សៀវភៅណែនាំអ្នកប្រើប្រាស់ Druck UPS4E Loop Calibrator

UPS4E Loop Calibrator

ព័ត៌មានអំពីផលិតផល

លក្ខណៈបច្ចេកទេស

  • Brand: UPS4E
  • Model: Loop Calibrator
  • Manufacturer: Druck.com

ការពិពណ៌នា

The Loop Calibrator is a precision instrument designed for
calibrating and testing current loops in various industrial
applications. It offers a range of features to ensure accurate
measurements and ease of use.

លក្ខណៈពិសេស

  • Power and USB Connection
  • ការព្រមានកម្រិតថ្ម
  • USB Virtual Comm Port (VCP) and Mass Storage Class (MSC)
    ពិធីការ
  • Current Source and Measurement
  • Primary and Secondary Readings
  • Positive and Negative Measurement
  • Current Range Selection
  • Date, Time, and Calibration Date
  • ពន្លឺខាងក្រោយ
  • បិទដោយស្វ័យប្រវត្តិ
  • Selectable Internal Loop Resistor and 24 V Supply
  • Calibrator Information
  • ការកត់ត្រាទិន្នន័យ
  • ការក្រិតតាមខ្នាត
  • អាប់ដេតកម្មវិធីបង្កប់

ការណែនាំអំពីការប្រើប្រាស់ផលិតផល

1. Powering On the Calibrator

To power on the calibrator, press and hold the power button for
a few seconds until the display lights up.

2. Setting Current Range

Use the navigation buttons to select the desired current range
for calibration or testing.

3. Connecting to a Device Under Test (DUT)

Connect the calibrator to the DUT using the appropriate cables
និងឧបករណ៍ភ្ជាប់។

4. Performing Calibration

Follow the provided calibration procedures to ensure accurate
readings and calibrate the current loop as needed.

៦.២. ការកត់ត្រាទិន្នន័យ

If data logging is required, enable the feature and specify the
logging parameters before initiating the calibration process.

សំណួរគេសួរញឹកញាប់

សំណួរ: តើខ្ញុំគួរធ្វើដូចម្តេចប្រសិនបើថ្មមានកម្រិតទាប?

A: When the battery is low, replace it immediately with a new
one to ensure uninterrupted operation of the calibrator.

Q: How can I update the firmware of the calibrator?

ចម្លើយ៖ ដើម្បីធ្វើបច្ចុប្បន្នភាពកម្មវិធីបង្កប់ សូមចូលទៅកាន់គេហទំព័ររបស់ក្រុមហ៊ុនផលិត webគេហទំព័រ និង
ទាញយកកំណែកម្មវិធីបង្កប់ចុងក្រោយបំផុត។ ធ្វើតាមការណែនាំ
ផ្តល់ជូនដើម្បីបញ្ចប់ដំណើរការអាប់ដេត។

Q: Can the calibrator be used for both current sourcing and
ការវាស់វែង?

A: Yes, the calibrator is capable of both sourcing and measuring
current, making it a versatile tool for calibration tasks.

UPS4E
Loop Calibrator User Manual
Druck.com

អំពីសៀវភៅណែនាំនេះ។
INFORMATION Read this manual carefully before use. Keep for future
យោង
This manual contains operating instructions that you must follow to make sure of safe operation and to maintain the equipment in a safe condition.
សុវត្ថិភាព
RISK OF ELECTRIC SHOCK Do not apply voltage or currents to the
Calibrator that are greater than shown in the specifications.
WARNING Do not store or use the Calibrator in temperatures outside its
specified range. Temperatures outside the specified range can cause damage to the batteries and Calibrator. The batteries can leak.
CAUTION Remove or replace the batteries immediately if they are exhausted.
Recycle them according to your local regulations. Check the Calibrator for damage before each use. Do not use if the Calibrator has damage. The Calibrator is a portable instrument for use in indoor environments and temporary use in outdoor environments. It is not for permanent installation in an outdoor environment. We have designed this equipment to be safe when operated using the procedures shown in this manual. Do not use this equipment for any other purpose than that stated, as the protection provided by the equipment may not work. For a complete list of safety information, refer to the Quick Start and Safety Manual supplied with the Calibrator. Refer to our website for supporting documents.
www.druck.com
ដំបូន្មានបច្ចេកទេស
Contact the manufacturer for technical advice. See rear pages for contact details.
Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | i

និមិត្តសញ្ញា
ការពិពណ៌នានិមិត្តសញ្ញា
ឧបករណ៍នេះបំពេញតាមតម្រូវការនៃការណែនាំសុវត្ថិភាពអឺរ៉ុបពាក់ព័ន្ធទាំងអស់។ ឧបករណ៍មានសញ្ញា CE ។
Druck is an active participant in Europe’s Waste Electrical and Electronic Equipment (WEEE) take-back initiative (directive 2012/19/EU).
គ្រឿងបរិក្ខារដែលអ្នកបានទិញបានទាមទារការទាញយក និងប្រើប្រាស់ធនធានធម្មជាតិសម្រាប់ផលិតរបស់វា។ វាអាចមានសារធាតុគ្រោះថ្នាក់ដែលអាចប៉ះពាល់ដល់សុខភាព និងបរិស្ថាន។
ដើម្បីជៀសវាងការផ្សព្វផ្សាយសារធាតុទាំងនោះនៅក្នុងបរិស្ថានរបស់យើង និងកាត់បន្ថយសម្ពាធលើធនធានធម្មជាតិ យើងសូមលើកទឹកចិត្តឱ្យអ្នកប្រើប្រព័ន្ធដកយកមកវិញដែលសមស្រប។ ប្រព័ន្ធទាំងនោះនឹងប្រើឡើងវិញ ឬកែច្នៃសម្ភារៈភាគច្រើននៃឧបករណ៍ជីវិតចុងក្រោយរបស់អ្នកតាមរបៀបដ៏ត្រឹមត្រូវមួយ។ និមិត្តសញ្ញាធុងសំរាមដែលកាត់ចេញ អញ្ជើញអ្នកឱ្យប្រើប្រព័ន្ធទាំងនោះ។
ប្រសិនបើអ្នកត្រូវការព័ត៌មានបន្ថែមអំពីប្រព័ន្ធប្រមូល ប្រើប្រាស់ឡើងវិញ និងកែច្នៃឡើងវិញ សូមទាក់ទងរដ្ឋបាលកាកសំណល់ក្នុងតំបន់ ឬតំបន់របស់អ្នក។
សូមចូលទៅកាន់តំណខាងក្រោមសម្រាប់ការណែនាំអំពីការដកថយ និងព័ត៌មានបន្ថែមអំពីគំនិតផ្តួចផ្តើមនេះ។

https://druck.com/weee

សទ្ទានុក្រម
This manual uses the following terms. Abbreviations are the same in the singular and plural.

Term CSV DC FS DUT HART mA max min MSC PC ppm

Description Comma-separated Values Direct current Full Scale Device Under Test Highway Addressable Remote Transducer Milliampere Maximum Minute or minimum Mass Storage Class – USB Personal Computer Parts per million

Copyright 2025 Baker Hughes Company. ii | UPS4E Instruction Manual­English

Term PIN Rdg USB V VCP °C °F

Description Personal Identification Number Reading Universal Serial Bus Volts Virtual Com Port – USB Degrees Celsius Degrees Fahrenheit

តើមានអ្វីនៅក្នុងប្រអប់
When you receive the product, check the box contents for these items: · UPS4E Calibrator. · 4 x AA Alkaline batteries. · 2 m USB cable type A to C. Druck Part IO-USB-C-CABLE · Test leads. Druck Part 209-359. · Quick Start Guide. Druck Part 183M0493-1. Note: We recommend that you keep the box and packaging for future use.
ធាតុស្រេចចិត្ត
Refer to Datasheet for any spares or accessories.

Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | iii

Copyright 2025 Baker Hughes Company. iv | UPS4E Instruction Manual­English

មាតិកា

១.១. សេចក្តីផ្តើម និងការពិពណ៌នា

1

1.1 ការណែនាំ

1

1.2 ការពិពណ៌នា

2

1.3 Screen Symbols and Terms

3

1.4 Calibrator Features

5

1.4.1 Power and USB Connection

5

1.4.2 Battery Low Warning

5

1.4.3 USB Virtual Comm Port (VCP) and Mass Storage Class (MSC)

ពិធីការ

6

1.4.4 Current Source and Measurement

6

1.4.5 Primary and Secondary Readings

6

1.4.6 Positive and Negative Measurement

7

1.4.7 Current Range Selection

7

1.4.8 Date, Time and Calibration Date

7

៤.៥ អំពូល Backlight

7

1.4.10 ការបិទដោយស្វ័យប្រវត្តិ

7

1.4.11 Selectable Internal Loop Resistor and 24 V Supply

7

1.4.12 Calibrator Information

8

៥.៣ ការកត់ទិន្នន័យ

8

1.4.14 ការក្រិតតាមខ្នាត

8

៣.១១ ការធ្វើបច្ចុប្បន្នភាពកម្មវិធីបង្កប់

8

1.5 Operation Modes and Advanced Options

8

1.5.1 Continuity Test Mode

9

1.5.2 អរAMP, `AUTO’ RAMP, STEP and `AUTO’ STEP

9

1.5.3 `SPAN’ Check

10

2. ប្រតិបត្តិការ

11

2.1 បើក និងបិទថាមពល

11

2.2 Connecting by USB

11

2.3 First Use – Date and Time

12

2.4 Typical Opening Screen

13

2.5 ការកំណត់

14

2.5.1 To Open the Settings

15

2.5.2 Settings – Range and 250 ohm Internal Resistor

15

2.5.3 Setting – Data Logging

16

2.5.4 Setting – USB

16

2.5.5 Settings – Backlight and Auto Power Off

16

2.5.6 ការកំណត់ – Viewing Calibrator Information

17

2.5.7 Setting – Calibration

18

2.5.8 Setting – Calibration Date

18

2.5.9 Setting – Firmware Update

18

2.5.10 Setting – Date and Time

18

2.6 Operation Modes and Advanced Options

19

2.6.1 Current Measuring Mode – External or Internal Loop Supply

20

2.6.2 Current Source Mode – External or Internal Loop Supply

21

៣.១ វ៉ុលtage Measuring and Continuity Test Modes

22

2.6.4 Advanced Options (MAN, LIN, FLOW and VALVE)

23

2.6.5 MAN (Manual) Operation

24

2.6.6 Advanced Options (`AUTO’ STEP, STEP, `AUTO’ RAMP, រAMP

Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | v

and `SPAN’ check)

25

3. Connections and Tasks

27

3.1 DC Current Measurement or Source – External Supply

27

3.2 DC Current Measurement or Source – Internal 24 V Supply

28

3.3 DC Current Measurement or Source – External Supply and Resistor

29

3.4 DC Current Measurement or Source – Internal 24 V Supply and Resistor

30

៣.២.៣ ឌីស៊ីវ៉ុលtage Measurement 0 – 30 VDC

31

3.6 ការធ្វើតេស្តបន្ត

31

៥.៣ ការកត់ទិន្នន័យ

32

3.7.1 Setting – Data Log Time (Interval and Duration)

32

3.7.2 Viewing the Data Log

33

4. Calibration Procedures

35

4.1 មុនពេលចាប់ផ្តើម

35

4.2 `ADJ’ (ADJUST) function

35

4.3 ADV (ADVANCED) button polarity function

35

4.4 Calibration Notes

35

4.5 Preset Calibration Values

36

4.6 Deviation Tolerances

36

4.6.1 វិធានការបច្ចុប្បន្ន

36

4.6.2 ប្រភពបច្ចុប្បន្ន

36

៣.១ វ៉ុលtage រង្វាស់

37

4.7 Calibration Flow

38

4.8 Procedure 1: Unlock PIN for Calibration

39

4.9 Procedure 2: Current (Measure)

40

4.10 Procedure 3: Current (Source)

41

4.11 Procedure 4: Voltage (Measure)

42

4.12 Calibration Restore

43

5. លក្ខណៈបច្ចេកទេស

45

6. ការថែទាំ

47

6.1 ការសម្អាត

47

6.2 Installing and Changing the Batteries

47

6.3 ការត្រួតពិនិត្យការបន្ត

48

6.4 ការធ្វើបច្ចុប្បន្នភាពកម្មវិធីបង្កប់

48

6.4.1 Method 1.

49

6.4.2 Method 2.

49

6.5 Error Codes and Warnings

50

6.5.1 លេខកូដកំហុស

50

6.5.2 ការព្រមាន

50

6.6 Instrument Return

50

6.6.1 Return Goods/Material Procedure

50

6.7 ការវេចខ្ចប់សម្រាប់ការរក្សាទុក ឬដឹកជញ្ជូន

51

Copyright 2025 Baker Hughes Company. vi | UPS4E Instruction Manual­English

១.១. សេចក្តីផ្តើម និងការពិពណ៌នា
1.1 ការណែនាំ

សេចក្តីផ្តើម

Figure 1-1: The UPS4E The Druck UPS4E Calibrator is a pocket-size battery-operated hand-held instrument for voltage measurement and measurement and sourcing of electrical current for test and calibration of devices such as flow and pressure sensors or valves. It includes a simple continuity test function and can log test data for viewing on a suitable PC. The Calibrator also has advanced current output features.
Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | 1

Chapter 1. Introduction and Description 1.2 Description

5

4

៦៧ ៨
3

6
១២៣ ៤

៦៧ ៨

1 Screen 3 USB Type C Connector 5 Buzzer 7 Power button. 9 Up, down, left and right Navigation Pad
ប៊ូតុង។

2 Buttons and Navigation Pad 4 Red and Black Connection Sockets 6 Mode button. 8 Advanced button. 10 Enter button.

Figure 1-2: Calibrator Primary Parts

The Calibrator has a custom segment monochrome LCD screen that shows the measured values of voltage and current. It has a Navigation Pad and three buttons for user input and operation of the Calibrator. The top of the Calibrator case has two 4 mm connection sockets and a buzzer. The buzzer operates for continuity tests and to confirm some actions – for example, when energizing the Calibrator. The bottom of the case has a USB type C socket for connection to a suitable PC or USB power source.

Copyright 2025 Baker Hughes Company. 2 | UPS4E Instruction Manual­English

Screen Symbols and Terms
1.3 Screen Symbols and Terms
The display screen has pre-defined symbols and terms. You will not see all these symbols and terms at the same time. They will change, determined by how you use the Calibrator. Figure 1-3 shows the screen with all symbols and terms.

បច្ចុប្បន្នភាព FW

Figure 1-3: Screen Showing all Symbols and Terms

Table 1-1: Screen Symbols and Terms

Symbol/ Term

ការពិពណ៌នា

Symbol/ Term

ការពិពណ៌នា

Battery level indicator – shown when using battery power.
Flashes empty when the battery level is low.

Shown when setting the date. DD = Date MM = Month
YYYY = Year For example: 24 10 2025

Shown when in calibration mode or setting the calibration date.

Shows the amount of days until calibration is due.

Shown when the calibration is due.

Backlight – shown when setting the backlight in the settings.

Setting – shown when you enter the Setting menu.
Shown when setting the Data log interval.
មីលីampere or current measuring enabled.

Information – shown when viewing information in the settings.
Auto Off – shown when setting the Auto Power off options.
ភេនសិនtage.

Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | 3

Chapter 1. Introduction and Description

Symbol/ Term

Table 1-1: Screen Symbols and Terms

ការពិពណ៌នា

Symbol/ Term

ការពិពណ៌នា

វ៉ុលtagអ៊ី ឬវ៉ុលtage measuring enabled.

Current Range: 0 to 20 or 4 to 20 milliampere ។

Measure mode enabled.

Shown when setting the 0 to 20 mA or 4 to 20 mA range.

Source mode enabled.

អប្បបរមា។

24 V loop power enabled.

អតិបរមា។

Continuity test enabled. Manual current output enabled.
Linear profile enabled. Flow profile enabled. Valve profile enabled. Automatic current output enabled used with Step and Ramp function. Step function enabled.
Ramp function enabled.

Internal 250 ohm resistor enabled.
Adjust – used in the calibration procedure.
Used in the calibration procedure and some other settings.
Enabled when setting data log and while logging data.
Enabled when setting the time.
Used in the calibration process.
Hours, minutes and seconds. For example 16:50:32
Application version.

Span check profile បានបើកដំណើរការ។
Up and down indicators to show the direction of the changing primary reading. Also work as maximum and minimum indicators.
PIN Lock – enabled when entering the PIN in restricted modes.

Universal Serial Bus Port.
Virtual Comm Port – USB mode. The mode is MSC if VCP is not
បានបង្ហាញ។
FW UPDATE Firmware Update – enabled when updating the firmware.

Copyright 2025 Baker Hughes Company. 4 | UPS4E Instruction Manual­English

1.4 Calibrator Features
1.4.1 Power and USB Connection

Calibrator Features

៦៧ ៨
2

1 Mains-to-USB power adapter (not supplied)

2 USB Cable Type A to C

Figure 1-4: Power and USB Connection

The Calibrator can be battery or USB-powered. The internal batteries supply power to the Calibrator if it is not supplied through the USB connection. When using battery power, you will see the battery level indicator. When using USB power, you will see the USB symbol in the screen and the battery symbol will go off.

You can supply power to the Calibrator through the USB connection using a suitable mains-toUSB adapter, or a connected computer. See “Specifications” for USB details.

When connected to a computer, the computer can supply power to the Calibrator and can access the internal memory of the Calibrator to open measurement data logs or update the Calibrator firmware.

INFORMATION The measurement circuits of the UPS4E are referenced to
it’s USB socket, so if connected to a ground referenced USB supply, this may affect measurements. For optimum performance we recommend disconnecting the USB connection and using battery power only when doing measurements.

កំណត់ចំណាំ៖
· The USB connection does not recharge the Calibrator batteries, so do not use rechargeable batteries.
· We do not supply a mains-to-USB adapter with the Calibrator. Refer to “Specifications” on page 45 for USB power details.

1.4.2 Battery Low Warning

The Calibrator shows a warning when the battery drops to a low level. See the “Specifications” on page 45 for the low level voltage. In this case, the battery symbol will flash.
CAUTION Remove or replace exhausted batteries immediately. Recycle them
យោងតាមបទប្បញ្ញត្តិក្នុងស្រុករបស់អ្នក។

Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | 5

Chapter 1. Introduction and Description
1.4.3 USB Virtual Comm Port (VCP) and Mass Storage Class (MSC) Protocols
= MSC
VCP
= VCP
Figure 1-5: MSC and VCP mode The USB connection can work in a choice of two USB protocols. VCP is the default mode. It sets the USB socket to work as a communications port for direct communications with your PC to make the initial connection. MSC mode lets your PC see the internal memory of the Calibrator as a storage device, so you may see and download data log files ឬផ្ទេរ files. 1.4.4 Current Source and Measurement The Calibrator can work as a current source to control and measure the current in a loop containing a device such as a valve, with or without an external supply. It will also work as a measurement-only instrument where external devices control the loop supply current. 1.4.5 Primary and Secondary Readings
1
2
Figure 1-6: Two Readings The Calibrator has two sets of readings; the primary reading (1) and the secondary reading (2). The primary reading shows the measured value in units of milliamperes or volts. It can also show text for some operations, settings and calibration. The primary reading can flash for few seconds during some measurements until the Calibrator measurement circuits determine that the value has stabilized or the value is out of range or that the loop is open. The secondary reading shows
Copyright 2025 Baker Hughes Company. 6 | UPS4E Instruction Manual­English

Calibrator Features
the measured value as a percentage of the range you have selected. This is also true for reversed (negative) readings. The calibration procedures use the secondary reading when setting calibration values. Note: The advanced options can affect the secondary reading percentage value relationship with the measured value shown in the primary reading. For example 8 mA = 50% with FLOW selected or 8 mA = 25% with LIN selected in the 4-20 mA range.
1.4.6 Positive and Negative Measurement The Calibrator has red (positive) and black (negative) connections to show polarity of connection, but it will measure in both positive and negative polarity. The primary and secondary readings will show a negative symbol for reverse polarity readings. Note: Current sourcing is only in positive polarity.
1.4.7 Current Range Selection The Calibrator has two current range options: 4-20 mA or 0-20 mA.
1.4.8 Date, Time and Calibration Date You can set the current date and time in the Calibrator and the calibration date. The Calibrator can then calculate how many days until calibration is due. The screen will show `CAL DUE’ when calibration is due.
៤.៥ អំពូល Backlight

Figure 1-7: Backlight Symbol The screen of the Calibrator has a backlight for use in dark areas. The backlight has three settings, ON, OFF or AUTO. See “Settings – Backlight and Auto Power Off” on page 16 for more detail.
1.4.10 Auto Power Off To save battery energy, this feature de-energizes the Calibrator after ten minutes of no button pushes. Note: The Calibrator disables the Auto Power Off feature when you use data logging or when using USB power.
1.4.11 Selectable Internal Loop Resistor and 24 V Supply
3

4

4

1
+

០៣៖

ហាត

2

24 វី +

០៣៖

ហាត

2

4 – 20 mA –

I

1 Loop supply 3 Calibrator

2 Device under test 4 HART instrument

Figure 1-8: Loop Resistor and 24 V Supply

Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | 7

Chapter 1. Introduction and Description
In a process control or measurement loop, a resistor of between 230 and 600 ohm is necessary for communications to work correctly with HART instruments. The Calibrator has an internal 24 V supply and a nominal 250 ohm loop resistance in series that you can enable or disable. The Calibrator can then work as a loop supply, resistor and measuring instrument, so an external resistor and loop supply is not necessary. You can disable one or both of these features if you have an external voltage source and resistor. Note: You must enable the internal 24 V supply or have an external loop supply for the Calibrator to control current. 1.4.12 Calibrator Information
Figure 1-9: Information Symbol Use the Calibrator Settings to show important Calibrator information such as the firmware version (APP VER) or battery voltage of the Calibrator. See “Settings” for more details. 1.4.13 Data Logging The Calibrator can log the data from your tests at set intervals and durations. See “Data Logging” on page 32 for more details. 1.4.14 Calibration The Calibrator has a PIN-protected procedure for checking and adjusting its calibration using suitable external calibration equipment. See “Calibration Procedures” on page 35 for more details. 1.4.15 Firmware Update The Calibrator has a PIN-protected procedure for upgrading its firmware using a suitable PC and the USB cable supplied with the Calibrator. See “Updating the Firmware” on page 48.
1.5 Operation Modes and Advanced Options
បច្ចុប្បន្នភាព FW
Figure 1-10: Mode and Advanced Option Symbols The upper part of the screen shows the operation mode and advanced option symbols. You can select basic operation modes of measure or source, with or without the internal 24 V supply
Copyright 2025 Baker Hughes Company. 8 | UPS4E Instruction Manual­English

Operation Modes and Advanced Options
enabled. The advanced options give different ways to change the source current. See “Operation Modes and Advanced Options” on page 19 for more details. 1.5.1 Continuity Test Mode
Figure 1-11: Continuity Symbol The Calibrator can work as a simple continuity tester. When connected to a device under test (DUT) the Calibrator will operate its buzzer if the DUT has a resistance of less than approximately 100 ohms.
The primary reading will show opEn or Short circuit.
1.5.1.1 Manual, Linear, Flow and Valve Profiles

LIN mA

FLOW mA

VALVE mA

s

s

s

Figure 1-12: Linear, Flow and Valve Profiles (s = time) You can select different profiles for the Calibrator:
· `MAN’ – Manual, where you manually set the loop current to any value. · `LIN’ – Linear, where the loop current changes in a linear way to simulate linear transmitters. · `FLOW’ and `VALVE’ – where the loop current changes in a non-linear way to simulate flow
transmitter and valve control signals.
See “Specifications” on page 45 for the preset values.
Note: These options can affect the secondary reading percentage value relationship with the measured value shown in the primary reading, even in measure mode. For example 8 mA = 50% with `FLOW’ selected or 8 mA = 25% with `LIN’ selected in the 4-20 mA range.
1.5.2 អរAMP, `AUTO’ RAMP, STEP and `AUTO’ STEP

Figure 1-13: RAMP, AUTO RAMP, STEP and AUTO STEP functions The RAMP symbol shows that the output current will gradually change in value from minimum to maximum of the selected range over a specified time that you can set. You start the ramp using the Navigation Pad buttons. `AUTO’ RAMP automatically cycles the output current over the specified time. With RAMP selected, the Calibrator calculates the values and rate of change, based on the specified time.
The STEP symbol shows that the output current will change in steps of preset values from minimum to maximum of the selected range over a specified time that you can set. You change
Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | 9

Chapter 1. Introduction and Description
the steps using the Navigation Pad buttons. `AUTO’ STEP automatically cycles the output current in preset values over the specified time. See “Advanced Options (`AUTO’ STEP, STEP, `AUTO’ RAMP, រAMP and `SPAN’ check)” on page 25 for more details. See “Specifications” on page 45 for the preset step values. Note: The `VALVE’ profile option does not have the RAMP function. 1.5.3 `SPAN’ Check This option lets you manually set the current source output directly from minimum to maximum of the range and back using the buttons of the Navigation Pad. This will do a full span check. See “Advanced Options (`AUTO’ STEP, STEP, `AUTO’ RAMP, រAMP and `SPAN’ check)” on page 25 for more details.
Copyright 2025 Baker Hughes Company. 10 | UPS4E Instruction Manual­English

បើក និងបិទ
2. ប្រតិបត្តិការ
This chapter shows how to set and operate the features of the Calibrator.
2.1 បើក និងបិទថាមពល
UpS4E
MODE ADV
Figure 2-1: To Energize the Calibrator 1. To energize the Calibrator, push the Power button until the screen operates. It will normally
operate the buzzer, show the UPS4E text then show a typical opening screen. See Figure 2-3. If this is the first use of the Calibrator after fitting new batteries, it will ask you to enter the date and time. See “First Use – Date and Time”. 2. To de-energize the Calibrator, push the Power button for more than two seconds and release. The screen will go off.
2.2 Connecting by USB
1. Use the USB cable supplied with the Calibrator. Connect its smaller USB C plug into the USB connection on the bottom of the Calibrator. See Figure 1-4 on page 5.
2. Connect the USB A plug into a USB type A socket of your computer, or into the output socket of a suitable mains to USB adapter. See “Specifications” for power details.
3. Energize the PC or USB adapter and the Calibrator.
Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | 11

Chapter 2. Operation 2.3 First Use – Date and Time

MODE ADV

MODE ADV

MODE ADV

MODE ADV

MODE ADV

MODE ADV

Figure 2-2: Set Date and Time
When you first energize a Calibrator after fitting new batteries (or the batteries have been removed for longer than approximately seven minutes), the Calibrator will automatically enter the Settings to ask you to set the current date and time. It can then compare this with the calibration date to calculate and show how many days until calibration. The default date will be the build date set in the factory. You can adjust the calibration date in the Settings. Refer to “Setting Calibration Date” on page 18.

1. The screen will show the Settings symbol

and flash the DATE day (DD). Use the

Navigation up and down buttons to change the date.

2. Push the Enter button to accept the date.

3. The screen will sequentially show the month (MM), year (YYYY), TIME hours (HH) and minutes (MM). At each step, use the Navigation Pad up and down buttons to change the values. Push the Enter button to accept the values and move onto the next value.

4. Finally, the screen will show the amount of `DAYS’ until the calibration is due (CAL DUE), based on the calibration date. See “Setting – Calibration Date” on page 18.

Copyright 2025 Baker Hughes Company. 12 | UPS4E Instruction Manual­English

2.4 Typical Opening Screen

Typical Opening Screen

Figure 2-3: Typical Opening Screen
After you have set the date and time, the Calibrator will show an opening screen that flashes the mA and source symbols, asking you to select an operation mode such as mA source or measure. It will show the same opening screen each time you energize it.

Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | 13

Chapter 2. Operation 2.5 Settings

RANGE · 4-20 mA · 0-20 mA

០៣៖

· on · off

កំណត់ហេតុ

· HH · MM · SS

· AUto · off · on

SET

··

VCP ` ‘(MSC)

AUTO OFF · on · off

· APPVER · V · 250 : · CAL DATE

កាល់

កាល់

· PIN · DD · MM · YYYY

· DD DATE · MM
· YYYY

បច្ចុប្បន្នភាព FW

របៀប

Figure 2-4: Settings Options Flow The Calibrator has settings that you can change, such as the backlight or measuring range. When

you open the settings, the screen shows the Settings

symbol in the top right.

Copyright 2025 Baker Hughes Company. 14 | UPS4E Instruction Manual­English

2.5.1 To Open the Settings

ការកំណត់

របៀប
ADV
Figure 2-5: To Open the Settings 1. To open the settings, push both the up and down Navigation Pad buttons together. The
screen will show the Settings symbol. 2. Use the Navigation Pad buttons to go through the settings and make any changes as shown
in the following pages. 3. Push the `MODE’ button at any time to escape the Settings or to return to the opening
screen. 2.5.2 Settings – Range and 250 ohm Internal Resistor

MODE ADV

MODE ADV

Figure 2-6: Settings – Range and 250 ohm Resistor 1. Open the Settings. 2. The screen will show `RANGE’. If not, then push the right hand Navigation Pad button until
the screen shows `RANGE’. It will flash. 3. Push the Enter button to accept that you are to select the range. 4. Push the up and down Navigation Pad buttons to select 0-20 mA or 4-20 mA. 5. Push the Enter button to save the range setting. 6. Push the right hand Navigation Pad button until the screen shows 250 .
Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | 15

ជំពូកទី 2. ប្រតិបត្តិការ
7. Push the Enter button to accept that you are to select the resistor. 8. Push the up and down Navigation Pad buttons to select `ON’ or `OFF’. 9. Push the Enter button to save the setting. 2.5.3 Setting – Data Logging Refer to Section 3.7 on page 32. 2.5.4 Setting – USB

MODE ADV

MODE ADV

Figure 2-7: Settings – USB 1. Open the Settings. 2. Push the right hand Navigation Pad button until the screen shows `SET’ and the USB

និមិត្តសញ្ញា

. It will flash.

3. Push the Enter button to accept that you are to set the USB mode.

4. Push the up or down Navigation Pad buttons to select VCP on or off. If VCP is not shown, then the mode is MSC. VCP is the default mode.

5. Push the Enter button to accept your changes.

2.5.5 Settings – Backlight and Auto Power Off

ស្វ័យប្រវត្តិ

ON

MODE ADV

MODE ADV

Figure 2-8: Settings – Backlight and Auto Power Off
Copyright 2025 Baker Hughes Company. 16 | UPS4E Instruction Manual­English

ការកំណត់
1. Open the Settings. 2. Push the right hand Navigation button until the screen shows the backlight symbol. It will
flash. 3. Push the Enter button to accept that you are to set the backlight. 4. Push the up and down Navigation buttons to select `Auto’, `ON’ or `OFF’.
· ON – the backlight is on at all times while the Calibrator is energized · OFF – the backlight is never on · AUTO – the backlight remains on for ten seconds after the last button has been pushed
while the Calibrator is energized 5. Push the Enter button to save the Backlight setting. 6. Push the right hand Navigation button until the screen shows `AUTO OFF’. It will flash. 7. Push the Enter button to accept that you are to set the auto power off. 8. Push the up and down Navigation buttons to select `ON’ or `OFF’. 9. Push the Enter button to save the setting.
2.5.6 ការកំណត់ – Viewing Calibrator Information

OO.O3

MODE ADV

MODE ADV

MODE ADV

MODE ADV

MODE ADV

Figure 2-9: Settings – Calibrator Information 1. Open the Settings.
2. Push the right hand Navigation button until the screen shows the Information symbol . It will flash.
3. Push the Enter button to accept that you are to view information. 4. Push the up and down Navigation buttons to select the information that you need to see, for
example: · Software version APPVER · Battery voltage V · 250 ohm resistor on or off · Calibration date 5. Select the `MODE’ button to return to normal operation.

Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | 17

ជំពូកទី 2. ប្រតិបត្តិការ
2.5.7 Setting – Calibration Refer to Section 4 on page 35. 2.5.8 Setting – Calibration Date

MODE ADV

MODE ADV

MODE ADV

Figure 2-10: Settings – Calibration Date 1. Open the Settings. 2. Push the right hand Navigation Pad button until the screen shows the `DATE’ and `CAL’
symbols. They will flash.
3. Push the Enter button to accept that you are to set the calibration date. 4. The screen shows the PIN symbol to let you know that you must enter the four digit PIN 4321
to continue to set the calibration date. Use the left and right Navigation Pad buttons to select each digit and use the up and down Navigation Pad buttons to change the number of each digit. Then push Enter. 5. Push the left and right Navigation Pad buttons to select day (DD), month (MM) or year (YYYY). They will flash. 6. Push the up and down Navigation Pad buttons to change the value of the day, month or year.
7. Push the Enter button each time to accept your changes. 8. The Calibrator will now use this date to calculate the days until the calibration is due.
2.5.9 Setting – Firmware Update Refer to “Updating the Firmware” on page 48.
2.5.10 Setting – Date and Time Refer to “First Use – Date and Time” on page 12.

Copyright 2025 Baker Hughes Company. 18 | UPS4E Instruction Manual­English

Operation Modes and Advanced Options 2.6 Operation Modes and Advanced Options
របៀប

របៀប

mA

របៀប

mA 24 V

របៀប

mA

mA MODE

១២ វ

របៀប

របៀប

V

Figure 2-11: MODE Options Flow
You use the `MODE’ button to select the operation modes of the Calibrator. For example, current or voltage measuring or current sourcing. Select the Enter button to accept the mode and escape to the Advanced Options.

Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | 19

ជំពូកទី 2. ប្រតិបត្តិការ
2.6.1 Current Measuring Mode – External or Internal Loop Supply

MODE ADV

MODE ADV

Figure 2-12: Current Measure Mode
This is for use when the external device under test controls the loop current.
1. Push the `MODE’ button until the screen shows the current symbol mA and the measure mode symbol.
2. For internal loop supply, push the `MODE’ button again until the screen shows the 24 V symbol.
3. Push the Enter button to accept your changes.
4. Now select the Advanced Options of `LIN’ or `FLOW’ to change the way the secondary reading shows percentage value as necessary. See “Advanced Options (MAN, LIN, FLOW and VALVE)” on page 23.

Copyright 2025 Baker Hughes Company. 20 | UPS4E Instruction Manual­English

Operation Modes and Advanced Options
2.6.2 Current Source Mode – External or Internal Loop Supply

MODE ADV

MODE ADV

Figure 2-13: Current Source Mode
This is for use when the Calibrator controls the loop current.
1. Push the `MODE’ button until the screen shows the source mode symbol and the current symbol mA.
2. For internal loop supply, push the `MODE’ button again until the screen shows the 24 V symbol.
3. Push the Enter button to accept your changes.
4. Now select the Advanced Options. See “Advanced Options (MAN, LIN, FLOW and VALVE)” on page 23.

Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | 21

ជំពូកទី 2. ប្រតិបត្តិការ
៣.១ វ៉ុលtage Measuring and Continuity Test Modes

MODE ADV

MODE ADV

រូបភាពទី 2-14: វ៉ុលtage Measure and Continuity Modes
1. Push the `MODE’ button until the screen shows the measure mode symbol and the voltage symbol V.
2. Push the Enter button to accept your changes.
3. To change to continuity test, select the `MODE’ button again until the screen shows the Continuity Buzzer symbol.
4. Push the Enter button to accept your changes. There are no Advanced Options for these modes.

Copyright 2025 Baker Hughes Company. 22 | UPS4E Instruction Manual­English

Operation Modes and Advanced Options
2.6.4 Advanced Options (MAN, LIN, FLOW and VALVE)

mA

mA 24V

ADV

បុរស

ADV

លីន

ADV

ហូរ

ADV

វ៉ាល់

mA

ADV

លីន

mA

24V

ADV

ហូរ

Figure 2-15: Advanced Options MAN, LIN, FLOW and VALVE
បុរស

MODE ADV

MODE ADV

MODE ADV

MODE ADV

MODE ADV

Figure 2-16: Selecting Advanced Options
Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | 23

ជំពូកទី 2. ប្រតិបត្តិការ
From a current source mode: 1. Push the ADV button.’ 2. Use the ADV (Advanced) button to go through the profile options of `MAN’, `LIN’, `FLOW `or
`VALVE’. This will change how the Calibrator controls the output current and how the secondary reading shows the percentage. 3. Push the Enter button to accept your selection. From a current measure mode: 1. Push the ADV button. 2. Use the ADV (Advanced) button to go through the profile options of `LIN’ or `FLOW’. This will change how the secondary reading shows the percentage. 3. Push the Enter button to accept your selection. 2.6.5 MAN (Manual) Operation 1. Select the `MAN’ profile. 2. You can now manually enter a value to output. Use the Navigation Pad up, down, left and right buttons to select each digit in the primary reading and change it’s value. 3. Push the Enter button to accept the value. The secondary reading will show the value as a percentage of range. 4. The Calibrator can now control the loop current at the value you have set. Remember that you must select the internal 24 V supply or have an external supply in the loop. The primary reading will flash if there is no loop current.
Copyright 2025 Baker Hughes Company. 24 | UPS4E Instruction Manual­English

Operation Modes and Advanced Options
2.6.6 Advanced Options (`AUTO’ STEP, STEP, `AUTO’ RAMP, រAMP and `SPAN’ check)

លីន

ហូរ

វ៉ាល់

ស្វ័យប្រវត្តិ

ស្វ័យប្រវត្តិ

ស្វ័យប្រវត្តិ

SPAN

SPAN
Figure 2-17: Advanced Options – `AUTO’ STEP, STEP, `AUTO’ RAMP, រAMP and `SPAN’ check

MODE ADV

MODE ADV

MODE ADV

MODE ADV

MODE ADV

MODE ADV

Figure 2-18: Selecting Advanced Options
Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | 25

ជំពូកទី 2. ប្រតិបត្តិការ
1. From a current source profile option of `LIN’, `FLOW’ or `VALVE’ profile, push the Enter button.
2. Use the Navigation pad left and right buttons to go through the options of `AUTO’ STEP, STEP, `AUTO’ RAMP, រAMP and `SPAN’ check.
3. Push the Enter button to accept your selection. Note: The `VALVE’ profile option does not have the Ramp function. 4. For the `AUTO’ STEP, RAMP and `AUTO’ RAMP functions you must then use the
Navigation Pad buttons to set the time interval, followed by the Enter button to accept your selection. 5. For the `AUTO’ STEP functions, the output will now change through the preset steps over the time interval you have set. It will continue the cycle until you de-energize the Calibrator or change mode. You may push the Enter button to pause the function. Push again to restart. 6. For the `AUTO’ RAMP function, the output will now change over the time interval you have set. It will continue the cycle until you de-energize the Calibrator or change mode. You may push the Enter button to pause the function. Push again to restart. · In STEP mode, use the Navigation Pad up and down buttons to manually step through the preset output current values. · In RAMP mode, use the Navigation Pad up and down buttons to manually start the ramp which will automatically go up or down through the calculated output current values. · In `SPAN’ check mode, use the Navigation Pad up and down buttons to change directly from the maximum to minimum output current values and back. Notes: · Remember that you must either select the internal 24 V supply or have an external supply in the loop for the Calibrator to work. · See “Specifications” on page 45 for `LIN’, `FLOW’ and `VALVE’ preset values of current.
Copyright 2025 Baker Hughes Company. 26 | UPS4E Instruction Manual­English

DC Current Measurement or Source – External Supply
3. Connections and Tasks
The section gives typical details of how to connect the Calibrator for a given task. Before starting, read the safety precautions contained in the `Safety & Quick Start Guide’.
3.1 DC Current Measurement or Source – External Supply CAUTION External supplies must be no more than 30 VDC.
Set the Calibrator correctly to measure or source as necessary before connecting it to a device under test.

របៀប

+-

30 VDC
ADV

Figure 3-1: DC Current Measurement or Source – External Supply
1. Set the necessary RANGE in the Calibrator. For example 4-20 mA.
2. Set the Calibrator to measure or source current. Do not enable the internal 24 V loop power.
3. Select the Advanced Options you need, for example `LIN’ output, and `FLOW’.
4. Connect the cables to the unit under test and the external supply as shown.
5. The primary reading will show you the loop current.
6. If sourcing current, use the Navigation Pad buttons to change the source current, then press the Enter button.
7. The primary reading can flash for a few seconds until the Calibrator measurement circuits determine that the value has stabilized.

Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | 27

Chapter 3. Connections and Tasks 3.2 DC Current Measurement or Source – Internal 24 V Supply
MODE ADV
Figure 3-2: DC Current Measurement or source – Internal 24 V Supply 1. Set the necessary RANGE in the Calibrator. For example 4-20 mA. 2. Set the Calibrator to measure or source current with internal 24 V loop power. 3. Select the Advanced Options you need, for example `LIN’ output, and `FLOW’. 4. Connect the cables to the unit under test as shown. 5. The primary reading will show you the loop current. 6. If sourcing current, use the Navigation Pad buttons to change the source current, then press
the Enter button. 7. The primary reading can flash for a few seconds until the Calibrator measurement circuits
determine that the value has stabilized.
Copyright 2025 Baker Hughes Company. 28 | UPS4E Instruction Manual­English

DC Current Measurement or Source – External Supply and Resistor
3.3 DC Current Measurement or Source – External Supply and Resistor CAUTION External supplies must be no more than 30 VDC.

ហាត

០៣៖

របៀប

+-

30 VDC
ADV

Figure 3-3: DC Current Measurement – External Supply and Resistor
1. Set the necessary RANGE in the Calibrator. For example 4-20 mA.
2. Set the Calibrator to measure or source current with an external supply. Do not enable the internal 24 V source or 250 ohm resistor.
3. Select the Advanced Options you need, for example `LIN’ output, and `FLOW’.
4. Connect the cables to the unit under test, the HART device and the external source as shown.
5. The primary reading will show you the loop current.
6. If sourcing current, use the Navigation Pad buttons to change the source current, then press the Enter button.
7. The primary reading can flash for a few seconds until the Calibrator measurement circuits determine that the value has stabilized.

Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | 29

Chapter 3. Connections and Tasks 3.4 DC Current Measurement or Source – Internal 24 V Supply and Resistor
ហាត
MODE ADV
Figure 3-4: DC Current Measurement or Source – Internal 24 V Supply 1. Set the Calibrator to measure or source current with internal 24 V supply and 250 ohm
resistor. 2. Set the necessary RANGE in the Calibrator. For example 4-20 mA. 3. Select the Advanced Options you need, for example `MAN’ or `LIN’ output, and `AUTO’
STEP or `SPAN’. 4. Connect the cables to the unit under test and the HART device (if used) as shown. 5. The primary reading will show you the loop current. 6. If sourcing current, use the Navigation Pad buttons to change the source current, then press
the Enter button. 7. The primary reading can flash for a few seconds until the Calibrator measurement circuits
determine that the value has stabilized.
Copyright 2025 Baker Hughes Company. 30 | UPS4E Instruction Manual­English

ឌីស៊ីវ៉ុលtage Measurement 0 – 30 VDC
៣.២.៣ ឌីស៊ីវ៉ុលtage Measurement 0 – 30 VDC CAUTION Voltage sources must be no more than 30 VDC.

+

របៀប

30 VDC

ADV

Figure 3-5: DC Voltage Measurement 1. Set the Calibrator to measure voltage. See “Voltage Measuring and Continuity Test Modes”
on page 22. 2. Connect the cables to the voltage supply as shown. 3. The primary reading shows the measured voltage.
3.6 ការធ្វើតេស្តបន្ត
ឌូត

MODE ADV

Figure 3-6: Continuity Test

Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | 31

Chapter 3. Connections and Tasks
1. Set the Calibrator to Continuity test. See “Voltage Measuring and Continuity Test Modes” on page 22.
2. Connect the cables to the device under test (DUT) as shown. 3. If the DUT has a resistance of less than 100 ohms, the Calibrator buzzer will operate and the
primary reading will show `Short’ (short-circuit). If the DUT has a resistance higher than 1000 ohms, the buzzer will not operate and the primary reading will show `Open’ (open circuit).
៥.៣ ការកត់ទិន្នន័យ
Use the Calibrator Settings to log the data from your tests until you manually stop the logging or de-energize the Calibrator. The Calibrator will store the data internally in *.CSV file format, so you can then load it to a suitable PC for inspection. To read the log files, your computer must have a suitable spreadsheet application installed, such as Microsoft Excel. Note: The Calibrator disables the Auto Power Off feature when you use data logging.
3.7.1 Setting – Data Log Time (Interval and Duration)

INT
O:OI

MODE ADV

MODE ADV

OO:OI
MODE ADV

4.OOO
O.O
MODE ADV

1. Set the Calibrator into the measurement or source mode you wish to use and connect it to the test loop.
Note: You cannot change the operation mode while the Calibrator logs data. You must stop and restart the data log if you need to change the operation mode.
2. Open the Settings. 3. Push the right hand Navigation button until the screen shows LOG. It will flash. 4. Push the Enter button to accept that you are to set the log interval and duration. 5. You first set the interval. The screen shows INT. Push the left and right Navigation buttons
to select minutes (MM) and seconds (SS). Push the up and down Navigation buttons to change the digits for minutes or seconds.
6. Push the Enter button to save the interval time. 7. Now set the duration. Push the left and right Navigation buttons to select minutes (MM) and
hours (HH). Push the up and down Navigation buttons to change the minutes or hours. 8. Push the Enter button to save the interval and duration and start the Data Logging. The
Calibrator will return to the opening screen.
9. The LOG symbol will flash and the Calibrator will automatically log your measurements at the interval and duration of time that you have set. You can manually stop it by pressing the MODE or ADV buttons or de-energize the Calibrator.
Copyright 2025 Baker Hughes Company. 32 | UPS4E Instruction Manual­English

ការកត់ត្រាទិន្នន័យ
3.7.2 Viewing the Data Log 1. Disconnect the Calibrator from the test loop. 2. Use the cable supplied to connect the Calibrator to a suitable PC. 3. In the Calibrator Settings, set the USB connection to MSC mode. See “Setting – USB” on
page 16. 4. The PC will see the Calibrator internal memory as a mass storage device. 5. Use a file explorer application on the PC to examine the Calibrator internal memory for the
DataLog folder that contains the files.
Figure 3-7: Typical PC File Explorer Application Showing Datalog files 6. The Calibrator automatically gives each Datalog file a date and time that it was recorded, in
the format YYYY-MM-DD_HHMMSS. For example 2025-04-08_002113 is the year 2025 and the fourth month and the date of the 8th. The time is 21 minutes and 13 seconds past midnight. 7. Select the necessary file and open it with your spreadsheet application to examine it. 8. After use, use the PC to delete any unwanted Datalog files from the Calibrator to empty its memory.
Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | 33

Chapter 3. Connections and Tasks
Copyright 2025 Baker Hughes Company. 34 | UPS4E Instruction Manual­English

មុនពេលចាប់ផ្តើម
4. Calibration Procedures
Druck can provide a calibration service that is traceable to international standards. We recommend that you return the instrument to us or an approved service agent for calibration. If using an alternative calibration facility, check that it uses the same standards.
4.1 មុនពេលចាប់ផ្តើម
1. Check that the Calibrator battery level is at maximum. Consider changing to new batteries if not.
2. Make sure the calibration environment has a stable temperature of 21 ± 1°C (70 ± 2°F). 3. Make sure your calibration standard instrument has these ranges:
· Current (mA) input and output range +/- 0 to 25 mA. · Voltage (V) output range 0 to +/- 30 VDC. 4. Put the Calibrator and calibration standard instrument in the calibration environment for a minimum of two hours. 5. Connect and energize the Calibrator and the calibration standard instrument using the cables supplied or equivalent high-quality cables. 6. Wait a few minutes for the Calibrator and other instruments to thermally stabilize. Consider changing the AUTO OFF setting to off to avoid disruption while waiting for readings to stabilize during calibration.
4.2 `ADJ’ (ADJUST) function
This function adjusts the Calibrator primary reading to the same value as the external calibration instrument when doing current source calibration. It is useful in current and voltage measure calibration when you use calibration values that are different to the preset values. The Calibrator will adjust itself to the external instrument value. For example, if you need to calibrate measurement at a value of 18 mA instead of the preset 19.5 mA, set your external instruments to 18 mA and use the `ADJ’ function to adjust the calibrator to set the Primary reading to 18 mA.
4.3 ADV (ADVANCED) button polarity function
You can use ADV button to change the sign (polarity) of the primary reading when using the ADJ (ADJUST) function while doing calibration.
4.4 Calibration Notes
Push the `MODE’ button or de-energize the Calibrator to escape the calibration setting at any time. If the calibrator shows an error code at any time during the calibration (see “Error Codes and Warnings” on page 50) then it is likely that the calibrated value is incorrect. Push the Enter button to accept the error and redo the calibration.
Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | 35

Chapter 4. Calibration Procedures

4.5 Preset Calibration Values
INFORMATION The Calibrator has preset calibration values of current and
វ៉ុលtage, but you may apply different values and use the ADJ (adjust) to change the reading in the Calibrator to match the applied value.

Current Measure (mA)
ជួរ 20 mA
-៤០
0
+19.5
Check 1 mA

Table 4-1: Preset Calibration Values

Current Measure (mA)
ជួរ 24 mA

Current Source (mA)

វ៉ុលtage Measure (V)
ជួរ 20 V

-៤០

+0.2

-៤០

0

+19.5

0

+24.0

+20

Check 24 mA

Check 1 mA

Check 20 V

វ៉ុលtage Measure (V)
ជួរ 30 V
-៤០
0
+30
Check 30 V

4.6 Deviation Tolerances
After calibration, go through the Set Point values as suggested in these tables and check that the deviation is within the permissible limits.

4.6.1 Current Measure · 20 mA range: 12 ppmRdg + 56 ppmFS · 24 mA range: 22 ppmRdg + 86 ppmFS

Table 4-2: Current Measure Set Point and Deviation Tolerance

Set Point (mA) -24 -22 -20 -10 -5 0 5 10 20 22 24

Permissible Deviation (mA) ±0.002592 ±0.002548 ±0.002504 ±0.00124 ±0.00118 ±0.00112 ±0.00118 ±0.00124 ±0.002504 ±0.002548 ±0.002592

4.6.2 Current Source · 20 mA range: 13 ppmRdg + 68 ppmFS

Copyright 2025 Baker Hughes Company. 36 | UPS4E Instruction Manual­English

Deviation Tolerances

· 24 mA range: 22 ppmRdg + 86 ppmFS

Table 4-3: Current Source Set Point and Deviation Tolerance

Set Point (mA) Permissible Deviation (mA)

0.6

±0.00136

6

±0.00143

12

±0.00151

18

±0.00159

24

±0.0025

៣.១ វ៉ុលtage Measure · 20 V range: 10 ppmRdg + 69 ppmFS · 30 V range: 10 ppmRdg + 79 ppmFS

តារាង 4-4: វ៉ុលtage Measure Set Point and Deviation Tolerance

Set Point (V)

Permissible Deviation (V)

-៤០

±0.00267

-៤០

±0.00258

-៤០

±0.00257

-៤០

±0.00148

-5

±0.00143

0

±0.00138

5

±0.00143

10

±0.00148

20

±0.00257

21

±0.00258

30

±0.00267

Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | 37

Chapter 4. Calibration Procedures 4.7 Calibration Flow
កាល់
លេខសម្ងាត់
* >DD CAL >MM
>YYYY

20 mA

mA 24 V

១២ វ

rESto

SET ADJ

SET ADJ

SET ADJ

លេខសម្ងាត់

CHECK 24 mA

ពិនិត្យ

CHECK 30 V

របៀប

VV

SET ADJ

SET ADJ

ពិនិត្យ

ពិនិត្យ

Figure 4-1: Calibration Flow Note: *The Calibrator will ask for the CAL date and time if it has not already been set.

Copyright 2025 Baker Hughes Company. 38 | UPS4E Instruction Manual­English

Procedure 1: Unlock PIN for Calibration 4.8 Procedure 1: Unlock PIN for Calibration

អូអូអូ

MODE ADV

MODE ADV

Figure 4-2: Unlock PIN for Calibration
1. On the Calibrator, enter the Settings and select the CAL (calibration) option. See “Setting Calibration” on page 18.
2. The CAL symbol will flash. Push the Enter button.
3. The screen will show the Lock symbol and the primary reading will change to a set of four digits for the PIN.
4. Use the Navigation Pad to enter the correct PIN 4321. The left and right buttons will select the digit to be changed and the up and down buttons will change the value of the digit.
5. Push the Enter button to accept the correct PIN. The screen will cancel the Lock symbol.
Note: If you have not already entered the calibration date, the Calibrator will ask you to enter it before doing the calibrations. See “Setting – Calibration Date” on page 18.
6. You will now have the choice of calibrating current measure, current source or voltage.

Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | 39

Chapter 4. Calibration Procedures 4.9 Procedure 2: Current (Measure)

MODE ADV

MODE ADV

Figure 4-3: Calibration ­ Current Measure
1. When you select this option the mA and measure symbols will flash. Use the Navigation left and right buttons to move to the next calibration procedure or push the Enter button to accept. The screen will show the SET symbol.
2. The Calibrator will take you through two ranges; one uses steps of the preset calibration values for the 20 mA range, the second uses steps of the preset calibration values for the 24 mA range. The secondary reading will show the preset calibration values. At each step, set your external calibrator to the preset value, or to your preferred chosen value of calibration.
3. The primary reading will show the measured values at each step. Wait a few minutes to make sure the measured value remains stable before you push Enter to accept. The Calibrator will accept the applied current as a calibrated value and readjust itself.
4. The ADJ symbol will flash to allow you to adjust the primary reading to the same value as the current from the calibrated instrument using the Navigation Pad up and down and left and right buttons. This is useful if using calibrated values that are different to the preset values. You do not need to use this adjustment if you are using the preset values.
5. Push the Enter button to accept.
6. At the last steps, the Calibrator will recommend values of 1 mA and 24 mA and show CHECK. Apply the recommended value and push Enter to accept.

Copyright 2025 Baker Hughes Company. 40 | UPS4E Instruction Manual­English

4.10 Procedure 3: Current (Source)

Procedure 3: Current (Source)

24V
OO.z
MODE ADV

24V
OO.zOO
OO.z
MODE ADV

Figure 4-4: Calibration ­ Current Source
1. When you select this option the mA, source and 24 V symbols will flash. Use the Navigation left and right buttons to move to the next calibration procedure or push the Enter button to accept. The screen will show the SET symbol.
Note: The Calibrator automatically selects the internal 24 V source for this calibration. You cannot deselect it.
2. The Calibrator will take you through steps of the preset calibration values. The secondary reading will show the preset calibration values. At each step, set your external calibrator to view តម្លៃកំណត់ជាមុន។
3. This procedure is different to the others – the secondary screen will flash at each step to allow you to change the source current for calibration to a value different from the preset value if necessary.
4. Wait a few minutes to make sure the measured value at your external instrument remains stable before you push Enter to accept.
5. The ADJ symbol will flash to allow you to adjust the primary reading to the same value as the current measured at the external calibrated instrument using the Navigation Pad up and down and left and right buttons.
6. Push the Enter button to accept.

Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | 41

Chapter 4. Calibration Procedures 4.11 Procedure 4: Voltage (Measure)

MODE ADV

MODE ADV

Figure 4-5: Calibration ­ Voltage រង្វាស់
1. When you select this option the V and measure symbols will flash. Use the Navigation left and right buttons to move to the next calibration procedure or push the Enter button to accept. The screen will show the `SET’ symbol.
2. The Calibrator will take you through two ranges; one uses steps of the preset calibration values for the 20 V range, the second uses steps of the preset calibration values for the 30 V range. The secondary reading will show the preset calibration values. At each step, set your external calibrator to the preset value, or to your preferred chosen value of calibration.
3. The primary reading will show the measured values at each step. Wait a few minutes to make sure the measured value remains stable before you push Enter to accept. The Calibrator will accept the applied voltage as a calibrated value and readjust itself.
4. The ADJ symbol will flash to allow you to adjust the primary reading to the same value as the voltage from the calibrated instrument using the Navigation Pad up and down and left and right buttons. This is useful if using calibrated values that are different to the preset values. You do not need to use this adjustment if you are using the preset values.
5. Push the Enter button to accept.
6. At the last steps, the Calibrator will recommend values of 20 V and 30 V and show CHECK. Apply the recommended value and push Enter to accept.

Copyright 2025 Baker Hughes Company. 42 | UPS4E Instruction Manual­English

4.12 Calibration Restore

កាល់
rESto

Calibration Restore

MODE ADV

Figure 4-6: Calibration Restore
This option restores the Calibrator calibration to it’s factory conditions. If the factory calibration conditions have been damaged or are missing, then the screen will show `FAIL’. If this happens, contact Druck for more details. If the restore works correctly, then the Calibrator will return to its factory calibration conditions.
1. When you select this option the primary reading will show `rESto’. Push the Enter button and the Calibrator will ask for a four digit PIN 4321.
2. Use the Navigation Pad up, down and left and right buttons to select and change the digits. Select Enter to accept. The primary reading will show `PASS’.

Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | 43

Chapter 4. Calibration Procedures
Copyright 2025 Baker Hughes Company. 44 | UPS4E Instruction Manual­English

5. លក្ខណៈបច្ចេកទេស
Refer to the product Datasheet for a full specification of the Calibrator:
Druck.com
The Calibrator is suitable for indoor use with the following environmental requirements. It is also permitted to use outdoors as a portable instrument if the environmental requirements are met.

ធាតុ

ការពិពណ៌នា

វិមាត្រ

145 mm high x 73 mm wide x 25 mm

ទម្ងន់

318 ក្រាមរួមទាំងថ្ម

អេក្រង់បង្ហាញ

Custom Segment Monochrome Liquid Crystal Display

ជួររង្វាស់

0 mA to +/-24 mA +/-4 mA to +/-24 mA 0 V to +/-30 VDC

Internal Current Source

24 V and 24 mA maximum

External Loop Supply limit

អតិបរមា 30 V

Continuity Range

<100 ohm short circuit >1000 ohm open circuit

Internal HART loop resistor

Nominal 250 ohm

ជួរសីតុណ្ហភាពប្រតិបត្តិការ

-10 ទៅ 50°C (14 ទៅ 122°F)

ជួរសីតុណ្ហាភាពផ្ទុក

-20 ទៅ 70°C (-4 ទៅ 158°F)

ការការពារការចូល

IP54

ជួរសំណើម

0 to 90% relative humidity (RH) non-condensing.

ញ័រ / ញ័រ

MIL-PRF-28800F for class 2 equipment.

សញ្ញាបត្របំពុល

2

EMC

Electromagnetic Compatibility: EN 61326-1:2013

សុវត្ថិភាពអគ្គិសនី

EN 61010-1:2010 EN 61010-2-030:2023

ការអនុម័ត

CE សម្គាល់

ថាមពលថ្ម

4 x Type AA Cells. Nominal voltage: 1.5 V each cell. Alkaline only. Not rechargeable.

ឧបករណ៍ភ្ជាប់

2 x 4 mm sockets

Linear Function preset steps

4, 8, 12, 16 and 20 mA (4-20 mA range) 0, 5, 10, 15 and 20 mA (0-20 mA range)

Flow Function preset steps

4, 5, 8, 13 and 20 mA (4-20 mA range) 0, 1.25, 5, 11.25 and 20 mA (0-20 mA range)

Valve Function preset steps

3.8, 4, 4.2, 12, 19, 20 and 21 mA (4-20 mA range) 0, 0.2, 10, 19, 20, 21 mA (0-20 mA range)

Auto Step, Ramp and Auto Ramp 1 to 599 seconds time

Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | 45

ជំពូកទី 5. លក្ខណៈជាក់លាក់
Item Auto Power Off
កំណត់ហេតុទិន្នន័យ
Calibration Enter PIN Calibration Restore PIN Firmware Update PIN USB

Description After 10 minutes of no button pushes (if enabled). Disabled if data logging or if USB power connected.
Interval up to 1 min Duration up to 99 hours and 59 minutes CSV file ទម្រង់
4321
4321
5487
Type C connector. USB 2 specifications. Power source to UPS4E must be 500 mA or greater.

Copyright 2025 Baker Hughes Company. 46 | UPS4E Instruction Manual­English

ការសម្អាត
6. ការថែទាំ
THE EQUIPMENT CONTAINS NO USER SERVICEABLE PARTS. INTERNAL COMPONENTS MAY BE UNDER PRESSURE OR PRESENT OTHER HAZARDS. SERVICING, MAINTAINING, OR REPAIRING THE EQUIPMENT MAY RESULT IN DAMAGE TO PROPERTY AND SERIOUS PERSONAL INJURY (INCLUDING DEATH). THEREFORE IT IS PARAMOUNT THAT SERVICE ACTIVITIES ARE UNDERTAKEN ONLY BY A DRUCK AUTHORIZED SERVICE PROVIDER. REPAIR ACTIVITIES UNDERTAKEN BY UNAUTHORIZED PERSONNEL MAY INVALIDATE THE EQUIPMENT WARRANTY, SAFETY APPROVALS AND DESIGN CONDITION. DRUCK CANNOT BE HELD LIABLE FOR ANY DAMAGES (INCLUDING DAMAGE TO THE EQUIPMENT), MONETARY FINES, PROPERTY DAMAGE OR PERSONAL INJURY (INCLUDING DEATH) THAT MAY OCCUR DURING OR AS A RESULT OF SERVICE MAINTENANCE OR REPAIR WORK UNDERTAKEN BY AN UNAUTHORIZED SERVICE PROVIDER.
6.1 Cleaning CAUTION Do not use solvents or abrasive materials.
Clean the case and screen with a lint-free cloth and a weak detergent solution.
6.2 Installing and Changing the Batteries
Figure 6-1: Installing the Batteries Note: Replace all four batteries at the same time. 1. Use a Pozidriv screwdriver to unscrew the captive screws that secure the battery
compartment cover. See Figure 6-1. 2. Remove exhausted batteries and insert new batteries into the compartment. Make sure
battery polarities are as shown on the battery cover. Refer to “Specifications” on page 45 for battery types.
Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | 47

ជំពូកទី 6. ការថែទាំ
3. Replace the battery cover and secure the screws. 4. Push the Power button to check that the Calibrator works.
CAUTION Remove or replace the batteries immediately if they are exhausted.
Recycle them according to your local regulations.
6.3 ការត្រួតពិនិត្យការបន្ត
ខ្លី
MODE ADV
Figure 6-2: Continuity Check There is no adjustment for the continuity calibration, so this is a simple procedure to make sure the continuity check works correctly. 1. Set the Calibrator to do a Continuity test. See “Voltage Measuring and Continuity Test
Modes” on page 22. 2. Using the leads supplied with the Calibrator, connect both leads together and check that the
Calibrator indicates a short circuit and the buzzer operates. 3. Disconnect the leads and make sure the Calibrator shows open circuit. 4. If the test fails, check the leads for damage and replace if necessary.
6.4 ការធ្វើបច្ចុប្បន្នភាពកម្មវិធីបង្កប់
Copyright 2025 Baker Hughes Company. 48 | UPS4E Instruction Manual­English

ការធ្វើបច្ចុប្បន្នភាពកម្មវិធីបង្កប់
There are two ways to update the firmware in the Calibrator. For both, you will need to connect the Calibrator to a suitable PC using the USB cable supplied with the Calibrator. 6.4.1 Method 1. 1. Your PC must have an internet connection. 2. Go to the UPS4E download portal on the Druck website and follow the instructions you find
there. https://druck.com/software
6.4.2 Method 2. 1. You must have a copy of the latest firmware downloaded from the Druck webគេហទំព័រនៅ
https://druck.com/software and stored on your PC. The files can be in a zipped format, so you must unzip (extract) them before use. 2. Turn on the Calibrator. 3. Connect the Calibrator to the PC. The screen of the Calibrator will show the USB symbol
and `VCP’.
4. Check the Settings on the Calibrator to make sure it is set for MSC. See “Setting – USB” on page 16.
Figure 6-3: Typical PC Explorer Application Showing the Firmware file 5. នៅលើកុំព្យូទ័រ សូមបើក a file explorer application. You will see the Calibrator as a mass storage
device (UPS4E (D:) in the image), containing the Datalog folder and the firmware (*.raw) file (DK0543.raw in this example). Delete the existing firmware file from the Calibrator and replace with the latest firmware file. 6. On the Calibrator, press the up and down Navigation Pad buttons together to open the Settings. Refer to “To Open the Settings” on page 15. 7. Use the Navigation buttons to select the FW UPDATE. 8. The screen will show the lock symbol asking for you to enter the four digit PIN number 5487. Use the Navigation Pad to select each digit and change its value. Then push Enter. 9. The primary reading will show Enter. 10. Push the Enter button. The Calibrator will restart, then start to load the new firmware while showing a counter that goes from 0 to 99%. It will then restart again.
Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | 49

ជំពូកទី 6. ការថែទាំ

6.5 Error Codes and Warnings
The Calibrator can show one of several error codes and warnings on the screen. These codes and warnings can warn you that you are using the Calibrator incorrectly or need to retry an operation.
6.5.1 លេខកូដកំហុស

Error Code E01, E02, E03, E06 E04 E05 E07 E13 E14 E15
អ៊ី២៦

មូលហេតុនិងវិធីព្យាបាល
Software and hardware failures. Contact Druck service.
វ៉ុលtage out of range. Apply 0 to +/- 30 V
Current out of range. Apply 0 to +/- 24 mA
Firmware upgrade error. Try the upgrade again.
Calibration error. Try the calibration again.
PIN error. Try again with the correct PIN.
Data logging error. Check that the Datalog folder (memory) has sufficient space for the Datalog file you are about to create. Delete unwanted Datalog files.
Check the external voltage applied, it should be within +/30 V. Push Enter to acknowledge the error. Contact Druck Service if the error persists.

6.5.2 ការព្រមាន

Warnings Lo Bat CAL DUE
No Fil

មូលហេតុនិងវិធីព្យាបាល
Battery level critical. Change to new batteries or connect USB power source.
Shown on the screen if it has been more than 365 days from the last device calibration date. We recommend that you calibrate the instrument when you see this, but this warning does not stop device operation.
(ទេ file found) errors. Shown on the screen if the firmware update cannot be done either due to low battery or no firmware file found in UPS4E.
Change for fresh batteries if necessary or connect to a USB power source. Copy the correct firmware file again to the device.

6.6 Instrument Return
6.6.1 Return Goods/Material Procedure If the unit requires calibration or is unserviceable, return it to the nearest Druck Service Centre. See rear page. Contact the Service Department to obtain a Return Goods/Material Authorization (RGA or RMA). Provide the following information for a RGA or RMA: · Product (e.g. UPS4E) · Serial number.

Copyright 2025 Baker Hughes Company. 50 | UPS4E Instruction Manual­English

Packaging for Storage or Transportation
· Details of defect/work to be undertaken. · Calibration traceability requirements. · Operating conditions.
6.7 ការវេចខ្ចប់សម្រាប់ការរក្សាទុក ឬដឹកជញ្ជូន
1. Suitably pack the Calibrator and its cables. Re-use the original packaging if possible. 2. To return the instrument for calibration or repair, complete the return goods procedure. See
Section 6.6. Return the instrument to the manufacturer or an approved service agent for all repairs. Refer to “Specifications” for storage and transportation conditions.
Copyright 2025 Baker Hughes Company. English­UPS4E Instruction Manual | 51

ជំពូកទី 6. ការថែទាំ
Copyright 2025 Baker Hughes Company. 52 | UPS4E Instruction Manual­English

ទីតាំងការិយាល័យ

https://druck.com/contact
សេវាកម្ម និងទីតាំងជំនួយ

https://druck.com/service

Copyright 2025 Baker Hughes Company. This material contains one or more registered trademarks of Baker Hughes Company and its subsidiaries in one or more countries. All thirdparty product and company names are trademarks of their respective holders.
183M0498 Revision – | English