លក្ខណៈបច្ចេកទេស

សមាសភាគ ដំណើរការ ដែនកំណត់
PRO Power Unit ការលាងសម្អាតដៃ 50°C; 30 mins
ULTRA Power Unit Autowasher / Disinfector / Dishwasher 51 – 65°C; 12 mins
All CleanSpace Filters Thermal Disinfection or Washing Cycle 66 – 80°C; 10 mins
ការគាំទ្រក 81 – 90°C; 3 mins Contact CleanSpace
របាំងពាក់កណ្តាល 91 – 95°C; 5 mins Contact CleanSpace

CleanSpace Respirators

Free the way you breathe™

សេចក្តីផ្តើម

This supplement is used to check washing and disinfecting processes for material compatibility with CleanSpace respirator components. Where biological contaminants are a risk, cleaning and disinfecting a CleanSpace respirator is advisable. CleanSpace recommends deferring to the regulatory authority or standards in your region to ascertain the efficacy of microbiological removal. CleanSpace has not specifically evaluated the effectiveness of any agents with regards to inactivating viruses on the equipment.

សេចក្តីណែនាំអំពីការសម្អាត

  • Consider rinsing and drying components thoroughly to avoid any residue that may cause irritation to the wearer after use of any non-neutral cleaning or disinfectant agent treatment (including alcohol).
  • CleanSpace filter media may not be cleaned or disinfected. If a filter change is not required, in a healthcare setting, it is recommended to keep the filter in place. The external hard surface of the filter may be wiped down with a clean damp cloth soaked in a compatible solution or wipe.
  • Cleaning with solvents is not recommended. Solvents could cause damage to plastic components including cracking, fogging, fading, and decreased strength. If equipment is exposed to solvents, rinse well and check parts for cracking or fogging.

ការណែនាំសម្រាប់ការប្រើប្រាស់

  • This Supplement is used to check washing and disinfecting processes for material compatibility with CleanSpace respirator components.
  • Where biological contaminants are a risk, cleaning and disinfecting a CleanSpace respirator is advisable. CleanSpace recommends deferring to the regulatory authority or standards in your region, to ascertain efficacy of microbiological removal.
  • CleanSpace has not specifically evaluated the effectiveness of any agents with regards to inactivating viruses on the equipment.
  • Consider rinsing and drying components thoroughly to avoid any residue that may cause irritation to the wearer after use of any non-neutral cleaning or disinfectant agent treatment (including alcohol).
  • CleanSpace filter media may not be cleaned or disinfected. If a filter change is not required, in a healthcare setting, it is recommended to keep the filter in place. The external hard surface of the filter may be wiped down with a clean damp cloth soaked in a combatable solution or wipe.
  • Cleaning with solvents is not recommended. Solvents could cause damage to plastic components including cracking, fogging, fading and decreased strength. If equipment is exposed to solvents, rinse well and check parts for cracking or fogging.
  • Please contact the CleanSpace team if you have any questions: sales@cleanspacetechnology.com.
  • Rinse Aid or Drying Aid is not recommended to be used as an additive to the cleaning cycle. Use of a drying aid may discolour or damage the polycarbonate on the mask valve set.
  • The prolonged use of any cleaning/disinfecting agent could potentially shorten the useful life of the products. It is imperative that users perform a visual inspection prior to each use to ensure equipment is free from damage. If any visible deterioration is apparent (cracking, crazing, tears etc.), the component should be discarded and replaced. A slight discolouration of silicone components may occur and is acceptable.
  • Reassemble the mask (including the valve assembly) prior to use according to the User Instructions.
  • Store CleanSpace respirator components in a dry, dust-free environment away from direct sunlight.
  • CleanSpace cannot give assurances that deviations from the procedures listed in this supplement and their effects on the performance or safety of the product will be acceptable.
  • When using detergents, disinfectants, or sterilisation agents, always follow the manufacturer’s User Instructions. In the event of conflict, this supplement takes precedence.

សមាសធាតុ

CleanSpace-CST1001-PAPR-Power-System-image (1) CleanSpace-CST1001-PAPR-Power-System-image (2)

ចំណាំ ១. Do not stack on top of each other or these may deform when heated.

ទិន្នន័យបច្ចេកទេស

HALF MASK
Thermal Disinfection or Washing Cycle Temp ≤ 3 MINS ≤ 5 MINS ≤ 7 MINS ≤ 10 MINS ≤ 12 MINS ≤ 30 MINS
≤ 50 ̊C wash បាទ បាទ បាទ បាទ បាទ បាទ
51 – 65 ̊C wash បាទ បាទ បាទ បាទ បាទ ទេ
66 – 80 ̊C wash បាទ បាទ បាទ បាទ ទេ ទេ
81 – 90 ̊C wash បាទ ការប្រុងប្រយ័ត្ន* ទេ ទេ ទេ ទេ
91 – 95 ̊C wash ការប្រុងប្រយ័ត្ន* ការប្រុងប្រយ័ត្ន* ទេ ទេ ទេ ទេ
Drying Cycle Temp ≤ 30 MINS
≤ 90 ̊C drying បាទ
≤ 95 ̊C drying ការប្រុងប្រយ័ត្ន*
 Contact CleanSpace for further information.

ព័ត៌មានទំនាក់ទំនង

  • អូស្ត្រាលី (ការិយាល័យកណ្តាល): +61 2 8436 4000 | sales@cleanspacetechnology.com
  • មិនអាចគ្រប់គ្រងបាន ប្រសិនបើទាញយក ឬបោះពុម្ពពីអ៊ីនធឺណិត។
  • www.cleanspacetechnology.com

CleanSpace-CST1001-PAPR-Power-System-image (3)

ព័ត៌មានទំនាក់ទំនង
Please contact the CleanSpace team if you have any questions: sales@cleanspacetechnology.com.

សំណួរគេសួរញឹកញាប់

Can CleanSpace filters be cleaned or disinfected?

No, CleanSpace filter media may not be cleaned or disinfected. If a filter change is not required, it is recommended to keep the filter in place and wipe the external hard surface with a clean damp cloth soaked in a compatible solution or wipe.

What should I do if my equipment is exposed to solvents?

If equipment is exposed to solvents, rinse well and check parts for cracking or fogging.

Is it safe to use Rinse Aid or Drying Aid in the cleaning cycle?

No, Rinse Aid or Drying Aid is not recommended as it may discolour or damage the polycarbonate on the mask valve set.

What should I do if I notice visible deterioration on my equipment?

If any visible deterioration is apparent (cracking, crazing, tears, etc.), the component should be discarded and replaced.