Operating Instructions for PCE-UX 3081 / PCE-UX 3081WQ Moisture Analyzers
Model Numbers: PCE-UX 3081, PCE-UX 3081WQ, PCE-UX 3011, PCE-UX 3011Q, PCE-UX 3011HQ, PCE-UX 3031
Manufacturer: PCE Instruments
Introduction and Intended Use
PCE-UX moisture analyzers are designed for the rapid measurement of moisture and dry matter content in solid, viscous, and fluid substances using thermogravimetric analysis. The process involves weighing a sample while it is dried by infra-red heat.
Common Features for all PCE-UX Devices:
- Short measurement time
- Gentle and even drying of samples
- High reproducibility
- Automatic or time-controlled measurement shut-off
Device Variants
PCE-UX 3081 / PCE-UX 3081WQ: Special devices for extra-large samples, suitable for industries like chipboard manufacturing. Features include:
- 2 temperature-controlled ceramic heaters
- Temperature range: 40 – 220 °C / 300 °C
- Stainless steel drying tray: 240 x 110 mm
PCE-UX 3011 / PCE-UX 3011Q / PCE-UX 3011HQ:
- Temperature-controlled infra-red heater: 250 W / 375 W
- Temperature range: 40 – 200 °C / 360 °C / 600 °C
- Stainless steel drying tray: 111 mm diameter
PCE-UX 3031: Special device for the gypsum industry. Features:
- 2 temperature-controlled heaters for separate measurement of free and chemically bound moisture in gypsum products
- Temperature range: 40 – 360 °C
- Stainless steel drying tray: 111 mm diameter
Safety Instructions
- Always transport the device in an upright position and protect it from impact.
- Always operate the device on a level surface, using a vibration-free base.
- Avoid draughts.
- Never touch the (luminous) heater when it is hot, as there is a risk of injury! ❗
- Always disconnect the plug from the mains supply before performing any work on the device (e.g., changing the heater). Repairs should only be carried out by a qualified technician.
- ❗ Please take extra care when using flammable samples! You may need to contact the manufacturer for advice.
- ❗ Please be careful not to drop any sample particles next to the tray support plate. Carefully remove any particles with a vacuum cleaner.
- Only use accessories specifically designed for use with PCE-UX devices.
- Do not move the device during measurement, as this will result in measuring errors.
- Do not leave any substances in the measuring tray if they are not being measured.
- Devices should only be shipped in the original packaging using the original packaging material.
- PCE-UX moisture analyzers are stand-alone units and are not designed to be used in continuous production runs. Any such use is at your own risk.
Warranty, Transport, and Installation
Warranty
PCE Instruments undertakes to repair faulty devices within the warranty period, free-of-charge, if the defect is a manufacturing fault. This warranty covers the repair of spare parts and working hours required. It does not cover normal wear-and-tear, soiling, or staining of the device. PCE Instruments will not accept any freight costs.
This warranty does not apply to:
- Improper use of the device.
- Use of the device for purposes other than those stated by PCE Instruments.
- Mechanical damage or damage caused by fluids or sample matter.
- Incorrect setup or wrong electrical wiring.
- Mechanical damage of the load cell, e.g., as a result of being overloaded.
Transport, Packaging, Installation Site
- Please check the packaging and device for any damage upon delivery.
- Please retain the original packaging in case you need to return the goods to PCE Instruments.
- Please only use the original packaging when returning the goods.
- Place the device on a stable, vibration-free surface, as level as possible and protected from draughts. The most suitable locations are the corners of a room, rooms with only one entrance, and heavy tables on concrete floors.
Start-up Procedure
- Place the device on a stable, vibration-free surface, as level as possible and protected from draughts. Remove the protective film from the front foil.
- Swing the heater back until resistance is felt.
- Place the tray holder on the device: First place the tray holder carefully on the two pins, ensuring the distance between the housing and the tray holder is approx. 1.5 – 2 mm. Do not exert excessive force on the balance. Then put on the shielding plate unit. To remove the tray holder, first carefully lift up the shielding plates and then carefully pull out the tray holder in the middle with very slight tilting movements.
- Next, place one of the two drying trays supplied on the tray holder using one or two sets of tongs. The notch in the reflector tray acts as an access gap for easy positioning of the tray.
- Check that the voltage indicated on the type plate matches the local mains voltage.
- Insert the power cable into the device's power jack and connect to a suitable earthed mains supply. After the power supply has been connected, the device can be switched on or off using the on/off button on the back of the device. Once the device is switched on, the current software version is shown, followed by the weight display.
- To prevent weighing system measuring errors, the device should be switched on 20 minutes before use to allow for heat compensation. If weight values vary in the display's milligram range during this period, this is considered normal. If the device is switched off using the reset (on/off) button, it does not require a warm-up period if switched on again with the same button.
- PCE-UX moisture measuring devices are supplied ready for immediate use in most cases. If you are not satisfied with the result, the measurement parameters can be adjusted.
Printer Setup
- Open Cover: Hold the green glowing button and lift the plastic cover.
- Change Paper: Follow the procedure for loading new paper.
- Line feed: Press the line feed button for paper advance.
The PCE-UX Thermo paper is stable for 10 years.
Operation
Display and Buttons
The display shows:
temp. °C: Current or set temperature.time: Remaining or elapsed time.mode: Current operating mode (e.g., measurement, menu).value (weight): Measured weight or value.unit: Unit of measurement (e.g., g, %, mg).
Buttons:
T(Tare): Button for setting the weight readout back to zero.Reset(rooesfnfet): Button for resetting the processor program.esc: Button for exiting menu levels.up/down: Buttons for accessing and scrolling through the menu.set: Button for confirming entries.P1: Percentage purity (only for PCE-UX 3031).P2: Program memory (optional).P3: Residue on ignition (only for PCE-UX 3011G).
Ports
- Power supply connection
- Print connection for separate printer
- Interface RS 232
- USB-Interface
- Power supply switch
Menu Settings
The device's Preferences menu can be accessed using the 'up' or 'down' buttons. The menu is operated using a standard structure:
UPorDOWN: Navigate menu items.SET: Confirm menu level selection and, if required, return to the previous menu level.ESC: Return to the previous menu level and exit the menu.
If you confirm a wrong entry with 'set', press 'esc' and then 'set' again to reselect and change the menu item.
Temperature Control
The temperature set on the device relates to the thermal radiation that heats the sample. This energy causes moisture to evaporate. The set temperature dictates the energy level for evaporation. Note: The set temperature is not the material or ambient temperature, which is a consequence of the drying process.
Automatic Shut-off or Timer
Automatic Mode: The device dries the sample until a constant weight is reached and then switches off the heater.
Configurable settings:
- Weight loss (mg): Define a threshold for completion (e.g., 10 mg for a 10 g sample, corresponding to approx. 0.1%). Measurement completes when weight loss is less than or equal to this value within the defined interval, provided the start-of-query time has expired.
- Query interval (secs): Select an interval for checking weight (None / 10-99 secs). A period of 20 to 30 seconds is typically adequate.
- Start of query (mins): Specify the minimum drying period before the automatic shut-off check begins (factory setting: 3 minutes). This value may need to be increased for very low temperatures or specific materials (e.g., free moisture in gypsum at 45°C). Range: 1-99 mins.
Timer Setting: Allows selection of a specific drying time (1 – 180 mins).
Measurement Display Options
During measurement, the device can optionally display:
- % moisture
- % dry matter (TS)
- % moisture Atro (moisture related to dry matter)
- Dry matter (TS) in g/1000g
- Weight loss g (V)
- Residual weight g (R) (current weight)
Atro Calculation Example:
- 50 % moisture = 100 % Atro
- 25 % moisture = 33.3 % Atro
- 75 % moisture = 300 % Atro
Printer Options
For devices without a printer, set the number of prints to '1'.
- Number of prints: Selectable from 1 to 9.
- Interval printout: Selectable for no printout or every 10 – 99 seconds.
- Data output for external printer: Option to send an additional Line Feed (LF) via the interface.
Device ID Number
Enter a number up to 9999 to identify the device. Each digit can be selected separately.
Date and Time, Buzzer
Date and Time
Set the date (Day, Month, Year) and time (Hours, Minutes) using the 'up'/'down' and 'set' buttons. The date and time are shown on the printout.
Buzzer
Use 'up'/'down' and 'set' to activate or deactivate the buzzer, then press 'set' followed by 'esc'.
Calibrating the Scales
Follow the menu instructions and use a calibration weight (e.g., F1 weight) to calibrate the scales. Ensure the tray carrier and tray are on the balance and the tray is empty. If the display shows 'ERR', switch the power off and on, and try again. Refer to the "Potential Errors" section for reasons.
Saving Drying Parameters in Program Memory
If your device has program memory, 'QP' will appear in the display, followed by a number (1-9). The standard parameters are factory-set for program memory place 1. Different drying parameters (temperature, automatic/timer settings) can be saved or changed using the P2 button. Note that changes to unit of measurement, printer settings, data output, device ID, date/time, buzzer, and language affect all program memories simultaneously.
Moisture Analysis Process
- Connect the device to the mains supply and switch it on with a drying tray in place. A weight value will be displayed.
- Set the scales to zero using the 'tare' button, displaying '0.000 g'.
- Evenly distribute approximately 50 - 200 g of sample material in the tray and wait for the readout to stabilize.
- Swing the heater over the sample, switch on the heater switch, and the drying phase begins. Header data is printed for devices with a built-in printer. The sample is dried until it reaches a constant weight; the heater then switches off automatically. The time and measurement value are printed for devices with a printer. If you do not wish to make any measurements, always ensure the heater is swung back to its rearmost position.
LED Indicators:
Red glow: Heater is on.Green flashes: Measurement is ending.LEDs are off: Device is switched off.
Measurement values are displayed for the entire time the heater is positioned above the sample. Do not move the device during measurement to avoid errors.
The external printer outputs header data (Model, device ID, date, time, heater temperature, timer/shut-off parameters, initial weight) at the start of the measurement, and measurement value, measurement time, substance, and initials after the measurement is complete and the heater is swung back.
Please ensure drying trays have cooled down before using them for measurement.
For substances difficult to remove from the tray after drying, PCE Instruments recommends using foil trays, which can be made using the PCE-UX foil press.
If the measurement ends prematurely, the display will show an error related to the sample weight.
Data Interface
Weighing, temperature, and drying data can be accessed via a standard USB flash drive and serial interface (V24 RS 232). Depending on the model, a separate PCE-UX 3092 printer or a computer can be connected.
Interface Description
- Model: V24 RS 232 serial data transmission and USB flash drive.
- Transmission rate: 9600 baud.
- Data bits: 8 data bits, 1 stop bit.
- Parity check: None.
- Handshake: None.
Data transmission using a USB flash drive: If the receiving computer has an internet connection, a driver is automatically installed, creating a COM interface. Otherwise, the driver must be manually installed.
PC Data Transfer: Special software, ULTRA X DataBridge, is required for transmitting data to a PC. Data can be exported in .xls or .PDF formats. Installation is straightforward.
Accessories, Consumables, and Spare Parts
Accessories
- AV drying tray, large: Order no. 10000288 (Sample carriers, 2 trays supplied).
- Tray tongs: Order no. 10000230 (for lifting the tray, 2 supplied).
- Mains cable: Order no. H1000019 (Mains connection, supplied).
- Calibration weight F1 200 g: Order no. 10000403 (for calibrating the scales).
- Aluminium foil 130 mm x 260 mm: Order no. H1000013 (for placing on the drying tray).
- Foil press: Order no. 10000062 (for shaping aluminium foil).
- Software DataBridge: For transmitting data to a PC.
Consumables
- Printing paper 58 mm: Order no. 10000234 (for printer, thermal, stable for 10 years).
Spare Parts
- Ceramic heater 250 W: Order no. 10000405 (for PCE-UX 3081).
- Ceramic heater 400 W: Order no. 10000318 (for PCE-UX 2081WQ).
Potential Errors
| Error Display | Reason | Solution |
|---|---|---|
| Display blank | Initial load too small or large, Tray support and/or tray missing, Location with too much disturbance, Draught, Area around tray support is dirty, Scale is faulty, Scale without power, Scale is not connected to control card | Place empty drying tray on centre of tray support, Place on a vibration-free surface, Protect from draughts, Carefully clean area around tray support, Contact PCE Instruments. |
| Scale continuing to run | Location with too much disturbance, Draught, Area around tray holder is dirty, Scale is faulty | Place on a vibration-free surface, Protect from draughts, Carefully clean area around tray holder, Contact PCE Instruments. |
| Drying parameters cannot be entered | Menu is locked | Unlock menu lock, contact the person responsible or PCE Instruments. |
| Menu lock cannot be unlocked | 5-digit PIN for unlocking menu lock is incorrect | Contact PCE Instruments. |
| Scale cannot be tared | Weight within minus range | Switch off device, place empty tray on balance and switch on again. |
| Scale cannot be calibrated | Weight difference less than 16 mg or more than 16 mg, Initial load too small or large, tray holder and/or tray missing, Location with too much disturbance, Draught, Area around tray holder is dirty, Scale setting is incorrect, Scale is faulty | Less than 16 mg: Tare, press and hold button briefly. More than 16 mg: Tare. Place empty drying tray on centre of tray holder, Place on a vibration-free surface, Protect from draughts, Carefully clean area around tray holder, Contact PCE Instruments. |
| System error | Calibration process has stopped | Disconnect device from mains and switch on again. |
| Heater will not switch on | Heater is faulty | Change heater. |
| Infrared heater will not switch off | Temperature sensor is faulty, S202S01 relay is faulty, Automatic shut-off is set incorrectly, Timer is set incorrectly | Contact PCE Instruments. Check shut-off settings. Check timer settings. |
| Measurement not completing | Shut-off settings incorrect, Timer settings incorrect | Check shut-off settings. Check timer settings. |
| Printer: No printout | Printer is faulty or no data transmission, No connection to printer | Contact PCE Instruments. |
| Printer: No linefeed | Linefeed button or printer is faulty | Contact PCE Instruments. |
| Printer: Printout invisible | No thermal paper, Printing unit or printer board is faulty | Use the thermal paper from PCE. Contact PCE Instruments. |
| Printer: Printout too light | Printing unit is dirty | Contact PCE Instruments. |
Care and Maintenance
Please ensure that no pollutants enter the device. Always keep the area around the tray holder clean. Do not use compressed air to clean the device. Do not overload the scales; maximum load is 400 g. Protect the device from moisture and impact.
For ideal care and maintenance, PCE Instruments recommends taking out a service contract.
Technical Data
| Parameter | Specification |
|---|---|
| Scale: | |
| Weight range | 400g |
| Resolution | 0.1 g |
| Sample volume | max. 360 cm³ |
| Weighing tray | 240 x 110 mm |
| For moisture analysis: | |
| Initial weight | Any weight from approx. 5 g, recommended from 10 g |
| Heater capacity | Ceramic heater: 2x 250 watts (3081); 2x 400 watts (3081 WQ) |
| Automatic shut-off | Automatic drying to a constant weight, using 3 factors for IR heaters and Q heaters, which can be set separately: Query start, query interval, weight loss/query interval. |
| Temperature | Ceramic heater: adjustable from 40°C to 180°C (3081); 105°C to 360°C (3081QW) |
| Measuring units | % moisture, % dry matter, % ATRO moisture, g solid matter/kg |
| Reading accuracy | 0.1% readable |
| Communication: | |
| Data terminal | V24 RS 232/ USB |
| Data output | All weight and measuring data meet GLP requirements with a start time, optional variable interval printout, date, time, total measuring time etc. Using ULTRA X DataBridge software, data can be transmitted to a PC as a .xls file (Microsoft Excel) or .PDF file. |
| For operating the device: | |
| Mains connection | 230 V ± 10% / + 15 % |
| Mains frequency | 48-63 Hz |
| Power input | 250 watts – 850 watts, depending on the model |
| Dimensions | approx. W 385 x D 275 x H 420 mm |
| Weight | approx. 9 kg / 15 kg (3081 model) |
Technical data subject to change without notice.
PCE-UX moisture analyzers are stand-alone units and not suitable for use in continuous production runs. PCE Instruments shall not accept any liability for incorrect readings!
Disposal
For the disposal of batteries in the EU, Directive 2006/66/EC applies. Due to contained pollutants, batteries must not be disposed of as household waste; they must be given to designated collection points.
In accordance with EU Directive 2012/19/EU, PCE Instruments takes back devices for re-use or recycling by authorized companies.
For countries outside the EU, batteries and devices should be disposed of in accordance with local waste regulations.
If you have any questions, please contact PCE Instruments.
PCE Instruments Contact Information
Germany
PCE Deutschland GmbH
Im Langel 26
D-59872 Meschede
Deutschland
Tel.: +49 (0) 2903 976 99 0
Fax: +49 (0) 2903 976 99 29
info@pce-instruments.com
www.pce-instruments.com/deutsch
United Kingdom
PCE Instruments UK Ltd
Trafford House, Chester Rd, Old Trafford
Manchester M32 0RS
United Kingdom
Tel: +44 (0) 161 464902 0
Fax: +44 (0) 161 464902 9
info@pce-instruments.co.uk
www.pce-instruments.com/english
The Netherlands
PCE Brookhuis B.V.
Institutenweg 15
7521 PH Enschede
Nederland
Telefoon: +31 (0)53 737 01 92
info@pcebenelux.nl
www.pce-instruments.com/dutch
France
PCE Instruments France EURL
23, rue de Strasbourg
67250 Soultz-Sous-Forets
France
Téléphone: +33 (0) 972 3537 17
Numéro de fax: +33 (0) 972 3537 18
info@pce-france.fr
www.pce-instruments.com/french
Italy
PCE Italia s.r.l.
Via Pesciatina 878 / B-Interno 6
55010 Loc. Gragnano Capannori (Lucca)
Italia
Telefono: +39 0583 975 114
Fax: +39 0583 974 824
info@pce-italia.it
www.pce-instruments.com/italiano
United States of America
PCE Americas Inc.
1201 Jupiter Park Drive, Suite 8
Jupiter / Palm Beach 33458 FL
USA
Tel: +1 (561) 320-9162
Fax: +1 (561) 320-9176
info@pce-americas.com
www.pce-instruments.com/us
Spain
PCE Ibérica S.L.
Calle Mula, 8
02500 Tobarra (Albacete)
España
Tel. : +34 967 543 548
Fax: +34 967 543 542
info@pce-iberica.es
www.pce-instruments.com/espanol
Turkey
PCE Teknik Cihazlari Ltd.ti.
Halkali Merkez Mah. Pehlivan Sok. No.6/C
34303 Küçükçekmece - Istanbul
Türkiye
Tel: 0212 471 11 47
Faks: 0212 705 53 93
info@pce-cihazlari.com.tr
www.pce-instruments.com/turkish
Denmark
PCE Instruments Denmark ApS
Birk Centerpark 40
7400 Herning
Denmark
Tel.: +45 70 30 53 08
kontakt@pce-instruments.com
www.pce-instruments.com/dansk