫ Save
To save the comparison result, press “Save” on the upper right corner. You can choose default or name the record manually.
⚫ History
If users need to view the saved results, click “History” to view the records, and records can be paged, deleted, renamed and searched.
3.2 Compare on Linshang LS155 Color Haze Meter - Compare on

In the measurement interface, press the “Scan Standard” button to set current value as standard. Press the “Scan Sample” button to use current value as sample. If short press the button on the right side, another measurement will also be taken and the current value will replace the previous one. When measuring or adding a standard and measuring a sample, the difference or color difference between the two measured values will be calculated with the following results:
◆ The difference value
The difference Δ is the sample data minus the standard data;
◆ Color Bias
When the measurement parameter is selected as Lab, the color bias will be judged by ΔL, Δa, and Δb;
◆ Δ E value
When the measurement parameter is selected as Lab, LCh, or Yxy, ΔE is calculated using the ΔE*ab color difference formula;
◆ QC test
Suitable for difference inspection in production process. If ΔE is higher than the set threshold, it will read “NG” and sown in red; if ΔE is less than or equal to the set threshold, it will read “Pass” and display in green.

Linshang LS155 Color Haze Meter - QC testLinshang LS155 Color Haze Meter - QC test 1

The following steps can be performed after completing compare measurement:

  • Save
    To save the comparison result, press “Save” on the upper right corner. You can choose default or name the record manually.
  • History
    If users need to view the saved results, click “History” to view the records, and records can be paged, deleted, renamed and searched.
  • Save standard
    After the user has finished measuring the standard value, if need to save it for later calling, click the ” Save standard” to save it, and the saved value can be queried in the add standard interface.
  • Add standard
    After clicking “Add standard”, the interface will jump to “Add standard”, click the value you want to add, and the selected value will be used as the standard, and will automatically return to the measurement interface to compare with the current sample or the upcoming sample.
  • Input standard
    The user can manually input the Haze, Transmittance, Lab, LCh, Yxy value as standard.
  • Tolerance Setting
    Setting the threshold for QC inspection.

Menu bar

In the measurement interface, click the “Linshang LS155 Color Haze Meter - icon 2 ” icon on the upper left corner to pop up the menu bar, with the following options: Language, Calibration, Standard, Param Settings, Device info, Reset to defaults and Power off.
4.1 Calibration
Can enter the calibration interface to perform calibration operation of the instrument.
4.2 Language
Support Simplified Chinese and English.
4.3 Standard
Selection of ASTM/ISO standards, switching of measurement aperture 14mm/5mm, and setting of light source and observer’s field of view.
4.4 Param Settings
Select the parameter desired to be measured and displayed. Shortcut keys for the same functions are available at the top of the measurement interface.
4.5 Device Info
“Device Info” displays the basic information of the instrument, including: Device Model, Firmware Version, and Device Status.
4.6 Reset to defaults
To restore parameters to the factory settings and clear the data in the measurement interface.
4.7 Power Off
Press “Power off” to power off the instrument.
V PC Software
You can use the PC software to connect the computer via USB. The software has functions of compare measurement, reading compare record, export compare data to Excel, qualified number, unqualified number, total number statistics, report generation and printing, etc. For specific operation, please refer to the “Color Haze Meter Software Operation Manual”.

Linshang LS155 Color Haze Meter - Software

Precaution

  1. When the colorimeter has not been used for a long time, it is recommended to perform calibration before using.
  2. Please ensure that the sample is evenly colored with a flat and clean surface, otherwise it will affect the measurement accuracy.
  3. Avoid liquids entering the instrument through the measurement aperture, as this will damage the instrument and affect measurement accuracy and operational safety.
  4. When the instrument is not in use, put on the dust cover to prevent dust from entering or prolonged humidity, which may affect the measurement accuracy.
  5. It’s recommended to calibrate the instrument once each year. And we offer calibration service.
  6. Due to color difference in the LCD display, the color displayed on the instrument screen is for reference only.
  7. When measuring the haze, transmittance, color, clarity, turbidity, pt-co, in order to measure more accurately, it is necessary to fill pure water into the cuvette to calibrate it before measurement; at the end of the measurement, it is necessary to remove the test object to recalibrate again.
  8. Modify the standard settings (standard, measurement aperture, measurement conditions) parameters of the instrument, need to recalibrate.

Packing List

No. Description  Quantity  Unit 
1 Color Haze Meter 1 Set
2 USB cable 1 pcs
3 Power cable 1 pcs
4 5mm aperture cover 1 pcs
5 Screwdrivers 1 pcs
6 User manual 1 pcs
7 Certificate / Warranty card 1 pcs

Service

  1. The colorimeter has one-year warranty. If the meter works abnormally, please send the whole instrument to our company for maintenance.
  2. Provide users with spare parts and lifelong maintenance services.
  3. Provide the users with the gauge calibration service.
  4. Free technical support for long term.

Manufacturer: Shenzhen Linshang Technology Co., Ltd.
Website: www.linshangtech.com
Service hotline: +86-755-86263411