ZEISS logoA6 Single Use Injector with AT Smart Cartridge ZEISS A6 Single Use Injector with AT Smart Cartridge SetSet Instructions

A6 Single Use Injector with AT Smart Cartridge Set

Handling instructions for non-preloaded ZEISS MICS IOLs
Preparing the non-preloaded ZEISS MICS IOLs for use.

ZEISS A6 Single Use Injector with AT Smart Cartridge Set - fig
Fill the barrel halfway with viscoelastic substance, preferably with hypromellose.
Put two lines of viscoelastic to the plane of the  cartridge. Do not touch the tip of the cartridge.
Pull the lens gently from its holder using a forceps with fine tips.
Do not let the IOL dry up.
Do not use the fluid from the bottle for ocular irrigation.
Place the lens in the center of the
opened cartridge.
ZEISS A6 Single Use Injector with AT Smart Cartridge Set - fig 1
Hold optic down, center the IOL on the loading  deck of the cartridge.
Close the cartridge without clamping the lens.
Control the position of the lens within the  channel of the closed cartridge. Haptic must move freely. Injector A6 with groove for introduction of the  *AT.Smart cartridge.
ZEISS A6 Single Use Injector with AT Smart Cartridge Set - fig 2
Push the loaded cartridge from the front into the  groove until stop position. Lock cartridge by holding the tip and rotating the injector. Gently push plunger forward and check regular  and continuous forward movement of the lens.

Please refer to our webpages for diopter range applicable for each MICS IOL.ZEISS A6 Single Use Injector with AT Smart Cartridge Set - fig 3

*AT.Smart Cartridge
Single-use injector A6 with silicone tip

EN_32_022_0025I
Single Injector A6 is a reference of MDJ (120). The contents of the brochure may differ from the current status of approval of the product in your country.
Please contact our regional representative for more information. Subject to change in design and scope of delivery and as a result of ongoing technical development. Printed  on elemental chlorine-free bleached paper.
© 2015 by Carl Zeiss MediaTek AG. All copyrights reserved.

 

SMART METER SMPO1000 US iPulseOx Pulse Oximeter - icon 5 Carl Zeiss Meditec AG
Goeschwitzer Str. 51–52
07745 Jena
Germany
www.zeiss.com/iol
www.zeiss.com/med/contacts
CE SYMBOL AT.Smart Cartridge Set

Documents / Resources

ZEISS A6 Single Use Injector with AT Smart Cartridge Set [pdf] Instructions
A6 Single Use Injector with AT Smart Cartridge Set, A6, Single Use Injector with AT Smart Cartridge Set, Smart Cartridge Set
ZEISS A6 Single-Use Injector with AT.Smart Cartridge Set [pdf] Instruction Manual
A6 Single-Use Injector with AT.Smart Cartridge Set, Single-Use Injector with AT.Smart Cartridge Set, Injector with AT.Smart Cartridge Set, AT.Smart Cartridge Set, Cartridge Set
ZEISS A6 Single-Use Injector with AT.Smart Cartridge Set [pdf] Instruction Manual
A6, A6 Single-Use Injector with AT.Smart Cartridge Set, Single-Use Injector with AT.Smart Cartridge Set, Injector with AT.Smart Cartridge Set, AT.Smart Cartridge Set, Cartridge Set

References

Leave a comment

Your email address will not be published. Required fields are marked *