Specifications

  • Product Name: Reference 620TM Potentiostat/Galvanostat/ZRA
  • Manufacturer: Gamry Instruments, Inc.
  • Revision: 1.31
  • Copyright Year: 2023
  • Serial Number: N/A (unique to each product)
  • Warranty: 2 years from the original shipment date

Product Usage Instructions

  • Installation
    • Ensure all components are included in the package.
    • Refer to the installation guide provided for detailed step-by-step instructions.
    • Connect the potentiostat/galvanostat/ZRA to a compatible power source using the provided cables.
    • Install any necessary software updates from the manufacturer’s website.
  • Operation
    • Power on the instrument using the designated power button.
    • Follow the user manual guidelines to set up your desired experiment parameters.
    • Connect your samples or electrodes as per the experimental requirements.
    • Initiate the experiment and monitor the results displayed on the screen.
  • Maintenance
    • Regularly clean the instrument using a soft, dry cloth to remove dust and debris.
    • Avoid exposing the instrument to extreme temperatures or humidity levels.
    • Store the instrument in a cool, dry place when not in use.
    • Refer to the user manual for specific maintenance tasks recommended by the manufacturer.

FAQs

  • Q: What is the warranty period for the Reference 620TM Potentiostat/Galvanostat/ZRA?
    • A: The product comes with a 2-year warranty from the original shipment date ensuring it is free of defects resulting from faulty manufacture.
  • Q: How can I request service or support for the product?
    • A: You can visit Gamry Instruments, Inc.’s service and support page at https://www.gamry.com/support-2/ for information on installation, software updates, and training. You can also contact them via email or telephone using the provided contact details.
  • Q: Are software updates included in the warranty?
    • A: Software updates are not included in the standard warranty and may be available at an additional cost.

Reference 620TM Potentiostat/Galvanostat/ZRA
Operator’s Manual
Copyright 2023 Gamry Instruments, Inc. Revision 1.31 July 21, 2023 988-00086

If You Have Problems
Please visit our service and support page at https://www.gamry.com/support-2/. This page contains information on installation, software updates, and training. It also contains links to the latest available documentation. If you are unable to locate the information you need from our website, you can contact us via email using the link provided on our website. Alternatively, you can contact us one of the following ways:

Internet Telephone

https://www.gamry.com/support-2/ 215-682-9330 9:00 AM-5:00 PM US Eastern Standard Time 877-367-4267 Toll-free US & Canada Only

Please have your instrument model and serial numbers available, as well as any applicable software and firmware revisions.
If you have problems in installation or use of a system containing a Reference 620, please call from a phone next to your computer, where you can type and read the screen while talking to us.
We will be happy to provide a reasonable level of free support for registered users of the Reference 620 Potentiostat/Galvanostat/ZRA. Reasonable support includes telephone assistance covering the normal installation, use and simple customization of a computerized system containing a Reference 620 connected to a Windows®-compatible computer.
A service contract that extends both the hardware warranty and software update period is available at an additional charge. Software updates do not include software enhancements offered to our customers at additional cost.
Enhancements to the Reference 620 and Gamry’s standard applications software that require significant engineering time on our part can be performed on a contract basis. Contact us with your requirements.
Limited Warranty
Gamry Instruments, Inc. warrants to the original user of this product that it shall be free of defects resulting from faulty manufacture of the product or its components for a period of two years from the original shipment date of your purchase.
Gamry Instruments, Inc. makes no warranties regarding either the satisfactory performance of the Reference 620 Potentiostat/Galvanostat/ZRA including the software provided with this product or the fitness of the product for any particular purpose. The remedy for breach of this Limited Warranty shall be limited solely to repair or replacement, as determined by Gamry Instruments, Inc., and shall not include other damages.
Gamry Instruments, Inc. reserves the right to make revisions to the system at any time without incurring any obligation to install same on systems previously purchased. All system specifications are subject to change without notice.
There are no warranties which extend beyond the description herein. This warranty is in lieu of, and excludes any and all other warranties or representations, expressed, implied or statutory, including merchantability and fitness, as well as any and all other obligations or liabilities of Gamry Instruments, Inc; including but not limited to, special or consequential damages.
This Limited Warranty gives you specific legal rights and you may have others, which vary from state to state. Some states do not allow for the exclusion of incidental or consequential damages.
No person, firm or corporation is authorized to assume for Gamry Instruments, Inc., any additional obligation or liability not expressly provided herein except in writing duly executed by an officer of Gamry Instruments, Inc.
3

Disclaimers
Gamry Instruments, Inc. cannot guarantee that the Reference 620 Potentiostat/Galvanostat/ZRA will work with all computer systems, operating systems, or third-party software applications hardware/software. The information in this manual has been carefully checked and is believed to be accurate as of the time of printing. However, Gamry Instruments, Inc. assumes no responsibility for errors that might appear.
Copyrights
Reference 620TM Potentiostat/Galvanostat/ZRA Operator’s Manual copyright 2007-2023, Gamry Instruments, Inc., all rights reserved. Gamry Framework copyright 1989-2022, Gamry Instruments, Inc., all rights reserved. Interface 1010, Interface 5000, Interface Power Hub, EIS Box 5000, Reference 620, Reference 3000TM, Reference 3000AETM, Reference 30K, LPI1010, eQCM 15M, IMX8, Gamry Framework, Faraday Shield, and Gamry are trademarks of Gamry Instruments, Inc. Windows® is a registered trademark of Microsoft Corporation. No part of this document may be copied or reproduced in any form without the prior written consent of Gamry Instruments, Inc.

Safety Considerations

Chapter 1: Safety Considerations
Your Reference 620 Potentiostat/Galvanostat/ZRA has been supplied in a safe condition. This chapter of the Reference 620 Operator’s Manual contains some information and warnings that you must follow to insure continued safe operation of the Reference 620.
Inspection
When you receive your Reference 620 Potentiostat/Galvanostat/ZRA, inspect it for evidence of shipping damage. If you note any damage, please notify Gamry Instruments Inc. and the shipping carrier immediately. Save the shipping container for possible inspection by the carrier.
Warning: A Reference 620 damaged in shipment can be a safety hazard. Do not operate
damaged apparatus until a qualified service technician has verified its safety. Tag a damaged Reference 620 to indicate that it could be a safety hazard.
Product Safety
The Reference 620 has been designed, tested, and certified to meet the requirements of an international standard, EN 61010, Safety requirements for electrical equipment for measurement, control, and laboratory use. As defined in this standard, it is a Category II apparatus, with any “hazardous live voltages” protected by “reinforced insulation”. Most of the Reference 620 circuitry operates at voltages low enough to be considered safe. The Reference 620 contains a limited amount of internal circuitry that is at “hazardous live” voltages as defined in EN 61010 (the standard mentioned above). “Reinforced insulation” (again defined in EN 61010) is used to reduce the risk of electrical shock due to this “hazardous live” voltage. The majority of the Reference 620’s circuitry does not contain voltages higher than 42 V DC. As a generalization, input and output voltages in the Reference 620 are limited to 36 V. This voltage level is considered safe. The “AC Adapter” supplied with the Reference 620 is certified under EN 60950. The AC Adapter converts the AC mains voltage to 24 V DC, which is used to power the Reference 620. Always use the AC adapter (power brick) supplied with your Reference 620 to supply DC power to the instrument.
Caution: Do not use a DC power source other than the AC adapter model provided with
your Reference 620. Other replacements may void the performance and/or safety characteristics of the Reference 620.
AC Mains Connection to the Power Brick
The Reference 620 does not connect directly to an AC Mains supply. Instead, the mains are connected to the desktop AC adapter (power brick), which outputs 24 V DC, which in turn powers the Reference 620.
9

The Reference 620’s AC Adapter is rated for operation from 100 to 240 V AC, 47 to 63 Hz. It should therefore be useful throughout the world. The Reference 620 is normally provided with an AC line cord suitable for your location. This AC line cord connects the AC mains to the AC power adapter. If your Reference 620 has been provided without an AC line cord, or a cord that is not compatible with your local AC mains socket, obtain a line cord certified for use in your country. Contact your local Gamry Representative or Contact Technical Support if you are uncertain what AC line cord to use.
Grounding in the Reference 620
The circuitry and the metal case of the Reference 620 are not connected to an earth ground. If they were connected to earth ground, it would compromise the Reference 620’s ability to make measurements in electrochemical cells that contain earth grounded conductors. A few examples of such cells include autoclaves, metallographic stress apparatus, and detectors for capillary electrophoresis. Most electrochemical cells are isolated from earth ground, so isolation of the Reference 620 from earth is not required. In these cases, connection of the Reference 620 chassis to an earth ground may lower noise seen in electrochemical tests. A Chassis Ground binding post on the rear panel of the Reference 620 makes for easy implementation of this connection. Simply run a wire from this binding post to a suitable source of earth ground. A black 1.2-meter wire is provided with the Reference 620 to ease this connection.
Sources of earth ground include · Most metal water pipes, · The chassis of most electronic apparatus (which are generally earth-grounded), and · The protective ground terminal of an AC mains power plug. We recommend that you discuss grounding with an electrical or electronics professional prior to making this earth-ground connection. Note this connection of the Reference 620 to an earth ground is not a “Protective Earth Ground” as defined in EN 61010. The Reference 620 is safe in the absence of this connection. This binding post is not intended for any use other than connecting the Reference 620 to an earth ground to improve shielding against noise. Connecting this binding post to a hazardous voltage can create a significant safety hazard.
Warning: Do not connect the chassis ground binding post to any voltage other than
earth ground. An improper connection can create a safety hazard, which could result in personal injury or death.
10

Caution: Do not earth ground the Reference 620 if your electrochemical cell has an earth
ground connection. This is particularly hazardous when the cell can source significant energy. The cell can attempt to discharge through the Reference 620’s circuitry, even when the cell switch is open. Many electrochemical cells involved in energy storage and conversion, including batteries, fuel cells, supercapacitors and photovoltaic cells can source enough energy to form a hazard. Most Reference 620 cell cables contain fuses that will prevent instrument damage when earth ground connections form a low impedance path that attempts to discharge the cell. The blown fuse will prevent normal Reference 620 operation, until the fuse is replaced.
Operation with Earth-Grounded Cells and Auxiliary Apparatus
As described above, the Reference 620 circuitry is isolated from earth ground, allowing it to make measurements on cells that include an earth ground. This ground isolation is often called floating operation. Cells with earth ground include many autoclaves, pipelines and storage tanks, and many fuel-cell systems. Connection of the Reference 620 to auxiliary apparatus will often earth-ground the Reference 620, destroying its ability to float and make measurements on earth-grounded cells. Connection of the Monitor BNCs to an oscilloscope is an example where the instrument is earthed. The User I/O connector can be connected to earth-grounded apparatus without earth-grounding the Reference 620, if the cable connections are done carefully. The metal shell on the Reference 620 User I/O Connector is connected to the instrument’s chassis, which is a floating ground. In a system that needs isolation from earth ground, the shield of a User I/O cable must not connect the D-connector’s metal shell to earth ground. Reference all User I/O signals to pin 6 of the D-connector, which is the ground reference for the signals on the Reference 620 User I/O connector.
Caution: Floating operation of the Reference 620 can be compromised by improper cables
to your I/O Connector. We do not recommend use of standard 15-pin shielded cables with this connector. Custom cables with the shield connected to pin 6 of the D-connector are always preferred and absolutely required when the cell is earth grounded.
Warning: Do not connect the chassis ground binding post to any voltage other than
earth ground. An improper connection can create a safety hazard, which could result in personal injury or death. The Reference 620 contains a TVS diode that limits the voltage difference between the Reference 620’s chassis ground and earth ground to about 28 V. This voltage limiter is not part of the safety mechanisms in the Reference 620, instead it is present to limit the possibility of improper instrument operation or instrument damage caused by electrostatic discharge (static electricity) and other surge events such as lightning.
11

Temperature and Ventilation
Your Reference 620 Potentiostat/Galvanostat/ZRA was designed for indoor use at ambient temperatures between 0C and 45C.
The Reference 620 uses forced air-cooling to keep its electronic components within their recommended operating temperature range. Air is drawn into the Reference 620 chassis through slots located on the sides of the chassis near the front panel. The air exits through the two fans at the rear of the unit.

Caution: Do not block the airflow into or out of the Reference 620 chassis. While the
circuitry should shut down before it is damaged by excessive heat, the Reference 620 enclosure may become uncomfortably hot to the touch if insufficient air flows through the chassis. Running the Reference 620 without adequate cooling could shorten the time to failure of some of the circuitry.

Be careful when operating the Reference 620 in an enclosed space (such as an enclosed relay rack or NEMA
enclosure). The temperature within the enclosure must not exceed 45C. You may need to provide ventilation holes or even forced air-cooling for the enclosed space if you determine that there is an excessive temperature rise.

Defects and Abnormal Stresses
Treat your Reference 620 as potentially hazardous if any of the following is true of the unit: It shows visible damage, It does not operate properly, It has been stored for a long period of time under unfavorable conditions, It has been dropped or subjected to severe transport stress, It has been subjected to environmental stress (corrosive atmosphere, fire, etc.).
Do not use your Reference 620 or any other apparatus if you think it could be hazardous. Have it checked by qualified service personnel.

Environmental Limits

There are environmental-limit conditions on the storage, shipping, and operation of this equipment. The Reference 620 is not designed for outdoor use.

Storage

Ambient Temperature

-40C to 75C

Relative Humidity

Maximum 90% non-condensing

Shipping

Same as storage plus Acceleration

Maximum 30 G

Operation

Ambient Temperature Relative Humidity

0C to 45C Maximum 90% non-condensing

12

Warning: The Reference 620 is not designed for operation in conditions where liquid
water may enter the chassis, or water vapor may condense within the chassis. Operation of a Reference 620 that has water within the chassis can create a safety hazard, which could result in personal injury or death.
Cleaning
Disconnect the Reference 620 from all power sources prior to cleaning. Use a cloth lightly dampened with either clean water or water containing a mild detergent to clean the outside of the Reference 620 enclosure. Alternatively, you can use isopropyl alcohol. Do not use a wet rag or allow fluid to enter the Reference 620 enclosure. Do not immerse the Reference 620 in any type of cleaning fluid (including water). Do not use any abrasive cleaners.
Service
Your Reference 620 Potentiostat/Galvanostat/ZRA has no user-serviceable parts inside. Refer all service to a qualified service technician.
Warning: Never operate the Reference 620 with any cover or panel on the chassis
open. Dangerous voltages may be present at several points within the Reference 620 chassis, including PC board traces. Always remove the power connection before opening the Reference 620 case.
RF Warning
Your Reference 620 Potentiostat/Galvanostat/ZRA generates, uses, and can radiate radio-frequency energy. The radiated levels are low enough that the Reference 620 should not create an interference problem in most industrial laboratory environments. The Reference 620 has been tested for both radiated and conducted RF interference and has been found to be in compliance with FCC Part 18 and EN 61326:1998–Electrical equipment for measurement, control, and laboratory use– EMC Requirements.
Electrical Transient Sensitivity / Electrostatic Discharge
Your Reference 620 Potentiostat/Galvanostat/ZRA was designed to offer reasonable immunity from electrical transients, including transients on the incoming AC Mains supply and Electrostatic Discharge. It has been tested for compliance with EN 61326:1998–Electrical equipment for measurement, control, and laboratory use– EMC Requirements describing acceptable limits for Electrical Transient susceptibility in Laboratory Test equipment. The Reference 620 is not rated for continuous use when subject to ESD events. It should suffer no permanent damage when subject to the standard ESD events defined in EN 61326 but may cease normal operation until it is powered down and restarted. In severe cases, the Reference 620 could malfunction as a result of electrical transients such as a static discharge. If you are having problems in this regard, the following steps may help: If the problem is static electricity (sparks are apparent when you touch the Reference 620 or its cables):
13

Placing your Reference 620 on a static control work surface may help. Static-control work surfaces are now generally available from computer-supply houses and electronics tool suppliers. An antistatic floor mat may also help, particularly if a carpet is involved in generating the static electricity.
Reinstall insulative rubber covers on BNC jacks while connections are not being used. These covers are present to prevent disruption during use of device, should a discharge occur.
Air ionizers or even simple air humidifiers can reduce the voltage available in static discharges. If the problem is AC power-line transients (often from large electrical motors near the Reference 620):
Try plugging your Reference 620 into a different AC power branch circuit. Plug your Reference into a power-line surge suppressor. Inexpensive surge suppressors designed for
use with computer equipment are now generally available. Contact Gamry Instruments, Inc. if these measures do not solve the problem.
CE Compliance
The European Community has instituted standards limiting radio-frequency interference emitted by electronic devices, setting limits for susceptibility of apparatus to RF energy and transient events, and mandating safety requirements. Gamry Instruments, Inc. has designed and tested the Reference 620 to comply with these standards. The relevant CE regulations include EN 61010 and EN 61326.
RoHS Compliance
The Reference 620 has been built using lead-free components and lead-free solder. It is in compliance with the European RoHS initiative.
14

Introduction

Chapter 2: Introduction
About this Manual
This manual covers the installation, safety, and use of the Gamry Instruments Reference 620 Potentiostat/Galvanostat/ZRA. This manual describes use of a Reference 620 with Revision 7.9 (and later revisions) of the Gamry Framework software. It is equally useful when setting up a newly purchased potentiostat or modifying the setup of an older potentiostat for use with new software. You will find dry technical material such as specifications and connector pin-out diagrams in the Appendices. This manual does not discuss software installation or software operation in any detail. Software support for the Reference 620 is described in the Gamry Framework’s Help system. All the Gamry Instruments’ applications running under the Gamry Framework control the Reference 620 via a PSTAT object. See the Framework’s Help system for information concerning PSTAT objects and their functions.
About the Reference 620
The Reference 620 Potentiostat is a research-grade electrochemical instrument packaged in a small, easy-tohandle case. It offers measurement capabilities similar to instruments that are more than ten times its size and weight and more than twice its price. The Reference 620 can operate as a potentiostat, a galvanostat, or a ZRA (zero-resistance ammeter). Reference 620 features include:
Eleven-decade current auto-ranging, Electrical isolation from earth ground, Current-interrupt iR-compensation, and Both analog and digital filtering. A sine-wave generator on the Reference 620 allows its use for impedance measurements at frequencies up to 5 MHz. Data can be acquired at frequencies up to 300 000 points per second, allowing cyclic voltammetry at scan rates of 1500 V/s with 5 mV per point resolution. A unique DSP (Digital Signal Processing) data-acquisition mode allows the Reference 620 to reject noise from the instrument itself, from the electrochemical cell, and from the laboratory environment. In many cases where other instruments require a cell in a Faraday shield to make quiet measurements, the Reference 620 can be used with the cell exposed on a bench top. The Reference 620 offers an unprecedented combination of high speed, high sensitivity, and low noise. At the time this manual is written, the Reference 620 offers the widest EIS performance area (in a plot of impedance versus frequency) of any potentiostat. State-of-the-art analog components were used throughout the design. In all design decisions, performance weighed more heavily than product cost. The Reference 620, like all Gamry potentiostats, requires a computer for its use. The Reference 620 connects to the computer through a USB connection. The USB connection has become truly universal, with USB ports found on all modern computers. Gamry’s software currently supports up to 16 Reference 620 Potentiostats connected to one computer. The Reference 620 is isolated from earth ground. It can therefore be used to make measurements on cells that contain an earth-grounded metal. A few of examples of such systems include are autoclaves, large metal storage tanks, stress apparatus, and capillary electrophoresis detectors.
15

Notational Conventions
In order to make this manual more readable we have adopted some notational conventions. These are used throughout this manual and all other Gamry Instruments manuals:
Numbered lists. A numbered list is reserved for step-by-step procedures, with the steps always performed sequentially.
Bulleted list. The items in a bulleted list, such as this one, are grouped together because they represent similar items. The order of items in the list is not critical.
File names and folders. Inside paragraphs, references to computer files and Windows® folders are capitalized and placed within quotes, for example: “C:MYGAMRYDATACV.DTA” and “GAMRY.INI”.
16

Instrument Circuitry

Chapter 3: Instrument Circuitry
Reference 620 Schematic/Block Diagrams
If you are not familiar with electronic schematics or potentiostats, you probably want to skip this chapter. This information is for expert use only and is not required for routine use of the Reference 620. The following figures are partly schematic diagrams and partly block diagrams. They are intended to show the basic principles of the Reference 620 circuitry without the confusion of the full circuitry details. The complexity of the Reference 620 can be quite daunting: the Reference 620 circuit boards contain more than 3000 components connected by almost 2500 circuit nets! The schematic/block diagram figures show:
The potentiostat in potentiostatic control mode, The control-board circuits for signal generation, The control-board circuits for signal conditioning and A/D conversion, The Auxiliary ADC channel-input switching, The processors in the Reference 620, DC-DC power conversion.
Figure 3-1 Reference 620 Potentiostat Board in Potentiostat Mode
Simplified Schematic/Block DiagramGAMRY-INSTRUMENTS-Reference-620-Potentiostat,-Galvanostat,-ZRA-FIG- (1)
17

Notes for Figure 3-1: Only Potentiostat Mode circuitry is shown in this figure. In this mode the voltage difference between the Reference and Working Sense leads (called Esig) is feedback into the control amplifier. In Galvanostat Mode, the feedback is from Isig. In the ZRA mode, the feedback is from a differential amplifier (not shown) measuring the difference between the Counter Sense and Working Sense leads of the cell cable. It is conceptually similar to the voltage-sensing circuit that generates Esig. Computer bus connections to Bias DAC and PFIR (Positive Feedback IR compensation) DAC are not shown. Switches are either reed relays or MOS switches as appropriate. All components shown as variable (resistance Rm and capacitors IEStab & CASpeed) are fixed-value components switched into the circuit. They are not continuously variable as implied in the figure. The monitor BNC connectors for Isig and Esig are lightly filtered using an RLC circuit. The programmable attenuator on Esig prior to the ADC channel scales Esig so that it is compatible with the A/D channel’s ±3 V input range. The 0.25 gain setting allows the Reference 620 to measure potential signals slightly in excess of 10 V (on a 12 V full-scale range). Isig is gained to be 3 V full-scale so it does not require a similar attenuation function. All the resistors summing voltages into the Control Amplifier input do not have values shown; their values depend on scaling factors too complex to discuss in this simplified diagram. Calibration components are not shown. Overload protection and overload detection are not shown. Good engineering practice demands that any possible misconnection of the cell leads will not damage the instrument. This practice has been followed in the Reference 620 design.
18

Figure 3-2 Reference 620 Signal Generation CircuitryGAMRY-INSTRUMENTS-Reference-620-Potentiostat,-Galvanostat,-ZRA-FIG- (2)
Notes for Figure 3-2: All the resistors summing voltages into the Summing Amplifier input do not have values shown on the diagram. Their values depend on scaling factors too complex for this simplified diagram. The IR DAC has a 8 V full-scale range. Calibration components are not shown. The DDS can generate fixed-amplitude sine waves with frequencies between 5 MHz and 1 mHz. In practice, Gamry’s electrochemical-impedance spectroscopy software uses the Scan DAC to generate sine signals if the frequency is below 20 Hz. The low-pass filter removes high-frequency distortion in the “raw” DDS output. The attenuator scales the DDS. The maximum output signal is 4.096 V peak-to-peak, and the minimum is approximately 20 V peak-to-peak. The BNC connector for Sig Gen out is lightly filtered using an RLC circuit.
19

Figure 3-3 One A/D Signal Chain in the Reference 620GAMRY-INSTRUMENTS-Reference-620-Potentiostat,-Galvanostat,-ZRA-FIG- (3)
Notes for Figure 3-3: This diagram shows one of three identical ADC channels. One channel is dedicated to measurement of the potentiostat’s current signal, another is used to measure the cell or stack voltage, and the third is switched between a wide selection of possible signals. All three A/D converters are triggered simultaneously to start a conversion. This trigger and the pulse updating the Scan DAC voltage are under the control of a hardware state-machine. This insures that all waveform and data-acquisition timing is tightly controlled and reproducible point-to-point. By default, the data acquisition is synchronized with the 300 kHz power-supply switching frequency, to reduce noise from the power supply. Data-acquisition times that are a multiple of 3.333 s maintain this synchronization. All analog signals that cross from the Potentiostat Board to the Control Board or vice versa are received differentially as shown here. The 5 Hz, 1 kHz, and 200 kHz filters are two-pole Butterworth filters. The 3 MHz RLC filter has an arbitrary transfer function. All signal-channel components are selected for optimal DC accuracy, low noise, and high bandwidth.
20

Figure 3-4 Aux Channel Input SwitchingGAMRY-INSTRUMENTS-Reference-620-Potentiostat,-Galvanostat,-ZRA-FIG- (4)
Notes for Figure 3-4: Two ADC Channels are dedicated to the potentiostat’s voltage and current signals. This diagram shows the signals that can be connected to the third channel. The Aux ADC BNC input is a differential input. It by default it is configured for 50 k input impedance and less than 1 kHz bandwidth. Software controlled switches can reconfigure it for higher input impedance faster response. The thermocouple input connects a K-type thermocouple. The type of thermocouple is shown next to the connector on the rear of the instrument. Note that this circuit must be calibrated to obtain reasonable accuracy. A selection on the Framework Utility menu allows user calibration of this input. ZSIG is the voltage of the cell cable’s counter sense lead minus the voltage on the Work Sense lead. It is call Zsig because this is the feedback voltage used in ZRA (zero resistance ammeter) mode. Connection of the potentiostat’s CA input voltage to the Aux channel is reserved for use in future Gamry Instruments’ software and/or hardware projects.
21

Figure 3-5 Microprocessors in the Reference 620GAMRY-INSTRUMENTS-Reference-620-Potentiostat,-Galvanostat,-ZRA-FIG- (5)
Notes for Figure 3-5: Note the lack of a ground connection between the USB bus and the Reference 620 circuitry. The USB Controller firmware is loaded into microprocessor RAM on power-up. The Gamry Instrument Manager can update the USB firmware via the USB. The Power PC firmware is also transferred from ROM into RAM on power-up. The Power PC firmware can also be updated over the USB via a selection in the Reference 620 section of the Instrument Manager. Time-critical sections of the Power PC code are kept in the processor’s fast-cache memory. The term USART refers to a Universal Synchronous/Asynchronous Receiver Transmitter. It converts parallel data to a self-clocking serial bit stream. The USARTs send data at 6 Mbits/s. The Bus Transceiver isolates bus activity on the Controller and Potentiostat boards. Only reads and write-to locations on these boards generate bus activity. This reduces noise pick-up. The Bus Transceiver isolates bus activity on the Controller and Potentiostat boards. Only reads and writes to locations on these boards generate bus activity. This reduces noise pick-up. Unlike previous Gamry Instruments’ Potentiostats, the Reference 620 calibration data are stored in the instrument, not in a data file. When a Reference 620 is moved from one computer to another, its calibration remains valid.
22

Figure 3-6 DC-DC Power Conversion
Notes for Figure 3-6: Note the ground isolation between the input power and the Reference 620 circuitry. The Reference 620 chassis is connected to the Floating Instrument Ground. Transformers and isolators are the only components connected between the grounds. The 300 kHz power-supply sync signal is derived from the same clock used to control data acquisition. Data points taken at an integer multiple of 3.333 s/point are synchronized with the power supply, minimizing the effect of power-supply noise on the data. Additional circuitry that is not shown protects the Reference 620 against ESD (electrostatic discharge) and electrical surges. Note that the Reference 620 is also protected against damage if an incorrect polarity power input is connected to the unit. The incoming DC voltage must be between 22 and 26 V. With inputs below 22 V, the PWM (Pulse Width Modulator) may be unable to regulate the supply. Above 26 V, the PWM may not start up.
23

Installation

Chapter 4: Installation
This chapter of the Gamry Instruments Inc. Reference 620 Operator’s Manual covers normal installation of the Reference 620. We assume the Reference 620 is installed as part of a Gamry Framework-based electrochemical measurement system containing a Microsoft Windows®-compatible computer. These instructions assume use with Gamry’s Framework Software Revision 7.9 or higher.
Figure 4-1 Front View of a Reference 620
Initial Visual Inspection
After you remove your Reference 620 from its shipping carton, check it for any signs of shipping damage. If you find any damage, please notify Gamry Instruments, Inc., and the shipping carrier immediately. Save the shipping container for possible inspection by the carrier.
25

Warning: The “reinforced insulation” that keeps the operator from accessing the
“hazardous live” voltages in the Reference 620 can be rendered ineffective if the Reference 620 is damaged in shipment. Do not operate damaged apparatus until a qualified service technician has verified its safety. Tag a damaged Reference 620 to indicate that it could be a safety hazard. If the Reference 620 is taken from a cold location (for example outdoors in winter conditions) to a warm, humid location, water vapor may condense on the cold surfaces inside the Reference 620, possibly creating a hazardous condition. The “reinforced insulation” that keeps the operator from accessing the “hazardous live” voltages in the Reference 620 can be rendered ineffective if the Reference 620 has condensed water inside its case. Before connecting power to a “cold” Reference 620, allow at least one hour for the Reference 620 to warm at room temperature.
Physical Location
Normally you place your Reference 620 on a flat workbench surface. You need access to the rear of the instrument because some cable connections are made from the rear. The Reference 620 is generally operated in an upright position (see Figure 4-1). Operation in other positions is possible as long as you insure that air movement through the chassis is not restricted.
Caution: Do not block the airflow into or out of the Reference 620 chassis. While the
circuitry should suffer no damage due to excessive heat, the Reference 620 enclosure may become uncomfortably hot to the touch if no air flows through the chassis. Running the Reference 620 without adequate cooling could shorten the time to failure of some of the circuitry.
If you place your Reference 620 within an enclosed space, make sure that the internal temperature within that space does not exceed the 45C ambient temperature limit of the Reference 620. Be particularly careful if a computer or other heat-dissipating equipment is mounted in the same enclosure as the Reference 620. The Reference 620 has not been designed for outdoor use.
Computer Requirements
Before you connect a Reference 620 to a host computer, you must ensure that your computer meets these simple requirements:
A computer based on one of the x86family ofIntelTM microprocessors or a 100%-compatible processor from another vendor.
Microsoft Windows® 10 or newer is required with Gamry Framework software version 7.9.1 and above. Only the 64-bit version of these operating systems is supported.
A USB port that supports Full Speed (12 Mbits/second) or High Speed (480 Mbits/second) USB transfers. It must be compatible with Revision 1.1 or Revision 2.0 of the USB specification.
Gamry’s Windows®-based application software packages may impose additional, more stringent requirements. See the software documentation or contact a Gamry representative for details.
26

Quick-start Guide for System Installation
Your shipment includes a short document entitled Quick-start Installation Guide­USB Potentiostat. It contains the latest instructions for installing Gamry hardware and software onto a computer system. If this document is missing, the following information should be sufficient for you to install the Gamry Framework Software and Gamry potentiostat onto your computer.
Software Installation
The Reference 620 is compatible with the Windows® Plug & Play configuration system. Like most Plug & Play hardware, it is best if you install the software for the Reference 620 before you install the potentiostat hardware. The latest Gamry software is not provided on a DVD anymore but is now available for download as *.exe or *.iso file on Gamry Instruments’ Client Portal after creating an account and registering your instrument: https://www.gamry.com/client-portal/my-account/ An internet download of Gamry Instruments’ software will download a self-extracting file. Running this file will extract the software and begin the installation.
Reboot your Computer after Software Installation
Reboot your computer after the Gamry Setup program is done. The Setup program normally offers you the opportunity to do so. USB device drivers are usually loaded when Windows boots up. Following Setup, you may not be able to use your Reference 620 until the drivers are loaded.
Device-driver installation may not occur until a while after the Windows® Desktop appears. On a slow computer, or a busy computer with lots of active applications, the delay before driver installation can be a minute or more.
27

Figure 4-2 Rear Panel of the Reference 620
Power Cord and Power Connection
The Reference 620 does not plug directly in the AC mains supply. Instead, the mains are connected to an external power supply, which supplies a regulated 24-V DC output. This regulated DC is then connected to the DC Power In jack on the rear of the Reference 620 (see Figure 4-2). The external power supply provided with the Reference 620 is rated for operation from 100 to 240 V AC, at frequencies from 47 to 63 Hz. It should be usable worldwide.
Caution: If your facility owns both Reference 620 and Reference 3000, you must insure
that the smaller power adapter from the Reference 620 is not used to power a Reference 3000. The Reference 3000 will not power up with the smaller adapter. Fortunately, neither the Reference 3000 nor the small power adapter will be damaged if connected in error.
28

The Reference 620 external power supply is delivered with a line cord suitable for use in the United States. In other countries, you may have to replace the line cord with one suitable for your type of electrical outlet. You must always use a line cord with a CEE 22 Standard V (IEC 320 C13) female connector on the apparatus end of the cable. This is the same connector used on the US standard line cord supplied with your Reference 620. See Chapter 1 for specific safety information regarding line cord selection.
Power-up Test
Before you make any other connections to your Reference 620, check that the Reference 620 is at least nominally functional. After connecting DC power to the Reference 620, turn on the Power switch on the rear panel of the Reference 620 (see Figure 4-2). One quick test is to power up the Reference 620 and watch the blue Power LED indicator on the front panel of the Reference 620 (see Figure 4-1). The Power LED should illuminate for a second or so, flash three times, then remain on. If this sequence doesn’t happen, see Appendix E for a list of blink codes. The status of the other LED indicators is not important at this time. Each flash of the Power LED upon startup corresponds to successful conclusion of one portion of the power-up self-test.
Caution: If the Power LED goes on, then turns off and stays off, the Reference 620 is
not working properly! Contact Gamry Instruments or your local Gamry Instruments representative as soon as possible if this power-up test fails.
USB Cables
The Reference 620 connects to the computer using a standard High Speed USB A/B cable. A suitable cable was shipped with your Reference 620. If this cable is lost, you can get a replacement at almost any computer retailer. The replacement cable should be rated for USB 2.0 High Speed USB operation. An A/B USB cable has different connectors on each end. The end with a wider, rectangular connector plugs into a USB port on your computer (or a similar port on a USB hub). The end with a nearly-square connector plugs into the USB port on the Reference 620 (see Figure 4-2). The USB connection can be “hot-plugged”. This means that both the computer and the Reference 620 can be powered up before the USB cable is plugged in. Unlike many other instrument system connections, you need not power down the system before plugging in the USB. You may also safely remove the USB cable without powering down the Reference 620 and your computer. Be aware however, that this may have undesirable consequences if the system is currently taking data or performing an electrochemical experiment.
29

Front Panel USB LED
The front panel USB LED provides a simple test of two aspects of normal Reference 620 USB operation. It has three normal states:

Unlit

The USB cable is disconnected, or the USB connection is disabled by the host computer.

Continuous green

A valid cable connection has been made and the Reference USB processor is receiving power from the USB cable.

Flashing yellow Valid USB messages are being transferred between the computer and the Reference 620.

Continuous red

A software download is in progress, or a USB communications error. Do not turn off the Reference 620.

The flashing state only occurs when Gamry Instruments application software is running. The firmware update process is described later in this chapter.

Cell Cable Installation
The Cell Cable connector is a 25-pin female D-type connector on the front of the Reference 620.
All of Gamry’s standard cell cables have a 25-pin D-type connector on one end and a number of leads terminated with banana plugs on the other. The D-type connector end of the cable is connected to the Cell Cable connector on the front of the Reference 620. Always use the knurled screws on this cable to hold the cable in place.
A wide variety of cell cables are available for the Reference 620. These include unshielded cables designed for maximum AC performance, shielded cables in various lengths, and special cables for use in EIS systems where very low impedances must be measured. The Reference 620 can automatically detect which Gamry Instruments cable is connected and the Gamry Framework software can then adjust the system performance for the characteristics of that cable.

Multiple Potentiostat Systems
Gamry’s current Framework software (Revision 7.9) allows a computer to operate up to 16 Gamry Instruments potentiostats simultaneously. The 16 potentiostats can include Reference-family and Interface-family devices with USB connections.
A system with multiple Reference 620 potentiostats just needs them all plugged into the computer.

We do not recommend using bus-powered hubs to expand your USB network: an externally powered USB hub is required. Suitable hubs are available at most computer retailers. Contact our main office or your local Gamry Instruments representative if you need assistance configuring a system containing Reference 620 hardware.

30

First-time Device Installation in Windows®

These instructions presume you have already installed Gamry software Revision 7.9 or higher.

Running the Framework
Regardless of your electrochemical application, Gamry recommends running the Gamry Framework after you install new Framework software or add a potentiostat to your system. The Framework Instrument Manager allows you to:
Rename potentiostats, Calibrate potentiostats, Manage potentiostat firmware, Authorize specific applications for use with specific potentiostats Run the Gamry Framework by clicking on the icon it installed on the Windows® desktop. You can connect and power any Gamry potentiostats either before or after you start the Framework.
Framework Device Status Bar
By default, the Gamry Framework shows a Device Status Bar under its main menu (see Figure 4-3). If you don’t see the Device Status Bar when you run the Gamry Framework, it has been disabled in the Framework Options menu. Potentiostat Devices (instruments) that are connected to the computer appear on this bar. The round indicator associated with each device shows its status:

Green The device is available to run experiments.

Orange The device is currently running an experiment.

White

The device is connected to the system but is not usable. This is generally the result of a mismatch between the Framework software and the device’s firmware. You can use the Gamry Instrument Manager to fix the mismatch.

The screen capture below shows a Framework screen with three USB instruments connected.

31

Figure 4-3 Framework with Three Potentiostats and One Running Test
The Interface 1000 (IFC 01004) in this system is shown with a green indicator because it is installed and ready to run. The Reference 600 labeled My Ref600 has a yellow indicator because it is recording the EIS spectrum shown on the screen. The Reference 600 labeled Jims Ref600 has a white indicator, showing it is plugged in but cannot be used. This is an indication of obsolete firmware. Though no Reference 620 was used in this example, its status indicator behaves in the same manner described above.
Gamry Instrument Manager
Use Gamry’s Instrument Manager (GIM) application make changes to the configuration of your Reference 620 system. Get access to this dialog box via the Options menu in the Gamry Framework. Use the Instrument Manager to:
Rename potentiostats Delete potentiostats that are not currently connected to the computer Select the order in which potentiostats appear in menus Authorize potentiostat use with applications packages Update firmware within potentiostats Figure 4-4 shows an example of the Gamry Instrument Manager window with a Reference 600.
32

Figure 4-4 Instrument Manager Dialog Box
GIM appears automatically when a new instrument is connected to the computer. You can start GIM anytime by opening Framework and selecting Options > Instrument Manager…. GIM is always in its own window, separate from the Framework window. Each Gamry potentiostat in the system appears in the list on the left. All Gamry Instruments potentiostats that are known to the system are displayed in the Instrument Manger. Select an instrument by clicking on its name. Selecting an instrument that is connected and idle blinks its Power LED. After a moment, your potentiostat should appear next to Devices Present along with a green virtual LED. Repeat for additional potentiostats. Though no Reference 620 was used in this example, its functions apply in the same manner as described in this chapter.
Change Tier button
Use the Change Tierbutton when you upgrade an instrument between Framework revisions or when you buy an instrument upgrade. Do the following:
1. Click the Change Tier button in the Device Info section of GIM. A pop-up box appears asking you if you want to proceed.
2. Click the Yes button. The Authorization Code Entry window appears. This consists of a drop-down Package menu and the Authorization Code field.
3. Choose the package to which you will be upgrading in the Package menu, and then enter the corresponding ten-digit Authorization Code. Gamry Instruments gives you the Authorization Code. Click the OK button.
4. A status box opens and the instrument restarts. When this is complete the instrument is at the new tier. 33

Firmware Update

Your Reference 620 was shipped with the latest version of all its firmware. From time to time, Gamry makes changes to the instrument’s firmware code, and a firmware update is required to make use of the new or improved code.

There are two firmware images that you can update on your Reference 620.

Instrument Firmware Handles most of the functions of the Reference 620.

Communications Firmware

Handles the USB communications between your Reference 620 and the host computer

Appropriate update files (firmware images) can be obtained from the Gamry Instruments website at www.gamry.com. Download the file containing the new image and save it onto your computer’s hard disk. Alternatively, every Gamry Software disk (or Gamry Software Flash Drive) contains a firmware folder that contains firmware files compatible with that DVD’s Framework revision.

GIM automatically determines if the software and firmware are compatible. The status of compatibility is shown in the Device Info area.

If there is a conflict, this area shows the following (here using a Reference 3000 as an example):

If the firmware is compatible with software version, the Device Info area appears as:

Auto Update All
If no updates are required to use the instrument, this tool is grayed out. When one or more of the firmware revisions are incompatible, the Auto Update All button is active. This tool automatically selects the compatible firmware and installs them. Gamry Instruments recommends that you use this button. To use this tool, take the following steps:
34

Caution: Do not power off or disconnect the instrument during this process. Doing so
can cause irreversible damage to the instrument. Click the Auto Update All button. A window appears giving information on what version of firmware is currently installed, and what version will be installed. The install process starts automatically. After all the relevant firmware are updated, a window appears confirming the process is complete and successful. Click the OK button. The instrument is now ready to be used.
Select And Update
The Select And Update button is available at all times, even if an update is not required. Gamry recommends not altering the firmware if it is not necessary. This tool allows you to change individual firmware (instrument, PLD, Comm).
Please contact Gamry Tech Support if you are unsure if firmware modification is required.
Blink Button
Clicking the Blink button causes the Power LED on the selected instrument to blink five times. This may help you identify the specific instrument chosen in the software from among a bank of similar looking potentiostats.
Changing the Label
If you purchase additional Gamry Instruments application software, you can do so using the Framework software’s Instrument Manager. The dialog box also allows you to change the label for your device:
1) In Framework, choose Options.
A drop-down menu appears. 2) Choose Instrument Manager…
The Instrument Manager window appears. 3) Click the instrument whose label you wish to change, to highlight that instrument.
35

Click the label itself, and then enter a new, unique name. Press the green checkmark to confirm the new name.
Introduction to the Reference 620 Customization Label
Battery tests are a good example of when multiple tests of a particular battery composition and/or construction are required. Multiple potentiostat systems are often used to speed up the testing process. If the multiple potentiostats can be randomly located in the lab, potentiostat identification can be a problem. You may need to know “Is the potentiostat to the left of my host computer the second or third potentiostat in the system? The Reference 620 includes a customization label that makes each unit visually unique. This allows you to know exactly which potentiostat will be used for each test. You can easily place other labels in the instrument’s Customization Label area. In a system containing eight potentiostats, the potentiostats might be labeled Pstat 1, Pstat 2, Pstat 3, … Pstat 8. Insert simple paper labels in an area behind the clear plastic outer layer of the front panel. The paper is behind the plastic, so it is not affected by the lab environment. Best of all, you can edit or print the labels so personal labels, like Bob’s Pstat, are also possible.
Label Sheet Provided with Each Reference 620
Every Reference 620 is shipped with a pre-printed label sheet containing: 1) Instructions on how to change the label 2) The first eight letters of the Greek alphabet (alpha through theta) 3) The eight planets in the solar system (Mercury through Neptune) 4) Several blank white labels 5) Pstat 1 through Pstat 16 on a blue background 6) Pstat 1 through Pstat 16 on a red background
You can cut any of these labels from the sheet and insert as the Reference 620 Customization Label. The white labels are provided to allow you to handwrite a label.
36

Procedure to Change a Label
If you are making a custom-printed label, we recommend you edit the Excel® file available on Gamry.com. Once you are happy with the edits, print a label sheet as described above. If you are using the Gamry-supplied label sheet, identify the location on the sheet where the new label is found.
1) If you are writing by hand on a blank label, do it now. 2) Use scissors to cut out the new label. Cut white labels on the black lines. 3) Unscrew the four Philips-head screws on the front bezel and remove the instrument’s front bezel (the
black frame around the front panel). See Figure 4-5. Figure 4-5
Removing the Front Bezel
4) Your instrument should now look like Figure 4-6 (the front bezel is removed). Notice the white tab on the left side on the instrument. This is part of the old label.
37

Figure 4-6 Removing the Front Bezel
Tab of instrument label extends beyond front panel.
5) Pull gently on the paper tab on the left side of the front panel to remove the existing label. You need to pull out about 5 cm of label.
6) Insert the new label in the same spot as the old label. The text should face the front of the instrument. A small tab of paper extends beyond the instrument’s front panel.
7) Examine the new label’s position in the Front Panel. Adjust the label if needed. 8) Replace the black bezel. Reattach the four screws to fix the bezel in place. 9) Rename the Reference 620 in the Gamry Framework software to match the new label. Use the
Framework Options > Instrument Manager… dialog box. In the resulting dialog box, select the device you want to rename and then select the Device Settings… button.
38

Calibration

Chapter 5: Calibration
Introduction
There are five different types of calibration for the Reference 620 potentiostat: Factory Calibration User: Offset calibration User: Cable calibration User: Low I Range DC calibration User: Calibrate Temperature Measurement
Factory Calibration is a lengthy process that uses specialized fixtures and very high accuracy test equipment. At this time, Factory Calibration cannot be performed at a user’s facility. Factory calibration is traceable to NIST standards. To maintain this traceability, you must return your instrument to a Gamry calibration site every one or two years. Gamry recommends User Calibration of the Reference 620 at least once per year, or when the quality of your data is in question.
Factory Calibration
The Reference 620 is Gamry’s first product where calibration comprehensively minimizes gain and offset errors on all the circuits in the instrument. This is done in a Factory Calibration. The Factory Calibration typically achieves better than 0.01% gain accuracy and offset errors less than 0.01% (of full scale). In addition, Factory Calibration for the Reference 620 already includes AC calibration to remove errors in the signal chain filters and gains. This eliminates problems from earlier Gamry systems, where AC Calibration used a cable that did not have adequate bandwidth for high frequency measurements. If you see a problem with your system that is related to AC calibration, please contact Gamry Instruments’ support for further assistance.
User Calibration in General
All User calibrations can be performed at a user’s facility. User calibrations are initiated through FrameworkTM software, or directly through Gamry Instrument Manager (GIM). There are four User Calibrations applicable to the Reference 620.
Calibrate Instrument (which is a DC Offset calibration) Calibrate Cable Capacitance DC Low I Calibration Calibrate Temperature Measurement Follow these step-by-step instructions to run one of these calibrations: 1. To initiate a User Calibration, first choose the type of calibration you want. 2. Click Experiment > Utilities in Gamry’s Framework software. On the resulting pull down menu,
select the calibration you want. This launches the appropriate calibration routine. 3. The first dialog box in each Calibration allows you to select one instrument in a multi-instrument
system. This selector is irrelevant for a system with only one instrument. 4. The dialog box may also allow for selection of Calibration Type: DC, AC, or Both.
39

Note: For Reference 620 instruments, only DC calibration can be performed. 5. Select OK. 6. The details of what to do next depend on the calibration you have selected. Follow the instructions
given in Framework. See the relevant Calibration Quick Start Guide for additional details. The calibration for the Reference 620 has been divided into three sections: Instrument DC , cable calibration, and Low I DC Calibration. Gain access to all three calibration procedures via the Utility selection on the Framework’s Experiment drop-down menu. The Reference 620 recognizes the type of cable connected to its cell connector. It maintains a separate AC calibration table for each type of cable. The Gamry Framework will not let you use AC calibration data recorded using a 60 cm shielded cable for experiments run using a Twisted Pair cable optimized for lowimpedance EIS measurements.
Calibrate Instrument
Calibrate Instrument is less important for the Reference 620 than it was for older Gamry Potentiostats. Many of the corrections that were previously measured in Calibrate Instrument have been moved into Factory Calibration. Gamry’s engineers analyzed the stability of the circuitry and tried to calibrate all stable circuits with a high accuracy NIST traceable DVM. Some circuits are still subject to drift when environmental conditions change. A common example is the current offset on the 600 pA and 60 pA IE-Converter ranges. Changes in humidity and temperature can create offset current drifts of hundreds of femtoamps (fA) or even a handful of picoamps (pA). The drift is caused small voltages in the I/E circuit and non-infinite PCB (printed circuit board) resistance between these voltages and the Working electrode connection. Both the voltage magnitudes and the PCB resistance depend on time, temperature, and ambient humidity. User Calibration measures DC Offsets in the Reference 620 and corrects for them. It is most useful in eliminating current drift on the sensitive IE Converter ranges. User experiments and Calibrate Instrument run under similar environmental conditions will minimize errors. Calibrate Instrument requests connection of an external Calibration Cell. The 200 Calibration Cell provided with a Reference 620 is shown below.
Figure 5-1 200 Calibration Cell
40

Caution: The Reference 620 calibration calls for an external resistive dummy cell. Your
Reference 620 was shipped with a 200 Calibration Cell, which includes a 200 , 0.02% accurate resistor. After calibration, please place this Calibration Cell in a safe place where you can find it if your unit requires recalibration. If you do need to recalibrate and you cannot find your Calibration Cell, you can perform DC Calibration using a different 200 resistor. Its power-rating is unimportant. Some performance checks in the calibration process may fail if the resistors inaccuracy exceeds 0.2% (0.4 ). Potentiostat calibration is only required infrequently. Recalibrate your Reference 620 under the following circumstances:
1) It is at least one year since your last calibration. 2) Your potentiostat has been serviced. 3) You notice breaks or discontinuities in the data curves recorded with your system. 4) The system is being run in an environment that is very different from the previous operating
environment. For example, if the Reference 620 was calibrated at 15°C and you are now operating it at 30°C, you should recalibrate.
Figure 5-2 200 Calibration Cell with Leads attached for Calibration
Calibrate Cable Capacitance
Gamry’s Cell Cables are build using low leakage, low capacitance coaxial (coax) cables. Small variation in the capacitance of the Working lead coax can cause glitches in the EIS spectra recorded with different, nominally identical cell cables. A glitch is a sharp discontinuity in an EIS spectrum’s Bode or Nyquist plot. This calibration measures the EIS spectrum of a resistive cell on a specific cell cable and compares it to the response expected with an ideal cell cable. It then calculates a correction term that corrects the EIS spectrum of all cells measured with that specific cable. This procedure uses an external resistive 20 k Calibration Cell.
41

Figure 5-3 20 k Calibration Cell
Caution: The Reference 620 cable calibration calls for an external resistive dummy cell.
Your Reference 620 was shipped with a 20 k Calibration Cell including a 20 k, 0.05% accurate resistor. After calibration, please place this Calibration Cell in a safe place where you can find it if your unit requires recalibration. The cable calibration cell is a different cell (20 k) than standard instrument calibration (200 ). Cable calibration is only required infrequently, if you feel that the cable capacitance is a problem for your experiments, especially if you see excessive phase glitches on high-impedance samples.
Procedure to calibrate the cable
1) Connect the Chassis Ground on the back of your potentiostat to a known, good earth ground.
2) Connect the cell cable to the correct color-coded receptacles on the 20 k Calibration Cell. 42

3) Place the Calibration Cell inside the Calibration Shield, close the lid, and connect the black floatingground lead of your cell cable to the Shield’s grounding post. (Note that the photograph below left is of a 200 cell, so is only a general representation of the procedure.)
4) Open Gamry FrameworkTM software. Select Experiment > Named Script…
The Select a script to run window appears. 5) From the list of scripts, choose “calcable.exp”, then click the Open button.
43

The Cable Capacitance Calibration window appears.
6) In the Cable Tag field, enter a unique name for the cable you are calibrating. Choose the desired Action radio button: o To calibrate the cable, choose Cal Cable. o To reset the vales to zero (if, say, the calibration doesn’t work), choose Zero Values. Click the OK button. The Performance Tips window appears.
7) Make sure that all of the tips are true, then click the OK button. The Cell Required window appears. 44

8) Make sure that the correct Calibration Cell is attached, then click the OK button. The calibration runs. The Done window appears when the calibration successfully finished.
9) Click the OK button to confirm completion.
Low I Range DC Calibration
The standard Reference 620 calibration is performed with the cell leads connected to a 200 resistor. During the calibration procedure, DC current-range offsets are recorded with the cell switch turned off. A DC current measurement is made on each of the eleven current ranges in the Reference 620. The measured current on each range is the sum of current contributions from:
The input current of the I/E Converter input amplifier, The input current of the Working sense input amplifier, The input current of the Reference input amplifier, The input current of a Counter Sense input amplifier, and Current leakage in the cell switch. In most real-world experiments, the cell is turned on, and the I/E converter does not measure the last three terms listed above. These currents still exist, but they are generally sourced by the low-impedance counter electrode lead. During Low I DC Calibration, the first two terms are directly measured in order to improve picoampere accuracy of the instrument. The current contributions from each source on the above list are (at most) a few pA, so they are insignificant on all but the most sensitive Reference 620 current ranges: the 60 pA and 600 pA ranges. This effect can cause differences of up to 8 pA between DC current measured with the cell turned off and current measured with the cell turned on. In most cases the difference is smaller: one or two pA Only two Reference 620 applications are sensitive enough that this current-measurement offset causes problems: Physical Electrochemistry (Cyclic Voltammetry for example) with small electrodes
45

EIS on high-impedance samples such as barrier coatings Incorrect DC-current readings in EIS can slow the experiment, because the automatic ranging algorithms in the EIS software can make poor choices regarding ranges, giving incorrect data. This can significantly lengthen the time needed to measure an EIS spectrum. Most corrosion experiments or macro-electrode measurements involve currents much too large to be affected by this difference. The Gamry Framework includes a special “Low I DC Current” calibration procedure that corrects the Reference 620 offsets to minimize this problem. The procedure uses a script that:
1. Asks you to disconnect the reference, counter, and counter sense leads from the calibration cell, 2. Measures the I/E input and Working sense amplifier input currents on the 60 pA and 600 pA ranges, 3. Replaces the 60 pA and 600 pA current range offsets measured in the full DC calibration with these
improved values. The error sources listed above are both time- and temperature-dependent, so we recommend frequent “Low I DC Calibration” — if you need accurate measurement of absolute currents at pA levels. The procedure runs fairly quickly, so daily or weekly calibration ought not be too inconvenient.
The Low I DC Calibration is not a full calibration. You must run a full DC Calibration on your Reference 620 before you run the Low I DC calibration. Remember that the Reference 620 must have a full DC calibration on the same type of cable you use for the Low I DC Calibration.
46

Cell Connections

Chapter 6: Cell Connections

Normal Cell Connections
This section assumes that you are using standard, shielded cell cables. This information does not depend on the length of the cables.
Each Reference 620 in your system was shipped with a standard, shielded cell cable (part number 985-00151). It is a 60 cm complex cable with a 25-pin D-type connector on one end and five banana plugs and one pin jack on the other end.
In some cases, your system may also include a special-purpose cell cable. The special-purpose cell cable includes documentation describing its use.
The 25-pin male end of the standard cell cable connects to the Reference 620’s front panel Cell Cable connector. Always securely screw the cell cable into place so the cable cannot fall off the unit. Disconnection can be disastrous if it occurs during an experiment.
The other end of the cell cable terminates in a number of 2 mm banana plugs and one pin jack. Each termination comes with a removable alligator clip. Table 6-1 identifies each terminal of the cable.
Table 6-1 Cell-cable Terminations: Potentiostat and Galvanostat Modes

Color

Type

Name

Normal Connection

Blue Green White Red Orange Black

Banana plug Banana plug Banana plug Banana plug Banana plug Pin jack

Working Sense Working Electrode Reference Counter Electrode Counter Sense Floating Ground

Connect to working electrode Connect to working electrode Connect to reference electrode Connect to counter electrode Used in ZRA mode; connect to counter electrode Leave open or connect to a Faraday shield

Connect both the blue and green cell leads to the working electrode. The working electrode is the electrode being tested. The blue pin-jack connection senses the voltage of the working electrode. The green workingelectrode connection carries the cell current. The working electrode may be as much as 150 mV above the circuitry ground (floating ground).
Connect the white banana plug to the cell’s reference electrode, such as an SCE or Ag/AgCl reference electrode. The measured cell potential is the potential difference between the blue and white cell connectors. The accessory kit provided with your Reference 620 contains a converter that allows connection of a reference electrode’s pin plug to a cell cable’s banana jack.
Connect the red banana plug to the counter or auxiliary electrode. The counter electrode is usually a large inert metal or graphite electrode. The counter electrode terminal is the output of the Reference 620’s power amplifier.
The orange lead is only used in ZRA mode, where it senses the counter electrode potential (see following section). Automatic switching to ZRA mode is possible if this lead is connected to the counter electrode. If you are not using ZRA mode, this lead may be left open as long as you ensure that it will not short against any other electrode.
The black pin jack is connected on the Reference 620 end to Floating Ground. This is the circuitry ground for the analog circuits in the Reference 620. In most cases, leave this terminal disconnected at the cell end. When you do, take care that its metal contact does not touch any of the other cell connections.

47

If your cell is a typical glass laboratory cell, all of the electrodes are isolated from earth ground. In this case, you may be able to lower noise in your data by connecting the Reference 620’s Floating Ground to an earth ground.
Caution: If any electrode in your cell is at earth ground, never connect the
Reference 620 chassis to earth ground. Autoclaves, stress apparatus, and field measurements may involve earth-grounded electrodes. A binding post on the rear panel of the Reference 620 is provided for this purpose. A water pipe can be a suitable earth ground.
Warning: Make sure that your earth-ground connection is made to a legitimate source
of earth ground. Consult a qualified electrician if you are uncertain how to obtain an earth ground. Connecting the Reference 620 to an incorrect and unsafe voltage can create a safety hazard (see Chapter 1 for details). If you are measuring very small currents, you may find that a metal enclosure completely surrounding your cell (a Faraday shield) significantly lowers measured current-noise. Usually, you connect this Faraday shield to both earth ground and floating ground. The floating ground on the black cell lead is a convenient source of ground. If any electrode in your cell is connected to earth ground, only connect your Faraday shield to the black cell lead (Floating Ground). In either potentiostatic or galvanostatic mode, if you do not have a reference electrode in your cell, the reference lead can be connected to the counter electrode for a two-electrode experiment. The potential reading will be the difference between the counter electrode and the working electrode.
ZRA Mode Cell Connections
The Reference 620 can function as a precision Zero-resistance Ammeter (ZRA), which maintains two metal samples at the same potential and measures the current flow between the samples. It can also measure the potential of the samples versus a reference electrode. The cell cable connections for ZRA mode are shown in Table 6-2. The connections are very similar to those for the potentiostat and galvanostat modes. A second working electrode is substituted for the counter electrode, and the Orange Counter Sense lead must be connected.
48

Color Blue Green White Red Orange Black

Type Banana plug Banana plug Banana plug Banana plug Banana plug Pin jack

Table 6-2 Cell Cable Connections for ZRA Mode

Name

Normal Connection

Working Sense

Connect to metal sample #1

Working Electrode Connect to metal sample #1

Reference

Connect to a reference electrode

Counter Electrode

Connect to metal sample #2

Counter Sense

Connect to metal sample #2

Floating Ground

Leave open or connect to a Faraday shield

The counter sense and the working sense lead are each connected to different metal samples. In the ZRA mode, the Reference 620 is normally programmed to maintain zero volts between these leads. It therefore maintains the two metal samples at the same voltage.
The white banana plug on the cell cable is normally connected to a reference electrode. The potential between this lead and the working sense lead is reported as the cell potential.
If you don’t have a reference electrode in your cell, we recommend that you connect the white reference lead to the working electrode. In theory, the measured potential is exactly zero when this is done. In practice, A/D noise and offset will create a small potential signal with a value very close to zero.

Membrane Cell Connections

The Reference 620 can be used with membrane cells. In this type of cell, a membrane separates two electrolyte solutions. Two reference electrodes are used: one in each electrolyte. Each electrolyte also contains a counter electrode. The Reference 620 controls the potential across the membrane. Table 6-3 shows the cell connections used with a membrane type cell.

Table 6-3 Cell Cable Connections for a Membrane Cell

Color

Type

Name

Normal Connection

Blue

Banana plug Working Sense

Connect to reference electrode #1

Green

Banana plug Working Electrode Connect to counter electrode #1

White

Pin jack

Reference

Connect to reference electrode #2

Red

Banana plug

Counter Electrode

Connect to counter electrode #2

Orange

Banana plug Counter Sense

Leave open (only needed in ZRA mode)

Black

Pin jack

Floating Ground

Leave open or connect to a Faraday shield

Reference electrode #1 and counter electrode #1 must be on one side of the membrane, and reference electrode #2 and counter electrode #2 must be on the other side.

49

Panel Indicators and Connectors

Chapter 7: Panel Indicators and Connectors
Front Panel
The Reference 620 front panel includes one connector and four backlit LED indicators.
Cell Cable Connector
The Cell Cable is a 25-pin D-type connector that is used to connect the Reference 620 to an electrochemical test cell. It is normally used with a Gamry Instruments-supplied cell cable. In addition to the pins used for cell connections, the Reference 620 Cell Connector also uses five pins to read a cell cable ID. Gamry’s software compensates for the cell-cable characteristics for optimal system performance, especially in EIS (Electrochemical Impedance Spectroscopy). The cell connections are discussed at great length in Chapter 6. A pin-out diagram of the two cell cable connectors is in Appendix B.
Power LED
The Power LED is located on the lower left of the Reference 620 front panel. It normally glows a steady blue, when the Reference 620 is turned on and has passed some simple self-tests.
51

Cell On LED
Overload LED
Power LED
USB LED When the Reference 620 is first turned on, the Power LED glows steadily for a second or two, blink three times, and then go to its normal steady blue output. Each blink in this sequence indicates successful completion of a portion of the Power PC’s power-up self-test routine. A Reference 620’s Power LED blinks when that instrument is selected in the Framework’s Instrument Manager. This allows easy identification of a specific instrument in a MultEchem system without looking at the instrument Serial Label on the bottom of the instrument chassis. When the Power LED is off, one of the following is true:
The rear-panel power switch is off. There is no DC +24 V supply connected to the rear panel Power In connector. The external DC power supply has no input power or is malfunctioning. One part of the Power PC’s power-up self-test has failed.
Caution: The Power LED is used both to indicate power status and that power-up tests
have passed. It therefore cannot be relied on as a true power-status indicator. Always unplug the DC Power In connection if you suspect your Reference 620 is malfunctioning.
52

USB LED
The USB LED is located just below the Power LED. It is a tri-color LED, capable of glowing green, orange, or red. The USB LED is unlit when:
The Reference 620 is not powered. The Reference 620 does not have a USB cable plugged into its rear-panel USB port. The computer end of the USB cable is not plugged into a USB port on a computer or hub. The USB cable is not supplying USB power to the Reference 620. The computer has disabled the USB port going to the Reference 620. The USB LED glows a steady green if a valid USB connection is made and the Reference 620’s communication processor is receiving power through the USB cable. The USB LED flashes orange whenever the Reference 620 is receiving or transmitting valid USB messages to or from the host computer. It does not flash if there is USB traffic addressed to other devices on the USB bus, including messages aimed at a different Reference 620. The USB LED indicates a solid RED in one special condition: when a firmware download is taking place.
Caution: Interrupting a firmware update can cause a catastrophic failure of your system.
Do not turn off the Reference 620, do not unplug the USB cable, and do not stop the operation of the host computer when the USB LED is a continuous red color. Do not interrupt a firmware update that is in progress. An incomplete update can render a Reference 620 inoperable until it is returned to Gamry for reprogramming. Should the firmware update be interrupted, contact Gamry before starting the return process.
Cell On LED
The Cell On LED glows yellow whenever the Reference 620 is actively applying voltage or current to the electrochemical cell attached to the Cell Cable. Avoid touching the cell cable leads whenever the Cell On LED is lighted, because the quality of the data collected in your experiment may be compromised.
The Cell On LED does not indicate a dangerous condition when it is lighted. The voltages put out by a Reference 620 are generally considered safe. Still, avoid touching the cell leads when the cell is on. If you need to make changes to your cell leads, do this between experiments, when the Cell On LED is off and the potentiostat is inactive. In a typical experimental sequence, the Cell On LED is off between experiments and during any open-circuit potential measurements. It glows yellow whenever the cell is polarized.
Overload LED
The Overload LED is normally unlighted. When it glows red, this indicates that some circuit in the Reference 620 has exceeded its normal operating limit. Conditions that generate Overloads include:
53

The absolute value of the differential electrometer output voltage (the difference in voltage between the Working and Reference leads) exceeds 10 V. This condition is known as an E Overload.
The control amplifier has lost control of the cell. The absolute value of the cell current may be trying to exceed ±600 mA, or the absolute value of the counter electrode voltage may be trying to exceed ±22 V. Either condition is called a Control Overload.
The absolute value of the cell current has exceeded full scale on the current range presently in use. This condition is known as an I Overload.
Transient (temporary) overloads during an experiment in which the cell voltage or current is being stepped or swept are often normal. In most cases, they do not indicate a system or instrument malfunction. Consider the case of an infinitely fast voltage step into a perfect capacitor. In theory, charging the capacitor requires an infinite current. The current spike seen at each step in a stepped voltage waveform can easily illuminate the Overload LED. The current spike normally decays to near zero before the actual current and voltage readings are taken. Overload indications when the cell is being connected or disconnected are also common, and usually do not indicate a problem. Overloads can also be seen when one of the cell leads is disconnected from the other cell leads, even though the cell is off. Again, this does not indicate a problem. A steadily glowing Overload LED during an experiment most likely indicates a problem is occurring. Possible causes include:
One of the cell leads is disconnected (this is the most common cause), A gas bubble in the cell is blocking one of the electrodes, The potentiostat could be oscillating (see the next chapter).
A glowing red Overload LED does not necessarily indicate a system malfunction. The Overload LED can illuminate when one or more cell leads are disconnected, without indicating a problem with the system. The Overload LED can often illuminate momentarily during a swept or stepped experiment. The only Overload LED indication that definitely points towards a problem is a continuously glowing Overload LED during an experiment.
54

Rear Panel
The rear panel contains one switch and a large number of connectors. See photograph below.
Power In Jack
The Reference 620 derives all its power from a +24 V DC supply connected to the Power In jack. The input current is less than 3 A. Always use the external power supply supplied with your Reference 620 to supply DC power to the instrument. This supply is rated for operation from 100 to 240 V AC, at frequencies from 47 to 63 Hz. It should therefore be useable worldwide. While a Reference 620 may work with other power sources, we cannot guarantee it will work to its full specifications. If you have to use the Reference 620 with a different supply, make sure that the supply is regulated, has an output between 22 and 26 V, and supplies 3 A of load current.
55

Caution: Power input voltages less than 20 V or greater than 32 V can damage the
Reference 620’s power supply.
Chassis Ground Power In jack
Power Switch
Power Switch
The Power switch is located just below the Power In jack. It switches the power from this jack to the input of the Reference 620’s DC-DC converter. All protection circuitry, including the in-rush current limiter, are located after this switch. Normally, the DC power is connected before the Power switch is turned ON. However, no damage will occur if this switch is already in the ON position when the DC power is connected, or when the AC power input is connected to the external power supply.
56

Chassis Ground
The rear panel Chassis Ground is intended for one use only. When the Reference 620 is used with cells isolated from earth ground, connecting the Chassis Ground to earth ground may lower the noise measured in the system. Note that the chassis of the Reference 620 is connected to Floating Ground. See Chapter 1 for safety information concerning this connection.
Either a banana plug or the stripped end of a wire can be connected to the Chassis Ground binding post. The other end of the wire is then connected to earth ground.
A black banana-plug-to-banana-plug lead is provided with your Reference 620. You may find it useful when making this earth-ground connection.
USB Port
The USB port on the rear panel of the Reference 620 is a Type B connector as defined in Revision 1.1 and 2.0 of the USB Specification. Use a standard, shielded, Type A/B cable to connect this port to a computer’s USB port or a USB hub (preferably an externally powered hub). The two ends of a Type A/B cable are different: the more-rectangular end plugs into the computer, and the more-square end plugs into the Reference 620.

Thermocouple input

USB port

Misc. I/O connector 57

A suitable USB cable was included with your Reference 620 shipment. If this cable is lost, replace it with a cable from your local computer retailer. The Reference 620 is a High Speed USB 2.0 peripheral, capable of data-transfer at 480 Mbits/second. If it is plugged into a computer port incapable of High Speed operation it will downgrade to USB 1.1 full-speed operation (12 Mbits/second). Obviously, data-transfer speed will be slower if this occurs. The Reference 620 USB port is compatible with Revision 1.1 and 2.0 of the USB specification. It supports the Windows Plug and Play mechanism, including dynamic connect/reconnect. The front-panel USB LED should be green whenever a valid computer to Reference 620 connection has been made and both the computer and Reference 620 are fully powered.
Thermocouple Input
The Reference 620 has a T/C Type K jack for a type K thermocouple. The ISO standard calls for color-coded mini-thermocouple connections. Yellow is the color assigned to type K thermocouples. The mating connector on your thermocouple should therefore be yellow. Possible uses for temperature measurement in an electrochemical test include:
Looking for a temperature rise at end-of-charge on a battery. Measuring ambient temperature prior to a corrosion measurement. Measuring temperature in a cell before making a CV measurement that will be used to calculate
reaction kinetics. Gamry Instruments chose not to provide a thermocouple with the Reference 620. There is simply too much variety in the mechanical design of thermocouple probes. Commercial thermocouples designed for measurement in air, on solid surfaces, and in immersion service are available from a variety of vendors. Make sure you get the proper type of thermocouple as shown on the rear of your Reference 620. The Reference 620 uses a temperature-measurement IC to convert the thermocouple output to a usable voltage t outputs a voltage that is nominally 10 mV per degree Celsius. For the K-Type thermocouple, the Analog Device AD597AR is rated for an accuracy of 4°C. The specified accuracy is only achieved when the Reference 620 is calibrated. The scaling at the A/D converter is ±3 V full-scale, or ±300°C full-scale. The Reference 620 calibration script has an optional section for thermocouple calibration. An ice-water bath and a beaker of boiling water provide convenient standards for a two-point calibration.
Caution: One side of the thermocouple is connected to the Reference 620’s Floating
Ground. An improper connection to the thermocouple input can compromise the Reference 620’s ability to float, and invalidate data collected on earth-grounded cells. A connection to a non-insulated thermocouple immersed in your electrochemical cell can also cause erroneous readings.
Misc. I/O Connector
The Misc. (Miscellaneous) I/O connector is a multipurpose connector. It contains both digital and analog signals used to connect external devices to the Reference 620. All of its signals are isolated from both earth ground and the Reference 620’s Floating Ground. The device connected to this connector establishes a ground reference. This isolation allows the Misc. I/O connector to be connected to earth-grounded apparatus, without compromising the Reference 620’s ground isolation. A full description of this connector can be found in Appendix C of this manual. This appendix includes details such as connector pin-out, output and input voltage-levels, and full signal descriptions.
58

The following list is a short description of the signals in the Misc. I/O Connector and their uses: Sync Out and Sync In signal allow two or more Reference 620s to use one data-acquisition clock. Four digital outputs can be used to turn on external devices under control of an Explain experimental control script. Some of Gamry applications assign three of the digital output to control stirring, flow of de-aeration gas, and formation of mercury drops on a mercury drop electrode. Four digital inputs that can be read in an Explain experimental control script. A 12-bit D/A converter used to set “continuously variable” settings, such as electrode rotation rate on a rotating disk electrode. A 5 V isolated power supply that can provide up to 50 mA of current for external circuitry.
Caution: Floating operation of the Reference 620 can be compromised by improper
connections to the Misc. I/O Connector. We do not recommend use of standard 15-pin shielded cables with this connector. Custom cables with the shield connected to pin 6 of the D-connector are preferred.
I Monitor BNC
The I Monitor BNC connector represents the output of the Reference 620’s current measurement circuit. With the exception of the filtering described below, it is the raw signal. It is high bandwidth on the less-sensitive current ranges. The effective bandwidth of the current signal falls as you reach the nA and pA current ranges. IE Stability capacitors further slow the response. The outer shell of this BNC connector is connected to the Reference 620’s floating ground.
59

I Monitor BNC E Monitor BNC

Aux. In BNC Sig. Gen. Out BNC Ext. Sig. In BNC

Caution: The shell of the I Monitor BNC is connected to the Reference 620’s Floating
Ground. Connection of this BNC to a piece of earth-ground-referenced equipment can compromise the Reference 620’s ability to float, and invalidate data collected on earth-grounded cells. Scaling on this signal is 3 V for the nominal full scale current on the selected current range. Cathodic currents cause a positive output voltage. If the software is auto-ranging the current-range selection, this signal is discontinuous at each range change. The I Monitor BNC connector is lightly filtered using an RLC circuit. It has a bandwidth of approximately 3 MHz when connected to a high-impedance input. This bandwidth is further reduced if a coaxial cable is connected to the BNC. Its output impedance is approximately 200 in parallel with 220 pF.
60

E Monitor BNC
The E Monitor BNC connector is the output of the Reference 620’s differential electrometer circuit. With the exception of the filtering described below, it is a buffered representation of the voltage difference between the white and blue cell cable leads. It has a high bandwidth. The outer shell of the BNC connector is connected to the Reference 620’s floating ground.
Caution: The shell of the E Monitor BNC is connected to the Reference 620’s Floating
Ground. Connection of this BNC to a piece of earth-ground-referenced equipment can compromise the Reference 620’s ability to float, and invalidate data collected on earth-grounded cells. The E Monitor BNC connector is lightly filtered using an RLC circuit. It has a bandwidth of approximately 3 MHz when connected to a high-impedance input. This bandwidth is further reduced if a coaxial cable is connected to the BNC. Its output impedance is approximately 200 in parallel with 220 pF.
Ext. Sig. In BNC
The Ext. Sig. In BNC connector allows you to add a voltage to the Reference 620’s signal generator. This signal is summed with the other signal-generator sources, including the IR DAC, the Scan DAC, and the DDS output. The outer shell of the BNC is connected to the Reference 620’s floating ground.
Caution: The shell of the Ext. Sig. In BNC is connected to the Reference 620’s Floating
Ground. Connection of this BNC to a piece of earth-ground-referenced equipment can compromise the Reference 620’s ability to float, and invalidate data collected on earth-grounded cells. The signal generator output is usually directly connected to the potentiostat’s input. When the cell is turned on in potentiostat mode, the feedback is such that a negative signal-generator output creates a positive differential electrometer signal, which corresponds to a negative working electrode versus reference electrode voltage. The polarity of the Ext. Sig. In signal is inverted at the signal generator’s output. As described above, a negative input signal on this BNC creates a positive change in the working electrode versus reference electrode voltage. The input impedance of this signal is 3 k in parallel with 15 pF.
Sig Gen. Out BNC
The Sig Gen. Out BNC connector allows you to monitor the “signal generator” signal being sent from the Reference 620’s controller board to the potentiostat board. This signal has a high bandwidth. The signal output range is ­15 V to +15 V. The outer shell of the BNC is connected to the Reference 620’s floating ground.
Caution: The shell of the Sig. Gen. Out BNC is connected to the Reference 620’s
Floating Ground. Connection of this BNC to a piece of earth-ground-referenced equipment can compromise the Reference 620’s ability to float, and invalidate data collected on earth-grounded cells.
61

The Sig. Gen. Out BNC connector is lightly filtered using an RLC circuit. It has a bandwidth of approximately 3 MHz when connected to a high-impedance input. This bandwidth will be further reduced if a coaxial cable is connected to the BNC. Its output impedance is approximately 200 in parallel with 220 pF.
Aux. In BNC
The Aux. In BNC connector allows you to measure a voltage from outside the Reference 620 using the Reference 620’s internal A/D. The scaling is 3 V in equals 30 000 A/D counts. This is a resolution of 100 µV per bit. The results are reported in volts. The input is differential (see Appendix D).
The allowed input voltage range is 24 V.

circuitry.

Caution: Voltages outside the Aux. In range of ±5 V could damage the Reference 620

Consult Appendix D for additional information concerning this connector.

62

Stability in Potentiostat Mode

Chapter 8: Stability in Potentiostat Mode
Capacitive Cells and Stability
All potentiostats can become unstable when connected to capacitive cells. The capacitive cell adds phase-shift to the potentiostat’s feedback signal (which is already phase-shifted). The additional phase-shift can convert the potentiostat’s power amplifier into a power oscillator. To make matters worse, almost all electrochemical cells are capacitive because an electrical double-layer forms next to a conductor immersed in a solution. Potentiostat oscillation is an AC phenomenon. However, it can affect both AC and DC measurements. Oscillation often causes excessive noise or sharp DC shifts in the system’s graphical output. The Reference 620 potentiostat can be stable on less-sensitive current ranges and unstable on more-sensitive current ranges. Whenever you see sharp breaks in the current recorded on the system, you should suspect oscillation. The Reference 620 has been tested for stability with cell capacitors between 10 pF and 0.1 F. In all but its fastest control amp speed setting, it is stable on any capacitor in this range–as long as the impedance in the reference-electrode lead does not exceed 20 k. With reference-electrode impedances greater than 20 k the Reference 620 may oscillate. The RC filter formed by the reference-electrode impedance and the reference terminal’s input capacitance filters out the high-frequency feedback needed for potentiostat stability. Longer cell cables make the problem worse by increasing the reference terminal’s effective input capacitance. Even when the system is stable (not oscillating), it may exhibit ringing whenever there is a voltage step applied to the cell. The Reference 620’s D/A converters routinely apply steps, even when making a pseudo-linear ramp. While this ringing is not a problem with slow DC measurements, it can interfere with faster measurements. The steps taken to eliminate potentiostat oscillation also help to minimize ringing.
Improving Potentiostat Stability
There are a number of things that you can do to improve an unstable or marginally stable Reference 620 potentiostat/cell system. This list is not in any particular order. Any or all of these steps may help.
Slow down the potentiostat. The Reference 620 has five control-amplifier speed settings, which you select in the software. Slower settings are generally more stable.
Increase the Reference 620’s I/E stability setting. The Reference 620 includes three capacitors that can be placed in parallel with its I/E converter resistors. These capacitors are connected to relays that are under software control. Contact your local Gamry Instruments’ representative for more information concerning changes in these settings.
Lower the reference-electrode impedance. Make sure that you don’t have a clogged referenceelectrode junction. Avoid asbestos-fiber reference electrodes and double-junction electrodes. Avoid small-diameter Luggin capillaries. If you do have a Luggin capillary, make sure that the capillary’s contents are as conductive as possible.
Add a capacitively-coupled low-impedance reference element in parallel with your existing reference electrode. The classic fast combination reference electrode is a platinum wire and a junction-isolated SCE, see Figure 8-1. The capacitor insures that DC potential comes from the SCE, and AC potential from the platinum wire. The capacitor value is generally determined by trial and error.
63

Figure 8-1 Fast Combination Reference Electrode

White Cell Lead

100 pF to 10 nF

SCE

Platinum

Electrolyte

Provide a high-frequency shunt around the cell. A small capacitor between the red and white cell leads allows high-frequency feedback to bypass the cell, see Figure 8-2. The capacitor’s value is generally determined by trial and error. 1 nF (1000 pF) is a good starting point.
In a sense, this is another form of an AC-coupled low-impedance reference electrode. The counter electrode is the low-impedance electrode, eliminating the need for an additional electrode in the solution.
Figure 8-2 High Frequency Shunt

Red

100 pF to 10 nF

White

Green

Working

Reference

Counter

Add resistance to the counter-electrode lead, see Figure 8-3. This change lowers the effective bandwidth of the control amplifier. As a rule of thumb, select the resistor to give one volt of drop at the highest current expected in the test being run. For example, if you expect your highest current to be around 1 mA, you can add a 1 k resistor. This resistor has little or no effect on the DC accuracy of the potentiostat. It can create problems in high-speed experiments such as fast CV scans or EIS, which need high bandwidth.
64

Red White Green

Figure 8-3 Resistor Added for Stability

Resistor

Working

Reference

Counter

65

Measurement of Small Signals

Chapter 9: Measurement of Small Signals

Overview
The Reference 620 is a very sensitive scientific instrument. It can theoretically resolve current changes as small as 1 femtoampere (1 × 10­15 A). To place this current in perspective, 1 fA represents the flow of about 6400 electrons per second!
The small currents measured by the Reference 620 place demands on the instrument, the cell, the cables and the experimenter. Many of the techniques used in higher current electrochemistry must be modified when used to measure pA currents. In many cases, the basic physics of the measurement must be considered.
This chapter discusses the limiting factors controlling low-current measurements. It includes hints on cell and system design. The emphasis is on EIS (Electrochemical Impedance Spectroscopy), a highly demanding application for the Reference 620.

Measurement System Model and Physical Limitations

To get a feel for the physical limits implied by very sensitive current measurements, consider the equivalent circuit shown in Figure 9-1. We are attempting to measure the cell impedance given by Zcell.
This model is valid for analysis purposes even though the real Reference 620 circuit topology differs significantly.

Definition of parameters used in Figure 9-1:

ES

ideal signal source

Zcell

unknown cell impedance

Icell

“real” cell current

Rm

current measurement circuit’s current-measurement resistance

Rshunt unwanted resistance across the cell

Cshunt unwanted capacitance across the cell

Cin

current-measurement circuit’s stray input capacitance

Rin

current-measurement circuit’s stray input resistance

Iin

measurement circuit’s input current

In the ideal current-measurement circuit, Rin is infinite while Cin and Iin are zero. All of the cell current, Icell, flows through Rm.

With an ideal cell and voltage source, Rshunt is infinite and Cshunt is zero. All the current flowing into the current measurement circuit comes from Zcell.

The voltage developed across Rm is measured by the meter as Vm. Given the idealities discussed above, use Kirchhoff’s and Ohm’s laws to calculate Zcell:

Zcell = ES × Rm / Vm

67

Figure 9-1 Equivalent Measurement Circuit
R shunt
C shunt

Icell Rm

R in

C in

Unfortunately, technology limits high-impedance measurements because: Current measurement circuits always have non-zero input capacitance, i.e., Cin > 0. Infinite Rin cannot be achieved with real circuits and materials. Amplifiers used in the meter have input currents, i.e., Iin > 0. The cell and the potentiostat create both a non-zero Cshunt and a finite Rshunt.
Additionally, basic physics limits high-impedance measurements via Johnson noise, which is the inherent noise in a resistance.
Johnson Noise in Zcell
Johnson noise across a resistor represents a fundamental physical limitation. Resistors, regardless of composition, demonstrate a minimum noise for both current and voltage, per the following equations:
E = (4kTR F)1/2 I = (4kT F / R)1/2 where: k = Boltzmann’s constant 1.38 × 10­23 J/K T = temperature in K F = noise bandwidth in Hz R = resistance in . For purposes of approximation, the noise bandwidth, F, is equal to the measurement frequency. Assume a 1011 resistor as Zcell. At 300 K and a measurement frequency of 1 Hz this gives a voltage noise of 41 V rms. The peak-to-peak noise is about five times the rms noise. Under these conditions, you can make a voltage measurement of 10 mV across Zcell with an error of about 0.4%. Fortunately, an AC measurement can reduce the bandwidth by integrating the measured value at the expense of additional measurement time. With a noise bandwidth of 1 mHz, the voltage noise falls to about 1.3 V rms.
68

Current noise on the same resistor under the same conditions is 0.41 fA. To place this number in perspective, a 10 mV signal across this same resistor will generate a current of 100 fA, or again an error of up to 0.4%. Reducing the bandwidth helps. At a noise bandwidth of 1 mHz, the current noise falls to 0.013 fA. With Es at 10 mV, an EIS system that measures 1011 at 1 Hz is about 2½ decades away from the Johnsonnoise limits. At 10 Hz, the system is close enough to the Johnson-noise limits to make accurate measurements impossible. Between these limits, readings get progressively less accurate as the frequency increases. In practice, EIS measurements usually cannot be made at high-enough frequencies that Johnson noise is the dominant noise source. If Johnson noise is a problem, averaging reduces the noise bandwidth, thereby reducing the noise at a cost of lengthening the experiment.
Finite Input Capacitance
Cin in Figure 9-1 represents unavoidable capacitances that always arise in real circuits. Cin shunts Rm, draining off higher-frequency signals, limiting the bandwidth that can be achieved for a given value of Rm. This calculation shows at which frequencies the effect becomes significant. The frequency-limit of a current measurement (defined by the frequency where the phase error hits 45°) can be calculated from:
fRC = 1/(2f RmCin) Decreasing Rm increases this frequency. However, large Rm values are desirable to minimize the effects of voltage drift and voltage noise in the I/E converter’s amplifiers. A reasonable value for Cin in a practical, computer-controllable, low-current measurement circuit is 20 pF. For a 6 nA full-scale current range, a practical estimate for Rm is 107 .
fRC = 1/6.28 (1 × 107)(2 × 10­12) 8000 Hz In general, stay two decades below fRC to keep phase-shift below one degree. The uncorrected upper frequency-limit on a 6 nA range is therefore around 80 Hz. You can measure higher frequencies using the higher-current ranges (i.e., lower-impedance ranges) but this reduces the total available signal below the resolution limits of the “voltmeter”. This exercise forms one basis to the statement that high-frequency and high-impedance measurements are mutually exclusive. You can use software correction of the measured response to improve the useable bandwidth, but not by more than an order of magnitude in frequency.
Leakage Currents and Input Impedance
In Figure 9-1, both Rin and Iin affect the accuracy of current measurements. The magnitude error caused by Rin is calculated via:
Error = 1 ­ Rin/(Rm+ Rin) For an Rm of 107 , an error < 1% demands that Rin must be greater than 109 . PC-board leakage, relay leakage, and measurement-device characteristics lower Rin below the desired value of infinity. A similar problem is the finite input-leakage current Iin into the voltage-measuring circuit. It can be leakage directly into the input of the voltage meter, or leakage from a voltage source (such as a power supply) through an insulation resistance into the input. If an insulator connected to the input has a resistance of 1012 between +15 V and the input, the leakage current is 15 pA. Fortunately, most sources of leakage current are DC and can be tuned out in impedance measurements. As a general rule, the DC leakage should not exceed the measured AC signal by more than a factor of 10. The Reference 620 uses an input amplifier with an input current of around 1 pA. Other circuit components may also contribute leakage currents. You therefore cannot make absolute current measurements of very low pA currents with the Reference 620. In practice, the input current is approximately constant, so current differences or AC current levels of less than one pA can usually be measured.
69

Voltage Noise and DC Measurements
Often the current signal measured by a potentiostat shows noise that is not the fault of the currentmeasurement circuits. This is especially true when you are making DC measurements. The cause of the current noise is noise in the voltage applied to the cell. Assume that you have a working electrode with a capacitance of 40 µF. This could be represented by a 1 cm2 polished bare metal immersed in an electrolyte solution. A rough estimate of the capacitance of the electrical double-layer formed by a metal/electrolyte interface is 20 µF/cm2. The area is the microscopic area of the surface, which is larger than the macroscopic geometric area, because even a polished surface is rough. The impedance of this 40 F electrode, assuming ideal capacitive behavior, is given by:
Z = 1/j C At 60 Hz, the impedance magnitude is about 66 . Apply an ideal DC potential across this ideal capacitor and you get no DC current. Unfortunately, all potentiostats have noise in the applied voltage. This noise comes from the instrument itself and from external sources. In many cases, the predominant noise frequency is the AC power-line (mains) frequency. Assume a realistic noise voltage, Vn, of 10 µV (this is lower than the noise-level of most commercial potentiostats). Further, assume that this noise voltage is at the North American power-line (mains) frequency of 60 Hz. The noise voltage creates a current across the cell capacitance:
I = Vn/Z 10 × 10­6/66 150 nA This rather large noise current prevents accurate DC current measurement in the low nA or pA ranges. In an EIS measurement, you apply an AC excitation voltage that is much bigger than the typical noise voltage, so this is not a significant problem.
Shunt Resistance and Capacitance
Non-ideal shunt resistance and capacitance arise in both the cell and the potentiostat. Both can cause significant measurement errors. Parallel metal surfaces form a capacitor. The capacitance rises as the metal’s area increases and as the separation distance between the metals decreases. Wire and electrode placement have a large effect on shunt capacitance. If the clip leads connecting to the working and reference electrodes are close together, they can form a significant shunt capacitor. Values of 1 to 10 pF are common. This shunt capacitance cannot be distinguished from “real” capacitance in the cell. If you are measuring a paint film with a 100 pF capacitance, 5 pF of shunt capacitance is a very significant error. Shunt resistance in the cell arises because of imperfect insulators. No material is a perfect insulator (infinite resistance). Even PTFE, which is one of the best insulators known, has a bulk resistivity of about 1012 ·m. Worse yet, surface contamination often lowers the effective resistivity of good insulators. Water films can be a real problem, especially on glass. Shunt capacitance and resistance also occur in the potentiostat itself. The Reference 620 Potentiostat mode specifications in Appendix A contain equivalent values for the potentiostat’s Rshunt and Cshunt. These values can be measured by an impedance measurement with no cell. In most cases, the cell’s shunt resistance and capacitance errors are larger than those from the potentiostat.
Hints for System and Cell Design
The following hints may prove helpful.
70

Faraday Shield
A Faraday shield surrounding your cell is mandatory for very low-level measurements. It reduces both current noise picked up directly on the working electrode and voltage noise picked up by the reference electrode. A Faraday shield is a conductive enclosure that surrounds the cell. The shield can be constructed from sheet metal, fine-mesh wire screen, or even conductive paint on plastic. It must be continuous and completely surround the cell. Don’t forget the areas above and below the cell. All parts of the shield must be electrically connected. You will need an opening in the shield large enough to allow a cell cable to enter the shield. The shield must be electrically connected to the Reference 620’s floating-ground terminal. An additional connection of both the shield and the Reference 620 floating ground to an earth ground may also prove helpful.
Only connect the Reference 620 ground to earth ground if all conductive cell components are well-isolated from earth ground. A glass cell is usually well isolated. An autoclave is generally not well isolated.
Avoid External Noise Sources
Try to keep your system away from electrical noise sources. Some of the worst are: Fluorescent lights Motors Radio transmitters Computers and computer monitors
Try to avoid AC powered or computerized apparatus within your Faraday shield.
Cell Cable Length and Construction
The Reference 620 is shipped with a 60 cm shielded cell cable. We also offer extended length cables and unshielded cables as options at extra cost. Cell cables longer than 1 meter will result in degraded instrument performance. Increased noise and decreased stability both can occur. However, with most cells, the instrument will work acceptably with an extended cell cable, so our advice is to try it. As a rule, do not attempt to use current-interrupt IR-compensation with cell cables longer than 5 m. We do not recommend that you use the Reference 620 with any cables not supplied by Gamry Instruments. The Reference 620 cable is not a simple cable like a typical computer cable. The Reference 620 cable includes a number of individually shielded wires contained within an overall shield. We pay careful attention to issues such as shield isolation, isolation resistance, and capacitance. If you do need a special cable, contact us with your requirements.
Lead Placement
Many experiments with the Reference 620 involve cells with small capacitances, the value of which may be important. In these cases, the capacitance between the Reference 620’s cell leads can result in an error. The Reference 620 alligator clips can have 10 pF or more of mutual capacitance if they are run alongside each other. If you wish to avoid excessive capacitance:
71

Place the leads as far apart as possible. Pay special attention to physical separation between the working electrode/working sense leads and the counter/counter sense/reference electrode leads.
Have the leads approach the cell from different directions. Remove the alligator clips from the leads. In extreme cases you can replace the banana plugs and pin
jack with smaller connectors. If you do so, be careful not to compromise the isolation between the center conductor and the shield. The cell leads must not be moved during an experiment measuring small currents. Both microphonic and triboelectric effects can create spurious results when the cell cables are moved.
Cell Construction
If you need to measure small currents or high impedances, make sure that your cell construction does not limit your response. A cell where the resistance between the electrodes is only 1010 cannot be used to measure 1013 impedances. In general, glass and PTFE are the preferred construction materials for cells. Even glass may be a problem when it is wet. You also must consider Cshunt. Make the “inactive” portion of your electrodes as small as possible. Avoid placing electrodes close together or parallel with each other if you are measuring high impedances.
Reference Electrode
Keep your reference electrode impedance as low as possible. High-impedance reference electrodes can cause potentiostat instability and excessive pickup of voltage-noise. Try to avoid:
Narrow-bore or Vycor