Automated Thermal Block

Model DUP-1000
User Guide

hygiena DUP-1000 Automated Thermal Block

© Copyright 2022 Hygiena® or its affiliates. All rights reserved worldwide. No part of this publication may be reproduced or transmitted in any form, or by any information storage and retrieval system, without the prior agreement and written permission of Hygiena.

Information in this document is subject to change without notice. Hygiena assumes no responsibility for any errors that may appear in this document. This document is believed to be complete and accurate at the time of publication. In no event shall Hygiena be liable for incidental, special, multiple or consequential damages in connection with or arising from the use of this document.

Part Number 2C-075.1-0314

TRADEMARKS

The Hygiena logo and BAX® are registered trademarks of Hygiena or its affiliates.

All other trademarks are the property of their holders.

FIELD OF USE

The Automated Thermal Block is manufactured by Torrey Pines Scientific for use with the BAX System. Please see BAX System documentation for details on Field of Use. Please read the Limitation of Warranty and Liability before using the product.

WARRANTY

This product is warranted by Torrey Pines Scientific to be free from defects in material and workmanship for a period of one year from the date of purchase. If repair or adjustment is necessary and has not been the result of abuse or misuse within the one-year period, correction of the defect will be made without charge.

For warranty and repair issues, contact Torrey Pines Scientific at 760-930-9400.

TECHNICAL SUPPORT

If you have any questions or comments on the Automated Thermal Block, please contact your distributor or Hygiena at 800-863-6842 or email diagnostics.support@Hygiena.com for technical assistance.

User Guide

Instrument Overview

The Automated Thermal Block is a Peltier-driven heating and cooling device pre-programmed to perform the lysis steps of the BAX System protocols. It has been designed as a solid-state device with minimal moving parts to help ensure quality performance over time. The temperature of the plate is sensed by a platinum resistance temperature device mounted under the plate. The computer in the unit compares the plate temperature with the target temperature and instructs the module to heat or chill the plate as required.

The block automatically provides the sequential heating and cooling conditions for lysis protocols, eliminating the need to transfer samples between separate heating and cooling blocks. Once the lysis program is complete, the block holds the samples at 4 °C until they are removed.

This model has been optimized for use with the cluster tubes and metal tube holder supplied with the BAX System. Other tubes and holders are not recommended.

DUP-1000 Automated Thermal Block 2

Supplied Components

The purchase of the thermal block includes the following components:

  1. Thermal block unit
  2. Metal tube holders (2)
  3. Power supply and cord
  4. User documentation

Automated Thermal Block

Specifications and Requirements
Item Specification
Dimensions (with lid) 6.5″ W x 8.75″ D x 8″ H (16.5 x 22.25 x 20.3 cm)
Weight Approximately 8 lbs. (3.6 kg)
Power usage 12 volts, 8.4 amps
Power requirements 90 to 265 volts AC, 50/60Hz
Thermal range -10 °C to 110 °C (14 to 230 °F)
Thermal uniformity ±1 °C as measured at the block surface
Thermal accuracy ±1 °C as measured at the block surface
Sample throughput 1-96 samples per cycle
Room environment Temperature between 65 and 89 °F (18 and 32 °C)
Safety Symbols and Precautions

Safety Symbols

Symbol          Description


hygiena DUP-1000 - Safety Symbol 1               Indicates that you should consult the manual for further information and proceed with appropriate caution.


hygiena DUP-1000 - Safety Symbol 2               Indicates the presence of a hot surface or other high-temperature hazard and proceed with appropriate caution.


hygiena DUP-1000 - Safety Symbol 3               Indicates the presence of an electrical shock hazard and proceed with appropriate caution.



General Safety Information

Extreme Temperatures

hygiena DUP-1000 - Safety Symbol 2 The thermal block is capable of reaching 110 ºC (230 ºF), which can burn the skin if touched.
To prevent injury, remember the following:

  • Use extreme caution around the block at all times.
  • Never leave the block accessible to others when it is hot.
  • Do not touch the plate surface unless you are sure it has cooled.
  • Always keep the lid in place when the block is not in use.

Power Requirements

hygiena DUP-1000 - Safety Symbol 3 Always use the supplied AC power cord when connecting the heating block to a power source. Use the normal care and precaution that one would use with any electrical appliance.


Connecting the Thermal Block

The Automated Thermal Block is pre-programmed to perform the sequential heating and cooling steps needed for BAX System assays. No calibration or software installation is required when the thermal block is used as specified.

Follow these steps to connect the thermal block:

1. Place the thermal block unit on a level, dry laboratory workbench. Allow at least 4 inches (10 cm) on all sides of the unit for ventilation.

2. Insert one end of the provided power cord into the power supply and plug the other end of the power cord into a 3-wire outlet.

Note: Ensure the outlet is properly grounded and runs the appropriate voltage (see specifications on the previous page).

3. Insert the power supply jack into the power input port on the back panel of the thermal block.

DUP-1000 - Connecting the Thermal Block 1

  1. On/Off switch
  2. Power input
  3. IO port

4. Seat the metal tube holder over the metal plate surface on the top of the thermal block unit.

DUP-1000 - Connecting the Thermal Block 2a

  1. Metal tube holder
  2. Metal plate surface
  3. Thermal block unit
Front Panel Display

The front panel of the Automated Thermal Block contains the following items:

  1. LCD display – shows the program menu, program selection, settings and instructions.
  2. SCROLL button – changes the selected menu item in the LCD display. An audible beep sounds when the button is pressed.
  3. SELECT/CONTINUE button – sets the menu item in the LCD display as the selected program and is also used to confirm required actions. An audible beep sounds when the button is pressed.
  4. Blue ACTION NEEDED LED – indicates that the user must perform an action before the program can continue. This notification is accompanied by an audible alarm.
  5. Green POWER LED – indicates that the thermal block is turned on.
DUP-1000 - Front Panel DisplayProgram Menu

The Automated Thermal Block is pre-programmed with four lysis protocols:

Gram positive – use with Gram-positive bacteria, including Staphylococcus and Listeria*
*Note: For Listeria 24E assays, use the “24E” program. For real-time Listeria assays, use the “RT Listeria” program.

Gram negative – use with Gram-negative bacteria, including Salmonella, E. coli, Campylobacter, Vibrio and Cronobacter

24E – use with BAX System 24E assays for Listeria and L. monocytogenes

RT Listeria – use with BAX System real-time assays for Listeria and L. monocytogenes

Turning the Thermal Block On and Off

ON

Turn on the thermal block by toggling the on/off switch on the back panel of the unit.

At start-up, the LCD automatically displays the following series of messages:

DUP-1000 - Connecting the Thermal Block 1

  1. On/Off switch
  2. Power input
  3. IO port

The first screen shows the version of software currently installed on the thermal block.

DUP-1000 - Turning the Thermal Block On and Off 1


The second screen shows the seq file version number and the checksum number for validation purposes.

Note: A change in either of these values indicates a modification to the thermal block software. If this occurs, any validation previously performed on the instrument may no longer apply.

DUP-1000 - Turning the Thermal Block On and Off 2


The start-up sequence is complete when the LCD display shows the program menu.

DUP-1000 - Turning the Thermal Block On and Off 3


OFF

After finishing a run, you can either select another program or turn off the power to the unit by toggling the on/off switch on the back panel of the unit.

DUP-1000 - Connecting the Thermal Block 1

  1. On/Off switch
  2. Power input
  3. IO port

Running the Gram-Positive Program

Samples are heated to 55 °C for 60 minutes, then 95 °C for 10 minutes, then held at 4 °C for at least 5 minutes or until removed.


1 Turn on the thermal block, then press the SCROLL button until the arrow DUP-1000 - Arrow is next to the Gram Positive option.

Press the SELECT/CONTINUE button.

DUP-1000 - Gram-Positive Program 1


2 The unit beeps once and the LCD display changes as the thermal block begins equilibrating to 55 °C.

Note: Do not load samples into the thermal block until the load prompt appears on the LCD display.

DUP-1000 - Gram-Positive Program 2


3 When the thermal block has reached 55 °C, the unit sounds two audible beeps and the blue ACTION NEEDED LED activates. The LCD display changes to “Load Samples”.

DUP-1000 - Gram-Positive Program 3


4 After the “Load Samples” prompt appears, place a rack of prepared samples into the metal tube holder.

Note: See the BAX System User Guide for information on preparing samples.

Press the SELECT/CONTINUE button.

DUP-1000 - Gram-Positive Program 4


5 The LCD display changes to “GmPos Lysis 55C”. A timer in the bottom right corner counts down from 60:00 and shows the number of minutes remaining in this step.

DUP-1000 - Gram-Positive Program 5


6 After lysis at 55 °C is complete, the LCD display changes to “Heating to 95C”. A timer in the bottom right corner counts up from 0:00 during the temperature change. No user action is required.

DUP-1000 - Gram-Positive Program 6


7 After the thermal block reaches 95 °C, the LCD display changes to “Denaturing @ 95C”. A timer in the bottom right corner counts down from 10:00 and shows the number of minutes remaining in this step. No user action is required.

DUP-1000 - Gram-Positive Program 7


8 After denaturation at 95 °C is complete, the LCD display changes to “Cooling to 4C”, and the thermal block sounds two audible beeps to signal the change. A timer in the bottom right corner counts up from 0:00 during the temperature change. No user action is required.

DUP-1000 - Gram-Positive Program 8


9 After the samples have cooled for 5 minutes, the LCD display changes to “Sample PCR Ready”, the blue ACTION NEEDED LED activates and the thermal block sounds four audible beeps to signal that samples can be removed.

Press the SELECT/CONTINUE button.

DUP-1000 - Gram-Positive Program 9

Note: A timer in the bottom right corner counts up from 0:00 to show any additional time that samples are held at 4 °C.


10 The LCD display changes to “Completed” and the thermal block sounds an audible beep to signal that the program has been completed. The timer continues to show any additional time that samples are held at 4 °C.

DUP-1000 - Gram-Positive Program 10


11 Follow the instructions in the BAX System User Guide to hydrate the PCR tablets with these lysates.

Note: Lysates must be maintained at 4 °C during tablet hydration.


12 After the PCR tablets have been hydrated, remove the rack of cluster tubes from the metal tube holder.

Press the SELECT/CONTINUE button.

DUP-1000 - Gram-Positive Program 11


13 The LCD display changes to “Gram Positive Finished” and the thermal block sounds an audible beep to signal that the program has finished.

DUP-1000 - Gram-Positive Program 12


14 Press the SELECT/CONTINUE button to exit from the program. The LCD changes to display the program menu.

DUP-1000 - Gram-Positive Program 13


Running the Gram-Negative Program

Samples are heated to 37 °C for 20 minutes, then 95 °C for 10 minutes, then held at 4 °C for at least 5 minutes or until removed.


1 Turn on the thermal block, then press the SCROLL button until the arrow DUP-1000 - Arrow is next to the Gram Negative option.

Press the SELECT/CONTINUE button.

DUP-1000 - Gram-Negative Program 1


2 The unit beeps once and the LCD display changes as the thermal block begins equilibrating to 37 °C.

Note: Do not load samples into the thermal block until the load prompt appears on the LCD display.

DUP-1000 - Gram-Negative Program 2


3 When the thermal block has reached 37 °C, the unit sounds two audible beeps and the blue ACTION NEEDED LED activates. The LCD display changes to “Load Samples”.

DUP-1000 - Gram-Negative Program 3


4 After the “Load Samples” prompt appears, place a rack of prepared samples into the metal tube holder.

Note: See BAX System User Guide for information on preparing samples.

Press the SELECT/CONTINUE button.

DUP-1000 - Gram-Negative Program 4


5 The LCD display changes to “GmNeg Lysis 37C”. A timer in the bottom right corner counts down from 20:00 and shows the number of minutes remaining in this step.

DUP-1000 - Gram-Negative Program 5


6 After lysis at 37 °C is complete, the LCD display changes to “Heating to 95C”. A timer in the bottom right corner counts up from 0:00 during the temperature change. No user action is required.

DUP-1000 - Gram-Negative Program 6


7 After the thermal block reaches 95 °C, the LCD display changes to “Denaturing @ 95C”. A timer in the bottom right corner counts down from 10:00 and shows the number of minutes remaining in this step. No user action is required.

DUP-1000 - Gram-Negative Program 7


8 After denaturation at 95 °C is complete, the LCD display changes to “Cooling to 4C” and the thermal block sounds two audible beeps to signal the change. A timer in the bottom right corner counts up from 0:00 during the temperature change. No user action is required.

DUP-1000 - Gram-Negative Program 8


9 After samples have cooled for 5 minutes, the LCD display changes to “Sample PCR Ready”, the blue ACTION NEEDED LED activates and the thermal block sounds four audible beeps to signal that samples can be removed.

Press the SELECT/CONTINUE button.

DUP-1000 - Gram-Negative Program 9

Note: A timer in the bottom right corner counts up from 0:00 to show any additional time that samples are held at 4 °C.


10 The LCD display changes to “Completed” and the thermal block sounds an audible beep to signal that the program has been completed. The timer continues to show any additional time that samples are held at 4 °C.

DUP-1000 - Gram-Negative Program 10


11 Follow the instructions in the BAX System User Guide to hydrate the PCR tablets with these lysates.

Note: Lysates must be maintained at 4 °C during tablet hydration.


12 After the PCR tablets have been hydrated, remove the rack of cluster tubes from the metal tube holder.

Press the SELECT/CONTINUE button.

DUP-1000 - Gram-Negative Program 12


13 The LCD display changes to “Gram Negative Finished” and the thermal block sounds an audible beep to signal that the program has finished.

DUP-1000 - Gram-Negative Program 13


14 Press the SELECT/CONTINUE button to end the program. The LCD changes to display the program menu.

DUP-1000 - Gram-Negative Program 14


Running the 24E Program

Samples are heated at 37 °C for 30 minutes, then removed for the addition of lysis reagents.
Samples then are heated to 55 °C for 30 minutes, then 95 °C for 10 minutes, then held at 4 °C for at least 5 minutes or until removed.


1 Turn on the thermal block, then press the SCROLL button until the arrow DUP-1000 - Arrow is next to the 24E option.

Press the SELECT/CONTINUE button.

DUP-1000 - 24E Program 1


2 The unit beeps once and the LCD display changes as the thermal block begins equilibrating to 37 °C.

Note: Do not load samples into the thermal block until the load prompt appears on the LCD display.

DUP-1000 - 24E Program 2


3 When the thermal block has reached 37 °C, the unit sounds two audible beeps and the blue ACTION NEEDED LED activates. The LCD display changes to “Load Samples”.

DUP-1000 - 24E Program 3


4 After the “Load Samples” prompt appears, place a rack of prepared samples into the metal tube holder.

Note: See BAX System User Guide for more information on prepared samples.

Press the SELECT/CONTINUE button.

DUP-1000 - 24E Program 4


5 The LCD display changes to “37C 24E Lysis”. A timer in the bottom right corner counts down from 30:00 and shows the number of minutes remaining in this step.

DUP-1000 - 24E Program 5


6 After lysis at 37 °C is complete, the blue ACTION NEEDED LED activates and the thermal block sounds two audible beeps to signal that user action is required. The LCD display changes to “Remove & Press Continue”.

DUP-1000 - 24E Program 6


7 Remove the rack of cluster tubes from the metal tube holder.

Press the SELECT/CONTINUE button.

DUP-1000 - 24E Program 7


8 The unit beeps once and the LCD display changes as the thermal block begins equilibrating to 55 °C.

While the block is heating, follow the instructions in the BAX System User Guide to add additional lysing reagents to each sample.

DUP-1000 - 24E Program 8

Note: Do not re-load samples into the thermal block until the re-load prompt appears on the LCD display.


9 When the thermal block has reached 55 °C, the unit sounds two audible beeps and the blue ACTION NEEDED LED activates. The LCD display changes to “Reload Samples”.

DUP-1000 - 24E Program 9


10 After the “Reload Samples” prompt appears, place the rack of samples with additional lysing reagents into the metal tube holder.

Press the SELECT/CONTINUE button.

DUP-1000 - 24E Program 10


11 The LCD display changes to “55C 24E Lysis”. A timer in the bottom right corner counts down from 30:00 and shows the number of minutes remaining in this step.

DUP-1000 - 24E Program 11


12 After lysis at 55 °C is complete, the LCD display changes to “Heating to 95C”. A timer in the bottom right corner counts up from 0:00 during the temperature change. No user action is required.

DUP-1000 - 24E Program 12


13 After the thermal block reaches 95 °C, the LCD display changes to “Denaturing @ 95C”. A timer in the bottom right corner counts down from 10:00 and shows the number of minutes remaining in this step. No user action is required.

DUP-1000 - 24E Program 13


14 After denaturation at 95 °C is complete, the LCD display changes to “Cooling to 4C” and the thermal block sounds two audible beeps to signal the change. A timer in the bottom right corner counts up from 0:00 during the temperature change. No user action is required.

DUP-1000 - 24E Program 14


15 After samples have cooled for 5 minutes, the LCD display changes to “Sample PCR Ready”, the blue ACTION NEEDED LED activates and the thermal block sounds four audible beeps to signal that samples can be removed.

Press the SELECT/CONTINUE button.

DUP-1000 - 24E Program 15

Note: A timer in the bottom right corner counts up from 0:00 to show any additional time that samples are held at 4 °C.


16 The LCD display changes to “Completed” and the thermal block sounds an audible beep to signal that the program has been completed. The timer continues to show any additional time that samples are held at 4 °C.

DUP-1000 - 24E Program 16


17 Follow the instructions in the BAX System User Guide to hydrate the PCR tablets with these lysates.

Note: Lysates must be maintained at 4 °C during tablet hydration.


18 After the PCR tablets have been hydrated, remove the rack of cluster tubes from the metal tube holder.

Press the SELECT/CONTINUE button.

DUP-1000 - 24E Program 17


19 The LCD display changes to “24E Finished” and the thermal block sounds an audible beep to signal that the program has finished.

DUP-1000 - 24E Program 19


20 Press the SELECT/CONTINUE button to end the program. The LCD changes to display the program menu.

DUP-1000 - 24E Program 20


Running the RT Listeria Program

Samples are heated to 55 °C for 30 minutes, then 95 °C for 10 minutes, then held at 4 °C for at least 5 minutes or until removed.


1 Turn on the thermal block, then press the SCROLL button until the arrow DUP-1000 - Arrow is next to the RT Listeria option.

Press the SELECT/CONTINUE button.

DUP-1000 - RT Listeria Program 1


2 The unit beeps once and the LCD display changes as the thermal block begins equilibrating to 55 °C.

Note: Do not load samples into the thermal block until the load prompt appears on the LCD display.

DUP-1000 - RT Listeria Program 2


3 When the thermal block has reached 55 °C, the unit sounds two audible beeps and the blue ACTION NEEDED LED activates. The LCD display changes to “Load Samples”.

DUP-1000 - RT Listeria Program 3


4 After the “Load Samples” prompt appears, place a rack of prepared samples into the metal tube holder.

Note: See the BAX System User Guide for information on preparing samples.

Press the SELECT/CONTINUE button.

DUP-1000 - RT Listeria Program 4


5 The LCD display changes to “RTLis Lysis 55C”. A timer in the bottom right corner counts down from 30:00 and shows the number of minutes remaining in this step.

DUP-1000 - RT Listeria Program 5


6 After lysis at 55 °C is complete, the LCD display changes to “Heating to 95C”. A timer in the bottom right corner counts up from 0:00 during the temperature change. No user action is required.

DUP-1000 - RT Listeria Program 6


7 After the thermal block reaches 95 °C, the LCD display changes to “Denaturing @ 95C”. A timer in the bottom right corner counts down from 10:00 and shows the number of minutes remaining in this step. No user action is required.

DUP-1000 - RT Listeria Program 7


8 After denaturation at 95 °C is complete, the LCD display changes to “Cooling to 4C” and the thermal block sounds two audible beeps to signal the change. A timer in the bottom right corner counts up from 0:00 during the temperature change. No user action is required.

DUP-1000 - RT Listeria Program 8


9 After the samples have cooled for 5 minutes, the LCD display changes to “Sample PCR Ready”, the blue ACTION NEEDED LED activates and the thermal block sounds four audible beeps to signal that samples can be removed.

Press the SELECT/CONTINUE button.

DUP-1000 - RT Listeria Program 9

Note: A timer in the bottom right corner counts up from 0:00 to show any additional time that samples are held at 4 °C.


10 The LCD display changes to “Completed” and the thermal block sounds an audible beep to signal that the program has been completed. The timer continues to show any additional time that samples are held at 4 °C.

DUP-1000 - RT Listeria Program 10


11 Follow the instructions in the BAX System User Guide to hydrate the PCR tablets with these lysates.

Note: Lysates must be maintained at 4 °C during tablet hydration.


12 After the PCR tablets have been hydrated, remove the rack of cluster tubes from the metal tube holder.

Press the SELECT/CONTINUE button.

DUP-1000 - RT Listeria Program 12


13 The LCD display changes to “RT Listeria Finished” and the thermal block sounds an audible beep to signal that the program has finished.

DUP-1000 - RT Listeria Program 13


14 Press the SELECT/CONTINUE button to exit from the program. The LCD changes to display the program menu.

DUP-1000 - RT Listeria Program 14


Cleaning and Decontaminating the Thermal Block
Cleaning the Thermal Block
  • Do not attempt to clean the thermal block or metal tube holder when they are hot.
  • Remove dust and debris by wiping the instrument surfaces with a lint-free cloth.
  • Wipe all spills or condensation with a soft cloth or paper towel, if needed.
  • If needed, the casing can be wiped off with a damp cloth using mild soap or detergent.
  • Do not use solvents or cleansers containing iodine or acetone, as these solutions could
    damage the paint or display window of the block.
Decontaminating the Thermal Block

If contamination should occur, the following steps can be performed under sterile conditions to help remove the contaminants.

  • Wipe the surface of the thermal block unit with a cloth dampened with 20% bleach solution*, followed by a 70% ethanol rinse to prevent damage to the equipment.
  • Soak metal tube holders in 20% bleach solution* for about 5 minutes, then rinse thoroughly with water and allow to air dry.

* An alternate cleaner designed to remove free DNA may also be investigated. A 10% bleach solution may also be used for decontamination in accordance with your laboratory SOP but may require repetition of the cleaning to ensure the removal of all amplicons.

Calibrating the Thermal Block

The Automated Thermal Block should not require calibration when used as directed. If your laboratory requires calibration as part of your standard operating procedures, you can use the following procedure to calibrate the block to the nearest degree.

Materials

  • Resistance Temperature Detector (RTD) thermometer with microprobe – VWR #61220-601 or equivalent
  • Cluster tubes used for lysis (at least 48)

Note: If an RTD thermometer is not available, a dry bath alcohol thermometer (VWR #89095-782 or equivalent) may be used to perform the calibration. However, because results may be less accurate due to the lower resolution of the dry bath thermometer, calibration with an RTD thermometer is strongly recommended.

The buttons on the front panel display will serve different purposes during calibration:

  • SCROLL button – increases the value entered in the LCD display.
  • SELECT/CONTINUE button – decreases the value entered in the LCD display.
  • Pressing SCROLL and SELECT/CONTINUE buttons at the same time – sets the value in the LCD display as the selected value.

1 Arrange 48 cluster tubes in a rack and fill each tube with 200 µL of water.

DUP-1000 - Calibrating the Thermal Block 1


2 Poke a small hole in the cap of one cluster tube near the center of the cluster tube rack.

DUP-1000 - Calibrating the Thermal Block 2


3 With the power turned off, place the rack of cluster tubes in the metal holder.

DUP-1000 - Calibrating the Thermal Block 3


4 Insert the microprobe of the RTD thermometer through the hole in the cap of the cluster tube.

Make sure the tip of the probe touches the bottom of the tube to ensure that the probe is completely immersed.

DUP-1000 - Calibrating the Thermal Block 4


5 While pressing and holding the SELECT/CONTINUE button, turn on the power to the thermal block by toggling the on/off switch on the back panel. The LCD displays a message that “Default Settings Are Now Loaded”.

DUP-1000 - Calibrating the Thermal Block 5


6 Continue to hold the SELECT/CONTINUE button until the LCD displays “Set Point: off” (about 15 seconds).

Note: Plate temp is not important at this point.

DUP-1000 - Calibrating the Thermal Block 6


7 Press the SCROLL and SELECT/CONTINUE buttons at the same time. The LCD displays the Calibration menu.

DUP-1000 - Calibrating the Thermal Block 7


8 Press the SELECT/CONTINUE button until the arrow DUP-1000 - Arrow is next to the Clear Cal Pts option.

DUP-1000 - Calibrating the Thermal Block 8


9 Press the SCROLL and SELECT/CONTINUE push buttons at the same time to select the Clear Cal Pts option.

Calibration points previously saved in the thermal block are cleared, and the LCD display changes to read “Set Point: off”.

DUP-1000 - Calibrating the Thermal Block 9


10 Press the SCROLL button to increase the value of the set point to 37 °C. The block automatically begins the heating process.

If needed, use the SELECT/CONTINUE button to decrease the value of the set point.

DUP-1000 - Calibrating the Thermal Block 10

Note: Be sure to wait for the temperature reading of the thermometer to equilibrate before proceeding.


11 Press the SCROLL and SELECT/CONTINUE buttons at the same time. The calibration menu again appears on the LCD display.

DUP-1000 - Calibrating the Thermal Block 11


12 Press the SELECT/CONTINUE button until the arrow DUP-1000 - Arrow is next to the Calibrate option.

Press the SCROLL and SELECT/CONTINUE buttons at the same time to select the Calibrate option.

DUP-1000 - Calibrating the Thermal Block 12


13 Press the SELECT/CONTINUE push button until the arrow DUP-1000 - Arrow is next to the Lo Temp Cal Pt option.

Press the SCROLL and SELECT/CONTINUE buttons at the same time to select the Lo Temp Cal Pt option.

DUP-1000 - Calibrating the Thermal Block 13


14 The LCD display changes to a screen that compares the Displayed and Measured temperatures.

If needed, use the SCROLL button to increase the value of the Measured temperature or the SELECT/CONTINUE button to decrease the value of the Measured temperature to the value displayed on the thermometer.

DUP-1000 - Calibrating the Thermal Block 14

Note: When entering the Measured temperature, do not include decimal places.


15 Press the SCROLL and SELECT/CONTINUE buttons at the same time to set the Measured temperature.

If necessary, the thermal block will automatically adjust the Plate Temperature to align with the Set Point.

DUP-1000 - Calibrating the Thermal Block 15


16 Press the SCROLL button to increase the value of the set point to 95 °C. The block automatically begins the heating process.

If needed, use the SELECT/CONTINUE button to decrease the value of the set point.

DUP-1000 - Calibrating the Thermal Block 16

Note: Be sure to wait for the temperature reading of the thermometer to equilibrate before proceeding.


17 Press the SCROLL and SELECT/CONTINUE buttons at the same time. The calibration menu again appears on the LCD display.

DUP-1000 - Calibrating the Thermal Block 17


18 Press the SELECT/CONTINUE button until the arrow DUP-1000 - Arrow is next to the Calibrate option.

Press the SCROLL and SELECT/CONTINUE buttons at the same time to select the Calibrate option.

DUP-1000 - Calibrating the Thermal Block 18


19 Press the SELECT/CONTINUE push button until the arrow DUP-1000 - Arrow is next to the Hi Temp Cal Pt option.

Press the SCROLL and SELECT buttons at the same time to select the Hi Temp Cal Pt option.

DUP-1000 - Calibrating the Thermal Block 19


20 The LCD display changes to a screen that compares the Displayed and Measured temperatures.

If needed, use the SCROLL button to increase the value of the Measured temperature or the SELECT/CONTINUE button to decrease the value of the Measured temperature to the value displayed on the thermometer.

DUP-1000 - Calibrating the Thermal Block 20

Note: When entering the Measured temperature, do not include decimal places.


21 Press the SCROLL and SELECT/CONTINUE buttons at the same time to set the Measured temperature.

If necessary, the thermal block automatically adjusts the Plate Temperature to align with the Set Point.

DUP-1000 - Calibrating the Thermal Block 21


22 Calibration is complete. Turn off power to the thermal block by toggling the on/off switch on the back panel.

DUP-1000 - Calibrating the Thermal Block 22


Tips and Troubleshooting
Aborting a Program

The Automated Thermal Block automatically progresses through each sequential step until the program is complete. If necessary, you can cancel the program prior to completion via the Abort option.


1 While the program is running, press the SCROLL button to display the Abort menu.

DUP-1000 - Aborting a Program 1


2 Press the SCROLL push button until the arrow DUP-1000 - Arrow is next to the Abort option, then press the SELECT/CONTINUE push button.

DUP-1000 - Aborting a Program 2


3 The LCD display shows a message to confirm that the program was terminated.

DUP-1000 - Aborting a Program 3


4 To start another program, press the SELECT/CONTINUE button to return to the program menu.

To turn off the unit, toggle the on/off switch on the back panel.

DUP-1000 - Aborting a Program 4


Error Messages

If the power fails or any interruption occurs during a run, the LCD displays an error message.

Affected samples in the thermal block should not be used with the BAX System. Remove and dispose of these samples according to your standard operating procedures.


The letter “F” (Fault) appears in the lower right corner of the LCD display if an interruption occurs during any step in the program.

This fault indicator appears on the LCD display until the program completes or the program is aborted by the user.

Affected samples should not be used with the BAX System. Remove and dispose of these samples according to your standard operating procedures.

DUP-1000 - Error Messages 1


The Power Failure message appears when power is interrupted or fails during a run. Toggle the on/off switch on the back panel to reset the unit.

Affected samples should not be used with the BAX System. Remove and dispose of these samples according to your standard operating procedures.

DUP-1000 - Error Messages 2


www.hygiena.com                                                            INS2051 Rev A