Power Technology IRV3 Infrared Viewing ឧបករណ៍

ការណែនាំអំពីការប្រើប្រាស់ផលិតផល
សេចក្តីផ្តើម៖
- Power Technology’s IR viewers convert otherwise invisible infrared radiation into a visible green image. These devices support both CW and pulsed sources down to sub-microsecond regimes without external synchronization.
- Applications (typical): IR laser alignment; semiconductor wafer inspection (with microscope adapter); photo processing; thermal/industrial inspection; forensics; food sorting; fluid inspection; and biological illumination tasks.
ការដំឡើង៖
- Mounting (both models): Use the ¼-20 tripod interface for hands-free operation.
- Lens/Eyepiece: Install the supplied C-mount objective, then set coarse focus on the objective and fine focus at the eyepiece. (Locking screws as shown in figures.)
- Filters (NIR use): Fit the IR cut-off/attenuation filter for near-IR viewing; when inspecting reflections, prefer a metallic diffuser. Paper tends to absorb/attenuate IR.
ប្រតិបត្តិការ៖
- IRV3: Charge via USB-C at ≤ 5 VDC / 500 mA. (Do not exceed this rating.) An internal Li-ion pack powers the tube; after switching OFF, residual charge may keep the screen lit briefly—this is expected.
- IRX3: Install two 3.6 V 18650 Li-ion cells with correct polarity, close the cover securely, then power via the ACTIVATE button. Observe the LED battery indicator: solid = full, slow-flash = OK, fast-flash = low.
- Caution – Direct exposure: Avoid exposing the tube face to concentrated radiation; permanent sensor damage can occur. Important – These viewers are for indirect beam observation. Do not use for direct beam viewing or to look for“mid-air” beams. View a diffuse target (IR card, matte metal, etc.)—over-exposure can damage the tube or reduce photocathode response.
Operating Procedure & Control Description: IRX3 (Figure 1)
- Power: Insert two 18650 cells, close the cover, and press ACTIVATE.
- Brightness: Use screen brightness buttons to set a comfortable luminance without blooming.
- Focusing & Iris: Rotate the focus ring to sharpness; adjust the diaphragm (iris) for contrast and depth; set eyepiece focus to your vision.
- Filters & Targets: Use the cut-off filter for NIR; prefer metallic diffuse targets over paper.
- Battery Status: LED solid = full; gradual flash = acceptable; rapid flash = low. Auto power-off after inactivity (see data sheet).

- LED
- រង្វង់ផ្តោតភ្នែក
- Button ACTIVATE
- Screen brightness adjustment buttons
- តម្រង
- ដៃ
- គម្របថ្ម
- ខ្សែស្រឡាយជើងកាមេរ៉ា
- Diaphragm adjustment ring
- ចិញ្ចៀនលៃតម្រូវការផ្តោតអារម្មណ៍
- Lens locking screw
IRV3 (Figure 2)
Power/Charge: If needed, connect USB-C (≤ 5 V/500 mA). Press the ON/OFF button to start. Residual screen glow after power-down is normal.
- Focusing & Iris:
- Rotate the objective focus ring to bring the scene to focus (1×: F/1.4 25 mm; 2×: F/1.8 50 mm).
- Adjust the iris to balance brightness vs. depth of field; tighten the iris set screw after adjustment.
- Fine-tune the eyepiece focus to your vision.
- Filters & Targets: For NIR, install the IR cut-off filter. View a diffuse reflector (e.g., IR card or matte metal).
- Tripod Use: Thread onto ¼-20 tripod for stable, hands-free operation. Notes on Visualization: For wavelengths across the IRV model family (IRV1300/1700/2000), detection thresholds vary with wavelength and optical configuration; observing a diffuse target at 1 m typically requires the minimum irradiances indicated in the technical curves.

- USB-C សម្រាប់សាកថ្ម
- ដៃ
- ប៊ូតុងបើក/បិទ
- LED
- ចិញ្ចៀនលៃតម្រូវការផ្តោតអារម្មណ៍
- ខ្សែស្រឡាយជើងកាមេរ៉ា
- តម្រង IR
- Iris adjustment ring
- កែវភ្នែក
Troubleshooting (Both Models)
- Image too dim: Open iris moderately; verify correct filter; increase target reflectivity (matte aluminum card); ensure battery charge (IRX3 LED status).
- Blooming/saturation: Stop down the iris; add neutral density attenuation for very bright sources (IRX3 accessory ND filters).
- Cannot see free-space beam: Use a card/metallic target; do not attempt to “look into” the beam path.
ការថែទាំ និងសេវាកម្ម៖
- These products contain no user-serviceable parts. Keep optics clean and dry.
- Use clean compressed air first; if needed, clean glass carefully with alcohol (isopropyl) on a lint-free swab.
- Avoid solvents on plastics or painted surfaces.
- Handling: Protect from mechanical shock and moisture ingress.
- Battery safety (IRX3): Follow local regulations; keep out of reach of children; never expose cells to heat/flame.
Warranty and Repair Return Policy:
No return will be accepted without a factory-issued RMA number clearly marked on the package. Collect/COD returns are not accepted. For warranty returns, PTI coversoutbound shipment to the customer after repair. When requesting an RMA, provide a model, serial number(s), and a concise problem description. Opening or modifying the viewer outside documented procedures may void the warranty and can require re-certification under 21 CFR 1040.10/.11.
Laser Safety (Read Before Use):
- Use controls and procedures only as specified to avoid hazardous radiation exposure.
- Optical instruments increase eye hazard; never aim sources toward people; avoid eye-level beam paths.
- Always wear appropriate, wavelength-rated protective eyewear for the source under test.
Notes on Performance & Best Practices:
- Target choice dominates visibility: Use matte aluminum/ceramic IR viewing cards or rough metal, not paper, to maximize SNR.
- Keep irradiance within the tube’s comfort zone: If the scene blooms, stop down the iris and/or add ND attenuation (IRX3 accessories include ND filters).
- Free-space beams: Don’t rely on scattering; use projection at 1 m and consult the model’s minimum power-density vs. wavelength curves for realistic expectations.
IRV3 Performance Charts: Spectral Sensitivity & Minimum Power Density
POWER DENSITY
- Approximate minimum power density required to view an infrared laser beam from a distance of one meter:
IRX3 Performance Charts: Minimum Power Density & Brightness Levels
Minimal Power
- densityThreshold power density dependence on wavelength.
- The threshold power density is defined by measuring a laser beam spot on a paper, which exhibits 20% of the overall brightness (calculated a 255 x in contrast to the background.
- The measurements were taken with the viewer positioned 1.15 meters away from the piece of paper.

កម្រិតពន្លឺ
- Normalised brightness dependence on power difference from the minimum value.
- The power level of 0 signifies the theoretical minimal value at which the laser beam spot becomes observable on a piece of paper.
- It’s worth noting that the viewer exhibits lower sensitivity to laser light at 1450nm compared to 1550 nm or even 1900nm.
- Mailing: P.O. Box 191117, Little Rock, AR 72219-1117
- Shipping: 16302 N. Alexander Road, Alexander, AR, 72202
- ទូរស័ព្ទ៖ 501.407.0712
- ទូរសារ៖ 501.407.0036
- អ៊ីមែល៖ Sales@PowerTechnology.com
- Web: www.PowerTechnology.com
សំណួរគេសួរញឹកញាប់
1. Image Too Dim:
If the image is too dim, open the iris moderately, verify correct filter, increase target reflectivity, and ensure battery charge.
2. Blooming/Saturation:
To address blooming/saturation issues, stop down the iris and add neutral density attenuation for very bright sources.
3. Cannot See Free-Space Beam:
Use a card or metallic target and avoid looking directly into the beam path.
4. Maintenance & Service:
These products contain no user-serviceable parts. Keep optics clean and dry, use clean compressed air first for cleaning, and avoid solvents on plastics or painted surfaces.
5. Battery Safety (IRX3):
ollow local regulations, keep batteries out of reach of children, and never expose cells to heat or flame.
6. Warranty and Repair Return Policy:
No return will be accepted without a factory-issued RMA number clearly marked on the package. Collect/COD returns are not accepted.
ឯកសារ/ធនធាន
![]() |