IMPLEN NP80 NanoPhotometer Nano Volume and Cuvette
Description of the Qualification
Objectives
This qualification is for the documented validation of the NanoPhotometer , to show that all is conform with the requirements, specifications of the installation and function.
Instrument Description
The NanoPhotometer measures sprectra in the range of 200 nm to 900 nm. Due to the integrated beam deflection and the use of fiber-optic light conductors, the sample can be measured in standard volumes
(cuvettes) and/or NanoVolumes. For the detection there is a 3648 pixel CCD array used. The instrument is maintainance and recalibration free due to sealed optics, the True Path and Sample Compression Technology™. This guarantees for a high accuracy and linearity over the entire life-time.
The following methods are available:
- single and multi wavelengths
- wavescan in the range of 200 nm – 900 nm (adjustable)
- concentration calculations (pre-programmed for nucleic acids and proteins)
- frequency of incorporation/degree of labelling of various fluorescent dyes (pre-programmed for nucleic acids and protein)
- cell density measurements
- colorimetric assays
- standard curves
- absorbance ratio calculations
General Data
System: | UV/Vis spectrophotometer | ||||
Type: | NanoPhotometer | ||||
New device | YES | ||||
Used device | NO | ||||
Serial Number: | M80798 | ||||
Operating System: | Windows10 | ||||
Location: | Company: | Implen GmbH | |||
Department: | Product Management | ||||
Person in Charge: | Dr. Helena Funk | if | +49 89 726 371 80 | ||
User: | Dr. Helena Funk | if | +49 89 726 371 80 | ||
Location: | Production | ||||
Adress: |
Schatzbogen 52 |
||||
City: | Munich | ||||
State/Province & Postal: | 81829 | ||||
Country: | Germany |
Validation Tools
The certified validation tools for the NanoPhotometer is a Didymiumglassfilter calibrated against a traceable NIST standard by an accredited supplier and a potassium hydrogen phthalate solution calibrated against a certified UV/Vis spectrophotometer. Calibration certificates can be furnished upon request.
Installation Qualification
Check of delivered items
Test description:
Verification of the delivery content by checking the delivery note.
Enclose a copy of the delivery note to the qualification report.
Test plan | Test report | ||
Test parameter | Acceptance criteria | passed | remarks |
Check the delivery | Complete according to delivery
note |
YES |
Inspection of Documentation
Test plan | Test report | ||
Test parameter | Acceptance criteria | passed | remarks |
User Manuals | – Short Instructions NanoPhotometer NP80/N60/N50/C40
– NanoPhotometer User Manual in digital version on USB flash drive |
YES |
|
Certificates | NanoPhotometer Operation
Certificate: NanoVolume Application Cuvette Application |
YES |
Inspection of the Instrument
Test description:
Validation and approval of the NanoPhotometer installation in a technical correct manner according to the user manual of the supplier.
Test plan | Test report | ||
Test parameter | Acceptance criteria | passed | remarks |
Installation of the NanoPhotometer | Power supply plugged in (operating voltage: 90-250 V, 50/60 Hz, 90W, 19 VDC) and
load the battery pack at least for 3 hours prior to the first use |
YES |
|
By switching the instrument on a self-diagnostic check is automatically performed | pass all tests – no failure message appears – main display is shown |
YES |
|
Integrated vortexer | Activated if NanoPhotometer is switched on and deactivated if the NanoPhotometer is
switched off. |
YES |
Installation of Software
Test description:
Validation and approval of the software installation in a technical correct manner according to the user manual of the supplier.
Test plan | Test report | ||
Test parameter | Acceptance criteria | passed | remarks |
Built-in Touchscreen | home screen is shown after
starting the instrument |
YES | N/A when no touch screen is
connected |
computer (PC/MAC) | software can be opened and the home screen of the connected NanoPhotometer is shown |
YES |
N/A when using the built-in touch screen, a smart phone or tablet for controlling the NanoPhotometer. |
Tablet | software can be opened and the home screen of the connected NanoPhotometer is shown |
YES |
N/A when using only the built-in touch screen, a computer or tablet for controlling the
NanoPhotometer. |
Smart Phone | software can be opened and the home screen of the connected NanoPhotometer is shown |
YES |
N/A when using only the built-in touch screen, a computer or tablet for controlling the
NanoPhotometer. |
Installation of Printer
Installation of Printer | ||||
Test plan | Test report | |||
Test parameter | Acceptance criteria | passed | remarks | |
Printer via USB
cable/network/WiFi network |
Print icon is shown on all
method screens. |
YES | N/A when no printer is
available |
|
Printer is functional | Do a measurement and print
the results. |
YES | N/A when no printer is
available |
|
Printer installataion passed | YES |
Operation Qualification
Validation of the Calibration Standard
Test description Didymiumglassfilter:
As a validation tool, the Didymiumglassfilter is the secondary spectrometric calibration standard used. This filter is calibrated and checked against calibration- or control standards, which are NIST proven. For all validation tools which are used during the qualification, an actual and valid calibration certificate is necessary. Enclose a copy of the calibration certificate to the qualification report.
Test plan | Test report | ||||
Test parameter | Serial number | Acceptance criteria | passed | remarks | |
Didymiumglassfilter: | E2343 | serial number correct | YES | ||
Valid calibration certificate: | Next calibration: |
5.11.2021 |
YES |
recalibration of the Didymiumglassfilter every
24 months |
Test description potassium hydrogen phthalat solution:
As validation tool a potassium hydrogen phthalat solution used. This solution is calibrated and checked against a certified UV/Vis spectrophotometer.
Test plan | Test report | |||
Test parameter | Acceptance criteria | passed | remarks | |
Lot number | 1606 | lot number correct | YES | After opening a vial it can be
used for one hour. |
Absorbance @ 280 nm
(10 mm path) |
21,40 | Absorbance entered | YES | The absorbance can be
found on the vial. |
Validation of the Wavelength Accuracy – Cuvette
Test description:
Method: More Apps – Wavescan
Parameter Settings: Change to Cuvette, Pathlength 10 mm, Start Wavelength 200 nm, End Wavelength 900 nm, Baseline Correction Off, Smoothing 1
Blank: air
Sample: Didymiumglassfilter S/N E2343
approx. position: | 329 nm | 472 nm | 512 nm | 681 nm |
certified expected peaks: | 328,71 | 472,06 | 512,86 | 681,29 |
measured value: | 328 | 472 | 511 | 680 |
Tolerance is +/- 2nm
Test plan | Test report | |||
Test parameter | Serial number | Acceptance criteria | passed | remarks |
Didymiumglassfilter | E2343 | Peak positions (see also calibration certificate) |
YES |
Peak position must be visible in the graph, if not listed in the table please check manual for the corresponding peak |
Validation of Absorbance Accuracy – Cuvette
Test description:
Cuvette mode
Method: More Apps – Wavelength
Parameter Settings: Change to Cuvette, Pathlength 10 mm, add 4 Wavelengths: enter wavelengths 270 nm, 280 nm, 321 nm and 341 nm, Baseline Correction Off, Smoothing 1
Blank: air
Sample: Didymiumglassfilter S/N E2343
certified expected
absorbances: |
tolerance | measured values: | |
A270 | 1,145 | +/- 0.050 | 1,140 |
A280 | 1,143 | +/- 0.046 | 1,140 |
A321 | 0,790 | +/- 0.046 | 0,790 |
A341 | 0,495 | +/- 0.034 | 0,490 |
Test plan | Test report | |||
Test parameter | Serial number | Acceptance criteria | passed | remarks |
Didymiumglassfilter | E2343 | The measured absorbances (270, 280, 321 and 341 nm)
have to be in the tolerance range defined in the calibration certificate |
YES |
Validation of the Reproducibility – Cuvette
Test description:
Method: More Apps – Wavelength
Parameter Settings: Change to Cuvette, Pathlength 10 mm, enter wavelength 270nm, 280 nm, 321 nm and 341 nm, Baseline Correction Off, Smoothing 1
Blank: air
Sample: Didymiumglassfilter S/N E2343
certified expected
absorbances: |
tolerance | measured values: | measured values: | measured values: | measured values: | CV% | |
A270 | 1,145 | +/- 0.050 | 1,145 | 1,144 | 1,145 | 1,145 | 0,038 |
A280 | 1,143 | +/- 0.046 | 1,143 | 1,142 | 1,144 | 1,143 | 0,062 |
A321 | 0,790 | +/- 0.046 | 0,790 | 0,790 | 0,799 | 0,790 | 0,492 |
A341 | 0,495 | +/- 0.034 | 0,495 | 0,495 | 0,495 | 0,495 | 0,000 |
Test plan | Test report | |||
Test parameter | Serial number | Acceptance criteria | passed | remarks |
Didymiumglasfilter | E2343 | CV% = 0.75% |
YES |
Validation of the Baseline – Cuvette
Test description:
Method: More Apps – Wavelength
Parameter Settings: Change to Cuvette, Pathlength 10 mm, enter wavelength 280 nm, Baseline Correction 377 nm, Smoothing 1
Blank: air
Sample: empty cell holder (air)
tolerance | measured
values: |
measured
values: |
measured
values: |
measured
values: |
Absmax | � Abs | |
A280 | +/- 0.002 | 0,000 | 0,000 | 0,000 | 0,001 | 0,001 | 0,001 |
Test plan | Test report | ||
Test | Acceptance criteria | passed | remarks |
Measurement against air / empty cell holder | Tolerance is +/- 0.002 A according to the technical specifications |
YES |
Validation of Absorbance Accuracy – NanoVolume
Test description:
Method: More Apps – Wavelength
Parameter Settings: Dilution 15, Wavelength 280 nm, Baseline Correction 377 nm, Smoothing 1 Blank: water
Volume: 1.5 µl
Sample: vial with lot number 1606
certified
expected absorbances: |
measured values: | |
A280 | 21,40 | 21,390 |
Test plan | Test report | |||
Test parameter | Lot number | Acceptance criteria | passed | remarks |
Potassium hydrogen phthalate solution (PHP) | 1606 | The absorbance at 280 nm has to be in the certified range |
YES |
Validation of the Reproducibility – NanoVolume
Test description:
Method: Protein UV
Parameter Settings: BSA, Volume 1 µl- 2 µl, wavelength 280 nm, Background Correction: On, Manual Dilution: Off Blank: water
Volume: 1.5 µl
Sample: water/pipette once and measure four times
certified expected
absorbances: |
measured values: | measured values: | measured values: | measured values: | Absmax | ∆Abs | |
A280 | 0,000 | 0,000 | 0,000 | 0,002 | 0,000 | 0,002 | 0,002 |
Test plan | Test report | |||
Test parameter | Lot number | Acceptance criteria | passed | remarks |
Water | 1606 | Four repeated absorbance measurements tolerance is +/-
0.03 A (10 mm path) according to the technical specifications. |
YES |
Validation of the Baseline – NanoVolume
Test description:
Method: More Apps – Wavelength
Parameter Settings: Dilution 15; wavelength 280 nm, Baseline Correction 377 nm, Smoothing 1 Blank: air
Sample: air – no sample
tolerance | measured
values: |
measured
values: |
measured
values: |
measured
values: |
Absmax | � Abs | |
A280 | +/- 0.03 | 0,002 | 0,005 | 0,000 | -0,005 | 0,005 | 0,010 |
Test plan | Test report | ||
Test | Acceptance criteria | passed | remarks |
Measurement against air / empty cell holder | Tolerance is +/- 0.03 A (10 mm path) according to the technical specifications |
YES |
Results and Release of Installation and Operation Qualification
The below listed persons confirm with their signature the succsessful installation and operation qualification of the NanoPhotometer .
Valuation | passed | remarks |
All tests performed | YES | |
All acceptance criteria fullfiled | YES | |
System ready to use | YES |
Name (block letter): | Date: | Signature: | |
Performed by: |
Dr. Helena Funk |
13.03.2020 |
|
Approved by: |
Dr. Michael Riepl |
13.03.2020 |
Attachments
- Copy of delivery note
- Calibration Certificates for the NanoPhotomter
- Calibration Certificate for the Didymiumglassfilter
- Calibration Certificate for the Standard Solution
Documents / Resources
![]() |
IMPLEN NP80 NanoPhotometer Nano Volume and Cuvette [pdf] Instructions NP80, NanoPhotometer Nano Volume and Cuvette |