KERN OIV 345 Video Microscope

Specifications

  • Product: Video Microscope KERN OIV 345
  • Manufacturer: KERN & Sohn GmbH
  • Model: OIV-3-BA-d-2411
  • Version: 1.1 (July 2024)

Device Overview

The Video Microscope KERN OIV 345 is designed for magnifying and viewing objects with precision. It comes with a software interface for enhanced functionality.

Before Use

General Information: Familiarize yourself with the operating instructions before using the microscope.

Basic Information

  • Observe the notes in the operating instructions: Follow the guidelines provided in the manualfor safe usage.
  • Staff Training: Ensure that operators are  trained on the correct usage of the microscope.
  •  Safety: Always prioritize safety while handling the device.

Unpacking and Commissioning

Unpacking: Carefully unpack the microscope and its components.

Assembly

Operation and Function of the Microscope:

Understand how to set up and operate the microscope effectively.

Operation and Functionality of the Software
Toolbar: Explore the software toolbar for various functions and features.

Optional Accessories
Optional Accessories: Discover additional accessories that can enhance your microscopy experience.

FAQs

Q: How do I troubleshoot common issues with the video microscope?
A: Refer to section 12 (Troubleshooting) of the manual for guidance on resolving common problems.

Q: What is the recommended power supply for the video microscope?
A: Section 14 (Power Supply) provides information on the mains connection required for powering the device.

“`

CORE Item number/type Optical system Lighting Dimmable lighting Stand
Optical zoom Digital zoom Magnification ratio Field of vision Tilting the screen Screen size Screen sharing Camera resolution Sensor Real-time refresh rate Recording Output interface Input interface Storage of data Camera parameters

Measurements


Other functions Languages
Net weight Product dimensions WxD Gross weight Packaging dimensions WDH
4

OIV 345 TOIV 345-B Greenough System LED ring light 3.3 W Yes Mechanical, upright height 325 mm, base plate 260×320 mm 0.7x – 4.5x 17x – 110x 6,4 : 1 Ø 17.6 – 1.47 mm 48° 11.6 inch 1920×1080 HD 5 MP SONY CMOS 1/2.8″ 60fps / 1920×1080 60fps / 1920×1080 HDMI USB 2.0 2x Snapshot, Recording, USB Exposure, white balance, gain, gamma, brightness, contrast, hue, saturation, grayscale Point, line, parallel lines, dashed lines, angle, diameter circle, radius circle, three-point circle, distance from circle to line, concentric circles, double circle, vertical line, any line, polygon, arc, marking, screenshot, table data export, calibration, parameter settings
Mirror horizontally, mirror vertically, freeze German, English, French, Italian, Russian, Japanese, Chinese 4 kg 260×320 mm 6 kg 425x395x240 mm
OIV-3-BA-d-2411

2 Declaration of conformity
The current EC/EU Declaration of Conformity can be found online at:
https://www.kern-sohn.com/shop/de/DOWNLOADS/

5

OIV-3-BA-d-2411

3 Device overview
Screen
Microscope head with camera
LED ring light
Microscope stand
MicroSD card slot (back side)
HDMI interface Power connection On/off switch ring light
6

Zoom adjustment wheel Focus wheel
Dimmer LED ring light
Operating indicator (LED) USB port (2x)
On/off switch Microscope
OIV-3-BA-d-2411

4 Before use
4.1 General information
The packaging must be opened carefully to prevent the accessories inside from falling to the floor and breaking.
In general, a microscope should always be handled with great care, as it is a sensitive precision instrument. Avoiding abrupt movements during operation or transportation is therefore particularly important, especially to avoid endangering the optical components. You should also avoid dirt or fingerprints on the lens surfaces, as in most cases this impairs the sharpness of the image.
If the performance of the microscope is to be maintained, it must never be disassembled. Parts such as objective lenses and other optical components should therefore be left as they are at the start of operation.

5 Basic information (general)

5.1 General information on warnings

Warnings are used in these operating instructions to warn you of possible personal injury or damage to property in certain situations.

Signal word DANGER WARNING CAUTION NOTE

Description
Failure to observe the instructions will lead directly to serious injury, permanent impairment (e.g. loss of a limb) or death of the user or third parties
Failure to observe the instructions may result in serious injury, permanent impairment (e.g. loss of a limb) or death of the user or third parties
Failure to observe the instructions may result in minor injuries or temporary damage to the user or third parties (e.g. minor cuts)
Failure to observe the instructions may result in damage to property

Symbols in warning notices:
7

OIV-3-BA-d-2411

Icon Warnings

Meaning Warning signs warn you of dangers that could lead to injury. The symbol indicates the type of danger. Indicates general hazards or a danger point
Warning of electrical voltage

Icon Commandment sign

Meaning
Mandatory signs prescribe measures that you must take to avoid personal injury or damage to property. The symbol indicates the necessary actions or objects to prevent damage.

Indicates a prescribed action

5.2 Intended use The OIV 345 is very well suited for inspection and quality control.

5.3 Improper use The OIV 345 must not be used for medical purposes. Do not use the device in potentially explosive atmospheres or for measurements in liquids or on live parts. Unauthorized structural changes, additions and conversions to the appliance are prohibited.
5.4 Warranty The guarantee expires in the event of · Non-compliance with our specifications in the operating instructions · Use outside the described applications · Changing or opening the device · Mechanical damage and damage caused by media, liquids, natural wear and tear · Improper set-up or electrical installation · Improper assembly or electrical installation
OIV-3-BA-d-2411

Before connecting the device to the mains, you must ensure that you are using the correct input voltage. The information for selecting the correct mains cable can be found on the device, on the back of the product directly above the connection socket. It is essential that you follow this information. If you do not adhere to these instructions, this may result in fire or other damage to the appliance.
The main switch must also be switched off before the mains cable is connected. In this way, you avoid triggering an electric shock.
If you are using an extension cable, the mains cable you are using must be earthed. Only carry out work that comes into contact with the electrical system of the appliance when it is de-energized.

Storage and transportation
Avoid exposing the device to direct sunlight, high or low temperatures, shocks, dust and high humidity.
The suitable temperature range is 0 – 40 °C and a relative humidity of 85% should not be exceeded.
The appliance should always be placed on a firm, smooth and horizontal surface.


When the microscope is not in use, it is best to cover it with the dust cover supplied. Dust or dirt inside the optics of a microscope can in many cases lead to irreversible malfunctions or damage.
Accessories consisting of optical elements, such as additional lenses, are preferably stored in a drying box with desiccant.
Note If you store or transport the device improperly, the device may be damaged. Observe the instructions for transporting and storing the appliance.
Packaging / return transportation Returns are only possible within the framework of the General Terms and Conditions Keep all parts of the original packaging for any necessary return transportation.
· Only the original packaging may be used for return transportation. · Disconnect all connected cables and loose/movable parts before shipping. · Reattach any existing transport locks. · Secure all parts against slipping and damage.

9

OIV-3-BA-d-2411

6 Basic warnings and safety instructions
6.1 Observe the notes in the operating instructions
Read the operating instructions carefully before commissioning/using the device, even if you already have experience with KERN devices. Always
keep the instructions in the immediate vicinity of the appliance.
6.2 Staff training
The appliance may only be used by persons who have read and understood the operating instructions, in particular the chapter on safety.
6.3 Safety
WARNING
Read all safety information and instructions. Failure to observe the safety information and instructions may result in electric shock, fire and/or serious injury. Keep all safety information and instructions for future reference. The design of the device must not be modified. This can lead to incorrect measurement results, safety defects and destruction of the device Do not operate the appliance in potentially explosive rooms or areas and do not install it there. Do not operate the device in an aggressive atmosphere. Do not immerse the appliance in water. Ensure that no liquids penetrate the inside of the device. The device may only be used in a dry environment and under no circumstances in rain or relative humidity above the operating conditions. Protect the device from permanent direct sunlight. Do not expose the appliance to strong vibrations. Do not remove any safety signs, stickers or labels from the device. Keep all safety signs, stickers and labels in a legible condition Do not open the device The lamp generates a lot of heat during operation. Avoid touching the lamp housing during operation and for some time afterwards. Do not operate the device in an aggressive atmosphere.

10

OIV-3-BA-d-2411

WARNING
Risk of injury due to electric shock! Risk of short circuit due to penetration of liquids into the housing! Do not immerse the appliance or accessories in water. Make sure that no water or other liquids get into the housing. Work on electrical components may only be carried out by an authorized specialist company! Take care not to twist or kink the mains cable. Only use the original adapter supplied.
WARNING
There is a risk of suffocation! Do not leave the packaging material lying around carelessly. It could become a dangerous toy for children. The appliance is not a toy and does not belong in the hands of children. This appliance can be dangerous if it is used improperly or not as intended by untrained persons! Observe the personell qualification!
WARNING
Electrostatic sensitive device! The device can be destroyed by electrostatic discharges. Connectors for HF signals are particularly at risk. Please observe the handling instructions for electrostatically sensitive components.
CAUTION
Keep a sufficient distance from heat sources. Do not use the device in environments with high humidity or water mist
NOTE
To avoid damaging the device, do not expose it to extreme temperatures, extreme humidity or moisture. Do not use harsh cleaning agents, abrasives or solvents to clean the appliance.

11

OIV-3-BA-d-2411

7 Unpacking and commissioning
7.1 Unpacking
In the event of a return shipment, please observe the instructions in the chapter “Packaging/return transportation
On receipt of the device, please check in advance whether there is any damage caused during transportation and whether the outer packaging, the housing, other parts or even the device itself are damaged. If any damage is apparent, please notify KERN & Sohn GmbH immediately.

12

OIV-3-BA-d-2411

8 Assembly

Microscope stand
Attach the microscope stand with the base plate and the supplied M5 threaded screws (3x). Tighten the screws on the underside of the plate using the M5 hexagon tool supplied.
Set up the microscope stand after successful assembly.

1

2

Microscope head Then insert the microscope head (1) into the guide on the holder (2) of the microscope stand.
To attach the microscope head to the stand, tighten the screws (2x) on the outside of the housing using the Allen key supplied.

Power connection
Plug the mains cable into the connection on the microscope. Note: Take care not to twist or kink the mains cable. Only use the original adapter supplied.

13

OIV-3-BA-d-2411

8.1 Operation and function of the microscope
When the microscope is ready for use after assembly, it must first be connected to the mains via the mains cable. Only insert the mains plug into a suitable socket. Ensure that the mains cable is laid correctly. The following sections describe all the important functions that are useful for operating the device.

2

Switching on the device

To switch the microscope on, press the On/Off

button (1) until it is backlit in blue. The LED

operating indicator (2) (PL) lights up green as soon

1

as the microscope is switched on. The system now

starts in operating mode.

Switching the lighting on and off

The switch for the ring lighting is located on the side

3

of the ring light. To switch the lighting on/off, press

the switch (3).

Setting the illuminance
Turn the dimmer for the lighting (4) on the right rear side of the ring light to adjust the lighting.
4

5

6

Set magnification and focus
As the OIV series are zoom microscopes, the zoom factor (0.7x-4.5x) can be changed using the zoom adjustment wheel (5). Turn the dial to the maximum magnification. Focus the image using the focusing wheel (6). Repeat the process if necessary. Note: The max/min parfocality is already optimally preset on delivery.

14

OIV-3-BA-d-2411

8

7

Connections and interfaces
The device has an internal and external software application. Internal software via the screen: To operate the software, a mouse must be connected via the USB port. To do this, plug the USB receiver, which is located in the battery compartment of the wireless mouse supplied, into one of the two USB slots (7). A USB stick (max. 128 GB) must be connected for data storage. External software via a PC / laptop: The video microscope can be connected to a PC/laptop via a USB cable. To do this, connect the USB cable to one of the two USB slots (7) and to a USB port on the PC/laptop. The external software is available for download at https://www.kern-sohn.com/shop/de/downloads. The picture can be transmitted to another output device via the additional HDMI output (8).

MicroSD card connection

There is a slot for a microSD memory card on the

back of the device. The connection offers an

9

additional option for saving data (max. 128 GB). Use

a pointed object to push the MicroSD memory card

into the card slot until it is fully engaged.

15

OIV-3-BA-d-2411

9 Operation and functionality of the software
9.1 Camera menu
Click on the camera to switch to the camera page, set the relevant
camera parameters and use the corresponding tools.
9.2 Exposure and white balance

Automatic exposure is activated. The target value can be set

by

dragging the slider to the left or right.

This deactivates manual adjustment of the exposure time or the amplifier (gain)

Automatic exposure is deactivated. The target value cannot be changed

Exposure time and amplifier (gain) can be

set

manually using the sliders.

Click this button to reset the exposure parameters and activate automatic exposure.

16

OIV-3-BA-d-2411

automatic white balance is activated. Manual adjustment of the red, green and blue levels is not possible The RGB values are regulated automatically by the camera.

Automatic white balance is deactivated.

The parameters for red, green and blue levels can be adjusted manually. To do this,

drag the slider to the left or

right accordingly.

This function is a one-time white balance. Click once and the camera automatically sets the values for red, green and blue levels.
6.1 Adjust parameters

Drag the slider to the left or right to increase or decrease the values. Click on this button to reset all parameters to their default values.

17

OIV-3-BA-d-2411

9.3 Control elements
Magnification: Magnification of the camera. The middle of the magnification is determined by the position of the previous magnification of the scroll wheel; if the scroll wheel has not been magnified, it is set to the middle magnification by default. Note: The magnification is reset to 1.0X after opening the gallery. The magnification can also be increased using the mouse wheel.
Reduce: Reduce the magnification of the camera. The center of the magnification is determined by the position of the last magnification, the next magnification is determined by the position of the last magnification, etc. Note: The magnification is reset to 1.0X after opening the gallery. The mouse wheel can also reduce the magnification.
Move: If the current magnification factor is greater than 1.0, you can press and drag the left mouse button after clicking to move the current camera image.
Mirror horizontally: The entire image is mirrored horizontally. The result of this process is that the point in the top right-hand corner swaps position with the point in the top left-hand corner and the point in the bottom right-hand corner swaps position with the point in the bottom left-hand corner.

18

OIV-3-BA-d-2411

Vertical mirroring: The entire image is mirrored vertically. The result of this process is that the dot in the top right-hand corner swaps position with the dot in the bottom right-hand corner and the dot in the top left-hand corner swaps position with the dot in the bottom left-hand corner.
To prevent damage to the focusing system, the left and right adjustment wheels of the coarse and fine drive must never be turned in opposite directions at the same time.
Freeze: Freezes the image displayed by the current camera. Note: Increase in power, decrease in magnification, horizontal tilting, vertical tilting, Monochrome and quad split screen cannot be performed in the current state. Freezing is automatically switched off when the gallery is opened.
Monochromatic: The color of the current frame is removed.
Comparison: After clicking, first open the gallery and then the photo or video to display the left and right frame. The left frame is the image of the current camera, and the right frame is the selected photo or video. The slider at the bottom moves to the left and right to show the incomplete display area of the camera image or the photo or video. Note: The video contrast is only effective when the video is being played.
Four-part screen: After clicking, the screen is divided into four parts. Left-click on one of the screens to freeze it and click again to unfreeze it. Click on the quartered screen again to exit the quad screen. Exit the quad screen automatically when you open the gallery or the contrast. Note: The measurement cannot be carried out if the quartering mode is open.

19

OIV-3-BA-d-2411

9.4 Measurements
Click on Measure to switch to the measurement page, take measurements and change measurement attributes. As shown in the illustration on the right. 6.2 Measuring tools

Point: Move the mouse pointer to the position to be measured, click the left mouse button and raise it to complete the drawing and display the corresponding coordinates. Note: The coordinate values change depending on the position of the scale and the size of the calibration.
Line: The first click selects the starting point, the second click completes the line drawing and displays the line and its length. Note: The length of the line depends on the size of the calibration.
Parallel line: The first click selects the starting point, the second click identifies a line, and the third click draws parallel and vertical lines between the two lines. Note: The distance between the parallel lines depends on the size of the calibration.

17

20

OIV-3-BA-d-2411

Polyline: Click the left mouse button to get the different activities, after completion, click the right mouse button to stop, and show the total length of all lines. Note: The length of the dashed line depends on the size of the calibration.

Arc: Select three points at different positions and draws an arc according to the distance between the three points and displays the radius and angle of the arc. Note: The radius of the arc depends on the size of the calibration.

Radius circle: select the center of the circle, click the left mouse button and lift off, move the mouse, reach the point on the arc, click the left mouse button again, the operation is completed, select a circle corresponding to the two points, display the radius. Note: The radius of the circle depends on the size of the calibration.

Diameter of the circle: Select a point on the arc, click the left mouse button, move to another point on the arc, click the left button again, the operation is completed, after the two points mark a circle, show the radius. Note: The radius of the circle depends on the size of the calibration.

Three-point circle: With the left mouse button click three points in different positions, after the three points mark a circle, and show the radius of the circle. Note: The radius of the circle depends on the size of the calibration.

Circle to line: the left mouse button click on two points in different positions, after the two points mark a circle, click the left button again, the point and the center of the circle mark a line after the point to do the vertical line of the line, show the distance from the point to the center of the circle. Note: The distance between the circle and the line depends on the size of the calibration.

Concentric circles: Select the center of the circle, left-click and raise it, move the mouse to the point on the arc, click again, select the first circle corresponding to the two points, move the mouse to the end, left-click to raise it, select the second

21

OIV-3-BA-d-2411

circle corresponding to the center of the circle and the end point, and display the radius of the double circle. Note: The radius of the circle depends on the size of the calibration.
Double circle: Click three points at different positions with the left mouse button, select a circle corresponding to the three points, select three points at different positions again, click the left mouse button, select a circle in the same way, display the distance between the two circle centers. Note: The length between the two centers depends on the size of the calibration.
Angle: The left mouse button click on three different positions of the point, after the three points to form two lines, show the angle of the two lines.
Vertical line: click the first time to select the starting point, click the second time to define a line, click the third time after the third point to form a vertical line, display a short line perpendicular to the line, display the length of the vertical line. Note: The length of the vertical line depends on the size of the calibration.
Rectangle: Select the upper left corner of the rectangle, click the left mouse button to raise it, move the mouse to the lower right corner of the rectangle, click the left mouse button again to raise it, the operation is completed, the area and perimeter of the rectangle are displayed. Note: The area and circumference of the rectangle change depending on the size of the calibration.
Polygon: Click the left mouse button to get each active point, after completion, click the right mouse button to stop, show the area and perimeter of the polygon. Note: The area and circumference of the polygon change depending on the size of the calibration.
Any line: Move the mouse pointer to the position to be measured. Press the left mouse button and move it to draw any line. Press the left mouse button to complete the drawing.

22

OIV-3-BA-d-2411

Text: Move the mouse to the position to be measured, click on two points with the left mouse button, draw an arrow corresponding to the two points and open the input field. Click OK after the input in the input field is completed, and the entered text will be displayed from the end point of the input text. If you click Cancel, no text is entered, only the arrow is displayed.
Move: After clicking this icon, the mouse can capture the drawn points, line segments and circles. If the point clicked on is a measured point that has been drawn, you can change the size and position of the point. If this is not the case, switch to full movement. Note: Polylines, polygons and arbitrary lines cannot move as a whole. If the whole thing moves, the text moves with it. If the position of the click is text, the text can be moved separately.
Delete: Click on this icon and left-click on the drawn measurement to delete the measurement.
Delete all: After clicking on this symbol, a reminder message appears. If yes, delete everything. If no, delete everything.
Screenshot: Capture the current image. Take screenshots without adding them to the toolbar with various prompts.
Table: To view the details of all currently drawn line segments, click Export to export the Excel file saved in the MEASUREMENT folder on the USB stick.
9.5 Calibration setting

Select the currently saved calibration options.

23

OIV-3-BA-d-2411

Add calibration: Click to enter the “Add Calibration” mode, enter the name and length, and then click the “Save” button to save the newly created calibration.
Delete calibration: Deletes the currently selected calibration. Note: The first calibration cannot be deleted.
Edit calibration: Edit the currently selected calibration.
Straight line calibration: Switches the current calibration to a straight line. Vertical calibration: Switches the current calibration to vertical.
Three-point circle calibration: Switches the current calibration to three-point circle.

Unit: Switches the current calibration unit.

Add a new calibration: Click on

to add a new calibration, the delete calibration

cannot be clicked at this point, the selection field for the calibration is empty. Select the

calibration style

and click on the name input field, the keyboard

appears. Enter the name or click

to close the keyboard and finish entering the

name. Click on the input field for the length to call up the numeric keypad. After entering

, click on

to close the numeric keypad and complete the name entry. Click on the

drop-down field for the unit to complete the unit change. Click on

to save the

calibration. Adding the calibration completed.

Edit the order of calibration: Click Select calibration,

then the calibration

selection bar, new calibration and delete calibration cannot be clicked. Select the

desired calibration style,

click on the name and the keyboard will

appear. After completing the entry, click ENTER or

to close the keyboard and

finish entering the name. Click on the input field for the length to call up the numeric

keypad. When you have finished entering, click on

to close the numeric keypad

and complete the name entry.

24

OIV-3-BA-d-2411

Click on the

drop-down field for the unit to change the unit and click on

to save the calibration. Edit calibration completed.

Delete calibration sequence: Select the calibration you want to delete in the selection

bar

and click on

to delete the calibration.

25

OIV-3-BA-d-2411

9.6 Measurement attributes
:
Line Color (Line Color): After clicking, a drop-down field appears. After selecting the color, you can change the color. Note: The measured color that was drawn is also changed.
Line width (Line Width): After clicking, a drop-down field appears. Select the width and change the line width. Note: The width of the measured line width that was drawn is also changed.
Font color (Font Color): After clicking, a drop-down field appears. Select the color and change the font color. Note: The font color of the drawn measurements is also changed.
Font size (Font Size): After clicking, a drop-down field appears in which you can select the size and change the font size. Note: The font size of the drawn dimensions also changes.
Accuracy (Precision): The drop-down box appears after clicking, select the precision option and change the precision. Note: The accuracy of the plotted measurements will also change.

26

OIV-3-BA-d-2411

9.7 Scale ruler
Display of the scale (Scale Disp): If you switch to On, the scale is displayed; if you switch to Off, the scale is closed. Position: Switches the position displayed on the scale. If it is switched to the center, the reference point of the scale is in the center; if it is switched to the top left, the reference point of the scale is at the top left; if it is switched to the bottom right, the reference point of the scale is at the bottom right. Color (Line Color): Click on the drop-down box that appears and select the color you want to change. Width (Line Width): Switches the width of the scale of the scale. 9.8 Further settings
Click Others to switch to other pages to view and change the properties of the system. 9.9 Languages

Select language
27

OIV-3-BA-d-2411

9.10 Storage space
If the USB flash drive has several drives, you can select the drive on which the files are to be saved.
9.11 Resetting the data to the factory settings
After the factory settings have been restored, the exposure parameters are reset to the default values and automatic exposure is activated. Reset the white balance parameters to the default values and activate manual white balance. Contrast, saturation, brightness, sharpness, noise reduction, hue and gamma are reset to the default values. Magnification zoom, screen movement, screen flip, freeze, monochrome, four-screen does not participate in factory restore. The color of the measurement attribute is reset to red, the line width is reset to 2, the font color is reset to red, the font size is reset to 10, the accuracy is reset to 0.01 and the detection is reset to 10px. The group color of the grid line V1-H1 is reset to red, the line is set to Off and the line width is reset to 1. The group color of grid line V2-H2 is reset to orange, the line is reset to Off and the line width is reset to 1. The group color of grid line V3-H3 is reset to yellow, the line is reset to Off and the line width is reset to 1. The group color of the grid lines V4-H4 is reset to green, the lines are reset to Off and the line width is reset to 1. The group color of grid line V5-H5 is reset to cyan, the line is set to Off and the line width is reset to 1. The group color of the grid lines V6-H6 is reset to blue, the line is reset to Off and the line width is reset to 1. The group color of the grid lines V7-H7 is reset to violet, the line is set to Off, the line width is reset to 1. The color of the VB-HB group line is reset to white, the line to Off and the line width to 1.

28

OIV-3-BA-d-2411

The Group name option is reset to the first group V1-H1 and the display is reset to On. The scale display is reset to Off, the position is reset to the middle, the color is reset to red and the width is reset to 1.

9.12 Set date / time

Change the time in

Click the mouse to change the year or

month or day or hour or minute or second, the mouse wheel increases or decreases

after the cursor flickers at the top, also you can click the small arrow on the right to

change the setting system time. Click on

to modify the system time.

9.13 Switch between two displays
If you connect the monitor, you can control the display on the monitor or on the small screen here. If you connect HDMI before switching on, the system automatically displays the screen. Note: You cannot switch when you are recording the screen. After you have clicked on the switch, the time-lapse photo will end automatically.

9.14 Move and hide toolbar

This button hides the side toolbar. If the current toolbar page is the measurement

page.

If you click on the “Measurement tool”

button in

the left toolbar (with the exception of “Delete all”, “Screenshot” and “Table”) and move

the mouse over the left toolbar, the left toolbar is automatically hidden. If you want to

show it, you can click the button

in the upper left corner to show it or click

to show the left status bar.

There is a move button

at the bottom of the camera, measurement and other

pages in the left status bar, which can be clicked to move the left status bar to the right,

and the simple measurement tool on the right will be moved to the left when it is

displayed.

29

OIV-3-BA-d-2411

10 Toolbar

10.1 Camera When the system recognizes that the USB flash drive is inserted and the available

capacity is sufficient, click

to start recording. The top right corner will show that

the photo is being saved. During this time, the camera button cannot be clicked again.

The camera button can only be clicked when the message “Recording successful” is

displayed in the top right-hand corner. The captured photo is named according to the

current system time. Once the photos have been successfully saved, you can open the

gallery to view the photos taken.

10.2 Camera timer If the system recognizes that the USB flash drive is inserted and the available
capacity is sufficient, the timer page of the camera is displayed.

Total time mode (Total time mode): Set the timer mode to the number of photos taken in total time. Interval time mode (interval tome mode): Set the interval at which a photo is taken. Day (Day): Set the number of days. Click with the mouse, scroll wheel or click on the small button to control the increase or decrease. Hour (Hour): Set the hours. Click with the mouse, scroll wheel or click on the small control button to increase or decrease.

30

OIV-3-BA-d-2411

Minute: Set the minutes. Click the mouse, scroll wheel or click the small button to increase or decrease the control. Second (Second): Set the second speed. Click the mouse, scroll wheel or click the small control button to increase or decrease the speed. Frequency: Specify the number of recordings. Click the mouse, scroll wheel or click the small control button to increase or decrease the frequency. Interval Time (Interval Time): Displays the time interval for each photo taken in the current mode. Time for next photo (Next photo time): Shows how much time is left until the next photo is taken. Name of the last photo (Last photo name): Displays the name of the last photo taken. Number of shots (Numbers of shots): Number of timed photos taken this time Start: The camera timer starts. Note: Time photography cannot be activated if the interval is less than two seconds. Pause: The camera timer pauses. Stop: Camera timer is stopped Close: Closing the camera timer window and stopping the camera timer

31

OIV-3-BA-d-2411

10.3 Video recording

If the system recognizes that the USB flash drive is inserted and the available capacity

is sufficient, click on the Video button

to start recording. The status of the button

is

The recording time and the indicator

are displayed in the top right or left corner.

Click the video button again to end the video recording. At this time, the upper right or

left corner will display that the recording is being saved and the video recording button

cannot be clicked. When the saving process is complete, the success of the recording

and the. The recording button is restored.

You can resume recording at this

point.

If the storage space on the USB flash drive is less than 50 MB, the recording will stop

automatically. After saving, recording is automatically activated and continued until the

storage space on the USB flash drive is less than 50 MB.

Note: If you disconnect the USB flash drive during recording, the video file may be

damaged and can no longer be opened.

10.4 Gallery

If the system recognizes that the USB stick is inserted and is not recording, you can

call up the gallery to view

photos and videos.

Switch to the image preview.

Switch to the video preview. A appears you that the current file is a video file.

above each video to remind

to the first page

to the last page

32

OIV-3-BA-d-2411

to the previous page to the next page
Displays the current page number. After clicking, a drop-down field appears from which you can select the page you want to jump to.
Sort order, date or name. Sort order, ascending or descending. Select an image or video and click Delete to delete the image or video. You can also select multiple items and then click Delete to delete them from the batch. Home, Leave gallery.

10.5 Rename image and video files

After you have clicked on the image or video file, the selected image or video appears in the selected style.

Right-click on the file to open the menu and select “Rename” or doubleclick on the file name to be renamed.

The cursor appears at the file name and calls up the keyboard.

After you have entered the name, click on and complete the name change.

or to close the keyboard

33

OIV-3-BA-d-2411

10.6 Delete image and video files

You can select one or more image/video files for deletion. After you have selected the

file(s), the file(s) will have a blue background.

Then click on

to

delete the file(s). Deleting the menu bar, which is called up by right-clicking on the file,

can only delete the currently selected individual file.

10.7 Open image files Double-click on the image to open it or right-click on the file to call up the menu and click on “Open” to open the image. After opening, the name of the current file is displayed in the control bar.

to switch to the previous screen

to switch to the next picture

Leave gallery

10.8 Open video files

Double-click on the video to open it or right-click on the file, call up the menu and click

on “Open” to open the video. After opening, the name of the current file is displayed in

the control bar. After the video has been opened, it will play automatically and the “Play”

button will change to

Click

again to pause the current video and

use the “Play” button to change to

Click

to switch to the previous video

and play it automatically. Click on

to switch to the next video and play it

automatically. Click to

pause the current video. Click to

exit

the

gallery.

34

OIV-3-BA-d-2411

10.9 File details
Right-click on the desired file, call up the menu bar and click on “Details” to display the details of the current file.

35

OIV-3-BA-d-2411

11 Optional accessories
Accessories Auxiliary lens 0.5x Auxiliary target 2.0x Solder protection lens Cleaning set for microscopes

OBB-A3225 OBB-A3226 OBB-A3227 OCS 901

12 Troubleshooting

Problem

Cause

Solution

Stains or dust on the picture

Contaminants are present on the sample

Clean sample

The image is blurred
The video image is not displayed
The picture is too bright or too dark
The display does not respond

Impurities are on the lens surface

Clean lens

The focus is not correct Setting the focus

The external power cable is not connected The LED lighting is set incorrectly
Check the display settings

Connect the external power cable
Set the brightness of the LED ring light.
Open the “Settings” menu to set the brightness

Overlaps in the system (too many processes) Restart the system lead to error loops

The LED ring light does not light up when it is switched on

No power
The LED light has burned out

Check the connection of the mains cable
Replace them with new lighting

The LED light has suddenly burned out

The voltage is too high

Use a suitable external mains adapter

The lighting brightness is not sufficient

The voltage is too low

Use a suitable external mains adapter

36

OIV-3-BA-d-2411

13 Service
If, despite studying these operating instructions, you still have questions about commissioning or operation, or if, contrary to expectations, a problem should occur, please contact your specialist dealer. The device may only be opened by trained service technicians authorized by KERN.
14 Power supply
14.1 Mains connection The microscope may only be connected to the mains if the information on the microscope ( sticker ) and the local mains voltage are identical.
Important: – Check the mains cable for damage before commissioning – Ensure that the power supply unit does not come into contact with liquids – The mains plug must be accessible at all times.

37

OIV-3-BA-d-2411

15 Maintenance, servicing and disposal
Disconnect the appliance from the power supply before carrying out maintenance, cleaning or repair work.

15.1 Maintenance and cleaning
The appliance must always be kept clean and regularly freed from dust. Before wiping the appliance when it gets wet, make sure that the power is switched off.
Glass components should preferably be wiped lightly with a lint-free cloth if they become dirty.
To wipe off oil stains or fingerprints from lens surfaces, the lint-free cloth is moistened with a mixture of ether and alcohol (70/30 ratio) and then used for cleaning
Ether and alcohol must always be handled with care as they are highly flammable substances. It is therefore essential to keep them away from naked flames and electrical appliances that are switched on and off and only use them in well-ventilated rooms.
However, organic solutions of this type should not be used to clean other components of the appliance. This could cause changes to the paintwork. It is sufficient to use a neutral cleaning agent for this purpose.
Other cleaning agents for the optical components include :
· Special cleaner for optical lenses · Special optical cleaning cloths · Bellows · Brush
If handled correctly and checked regularly, the microscope will function smoothly for many years
Should a repair nevertheless be necessary, please contact your KERN dealer or our Technical Service.
15.2 Waste disposal

38

OIV-3-BA-d-2411

Old appliances and accessories must not be disposed of with household waste. The operator must dispose of the packaging and the device at the place of use in accordance with the applicable national or regional legislation The device consists of various components and materials, such as
· Electronic components (printed circuit boards, electrical cables
· Plastic (housing)
· Metal Improper disposal of the appliance can have harmful effects on people and the environment. Proper and environmentally friendly disposal can prevent harmful effects and recover raw materials.
Disposal of rechargeable batteries and batteries:
Rechargeable batteries and batteries do not belong in household waste. The disposal of rechargeable batteries and batteries must be carried out by the operator in accordance with the applicable national or regional law of the place of use.

39

OIV-3-BA-d-2411

16 Battery law Note in accordance with the Battery Act – BattG:
INFORMATION
· The following information is valid for Germany.
In connection with the sale of batteries and rechargeable batteries, we are obliged as a dealer under the Battery Act to inform end users of the following:
· End users are legally obliged to return used batteries and rechargeable batteries.
· After use, batteries and rechargeable batteries can be returned free of charge to municipal collection points or retailers. The normal end of use of the batteries/rechargeable batteries must be reached, otherwise precautions against short circuits must be taken.
· The return option is limited to batteries and rechargeable batteries of the type that we carry or have carried in our range and to the quantity that end consumers usually dispose of.
· A crossed-out waste garbage can means that you must not dispose of batteries or rechargeable batteries in household waste under any circumstances. Old batteries or rechargeable batteries may contain harmful substances that can damage people and the environment if not disposed of correctly.
· Batteries containing harmful substances are marked with a symbol consisting of a crossed-out dustbin and the chemical symbol (Cd = cadmium, Hg = mercury, or Pb = lead) of the heavy metal responsible for the classification as containing harmful substances.

40

OIV-3-BA-d-2411

17 Further information
The illustrations may differ slightly from the product. The descriptions and illustrations in these operating instructions are subject to change without notice. Further developments to the device may result in such changes.
All language versions include a non-binding translation.
The original German document is binding.