Product Overview

This document describes specific validation procedures for template A-5l data loggers, for the purpose to ensure proper operation and provide a reliable temperature and time record during transportation and storage.

tempmate-S1-Single-Use-Temperature-Logger-FIG-1

Technical Specifications

  • Recording Options: Single-Use
  • Temperature Range: -300C to 70 •c
  • Temperature Accuracy: ±0.5 (-200C/+400C); ±I.O (other range)
  • Temperature Resolution: 0.1 •c
  • Data Storage Capacity: 16.000 readiness
  • Shelf Life / Battery: 2 years / CR2450 button cell
  • Recording Interval: 10 minutes (standard, others on request)
  • Recording Duration: up to 1 0 days (standard, others on request)
  • Startup Mode: Button or software
  • Stop Mode: Button, software, or stop when full
  • Protection Class: IP67 / NEMA 6
  • Dimensions: 83 mm x 47 mm x 7 mm (L x w x H)
  • Weight: 14.6g

Operation Instructions

  1. Keep the button pressed for a minimum of S seconds to start recording. Then the RECORD LED flashes 10 times, which indicates a successful starting. If the STATUS LED is lit for 3 seconds, please do not use the logger (Indicating low battery)! Tear off the serial number label for your reference.tempmate-S1-Single-Use-Temperature-Logger-FIG-2
  2. Press the button once at any time to check the current status of the logger. Double-click the button to set a mark in the reports. See details of these optional features in the tables below.tempmate-S1-Single-Use-Temperature-Logger-FIG-3
  3. Keep the button pressed for a minimum of S seconds to stop recording. Then the STATUS LED flashes 10 times, which indicates a successful stopping.tempmate-S1-Single-Use-Temperature-Logger-FIG-4
  4. After stopping the record, t ear Off the transparent plastic cover and connect the logger to a USB port of the computer. The PDF and CSV reports will be generated automatically and can be found in the device’s .,t empmate· folder.tempmate-S1-Single-Use-Temperature-Logger-FIG-5

Operation Indication

Start/ Stop

  • Status Action LED confirmation
  • Start Logger 5 sec pushing button RECORD LED flashing 10 timestempmate-S1-Single-Use-Temperature-Logger-FIG-6
  • Set Mark double-click button STATUS CED + RECORD LED flashing 5 timestempmate-S1-Single-Use-Temperature-Logger-FIG-7
  • Stop Logger 5-sec pushlng button STATUS LED flashing 10 timestempmate-S1-Single-Use-Temperature-Logger-FIG-8

Status Request by pushing the Button 1 time
Status Action LED confirmation

  • Not Started to push button 1-time STATUS LED + RECORD LED flashing 1 timetempmate-S1-Single-Use-Temperature-Logger-FIG-9
  • Recording – 0K push button 1-time RECORD LED flashing I timetempmate-S1-Single-Use-Temperature-Logger-FIG-10
  • Recording – Alarm push button 1-time STATUS LED flashing I timetempmate-S1-Single-Use-Temperature-Logger-FIG-11
  • Stopped – 0K push button 1-time RECORD LED flashing 2 timestempmate-S1-Single-Use-Temperature-Logger-FIG-12
  • Stopped – Alarm push button 1-time STATUS LED flashing 2 timestempmate-S1-Single-Use-Temperature-Logger-FIG-13