Specifications

  • Display module Type: LCD-H210
  • Input Current: 4-20 mA
  • Maximum current: 30 mA
  • Minimum current for operation: ~3.5 mA
  • Voltage drop: 4.5 V
  • Indication Display: 4-digit LCD with backlight
  • Indication range: -1999 to 9999
  • Digit Height: 8.9 mm / 0.35 inch
  • Decimals: Not specified
  • Underrange / Overrange: Not specified
  • Response time: Not specified
  • Temperature influence: Not specified
  • Configuration method: Not specified
  • Typical accuracy: Not specified
  • NAMUR NE 43 compliance: Not specified
  • HART transparent: Not specified
  • Electrical connection: Push-in spring connection
  • Connection head Type: D2
  • Material of body / window: Not specified
  • Cable gland thread D2: Not specified
  • Process connection thread D1 / Hole d1: Not specified
  • Paint type/colour: RAL 9006 (white aluminium)
  • Protection class: Not specified

Product Usage Instructions

1. Installation

Installation, assembly, start-up, and maintenance should only be
performed by appropriately trained personnel following regional
occupational health and safety directives.

2. Mounting the LCD-H210

The LCD-H210 is designed to be fitted to a temperature sensor.
Ensure proper alignment and secure mounting to the sensor.

3. Electrical Connection

Connect the LCD indicator directly in a 4-20 mA loop without the
need for an external power supply. Use the provided terminal block
for electrical connections with push-in spring connection.

4. Configuration

Use the push buttons located on the device for configuration.
Refer to the detailed handbook for Display Module LCD-D100 for more
information on configuration options.

Frequently Asked Questions (FAQ)

Q: Can the LCD-H210 be used with other types of sensors?

A: The LCD-H210 is designed to be fitted specifically to a
temperature sensor. It is recommended to consult with the
manufacturer for compatibility with other sensors.

Q: What is the warranty period for the LCD-H210?

A: The warranty period may vary. Please refer to the warranty
information provided by the manufacturer or distributor.

USER INSTRUCTIONS
LCD-H210
Digital Loop Powered LCD Indicator Integrated into a Connection Head
The user instruction must be read prior to adjustment and/or installation. All information subject to change without notice. MEASURE OF SUCCESS
INOR Process AB, PO Box 9125, SE-200 39 Malmö, Sweden, Phone: +46 40 312 560, Fax: +46 40 312 570, E-mail: support@inor.se
INOR Transmitter OY, Unikkotie 13, FI-01300 Vantaa, Finland, Phone:+358 10 421 7900, Fax: +358 10 421 7901, E-mail: myynti@inor.fi
INOR Transmitter GmbH, Am See 24, D-47279 Duisburg, Germany, Phone: +49-203 7382 762 0, Fax: +49-203 7382 762 2, E-mail: info@inor-gmbh.de
KROHNE Temperature Division INOR, 55 Cherry Hill Drive, Beverly, MA 01915, United States
Phone: +1 978 826 6900, Fax: +1 978 535 1720, E-mail: inor-info@krohne.com
www.inor.com, www.krohne-inor.se www.krohne-inor.fi, www.inor-gmbh.de
This product should not be mixed with other kind of scrap, after usage. It should be handled as an electronic/electric device.
MEASURE OF SUCCESS

© INOR 09/2024 – 4010579401 – MA LCD-210 R01 en

INTRODUCTION
LCD-H210 is a digital loop powered LCD indicator integrated into a connection head with window. The connection head is designed to be fitted to a temperature sensor and can also be equipped with a built-in DIN B head mount 2-wire transmitter. The LCD indicator is installed directly in a 4-20 mA loop without need for external power supply. The indicator show numeric values in the range from -1999 to 9999 proportional to the 4-20 mA input signal on a digital display.
GENERAL INFORMATION
The LCD indicator is integrated into a connection head and is used to show numerical values proportional to a 4-20 mA analogue process signal in the range from -1999 to 9999 on a digital display. The indicator is loop powered and is connected to the 4-20 mA loop of a 2-wire transmitter or any device producing a 4-20 mA signal. The aluminum connection head with polycarbonate window is designed to be fitted to a temperature sensor and can also be equipped with a built-in DIN B head mount 2-wire transmitter for direct process value visualization. A typical application is to use LCD-H210 fitted to a temperature sensor and connected to the 4-20 mA loop from a signal conditioner for local view of the measured temperature. A high-contrast, 4-digit LCD display with backlight makes the display easy to read in any lighting conditions. Configuration of the indicator is done either with NFC and the smartphone app INOR Connect or with three push buttons.
INFORMATION! This manual describes in short form the various functions and technical data for the display, for a more detailed description please consult the handbook for Display Module LCD-D100.
CAUTION! Installation, assembly, start-up and maintenance may only be performed by appropriately trained personnel. The regional occupational health and safety directives must always be observed.

DATA (short form)

Display module Type Input Current Maximum current Minimum current for operation Voltage drop Indication Display Backlight Indication range Digit Height Decimals Underrange / Overrange Response time Temperature influence Configuration method
Typical accuracy NAMUR NE 43 compliance HART transparent Electrical connection
Connection head Type Material of body / window
Cable gland thread D2 Process connection thread D1 / Hole d1 Paint type/colour Protection class

INOR model LCD-D100 4-20 mA 30 mA ~3.5 mA 4.5 V
7-segment black LCD with clear background White LED powered from the 4-20 mA loop 4 digits (-1999 to 9999) 8.89 mm / 0.35″ Selectable, 0 to 3 Flashing symbols Lo (I 3.6 mA) / HI (I 21.0 mA Appr. 1 s, Power on delay: 5 s and stable after 1 min ±0.01 % FS / °C 3 push buttons or NFC, located on the rear of the display ±0.05% of span ±1 digit Yes Yes Push-in spring connections, Wire cross section 0.25 mm2-1.5 mm2
INOR model BUZ-HW Aluminium pressure die-casting / Polycarbonate M20x1.5 M24x1.5 / Ø14 mm Polyester/white aluminium (RAL 9006) Up to IP65 (depending on applied cable gland and sealing for process connection)

ORDERING INFORMATION

Product LCD-H210 LCD-H210 – Customized LCD-D100 – Display module only

Part No. 70LCDH2101 On request 70D1000001

114 / 4.48 41.5 / 1.63

LCD-H210 D2

DIMENSIONS
Ø 43 / 1.69 Window (polycarbonate) Display module LCD-D100

17 / 0.6 26 / 1.02

D1

d1

D2

M24x1.5 Ø14 M20x1.5

d1 D1

4x M4X6

33 / 1.29

Display screen size 39.5×14.0 / 1.5×0.5 Digit hight 8.9 / 0.35

mm / inch

The display module can be rotated 180° INDICATOR DESIGN

Back view
1 2

Front view 7

Color RAL 9006 1 (white aluminium)

1 2 3 4 PS+ +TX

3

5

4

6

1

1

1. Fixing screws to connection head

Side view

(2x M4)

2. Push buttons for configuration (3x)

3. NFC antenna

1

4. Terminal block for electrical

connections, Push-in spring connection

5. 4-digit LCD with backlight

6. Engineering unit (°C as defalut, labels

for different units are included)

7. Pre-monted cable for connection of a

2-wire transmitter (red +, white – )

3

2

4

DIMENSIONS
LCD-D100 – Display module only 1
Back view

63 / 2.48

1 2 3 4 PS+ +TX

1
1 Front view

14 / 0.5

63 / 2.48 55 / 2.16

21 / 0.8 9.5 / 0.37

Side view
1 Fixing screw to connection head
mm / inch

1 39.5 / 1.5 Ø 63 / 2.4 1
CONNECTIONS

Loop Power Supply 2-Wire Transmitter

+ 4-20 mA ~24 VDC _

+ 4-20 mA _

OutpIuntput

Loop Power Supply 2-Wire Transmitter

+ ~24 VDC _

+ 4-20 mA _

InpOuuttput

1 2 3 4 PS + +TX
1 Current source
4-20 mA
_+

1 2 3 4 PS + +TX
2
Note Terminal connections number 2 and 3 are internally linked with each other

1 2 3 4 PS + +TX
3
1. Connection to the indicator with power supply (for the transmitter) and a 2-wire transmitter by using the internal jumper between terminals 2 and 3 to establish a 4-20 mA current loop.
2. Connection to the indicator with power supply (for the transmitter) and a 2-wire transmitter without using the internal jumper between terminals 2 and 3.
3. Connection to the indicator with a current source (an active 4-20 mA signal)

CONFIGURATION – VIA PUSH BUTTONS

5s

Save

Time out

-1 Step

Save -1
Step

Time out

Save

Time out

Step

0000 3333

Save

Time out

Step

F 00 06

Save

Time out

Step

0

F

Save

Time out

Step

Fn 00 05 10 15 30

Save

Time out

Step

HLo off on

Save

Time out

Step

on off

Step

Submenu

Step

No

dEF

Down

CONFIGURATION – VIA PUSH BUTTONS

DISPLAY FUNCTION, NAMUR NE 43

3

2

1

1

2

3

4

5

6

7

8

9

10

11

1 Pushbutton with step (decrease) function
2 Pushbutton with enter function 3 Pushbutton with step (increase)
function

1 Parameter name 2 Factory default value 3 The default setting value refers
to display TAG / ID is the last four digits in the serial number.

Parameter description:
1. Engineering value related to the lower point (4 mA), the default setting value refers to 0.0 at lower point.
2. Engineering value related to the higher point (20 mA), the default setting value refers to 100.0 at higher point.
3. Decimal point location, the default setting value refers to 1 decimal. 4. Filter, the default setting value refers to “No filter”. 5. TAG / ID number, the default setting value refers to the last four digits in the se-
rial number for the display. 6. Display update interval, the default setting value refers to update display 7. immediately. 8. Backlight mode, the default setting value refers to the backlight on but flashing
during alarm. 9. NFC mode, the default setting value refers to the configuration via NFC is on. 10. Error correction, the default setting refers to no error correction. 11. Reset back to factory default. 12. End setup and return to indication.

INFORMATION! Consult the handbook for Display Module LCD-D100 for a more detailed description of the various functions.

CONFIGURATION – VIA APP
Before making a configuration of LCD-H210 you need to do following: 1. Make sure that you have a mobile device with NFC communication activated 2. Download the app INOR Connect to your mobile device

Required versions: iOS iOS 13 or later and Iphone 7 or later for NFC

Android Android 4.4 or later

Configuration procedure: 1. Launch the app by clicking on the App icon or
holding your mobile device against the display on the part of the device where NFC is located (only possible with Android).

2. Click on “Read Configuration” and hold your mobile device against the display as explained in the first section.

3. In the app you can edit the following: · Display indication range · Decimal points position · Backlight mode · Filter settings · TAG-no

Scan the QR code to download and install INOR Connect
Place the smartphone´s NFC antenna directly on the display´s NFC antenna for configuration via the app INOR Connect

4. Once you have set the desired values in the app, you transfer it to the display by clicking the transfer button and holding the mobile device against the display until a green check box appears confirming that the transfer has been completed.

Input current (mA) 21,0 mA 20,5 to <21,0 mA 20,0 to 20,5 mA 4,0 to 20,0 mA
3,8 to 4,0 mA >3,6 to 3,8 mA 3,6 mA

Signal description
Failure Overrange Normal operating range Underrange Failure

Indication on LCD-H210
Flashing HI Fixed value (max. range value +3.1 %) Extended indication range The configured indication range
Extended indication range Fixed value (min. range value -1.25 %) Flashing Lo

LIMITED WARRANTY
INOR Process AB, or any other affiliated company within the Inor Group (hereinafter jointly referred to as “Inor”), hereby warrants that the Product will be free from defects in materials or workmanship for a period of five (5) years from the date of delivery (“Limited Warranty”). This Limited Warranty is limited to repair or replacement at Inor’s option and is effective only for the first end-user of the Product. Upon receipt of a warranty claim, Inor shall respond within a reasonable time period as to its decision concerning:
1 Whether Inor acknowledges its responsibility for any asserted defect in materials or workmanship; and, if so,
2 the appropriate cause of action to be taken (i.e. whether a defective product should be replaced or repaired by Inor).
This Limited Warranty applies only if the Product: 1 is installed according to the instructions furnished by Inor; 2 is connected to a proper power supply; 3 is not misused or abused; and 4 there is no evidence of tampering, mishandling, neglect, accidental damage, modification or repair without the approval of Inor or damage done to the Product by anyone other than Inor.
This Limited Warranty is provided by Inor and contains the only express warranty provided. INOR SPECIFICALLY DISCLAIMS ANY EXPRESS WARRANTY NOT PROVIDED HEREIN AND ANY IMPLIED WARRANTY, GUARANTEE OR REPRESENTATION AS TO SUITABILITY FOR ANY PARTICULAR PURPOSE, PERFORMANCE, QUALITY AND ABSENCE OF ANY HIDDEN DEFECTS, AND ANY REMEDY FOR BREACH OF CONTRACT, WHICH BUT FOR THIS PROVISION, MIGHT ARISE BY IMPLICATION, OPERATION OF LAW, CUSTOM OF TRADE OR COURSE OF DEALING, INCLUDING IMPLIED WARRANTIES OF MER-CHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. EXCEPT AS PROVIDED HEREIN, INOR FURTHER DISCLAIMS ANY RESPONSIBILITY FOR LOSSES, EXPENSES, INCONVENIENCES, SPECIAL, DIRECT, SECONDARY OR CONSEQUENTIAL DAMAGES ARISING FROM OWNERSHIP OR USE OF THE PRODUCT.
Products that are covered by the Limited Warranty will either be repaired or replaced at the option of Inor. Customer pays freight to Inor, and Inor will pay the return freight by post or other “normal” way of transport. If any other type of return freight is requested, customer pays the whole return cost.

Documents / Resources

INOR LCD-H210 Digital Loop Powered LCD Indicator Integrated [pdf] User Manual
LCD-H210 Digital Loop Powered LCD Indicator Integrated, LCD-H210, Digital Loop Powered LCD Indicator Integrated, Powered LCD Indicator Integrated, LCD Indicator Integrated, Indicator Integrated

References

Leave a comment

Your email address will not be published. Required fields are marked *