HT Instruments PVCHECKs Electric Safety and Performance Of a Photovoltaic System
Specifications
- Product Name: PVCHECKs
- Release Version: EN 3.02 – 14/05/2025
Product Information
PVCHECKs is a versatile instrument designed for measuring efficiency, parameters, insulation, and continuity on PV plants and modules.
General Description
The PVCHECKs instrument offers various functions including defining new PV modules, modifying existing modules, and deleting modules.
Measuring Efficiency on PV Plants
To measure efficiency on PV plants using the remote unit SOLAR-02, follow the instructions provided in section 6.1 of the user manual.
Measuring Parameters without SOLAR-02
If you need to measure parameters on PV plants without using SOLAR-02, refer to section 6.2 for detailed instructions.
Quick Check on PV Modules and Strings (IVCK)
For a quick check on PV modules and strings using IVCK, follow the guidelines outlined in section 6.3 of the manual.
Measurement of Insulation
To measure insulation on PV modules, strings, or fields, refer to section 6.4 for step-by-step instructions.
Measurement of Continuity
For measuring continuity on PV modules, strings, or fields, follow the calibration steps for measuring cables and then proceed with continuity measurement as described in section 6.5.
“`
PRECAUTIONS AND SAFETY MEASURES
The instrument has been designed in compliance with directive IEC/EN61010-1 relevant to electronic measuring instruments. Before and while carrying out measurements, observe the following indications and read all notes preceded by the symbol with the utmost attention.
· Do not carry out any voltage or current measurement in humid environments · Do not carry out any measurements in case gas, explosive materials or flammables are
present, or in dusty environments · Avoid contact with the circuit being measured if no measurements are being carried out · Avoid contact with exposed metal parts, with unused measuring probes, etc. · Do not carry out any measurement in case you find anomalies in the instrument such
as deformation, breaks, absence of display on the screen, etc. · Pay special attention when measuring voltages higher than 25V in special
environments and 50V in normal environments, since a risk of electrical shock exists.
In this manual, and on the instrument, the following symbols are used:
CAUTION: observe the instructions given in this manual; an improper use could damage the instrument or its components
High voltage danger: electrical shock hazard
Double insulation
DC voltage or current
Connection to earth
1.1. PRELIMINARY INSTRUCTIONS · This instrument has been designed for use in the environmental conditions specified in
§ 10.6. Do not use in different environmental conditions. · The instrument may be used for measuring VOLTAGE and CURRENT in CAT III 300V
with a maximum voltage of 1000VDC between inputs. Do not use on systems exceeding the limit values specified in § 10.1, § 10.2 and § 10.3
CAUTION
Do not use the instrument for IVCK tests on PV modules with efficiency >19%. Check the technical characteristics of the PV modules before carrying out the tests in order to avoid possible damage to the instrument
· We recommend following the normal safety rules devised to protect the user against dangerous currents and the instrument against incorrect use.
· Only the accessories provided together with the instrument will guarantee safety standards. They must be in good conditions and replaced with identical models, when necessary.
· Check that the batteries are correctly inserted. · Before connecting the measuring cables to the circuit being measured, check that the
desired function has been selected.
EN – 3
PVCHECKs
1.2. DURING USE Please carefully read the following recommendations and instructions:
CAUTION
· Failure to comply with the caution notes and/or instructions may damage
the instrument and/or its components or be a source of danger for the
operator.
· The symbol ”
” indicates a full charge level of the internal batteries.
When battery charge decreases to a minimum level, the symbol ”
”
appears on the display. In this case, stop testing and replace the batteries
according to the indications given in § 9.2
· The instrument is capable of keeping data stored even without
batteries.
1.3. AFTER USE When measurements are completed, turn off the instrument by pressing and holding the ON/OFF key for some seconds. If the instrument is not to be used for a long time, remove the batteries and follow the instructions given in § 3.3
1.4. DEFINITION OF MEASUREMENT (OVERVOLTAGE) CATEGORY Standard “IEC/EN61010-1: Safety requirements for electrical equipment for measurement, control and laboratory use, Part 1: General requirements” defines what measurement category, commonly called overvoltage category, is. § 6.7.4: Measured circuits, reads:
Circuits are divided into the following measurement categories:
· Measurement category IV is for measurements performed at the source of the lowvoltage installation Examples are electricity meters and measurements on primary overcurrent protection devices and ripple control units.
· Measurement category III is for measurements performed on installations inside buildings Examples are measurements on distribution boards, circuit breakers, wiring, including cables, bus-bars, junction boxes, switches, socket-outlets in the fixed installation, and equipment for industrial use and some other equipment, for example, stationary motors with permanent connection to fixed installation.
· Measurement category II is for measurements performed on circuits directly connected to the low-voltage installation
Examples are measurements on household appliances, portable tools and similar equipment.
· Measurement category I is for measurements performed on circuits not directly connected to MAINS
Examples are measurements on circuits not derived from MAINS, and specially protected (internal) MAINS-derived circuits. In the latter case, transient stresses are variable; for that reason, the standard requires that the transient withstand capability of the equipment is made known to the user.
EN – 4
PVCHECKs
2. GENERAL DESCRIPTION
2.1. INTRODUCTION This instrument has been designed to carry quick checks (IVCK) on photovoltaic modules/strings in order to verify the parameters declared by the manufacturer. In addition, this instrument measures insulation/continuity on PV modules/strings/fields and evaluates the efficiency of a PV field.
IVCK and insulation/continuity measurements may be performed in sequential order IVCK Insulation Continuity, or be carried out separately in manual mode.
2.2. INSTRUMENT FUNCTIONS The instrument has the following features: Continuity test of protective conductors (LOW) · Test with test current > 200mA in compliance with standard IEC/EN62446-1 · Manual calibration of measuring cables.
Measurement of insulation resistance on PV modules/strings (M) · Test voltages of 250V, 500V, 1000VDC in compliance with standard IEC/EN62446-1 · 3 measuring modes: Field, Timer, String · Insulation test of unearthed metal masses
Efficiency evaluation of a PV installation in the short/medium term (EFF) · Measurement of DC voltage, DC current and DC power at PV modules/strings’ outputs · Measurement of irradiation [W/m2] by means of a reference cell connected to the
optional remote unit SOLAR-02 · Measurement of module and environmental temperature by means of probe connected
to the optional remote unit SOLAR-02 · Application of compensation relationships to DC efficiency. · Immediate evaluation of DC efficiency according to the limits set by the user · Parameter recording of a PV system with 5s to 60min programmable IP
Quick checks (IVCK) in compliance with standard IEC/EN62446-1 · Measurement of open circuit voltage Voc on PV modules/strings up to 1000VDC · Measurement of short-circuit current Isc on PV modules/strings up to 15A · Measurement of irradiation by means of an optional reference cell · Immediate evaluation (OK/NO) of test results · Possible connection with the optional remote unit SOLAR-02 · Internal customizable database for managing up to 30 PV modules · Display of results in OPC and STC conditions
The instrument is provided with backlit display, internal contrast adjustment and a HELP key able to give a valid help to the operator while connecting the instrument to the installation. An Auto Power OFF function, which can also be deactivated, is available after approx. 5 minutes’ idleness.
EN – 5
PVCHECKs
3. PREPARATION FOR USE
3.1. INITIAL CHECKS Before shipping, the instrument has been checked from an electric as well as mechanical point of view. All possible precautions have been taken so that the instrument is delivered undamaged. However, we recommend checking it to detect any damage possibly suffered during transport. In case anomalies are found, immediately contact the Dealer.
We also recommend checking that the packaging contains all components indicated in §
10.7. In case of discrepancy, please contact the Dealer. In case the instrument should be returned, please follow the instructions given in § 12.
3.2. INSTRUMENT POWER SUPPLY The instrument is battery supplied. For battery type and life, see § 10.5.
The symbol ”
” indicates a full charge level of the internal batteries. When battery
charge decreases to a minimum level, the symbol ”
” appears on the display. In this
case, stop testing and replace the batteries according to the indications given in § 9.2.
The instrument is capable of keeping data stored even without batteries.
The instrument is provided with advanced algorithms to maximize the batteries’ life.
During instrument operation, a short pressing of the key turns on the display’s backlighting. In order to save battery efficiency, backlighting automatically turns off after approx. 30 seconds.
A frequent use of backlighting reduces the batteries’ life.
3.3. STORAGE In order to guarantee precise measurement, after a long storage time under extreme
environmental conditions, wait for the instrument to come back to normal condition (see §
10.6).
EN – 6
4. NOMENCLATURE
4.1. INSTRUMENT DESCRIPTION
PVCHECKs
CAPTION: 1. Inputs 2. Display 3. Connector for optical output/USB
port 4. Arrow keys/ENTER key 5. GO/STOP key 6. SAVE key 7. ON/OFF key 8. HELP / key 9. ESC/MENU key
Fig. 1: Description of the front part of the instrument
CAPTION: 1. Input for irradiation measuring probe 2. Input for auxiliary temperature
measuring probe / DC current clamp (IVCK, EFF) 3. P, N inputs for DC voltage (IVCK, EFF) / Insulation (M) measurement 4. E, C inputs for continuity tests (LOW)
Fig. 2: Description of the upper part of the instrument CAPTION: 1. Connector for opto-insulated/USB output cable connection
Fig. 3: Description of the instrument’s side
EN – 7
PVCHECKs
4.2. KEYBOARD DESCRIPTION The keyboard includes the following keys:
ON/OFF key to switch on/off the instrument
ESC/MENU key to exit the selected menu without confirming and to go back to the main menu keys to move the cursor inside the various screens in order to select programming parameters ENTER key to confirm modifications, selected programming parameters and to select the function to be accessed from the menu.
GO/STOP key to start measurements
SAVE key to save the measured values
HELP key (long pressing) to access the on-line help, to view the possible connections between the instrument and the installation.
key (simple pressing) to activate the display’s backlighting
4.3. DISPLAY DESCRIPTION
The display is a graphic module with a resolution of 128 x 128 15/05/12 15:34:26
dots. The display’s first line indicates the system’s date/time
and contains the battery charge indicator.
At the bottom of the display, the function of the ENTER key
and the active mode are displayed.
The symbol
indicates the presence of an active radio
connection with the remote unit SOLAR-02.
The flashing symbol
indicates that the instrument is
currently searching for a radio connection with the remote unit
SOLAR-02.
Selection
4.4. INITIAL SCREEN When switching on the instrument, the initial screen appears for a few seconds. It shows:
· The instrument model (PVCHECKs)
· The manufacturer’s name · The presence of the enabled internal module for RF
connection
· The serial number (SN:) of the instrument
· The firmware version (FW:) in the instrument’s memory
· The date of the last instrument calibration (Calibration date:)
PVCHECKs
HT
RF SN: 15345678
FW: 1.10 Calibration date:
09/04/2025
After a few seconds, the instrument switches to the last function selected.
EN – 8
PVCHECKs
5. GENERAL MENU
Pressing the ESC/MENU key in any condition of the 15/05/12 15:34:26
instrument displays the general menu screen, in which the I V C K T e s t m o d . / s t r i n g s
instrument may be set, the saved measures can be displayed
and the desired measuring function may be selected.
LOW PE Continuity test
Use the cursor to select one of the options and confirm with M I n s u l a t i o n t e s t
ENTER to access the desired function.
EFF Efficiency test
SET Settings
DB
Archive modules
MEM Recall Memory
PC
Trasf. data to PC
Press ENTER to sel.
MENU
5.1. SET INSTRUMENT SETTINGS
Position the cursor onto SET by using the arrow keys (,) 15/05/12 15:34:26
and confirm with ENTER. The display shows the screen G e n e r a l
which lists the different settings of the instrument.
Measuring unit
Settings will be maintained also after switching off the D a t e a n d T i m e
instrument.
Remote Unit – Pyranometer
Irradiance
DC Clamp
Press ENTER to sel. SET
5.1.1. General
1. Position the cursor onto General by using the arrow keys 15/05/12 15:34:26 (,) and confirm with ENTER.
2. The display shows the screen which allows: Setting the instrument language Enabling/disabling the auto power off function Adjusting display contrast
Language :English
Auto Power Off : NO
Contrast:
: 40
Key Beep
:NO
Enabling the acoustic signal when pressing a key.
3. Use the arrow keys (,) for option setting, and select the desired option by means of the arrow keys ( , ).
4. Press the SAVE key to save the settings made; the
SAVE to store SET
message “Data saved” will be displayed for a few
seconds. Press the ESC/MENU key to exit without saving
and go back to the previous screen.
EN – 9
PVCHECKs
5.1.2. Measuring unit This section allows setting the measuring units of some parameters included in the
database (DB) for the management of PV modules (see § 5.6) when measuring IVCK.
1. Position the cursor onto “Measuring Unit” by using the 15/05/12 15:34:26
2.
arrow keys (,) and confirm with ENTER. The display shows the screen which allows
setting
the
Alpha
:mA/°C
measuring units of the following parameters:
Beta
: %/°C
· Alpha possible selection: “%/°C” and “mA/°C” · Beta possible selection: “%/°C” and “mV/°C” · Gamma expressed in “%/°C”
Gamma Tolerance
: %/°C : %
· Voc and Isc tolerance expressed in “%” 3. Set the desired measuring units by means of the arrow
keys ( , ).
SAVE to store SET
4. Press the SAVE key to save the settings made; the
message “Data saved” will be displayed for a few seconds.
5. Press the ESC/MENU key to exit without saving and go
back to the previous screen.
5.1.3. Date and time
1. Position the cursor onto “Date and Time” by using the 15/05/12 15:34:26
2.
arrow keys (,) and confirm with ENTER. The display shows the screen which allows
setting
the
Year Month
:2012 : 05
system’s date/time USA (US) format.
both
in the
European
(EU)
or in
the
Day Hours Minutes
: 15 : 09 : 53
3. Set the desired measuring units by means of the arrow F o r m a t
: EU
keys ( , ).
4. Press the SAVE key to save the settings made; the message “Data saved” will be displayed for a few
seconds. Press the ESC/MENU key to exit without saving
SAVE to store SET
and go back to the previous screen.
EN – 10
PVCHECKs
5.1.4. Remote unit/Pyranometer This section allows selecting the type of remote unit to be used (if available) and setting
the values of typical parameters (Sensitivity and Alpha) of the reference irradiance cell
(HT304k optional accessory). The values of these parameters, which are printed on the back label of the cell, depend on the type of PV module being tested.
1. Position the cursor onto Remote Unit-Pyranometer by 15/05/12 15:34:26
2.
using the arrow keys (,) and confirm with ENTER. The display shows the screen which allows selecting the
Remote
Unit
EFF:
YES
use of the remote unit for EFF or IVCK tests. Possible R e m o t e U n i t I V C K : N O
options are:
Sens.
YES (use of SOLAR-02)
Alpha
NO (no use of SOLAR-02) In case the remote unit
:31.0mV/kW/m2 : 0.060 %/°C
SOLAR-02 is not used for IVCK tests, it is necessary to
set the values of Sensitivity (Sens.) and Alpha
parameter of the irradiance cell on the instrument. 3. Set the desired values by means of the keys ( , ) 4. Press the SAVE key to save the settings made; the
SAVE to store SET
message “Data saved” will be displayed for a few seconds.
Press the ESC/MENU key to exit without saving and go
back to the previous screen
EN – 11
PVCHECKs
5.1.5. Irradiance This section allows setting the minimum irradiance threshold both for IVCK measurement
and for efficiency test on a PV installation.
1. Position the cursor onto “Irradiance” by using the arrow 15/05/12 15:34:26
2.
keys (,) and confirm with ENTER. The display shows the screen containing
“Min
Irr
IVCK”,
Min Min
IrrIVCK Irr EFF
:000W/m2 :600W/m2
which allows setting the minimum irradiance threshold
expressed in W/m2, used by the instrument as a reference
during IVCK measurement, and “Min Irr EFF”, which
allows setting the minimum irradiance threshold expressed
in W/m2, used by the instrument as a reference while measuring the efficiency of a PV installation. Use the arrow keys (,) to switch between the two items.
SAVE to store SET
3. To set the minimum irradiance threshold, use the arrow
keys ( , ). To obtain results with the accuracy
guaranteed in this manual, we recommend following the
instructions given in § 10. The value can be set within the
following range: 0 800 W/m2
4. Press the SAVE key to save the settings made; the
message “Data saved” will be displayed for a few seconds.
Press the ESC/MENU key to exit without saving and go
back to the previous screen.
CAUTION
The “0 W/m2” setting for parameter “Irr min IVCK” allows carrying out IVCK measurements without the following conditions being checked: Connection of reference cell to IRR input of the instrument Unsteady irradiation values Number of modules consistent with the measured open circuit voltage
5.1.6. DC Clamp This option allows setting a possible correction factor K for the DC clamp, in order to improve current measurement. If present, the correction factor is indicated on the rear label of the clamp
itself, indicated as:
K= X.xxx
In case no label is present, set k = 1.000
1. Position the cursor onto “DC Clamp” by using the arrow 15/05/12 15:34:26
2.
keys (,) and confirm with ENTER. The display shows the item “DC clamp
K”,
which
allows
DC
clamp
K
:1.000
setting the corrective factor in the following range 0.950 to
1.050. Set the desired values by means of the arrow keys ( , ).
3. Press the SAVE key to save the settings made; the
message “Data saved” will be displayed for a few seconds.
Press the ESC/MENU key to exit without saving and go back to the previous screen.
SAVE to store SET
EN – 12
PVCHECKs
5.2. EFF EFFICIENCY TEST SETTINGS FOR PV INSTALLATIONS The purpose of this measurement is to evaluate the DC efficiency of a photovoltaic
installation, with the possibility of obtaining a positive or negative result of the
check/recording depending on a limit on the nDC parameter, freely set by the user. For this test, the use of the optional remote unit SOLAR-02 is necessary (see § 6.1)
5.2.1. Instrument settings
1. Position the cursor onto EFF by using the arrow keys 15/05/12 15:34:26
(,) values
and confirm with of the output
ENTER. The display shows the electrical parameters of the
Irr Pnom
– – 3.500
photovoltaic generator.
Tc
– – –
Te
– – –
Pdc
0.0
Vdc
0.000
Idc
0.0
ndc
– – –
W/m2
kW
°C °C kW V A
GO to Start
Selection
EFF
2. Press the ENTER key. The instrument shows the 15/05/12 15:34:26
following options: System parameters and Instrument I r r
settings.
Pnom
3.
Use the settings”
arrow keys and confirm
(,) to select “Instrument with ENTER. The instrument
Tc Te Pdc
shows the following screen:
Vdc Idc
ndc
– – 3.500
– – – – –
0.0 0.000
0.0 – – –
W/m2
kW
°C °C kW V A
System parameters
Instrument settings
Selection
EFF
4. By using the arrow keys ( , ) it is possible to set:
15/05/12 15:34:26
the integration period (IP) instrument when testing
which can be used the parameters of
by a
the PV
IP FS
DC
clamp
: 5s : 1000 A
system. The following values can be selected: 5s, 10s,
30s, 60s, 120s, 300s, 600s, 900s, 1800s, 3600s. The FS of the DC clamp used for measuring DC
current with value selectable between 1A 3000A
5. Press the SAVE key to save the settings made; the message “Data saved” will be displayed for a few
seconds. Press the ESC/MENU key to exit without saving and go back to the previous screen.
SAVE to store EFF
EN – 13
PVCHECKs
5.2.2. System parameters
1. Position the cursor onto EFF by using the arrow keys 15/05/12 15:34:26
(,) and confirm with ENTER. The display shows the I r r values of the output electrical parameters of the P n o m
– – 3.500
photovoltaic generator.
Tc
– – –
Te
– – –
Pdc
0.0
Vdc
0.000
Idc
0.0
ndc
– – –
W/m2
kW
°C °C kW V A
GO to Start
Selection
EFF
2. Press the ENTER key. The instrument shows the 15/05/12 15:34:26
following options: System parameters and Instrument I r r
settings
Pnom
3.
Use the arrow Parameters” and
keys (,) to select “System confirm with ENTER. The instrument
Tc Te Pdc
shows the following screen:
Vdc Idc
ndc
– – 3.500
– – – – –
0.0 0.000
0.0 – – –
W/m2
kW
°C °C kW V A
System parameters
Instrument settings
Selection
EFF
4. By using the arrow keys ( , ) it is possible to set:
15/05/12 15:34:26
Pmax maximum total rated power of PV installation expressed in kW
Range coefficient of power variation with
Pmax: Gamma
temperature,
characteristic
parameter
of
PV
modules
NOCT Te
:3.500kW
:
-0.45 %/°C
: 45 °C
: 40 °C
(range: -1.00 -0.01%/C)
Tc
: 45 °C
NOCT
rated
operating
temperature
of
the
cell,
nDC Lim: Corr.Type.
characteristic parameter of PV modules (typically in
: 0.85 : T. Env.
the following range: 42 48°C) Te, Tc setting of default values of environmental
and (Te) and PV module (Tc) temperatures. These
values are taken into consideration by the instrument
SAVE to Store EFF
only when no auxiliary probe is connected to the
remote unit SOLAR-02 (range: Te = 0°C 80°C; Tc =
0°C 100°C) nDC Lim minimum limit of DC efficiency (default
value: 0.85 , range: 0.01 1.15) Corr. Type Setting of the compensation
relationship on the calculation of Pdc power and the
maximization of DC efficiency (see § 5.2.3).
EN – 14
PVCHECKs
5.2.3. Selection of the compensation relationship of temperature effects
This option allows selecting the relationship to be used to correct the measurements
performed according to the modules’ temperature as regards the calculation of nDC
efficiency. The following modes are available:
– T.Mod.: Correction factor Rfv2 related to PV module Temp. (Italian guideline CEI-82-25)
– T.Env: Correction factor Rfv2 related to environmental Temp. (Italian guideline CEI-82-25)
– nDC:
nDC correction according to module temperature
CAUTION
If it’s performed a test according to Italian guideline it is recommended to select “T.Env.” correction
Corr. Type
T.Mod.
Temperature used (Tcel)
Tcel = Tm oduli_Mis
T.Env.
Tc e l =
Ta m b+
(NOCT
–
20)
Irr 800
nDC
Tcel = Tm oduli_Mis
Calculation of nDC
Ref.
1
Rfv2
=
1
–
(Tcel-
40)
100
(se Tcel 40C) (se Tcel 40C)
hence
CEI 82-
nDC =
Pdc
25
Rfv2
Gp GSTC
Pn
( ) nDC
=
GSTC Gp
1 +
100
Tcel
– 25
Pdc Pn
– – –
where:
Symbol
G p
GS T C
P n Pdc
Rfv 2
NOCT
Description
Irradiance measured on PV module surface
Standard Irradiance = 1000
Rated Power = sum of all Pmax values of the section of the PV plant being measured
DC power measured at the PV generator’s output Correction factor depending on the Temperature of PV Cells (Tcel) measured or calculated according to the type of selected correction relationship Absolute value of the thermal coefficient of Pmax of the PV modules included in the plant section being measured. (Normal Operating Cell Temperature) = Temperature the cells under reference conditions are brought to (800W/m2, 20°C, AM=1.5, air speed =1m/s).
Measurin g unit
W/m2 W/m2
kW kW
%/C %/C
For further details see § 11.1
EN – 15
PVCHECKs
5.3. LOW SETTINGS FOR CONTINUITY TEST WITH 200MA The purpose of this measurement is to test the continuity of protective conductors and equipotential ones (e.g. from rod to earth and connected foreign earth) and earth rods of SPDs on PV installations. The test must be carried out using a test current > 200mA according to the prescriptions of IEC/EN62446-1 guideline
5.3.1. Instrument settings
1. Position the cursor onto LOW by using the arrow keys 15/05/12 15:34:26
(,) and confirm with ENTER. The display shows the R P E m a x :
1
following screen:
Rcal
: —
Rpe
= —
Itest
= – – – mA
Selection
LOW
2. Press the ENTER key. The instrument shows the 15/05/12 15:34:26
3.
following options: Settings and Leads calibration Use the arrow keys (,) to select “Settings”
and
RPE max Rcal
: :
1 —
confirm with ENTER. The instrument shows the following screen:
Rpe
= —
Itest
= – – – mA
Leads calibration Settings Selection
LOW
4. Using the arrow keys ( , ) it is possible to set the 15/05/12 15:34:26
maximum limit value of Rpe resistance which instrument uses as a reference while measuring.
the R P E m a x
:01
5. Press the SAVE key to save the settings made; the message “Data saved” will be displayed for a few
seconds. Press the ESC/MENU key to exit without saving and go back to the previous screen.
SAVE to store
Selection
LOW
CAUTION
The settings saved for max RPE also affect the settings for the Continuity test contained in the IVCK measurement (MENU IVCK)
EN – 16
PVCHECKs
5.4. M SETTINGS FOR INSULATION MEASUREMENT
5.4.1. Instrument settings
1. Position the cursor onto M by using the arrow keys 15/05/12 15:34:26
(,) and confirm with ENTER. The display shows the I n s . T e s t : 1 0 0 0 V
following screen:
Ri min
: 1.0
M
Mode
: Field
Vtest – – – V – – – V
Ri(+)
– – –
M
Ri(-)
– – –
M
Rp
– – –
M
Selection
M
2. Press the ENTER key. The instrument shows the option: 15/05/12 15:34:26
3.
Settings: Confirm with ENTER. The instrument shows the following
Ins. Test Ri min
: 1000 : 1.0
V M
screen:
Mode
: Field
Vtest – – – V – – – V
Ri(+)
– – –
M
Ri(-)
– – –
M
Rp Settings Selection
– – –
M
M
4. Use the arrow keys (,) for option setting, and select 15/05/12 15:34:26
the desired option by means of the arrow keys ( , ). The following parameters can be set:
Ins. Test
: 1000V
Ri min
: 1.00 M
Ins. Test test voltage: 250, 500, 1000VDC Mode operating modes: Field, Timer, String
Mode
: TIMER
Ri min minimum limit value for insulation resistance T e s t t i m e : 1 0 s
Test time maximum value of test time (only for
TIMER mode) with a maximum value of 300s in steps
SAVE to store
of 1s.
M
5. Press the SAVE key to save the settings made; the
message “Data saved” will be displayed for a few
seconds. Press the ESC/MENU key to exit without saving
and go back to the previous screen.
CAUTION
The settings saved for test voltage also affect the settings for the insulation measurement contained in the IVCK measurement (MENU IVCK)
EN – 17
PVCHECKs
5.5. IVCK SETTINGS FOR IVCK QUICK TEST The purpose of this measurement is to verify the functionality of the connections and of the strings in a photovoltaic field, according to the provisions of standard IEC/EN62446-1 by measuring the open circuit voltage and the short-circuit current at operating conditions and referred to STC (via the optional measurement of irradiation) and providing immediate results regarding the measurement just carried out, both in absolute terms and by comparison with the strings previously tested. The test also allows carrying out (if selected) the continuity test and an insulation measurement in a sequence.
5.5.1. Instrument settings
1. Position the cursor onto IVCK by using the arrow keys 15/05/12 15:34:26
(,) and confirm with ENTER. The display shows the following screen:
Module
Vdc Irr
: SUNPWR318
: 0.0
V
: 0
W/m2
Tc
: Auto °C
Voc,Isc:
Ri(1000V) = – – – M
Rpe (Cal) = – – –
Selection
IVCK
2. Press the ENTER key. The instrument shows the 15/05/12 15:34:26
following options: Settings, Reset Averages 6.3.4) and Leads calibration (see § 6.5.2).
(see
§
Module
Vdc Irr
3. Use the arrow keys (,) to select “Settings” and T c
: SUNPWR318
: 0.0
V
: 0
W/m2
: Auto
°C
confirm with ENTER. The instrument shows the following V o c , I s c :
screen:
Ri(1000V) = – – – M
Rpe (Cal) =
Leads calibration Reset Averages Settings Selection
—
IVCK
4. Use the arrow keys (,) for option setting, and select 15/05/12 15:34:26
the desired option by means of the arrow keys ( , ). The following parameters can be set:
Module: Mod.Nx Str Temp
SUNPWR210
: 01 : Manual
Module type of module being tested Mod. N x Str. number of modules
in
the
string.
Tol. Tol.
Voc Isc
25
°C
: 05%(+4%)
: 05%(+4%)
Allowable values 1 50
Ins. Test : 1000
V
Temp
temperature
measuring
method.
Possible
Ri min RPE Test
: 1.0 : 2
M
options:
“Auto” automatic measurement (recommended) carried out according to the measured value of the
modules’ Voc.
SAVE to store IVCK
EN – 18
PVCHECKs
“Manual” setting, made by the operator, of the known temperature value of the module in the line below. Aux temperature measurement with auxiliary probe PT1000. Tol. Voc (%) percentage value of the desired limit tolerance (set by the operator according to his/her own requirements) for the measurement of Voc carried out by the instrument. Allowable values: 0% 25%. The value between brackets (4%) indicates the reading error of the instrument in Voc measurement. Tol. Isc (%) percentage value of the desired limit tolerance (set by the operator according to his/her own requirements) for the measurement of Isc carried out by the instrument. Allowable values: 0% 25%. The value between brackets (4%) indicates the reading error of the instrument in Isc measurement. Ins Test enabling/disabling of insulation measurement and setting of test voltage. Possible options: OFF, 250V, 500V, 1000V. With enabled function, the line “Ri min” appears, which allows setting the minimum limit threshold in the range 0.1 100M RPE Test enabling/disabling of continuity test and setting of the limit threshold value for measurement. Possible options OFF, 1 5 in steps of 1 5. Press the SAVE key to save the settings made; the message “Data saved” will be displayed for a few seconds. Press the ESC/MENU key to exit without saving and go back to the previous screen.
CAUTION
The settings saved for test voltage for the insulation measurement carried out within the IVCK function also affect the settings for the single measurement (MENU M).
EN – 19
PVCHECKs
5.6. DB DATABASE MODULE MANAGEMENT The meter allows managing up to a maximum of 30 different types of PV modules, further to a DEFAULT module (not editable and not erasable) which can be used as
reference case when no piece of information about the type of module being tested is
available.
The parameters, referred to 1 module, which can be set are described below in Table 1 together with their measuring range, resolution and validity condition.
Symbol Nms Pmax
Voc Vmpp
Isc Impp Toll –
Toll +
Alpha
Beta Gamma NOCT
Tech. Rs
Description Number of modules per string
Range 1 50
Resol. 1
Maximum rated power of module
50 4800W
1W
Open-circuit voltage Voltage on maximum power point
Short-circuit current
15.00 99.99V 100.0 320.0V 15.00 99.99V 100.0 320.0V
0.5 15.00A
0.01V 0.1V 0.01V 0.1V
0.01A
Current on maximum power point
0.5 15.00A
0.01A
Negative tolerance for Pmax provided by the module manufacturer
0% 25.0% 0 99W
0.1% 1
Positive tolerance for Pmax provided by the module manufacturer
0 25% 0 99W
0.1% 1
Isc temperature coefficient
-0.1000.100%/°C 0.001%/°C -15.00 15.00mA/°C 0.01mA/°C
Voc temperature coefficient
-0.99 -0.01%/°C 0.01%/°C –0.999 -0.001V/°C 0.001V/°C
Pmax temperature coefficient
-0.99 -0.01%/°C 0.01%/°C
Rated working temperature of cell
0 100°C
1°C
Effects due to module technology
STD (standard) CAP(capacitive eff.)
Internal serial resistance
0.00 10.00
0.01
Table 1: Typical parameters of PV modules
Conditions Pmax – Vmpp I mpp 0.01
Pm a x Voc Vmpp Voc Vmpp Isc Impp Isc Impp 100*Tol-/Pnom< 25
100*Tol+/Pnom< 25
100*Alfa / Isc 0.1
100*Beta/Voc 0.999
CAUTION
The “Tech” item is referred to the choose of the technology of the module on test. Select the “STD” option for test on “STANDARD” PV modules, “CAP” option for test on higher capacitive effects PV modules
A wrong choice of the type of technology can lead to a negative outcome of the final test.
EN – 20
PVCHECKs
5.6.1. How to define a new PV module
1. Position the cursor onto DB by using the arrow keys 15/05/12 15:34:26
(,) and confirm with screen which contains:
ENTER.
The
display
shows
the
Model
Pmax
:DEFAULT
=
185
W
· ·
The type of PV module The values of the parameters associated to the module
Voc Vmpp Isc
= = =
44.5 37.5 5.40
V V A
2.
(see Table 1) Use the arrow keys
(
,
)
to
select
the
“DEFAULT”
Impp Toll-
= =
4.95 0
A %
module type and confirm with ENTER.
Selection
DB
3. Press the ENTER key, select the “New” command (which 15/05/12 15:34:26
allows defining a new module) and ENTER. Use the arrow keys (,)
confirm again with to scroll the list of
Model
Pmax
:DEFAULT
=
185
W
parameters.
Voc
=
44.5
V
Vmpp =
37.5
V
Isc
=
5.40
A
Impp =
4.95
A
Toll- =
0
%
New
Selection
DB
4. By using the internal virtual keyboard it is possible to 15/05/12 15:34:26
define the name of the module (e.g. SUNPOWER 210) by means of the arrow keys (,, , ). Press ENTER to
Model
Pmax
: =
185
W
digit any character of the desired name.
Voc
=
44.5
V
5.
Press the SAVE key to save the name of the new module thus defined or the ESC/MENU key to exit without saving.
KEYBOARD SUNPOWER 210
A B CD E F GH I J K L MNOP
Q R S T U V WX Y Z – + 0 1 2 3
4 5 6 7 8 9 SPACE
DEL
SAVE / ESC
6. Insert the value of each parameter (see Table 1) 15/05/12 15:34:26
according to the possible datasheet of the Manufacturer.). Position the cursor onto the parameter to be defined by using the arrow keys (,) and set the value using the
Model
Pmax Voc
: SUNPWR 210
= 0 W
=
0.0
V
arrow keys ( , ). Keep the arrow keys ( , ) pressed
Vmpp Isc
= =
0.0 0.00
V A
to carry out a quick setting of the values.
Impp =
0.00
A
7. Press
the
SAVE
key
to
save
the
settings
or
the
Toll-
=
0
%
ESC/MENU key to exit without saving.
DB
CAUTION
· If the value of any parameter is unknown, press and hold the HELP key for a few seconds to set the default values.
· Upon pressing the SAVE key, the instrument checks the conditions indicated in Table 1
· and, if one or more of these conditions are not fulfilled, one of the error messages listed in § 6.6 is shown on the display. The instrument does not save the configuration set before any error is solved.
EN – 21
PVCHECKs
5.6.2. How to modify an existing PV module
1. Select the PV module to be modified from the internal 15/05/12 15:34:26
2.
database by means of the arrow keys ( , ). Press the ENTER key and select the “Modify” command using the arrow key ().
Model
Pmax Voc
SUNPWR210
=
210
W
= 47.70
V
3. Confirm selection with ENTER.
Vmpp = 40.00
V
Isc
=
5.75
A
New
Modify
Delete
Delete all
Selection
DB
4. By using the internal virtual keyboard it is possible to 15/05/12 15:34:26
define again the name of unchanged by means of the
the module or leave it arrow keys (,, , ).
Model:
Pmax
SUNPWR210
=
185
W
Press ENTER to digit any character of the desired name. V o c = 4 4 . 5 V
5.
Press the SAVE key to save the name of the new module thus defined or to access the new parameter setting.
KEYBOARD SUNPOWER 210
A B CD E F GH I J K L MNOP
Q R S T U V WX Y Z – + 0 1 2 3
4 5 6 7 8 9 SPACE
DEL
SAVE / ESC
6. Change the value of the desired parameters by using the 15/05/12 15:34:26
arrow keys (,) and set the value using the arrow keys ( , ). Keep the arrow keys ( , ) pressed to carry out
Model
Pmax
a quick setting of the values. If the value of any parameter V o c
is
unknown,
press
and
hold
the
HELP
key
for
a
few
Vmpp Isc
seconds to set the default values.
Impp
7.
Press
the
SAVE
key
to
save
the
settings
made
or
the
Toll-
: SUNPWR 210
= 210 W
=
47.70
V
=
40.00
V
=
5.75
A
=
5.25
A
=
5
%
ESC/MENU key to exit without saving. In this case, the
instrument shows the message “Data not saved”.
DB
5.6.3. How to delete an existing PV module
1. Select the PV module from the internal database by 15/05/12 15:34:26
2.
means of the arrow keys ( , ). Press the ENTER key and select the “Delete” command by means of the arrow key () to delete the selected
Model
Pmax Voc
:SUNPWR210
=
210
W
= 47.70
V
module.
Vmpp = 40.00
V
Isc
=
5.75
A
3. Press the ENTER key and select the “Delete all” N e w
command by means of the arrow key () to delete all M o d i f y
modules in the database.
Delete
4. Confirm the selection with ENTER or press ESC/MENU to D e l e t e a l l
exit.
Selection
DB
CAUTION
It is not possible to modify nor delete the “DEFAULT” PV module which is a factory setting.
EN – 22
PVCHECKs
6. OPERATING INSTRUCTIONS
6.1. MEASURING EFFICIENCY PV PLANTS BY USING REMOTE UNIT SOLAR-02 For the sake of simplicity, further in this section, the word “string” will be used, although often the term “photovoltaic field” would be more correct. From the point of view of the instrument, the management of a single string or of more parallel strings (PV field) is identical. The instrument PVCHECKs (Master) allows measuring the efficiency of PV installations in combination with the optional remote unit SOLAR-02, which the irradiance and temperature probes are connected to. This remote unit is capable of communicating with the Master unit (to manage data synchronization and download) via a radiofrequency (RF) connection, which is active up to a maximum distance of 1m between the units.
CAUTION
· The maximum voltage between inputs P and N is 1000VDC. Do not measure voltages exceeding the limits given in this manual. Exceeding these limits could result in electrical shocks to the user and damage to the instrument.
· In order to guarantee the operator’s safety, while making the connections, disable the system being measured by means of the switches/breakers upstream and downstream of the DC/AC converter (inverter).
Fig. 4: Connection of instrument for efficiency measurement on a PV installation 1. Check and, if necessary, set the sensitivity of the reference cell on SOLAR-02
consistently with the type of PV modules which will be measured (please refer to the user manual of SOLAR-02). 2. We recommend making a preliminary evaluation of the value of irradiation on the surface of the PV modules being tested by means of unit SOLAR-02 (operating independently) and the reference cell. 3. Switch on PVCHECKs, check and, if necessary, change the instrument settings relevant to the type of remote unit, to the minimum irradiation threshold, to the full scale of the DC clamp, to the integration period and to the parameters of the system being measured (see § 5.1.4, § 5.1.5, § 0, § 5.2.1 and § 5.2.2).
EN – 23
PVCHECKs
4. In order to guarantee the operator’s safety, disable the system being measured by means of the switches/breakers upstream and downstream of the DC/AC converter (inverter).
5. Bring PVCHECKs and SOLAR-02 nearer (maximum distance of 1m between them). All instruments must be switched on (see the User Manual of SOLAR-2 for further details).
6. On PVCHECKS, press the MENU key, select the function EFF and press ENTER; wait for the two units to start communicating with each other. This condition is highlighted by
the contemporary presence of the following indicators:
Symbol Symbol
steady (not flashing) on the display of PVCHECKs steady (not flashing) on the display of SOLAR-02
7. Connect the inputs P and N respectively to the positive and negative output of the string, respecting the colours indicated in Fig. 4
8. Connect the output connector of the DC clamp to the IDC input.
CAUTION
BEFORE CONNECTING THE DC CLAMP TO THE CONDUCTORS Switch on the clamp, check the LED indicating the status of the clamp’s internal batteries (if present), select the correct range, press the ZERO key on the DC clamp and check on the display of PVCHECKs the actual zeroing of the corresponding Idc value (values up to 0.02A are acceptable).
9. Connect the DC current clamp to the positive output conductor of the string, respecting the direction of the arrow found on the clamp itself as indicated in Fig. 4
10. Position the clamp so that the jaw is not near the negative conductor.
11. The display shows the first screen which contains the 15/05/12 15:34:26
values of the output electrical parameters of the I r r
module/string.
Pnom
– – 3.500
12. Before symbol
starting measuring, check for the presence of ” ” steady on the display, which indicated a
Tc Ta Pdc
correct RF connection with the remote unit SOLAR-02.
Vdc Idc
ndc
45 30 3.125 389 8.01 – – –
W/m2 kW °C °C kW V A
Go to Start
Selection
EFF
13. Keeping the remote unit SOLAR-02 always close to the 15/05/12 15:34:26
main unit, press start testing. The
the GO/STOP key on PVCHECKS to message “Waiting start recording…”
Irr Pnom
– – 3.500
appears on the display of the main unit and “HOLD” is shown on the display of SOLAR-02, together with an
Tc Ta Pdc
indication of the time until “00” is reached.
remaining,
expressed
in
seconds,
Vdc Idc ndc
45 30 3.125 389 8.01 – – –
W/m2 kW °C °C kW V A
Waiting start recording…
Selection
EFF
EN – 24
PVCHECKs
14. Upon reaching “00” after pressing the GO/STOP key, the 15/05/12 15:35:00
test is started and the two units are synchronized with I r r
– – –
each other. In these conditions the message “Recording P n o m 3 . 5 0 0
running…” is displayed on the main “Recording…” is displayed by SOLAR-02.
unit
and
Tc Te
Pdc
45 30 3.125
Vdc
389
Idc
8.01
ndc
– – –
W/m2 kW °C °C kW V A
Recording running…
Selection
EFF
15. At any time it will be possible to analyze the current 15/05/12 15:35:00
recording status by pressing the MENU key. The following information will be shown:
Start 15/05/12 15:30:00
Period:
5s
Starting date and time of recording The value set for the integration period (IP)
IP Number Rec. Time
61 0d 1h
The number of periods elapsed from the beginning of the
recording The remaining memory capacity for recording
Recording running…
Press the ESC key to exit the screen 16. Now it is possible to bring the unit SOLAR-02 near the PV strings to measure irradiation and
temperature by means of the relevant probes. When the distance between the unit SOLAR02 and PVCHECKs does not allow a RF connection, on the display of SOLAR-02, the
symbol ” ” flashes for approx. 30s and then disappears, while PVCHECKs continues searching for the connection for approx. 1 minute. 17. Position the reference cell onto the surface of the PV modules. Please refer to the relevant User Manual for a correct assembly. 18. Put the temperature sensor in contact with the rear side of the module and fasten it with some tape; avoid touching it with your fingers in order not to alter the measurement result. 19. Wait for a few seconds to allow the probes to reach a steady measure and then connect the irradiation probe to input PYRA/CELL and the temperature probe to input TEMP of unit SOLAR-02. 20. Wait for the message “READY” to appear on the display of SOLAR-02 to indicate that the unit has detected the data with solar irradiation > minimum threshold set (see § 5.1.5). 21. With the message “READY” shown on the display, wait for approximately 1 minute in order to take a certain number of samples. 22. Disconnect the irradiation and temperature probes from unit SOLAR-02 and bring the unit near PVCHECKs (max distance 1m).
23. The main unit PVCHECKs must be in EFF mode. If there is no flashing symbol ” “, press key to activate the RF connection search again.
24. Press key on SOLAR-02 to activate the RF connection again. Consequently, the main unit will show the message “active radio connection”.
25. To stop testing, press the GO/STOP key on the instrument and confirm with ENTER that you want to stop recording.
26. The display of the unit SOLAR-02 will display the message “SEND” to indicate that the data will be transferred to the main unit.
EN – 25
PVCHECKs
27. After the automatic data transfer phase, the instrument will 15/05/12 15:35:00
display: Do not
show
any
results
if
do
not
exist
on
the
PV
Irr Pnom
712 3.500
W/m2 kW
installation a “stable irradiance” the minimum irradiance threshold
condition
more
than
Tc Te Pdc
Display the best performance values if during the recording, the Irradiance values reached the “stable”
Vdc Idc ndc
45 °C 30 °C 3.125 kW 389 V 8.01 A 0.88
condition and its values were higher than the minimum
irradiance threshold
Analys Results
Unable to carry out the analysis if irradiance has never Selection
EFF
reached a steady value higher than the minimum threshold
set, or if no value exists which is valid during the whole
recording (nDC > 1.15).
28. Press SAVE to save the results obtained (see § 7.1) or ESC to exit the screen and go back
to the initial screen.
EN – 26
PVCHECKs
6.2. MEASURING PARAMETERS ON PV PLANT WITHOUT USING SOLAR-02 The “PV installation efficiency” test carried out without using the optional remote unit SOLAR-02 only allows evaluating output electrical parameters of a string or a photovoltaic field (quantities, Vdc, Idc and Pdc), which can be periodically recorded with a programmable integration period (see § 5.2.1). In this mode, the values of Irradiation, Te, Tc, the nDC efficiency value are not evaluated and no result is provided by the instrument.
CAUTION
· The maximum voltage between inputs P and N is 1000VDC. Do not measure voltages exceeding the limits given in this manual. Exceeding these limits could result in electrical shocks to the user and damage to the instrument.
· In order to guarantee the operator’s safety, while making the connections, disable the system being measured by means of the switches/breakers upstream and downstream of the DC/AC converter (inverter).
Fig. 5: Instrument connection for measuring PV installation parameters without SOLAR-02 1. Select the option “NO” as regards the type of remote unit to be used during EFF test
(see § 5.1.4), set the full scale of the DC clamp (see § 5.2.1), a possible correction factor of the DC clamp (see § 0), the integration period and the rated power of the system (see § 5.2.1 and § 5.2.2). 2. In order to guarantee the operator’s safety, disable the system being measured by means of the switches/breakers upstream and downstream of the DC/AC converter (inverter) 3. Connect the inputs P and N respectively to the positive and negative output of the string, respecting the colours indicated in Fig. 5 4. Connect the output connector of the DC clamp to the IDC input.
CAUTION
BEFORE CONNECTING THE DC CLAMP TO THE CONDUCTORS Switch on the clamp, check the LED indicating the status of the clamp’s internal batteries (if present), select the correct range, press the ZERO key on the DC clamp and check on the display of PVCHECKs the actual zeroing of the corresponding Idc value (values up to 0.02A are acceptable).
EN – 27
PVCHECKs
5. Connect the DC current clamp to the positive output conductor of the string,
respecting the direction of the arrow found on the clamp itself as indicated in Fig. 5. Position the clamp so that the jaw is not near the negative conductor.
6. The display shows the first screen which contains the 15/05/12 15:34:26
values of the module/string.
output
electrical
parameters
of
the I r r
Pnom
– – 3.500
Tc Te Pdc Vdc Idc ndc
– – – – 3.125 389 8.01 – – –
W/m2 kW °C °C kW V A
Go to Start
Selection
EFF
7. Press GO/STOP key on PVCHECKs to start the test. The 15/05/12 15:34:26
message “Waiting start recording…” appears on the I r r
display of the instrument until “00” is reached.
Pnom
– – 3.500
Tc Te Pdc Vdc Idc ndc
45 30 3.125 389 8.01 – – –
W/m2 kW °C °C kW V A
Waiting start recording…
Selection
EFF
8. Upon reaching “00” after pressing the GO/STOP key, the 15/05/12 15:35:00
test is started. Unser these conditions, the message I r r “Recording running…” appears on the instrument’s P n o m
– – 3.500
display.
Tc Te Pdc Vdc Idc ndc
45 30 3.125 389 8.01 – – –
W/m2 kW °C °C kW V A
Recording running…
Selection
EFF
9. At any time it will be possible to analyze the current 15/05/12 15:35:00
recording status by pressing the MENU key. The following information will be shown:
Start : 14/02/00 17:18:00
Period:
5s
Starting date and time of recording The value set for the integration period
IP Number Rec. Time
61 0d 1h
The number of periods elapsed from the beginning of the
recording The remaining memory capacity for recording
Recording running…
Press the ESC key to exit the screen
10. To stop testing, press the GO/STOP key on the instrument and confirm with ENTER that you want to stop recording.
11. Press SAVE to save the results obtained (see § 7.1) or ESC to exit the screen and go back to the initial screen.
EN – 28
PVCHECKs
6.3. QUICK CHECK ON PV MODULES AND STRINGS (IVCK) 6.3.1. Foreword This function carries out a series of quick tests on a PV module/string, measuring in a sequence:
The Voc voltage and the short-circuit current Isc according to the prescriptions of standard IEC/EN62446-1 with the possibility of measuring (by using the relevant probes) also the irradiance and module temperature values.
Insulation resistance measurement (if enabled see § 5.5.1), only carried out in STRING mode (see § 6.4.4) i.e. by automatically producing an internal short circuit between the input terminals P and N and measuring between this short-circuit point and the input terminal E.
Continuity test of protective conductors (if enabled see § 5.5.1) with 200mA between the input terminals E and C of the instrument.
Irradiance measurement can be carried out in one of the following modes:
· Reference cell directly connected to PVCHECKs · Reference cell connected to SOLAR-02 in RF connection with PVCHECKs
Irradiance measurements are always carried out in real time; it is therefore impossible to start a “remote” recording of irradiance values through SOLAR-02.
If the minimum irradiance threshold (see § 5.1.5) is: · = 0 the instrument does not control the presence of the reference cell, irradiance
changes, number of modules and does not display error messages is case it is impossible to calculate Voc and Isc values transposed under STC. This mode is suitable to carry out a test session on a considerable number of strings as quickly as possible. · > 0 (recommended >700W/m2) the instrument carries out all tests provided for I-V test, manages all conditions and error messages for I-V test (wrong num. of mod., temp. out of range, cell presence, min. irr., etc.) and calculated the values of Voc and Isc under STC. This mode is recommended whenever accurate tests are to be carried out on modules/strings being examined.
In general, the result page will include:
· Description of the module used · Irradiance and temperature values (if available) · Average values of Voc and Isc calculated as average of the corresponding values under
OPC on the last 10 tests saved in the memory. If the number of tests is < 10, the average is calculated on the number of available tests. The first test will display dashes in the field “average values” as there are no previous tests on which to calculate an average. · The values of Voc and Isc measured under OPC as well as any partial result (only available if STC values are not available) obtained by comparison with average values. · The values of Voc and Isc calculated under STC (if available) and any partial result obtained by comparing the values calculated under STC with the nominal ones (inserted in DB modules). · The overall test outcome (OK(NO). The overall result will be calculated based on the partial results obtained: Based on partial results under STC (if these are available) Based on partial results under OPC (if STC values are not available)
The instrument will not display any overall result if no partial result is available.
EN – 29
PVCHECKs
6.3.2. Carrying out an IVCK quick test without measuring irradiance
CAUTION
Do not use the instrument for IVCK tests on PV modules with efficiency >19%. Check the technical characteristics of the PV modules before carrying out the tests in order to avoid possible damage to the instrument
CAUTION
· The maximum voltage between inputs P, N, E and C is 1000VDC. Do not measure voltages exceeding the limits given in this manual
· Do not perform test on PV modules or strings connected to the DC/AC converter · The maximum current which can be measured by the instrument is 15A.
Before carrying out the measures of IVCK and Insulation in “STRING” mode, always ensure that the instrument is connected to ONE STRING and not more strings in parallel to avoid possible damage of it
1. Switch on the instrument by pressing the ON/OFF key.
2. Check that the remote unit SOLAR-02 is not selected (see § 5.1.4 setting NO)
3. Check that the value of minimum irradiance set in section “Irradiance” (see § 5.1.5) is
equal to 0.
2. Position the cursor onto IVCK by using the arrow keys 15/05/12 15:34:26
(,) and confirm with ENTER. The display shows the screen to the side: The meaning of the parameters is the
Module Vdc
Irr
: SUNPWR318 : 0.0 V
: – – – W/m2
following:
Tc
: —
Module type of module being tested
Voc,Isc:
Vdc value of output measured in real time
voltage
from
module/string
Ri(1000V)
– – – M
Irr value of irradiation measured in real time
Rpe (Cal)
—
Tc value of module temperature (see § 5.5.1) Voc, Isc section with display of result OK/NO of
Voc and Isc measurement
Selection
IVCK
Ri() the value in brackets may be NO/selected test
voltage (see § 5.5.1). The value of Ri indicates the
insulation resistance
Rpe() the value in brackets may be NO, Cal or
NoCal (see § 5.5.1). The value of Rpe indicates the
result of the continuity test.
3. Press the ENTER key, select “Settings” and confirm again with ENTER. Set the
instrument as indicated in § 5.5.1
4. If necessary, press the ENTER key, select “Reset Averages” and confirm again with
ENTER. If necessary, carry out this operation as indicated in § 6.3.4
5. If necessary, press the ENTER key, select “Leads calibration” and confirm again with
ENTER. If necessary, carry out this operation as indicated in § 6.5.2
6. Connect the instrument to the module/string being tested and, if necessary, to the main
earth node of the system and to the earthed metal masses as shown in Fig. 6. In
particular, connect the negative output pole of the module/string to terminal N and the
positive output pole of the module/string to terminal P.
EN – 30
PVCHECKs
CAPTION:
E: Green cable C: Blue cable P: Red cable N: Black cable
1. PV module/string 2. Main system earthing 3. Earthed metal structure in
the system
Fig. 6: Connection for IVCK test without irradiance measurement
CAUTION
Upon pressing the GO/STOP key, different error messages can be displayed by the instrument (see § 6.6) and, therefore, the test cannot be started. Check and eliminate, if possible, the problem causing the error message before going on with the test.
9. Press GO/STOP to start the test. In case no error15/05/12 15:34:26
conditions occur, the instrument “Measuring…” and the measure
displays the message of open-circuit voltage
Mod Vdc
Irr
u
l
e
: SUNPWR318
: 548.0 V
: 0
W/m2
between the terminals P and N and of short-circuit current T c
: Auto °C
(for Isc values 15A).
Voc,Isc:
Ri(1000V)
– – – M
Rpe (Cal)
—
Measuring…
Selection
IVCK
10.When Voc and Isc measurements are complete, the15/05/12 15:34:26
message “OK” is shown in case the result of positive (measured values within the tolerances
the set
test is on the
Modu Vdc
Irr
l
e
instrument)
Tc
: SUNPWR318 : 548.0 V
: 0
W/m2
: Auto °C
11.With insulation measurement selected, the instrument V o c , I s c : goes on with the test keeping terminals P and N shortcircuited, and carrying out the test between this point and R i ( 1 0 0 0 V )
116
OK M OK
terminal E for the time necessary to obtain a steady value. R p e ( C a l )
—
12.The value of insulation resistance is shown in field “Ri”
and the message “OK” appears in case the result of the
Measuring…
test is positive (measured value higher than the minimum Selection
IVCK
limit value set on the instrument).
EN – 31
PVCHECKs
13.With continuity measurement selected, the instrument15/05/12 15:34:26
goes on by opening the test between terminals E
short-circuit and C.
and
carrying
out
the
Module Vdc
Irr
: SUNPWR318
: 548.0 V
: 0
W/m2
14.The value of resistance in the continuity test is shown in T c
: Auto °C
field “Rpe” and the message “OK” appears in case the V o c , I s c :
result of the test is positive (measured value lower than
the maximum limit value set on the instrument).
Ri(1000V)
116
OK M OK
15.The message “Outcome:OK” is finally shown by the R p e ( C a l )
2.00 OK
instrument in case the result of all performed tests is
positive.
Outcome: OK
IVCK
16.Press the arrow key to show the following page in 15/05/12 15:34:26
which the values of parameters Voc and Isc are shown.
Module: Irr
It shows:
Tc (AUTO)
SUNPWR210 —W/m2 –°C
The module being used The average values of
Voc
and
Isc
under
OPC
VocAvg@OP C IscAvg@OPC
647 V 5.43 A
conditions The values of Voc and Isc measured under OPC and
Voc@OPC Isc@OPC Voc@STC
647V OK 5.35A OK
—V
the relevant partial results obtained by comparison with the average values.
Isc@STC
—
Outcome: OK
IVCK
A
In general:
EsitoVoc@OPC = OK
se 100 VocAvg@OPC -Voc@OPC (Tol Voc + 4%)
V ocA v g@ OPC
Esito Isc@OPC = OK
se 100 IscAvg@OPC – Isc@OPC (Tol Isc+ 4%)
IscA v g@ OPC
The overall value of the results:
o OK: if all results under OPC are OK, o NO if one of the results under OPC is NO
17.Press the key to go back to the previous screen. 18.Press the SAVE key to store the test result in the instrument’s memory (see § 7.2) or
the ESC/MENU key to exit the screen without saving and go back to the main measuring screen.
CAUTION
The average values of Voc and Isc are displayed in the results’ page. These values include the average values of Voc and Isc under OPC conditions calculated as mean of the last 10 tests previously saved. If the user has carried out and saved a number of tests <10 or reset the average values (see § 6.3.4) the average value displayed during test N+1 will be calculated on the available N values.
EN – 32
PVCHECKs
6.3.3. Carrying out an IVCK quick test and measuring irradiance
CAUTION
Do not use the instrument for IVCK tests on PV modules with efficiency >19%. Check the technical characteristics of the PV modules before carrying out the tests in order to avoid possible damage to the instrument
CAUTION
· The maximum voltage between inputs P, N, E and C is 1000VDC. Do not measure voltages exceeding the limits given in this manual.
· Do not perform test on PV modules or strings connected to the DC/AC converter
· The maximum current which can be measured by the instrument is 15A. Before carrying out the measures of IVCK and Insulation in “STRING” mode, always ensure that the instrument is connected to ONE STRING and not more strings in parallel to avoid possible damage of it
1. Switch on the instrument by pressing the ON/OFF key. 2. Irradiance measurement is carried out in one of the following modes:
Measurement by reference cell directly connected to PVCHECKs Measurement by reference cell connected to SOLAR-02 unit in RF with PVCHECKs
3. Check that the settings of remote unit SOLAR-02 are consistent with the type of
measurement to be carried out (see § 5.1.4). 4. Check the minimum irradiance value set (see § 5.1.5).
5. Position the cursor onto IVCK by using the arrow keys 15/05/12 15:34:26
(,) and confirm with ENTER. The display shows the screen to the side: The meaning of the parameters is the
Module Vdc Irr
: SUNPWR318
: 0.0 V
: 0
W/m2
following:
Tc
: Auto °C
Module type of module being tested
Voc,Isc:
Vdc value of output measured in real time
voltage
from
module/string
Ri(1000V)
– – – M
Irr value of irradiation measured in real time
Rpe (Cal)
—
Tc value of module temperature (see § 5.5.1)
Voc, Isc section with display of result OK/NO of
Voc and Isc measurement
Selection
IVCK
Ri() the value in brackets may be NO/selected test
voltage (see § 5.5.1). The value of Ri indicates the
insulation resistance Rpe() the value in brackets may be NO, Cal or
NoCal (see § 5.5.1). The value of Rpe indicates the
result of the continuity test.
6. Press the ENTER key, select “Settings” and confirm again with ENTER. Set the instrument as indicated in § 5.5.1
7. If necessary, press the ENTER key, select “Average Reset” and confirm again with ENTER. Carry out this operation as indicated in § 6.3.4
8. If necessary, press the ENTER key, select “Cable calibration” and confirm again with ENTER. Carry out this operation as indicated in § 6.5.2
9. Mount the rod onto the disc of the optional accessory M304 and keep it in touch with
the module’s surface. Make sure that the shadow cast by the rod onto the disc falls within the “limit internal circle” on the disc itself (see Fig. 7). If this is not the case, the incidence angle between the sun rays and the module surface is too high and,
therefore, the measurements carried out by the instrument CANNOT be considered
reliable. Repeat the same operations at another daytime.
EN – 33
PVCHECKs
10. Fasten the bracket to the module by using the set of screws supplied with the HT304k optional reference cell and mount the cell on it, if possible, with output terminals facing downwards. Rotate the cell until it rests on the tang found on the bracket so that the cell is perfectly parallel to the surface of the module, then fix it with the relevant screws provided.
Fig. 7: Positioning the inclinometer M304 11. Connect the output of the cell corresponding to the type of module being tested to the
IRR. input of the instrument using the cable supplied with the cell, or to the PYRA/CELL input of the remote unit SOLAR-02, if used (see Fig. 8 and Fig. 9) 12. If used, connect the temperature sensor to the TEMP input of the instrument and fasten it to the backside of the module under a cell using some adhesive tape. As an alternative, connect the sensor to the TEMP input of the remote unit SOLAR-02, if used (see Fig. 8 and Fig. 9) 13. Connect the instrument to the module/string being tested and, if necessary, to the main earth node of the system and to the earthed metal masses as shown in Fig. 8 and Fig. 9. In particular, connect the negative output pole of the module/string to terminal N and the positive output pole of the module/string to terminal P.
CAUTION
In case the remote unit SOLAR-02 is used for measuring irradiance, make sure that the RF communication with the master unit PVCHECKs is always active (symbol ” ” on and steady on the display).
CAPTION:
E: Green cable C: Blue cable P: Red cable N: Black cable
1. PV module/string 2. Main system earthing 3. Earthed metal structure in the
system 4. Reference cell for irradiance
measurement 5. Temperature sensor (if
required)
EN – 34
PVCHECKs
Fig. 8: Connection for IVCK test with direct irradiance measurement CAPTION:
E: Green cable C: Blue cable P: Red cable N: Black cable
1. PV module/string
2. Main system earthing
3. Earthed metal structure in the
system
4. Reference cell for irradiance
measurement
5. Temperature sensor (if
required)
6. Unità
remota
SOLAR-
02Remote unit SOLAR-02
Fig. 9: Connection for IVCK test with irradiance measurement through SOLAR-02
14.In the initial screen of IVCK mode, the following values 15/05/12 15:34:26
are displayed in real time: Module type of module being tested
Module Vdc
Irr
Vdc value of output voltage from module/string
Tc
Irr irradiance (from a direct measurement or V o c , I s c :
: SUNPWR318 : 548.0 V : 856 W/m2
: Auto °C
SOLAR-02 connected via RF) Tc modulo temperature (in MAN or AUX mode) and
Ri(1000V)
– – – M
the relevant measuring mode, or “- – -” in AUTO mode R p e ( C a l )
—
In case, the symbol ” ” of RF connection with
SOLAR-02
Selezione
IVCK
CAUTION
Upon pressing the GO/STOP key, different error messages can be displayed by the instrument (see § 6.6) and, therefore, the test cannot be started. Check
and eliminate, if possible, the problem causing the error message before going
on with the test.
15.Press GO/STOP to start the test. In case no error15/05/12 15:34:26
conditions occur, the instrument “Measuring…” and the measure
displays the message of open-circuit voltage
Mod Vdc
Irr
ule
: SUNPWR318 : 548.0 V
: 856 W/m2
between the terminals P and N and of short-circuit current T c
: Auto °C
(for Isc values 15A).
Voc,Isc:
Ri(1000V)
– – – M
Rpe (Cal)
—
Measuring…
Selection
IVCK
EN – 35
PVCHECKs
16.When Voc and Isc measurements are complete, the15/05/12 15:34:26
message “OK” is shown in case the result of positive (measured values within the tolerances
the set
test is on the
Mod Vdc
Irr
ule
instrument)
Tc
: SUNPWR318 : 548.0 V : 856 W/m2
: Auto °C
17.With insulation measurement selected, the instrument V o c , I s c : goes on with the test keeping terminals P and N shortcircuited and carrying out the test between this point and R i ( 1 0 0 0 V )
116
OK M OK
terminal E for the time necessary to obtain a steady value. R p e ( C a l )
—
18.The value of insulation resistance is shown in field “Ri” and
the message “OK” appears in case the result of the test is
Measuring…
positive (measured value higher than the minimum limit Selection value set on the instrument).
IVCK
19.With continuity measurement selected, the instrument15/05/12 15:34:26
goes on by opening the test between terminals E
short-circuit and C.
and
carrying
out
the
Module Vdc
Irr
20.The value of resistance in the continuity test is shown in T c
: SUNPWR318 : 548.0 V : 856 W/m2
: Auto °C
field “Rpe” and the message “OK” appears in case the V o c , I s c :
result of the test is positive (measured value lower than
the maximum limit value set on the instrument).
Ri(1000V)
116
OK M OK
21.The message “Outcome:OK” is finally shown by the R p e ( C a l )
2.00 OK
instrument in case the result of all performed tests is
positive.
Outcome: OK
IVCK
22.Press the arrow key to show the following page in 15/05/12 15:34:26
which the values of parameters Voc and Isc are shown. It shows:
Module: Irr Tc (AUTO)
SUNPWR210 —W/m2 57°C
The module being used The value of irradiance
VocAvg@OPC IscAvg@OPC Voc@OPC
647V 5.43 A 647V
The value of temperature of the module The average values of Voc and Isc under OPC
Isc@OPC Voc@STC Isc@STC
5.35 A 787V OK 5.72A OK
The values of Voc and Isc measured under OPC The values of Voc and Isc calculated under STC and the
Outcome: OK IVCK
relevant partial results obtained by comparison with the
nominal values.
In general:
( ) Esito Voc@STC = OK
se
100 VocNom@STC – Voc@STC Tol Voc + 4% VocNom@S TC
( ) Esito Isc@STC = OK
se
100 IscNom@STC – Isc@STC Tol Isc + 4% IscNom@S TC
The rated values of Voc and Isc are the values found in the instrument’s internal modules’ DB (see § 5.6).
The overall value of the results: o OK: if all results under STC are OK, o NO if one of the results under STC is NO
23.Press the key to go back to the previous screen.
EN – 36
PVCHECKs
24.Press the SAVE key to store the test result in the instrument’s memory (see § 7.2) or the ESC/MENU key to exit the screen without saving and go back to the main measuring screen.
CAUTION
The average values of Voc and Isc are displayed in the results’ page. These values include the average values of Voc and Isc under OPC conditions calculated as mean of the last 10 tests previously saved. If the user has carried out and saved a number of tests <10 or reset the average values (see § 6.3.4) the average value displayed during test N+1 will be calculated on the available N values.
6.3.4. Reset Averages If irradiance values are measured, the instrument provides a result by comparing the values measured with the average values calculated according to the previously saved measured. In this case, the average values calculated by the instrument are therefore particularly important. In case a new measurement campaign is started with significant variations in irradiance or temperature, it is advisable to set the average reference values to zero in order to make new calculations based on new measurements. To reset the average values, please keep to the following steps:
1. In IVCK mode, press the ENTER key, select “Reset 15/05/12 15:34:26
Averages” and confirm by pressing ENTER again to zero the average values calculated until that moment.
Module Vdc
Irr
: SUNPWR318
: 0.0 V
: 0
W/m2
Tc
: Auto °C
Voc,Isc:
Ri(1000V)
– – – M
Rpe (Cal)
Leads calibration Reset Averages Settings Selection
—
IVCK
The average values are automatically reset also modifying and then saving one of the following parameters:
Type of PV module Number of modules per string
The average values are not reset if the operator changes operating mode and then goes back to this mode.
EN – 37
PVCHECKs
6.3.4.1. Anomalous situations when performing IVCK tests
1. In case the instrument detects a voltage higher than 15/05/12 15:34:26
1000V at terminals P-N, P-E
and
N-E, it does not
carry
Module Vdc
out the test, gives out a long sound and displays the I r r
: SUNPWR318
: 0.0 V
: 0
W/m2
message “Vin > 1000”.
Tc
: Auto °C
Voc,Isc:
Ri(1000V)
– – – M
Rpe (Cal)
—
Vin > 1000
Selection
IVCK
2. In case the instrument detects a voltage lower than 15V 15/05/12 15:34:26
at terminals P and N, it does not carry out the test, gives
Module Vdc
out a long sound and displays the message “Low I r r
: SUNPWR318
: 0.0 V
: 0
W/m2
voltage”.
Tc
: Auto °C
Voc,Isc:
Ri(1000V)
– – – M
Rpe (Cal)
—
Low voltage
Selection
IVCK
3. In case the instrument detects a voltage higher than > 5V 15/05/12 15:34:26
at terminals E and C, it does not carry out the test, gives out a long sound and displays the message “Voltage >
Module Vdc
Irr
: SUNPWR318
: 0.0 V
: 0
W/m2
Lim”.
Tc
: Auto °C
Voc,Isc:
Ri(1000V)
– – – M
Rpe (Cal)
—
Voltage > Limit
Selection
IVCK
4. In case the instrument detects an Isc current higher than 15/05/12 15:34:26
15A, it does not carry out the test, gives out a long sound
Module Vdc
and displays the message “Isc current too high”.
Irr
: SUNPWR318
: 0.0 V
: 0
W/m2
Tc
: Auto °C
Voc,Isc:
Ri(1000V)
– – – M
Rpe (Cal)
—
Isc current too high
Selection
IVCK
EN – 38
PVCHECKs
6.4. MEASUREMENT OF INSULATION ON PV MODULES/STRINGS/FIELDS (M) 6.4.1. Foreword The purpose of this function is to measure the insulation resistance of the active conductors of a module, string, of a whole PV field and of possible unearthed metal masses according to the prescriptions IEC/EN62446-1 guideline. In general, the instrument measures insulation in the following modes:
FIELD mode used for measuring insulation resistance of a PV field (photovoltaic generator) consisting in one or more strings connected in parallel. The instrument performs the measurement on the Positive and Negative poles of the PV field.
TIMER mode the instrument measures continuously (with a max duration of 300s) on terminal “P” only, displaying the minimum resistance value obtained at the end of the selected time. It may be used for measuring the insulation resistance of more unearthed metal masses.
STRING mode used for measuring insulation only on single PV modules or strings, automatically creating an internal short-circuit between the Positive and Negative poles, with no need to use an external switch to short-circuit the positive and negative terminals and measuring between this short-circuit point and the installation earthing.
6.4.2. Measuring insulation FIELD mode
1. Position the cursor onto M by using the arrow keys 15/05/12 15:34:26
(,) and confirm with ENTER. The display shows the I n s . T e s t : 1 0 0 0 V
screen to the side:
Ri min
: 1.0
M
2. Press the ENTER key, activate “Settings” and possibly M o d e
: Field
change the desired parameters (see § 5.4). The following V t e s t – – – V – – – V
parameters are shown on the display:
Ri(+)
– – – M
Ins. Test selected test voltage (250, 500 or R i ( – )
– – – M
1000VDC)
Rp
Ri min minimum limit threshold for insulation
– – – M
measurement
Selection
Mode measuring mode: FIELD
Vtest real test voltages applied respectively
M
between the Positive pole and the Negative pole of 15/05/12 15:34:26
the field respect the earthing
Ri (+) insulation resistance measurement between the Positive pole of the PV field and the earthing Ri (-) insulation resistance measurement between
VPN VEP VEN
– – – – – – –
V V V
the Negative pole of the PV field and the earthing Rp final value of the measurement obtained by the
parallel of Ri(+) e Ri(-) values which is compared with
the Ri min limit by the instrument key access to the second
page
with
the
Selection
M
measured values of VPN, VEP and VEN voltages
3. Connect the instrument to the PV field being tested and to the main earth node of the system shown in Fig. 10. In particular, connect the negative output pole of the PV field to terminal N and the positive output pole of the PV field to terminal P.
EN – 39
PVCHECKs
CAPTION:
E: Green cable P: Red cable N: Black cable
1. Unearthed PV field 2. Main system earthing
Fig. 10: Instrument connection for insulation measurement in FIELD mode
CAUTION
Upon pressing the GO/STOP key, different error messages can be displayed by the instrument (see § 6.6) and, therefore, the test cannot be started. Check and eliminate, if possible, the problem causing the error message before going on with the test.
4. Press GO/STOP to start the test. In case no error 15/05/12 15:34:26
conditions occur, the instrument displays the message I n s . T e s t : 1 0 0 0 V
“Measuring…” as shown in the screen to the side.
Ri min
: 1.0
M
Mode
: Field
Vtest 1043 1057 V
Ri(+)
– – – M
Ri(-)
– – – M
Rp
– – – M
Measuring…
Selection
M
5. When measurement (with a fixed duration of 10s) is 15/05/12 15:34:26
complete, the instrument provides the values Ri (+) and T e s t I s o Ri (-), respectively the insulation resistance of the R l i m
: 1000 V
: 1.0
M
Positive and Negative poles of the PV field being tested. M o d o
: Campo
If both results are higher than the minimum limit set, the V t e s t 1 0 4 3 1 0 5 7 V
instrument displays the message “Outcome:OK”; R i ( + )
> 100 M
otherwise, it displays the message “Outcome: NO” as R i ( – )
6.
shown in the screen to the side. Press the SAVE key to store
the
test
result
in
the
Rp
> 100
69 Outcome: OK
M M
instrument’s memory (see § 7.2) or the ESC/MENU key to exit the screen without saving and go back to the main
M
measuring screen.
EN – 40
PVCHECKs
6.4.3. Measuring insulation TIMER mode
1. Position the cursor onto M by using the arrow keys 15/05/12 15:34:26
(,) and confirm with ENTER. The display shows the I n s . T e s t : 1 0 0 0 V
screen to the side.
Ri min
: 1.0
M
2. Press the ENTER key, activate “Settings” and possibly M o d e
: Timer
change the desired parameters (see § 5.4). The following V t e s t
— V
parameters are shown on the display:
Ri(+)
– – – M
Ins. Test selected test voltage (250, 500 or T e s t t i m e : 1000VDC)
Ri min minimum limit threshold for insulation Selection measurement
Mode measuring mode: TIMER Vtest real applied test voltage Ri (+) minimum insulation resistance between the
Positive pole and the earthing detected during the whole duration of measurement. Test time test duration which can be set within a
range of 10 300s key access to the second page with the
measured values of VPN, VEP and VEN voltages
10s M
3. Connect the instrument to the possible unearthed metal masses and to the main earth node of the system (see Fig. 11)
CAPTION:
E: Green cable P: Red cable N: Black cable
1. Unearthed PV module/string
2. Main system earthing
3. Unearthed metal mass
Fig. 11: Instrument connection for insulation measurement in TIMER mode
CAUTION
Upon pressing the GO/STOP key, different error messages can be displayed by the instrument (see § 6.6) and, therefore, the test cannot be started. Check and eliminate, if possible, the problem causing the error message before going on with the test.
EN – 41
PVCHECKs
4. Press GO/STOP to start the test. In case no error 15/05/12 15:34:26
conditions occur, the instrument displays the message “Measuring…” as shown in the screen to the side.
Ins. Test Ri min
: 1000 : 1.0
V M
Mode
: Timer
Vtest
1020 V
Ri(+)min
– – – M
Test time: 10s
Measuring…
Selection
M
5. When measurement is complete, the instrument provides 15/05/12 15:34:26
the Ri(+)min value, i.e. the minimum value of insulation I n s . T e s t : 1 0 0 0 V
resistance of the PV module/string (or of other metal R i m i n
: 1.0
M
masses) being tested, continuously measured during the M o d e
: Timer
whole duration of measurement. If the result is higher V t e s t
= 1020 V
than the minimum limit set, the instrument displays the R i ( + ) m i n = > 2 0 0 M message “Outcome:OK”; otherwise, it displays the
message “Outcome:NO” as shown in the screen to the side.
Test
time: 200s Outcome: OK
6. Press the SAVE key to store the test result in the
M
instrument’s memory (see § 7.2) or the ESC/MENU key
to exit the screen without saving and go back to the main
measuring screen.
EN – 42
PVCHECKs
6.4.4. Measuring insulation STRING mode
CAUTION
· The maximum current which can be measured by the instrument is 15A. Before carrying out the measures of Insulation in “STRING” mode, always ensure that the instrument is connected to ONE STRING and not more strings in parallel to avoid possible damage of it
· ALWAYS disconnect the string on test from the inverter before performing the measurement
1. Position the cursor onto M by using the arrow keys 15/05/12 15:34:26
(,) and confirm with ENTER. The display shows the I n s . T e s t : 1 0 0 0 V
screen to the side.
Rl min
: 1.0
M
2. Press the ENTER key, activate “Settings” and possibly M o d e
: String
change the desired parameters (see § 5.4). The following V t e s t
— V
parameters are shown on the display:
Rp
– – – M
Ins. Test selected test voltage (250, 500 or
1000VDC)
Ri min minimum limit threshold for insulation
measurement
Selection
Mode measuring mode: STRING
Vtest real applied test voltage
Rp final value of the measurement obtained by the
M
parallel of Ri(+) e Ri(-) values which is compared with
the Ri min limit by the instrument key access to the second page with the
measured values of VPN, VEP and VEN voltages
3. Connect the instrument to the PV module/string being tested and to the main earth
node of the system as shown in . In particular, connect the negative output pole of the
PV field to terminal N and the positive output pole of the PV field to terminal P.
CAPTION:
E: Green cable P: Red cable N: Black cable
1. Unearthed PV module/string 2. Main system earthing
Fig. 12: Instrument connection for insulation measurement in STRING mode EN – 43
PVCHECKs
CAUTION
Upon pressing the GO/STOP key, different error messages can be displayed
by the instrument (see § 6.6) and, therefore, the test cannot be started. Check
and eliminate, if possible, the problem causing the error message before going
on with the test.
4. Press GO/STOP to start the test. In case no error 15/05/12 15:34:26
conditions occur, the instrument displays the message “Measuring…” as shown in the screen to the side.
Ins. Test Ri min
: 1000 : 1.0
V M
Mode
: Stringa
Vtest
= 1020 V
Rp
= – – – M
Measuring…
Selection
M
5. When measurement is complete, the instrument provides 15/05/12 15:34:26
the Ri value, resistance of
i.e. the
the of
minimum value of insulation other objects being tested,
Ins. Test Ri min
: 1000 : 1.0
V M
continuously measured during the whole duration of M o d e
: Stringa
measurement. If the result is higher than the minimum V t e s t
= 1020 V
limit set, the instrument displays the message R p “Outcome:OK”; otherwise, it displays the message
= >100 M
“Outcome:NO” as shown in the screen to the side.
6. Press the SAVE key to store the test result in the instrument’s memory (see § 7.2) or the ESC/MENU key to exit the screen without saving and go back to the main
Outcome: OK M
measuring screen.
EN – 44
PVCHECKs
6.4.4.1. Anomalous situations
1. In any operating mode, in case the instrument detects a 15/05/12 15:34:26
voltage higher than 1000V at terminals P-N, P-E and N- I n s . T e s t : 1 0 0 0 V
E, it does not carry out the test, gives out a long sound R i m i n
: 1.0
M
and displays the message “Vin > 1000”.
Mode
: String
Vtest
= — V
Rp
= – – – M
Vin > 1000
Selection
M
2. In STRING mode, in case the instrument detects an Isc 15/05/12 15:34:26
current higher than 15A, it does not carry out the test, I n s . T e s t : 1 0 0 0 V
gives out a long sound and displays the message “Isc R i m i n
: 1.0
M
current too high”.
Mode
: Stringa
Vtest
= — V
Rp
= – – – M
Isc current too high
Selection
M
3. In STRING mode, in case the instrument detects a 15/05/12 15:34:26
current < 0.2A between terminals P and N, it does not carry out the test, gives out a long sound and displays
Ins. Test Ri min
: 1000 : 1.0
V M
the message “Current < Lim”.
Mode
: String
Vtest
= — V
Rp
= – – – M
Current < Lim
Selection
M
4. In STRING mode, in case the instrument detects a 15/05/12 15:34:26
voltage < 15V carry out the
between terminals P and N, it does not test and displays the message “Low
Ins. Test Ri min
: 1000 : 1.0
V M
voltage”.
Mode
: String
Vtest
= — V
Rp
= – – – M
Low voltage
Selection
M
EN – 45
PVCHECKs
6.5. MEASUREMENT OF CONTINUITY ON PV MODULES/STRINGS/FIELDS (LOW) 6.5.1. Foreword The purpose of this measurement is to test the continuity of protective conductors and
equipotential ones (e.g. from rod to earth and connected foreign earth) and earth rods of
SPDs on PV installations. The test must be carried out using a test current > 200mA
according to the prescriptions of IEC/EN62446-1 guideline
6.5.2. Calibration of measuring cables
1. Position the cursor onto LOW by using the arrow keys 15/05/12 15:34:26
(,) and confirm with ENTER. The display shows the R p e m a x :
1
screen to the side.
Rcal
: —
Fig. 13: Compensation of the measuring cables’ resistance Connect the measuring cables as shown in Fig. 13:
Rpe
= —
Compensation of the measuring cables’ resistance
Itest
= – – – mA
2.
Selection
LOW
Fig. 13: Compensation of the measuring cables’ resistance
2. Press the ENTER key. The instrument shows the 15/05/12 15:34:26
3.
following options: Settings and Leads calibration Use the arrow keys (,) to select “Leads calibration”
RPE max Rcal
: :
1 —
and confirm with ENTER.
Rpe
= —
Itest
= – – – mA
Leads calibration Settings Selection
LOW
4. Press GO/STOP to start calibration. The message 15/05/12 15:34:26
“Measuring…” is shown on the display.
RPE max :
1
5. At the end of the compensation procedure, in case the R c a l
: —
measured value of resistance is lower than 5, the instrument gives out a double sound to signal the positive
Rpe
= —
result of the test. The following screen appears on the I t e s t
= – – – mA
display:
Measuring…
Selection
LOW
EN – 46
PVCHECKs
6. The value of compensated resistance of the cables which 15/05/12 15:34:26
will be subtracted measures is shown
from all subsequent next to “Rcal” and the
continuity message
RPE max Rcal
: :
1.0 0.02
7.
“Calibration OK” appears on the display. To delete the value of compensated resistance, carry out a
Rpe
= —
new compensation procedure with a resistance higher than I t e s t
= – – – mA
5 such as, for example, with open leads. The value in Rcal
is zeroed on the display.
Calibration OK
Selection
LOW
6.5.3. Measuring continuity
1. Position the cursor onto LOW by using the arrow keys 15/05/12 15:34:26
(,) and confirm with ENTER. The display shows the R P E m a x : 1 . 0
screen to the side.
Rcal
: —
2.
Press the ENTER key, activate “Settings” change the desired parameters (see §
and possibly 5.3.1). The
Rpe
following parameters are shown on the display:
Itest
= — = – – – mA
RPE max maximum threshold for continuity
measurement, which can be selected in a range 1
5 in steps of 1 Rcal the value
of
resistance
of
the
measuring
Selection
cables after calibrating them
Rpe the result of continuity measurement
Itest real test current
LOW
3. Press the ENTER key, activate “Cable calibration” (see § 6.5.2) to carry out the initial calibration of the measuring cables.
4. Connect the instrument to the PV module/string being tested and to the main earth
node of the system as shown in Fig. 14
CAPTION:
E: Green cable C: Blue cable
1. PV module/string 2. Main system earthing 3. Earthed metal structure in the
system
Fig. 14: Connection for continuity measurement on structures of the PV installation EN – 47
PVCHECKs
CAUTION
Upon pressing the GO/STOP key, different error messages can be displayed by the instrument (see § 6.6) and, therefore, the test cannot be started. Check and eliminate, if possible, the problem causing the error message before going on with the test.
5. Press GO/STOP to start the test. In case no error 15/05/12 15:34:26
conditions occur, the instrument displays the message R P E m a x : 1 . 0
“Measuring…” as shown in the screen to the side.
Rcal
: —
Rpe
= —
Itest
= – – – mA
Measuring…
Selection
LOW
6. At the end of measurement, the instrument provides the 15/05/12 15:34:26
value of resistance of the object being tested. If the result R p e m a x : 1 . 0
is lower than the maximum limit set, the instrument R c a l
: —
displays displays
the message “Outcome:OK”; the message “Outcome:NO” as
otherwise, it shown in the
Rpe
= 0.23
screen to the side.
Itest
= 210 mA
7. Press the SAVE key to store the test result in the
instrument’s memory (see § 7.2) or the ESC/MENU key to
exit the screen without saving and go back to the main
Outcome: OK
measuring screen.
Selection
LOW
EN – 48
PVCHECKs
6.5.3.1. Anomalous situations
1. In case the instrument detects a voltage higher than 5V at 15/05/12 15:34:26
its terminals E and C, it does not carry out the test, gives R P E m a x :
1
out a long sound and displays the message “Voltage > R c a l
: —
Lim”.
Rpe
= —
Itest
= – – – mA
Voltage > Lim
Selection
LOW
2. In case the instrument detects that the calibrated 15/05/12 15:34:26
resistance instrument
is higher that the measured resistance, gives out a long sound and displays
the the
RPE max Rcal
: :
1 —
message: “Calibration not OK”
Rpe
= —
Itest
= – – – mA
Calibration not OK
Selection
LOW
3. In case the instrument detects a resistance higher than 15/05/12 15:34:26
5 at its terminals, it gives out a long sound, zeroes the R P E m a x :
1
compensated value and displays the message R c a l
: 0.00
“Calibration reset”.
Rpe
= —
Itest
= – – – mA
Calibration reset
Selection
LOW
EN – 49
PVCHECKs
6.6. LIST OF DISPLAY MESSAGES
MESSAGE Function not available
Data not saved Wrong date
RADIO transmission error SOLAR-02: Firmware incorrect
Firmware incorrect Error 4: contact service
Database full Module already added
Memory full Error: Vmpp >= Voc Error: Impp >= Isc Error: Vmpp * Impp >= Pmax Error: Alpha too high Error: Beta too high Error: Gamma too high Error: Toll too high
Wait data analysis
Download error Error while saving Remote unit undetected Unable to perform analysis Data unavailable Negative voltage
Low voltage Vin > 1000 Wrong no. of modules. Continue? Ref. cell temp. exceeding limits Cell temp. undetected (ENTER/ESC) Low battery Please wait for cooling… Irradiance too low NTC Error Isc current too high Current < Lim EEPROM error: contact service FRAM error: contact service RTC Error: contact service RADIO error: contact service FLASH error: contact service IO EXP error: contact service Voltage > limit Label already assigned
Isc current < Lim
CAUTION: internal short-circuit CAUTION: blown fuse
Reset calibration. Press ENTER. Calibration not OK
Error: Isc offset measurement Rcal > measured R
CAUTION: AC voltage at P-N terminals
Wait for condenser discharge
DESCRIPTION The selected function/characteristic is not available
The instrument was not able to save the data Set a correct system date
The instrument does not communicate via RF with external units Inconsistent SOLAR-02 FW. Update firmware. Unsuitable instrument FW. Update firmware. Internal instrument error
The number of modules added to the internal DB is > 30 The name of the module has already been added to the DB
Instrument memory full upon pressing the GO key Check module settings inside DB Check module settings inside DB Check module settings inside DB Check module settings inside DB Check module settings inside DB Check module settings inside DB Check module settings inside DB
Download data from SOLAR-02 and wait for the result of PV efficiency test
Contact service department Problems with the access to memory area The instrument does not detect any remote unit SOLAR-02 Problems with the data downloaded from SOLAR-02. Check settings.
Generic error. Repeat the test. Check the polarity of the instrument’s input terminals Check the voltage between input terminals P and N
Voltage between input terminals > 1000V Set no. of modules inconsistent with measured Voc Temperature measured by reference cell too high
Measurement not carried out on module cell
Low battery charge. Insert new batteries. Instrument overheated. Wait before resuming the test. Irradiance value lower than the minimum threshold set Internal NTC efficiency jeopardized. Contact service department
Measured Isc current > 15A Current between P and N lower than the minimum detectable value
Internal instrument error Internal instrument error Internal instrument error Internal instrument error Internal instrument error Internal instrument error Voltage between terminals E and C > 10V Change the numeric reference of the marker associated with the measure Isc current lower than the minimum detectable value. Contact service
department Contact service department Contact service department Resistance value of input cables > 2 Calibrated resistance value > measured resistance
Internal instrument error Calibrated resistance value > measured resistance
Presence of AC input voltage
Wait for discharge of tested object after insulation
EN – 50
PVCHECKs
7. STORING DATA
The instrument allows saving max 999 measured values. The saved data can be recalled at display and deleted at any moment, and can be associated to reference numerical markers relevant to the installation name, the PV string and the PV module (max 250 markers).
7.1. SAVING EFFICIENCY MEASURES
1. Press the SAVE key with a measured result shown on the 15/05/12 15:34:26
display. The instrument shows containing a virtual keyboard.
the
screen
to
the
side,
Irr Pnom
712 3.500
W/m2 kW
2.
Use the arrow keys (,) and ( , ) to add a short description (max 13 chars) relevant to the test performed.
Tc Te Pdc
45
°C
30
°C
3.125
kW
3.
Press the SAVE key again to confirm ESC/MENU to exit without saving.
data
saving
or
KEYBOARD PV PLANT
A B CD E F GH I J K L MNOP
Q R S T U V WX Y Z – + 0 1 2 3
4 5 6 7 8 9 SPACE
DEL
SAVE/ESC
7.2. SAVING IVCK, M AND LOW MEASURED RESULTS
1. Press the SAVE key with a measured result shown on the 15/05/12 15:34:26
display. The instrument shows containing the following items:
the
screen
to
the
side,
MEASURE Area Field
: : :
The first memory location available (“MEASURE”) The 1st marker level (ex: Area)
Inverter
:
Comment :
004 001
001 001
The 2nd marker level (ex: Field)
The 3rd marker level (ex: Inverter)
The field “Comment” in which the operator can include
a short description (max 13 char) for the system.
Modification SAVE
2. Each marker level can be assigned to different labels (5 Inverter
labels are built-in and 5 labels can be customized). Select the desired marker level using the arrow keys ( , ) and
Area Building Field
3.
press the ENTER key to get available strings. Select one of the items available using the
arrow
keys
Plant Comment
(,) and confirm with ENTER.
004 002
001 001 :
Modification SAVE
4. Before performing a test, by mean Topview PC software, Solar Power
it’s possible to add custom strings into the list of marker level strings. In this case, the new values may be selected
PVPlant factory HT instruments HT ITALIA
as an alternative to the buil-in ones, as shown in the PV installation
screen to the side.
Inverter Area
Building
Field
Plant
Modification
004 002
001 001 SAVE
EN – 51
PVCHECKs
CAUTION
· Marker strings can be uploaded into the instrument by mean of the TopView software through a PC connection (“Marker manager”).
· It is possible to add up to 5 custom strings for each marker, further to the 5 provided as default values.
· Deletion of custom strings can only occur through the TopView software. (The buil-in strings cannot be eliminated.)
5. Use the arrow keys (,) and ( , ) with the virtual 15/05/12 15:34:26
keyboard to add a short description (max 13 chars) in the “Comment” field. Press ENTER to digit any character of
MEASURE HT ITALIA StringT
: : :
004 002
001
the desired name.
Module1
:
001
Comment :
TASTIERA
TEST IMP. HT
A B CD E F GH I J K L MNOP
Q R S T U V WX Y Z – + 0 1 2 3
4 5 6 7 8 9 SPACE
DEL
SAVE/ESC
6. Press the SAVE key again to confirm data saving or ESC/MENU to exit without saving.
EN – 52
PVCHECKs
7.3. OPERATIONS WITH RESULTS
7.3.1. Recalling on the display the PV efficiency results
1. Press the ESC/MENU key to go back to the main menu, 15/05/12 15:34:26
select “MEM” and confirm with ENTER to access the MEM
TYPE
section where saved values are displayed. The screen to 001
SMP 08/04/2012
the side is shown by the instrument and contains the 002
REC 13/05/2012
saved tests.
003
2. Using the arrow keys (,) and the arrow key , select
*REC 14/05/2012
“Recall”, then “Efficiency” and confirm with ENTER to
Efficiency
3.
display the results of test only. Using the arrow key it is
possible
to
display
the
Rec IVCK,M, Open
LOW
following labels: TYPE indicates the type of data saved: “REC” for a
Delete Selection
MEM – EFF
check with a precise final result YES/NO, “*REC” when
the instrument does not have the irradiance and
temperature values recorded by SOLAR-02 and “IST”
for saving the instant conditions on the display. DATE indicates the date and time in which the
datum has been saved in the instrument. DESCRIPTION indicates the description provided by
the user when saving the datum.
4. Select the type of “SMP” datum, the item “Open” and
confirm with ENTER. The instrument shows the following
screen
5. Select the type of “REC” datum, the item “Open” and 15/05/12 15:35:00
confirm with ENTER. The containing the final results.
instrument
shows
the
screen
Irr Pnom
712 3.500
W/m2 kW
6.
Selecting the “*REC” datum type, the item “View” and confirming with ENTER, the message “Unable to perform
Tc Te Pdc
analysis” is displayed due to the values recorded by SOLAR-02
irradiance/temperature missing. The partial
Vdc Idc ndc
45 °C 30 °C 3.125 kW 389 V 8.01 A 0.88
values of this measurement are only visible after
downloading the data onto the PC (see § 8) using
Analysis Results
TopView.
Selection
EFF
EN – 53
PVCHECKs
7.3.2. Recalling on the display the results of IVCK, M and LOW measurements
1. Press the ESC/MENU key to go back to the main menu, 15/05/12 15:34:26
select “MEM” and confirm with ENTER to access the MEM DATE
TYPE
section where saved values are displayed. The screen to 001 08/04/2012 10:38 LOW
the side is shown by the instrument and contains the 002 13/04/2012 12:15 M
saved tests.
003 15/05/12 12:20 IVCK
2.
Using the arrow keys (,) and the arrow key , select
“Recall”, then “IVCK” and confirm with ENTER to display the results of the I-V curve only.
Efficiency Rec IVCK,M, LOW Open
3. The field “DATE” indicates the date/time in which the D e l e t e
result of measurement has been saved, the field “TYPE” Selection
MEM- IVCK
indicates the type of test carried out (LOW, M, IVCK).
4. Use the arrow key to go to the “Comments” label
15/06/12 15:34:26
5. The instrument will display the comment inserted by the MEM
Comments
operator during the saving procedure (see § 7.2) as 001
PV PLANT 1
regards the system.
002*
PV PLANT 2
6. The symbol “*” next to the measurement number indicates
that the instrument has performed the tests with recording
of irradiance and temperature values through the remote
unit, but these values have not been transferred or are not
available. For these measurements, the values Selection translated to STC will not be available.
MEM- IVCK
7. Press ESC/MENU to exit the screen and go back to the
main menu.
EN – 54
PVCHECKs
7.3.2.1. Accessing the data saved in the memory Numeric display
1. Select a line corresponding to a saved result and press 15/05/12 15:34:26
ENTER.
MEM
Comments
2. Select “Open” and press ENTER to open the section 001
PV PLANT 1
where measurement results are displayed as:
002*
Numeric screens of measured parameters under
PV PLANT 2
standard (STC) conditions and under operating
(OPC) conditions for IVCK tests Numeric screens of measured
parameters
when
Open Recall
measuring insulation (M) and continuity (LOW). D e l e t e
Selection
MEM-IVCK
3. For IVCK test, there are the values of the following 15/05/12 15:34:26
parameters: The module being used
Module: Irr
SUNPWR210
903W/m2
The value of irradiance The value of temperature of the module
Tc (AUTO) VocAvg@OPC IscAvg@OPC
57 °C – – -V – – -A
The
average
values
of
Voc
and
Isc
under
OPC
Voc@OPC Isc@OPC
647 V 5.35 A
conditions
Voc@STC
787V OK
The values of Voc and Isc measured under OPC conditions
The values of Voc and Isc calculated under STC and
Isc@STC
5.72 A
Outcome: OK
Selection
IVCK
OK
the relevant partial results obtained by comparison with
the nominal values.
4. For M test in FIELD mode there are the values of the 15/05/12 15:34:26
following parameters: Set rated test voltage Minimum threshold set on insulation measurement The type of mode·selected The real value of the applied test voltage The insulation value of positive pole Ri (+) The insulation value of positive pole Ri (-)
Ins. Test :
Ri min
:
Mode
:
Vtest 1065
Ri (+)
Ri (-)
Rp
1000 V
1.0
M
Field
1064 V
>100 M
>100 M
72 M
Outcome: OK
Selection
M
5. For M test in TIMER mode there are the values of the 15/05/12 15:34:26
following parameters: Set rated test voltage
Ins. Test : 1000 V
Ri min
: 1.0
M
Minimum threshold set on insulation measurement
Mode
: Timer
The type of selected mode
Vtest
1020 V
The real value of the applied test voltage
Ri(+)min
The Ri(+) minimum value of insulation resistance of
>200 M
the PV module/string (or of other metal masses) Test time:
10s
being tested, continuously measured during the
whole duration of measurement. The set measuring time
Oucome: OK
Selection
M
EN – 55
PVCHECKs
6. For M test in STRING mode there are the values of the 15/05/12 15:34:26
following parameters: Set rated test voltage Minimum threshold set on insulation measurement The type of mode selected The real value of the applied test voltage The Rp final value of the measurement obtained by
Ins. Test Ri min Mode
Vtest
Rp
: 1000 V
: 1.0
M
: String
1020 V
>200 M
the parallel of Ri(+) e Ri(-) values which is compared
with the Ri min limit by the instrument
Outcome: OK
Selection
M
7. For LOW test, there are the values of the following 15/05/12 15:34:26
parameters: Limit threshold set for continuity measurement
RPE max :
1.0
Rcal
: —
Value of calibration resistance of test cables The value of resistance of the object being tested
Rpe
= 0.99
The real value of the applied test current
Itest
= 212. mA
Outcome: OK
Selection
LOW
7.3.3. Deleting data from the memory
1. Within the list of saved results, press ENTER to display 15/05/12 15:34:26
the submenus.
MEM
2. Select “Delete”, then press . The instrument allows 001
TYPE SMP 08/04/2012
selecting the following items:
002
REC 13/04/2012
Delete Last deletes the last saved test Delete All deletes the whole memory content 3. Use the arrow keys (,) to select the desired option,
then press the ENTER key to confirm selection
Recall
4. Press ESC/MENU to exit the screen and go back to the O p e n D e l e t e L a s t
main menu.
Delete Delete All
Selection
MEM EFF
EN – 56
PVCHECKs
8. CONNECTING THE INSTRUMENT TO THE PC
CAUTION
· The connection between instrument and PC is realized by means of cable C2006
· To transfer data to a PC, you must have previously installed the Topview management software, which can be downloaded from the link: https://www.ht-instruments.com/download
· Before connecting, it is necessary to select the port to be used on the PC. To set these parameters, launch the TopView software and refer to the programme’s on-line help.
· The selected port must not be engaged by other devices or applications, e.g. a mouse, a modem, etc. Close any applications running using the Microsoft Windows Task Manager function.
· The optical port emits invisible LED radiations. Do not directly observe with optical instruments. Class 1M LED apparatus according to standard IEC/EN 60825-1.
To transfer data to the PC, follow this procedure:
1. Switch on the instrument by pressing the ON/OFF key. 2. Connect the instrument to the PC via the provided optical/USB cable C2006. 3. Press the ESC/MENU key to open the main menu. 4. Use the arrow keys (,) to select “PC” to access data transfer mode and confirm
with ENTER.
15/05/12 15:34:26
IVCK Test mod./strings
LOW PE Continuity Test
M Insulation Test
EFF Efficiency Test
SET Settings
DB
Archive modules
MEM Recall Memory
PC
Trasf. data to PC
Press Enter to sel.
MENU
5. The instrument shows the following screen:
15/05/12 15:34:26
Connecting to PC
MENU
6. Use TopView controls to activate data transfer (please refer to the on-line help of the programme).
EN – 57
PVCHECKs
9. MAINTENANCE
9.1. GENERAL INFORMATION The instrument you purchased is a precision instrument. While using and storing the instrument, carefully observe the recommendations listed in this manual in order to prevent possible damage or danger during use. Do not use the instrument in environments with high humidity levels or high temperatures. Do not expose to direct sunlight. Always switch off the instrument after use. In case the instrument is not to be used for a long time, remove the batteries to avoid liquid leaks that could damage the instrument’s internal circuits.
9.2. BATTERY REPLACEMENT When the low battery symbol ” ” appears on the LCD display, or if during a test the instrument shows the message “low battery”, it is necessary to replace the batteries.
CAUTION
Only expert and trained technicians should perform this operation. Before carrying out this operation, make sure you have disconnected all cables from the input terminals.
1. Switch off the instrument by pressing and holding the ON/OFF key. 2. Remove the cables from the input leads. 3. Loosen the battery compartment cover fastening screw and remove the cover. 4. Remove all the batteries from the battery compartment and replace them only with new
batteries of the same type only (see § 10.5) making sure to respect the indicated polarities. 5. Restore the battery compartment cover into place and fasten it by means of the relevant screw. 6. Do not scatter old batteries into the environment. Use the relevant containers for disposal.
9.3. CLEANING THE INSTRUMENT Use a soft and dry cloth to clean the instrument. Never use wet cloths, solvents, water, etc.
9.4. END OF LIFE CAUTION: this symbol indicates that the appliance, its accessories and the internal batteries must be collected separately and correctly disposed of.
EN – 58
PVCHECKs
10. TECHNICAL SPECIFICATIONS
10.1. TECHNICAL SPECIFICATIONS FOR PV INSTALLATION EFFICIENCY Accuracy is indicated as [%reading + (no. of digits) * resolution] at 23°C ± 5°C, <80%RH
DC Voltage Range [V] 5.0 199.9
200.0 999.9
Resolution [V] 0.1
0.5
Accuracy (1.0rdg + 2dgt)
DC Current (by means of external clamp transducer)
Range [mV]
Resolution [mV]
Accuracy
-1100 -5 5 1100
0.1
(0.5rdg + 0.6mV)
The value of current is ALWAYS displayed with a positive sign: the value of current transduced in a voltage lower than 5mV is zeroed.
FS DC clamp [A] 1< FS 10
10< FS 100 100< FS 1000
Resolution [A] 0.001 0.01 0.1
Minimum read value [A] 0.05 0.5 5
DC Power (Vmis > 150V) Clamp FS [A]
Range [W]
1< FS 10
0.000k 9.999k
Resolution [W] 0.001k
10< FS 100
0.00k 99.99k
0.01k
100< FS 1000
0.0k 999.9k
Vmis = voltage at which power is measured; Imis = measured current
0.1k
Accuracy
(1.5rdg + 3dgt) (Imis < 10%FS) (1.5rdg) (Imis 10%FS)
Irradiance (with HT304k reference cell)
Range [mV]
Resolution [mV]
1 40.0
0.02
Accuracy (1.0rdg + 0.1mV)
Temperature (with probe type PT300N)
Range [°C]
Resolution [°C]
-20.0 100.0
0.1
Accuracy (1.0rdg + 1°C)
EN – 59
PVCHECKs
10.2. TECHNICAL SPECIFICATIONS OF THE IVCK FUNCTION
DC Voltage @ OPC
Range [V]
Resolution [V]
Accuracy
5.0 199.9 200 999
0.1 0.5
(1.0rdg + 2dgt)
Minimum VPN voltage to start the test: 15V
DC Current @ OPC Range [A] 0.10 15.00
Resolution [A] 0.01
Accuracy (1.0rdg + 2dgt)
DC Voltage @ STC Range [V] 5.0 199.9
200 999
Resolution [V] 0.1
1
Accuracy (4.0rdg + 2dgt)
DC Current @ STC Range [A] 0.10 15.00
Resolution [A] 0.01
Accuracy (4.0rdg + 2dgt)
Irradiance (with HT304k reference cell)
Range [mV]
Resolution [mV]
1 40.0
0.02
Accuracy (1.0rdg + 0.1mV)
Temperature (with probe type PT300N)
Range [°C]
Resolution [°C]
-20.0 100.0
0.1
Accuracy (1.0rdg + 1°C)
CAUTION
Do not use the instrument for IVCK tests on PV modules with efficiency >19%. Check the technical characteristics of the PV modules before carrying out the tests in order to avoid possible damage to the instrument
10.3. TECHNICAL SPECIFICATIONS FOR ELECTRICAL SAFETY
Continuity of protective conductors (LOW)
Range []
Resolution []
Accuracy
0.00 1.99
0.01
2.0 19.9
0.1
(2.0rdg + 2dgt)
20 199
1
Test current >200mA DC up to 5 (cables included), resolution 1mA, accuracy (5.0%reading + 5digits) Open-circuit voltage 4 < V0 < 10V
Insulation resistance (M) Mode TIMER
Test voltage [V]
Range [M]
Resolution [M]
Accuracy
0.01 1.99
0.01
250, 500, 1000
2.0 19.9
0.1
(5.0rdg + 5dgt)
Open voltage Short circuit current Generated voltage
Rated current measured
20 199
1
<1.25 x nominal test voltage
< 15mA (peak) for each test voltage
resolution 1V, accuracy (5.0%reading + 5digits) @ Rmis> 0.5% FS
> 1mA with 1k @ Vnom
Insulation resistance (M) Mode FIELD (*), STRING (**)
Test voltage [V]
Range [M]
Resolution [M]
Accuracy (***)
250, 500, 1000
(*) For FIELD mode (**) For STRING mode Open voltage Short circuit current Generated voltage Rated current measured (***) For FIELD mode:
0.1 1.9
0.1
2 99
1
(20.0rdg + 5dgt)
if VPN >1V the minimum voltage VEP and VEN for the calculation of Ri(+) and Ri(-) is 1V
minimum VPN voltage to start the test: 15V
<1.25 x nominal test voltage
< 15mA (peak) for each test voltage
resolution 1V, accuracy (5.0%reading + 5digits) @ Rmis> 0.5% FS
> 1mA with 1k @ Vnom
add 5 dgts to the accuracy if [Max (R+,R-) / Min (R+,R-) 100]
EN – 60
PVCHECKs
10.4. REFERENCE STANDARDS
10.4.1. General
Instrument safety:
IEC/EN61010-1
EMC:
IEC/EN61326-1
Safety of measuring accessories:
IEC/EN61010-031
Measurements:
IEC/EN62446-1 (IVCK, LOW, M)
Insulation:
double insulation
Pollution level:
2
Measurement category:
CAT III 300V to earth
Max 1000V between inputs P, N, E, C
10.5. GENERAL CHARACTERISTICS
Display and memory
Type of display:
LCD custom, 128×128 pxl, with backlighting
Saved data:
max 999
PC interface:
optical / USB
Characteristics of radio module
Frequency range: R&TTE category: Max transmission power: Max RF connection distance:
PV system efficiency Integration Period: Memory of SOLAR-02:
Power supply Battery type: Low battery indication: Battery life: Automatic power off:
Mechanical characteristics Dimensions (L x W x H): Weight (batteries included): Mechanical protection:
2.400 2.4835GHz Class 1 30W 1m
5,10,30,60,120,300,600,900,1800,3600s approx. 1.5 hours (@ PI = 5s) approx. 8 days (@ PI = 600s)
6×1.5V alkaline type AA LR06 MN1500 the display shows symbol ” ” approx 120 hours (PV efficiency) after 5 minutes
235 x 165 x 75mm (9 x 6 x 3in) 1.2kg (2.5lv) IP40
10.6. ENVIRONMENTAL CONDITIONS FOR USE
Reference temperature:
23°C ± 5°C (73° ± 41°F)
Operating temperature:
0°C ÷ 40°C (32 ÷ 104°F)
Allowable relative humidity:
<80%RH
Storage temperature:
-10°C ÷ 60°C (14 ÷ 140°F)
Storage humidity:
<80%RH
Max operating altitude:
2000m
This instrument satisfies the requirements of Low Voltage Directive 2014/35/EU (LVD) and of EMC Directive 2014/30/EU
This instrument satisfies the requirements of 2011/65/EU (RoHS) directive and 2012/19/EU (WEEE) directive
10.7. ACCESSORIES See the attached packing list
EN – 61
PVCHECKs
11. APPENDIX THEORETICAL OUTLINE
11.1. EFFICIENCY TEST ON PV INSTALLATIONS According to the requirements of the laws in force, DC efficiency test on a PV installation
depends on the type of correction used to compensate the effects of the module’s
temperature and on the mathematical relationship used to calculate the nDC parameter (see § 5.2.3).
Corr.
Tcel value
Tmod
Tcel
=
Measured module value
temp.
Tcel = Calculated module temp. value
Tenv Tcel = Tamb + (NOCT – 20) Gp
800
Mathematical relationship for nDC calculation
1
Rfv2
=
1
–
(Tcel-
40)
100
(se Tcel 40C) (se Tcel 40C)
hence
nDC =
Pdc
Rf
v2
Gp GSTC
Pn
Standard Result
Italian guideline
CEI 82-25
OK/NO
nDC
Tcel = Measured module temp. value
( ) nDC
=
GSTC Gp
1+
100
Tcel
–
25
Pdc Pn
– – –
– – –
where:
Symbol
Description
G p Irradiance measured on PV module surface
Measurin g unit
W/m2
GSTC Standard Irradiance = 1000
W/m2
P n Pdc
Rfv 2
NOCT
Rated Power = sum of all Pmax values of the section of the PV plant being measured
DC power measured at the PV generator’s output
Correction factor depending on the Temperature of PV Cells (Tcel) measured or calculated according to the type of selected correction relationship Absolute value of the thermal coefficient of Pmax of the PV modules included in the plant section being measured. (Normal Operating Cell Temperature) = Temperature the cells under reference conditions are brought to (800W/m2, 20°C, AM=1.5, air speed =1m/s).
kW kW
%/C %/C
The previous relationships are valid if Irradiance > Min Irradiance (see the User Manual of the MASTER instrument) and the irradiance values are “steady”, i.e. for each sample taken, with IP 1min, the difference between the maximum and minimum irradiance values measured must be < 20W/m2
The final OUTCOME can be:
Non-displayable: if the obtained values are inconsistent (e.g. nDC >1.15) or if irradiation has never reached a steady value > minimum threshold set.
The maximum performance of the system
EN – 62
PVCHECKs
12. SERVICE
12.1. WARRANTY CONDITIONS This instrument is warranted against any material or manufacturing defect, in compliance with the general sales conditions. During the warranty period, defective parts may be replaced. However, the manufacturer reserves the right to repair or replace the product. Should the instrument be returned to the After-sales Service or to a Dealer, transport will be at the Customer’s charge. However, shipment will be agreed in advance. A report will always be enclosed to a shipment, stating the reasons for the product’s return. Only use original packaging for shipment; any damage due to the use of non-original packaging material will be charged to the Customer. The manufacturer declines any responsibility for injury to people or damage to property.
The warranty shall not apply in the following cases:
· Repair and/or replacement of accessories and battery (not covered by warranty). · Repairs that may become necessary as a consequence of an incorrect use of the
instrument or due to its use together with non-compatible appliances. · Repairs that may become necessary as a consequence of improper packaging. · Repairs which may become necessary as a consequence of interventions performed
by unauthorized personnel. · Modifications to the instrument performed without the manufacturer’s explicit
authorization. · Use not provided for in the instrument’s specifications or in the instruction manual.
The content of this manual cannot be reproduced in any form without the manufacturer’s authorization.
Our products are patented and our trademarks are registered. The manufacturer reserves the right to make changes in the specifications and prices if this is due to improvements in technology
12.2. SERVICE If the instrument does not operate properly, before contacting the After-sales Service, please check the conditions of battery and cables and replace them, if necessary. Should the instrument still operate improperly, check that the product is operated according to the instructions given in this manual. Should the instrument be returned to the After-sales Service or to a Dealer, transport will be at the Customer’s charge. However, shipment will be agreed in advance. A report will always be enclosed to a shipment, stating the reasons for the product’s return. Only use original packaging for shipment; any damage due to the use of non-original packaging material will be charged to the Customer.
EN – 63
HT ITALIA SRL Via della Boaria, 40
48018 Faenza (RA) Italy
T +39 0546 621002 | F +39 0546 621144 M ht@ht-instruments.com | www.ht-instruments.it
WHERE WE ARE
HT INSTRUMENTS SL C/ Legalitat, 89
08024 Barcelona Spain
T +34 93 408 17 77 | F +34 93 408 36 30 M info@htinstruments.es | www.ht-instruments.com/es-es/
HT INSTRUMENTS GmbH Am Waldfriedhof 1b
D-41352 Korschenbroich Germany
T +49 (0) 2161 564 581 | F +49 (0) 2161 564 583 M info@ht-instruments.de | www.ht-instruments.de
Documents / Resources
![]() |
HT Instruments PVCHECKs Electric Safety and Performance Of a Photovoltaic System [pdf] User Manual PVCHECKs Electric Safety and Performance Of a Photovoltaic System, PVCHECKs, Electric Safety and Performance Of a Photovoltaic System, Performance Of a Photovoltaic System, Photovoltaic System, System |