TIM Series Insertion Paddle Wheel Flow Meter Sensor
“
Specifications
- Operating Range: 0.1 to 10 m/s
- Pipe Size Range: DN15 to DN600
- Linearity: Provided
- Repeatability: Provided
Product Description
The Insertion Paddle Wheel Flow Meter Sensor features a high
impact NEMA 4X enclosure made of TIM Thermal Plastic. It includes a
vivid LED display for flow and total measurements. The design is
based on NASA Shape Effects on Drag and includes TI3M 316 SS
material, M12 quick connection, true union design, and Zirconium
ceramic rotor and bushings for enhanced wear resistance and
durability.
Product Usage Instructions
Safety Information
- De-pressurize and vent the system before installation or
removal. - Confirm chemical compatibility before use.
- Do not exceed maximum temperature or pressure
specifications. - Always wear safety goggles or face-shield during installation
and service. - Do not alter the product construction.
Installation
- Ensure the system is de-pressurized and vented.
- Confirm chemical compatibility with the sensor.
- Select appropriate installation fittings based on pipe
size. - Hand-tighten the sensor into place, do not use tools.
Rotor Pin | Paddle Replacement
- Line up the pin with the hole in the flow meter.
- Gently tap the pin until it is 50% out.
- Pull out the paddle carefully.
- Insert the new paddle into the flow meter.
- Push in the pin approximately 50% and tap gently to
secure. - Ensure holes are aligned properly.
FAQ
Q: What should I do if the sensor is under pressure?
A: Take caution to vent the system prior to
installation or removal to avoid equipment damage or injury.
Q: Can I use tools during installation?
A: Do not use tools as they may damage the
product beyond repair and void the warranty.
“`
Truflo® — TIM | TI3M Series (V1)
Insertion Paddle Wheel Flow Meter Sensor
Quick Start Manual
Read the user’s manual carefully before starting to use the unit. Producer reserves the right to implement changes without prior notice.
2F4i-n05d78Q©uaIcolintPyroPcersos Cdounctrtolss LOtdn. line at:
Valuetesters.com
info@valuetesters.com1
Truflo® — TIM | TI3M Series (V1)
Insertion Paddle Wheel Flow Meter Sensor
Safety Information
De-pressurize and vent system prior to installation or removal Confirm chemical compatibility before use DO NOT exceed maximum temperature or pressure specifications ALWAYS wear safety goggles or face-shield during installation and/or service DO NOT alter product construction
Warning | Caution | Danger
Indicates a potential hazard. Failure to follow all warnings may lead to equipment damage, injury, or death.
Hand Tighten Only
Over tightening may permanently damage product threads and lead to failure of the retaining nut.
Note | Technical Notes
Highlights additional information or detailed procedure.
Do Not Use Tools
Use of tool(s) may damage produced beyond repair and potentially void product warranty.
WARNING
Personal Protective Equipment (PPE)
Always utilize the most appropriate PPE during installation and service of Truflo® products.
Pressurized System Warning
Sensor may be under pressure. Take caution to vent system prior to installation or removal. Failure to do so may result in equipment damage and/or serious injury.
2F4i-n05d78Q©uaIcolintPyroPcersos Cdounctrtolss LOtdn. line at:
Valuetesters.com
info@valuetesters.com2
Truflo® — TIM | TI3M Series (V1)
Insertion Paddle Wheel Flow Meter Sensor
Product Description
The TI Series insertion plastic paddle wheel flow meter has been engineered to provide long-term accurate flow measurement in tough industrial applications. The paddle wheel assembly consists of a engineered Tefzel® paddle and micro-polished zirconium ceramic rotor pin and bushings. High performance Tefzel® and Zirconium materials have been selected due to their excellent chemical and wear resistant properties.
*
Rotates 330° *Optional
High Impact NEMA 4X Enclosure
TIM Thermal Plastic
Vivid LED Display
(Flow & Total)
Features ? ½” 24″ Line Sizes ? Flow Rate | Total ? Pulse | 4-20mA | Voltage Outputs (Optional)
New ShearPro® Design ? Contoured Flow Profile ? Reduced Turbulence = Increased Longevity ? 78% Less Drag than Old Flat Paddle Design*
*Ref: NASA “Shape Effects on Drag”
Tefzel® Paddle Wheel ? Superior Chemical And Wear Resistance vs PVDF
TI3M 316 SS
M12 Quick Connection
True Union Design
vs. Flat Paddle
Zirconium Ceramic Rotor | Bushings
? Up to 15x the Wear Resistance ? Integral Rotor Bushings Reduce Wear
and Fatigue Stress
360º Shielded Rotor Design
? Eliminates Finger Spread ? No Lost Paddles
TIM Thermal Plastic
TI3M 316 SS
vs. Competitor 2F4i-n05d78Q©uaIcolintPyroPcersos Cdounctrtolss LOtdn. line at:
Valuetesters.com
info@valuetesters.com3
Truflo® — TIM | TI3M Series (V1)
Insertion Paddle Wheel Flow Meter Sensor
Technical Specifications
General
Operating Range Pipe Size Range Linearity Repeatability
0.3 to 33 ft/s ½ to 24″ ±0.5% of F.S @ 25°C | 77°F ±0.5% of F.S @ 25°C | 77°F
0.1 to 10 m/s DN15 to DN600
Wetted Materials
Sensor Body O-Rings Rotor Pin | Bushings Paddle | Rotor
PVC (Dark) | PP (Pigmented) | PVDF (Natural) | 316SS FKM | EPDM* | FFKM* Zirconium Ceramic | ZrO2 ETFE Tefzel®
Electrical
Frequency
49 Hz per m/s nominal
15 Hz per ft/s nominal
Supply Voltage Supply Current
10-30 VDC ±10% regulated <1.5 mA @ 3.3 to 6 VDC
<20 mA @ 6 to 24 VDC
Max. Temperature/Pressure Rating Standard and Integral Sensor | Non-Shock
PVC PP PVDF 316SS
180 Psi @ 68°F | 40 Psi @ 140°F 180 Psi @ 68°F | 40 Psi @ 190°F 200 Psi @ 68°F | 40 Psi @ 240°F 200 Psi @ 180°F | 40 Psi @ 300°F
12.5 Bar @ 20°C | 2.7 Bar @ 60°F 12.5 Bar @ 20°C | 2.7 Bar @ 88°F 14 Bar @ 20°C | 2.7 Bar @ 115°F 14 Bar @ 82°C | 2.7 Bar @ 148°F
Operating Temperature
PVC PP PVDF
32°F to 140°F -4°F to 190°F -40°F to 240°F
0°C to 60°C -20°C to 88°C -40°C to 115°C
316SS
-40°F to 300°F
-40°C to 148°C
Output
Pulse | 4-20mA | Voltage (0-5V)*
Display
LED | Flow Rate + Flow Totalizer
Standards and Approvals
CE | FCC | RoHS Compliant See Temperature and Pressure Graphs for more information
* Optional
Model Selection
Size ½” – 4″ 6″ – 24″ 1″ – 4″ 6″ – 24″ 1″ – 4″ 6″ – 24″
PVC | PP | PVDF
Part Number TIM-P-S TIM-P-L TIM-PP-S TIM-PP-L TIM-PF-S TIM-PF-L
Add Suffix `E’ – EPDM Seals
Material PVC PVC PP PP PVDF PVDF
316 SS
Size ½” – 4″ 6″ – 24″
Part Number TI3M-SS-S TI3M-SS-L
Add Suffix `E’ – EPDM Seals
Material 316 SS 316 SS
2F4i-n05d78Q©uaIcolintPyroPcersos Cdounctrtolss LOtdn. line at:
Valuetesters.com
info@valuetesters.com4
Truflo® — TIM | TI3M Series (V1)
Insertion Paddle Wheel Flow Meter Sensor
Display Characteristics
LED Display
Flow Total
M12 Connection
Dimensions (mm)
Flow Rate
Unit | Output Indicators
91.7
91.7
106.4 210.0
179.0
2F4i-n05d78Q©uaIcolintPyroPcersos Cdounctrtolss LOtdn. line at:
Valuetesters.com
info@valuetesters.com5
Truflo® — TIM | TI3M Series (V1)
Insertion Paddle Wheel Flow Meter Sensor
Wiring Diagram
182
7
3
6
4
5
Terminal 1 2 3 4 5 6
M12 Female Cable
Description + 10~30 VDC Pulse Output
– VDC Pulse Output + 4-20mA or V* – 4-20mA or V*
Brown | 10~30VDC Black | Pulse Output
White | Pulse Output Grey | mABlue | -VDC Yellow | mA+
Color Brown White
Blue Black Yellow Grey
*Optional
Wiring – SSR* (Totalizer)
Set “Con n” in Pulse Output Control (Refer Pulse Control Programmming, Page 12)
Wire Color Brown White Blue
Description + 10~30VDC Pulse Output
-VDC * SSR – Solid State Relay
Wiring – One Pulse/Gal | Con E
Set “Con E” in Pulse Output Control (Refer Pulse Control Programmming, Page 12)
Wire Color Brown Black Blue
Description + 10~30VDC Pulse Output (OP2)
-VDC
Wiring – SSR* (Flow Rate)
Set “Con F/E/r/c” in Pulse Output Control (Refer Pulse Control Programmming, Page 12)
Wire Color Brown Black Blue
Description + 10~30VDC Pulse Output
-VDC * SSR – Solid State Relay
Wiring – To Flow Display | Con F
Set “Con F” in Pulse Output Control (Refer Pulse Control Programmming, Page 12)
Wire Color Brown White Blue
Description + 10~30VDC Paddle Pulse
-VDC
2F4i-n05d78Q©uaIcolintPyroPcersos Cdounctrtolss LOtdn. line at:
Valuetesters.com
info@valuetesters.com6
Truflo® — TIM | TI3M Series (V1)
Insertion Paddle Wheel Flow Meter Sensor
Installation
Retention Cap
Very Important
Lubricate O-rings with a viscous lubricant, compatible with the materials of construction.
Using an alternating | twisting motion, carefully lower the sensor into the fitting. | Do Not Force | Fig-3
Ensure tab | notch are parallel to flow direction | Fig-4
Hand tighten the sensor cap. DO NOT use any tools on the sensor cap or the cap threads or fitting threads may be damaged. | Fig-5
Lubricate with silicone inside of insertion fitting
Fig – 1
Fig – 2
Retention Cap
Flow Process Pipe
Fig – 3
Locating Pin
Ensure O-rings are well lubricated Notch
1¼” G
Sensor Blade Ensure tab is parallel to flow direction
Fig – 4 Top View
Correct Sensor Position
0011
Tab
Notch
VERY IMPORTANT Lubricate O-rings with a viscous 02 lubricant, compatible with the system 03
Fig – 5
Notch
Hand tighten using the retention cap
DO NOT use display to tighten
Locate the flow meter positioning tab and clamp saddle notch.
Engage one thread of the sensor cap, then turn the sensor until the alignment tab is seated in the fitting notch. Ensure tab is parallel to flow direction.
· Hand tighten the screw cap · DO NOT use any tools — threads may
be damaged · Ensure meter is firmly in place
2F4i-n05d78Q©uaIcolintPyroPcersos Cdounctrtolss LOtdn. line at:
Valuetesters.com
info@valuetesters.com7
Truflo® — TIM | TI3M Series (V1)
Insertion Paddle Wheel Flow Meter Sensor
Correct Sensor Position Setup
TI Series flow meters measure liquid media only. There should be no air bubbles and the pipe must always remain full. To ensure accurate flow measurement, the placement of the flow meters needs to adhere to specific parameters. This requires a straight run pipe with a minimum number of pipe diameters distance upstream and downstream of the flow sensor.
Flange
Inlet
Outlet
2x 90º Elbow
Inlet
Outlet
Reducer
Inlet
Outlet
10xID
5xID
25xID
5xID
15xID
5xID
90º Downward Flow
90º Elbow Downward Flow Upward
Inlet
Outlet
Inlet
Outlet
Ball Valve
Inlet
Outlet
40xID
5xID
Installation Positions
Figure – 1
20xID
5xID
Figure – 2
50xID
5xID
Figure – 3
Good if NO SEDIMENT present
Good if NO AIR BUBBLES present
*Maximum % of solids: 10% with particle size not exceeding 0.5mm cross section or length
2F4i-n05d78Q©uaIcolintPyroPcersos Cdounctrtolss LOtdn. line at:
Valuetesters.com
Preferred installation if SEDIMENT* or AIR BUBBLES
may be present
info@valuetesters.com8
Truflo® — TIM | TI3M Series (V1)
Insertion Paddle Wheel Flow Meter Sensor
Fittings and K-Factor
TEE FITTINGS
CLAMP-ON SADDLES
CPVC SOCKET WELD-ON ADAPTERS
Tee Fitting
IN
DN
½” (V1) 15
½” (V2) 15
¾”
20
1″
25
1½”
40
2″
50
2½”
65
3″
80
4″
100
K-Factor
LPM
156.1 267.6 160.0 108.0 37.0 21.6 14.4
9.3 5.2
GPM
593.0 1013.0 604.0 408.0 140.0
81.7 54.4 35.0 19.8
Sensor Length
S S S S S S S S S
Pressure vs. Temperature
bar psi 15.2 220
= PVC
= PP
= PVDF
13.8 200 12.4 180
11.0 160 9.7 140
8.3 120 6.9 100 5.5 80
4.1 60 2.8 40
1.4 20
00
°F 60
104
140
175
212
248
°C 20
40
60
80
100
120
Note: During system design the specifications of all components must be considered. | Non-Shock
Clamp Saddles
K-Factor
IN
DN
LPM GPM
2″
50
21.6
81.7
3″
80
9.3
35.0
4″
100
5.2
19.8
6″
150
2.4
9.2
8″
200
1.4
5.2
Sensor Length
S S S L L
*
Rotates 330°
PVC PP PVDF
316SS
Weld On Adapter
IN
DN
2″
50
2½”
65
3″
80
4″
100
6″
150
8″
200
10″
250
12″
300
14″
400
16″
500
18″
600
20″
800
24″
1000
K-Factor
LPM
14.4 9.3 9.3 5.2 2.4 1.4 0.91 0.65 0.5 0.4 0.3 0.23 0.16
GPM
54.4 35.5 35.0 19.8 9.2 5.2 3.4 2.5 1.8 1.4 1.1 0.9 0.6
Sensor Length
S S S S L L L L L L L L L
Min/Max Flow Rates
Pipe Size (O.D.)
½” | DN15 ¾” | DN20 1″ | DN25 1 ½” | DN40 2″ | DN50 2 ½” | DN60 3″ | DN80 4″ | DN100 6″ | DN150 8″ | DN200
LPM | GPM 0.3m/s min.
3.5 | 1.0 5.0 | 1.5 9.0 | 2.5 25.0 | 6.5 40.0 | 10.5 60.0 | 16.0 90.0 | 24.0 125.0 | 33.0 230.0 | 60.0 315.0 | 82.0
LPM | GPM 10m/s max 120.0 | 32.0 170.0 | 45.0 300.0 | 79.0 850.0 | 225.0 1350.0 | 357.0 1850.0 | 357.0 2800.0 | 739.0 4350.0 | 1149.0 7590.0 | 1997.0 10395.0 | 2735.0
*Optional
2F4i-n05d78Q©uaIcolintPyroPcersos Cdounctrtolss LOtdn. line at:
316SS PC
PVC
PP PVDF
Valuetesters.com
info@valuetesters.com9
Truflo® — TIM | TI3M Series (V1)
Insertion Paddle Wheel Flow Meter Sensor
Programming
STEPS
1
Home Screen
+
3 Sec.
2
Lock Settings
3
Flow Unit
4
K Factor
5
Filter Damping
6
Transmitter Range
3 Sec.
7
Transmitter Span
8
Transmitter Offset
Select/Save/Continue
DISPLAY
Move Selection Left
OPERATION
Home Screen
Change Digit Value
Lock Settings Factory Default: Lk = 10 Otherwise meter will enter Lockout Mode*
Flow Unit Factory Default: Ut.1 = Gallon Ut.0 = Liter | Ut.2 = Kiloliters
K Factor Value Enter K Factor value depending on pipe size. Refer to Page 9 for K-Factor Values
Filter Damping Factory Default: FiL = 20 | Range : 0 ~ 99 Secs (Filter Damping : Smooth out or “Dampen” the response of the Flow Meter to rapid fluctuations in flow.)
Transmitter Range | 20mA Factory Default: 4mA = 0 Enter 20mA Output Value Note: 20mA = 100** (Max. Flow Rate)
Transmitter Span Factory Default: SPn = 1.000 | Range : 0.000 ~ 9.999 (Span : Difference between Upper Range (UPV) & Lower Range (LRV))
Transmitter Offset Factory Default: oSt = 0.000 | Range : 0.000 ~ 9.999 (Offset : Actual Output – Expected Output)
2F4i-n05d78Q©uaIcolintPyroPcersos Cdounctrtolss LOtdn. line at:
Valuetesters.com
info@valuetesters.com10
Truflo® — TIM | TI3M Series (V1)
Insertion Paddle Wheel Flow Meter Sensor
Totalizer Reset
STEPS
1
Home Screen
+
3 Sec.
2
Totalizer Reset
DISPLAY
Home Screen
OPERATION
Totalizer Value will Reset to Zero
Setting Output Limits (SSR*)
Select/Save/Continue
Move Selection Left
STEPS
DISPLAY
1
Home Screen
Home Screen
OPERATION
Change Digit Value
Current Value (CV) Set Value (SV)
2 Flow Rate Pulse Output (OP1) 3 Totalizer Pulse Output (OP2)
Flow Rate Pulse Output (OP1) Limit Enter Flow Rate Pulse Output Value CV SV : Flow Rate Output (OP1) ON CV < SV : Flow Rate Output (OP1) OFF
Refer Page 6 for SSR* Wiring
Totalizer Pulse Output (OP2) Limit Enter Totalizer Pulse Output Value CV SV : Totalizer Output (OP2) ON CV < SV : Totalizer Output (OP2) OFF Note: Refer Pulse Control Programming (Pg 12)
Refer Page 6 for SSR* Wiring
*SSR – Solid State Relay
2F4i-n05d78Q©uaIcolintPyroPcersos Cdounctrtolss LOtdn. line at:
Valuetesters.com
info@valuetesters.com11
Truflo® — TIM | TI3M Series (V1)
Insertion Paddle Wheel Flow Meter Sensor
Pulse Control Programming
Select/Save/Continue
Move Selection Left
Change Digit Value
STEPS
DISPLAY
1
Home Screen
3 Sec.
Home Screen
OPERATION
2
Pulse Output Control
3 OP2 Auto Reset Time Delay
4
Alarm Mode Setting
Pulse Output Control Con = n : OP2 Manual Reset (When Totalizer = Set Value (SV)) Con = c | r : OP2 Auto Reset after (t 1) Secs Con = E : One Pulse/Gal (Default) Con = F : Paddle Pulse — Frequency Max 5 KHz (For TVF)
OP2 Auto Reset Time Delay Factory Default: t 1 = 0.50 | Range : 0.000 ~ 9.999 Secs (Displayed only when Con r | Con c is selected) Note: OP2 = Totalizer Output
Alarm Mode Setting Factory Default: ALt = 0 | Range: 0 ~ 3 Refer to Alarm Mode Selection
5
Hysterisis
Hysterisis Factory Default: HYS = 1.0 | Range: 0.1 ~ 999.9 (Hysterisis is a buffer around the Programmed Set Point)
6 OP1 Power On Time Delay
OP1 Power On Time Delay Factory Default: t2 = 20 Sec | Range: 0 ~ 9999 Secs Note: OP1 = Flow Rate Output
Relay Mode Selection
ALt No.
Description
ALt = 0 CV SV — Relay ON | CV < [SV – Hys] — Relay OFF
ALt = 1 CV SV — Relay ON | CV > [SV + Hys] — Relay OFF
ALt = 2 [SV + Hys] CV [SV – Hys] — Relay ON : CV > [SV + Hys] or CV < [SV – HyS] — Relay OFF
ALt = 3 [SV + Hys] CV [SV – Hys] — Relay OFF: CV > [SV + Hys] or CV < [SV – HyS] — Relay ON
Hys = Hysteresis — Acts like a buffer ± around (OP1) pulse output
CV: Current Value (Flow Rate) | SV = Set Value
2F4i-n05d78Q©uaIcolintPyroPcersos Cdounctrtolss LOtdn. line at:
Valuetesters.com
info@valuetesters.com12
Truflo® — TIM | TI3M Series (V1)
Insertion Paddle Wheel Flow Meter Sensor
Rotor Pin | Paddle Replacement
1
Line up pin with hole
2
GENTLY tap
Small Pin
3
Tap until pin is 50% out
Pin Hole
4
Pull out
5
6
Pull out Paddle
Insert new paddle in flow meter
7
Push in pin approx. 50%
8
GENTLY tap
9
Congratulations! Replacement procedure is complete!
Ensure holes are aligned
2F4i-n05d78Q©uaIcolintPyroPcersos Cdounctrtolss LOtdn. line at:
Valuetesters.com
info@valuetesters.com13
Truflo® — TIM | TI3M Series (V1)
Insertion Paddle Wheel Flow Meter Sensor
Installation Fittings
SA
Clamp-On Saddle Fittings
· PVC Material · Viton® O-Rings · Available in Metric DIN · Will Accept Signet® Type Flow Meter
Size 2″ 3″ 4″ 6″ 8″
PVC
Part Number SA020 SA030 SA040 SA060 SA080
PT | PPT | PFT
Installation Fittings
· PVC | PP | PVDF · Socket End
Connections · Will Accept Signet® Type
Flow Meter · True-Union Design
PVDF
PVC
Size ½” ¾” 1″ 1½” 2″
Part Number PFT005 PFT007 PFT010 PFT015 PFT020
Part Number PT005 PT007 PT010 PT015 PT020
Add Suffix `E’ – EPDM Seals `T’ – NPT End Connectors `B’ – Butt Fused End Connections for PP or PVDF
PP
Part Number PPT005 PPT007 PPT010 PPT015 PPT020
SAR
Clamp-On Saddle Fittings (SDR Pipe)
· PVC Material · Viton® O-Rings · Available in Metric DIN · Will Accept Signet® Type Flow Meter
Size 2″ 3″ 4″ 6″ 8″ 10″ 12″ 14″ 16″
PVC
Part Number SAR020 SAR030 SAR040 SAR060 SAR080 SAR100 SAR120 SAR140 SAR160
CT
CPVC Tee Installation Fitting
· 1″-4″ Pipe Sizes · Easy to Install · Will Accept Signet®
Flow Meter
CPVC
Size
Part Number
1″ 1 ½”
2″ 3″ 4″
CT010 CT015 CT020 CT030 CT040
Add Suffix –
`E’ – EPDM Seals
`T’ – NPT End Connectors
`B’ – Butt Fused End Connections for PP or PVDF
PG
Glue-On Adapter
· 2″-24″ Pipe Sizes · Easy to Install · Will Accept Signet® Flow Meter
Glue-On Adapter CPVC
Size
Part Number
2″- 4″ 6″- 24″
PG4 PG24
2F4i-n05d78Q©uaIcolintPyroPcersos Cdounctrtolss LOtdn. line at:
Valuetesters.com
info@valuetesters.com14
Truflo® — TIM | TI3M Series (V1)
Insertion Paddle Wheel Flow Meter Sensor
SWOL
Weld-On Adapter
· 2″-12″ Pipe Sizes · 316SS Weld-o-let with PVDF insert · Easy to Install · Will Accept Signet® Flow Meter
Weld-On Adapter – 316 SS
Size
Part Number
3″
SWOL3
4″
SWOL4
6″
SWOL6
8″
SWOL8
10″
SWOL10
12″
SWOL12
SST
316SS TI3 Series NPT Tee Fittings
· Will Accept Signet® Type Flow Meter
Threaded Tee Fitting – 316 SS
Size
Part Number
½” ¾” 1″ 1 ½” 2″ 3″ 4″
SST005 SST007 SST010 SST015 SST020 SST030 SST040
SSS
316SS TI3 Series Sanitary Tee Fittings
· Will Accept Signet® Type Flow Meter
Sanitary Tee Fitting – 316 SS
Size
Part Number
½” ¾” 1″ 1 ½” 2″ 3″ 4″
SSS005 SSS007 SSS010 SSS015 SSS020 SSS030 SSS040
SSF
316SS TI3 Series Flanged Tee Fittings
· Will Accept Signet® Type Flow Meter
Flanged Tee Fitting – 316 SS
Size
Part Number
½” ¾” 1″ 1 ½” 2″ 3″ 4″
SSF005 SSF007 SSF010 SSF015 SSF020 SSF030 SSF040
2F4i-n05d78Q©uaIcolintPyroPcersos Cdounctrtolss LOtdn. line at:
Valuetesters.com
info@valuetesters.com15
Truflo® — TIM | TI3M Series (V1)
Insertion Paddle Wheel Flow Meter Sensor
Warranty, Returns and Limitations
Warranty
Icon Process Controls Ltd warrants to the original purchaser of its products that such products will be free from defects in material and workmanship under normal use and service in accordance with instructions furnished by Icon Process Controls Ltd for a period of one year from the date of sale of such products. Icon Process Controls Ltd obligation under this warranty is solely and exclusively limited to the repair or replacement, at Icon Process Controls Ltd option, of the products or components, which Icon Process Controls Ltd examination determines to its satisfaction to be defective in material or workmanship within the warranty period. Icon Process Controls Ltd must be notified pursuant to the instructions below of any claim under this warranty within thirty (30) days of any claimed lack of conformity of the product. Any product repaired under this warranty will be warranted only for the remainder of the original warranty period. Any product provided as a replacement under this warranty will be warranted for the one year from the date of replacement.
Returns
Products cannot be returned to Icon Process Controls Ltd without prior authorization. To return a product that is thought to be defective submit a customer return (MRA) request form and follow the instructions therein. All warranty and non-warranty product returns to Icon Process Controls Ltd must be shipped prepaid and insured. Icon Process Controls Ltd will not be responsible for any products lost or damaged in shipment.
Limitations
This warranty does not apply to products which: 1. are beyond the warranty period or are products for which the original purchaser does not follow the warranty procedures
outlined above; 2. have been subjected to electrical, mechanical or chemical damage due to improper, accidental or negligent use; 3. have been modified or altered; 4. anyone other than service personnel authorized by Icon Process Controls Ltd have attempted to repair; 5. have been involved in accidents or natural disasters; or 6. are damaged during return shipment to Icon Process Controls Ltd
Icon Process Controls Ltd reserves the right to unilaterally waive this warranty and dispose of any product returned to Icon Process Controls Ltd where: 1. there is evidence of a potentially hazardous material present with the product; 2. or the product has remained unclaimed at Icon Process Controls Ltd for more than 30 days after Icon Process Controls Ltd
has dutifully requested disposition.
This warranty contains the sole express warranty made by Icon Process Controls Ltd in connection with its products. ALL IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, THE WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSLY DISCLAIMED. The remedies of repair or replacement as stated above are the exclusive remedies for the breach of this warranty. IN NO EVENT SHALL Icon Process Controls Ltd BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES OF ANY KIND INCLUDING PERSONAL OR REAL PROPERTY OR FOR INJURY TO ANY PERSON. THIS WARRANTY CONSTITUTES THE FINAL, COMPLETE AND EXCLUSIVE STATEMENT OF WARRANTY TERMS AND NO PERSON IS AUTHORIZED TO MAKE ANY OTHER WARRANTIES OR REPRESENTATIONS ON BEHALF OF Icon Process Controls Ltd. This warranty will be interpreted pursuant to the laws of the province of Ontario, Canada.
If any portion of this warranty is held to be invalid or unenforceable for any reason, such finding will not invalidate any other provision of this warranty.
by
2F4i-n05d78Q©uaIcolintPyroPcersos Cdounctrtolss LOtdn. line at:
Valuetesters.com
info@valuetesters.com16
Documents / Resources
![]() |
ICON Process Controls TIM Series Insertion Paddle Wheel Flow Meter Sensor [pdf] Instruction Manual TIM, TI3M, TIM Series Insertion Paddle Wheel Flow Meter Sensor, TIM Series, Insertion Paddle Wheel Flow Meter Sensor, Paddle Wheel Flow Meter Sensor, Wheel Flow Meter Sensor, Flow Meter Sensor, Meter Sensor |