HT4011 Clamp Meter User Manual

Brand: HT ITALIA

Release: EN 1.01 - 21/02/2014

1. Precautions and Safety Measures

The instrument is designed in compliance with directive IEC/EN61010-1 for electronic measuring instruments. For your safety and to prevent instrument damage, carefully follow the procedures in this manual and pay close attention to notes preceded by the ⚠️ symbol.

Before and after measurements, observe the following:

Symbols used in this manual and on the instrument:

1.1. Preliminary Instructions

1.2. During Use

Please read the following recommendations and instructions carefully:

CAUTION: Failure to comply with caution notes may damage the instrument, its components, or endanger the operator.

1.3. After Use

1.4. Definition of Measurement (Overvoltage) Category

Standard "IEC/EN61010-1" defines measurement categories. Circuits are divided as follows:

2. General Description

The instrument performs the following measurements:

Functions are selected via an 8-position rotary switch (including OFF). The instrument also features HOLD, MODE, Hz%, and REL keys. Refer to § 4.2 for their use. The LCD display shows the selected quantity, unit, and enabled functions.

2.1. Measuring Average Values and TRMS Values

Alternating quantity measuring instruments are divided into two families:

For sinusoidal waves, both families yield identical results. For distorted waves, TRMS meters provide higher values as they measure the entire wave including harmonics, while average-value meters measure only the fundamental wave.

2.2. Definition of True Root Mean Square Value and Crest Factor

The root mean square (RMS) value of current is defined as the equivalent DC current that dissipates the same energy in a resistor over a period.

The Crest Factor (CF) is the ratio between the Peak Value and the RMS value of a signal. For a pure sinusoidal wave, CF = √2 ≈ 1.41. Distortion increases the Crest Factor.

3. Preparation for Use

3.1. Initial Checks

The instrument is checked for electrical and mechanical integrity before shipping. While precautions are taken to prevent damage, it is recommended to check for damage incurred during transport. Contact the forwarding agent if anomalies are found. Verify that all components listed in § 6.3 are present; contact the dealer for discrepancies. For returns, follow instructions in § 7.

3.2. Instrument Power Supply

The instrument is powered by two 1.5V AAA LR03 batteries (included). The ? symbol on the display indicates a low battery. Replace batteries following § 5.2. The instrument features an Auto Power OFF function that switches it off approximately 30 minutes after the last operation (cannot be disabled).

3.3. Calibration

The instrument's performance is guaranteed for 12 months, based on the technical specifications in this manual.

3.4. Storage

To ensure measurement precision after long storage under extreme environmental conditions, allow the instrument to return to normal conditions (see § 6.2.1).

4. Operating Instructions

4.1. Instrument Description

4.1.1. Description of the Controls

Fig. 1: Instrument description

4.2. Description of Function Keys

4.2.1. HOLD Key

Pressing the HOLD key freezes the measured value on the display, showing "HOLD". This mode is deactivated by pressing the HOLD key again or operating the switch.

4.2.2. REL Key

When the switch is set to V, V, or AC current positions, the REL key zeros the displayed value for relative measurements. The first press saves the value as an offset, indicated by "REL". Subsequent readings show the current value minus the offset. This function is inactive for resistance, continuity, temperature, capacitance, duty cycle, and diode tests. Press REL again or turn the switch to exit.

4.2.3. Hz% Key

With the rotary switch set to V, V, or Hz, the Hz% key toggles between frequency (Hz) and duty cycle (%) measurements.

4.2.4. MODE Key

The MODE key selects resistance, continuity test (with buzzer), and diode test when the rotary switch is set to Ω. It also selects AC and DC voltage measurements when the switch is set to V, V, or Hz.

4.3. Description of Rotary Switch Functions

4.3.1. DC Voltage Measurement

CAUTION: Maximum input DC voltage is 600Vrms. Exceeding limits can cause electrical shock and instrument damage.

Fig. 2: Use of the clamp for DC voltage measurement

  1. Select the V position.
  2. Insert the red cable into input terminal V ⏚ CAP Hz% Temp and the black cable into input terminal COM (Fig. 2).
  3. Position test leads at the desired points in the circuit. The display shows the voltage value.
  4. The "O.L" symbol indicates overload status.
  5. Refer to § 4.2 for HOLD and REL functions.

CAUTION:

4.3.2. AC Voltage Measurement

CAUTION: Maximum input AC voltage is 600V. Exceeding limits can cause electrical shock and instrument damage.

Fig. 3: Use of the clamp for AC voltage measurement

  1. Switch on the instrument near an AC source. The red LED at the clamp's base (Fig. 1 – Part 2) should light up, indicating AC source presence.
  2. Select the V, V, Hz position.
  3. Insert the red cable into input terminal V ⏚ CAP Hz% Temp and the black cable into input terminal COM (Fig. 3).
  4. Position test leads in the circuit. The display shows the voltage value.
  5. The "O.L" symbol indicates overload status.
  6. Refer to § 4.2 for HOLD and REL functions.

CAUTION:

4.3.3. Frequency and Duty Cycle Measurement

CAUTION:

Fig. 4: Use of the clamp for frequency measurement and duty cycle

  1. Select position V, V, Hz for frequency measurement with leads.
  2. Press the Hz% key to cycle through "Hz" for frequency or "%" for duty cycle measurement.
  3. Insert the red cable into input terminal V ⏚ CAP Hz% Temp and the black cable into input terminal COM (Fig. 4 – left side). The display shows frequency (Hz) or duty cycle (%).
  4. The "O.L" symbol indicates overload status.
  5. Refer to § 4.2 for the HOLD function.

4.3.4. Resistance Measurement

CAUTION: Before measuring resistance, remove power from the circuit and discharge all capacitors.

Fig. 5: Use of the clamp for Resistance measurement

  1. Select the Ω position.
  2. Insert the red cable into input terminal V ⏚ CAP Hz% Temp and the black cable into input terminal COM.
  3. Position test leads in the circuit (Fig. 5). The display shows the resistance value.
  4. The "O.L" symbol indicates overload status.
  5. Refer to § 4.2 for the HOLD function.

4.3.5. Capacitance Measurement

CAUTION: Before measuring capacitance, cut off power supply and discharge all capacitance in the circuit or capacitor.

Fig. 6: Use of the clamp for capacitance measurement

  1. Select the CAP position.
  2. Press the MODE key to cycle until "nF" is displayed.
  3. Insert the red cable into input terminal V ⏚ CAP Hz% Temp and the black cable into input terminal COM.
  4. Position test leads in the circuit (Fig. 6). The display shows the capacitance value. The analog graphic bar is deactivated during capacitance measurement.
  5. The "O.L" symbol indicates overload status.
  6. Refer to § 4.2 for HOLD and REL functions.

4.3.6. Continuity Test and Diode Test

CAUTION: Before measuring resistance, remove power from the circuit and discharge all capacitors.

Fig. 7: Use of the clamp for continuity test and diode test

  1. Select the Ω position.
  2. Press the MODE key to cycle until the "))" symbol is displayed for continuity test.
  3. Insert the red cable into input terminal V ⏚ CAP Hz% Temp and the black cable into input terminal COM. Perform continuity test (Fig. 7 – left side). A buzzer sounds if resistance is below 30Ω.
  4. Press the MODE key to select diode test. The "+" symbol appears.
  5. For direct polarization, connect the red lead to the anode and the black lead to the cathode of the diode (Fig. 7 – right side). For reverse polarization, invert the leads.
  6. Display values between 0.4V and 0.7V (direct) or "OL" (reverse) indicate correct connection. "0mV" indicates a short circuit, while "OL" in both directions indicates an interrupted device.

4.3.7. Temperature Measurement in °C and °F

CAUTION: Do not touch live surfaces with the temperature probe. Voltages higher than 30Vrms or 60VDC imply electrical shock hazard ().

Fig. 8: Use of the clamp for temperature measurement

  1. Select the Temp°C or Temp°F position.
  2. Insert the K wire probe into input terminals V ⏚ CAP Hz% Temp and COM using the adapter, respecting polarity (Fig. 8). The display shows the temperature value.
  3. Refer to § 4.2 for the HOLD function.

4.3.8. AC Current Measurement

CAUTION: Before any measurement, disconnect all test leads from the circuit and meter input terminals.

Fig. 9: Use of the clamp for AC current measurement

  1. Select position 40A or 400A.
  2. Insert the cable into the middle of the clamp jaws for accurate measurements. The display shows the AC current value.
  3. The "O.L" symbol indicates overload status. Switch to a higher range if this occurs.
  4. Refer to § 4.2 for HOLD and PEAK functions.

CAUTION: Displayed values when the instrument is not in measuring mode are normal and not an instrument fault.

5. Maintenance

5.1. General Information

  1. The instrument is a precision device. Observe manual recommendations during use and storage to prevent damage or danger.
  2. Do not use in high humidity or high temperatures. Avoid direct sunlight.
  3. Always switch off the instrument after use. If not used for a long time, remove the battery to prevent leaks that could damage internal circuits.

5.2. Replacing the Battery

When the LCD displays the ? symbol, replace the battery.

CAUTION: Only expert technicians should perform this. Ensure all cables are removed from terminals or the tested cable from the clamp jaw before proceeding.

  1. Turn the rotary switch to OFF.
  2. Disconnect cables from terminals and the tested cable from the clamp jaw.
  3. Loosen the battery cover screw and remove the cover.
  4. Disconnect the battery from the connector.
  5. Connect new batteries, observing polarity (see § 6.1.2).
  6. Replace and fasten the battery cover.
  7. Dispose of old batteries properly in designated containers.

5.3. Cleaning the Instrument

Use a soft, dry cloth. Never use wet cloths, solvents, or water.

5.4. End of Life

CAUTION: The ♻️ symbol indicates that the appliance, accessories, and battery must be collected and disposed of separately and correctly.

6. Technical Specifications

6.1. Technical Characteristics

Accuracy is calculated as ± [% reading + (number of digits x resolution)]. Specifications are valid at 18°C ÷ 28°C with relative humidity <75% RH.

AC Voltage (Autorange)

RangeResolutionAccuracyInput ImpedanceBandwidthOvervoltage Protection
4.000V0.001V±(1.8%rdg + 8digits)10MΩ50-400Hz600V DC/ACrms
40.00V0.01V
400.0V0.1V±(2.5%rdg + 8digits)
600V1V

Integrated sensor for AC voltage detection: LED on for phase-earth voltage > 100V, 50/60Hz.

DC Voltage (Autorange)

RangeResolutionAccuracyInput ImpedanceOvervoltage Protection
400.0mV0.1mV±(0.8%rdg+2digits)10MΩ600AACrms
4.000V0.001V±(1.5%rdg + 2digits)
40.00V0.01V
400.0V0.1V±(2%rdg + 2digits)
600V1V

AC Current

RangeResolutionAccuracyBandwidthOvervoltage Protection
40.00A0.01A±(2.5%rdg + 8digits)50-60Hz400AACrms
400.0A0.1A±(2.8%rdg + 8digits)

Resistance and Continuity Test (Autorange)

RangeResolutionAccuracyBuzzerOvervoltage Protection
400.0Ω0.1Ω±(1.0%rdg + 4digits)<30Ω600VDC/ACrms
4.000kΩ0.001kΩ±(1.5%rdg+2digits)
40.00kΩ0.01kΩ
400.0kΩ0.1kΩ±(2.5%rdg+3digits)
4.000MΩ0.001MΩ±(3.5%rdg+5digits)
40.00MΩ0.01MΩ

Continuity test current: < 0.5mA

Capacitance (Autorange)

RangeResolutionAccuracyOvervoltage Protection
40.00nF0.01nF±(4.0%rdg+20digits)600VDC/ACrms
400.0nF0.1nF±(3%rdg+5digits)
4.000μF0.001μF
4000μF0.01μF±(4.0%rdg+10digits)
100.0μF0.1μF

Diode Test

RangeTest CurrentOpen-circuit Voltage
0.3mA typical1.5VDC

Frequency with Leads (Autorange)

RangeResolutionAccuracySensitivityOvervoltage Protection
10.00Hz - 49.99Hz0.01Hz±(1.5%rdg+2digits)≥15Vrms600VDC/ACrms
50.0Hz - 499.9Hz0.1Hz
0.500kHz - 4.999kHz0,001kHz
5.00kHz - 10.0kHz0.01kHz

100μs ≤ Pulse width ≤ 100ms ; Pulse frequency: 100Hz ÷ 150kHz; Sensitivity >10Vrms

Duty Cycle (Autorange)

RangeResolutionAccuracy
0.5% - 99.0%0.1%±(1.2%rdg + 2digits)

Temperature with K probe (Autorange)

RangeResolutionAccuracy (*)Overvoltage Protection
-20.0 ÷ 399°C0.1°C±(3%rdg+5°C)250VDC/ACrms
400 ÷ 760°C1°C
-4 ÷ 400°F0.1°F±(3%rdg+9°F)
400 ÷ 1400°F1°F

(*) Accuracy of K-probe not considered

6.1.1. Reference Standards

6.1.2. General Characteristics

6.2. Environment

6.2.1. Environmental Conditions for Use

This instrument satisfies the requirements of Low Voltage Directive 2006/95/EC (LVD) and of EMC Directive 2004/108/EC.

6.3. Accessories Provided

6.4. Optional Accessories

ModelDescriptionTemperature rangeAccuracy (at 100°C)Probe length (mm)Probe diameter (mm)
TK107Air and gas temperature-40 ÷ 800 °C± 2.2rdg2001.5
TK108Internal temperature of semisolid substances-40 ÷ 800 °C± 2.2rdg2003
TK109Internal temperature of liquids-40 ÷ 800 °C± 2.2rdg2004
TK110Surface temperature-40 ÷ 400 °C± 2.2rdg2005
TK111Surface temperature with fixed tip at 90°C-40 ÷ 400 °C± 2.2rdg2605

7. Service

7.1. Warranty Conditions

This instrument is warranted against material or manufacturing defects under general sales conditions. Defective parts may be replaced during the warranty period; the manufacturer reserves the right to repair or replace the product. Return transport to After-sales Service or Dealer is at the customer's charge, but shipment must be agreed in advance. Include a report stating the reason for return. Use original packaging; damage from non-original packaging is charged to the customer. The manufacturer is not responsible for injury to people or damage to property.

The warranty does not cover:

Manual content cannot be reproduced without authorization. Products are patented and trademarks registered. Manufacturer reserves the right to change specifications and prices due to technological improvements.

7.2. Service

If the instrument malfunctions, check battery and cable conditions first. If issues persist, ensure the product is operated according to this manual. For returns, transport is at the customer's charge, and shipment must be agreed in advance. Include a report stating the reason for return. Use original packaging; damage from non-original packaging is charged to the customer.

Models: HT4011 AC Clamp Meter, HT4011, AC Clamp Meter, Clamp Meter, Meter

File Info : application/pdf, 21 Pages, 273.57KB

elektro-lijn-ht4011-hl-en Acrobat Distiller 9.5.5 (Windows)

Related Documents

Preview HT9014 & HT9015 AC/DC TRMS Clamp Meter User Manual
Comprehensive user manual for the HT9014 and HT9015 AC/DC TRMS clamp meters, detailing safety precautions, instrument description, operating instructions, maintenance procedures, technical specifications, and warranty information.
Preview HT100 User Manual - HT Italia
User manual for the HT100 clamp meter from HT Italia, detailing safety precautions, instrument description, operating instructions, maintenance, technical specifications, and service information.