ORL 94BS Digital Refractometer

Product Information

Specifications:

  • Material: Stainless steel
  • Measurement Range: 0-100
  • Accuracy: ±0.1
  • Power Source: Battery
  • Connectivity: USB

Product Usage Instructions

Preparations before Operation

  1. Install the Battery:
    • Use the screwdriver to open the battery hatch.
  2. USB Connection:
    • Connect to USB for charging/connecting to the computer.

Booting and Measurement

  1. Booting:
    • Press the On/Off button to turn on the device.
    • If no sample is present, it will display a dash.
    • Avoid strong light when using outdoors to maintain measurement
      accuracy.
    • Clean the sample plate and prism before dripping the sample
      liquid.
    • Ensure stability during measurement.
    • Ensure uniform temperature levels for instrument, environment,
      and sample.
  2. Measurement:
    • Clear distilled water and dry the sample plate.
    • Drip 0.3~0.4ml of the sample, then close the cover for
      measurement.
    • To get an average measurement, press the Read button for 2
      seconds.

The Calibration

The meter supports pure water calibration. Follow these
steps:

  1. Drip 0.4ml pure water, then close the cover for
    measurement.
  2. To enter calibration status, press the CAL button for 2-3
    seconds until ‘CAL’ flashes.
  3. Press CAL button again during flashing to start
    calibration.

Frequently Asked Questions (FAQ)

Q: What should I do if the instrument displays ‘CAL’
flashing?

A: Press the CAL button again during flashing to start the
calibration process.

Q: Can the digital refractometer measure highly corrosive
liquids?

A: No, it cannot measure liquids highly corrosive to metal or
glass.

“`

Catalogue

1.Introduction

1

2.Display and Buttons

2

3.Preparations before Operation

2

4.Booting and Measurement

3

5.The Calibration

6

6.Scales Converting and Temperature Systems 7 Converting

7.Turn Off

8

8.Maintenance and Preservation

8

9.Appendix

9

Before operating your instrument, please read this manual properly.

1.Introduction Panel Descriptions

2.Display and Buttons Display Areas and Buttons

LCD Display Screen

Cover Stainless Steel Sample Groove
Prism

USB Connect

Multi-function Display Area Host Display Area
Power on / off

Battery Capaclty

USB Port Charge Lamp/Charging lndicaor

Button

Calibrating “Zero Point”
Converting Different Scales& Temperature Unit
Measurement

Temperature Compensation
Temperature Display Area

Refractive Index

Battery Hatch

If is displayed, please charge it.

This digital refractometer cannot measure any liquid that is highly corrosive to metal or glass. When measuring liquids that are corrosive to plastics or react chemically with plastics, be careful not to drop the measured liquid onto the shell. Otherwise it will corrode the shell.

The Packing Accessories Packaging x1 The instructions x1 Dropper x1 Screwdriver x1 USB Line x1 Charger x1
1

3.Preparations before Operating 3.1 Install the Battery Use the screwdriver to open the battery hatch.
2

The multi-function display area would show the current scale number.

3.2 USB Connection Connect to USB for charging/connecting to the computer.
USB lnterface 4.Booting and Measurement 4.1 Booting
3

Press “On/Off” button to turn on
If there is no sample,it will display “dash” Note : 1. When used outdoors,please avoid strong light
so as not to affect the measurement accuracy. 2. Before dripping into the sample liquid, please
clean the sample plate and Prism with soft clean cloth or soft paper. 3. Please keep the instrument in a stable status to measure. 4. Please ensure instrument, environment and sample are in the same temperature level before measuring. 4.2 Measurement After tum on,clear the distilled water and dry the sample plate, drip 0.3~0.4ml sample then close the cover to measure.
4

Cyclic Switch Measuring Result Press CAL once

Press the “Read” one time the meter will measure one time.

Showing Beyond Measuring Range

If press the “Read” button for 2 seconds, the instrument would make the automatic measurements upon programmed times (default 15times), the final value is the average of 15 times’ measurements, After measurements, it will return to the normal state and the scales display area would display the average of the automated measurements
The multifunctional area displays the number of remaining automatic measurements.For each automatic measurement, this number is reduced by 1.
5

5.The Calibration The meter only supports pure water calibration. The calibration method is as following:drop 0.4ml pure water then close the cover to measure.
LED Flashing Display
Press “CAL” button for 2-3 seconds till see the `CAL’ flashing .to enter the calibration status.
Press “CAL” button once again during the `CAL’ flashing to start to calibrate. When the calibration is completed showing as following. If no any operations for 10seconds the instrument would return back to booting status.
If fail to complete the calibration, multi-function display area would show an error code.
Display error code A03,indicating error of sensor measurement.See the appendix for other error codes .
6

We recommend calibrating the refractometer, when commissioning after a strong shock after longer transport after a change of location with a large temperature
difference if the device has not been used for a long time Always use distilled water and make sure that the refractometer, the water and the environment are at the same temperature.
6.Scales converting and temperature systems converting
6.1Scales Converting
Current Scale Number
Switching
Press`Scale’once to convert the scales and the values.
6.2Temperature System Converting
7

Switching

Temperature Unit

Press button for 2 seconds, temperature unit would be converted.
If exceed the temperature limitations, the signs “HHH” or “LLL” would show.

7.Turn Off
1. After turn on if without any operations for 3 minutes, the instrument would be automatically turned off.
2. Press “on/ off ” button for 2-3 seconds, the instrument would be turned off.
8.Maintenance and Preservation
1. Please clean and wash the sample plate with distilled water and dry it with soft cleaning cloth or paper towel after finishing the measuring one kind sample.
2. Never left the remains and residuals of samples in the sample plate for long time.
8

3. After finishing measurements of the corrosive liquid, please clean the sample plate as quick as possible to avoid the irreparable damage of the prism and metal surface of the plate.
4. Please use soft cleaning cloth or paper towel to clean the sample plate to avoid scribing the prism’s glass.
5. When the dropper and dust-free cloth are not used, please clean it with clean water and put it in the packing box after drying.
6. If no using the instrument for a long time, please remove the battery, and preserved in a cool and dry environment.

9.Appendix

Performance:

Range

Accuracy

Resolution

Brix

0.0%~94.0% ±0.1%

0.1%

R.I.

1.3330~1.5290 ±0.0002 0.0001

0.0~40.0 Temperature
32.0~104.0

±0.3 ±0.6

0.1 0.1

Dimension

180*100*55mm

Net weight

365g (excluding battery)

The Error Codes Table:

Code

Instructions

A01

Beyond the scope of calibration temperature.1.0~40.0)

During calibration, no solution

A02

or non-pure water

A03

This instrument has a hardware failure.

9

Description of Scales Numbering:
2024/09 V3.2 10

Documents / Resources

KERN ORL 94BS Digital Refractometer [pdf] Instruction Manual
ORL 94BS, MSDR-D-KERN, ORL 94BS Digital Refractometer, ORL 94BS, Digital Refractometer, Refractometer