KERN TDS Platform Scale

Overview
The KERN & SOHN DS platform scale (TDS-A / TDS-B, operating instructions version 2) provides setup, weighing/counting applications, and data output. Read the document and follow the safety precautions before using the device.
Product specifications (from provided source text)
| Platform scale model series | DS |
| Type series | TDS-A; TDS-B |
| Document title/version | Operating instructions (Translation of the original operating instructions), Version 2 (2026-08) – en_GB |
Setup and use
- Select the place of use: choose a location meeting the environment requirements and stable operating conditions.
- Prepare installation: bring the device and the required scope-of-delivery parts/tools/documentation to the place of use.
- Remove transport safety lock: do not remove screws other than those of the transport safety lock. Remove the transport safety lock according to the model’s lock type. Ensure corner load stops are not removed (where applicable).
- Install weighing plate and level the device: position and level the scale by adjusting the levelling feet until the level indicator’s air bubble is within the specified circle.
- Choose power method:
- Mains operation: (a) fit the correct power plug adapter if provided, (b) connect the mains cable to the device, then insert the plug into the socket.
- Rechargeable battery pack (factory option): charge the battery before first use by connecting the power supply/charger to the scale.
- Commissioning: allow the scale to reach operating temperature; during acclimatisation, keep the scale connected to the power supply.
- Open setup menu: in operating mode, press and hold [TARE] and [ON/OFF] simultaneously until the setup menu is shown.
- Run adjustment (from the setup menu): ensure the weighing plate is free of any load, confirm, place the adjustment weight centrally, confirm, remove the adjustment weight; if an adjustment error occurs, restart and repeat.
- Return to operating mode: after successful adjustment the scale switches to operating mode.
ntroduction
1.1 Validity
This document is an integral part of the product. It is valid for
the product versions listed below.
You can recognise which product version your product is on
the type plate. The following illustration is an example. The
real type plate may differ from the illustration. The model
number is shown in <Model> and the type item no. is shown
in <Type>.
Fig. 1: Example of the type plate (real type plate may be
different)
Model Item no. / Type
DS 3K0.01S TDS 3K0.01S-A
DS 8K0.05 TDS 8K0.05-A
DS 10K0.1S TDS 10K0.1S-A
DS 16K0.1 TDS 16K0.1-A
DS 20K0.1 TDS 20K0.1-A
DS 30K0.1 TDS 30K0.1-A
DS 30K0.1L TDS 30K0.1L-B
DS 36K0.2 TDS 36K0.2-A
DS 36K0.2L TDS 36K0.2L-B
DS 60K0.2 TDS 60K0.2-B
DS 65K0.5 TDS 65K0.5-B
DS 100K0.5 TDS 100K0.5-B
DS 150K1 TDS 150K1-B
Introduction
8 TDS-A_TDS-B-BA-e-2620
References in the text
Instructions
1.2 Document purpose
This document provides the user with descriptions for the
safe and efficient use of the product. The document must be
carefully read and understood by all users before the product
is used.
1.3 Other versions of the document
The original version of this document was created in German.
We reserve the right to make subsequent changes to the
product documentation. The current version and other language
versions of the document can be found on our website
⮫ www.kern-sohn.com.
You can find more interesting information about our products
in the download area:
■ Technical data sheets
■ Product brochures and catalogues
■ Information on the KCP protocol and software
1.4 Presentation conventions
Element Meaning
Example: [TARE] Labelling of keys
Example: <SEtUP> Text on the device display
* Asterisk * indicates default
settings
Instructions are presented as step-by-step guidance.
1. This is step 1.
2. This is step 2.
? This is the result of step 2.
Introduction
TDS-A_TDS-B-BA-e-2620 9
Scale operation
Other representations
Element Meaning
left = action, right = result
(the scales and displays
shown may differ from this
example)
Navigation keys for input on
the device (the keys shown
may differ from those in this
example)
Here you will find additional information or tips.
General safety instructions
10 TDS-A_TDS-B-BA-e-2620
Meaning of the signal words
2 General safety instructions
Read this document thoroughly and carefully before using the
device and observe all safety precautions contained herein
to avoid bodily injury and damage to property. Keep this
document in a place that is always easily accessible so that it
is quickly available when needed.
2.1 Meaning of the warnings
Warnings warn you of possible safety risks during certain
activities. Read the warnings in full and follow the instructions.
Failure to do so may result in injury, damage to
property, environmental damage, malfunctions or incorrect
results.
Signal word Meaning
DANGER This combination of symbol and
signal word indicates an immediate
dangerous situation that will
result in death or serious injury if
it is not avoided.
WARNING This combination of symbol and
signal word indicates a possible
dangerous situation that can
result in death or serious injury if
it is not avoided.
CAUTION This combination of symbol and
signal word indicates a possible
dangerous situation that can
result in minor injury if it is not
avoided.
NOTICE This combination of symbol and
signal word indicates a possible
dangerous situation that can
result in material and environmental
damage if it is not avoided.
ENVIRONMENT This combination of symbol and
signal word indicates a possible
dangerous situation that can
result in material and environmental
damage if it is not avoided.
General safety instructions
TDS-A_TDS-B-BA-e-2620 11
Meaning of the warning signs
Safe handling of power supply
units and power supplies
Warning signs Type of danger
Warning – danger zone.
Warning – dangers due to batteries.
Warning – high-voltage.
2.2 Safe handling of packaging material
Always keep packaging material away from children and animals
to avoid the danger of suffocation.
2.3 Safe handling of electrical devices
Improper handling of mains-powered devices or the use
of damaged or unsuitable power supply units or mains
cables can lead to life-threatening injuries. Therefore, please
observe the following instructions:
■ If the electrical equipment is defective (e.g. damage to the
power supply unit or mains cable, deformation, smoke,
etc.), the device must not be used. If you notice any
damage, disconnect the device from the power supply
immediately and contact your dealer or the technical
service of KERN & SOHN.
■ Only connect the device to a properly installed and
undamaged power supply. The information on the device
(type plate) and the mains voltage at the place of use must
be identical.
■ To prevent short circuits, ensure that no moisture (e.g.
steam or liquids) is present near the electrical equipment
at the place of use.
General safety instructions
12 TDS-A_TDS-B-BA-e-2620
Safe handling of batteries and
rechargeable batteries
■ Only use original power supply units and connecting
cables. If you need a replacement, please contact your
dealer or the technical service of KERN & SOHN.
■ Any work on the electrical equipment of the device may
only be carried out by the technical service of KERN &
SOHN.
■ The mains cable must not be pinched or kinked and must
be laid in such a way that it does not pose a tripping
hazard.
■ In case of an emergency, it must be possible to disconnect
the device from the power supply without any problems.
Batteries and rechargeable batteries may catch fire, emit
toxic fumes, release corrosive liquids or explode if not used
correctly. Therefore, observe the instructions in the following
section.
Damage
■ Do not use damaged or deformed batteries and rechargeable
batteries.
■ Protect batteries and rechargeable batteries from mechanical
damage, heat, cold, water and chemicals.
■ Protect batteries and rechargeable batteries from short
circuits. Never touch the terminals with metal objects.
■ When inserting batteries and rechargeable batteries,
ensure that the polarity (positive and negative terminals) is
correct and that you are using suitable battery types. Only
use types recommended by the manufacturer.
■ Batteries and rechargeable batteries must not be modified.
Charging batteries
■ Chargers must not be modified.
■ Only use the original charger.
■ Observe the charging time specified in the technical data
and avoid overcharging.
■ Do not charge batteries that are deformed or show other
signs of damage.
■ In the event of heat build-up, deformation, smoke or burnt
smell, disconnect the charger immediately and remove it
from service.
General safety instructions
TDS-A_TDS-B-BA-e-2620 13
Transport
■ When transporting batteries or rechargeable batteries, the
applicable national and international regulations must be
observed. Ensure that batteries or rechargeable batteries
are protected against short circuits and damage. Do not
ship defective batteries or devices containing defective
batteries.
Storage
■ Charge the battery if you will not be using it for a long
time.
■ Remove the batteries or rechargeable batteries if the
device is not to be used for a longer period of time.
■ Protect batteries and rechargeable batteries from short
circuits (e.g. by covering the terminals) and store them
separately by type. Preferably use a fireproof container for
storage.
■ Store batteries and rechargeable batteries at room temperature
in a dry place. Avoid extreme cold or extreme
heat in the storage environment. Also observe the recommended
ambient temperature stated in the technical data.
Battery fluid
■ Avoid contact between leaking fluid and skin, eyes or
clothing.
■ Wear protective clothing/equipment when removing a
defective battery and ensure the room is well ventilated.
■ Clean skin or clothing that has come into contact with
battery fluid thoroughly with soap suds and then rinse with
clean water.
■ In case of contact with eyes, rinse immediately with clean
water and then get medical help.
Disposal
■ Do not dispose of batteries and rechargeable batteries
in household waste, but at collection points or recycling
centres.
■ Ensure that the poles are not short-circuited during disposal.
■ Further information on disposal: ⮫ ‘Disposal of batteries
and rechargeable batteries’ on page 87
General safety instructions
14 TDS-A_TDS-B-BA-e-2620
Intended use of the scale
Improper use of the scale
2.4 Notes on IT security
This device is equipped with an interface for data transmission.
Unauthorised persons may be able to gain access to the
data being exchanged. It is the operator’s responsibility to
identify the potential IT risks at the place of use and to take
appropriate countermeasures where necessary.
2.5 Intended use
This scale is used to determine the weight of items to
be weighed. This scale is a “non-automatic weighing instrument”.
This means that the material to be weighed must be
placed manually, with care and as centred as possible on the
weighing plate. Once a stable weight value is displayed, the
result can be read.
2.6 Improper use
■ Our scales with strain gauges are not intended for use in
dynamic weighing processes.
However, the scales can also be used for dosing applications
after checking the individual area of application and
the accuracy requirements.
You can find out whether your scale is equipped with
strain gauge technology by checking the technical data.
⮫ Chapter 18.1 ‘Technical data’ on page 92
■ To prevent damage, avoid impact loadings and do not
exceed the maximum load limit (Max.) of the scale.
■ Any modification to the design of the scale will invalidate
the warranty and may lead to incorrect weighing results
and safety-related defects. Such modifications are therefore
prohibited.
■ The scale may only be used in accordance with the specifications
described in this document.
■ The scale is not designed for weighing people. It is not
suitable for medical purposes.
2.7 Obligations of the operating company
It is the responsibility of the operating company to comply
with the specifications described in this document as well
as with legal and official regulations at the place of use.
Any failure to comply with the safety precautions and instrucGeneral
safety instructions
TDS-A_TDS-B-BA-e-2620 15
tions described in this document is considered improper use.
KERN & SOHN is not liable for any damage resulting from
improper use.
2.8 Required user qualifications
■ Assembly, commissioning, servicing and maintenance of
the device may only be carried out by qualified personnel.
A qualified person is a person with appropriate professional
training or qualifications for the activities described.
They have in-depth knowledge of the tasks and the relevant
safety standards.
■ Only trained users may operate, handle and maintain the
device. A trained user is a person who has been trained
in the use of the device and in the corresponding safety
instructions.
2.9 Requirements for the environment of use
In addition to the environmental information in the technical
data, observe the following requirements for the environment
of use:
■ The device must not be operated in environments where
there is a risk of corrosion.
■ The device must be protected from excessive humidity,
steam, spray and dust.
■ The device must be protected from excessive moisture
caused by condensation. Condensation can occur if the
device is moved from a cold environment to a much
warmer environment. If such a change of environment
cannot be avoided, allow the device to acclimatise for
approx. 2 hours before starting it up.
■ The device must not be used in environments with aggressive
chemicals, gases or liquids.
■ The device must not be used in potentially explosive
atmospheres or environments where explosives are
present.
■ The device must not be used in environments subject to
severe impacts or vibrations.
■ The device must not be exposed to extreme heat or strong
temperature fluctuations (e.g. by placing the device next
to a heater or in direct sunlight).
■ Protect the device from direct draughts.
General safety instructions
16 TDS-A_TDS-B-BA-e-2620
■ Select a location with a stable power supply.
■ When choosing the place of use, consider the surrounding
environment. Electromagnetic fields and electrostatic
charges can affect the weighing result or cause damage.
Description of the device
TDS-A_TDS-B-BA-e-2620 17
Description of the device

3.1 General product description
This scale is a platform scale. Platform scales are designed
for use in industry or logistics.
3.2 Overview of the device
1 Weighing plate
2 Mains adapter connection
3 KUP connection (KERN Universal Port)
4 Display
5 Keys
6 Battery compartment
7 Level indicator
Description of the device
18 TDS-A_TDS-B-BA-e-2620
TDS 36K0.2-A
TDS 8K0.05-A
TDS 16K0.1-A
TDS 20K0.1-A
TDS 30K0.1-A
TDS 3K0.01S-A
TDS 10K0.1S-A
TDS 30K0.1L-B
TDS 36K0.2L-B
TDS 60K0.2-B
TDS 65K0.5-B
TDS 100K0.5-B
TDS 150K1-B
3.3 Overview of the keys
Key Designation in the
text
Short keystroke Long keystroke
[ON/OFF]
or [ç]
■ Switch on the scale
■ Switch backlight on/off
(only on devices with
backlighting)
■ Move up one menu level /
Exit menu
■ Previous digit position
■ Switch off the scale
Description of the device
TDS-A_TDS-B-BA-e-2620 19
Key Designation in the
text
Short keystroke Long keystroke
[CHANGE]
or [é]
■ Previous menu item
■ Increase digit
Application: Weighing
■ Switch between weighing
units (when pressed for
the first time: set the secondary
weighing unit to
be displayed)
Application: Counting
■ Switch between number
of pieces and weight
(when pressed for the
first time: set the secondary
weighing unit to
be displayed. Note: If
pcs is selected, a reference
must then be established.)
Application: Tolerance
weighing
■ Switch between weighing
units (when pressed for
the first time: set the secondary
weighing unit to
be displayed)
Application: Weighing
■ Display gross weight
■ Display tare weight
Application: Counting
■ Display gross weight or
gross quantity
■ Display tare weight or
tare quantity
Application: Tolerance
weighing
■ Switch the display
between gross, net and
tare weight
[PRINT]
or [è]
■ Send data to a connected
peripheral device
■ Select menu item and
move to the next menu
level / Confirm selection
■ Confirm digit and move to
the next digit position
[ZERO] ■ Zero setting
Description of the device
20 TDS-A_TDS-B-BA-e-2620
Key Designation in the
text
Short keystroke Long keystroke
[TARE]
or [ê]
■ Tare
■ Next menu item
■ Decrease digit
■ Open the application
menu
3.4 Overview of the display
Display / Icon Designation Description
Indicator: Stability Displayed when the weight
value is stable
Indicator: Zero Displayed while the scale is at
zero
Indicator: Negative Displayed when the weight
reading is negative
“HI” tolerance mark Upper tolerance limit exceeded
“OK” tolerance mark Weight reading within the tolerance
limits
“LO” tolerance mark Value below the lower tolerance
limit
Indicator: Pre-tare Displayed when pre-tare is
active
Indicator: Hold Displayed when weighing
values are held
Indicator: Net The displayed value is the net
weight (shown after taring)
Indicator: Gross weight The displayed value is the
gross weight (shown after zero
setting)
Indicator: Auto-print Displayed when the auto-print
function is active
Indicator: Total Displayed when the totalizing
function is used
Description of the device
TDS-A_TDS-B-BA-e-2620 21
Display / Icon Designation Description
Indicator: Data transmission Displayed during data transfer
Indicator: Wi-Fi Displayed when connected to
Wi-Fi
Character on the main display The main display shows the
results or messages generated
by the scale
Units Active unit displayed on the
right-hand side of the display
Battery charging indicator Displayed when running on
battery power
Unpacking, checking and transporting the device
22 TDS-A_TDS-B-BA-e-2620
4 Unpacking, checking and transporting the device
4.1 Unpacking and checking the device
WARNING
Injuries due to heavy weight of the device
Lifting heavy loads can lead to injuries and cause
permanent health damage (e.g. damage to the
spine)
− Only remove the device from the packaging
with the help of another person.
− Use a suitable lifting device to remove or
transport the device. Only use lifting equipment
if you are appropriately qualified and
trained. If you do not have the necessary qualifications,
ask trained persons to help you.
Secure the device properly.
WARNING
Danger of suffocation due to packaging material
Unattended packaging material can become a
suffocation hazard for children or animals.
− Do not leave packaging material unattended,
especially if children or animals are around.
− Save the original packaging and its components
for transport, storage or a possible
return.
− Contact your dealer if you notice any
missing parts or damage. Information on
warranty and goods return can be found in
⮫ Chapter 15.1 ‘Warranty’ on page 86.
1. Check the packaging for visible external damage that
may have occurred during shipment.
Unpacking, checking and transporting the device
TDS-A_TDS-B-BA-e-2620 23
2. Unpack all parts of the delivery (⮫ Chapter 4.2 ‘Scope
of delivery’ on page 23) and remove the packaging
materials.
3. Check that all parts of the delivery are complete and
undamaged.
? You can now transport the device to the place of use.
⮫ Chapter 4.3 ‘Transporting the device’ on page 23
4.2 Scope of delivery
■ Scale
■ Power supply unit
■ Power plug adapter set
■ Operating instructions
4.3 Transporting the device
WARNING
Injuries due to heavy weight of the device
Lifting heavy loads can lead to injuries and cause
permanent health damage (e.g. damage to the
spine)
− Only remove the device from the packaging
with the help of another person.
− Use a suitable lifting device to remove or
transport the device. Only use lifting equipment
if you are appropriately qualified and
trained. If you do not have the necessary qualifications,
ask trained persons to help you.
Secure the device properly.
Unpacking, checking and transporting the device
24 TDS-A_TDS-B-BA-e-2620
NOTICE
Damage due to improper transport
Incorrect transport can damage the device.
General information on transporting scales
− Remove all loads from the weighing plate
before transporting the scale.
− Remove all loose components from the device
before transporting it.
− Re-attach the transport safety lock before
transporting the device. This applies in particular
to transport in vehicles.
− Use the original packaging to transport the
device over long distances.
− When transporting batteries or rechargeable
batteries, the applicable national and international
regulations must be observed. Ensure
that batteries or rechargeable batteries are
protected against short circuits and damage.
Do not ship defective batteries or devices
containing defective batteries.
− When transporting the scale, hold it by the
housing and not by the weighing plate or the
weighing plate support.
− Use suitable lifting and transport equipment
to move the scale. Secure the scale before
transport to prevent it from falling.
Assembly, installation and commissioning
TDS-A_TDS-B-BA-e-2620 25
5 Assembly, installation and commissioning
5.1 Selecting the place of use
The scales are designed to achieve reliable results under
normal operating conditions. You can work accurately and
quickly if you choose the right location for using your scale.
Please note when choosing the location:
■ ⮫ Chapter 18.1 ‘Technical data’ on page 92
■ ⮫ Chapter 2.9 ‘Requirements for the environment of use’
on page 15
5.2 Installing the device
5.2.1 Safety during installation
DANGER
Electric shock due to live parts
Contact with live parts can lead to electric shock
and burns.
− Installation, cleaning, maintenance and servicing
work may only be carried out by qualified
personnel.
− Disconnect live parts from the power supply
or unplug the power plug from the mains.
− Only carry out installation, cleaning, servicing
and maintenance work on devices that
are switched off and disconnected from the
power supply, unless otherwise specified in
the relevant instructions.
− Follow the instructions in this document.
5.2.2 Preparing for installation
1. Take the device to the place of use (⮫ Chapter 4.3
‘Transporting the device’ on page 23, ⮫ Chapter 2.9
‘Requirements for the environment of use’ on page 15)
2. Ensure that all parts included in the scope of delivery,
any tools that may be required, and the relevant documentation
are at hand.
Assembly, installation and commissioning
26 TDS-A_TDS-B-BA-e-2620
5.2.3 Removing the transport safety lock
NOTICE
Do not remove any screws other than those of
the transport safety lock, as this may damage the
scale.
. Remove the transport safety lock as shown in the table
below.
Models
■ DS 3K0.01S
■ DS 10K0.1S
Models
■ DS 30K0.1L
■ DS 36K0.2L
■ DS 60K0.2
■ DS 65K0.5
■ DS 100K0.5
■ DS 150K1
Transport lock with one screw
The transport safety lock is located on the
bottom side of the platform and is marked with
an arrow.
Assembly, installation and commissioning
Installing the weighing plate
1. Remove the weighing plate from its packaging.
2. Place the weighing plate on the scale. Make sure it is
positioned correctly.
? You have successfully installed the weighing plate.
5.2.5 Levelling the device
− Select a suitable place of use for the scale.
⮫ Chapter 2.9 ‘Requirements for the environment
of use’ on page 15
− Check the levelling regularly.
− Level the scale after moving it to a new
location.
In design type 1, the level indicator is mounted
on a bolted sheet metal plate. In design types
2 and 3, the level indicator is located on the
platform underneath the weighing plate.
For accurate measurements, it is necessary for the scale to
be levelled at its place of use.
1. Position the scale at the place of use.
2. Remove the weighing plate.
Assembly, installation and commissioning
28 TDS-A_TDS-B-BA-e-2620
Fitting the power plug adapter
3. Adjust the levelling feet until the air bubble of the level
indicator is within the specified circle.
5.2.6 Connecting wired peripheral devices
Please use only KERN accessories and printers.
These are perfectly suited to your scale.
Unplug the scale from the mains before connecting any
peripheral devices (e.g. a printer or a PC).
For information on configuring the interfaces, see ⮫ Chapter
7 ‘Interfaces and data output’ on page 43.
5.2.7 Mains operation
A power plug adapter set is included with the device so that it
can be used flexibly in different locations with different power
plug standards. The installation of the mains plug adapter is
described below.
1
2
1. Press the release tab 1.
Assembly, installation and commissioning
TDS-A_TDS-B-BA-e-2620 29
Connecting the power supply to
the mains
2. Slide the power plug adapter 2 into the power supply
unit using the guide until it clicks into place.
? The power supply unit is now ready for use.
? To remove the power plug adapter, press the release
tab and proceed in reverse order.
WARNING
Serious injuries due to damaged power
supply units
Damaged power supply units can lead to short
circuits, fires and serious injuries.
− Before each use, check the power supply unit
and mains cable for visible damage. Do not
use the power supply unit if it is damaged.
WARNING
Injuries due to incorrect handling of power
supply units
Incorrect handling of power supply units can lead
to short circuits, fires and serious injuries.
− Observe the safety precautions for handling
power supply units. ⮫ ‘Safe handling of power
supply units and power supplies’ on page 11
− Avoid moisture at the mains connection,
power supply unit and socket.
− Only connect the device to a properly
installed socket.
− Make sure that the power supply unit cable is
never kinked or pinched.
− Ensure that the power supply unit is accessible
at all times.
− Ensure that the mains cable does not pose a
tripping hazard.
− Only connect the power supply unit to
the mains if the information on the device
(sticker) and the mains voltage at the place of
use are identical.
Assembly, installation and commissioning
30 TDS-A_TDS-B-BA-e-2620
1. Plug the mains cable connection into the mains connection
of the device (for the mains connection, see
⮫ Chapter 3.2 ‘Overview of the device’ on page 17).
2. Insert the power plug into the socket.
? The device can now be put into operation. ⮫ Chapter
5.3 ‘Commissioning the device’ on page 32
5.2.8 Operation with rechargeable battery pack
WARNING
Injuries due to incorrect handling of batteries
or rechargeable batteries
Incorrect handling of batteries can lead to short
circuits, fires, leaks and injuries.
− Never use batteries or rechargeable batteries
that are damaged or deformed.
− Never touch the terminals of batteries or
rechargeable batteries with metal objects, as
this can lead to short circuits.
− When inserting batteries or rechargeable batteries,
ensure that the polarity is correct (positive
and negative polarity).
− Do not use different types of batteries or old
and new batteries together if several batteries
are required for operation. This can lead to
damage.
− Only replace batteries or rechargeable batteries
with types recommended by the manufacturer.
− Remove the batteries from the device if you
wish to store them for a longer period of time.
This protects the device from damage caused
by any leakage of battery fluid.
The rechargeable battery is only available as a
factory option. For further information, please
visit our website at ⮫ www.kern-sohn.com.
Assembly, installation and commissioning
TDS-A_TDS-B-BA-e-2620 31
Recharge the battery WARNING
Serious injuries due to damaged power
supply units/chargers
Damaged power supply units/chargers can lead
to short circuits, fires and serious injuries.
− Before each use, check the power supply
unit/charger and mains cable for any visible
damage. Do not use the power supply unit/
charger if it is damaged.
WARNING
Injuries due to incorrect handling of power
supply units/chargers
Incorrect handling of power supply units/
chargers can lead to short circuits, fires and
serious injuries.
− Observe the safety precautions for handling
power supply units/chargers. ⮫ ‘Safe handling
of power supply units and power supplies’
on page 11
− Avoid moisture at the mains connection,
power supply unit/charger and at the socket.
− Only connect the device to a properly
installed socket.
− Make sure that the power supply unit/charger
cable is never kinked or pinched.
− Make sure that the power supply unit/charger
is accessible at all times while the battery is
charging.
− Ensure that the mains cable does not pose a
tripping hazard.
− Only connect the power supply unit/charger
to the mains if the specifications on the
device (sticker) match the mains voltage at
the place of use.
Assembly, installation and commissioning
32 TDS-A_TDS-B-BA-e-2620
Measures before commissioning
Commissioning procedure
WARNING
Injuries caused by improper charging of
rechargeable batteries
If batteries are not charged correctly, they may
catch fire or explode.
− Chargers must not be modified.
− Only use the original charger.
− Observe the charging time specified in the
technical data and avoid overcharging.
− Do not charge batteries that are deformed or
show other signs of damage.
− In the event of heat build-up, deformation,
smoke or burnt smell, disconnect the charger
immediately and remove it from service.
− Please observe the general safety instructions
regarding batteries and rechargeable
batteries. ⮫ ‘Safe handling of batteries and
rechargeable batteries’ on page 12
The battery should be fully charged before first use. You can
find the charging time in the technical data ⮫ Chapter 18.1
‘Technical data’ on page 92.
. To charge the battery, connect the power supply or
charger to the scale.
5.3 Commissioning the device
In order to obtain accurate weighing results with electronic
scales, the scale must have reached its operating temperature.
During this acclimatisation period (⮫ Chapter 18.1
‘Technical data’ on page 92), the scale must be connected to
the power supply.
Complete all necessary installation work and preparations
before commissioning.
1. Switch on the device. ⮫ Further information on page 56
2. An adjustment may be required at the installation location
at the beginning. ⮫ Chapter 6.3 ‘Adjustment settings’
on page 35
Settings
TDS-A_TDS-B-BA-e-2620 33
Open the setup menu
Menu navigation
Numerical input
6 Settings
6.1 Opening and using the setup menu
. In operating mode, press and hold [TARE] and [ON/
OFF] simultaneously.
? <SEtuP> à <cAL> is displayed.
? The scale is in setup menu mode.
Key Function
[é] Previous menu item
[ê] Next menu item
[è] Select menu item and move
to the next menu level /
Confirm selection
[ç] Move up one menu level /
Exit menu
Key Function
[é] Increase digit
[ê] Decrease digit
[è] Confirm digit and move to
the next digit position
[ç] Previous digit position
Settings
34 TDS-A_TDS-B-BA-e-2620
Overview of the main menu
6.2 Setup menu overview
The setup menu consists of several levels. The settings apply
globally to all applications.
Menu level Description Further information
<cAL> Adjusting ⮫ Chapter 6.3 ‘Adjustment settings’
on page 35
<coM> Interface parameters ⮫ Chapter 7.3.2 ‘Setting interface
parameters’ on page 44
<Print> Data output and printing ⮫ Chapter 7.4.1 ‘Overview of
output settings’ on page 45
<kEYton> Keypad sound ⮫ Chapter 6.4 ‘Setting the
acoustic signal’ on page 37
<AutoFF> Automatic shut-off function ⮫ Chapter 6.5 ‘Setting the
automatic switch-off function’
on page 38
<buttons> Assignment of keys ⮫ Chapter 6.6 ‘Assigning of key
functions’ on page 38
<bLiGht> Backlighting ⮫ Chapter 6.7 ‘Adjusting the
backlight’ on page 40
<tArErG> Taring range ⮫ Chapter 6.8 ‘Setting the
taring range’ on page 40
<ZLiMit> Zero-setting range ⮫ Chapter 6.9 ‘Setting the zero
range (Zero Limit)’ on page 40
<ZtrAck> Automatic zero tracking ⮫ Chapter 6.10 ‘Setting
automatic zero tracking’
on page 40
<FiLtEr> Speed of display ⮫ Chapter 6.11 ‘Setting the display
speed / filter’ on page 41
<unitS> Available units ⮫ Chapter 6.12 ‘Setting the
availability of display units’
on page 41
<ModES> Available applications ⮫ Chapter 6.13 ‘Setting the
availability of applications’
on page 41
<rESEt> Restore factory settings
(except date and time)
⮫ Chapter 6.14 ‘Restoring factory
settings’ on page 42
Settings
TDS-A_TDS-B-BA-e-2620 35
6.3 Adjustment settings
The acceleration due to gravity varies at different locations
on earth. For exact weighing results, the scale must be
adjusted to the prevailing acceleration due to gravity at the
location where it is installed. The scale must have reached its
operating temperature before it can be adjusted.
Adjustment of the scale is required in the following cases:
■ Before initial commissioning, if the scale has not already
been adjusted to the installation location at the factory.
■ After every change of location.
■ After a significant change in environmental conditions (e.g.
temperature or humidity).
■ If the scale has been disconnected from the power supply.
To ensure accurate measurements, it is also recommended
that the scale is adjusted regularly.
NOTICE
Incorrect weighing results
− Avoid unstable environmental conditions
during adjustment (e.g. vibrations or air
draught).
− Ensure that the scale has reached its operating
temperature. ⮫ Chapter 18.1 ‘Technical
data’ on page 92
− Only carry out the adjusting procedure with
the standard weighing plate in place.
− During external adjusting, ensure that only the
adjustment weight is on the weighing plate
without any other load.
− Only handle adjustment weights with
tweezers, weight handles or gloves. Avoid
direct skin contact between your hands and
adjustment weights.
6.3.1 Adjusting procedure with an external adjustment weight
To carry out the following steps, the scale must be in the
setup menu mode.
Settings
36 TDS-A_TDS-B-BA-e-2620
1. Select <cAL> g <cALEXt> ([é] / [ê]) and confirm
([è]).
? The first applicable adjustment weight is displayed.
2. Select the required adjustment weight ([é] / [ê]) and
confirm ([è]).
(Information on adjustment weights ⮫ Chapter 18.1
‘Technical data’ on page 92).
? <ZEro> is displayed.
3. Make sure the weighing plate is free of any load and
confirm by pressing [è].
? <PutLd> à and the value of the adjustment
weight are displayed.
4. Place the adjustment weight in the centre of the scale.
5. Confirm by pressing [è].
? <WAit> à <rEMVLd> is displayed.
6. Remove the adjustment weight.
? The scale switches to operating mode.
? The adjusting process is complete.
If an adjustment error occurs (e.g. additional
items on the weighing plate), the error message
<WronG> will appear. In this case, restart
the scale and repeat the adjusting procedure.
6.3.2 Adjusting procedure with a user-defined adjustment weight
To carry out the following steps, the scale must be in the
setup menu mode.
Settings
TDS-A_TDS-B-BA-e-2620 37
<kEYton>
1. Select <cAL> g <cALEud> ([é] / [ê]) and confirm
([è]).
? A numerical input field appears, and the active digit
is displayed flashing.
2. Enter the weight value (select a number: [é] / [ê],
next number: [è]) and confirm ([è]).
? <ZEro> is displayed.
3. Make sure the weighing plate is free of any load and
confirm by pressing [è].
? <PutLd> à and the value of the adjustment
weight are displayed.
4. Place the adjustment weight in the centre of the scale.
5. Confirm by pressing [è].
? <WAit> à <rEMVLd> is displayed.
6. Remove the adjustment weight.
? The scale switches to operating mode.
? The adjusting process is complete.
If an adjustment error occurs (e.g. additional
items on the weighing plate), the error message
<WronG> will appear. In this case, restart
the scale and repeat the adjusting procedure.
6.4 Setting the acoustic signal
Settings
■ <on>*: keypad sound enabled
■ <off>*: keypad sound disabled
Settings
38 TDS-A_TDS-B-BA-e-2620
6.5 Setting the automatic switch-off function
<AutoFF>
Submenu Description Settings
<ModE> Set switch-off mode ■ <oFF>*: automatic switch-off function disabled
■ <Auto>: automatic switch-off after the time
set at <tiME>, provided there is no change of
load
■ <onLY0>: automatic switch-off only when
the display shows zero
<tiME> Set switch-off time ■ <30S>*
■ <1Min>
■ <2Min>
■ <5Min>
■ <30Min>
■ <60Min>
6.6 Assigning of key functions
Certain scale keys can be programmed to perform different
functions. This allows quick access to specific functions in
operating mode.
Settings
TDS-A_TDS-B-BA-e-2620 39
<buttonS>
Submenu Description Settings
Key labelling
(e.g. <CHANGE>
([CHANGE]) or
<F-kEY> ([F]),
depending on the
device)
Change the functionality
of the selected
key
<SPuSh>: short
keystroke
■ <toGGLE>: change
the unit displayed
■ <unit>: set weighing
unit
■ <ShoWGt>: display
gross and tare values
■ <hoLd>: execute hold
function (only displayed
when the “Weighing”
application is active)
■ <ModE>: change application
■ <PtArE>: open pretare
settings
■ <chEck>:
When the “Counting”
application is active: set
the limits for piece
counting
When the “Tolerance
weighing” application is
active: open the settings
menu for tolerances
■ <rEF>: set reference
quantity (only displayed
when the “Counting”
application is active)
■ <oFF>: key disabled
<LPuSh>: long
keystroke
Settings
40 TDS-A_TDS-B-BA-e-2620
<tArErG>
<ZLiMit>
<ZtrAck>
6.7 Adjusting the backlight
<bLiGht>
Submenu Description Settings
<ModE> Set the lighting
mode
■ <ALWAYS>*: backlight always on
■ <tiMEr>: backlight switches off automatically
after the time set at <tiME>, provided
there is no change of load
■ <nobL>: backlight off
<tiME> Set the lighting
switch-off time
■ <5S>*
■ <30S>
■ <1Min>
■ <2Min>
■ <5Min>
■ <30Min>
6.8 Setting the taring range
This allows the maximum taring range to be set between 10
% and 100 % (default value = 99 %). To do so, enter the taring
range as a numerical value.
6.9 Setting the zero range (Zero Limit)
This allows the maximum zero range to be set between 1 %
and 100 % (default value = 2 %). To do so, enter the zero
range as a numerical value.
6.10 Setting automatic zero tracking
Settings
■ <on>*: automatic zero tracking (<= 3d) enabled
■ <off>: automatic zero tracking disabled
Settings
TDS-A_TDS-B-BA-e-2620 41
<FiLtEr>
If small quantities of the weighing material are
removed or added, the stability compensation
may produce incorrect results (e.g. in cases
of slow liquid outflow or evaporation). This function
should therefore be deactivated for dosing
operations involving minor weight fluctuations.
6.11 Setting the display speed / filter
This allows the display speed to be set between 0 and 100
(0 = very fast for quiet environments, 100 = very slow for
dynamic environments). To set the filter, enter a numerical
value.
6.12 Setting the availability of display units
These settings allow you to determine which units are available
in the application menu. For example, you can hide units
that are not relevant to the application field.
<unitS>
Submenu Description Settings
For available units,
see technical data
Selection of unit ■ <on>: make a unit available
■ <oFF>: lock unit
6.13 Setting the availability of applications
These settings allow you to determine which applications are
available in the application menu. For example, you can hide
applications that are not relevant to the application field.
Settings for an application can only be adjusted
if it is not actively running in operating mode.
Settings
42 TDS-A_TDS-B-BA-e-2620
<rESEt>
<ModES>
Submenu Description Settings
For available applications,
see technical
data
Selection of application
■ <on>: make application available
■ <oFF>: lock application
6.14 Restoring factory settings
Once you have selected <rESEt>, the message
<SurE> will appear on the display. Once you have confirmed
this message, the scale will switch off automatically.
The next time you start the device, the settings will have
been reset to their factory defaults. However, date and time
will not be reset.
Interfaces and data output
TDS-A_TDS-B-BA-e-2620 43
BalanceConnection
EasyTouch
KUP (KERN Universal Port)
7 Interfaces and data output
7.1 Optional software
BalanceConnection enables communication between scales
and peripheral devices. This software supports various interfaces
and protocols to ensure flexible integration into existing
systems. It is particularly useful for applications where centralised
collection and processing of measurement data is
required.
With EasyTouch, your weighing data is sent directly to your
PC, laptop or tablet for immediate further processing. Easy-
Touch also allows you to add other useful functions to your
scale and tailor it perfectly to your specific field of application.
For further information about KERN software, please visit our
website at ⮫ www.kern-sohn.com.
7.2 Description of the interfaces
You can find all available KUP interface adapters in our online
store at ⮫ www.kern-sohn.com. Installation and setup are
described in the relevant interface adapter documentation.
7.3 Setting up interfaces
7.3.1 Connecting wired peripheral devices
1. Switch off the scale and the peripheral device.
2. Disconnect the scale from the mains voltage.
3. Connect the KUP adapter to the scale.
4. Connect the peripheral device to the scale using a
cable.
5. Reconnect the scale to the mains voltage.
6. Switch on the scale and the peripheral device.
? The interface is now ready for use. You may need to
configure the interface parameters first. ⮫ Chapter
7.3.2 ‘Setting interface parameters’ on page 44
Interfaces and data output
44 TDS-A_TDS-B-BA-e-2620
<coM>
7.3.2 Setting interface parameters
The settings options are only displayed once
the interfaces have been fully initialised. During
initialisation, the message <buSY> is displayed.
This may take a short while.
Only the settings for the connected interface
are displayed.
Submenu Submenu Description Settings
<rS232> or <uSb-d> <bAud> Baud rate ■ <600>
■ <1200>
■ <2400>
■ <4800>
■ <9600>*
■ <14400>
■ <19200>
■ <38400>
■ <57600>
■ <115200>
■ <230400>
<dAtA> Number of
data bits
■ <7dbitS>
■ <8dbitS>*
<PAritY
>
Parity ■ <nonE>*: none
■ <odd>: odd
■ <EVEn>: even
■ <SPACE>: space
■ <MArk>: padding character
<StoP> Number of
stop bits
■ <1Sbit>*
■ <2SbitS>
<hAndSh
>
Handshake ■ <nonE>*
3. Select <PtArE> g <ActuAL> by pressing [é] /
[ê] and confirm ([è]).
? The zero display and the “Pre-tare” indicator ( )
as well as the “Net weight” indicator ( ) are displayed.
? The tare weight has been saved.
1. Press and hold [TARE].
? The application menu opens.
2. Select <PtArE> g <MAnuAL> by pressing [é] /
[ê] and confirm ([è]).
? The display switches to numeric input.
3. Enter the tare weight (select a number: [é] / [ê], next
number: [è]) and confirm ([è]).
? The “Pre-tare” indicator ( ) and the “Net weight”
indicator ( ) are displayed.
? The tare weight has been saved and will be displayed
as a negative value.
? Weighing can now be carried out using the appropriate
container.
8.5 Selecting an application
1. Press and hold [TARE].
? <apcmen> followed by the currently active application
is displayed.
? The scale is in application menu mode.
2. Select <ModE> ([é] / [ê]) and confirm ([è]).
? The available applications are displayed.
Basic operation
60 TDS-A_TDS-B-BA-e-2620
3. Select the desired application ([é] / [ê]) and confirm
([è]).
<WEih> Weighing
⮫ Chapter 9 ‘Weighing
application’ on page 61
<count> Counting
⮫ Chapter 10 ‘Counting
application’ on page 66
<chEck> Tolerance weighing
⮫ Chapter 11 ‘Tolerance
weighing application’
on page 76
? The scale returns to operating mode.
? The application is now active.
Once an application has been selected, the
device returns to operating mode. You must
then open the application menu again to view
the application-specific settings.
Weighing application
TDS-A_TDS-B-BA-e-2620 61
9 Weighing application
Instructions for zero setting and taring are provided in
⮫ Chapter 8 ‘Basic operation’ on page 55. Further specific
setting options can be found in the following chapters.
9.1 Overview of the “Weighing” application menu
Level 1 Description Further information
<ptare> Save known tare weight (pretare)
⮫ Chapter 8.4.2 ‘Saving a
known tare weight (pre-tare)’
on page 58
<hold> Hold reading (data hold) ⮫ Chapter 9.4 ‘Hold function’
on page 62
<unit> Set weighing units ⮫ Chapter 6.12 ‘Setting the
availability of display units’
on page 41
<mode> Select application ⮫ Chapter 8.5 ‘Selecting an
application’ on page 59
9.2 Simple weighing
To use this function, the “Weighing” application must be
active.
1. Check whether the display reads zero and, if necessary,
reset it to zero using [ZERO].
2. Place the load on the weighing plate.
? The weight value and the “Stability” indicator ( )
are displayed.
? The result can be read.
9.3 Underfloor weighing
Underfloor weighing allows objects that cannot be placed on
the weighing plate due to their size or shape to be weighed.
Weighing application
62 TDS-A_TDS-B-BA-e-2620
CAUTION
Risk of breakage due to overloading of the
hook
Falling loads may cause injury.
− Never exceed the scale’s specified maximum
load (Max.).
− Make sure there are no living creatures or
objects under the load that could be damaged.
NOTICE
Damage to the scale caused by dirt getting
inside
Dirt (e.g. dust) entering the measuring system
can cause damage and lead to incorrect results.
− Close the opening for the weigh-below hook
once underfloor weighing is complete (to prevent
dust ingress).
1. Switch off the scale.
2. Turn the scale over.
3. Open the cover on the bottom side of the scale.
⮫ Chapter 3.2 ‘Overview of the device’ on page 17
4. Place the scale over an opening.
5. Screw the hook in tightly.
6. If necessary, attach the weighing container and perform
taring.
7. Attach the load and weigh it.
9.4 Hold function
The hold function allows you to ‘freeze’ the displayed weight.
Once the load is removed from the scale, the weight value
remains visible on the display for 10 seconds.
To use this function, the “Weighing” application must be
active.
Weighing application
TDS-A_TDS-B-BA-e-2620 63
1. Press and hold [TARE].
? The application menu opens.
2. Select <hoLd> by pressing [é] / [ê], but do not
confirm yet.
3. Place the load on the scale.
4. Confirm by pressing [è].
? The first stable weight reading is ‘frozen’ and indicated
by the “Hold” symbol ( ).
For quicker access, the hold function can be
assigned to a key.
9.5 Percentage weighing
The “Percentage weighing” function allows you to calculate
the percentage of the sample weight in relation to a reference
weight.
To use this function, the “Weighing” application must be
active.
1. Check whether the display reads zero and, if necessary,
reset it to zero using [ZERO].
2. Press and hold [TARE].
? The application menu opens.
Weighing application
64 TDS-A_TDS-B-BA-e-2620
3. Select <unit> by pressing [é] / [ê] and confirm
([è]).
? The display switches to the unit selection menu.
4. Place the reference weight, which corresponds to 100
%, on the scale.
5. Select “%” by pressing [é] / [ê] and confirm ([è]).
? The weight value of the reference weight will be displayed
flashing.
6. Press [è].
? The scale switches to operating mode and displays
100 %.
? When weighing additional loads, their weight is displayed
as a percentage of the reference weight.
9.6 Weighing with a multiplication factor
This function allows the weight value to be multiplied by any
factor. This means, for example, that a known error factor
can be taken into account immediately when determining the
weight.
To use this function, the “Weighing” application must be
active.
1. Press and hold [TARE].
? The application menu opens.
2. Select <unit> by pressing [é] / [ê] and confirm
([è]).
? The display switches to the unit selection menu.
3. Select “FFA” by pressing [é] / [ê] and confirm ([è]).
? The display switches to numeric input.
Weighing application
TDS-A_TDS-B-BA-e-2620 65
4. Enter the factor (select a number: [é] / [ê], next
number: [è]) and confirm ([è]).
? The scale switches to operating mode and displays
the weight multiplied by the factor.
Counting application
66 TDS-A_TDS-B-BA-e-2620
10 Counting application
The Counting application allows you to count several parts
placed on the weighing plate.
10.1 Overview of the “Counting” application menu
Level 1 Level 2 Level 3 Description Further
information
<rEF> Set the reference
quantity
⮫ Chapter
10.2 ‘Setting
the reference
quantity’
on page 68
<ptare> Save known
tare weight
(pre-tare)
⮫ Chapter
8.4.2 ‘Saving
a known tare
weight (pretare)’
on page 58
<chEck> <tYPE> <AbS> Check quantity
(limit
values)
⮫ Chapter
10.5 ‘Quantity
checking
(limit values)’
on page 73
<rEL> Weigh the
target
number of
items (deviations)
⮫ Chapter
10.4
‘Weighing the
target number
of items’
on page 71
——————————————————————-
With setting <tYPE> = <AbS>
<LiMuPP> Upper limit
<LiMLoW> Lower limit
With setting <tYPE> = <rEL>
<VALuE> Target
number of
items
Counting application
TDS-A_TDS-B-BA-e-2620 67
Level 1 Level 2 Level 3 Description Further
information
<ErruPP> Upper tolerance
<ErrLoW> Lower tolerance
——————————————————————-
<noSiG> Define a
range to be
excluded
from the tolerance
assessment
<bEEPEr> <ch-ok>
(value within
the tolerances
/
limits)
Sound when
tolerance is
exceeded
■ <oFF>*
: no sound
■ <SLoW>
: slow
beeps
■ <Std>:
normal
beeps
■
<FASt>
: rapid
beeps
■
<cont>
: continuous
beep
<ch-Lo>
(below the
lower tolerance
/ below
the lower
limit)
<ch-hi>
(upper tolerance
/ upper
limit
exceeded)
<rESEt> Reset values
<unit> Set weighing
units
⮫ Chapter
6.12 ‘Setting
the availability
of display
units’
on page 41
Counting application
68 TDS-A_TDS-B-BA-e-2620
Option 1: Select a predefined
reference quantity
Level 1 Level 2 Level 3 Description Further
information
<mode> Select application
⮫ Chapter
8.5 ‘Selecting
an application’
on page 59
10.2 Setting the reference quantity
Before the scale can determine the number of items, it needs
to calculate the average weight per item.
− The higher the reference quantity, the
greater the counting accuracy.
− For small items or items that vary significantly
one from another, the reference
value must be set particularly high.
− The minimum part weight for piece counting
is specified in the technical data.
1. If necessary, place the weighing container on the scale
and perform taring.
2. Place the required number of reference samples on the
table.
3. Press and hold [TARE].
? The application menu opens.
Tolerance weighing application
The tolerance weighing application allows you to quickly
check whether a weight value falls within certain limits.
11.1 Overview of the “Tolerance weighing” application menu
Level 1 Level 2 Level 3 Description Further information
<chEck> <tYPE> <AbS> Check weight
(upper and lower
limits)
⮫ Chapter 11.3
‘Weight checking
(limit values)’
on page 79
<rEL> Weigh target
weight (deviations)
⮫ Chapter 11.2
‘Weighing the
target weight’
on page 77
——————————————————————-
With setting <tYPE> = <AbS>
<LiMuPP> Upper limit
<LiMLoW> Lower limit
With setting <tYPE> = <rEL>
<VALuE> Target weight
<ErruPP> Upper tolerance
<ErrLoW> Lower tolerance
——————————————————————-
<noSiG> Define a range to
be excluded from
the tolerance
assessment
Tolerance weighing application
TDS-A_TDS-B-BA-e-2620 77
Level 1 Level 2 Level 3 Description Further information
<bEEPEr> <ch-ok>
(value within the
tolerances /
limits)
Sound when tolerance
is
exceeded
■ <oFF>*
: no sound
■ <SLoW>
: slow beeps
■ <Std>:
normal beeps
■ <FASt>:
rapid beeps
■ <cont>:
continuous
beep
<ch-Lo>
(below the lower
tolerance / below
the lower limit)
<ch-hi>
(upper tolerance
/ upper
limit exceeded)
<rESEt> Reset values
<ptare> Save known tare
weight (pre-tare)
⮫ Chapter 8.4.2
‘Saving a known
tare weight (pretare)’
on page 58
<unit> Set weighing
units
⮫ Chapter 6.12
‘Setting the availability
of display
units’ on page 41
<mode> Select application ⮫ Chapter 8.5
‘Selecting an
application’
on page 59
11.2 Weighing the target weight
1. Press and hold [TARE].
? The application menu opens.
2. Select <chEck> g <tYPE> g <rEL> by
pressing [é] / [ê] and confirm ([è]).
? Application mode is active.
Tolerance weighing application
78 TDS-A_TDS-B-BA-e-2620
Enter target weight
Enter upper tolerance
Enter lower tolerance
Weigh target weight
3. Select <chEck> g <VALuE> by pressing [é] /
[ê] and confirm ([è]).
? The display switches to numeric input.
4. Enter your target weight (select a number: [é] / [ê],
next number: [è]) and confirm ([è]).
? The display switches to <VALuE>.
5. Select <ErruPP> by pressing [é] / [ê] and confirm
([è]).
? The display switches to the unit selection menu.
6. Select the desired unit ([é] / [ê]) and confirm ([è]).
? The display switches to numeric input.
7. Enter the upper tolerance (select a number: [é] / [ê],
next number: [è]) and confirm ([è]).
? The display switches to <ErruPP>.
8. Select <ErrLoW> by pressing [é] / [ê] and confirm
([è]).
? The display switches to the unit selection menu.
9. Select the desired unit ([é] / [ê]) and confirm ([è]).
? The display switches to numeric input.
10. Enter the lower tolerance (select a number: [é] / [ê],
next number: [è]) and confirm ([è]).
? The display switches to <ErrLoW>.
11. Press [ç] repeatedly to exit the menu.
1. If necessary, place the weighing container on the scale
and perform taring.
Tolerance weighing application
TDS-A_TDS-B-BA-e-2620 79
Enter the upper limit
2. Weigh the material to be weighed.
? Weight, “Stability” indicator ( ) and tolerance
marks are displayed.
? The result can be read.
Tolerance mark Description
Upper tolerance limit
exceeded
Weight within tolerance
limits
Below the lower tolerance
limit
− In addition to the tolerance marks, an
audible signal can also be set to check the
limits. Further information on this can be
found in the application menu.
− To delete the values, proceed as follows:
Open the application menu and go to
<chEck> g <rESEt>.
11.3 Weight checking (limit values)
1. Press and hold [TARE].
? The application menu opens.
2. Select <chEck> g <tYPE> g <AbS> by
pressing [é] / [ê] and confirm ([è]).
? Application mode is active.
3. Select <chEck> g <LiMuPP> by pressing [é] /
[ê] and confirm ([è]).
? The display switches to numeric input.
4. Enter the upper limit (select a number: [é] / [ê], next
number: [è]) and confirm ([è]).
The display switches to <LiMuPP>.
Tolerance weighing application
80 TDS-A_TDS-B-BA-e-2620
Enter the lower limit
Check weight
5. Select <LiMLoW> by pressing [é] / [ê] and confirm
([è]).
? The display switches to numeric input.
6. Enter the lower limit (select a number: [é] / [ê], next
number: [è]) and confirm ([è]).
? The display switches to <LiMLoW>.
7. Press [ç] repeatedly to exit the menu.
1. If necessary, place the weighing container on the scale
and perform taring.
2. Weigh the material.
? Weight, “Stability” indicator ( ) and tolerance
marks are displayed.
? The result can be read.
Tolerance mark Description
Upper limit exceeded
Weight within the limits
Below the lower limit
− In addition to the tolerance marks, an
audible signal can also be set to check the
limits. Further information on this can be
found in the application menu.
− To delete the values, proceed as follows:
Open the application menu and go to
<chEck> g <rESEt>.
Other features
12.1 Applying the totalizing function
The totalizing function allows you to add together the weight
values from individual weighings.
To use the totalizing function, your scale must
be connected to a printer or a computer.
1. Open the setup menu.
2. Select <Print> g <SuM> g <on> by pressing
[é] / [ê] and confirm ([è]).
3. Press [ç] repeatedly to exit the menu.
1. If necessary, place the weighing container on the scale
and perform taring.
2. Place the first load to be weighed on the weighing plate.
? The “Stability” indicator ( ) appears.
3. Press [PRINT].
? The value is stored in the total memory and the display
switches to <SuM1>, followed by the current
weight value.
? The “Total” indicator ( ) appears.
? The result is sent to a printer or PC.
4. Remove the weighed material.
5. Add any further loads to be weighed, as in steps 2 to 4.
? The number of weighings is displayed via <SuM1>,
<SuM2>, <SuM3> and so on.
. Press and hold [PRINT].
? The number of weighings and the total weight are
displayed.
? The total memory is cleared and the “Total” indicator
( ) disappears.
? The result is sent to a printer or PC.
Care, maintenance and servicing
13.1 Cleaning
13.1.1 Safety during cleaning
DANGER
Electric shock due to live parts
Contact with live parts can lead to electric shock
and burns.
− Installation, cleaning, maintenance and servicing
work may only be carried out by qualified
personnel.
− Disconnect live parts from the power supply
or unplug the power plug from the mains.
− Only carry out installation, cleaning, servicing
and maintenance work on devices that
are switched off and disconnected from the
power supply, unless otherwise specified in
the relevant instructions.
− Follow the instructions in this document.
NOTICE
Material damage due to incorrect cleaning
Unsuitable cleaning agents can cause damage to
the device.
− Do not use any aggressive cleaning agents
(e.g. solvents).
− Do not use cleaning agents containing caustic
soda, acetic, hydrochloric, sulphuric or citric
acid on stainless steel parts.
− Do not use metal brushes or cleaning sponges
made of steel wool.
− Do not open the device. There are no parts
inside that may be cleaned.
13.1.2 Recommended cleaning
1. Switch off the device.
Disconnect the device from the power supply.
3. Immediately remove any spilled material (e.g. loose
sample residues or powder) with a brush or a handheld
vacuum cleaner.
4. Clean the device with a gentle cleaning agent (e.g. soap
suds) and a soft cloth.
5. Use a dry, lint-free cloth to wipe the device.
13.2 Servicing
13.2.1 Safety during servicing
DANGER
Electric shock due to live parts
Contact with live parts can lead to electric shock
and burns.
− Installation, cleaning, maintenance and servicing
work may only be carried out by qualified
personnel.
− Disconnect live parts from the power supply
or unplug the power plug from the mains.
− Only carry out installation, cleaning, servicing
and maintenance work on devices that
are switched off and disconnected from the
power supply, unless otherwise specified in
the relevant instructions.
− Follow the instructions in this document.
13.2.2 Maintenance schedule
Interval Servicing activity Description
Regularly Levelling ⮫ Chapter 5.2.5 ‘Levelling
the device’ on page 27
■ Level the scale after moving
it to a new location.
■ Check the levelling regularly.
If necessary Charge battery Charge the rechargeable battery
when it is flat.
Care, maintenance and servicing
84 TDS-A_TDS-B-BA-e-2620
Interval Servicing activity Description
Regularly / after relocating the
device
Adjusting ⮫ Chapter
6.3 ‘Adjustment settings’
on page 35
Adjust the scale at regular
intervals and after each change
of location. KERN & SOHN
offers test weights for this purpose.
13.3 Control of inspection, measuring and test equipment
As part of quality assurance, the metrological characteristics
of the scale and any test weights must be checked at regular
intervals. The responsible user must define a suitable
interval for this, as well as the nature and scope of the test.
Information regarding the monitoring of scales and the test
weights required for this purpose is available on the KERN
website (⮫ www.kern-sohn.com). KERN’s accredited Calibration
Laboratory offers a fast and cost-effective service for
the calibration of test weights and scales (traceability to the
national measurement standard).
Decommissioning and storage
TDS-A_TDS-B-BA-e-2620 85
14 Decommissioning and storage
14.1 Taking the device out of operation
1. When operating with a rechargeable battery: To prevent
deep discharge, charge the battery if the device is to be
stored for a longer period of time.
2. Switch off the device.
3. Disconnect the device from the power supply:
■ Disconnect the power supply from the mains and
unplug it from the device.
■ Remove the rechargeable battery from the device.
4. Remove all loose / moving parts from the device.
5. Remove all interface cables.
6. Pack the device and all accompanying parts neatly in
the original packaging.
? The device can now be stored ⮫ Chapter 14.2
‘Storing the device’ on page 85.
14.2 Storing the device
NOTICE
Damage caused by incorrect storage
Improper storage can damage the device.
− Take the device out of operation according to
the instructions in this document ⮫ Chapter
14.1 ‘Taking the device out of operation’
on page 85.
− Store the device in the original packaging.
− Do not store the device outdoors.
− Observe the environmental conditions in the
technical data.
Warranty and goods return
86 TDS-A_TDS-B-BA-e-2620
Procedure for returns
15 Warranty and goods return
15.1 Warranty
The warranty claim only applies to defects that were already
present at the time of purchase. Damages or defects caused
by improper use, incorrect handling or normal wear and tear
are excluded from the warranty.
15.2 Goods return
− A return is not possible if you have used the
device improperly and it has been damaged
as a result.
− The original packaging, including all associated
parts, must be used for returns.
− In cases of returns, the complete scope of
delivery must be sent back.
1. Check that all parts of the delivery are present.
2. Remove all loose / moving parts and connected cables
from the device.
3. Reattach the transport safety lock.
4. Reapply any protective film that may be present.
5. Pack all parts of the delivery in the original packaging
and secure them so that they cannot slip and are protected
from damage.
Disposal
TDS-A_TDS-B-BA-e-2620 87
Disposal of electronic devices
Disposal of batteries and
rechargeable batteries
16 Disposal
ENVIRONMENT
Damage to people and the environment due
to improper disposal
Non-degradable materials (e.g. plastics or metals)
can be released into the environment and cause
damage if disposed of improperly.
Materials that are not recycled are a waste of
resources and lead to increased environmental
pollution.
Electrical devices may contain substances that
are not biodegradable and are harmful to the
environment and human health.
Batteries and rechargeable batteries may contain
substances that are harmful to the environment
and human health.
− Dispose of the device via appropriate recycling
measures or recycling centres in accordance
with the applicable national or regional
law of the place of use.
The crossed-out wheelie bin symbol indicates that waste
from electronic equipment must not be disposed of with
household waste. Disposal must be arranged by the operator
in accordance with the applicable national or regional law of
the place of use.
If this device contains confidential data, it is the responsibility
of the operator to delete this data before disposal.
Remove batteries and rechargeable batteries before disposal
of the device and dispose of them separately ⮫ ‘Disposal of
batteries and rechargeable batteries’ on page 87.
The crossed-out wheelie bin symbol indicates that batteries
and rechargeable batteries must not be disposed of with
household waste. Disposal of batteries and rechargeable batteries
must be arranged by the operator in accordance with
the applicable national or regional law of the place of use.
We are obliged to inform end users in accordance with the
EU Battery regulation and the applicable national regulations:
Disposal
88 TDS-A_TDS-B-BA-e-2620
■ End users are legally obliged to return all used (rechargeable)
batteries.
■ Batteries and rechargeable batteries can be returned free
of charge after use to municipal collection centres or
retailers. Non-used batteries/rechargeable batteries must
be protected against short circuits.
■ Returns to us are limited to usual household quantities
of batteries and rechargeable batteries that we supply or
have supplied.
■ If batteries or rechargeable batteries contain cadmium
(Cd), mercury (Hg) or lead (Pb), the respective symbol is
shown below the waste bin symbol.
Documents / Resources
![]() | TDS Platform Scale |
References
- sohn.comsohn.com
- kern-sohn.comwww.kern-sohn.com
- kern-sohn.com/shop/en/downloadswww.kern-sohn.com
- kern-sohn.com.www.kern-sohn.com.
- User Manualmanual.tools