Hillrom MP Temperature Management Controller User Manual

Tech support:
Phone: 800-445-3720 Option 2
Email: technical.support@hillrom.com
General questions or re-ordering:
Phone: 800-433-5774, Option 1
Email: orderssurgical@hillrom.com
INTRODUCTION
Device Description
General Description:
The Temperature Management Controller MP (Controller), when used with reusable Warming Devices (e.g., Warming Blankets, Integrated Warming Pad) constitutes a Temperature Management System (System).
It is the responsibility of the clinician to determine whether warming is appropriate for each individual patient. The System should not be used when clinical considerations indicate that warming of the patient is not advisable.
Indications for Use
General Indications for Use:
The Temperature Management System is intended to prevent or treat hypothermia and to provide warmth to patients. The System should be used in circumstances in which patients may not maintain a state of normothermia. The System can be used with adult and pediatric patients.
The System is intended primarily for use in hospitals and surgical centers including, without limitation, operating rooms, recovery rooms, emergency rooms, burn units and on other medical/surgical floors.
Contraindications
- Do not warm ischemic or non-perfused tissue; thermal injury may result. Examples include tissue distal to aortic cross clamping or when vasoconstrictive drugs would lead to prolonged vasoconstriction.
- Do not warm patients receiving transdermal medication; increased drug delivery may occur.
WARNINGS
General
- Explosion Hazard – Do not use the System in the presence of flammable anesthetics or highly oxygen-enriched environments such as hyperbaric chambers, oxygen tents, etc.
- Inspect System components prior to each use for signs of damage or excessive wear such as cuts, holes, loose electrical connections, or cold areas. If signs of wear are evident or if the Warming Device has been subjected to extreme physical force (e.g. pinched by clamps or run over by carts), do not use the device until it has been inspected by technical staff.
- Do not continue to use the System if the over-temperature indicator and/or any other alarms continue to sound after reset. Refer to the “Alarms and Alerts” section of this manual for more information.
- Warning required per IEC 2-35 Standard: use of materials of good thermal conductivity, such as water, gel, and similar substances, with the warming device not switched on can decrease the temperature of the body of a patient.
- California Proposition 65 Warning: The medical devices and products mentioned in this IFU may contain chemicals including Urethane or PVC, which is known to the State of California to cause cancer, birth defects, or other reproductive harm. For more information go to, www.p65warnings.ca.gov
CAUTIONS
Federal law (USA) restricts this device to sale by or on the order of a licensed healthcare professional.
PRECAUTIONS
General
- Use under the direct supervision of a clinician.
- Monitor the patient’s vital signs regularly during warming according to institutional protocol. If vital sign instability occurs, notify the clinician.
- Care should be taken when using multiple warming methods.
- The risk of skin irritation caused by pooling of surgical prep solutions under the patient may increase with warming; ensure that surgical prep solution instructions for use are followed.
- Contact your local sales representative regarding disposal. This device should not be disposed of with general waste at end of life. The device does not pose any potential hazard.
- Any serious incident that has occurred in relation to this Device should be reported to the manufacturer and the competent authority of the country in which it occurred.
- Do not open the Controller. Only approved personnel can open the Controller for service. There are no user- serviceable parts. If service is required, contact Customer Service. The manufacturer assumes no responsibility for the reliability, performance, or safety of the System if any of the following occur:
- The Controller is disassembled or serviced by an unauthorized person.
- The System components are used in a manner other than described in the User Manuals.
- The Controller is installed in an environment that does not meet the appropriate electrical and grounding requirements.
- The Controller is grounded and attached to an un-grounded table intended for use with a hyfrecator or equivalent devices.
INITIAL SETUP & ASSEMBLY
Contents
The following components are included in the Controller shipping carton:
- 1—Controller Model Temperature Management Controller MP
- 1—Mains power cord
- 1—IV pole adapter and mounting hardware
Reusable accessories are sold separately.
Mounting the Controller to an IV Pole
To mount the Controller to an IV pole, first secure the IV pole adapter to the Controller with the provided screws. Place the IV pole adapter around the IV pole and turn the clamp handle clockwise until securely tightened (Figure 1).
Caution
To prevent the IV pole from tipping, attach the Controller at a height that provides stability. It is recommended to use an IV pole with a minimum wheelbase radius of 35.6 cm (14 in) and to mount the Controller no higher than 112 cm (44 in) from the floor. Failure to properly mount the Controller may cause the IV pole to tip, possibly resulting in injury.

The back of the Controller features a standard VESA 75mm x 75mm interface, allowing for additional mounting options when using top and bottom holes. The provided IV pole clamp only works with the bottom four holes.
Rotate the clear cable-retention loop down on the side of the Controller. Use the loop to assist with cable management when the Controller is mounted to an IV pole.
INSTRUCTIONS FOR USE
The instructions below describe how to operate the Controller. For information about Warming Devices and accessories, refer to the User Manual provided with each item.
- Insert the Controller power plug into a properly grounded hospital-grade electrical outlet. Turn on the Controller via the rocker switch at rear.
Note: Turn on the Controller with no blankets or mattress connected to prevent possible alarm.
WARNING: To avoid risk of electric shock, this equipment must only be connected to a supply mains with protective earth grounding.
Note: The Controller is grounded and should not be attached to un-grounded tables intended for use with a hyfrecator or equivalent devices. - Position and secure the Device (e.g., Warming Blanket, Integrated Warming Pad) following instructions in the User Manual for each device.
- Insert the Device’s connecting cable into the proper port on the Controller.

Note: When the connecting cable is inserted into the Controller, an audible sound indicates that the control sensor and over-temperature thermistor are present and functioning properly. The Port becomes available on the touchscreen. - Control Warming Devices using the touchscreen.

- When patient warming therapy is complete, turn the Mains Power Switch to OFF.
- After use, disconnect the Controller from the electrical outlet.
CONTROL SCREEN OVERVIEW

ALARMS AND ALERTS
NOTE: WARMING BLANKET OR INTEGRATED WARMING PAD WILL STOP WARMING IF AN ALARM CONDITION APPEARS


Contact Technical Support @ 800-445-3720 or technical.support@hillrom.com
Automatic Shutoff Timer – This feature prevents the System from unintentionally being left on. The System is not intended for use without a clinician present. The System defaults to a 6-hour automatic shutoff, if there is no interaction with the Controller. This time period can be adjusted for up to 24 hours, should it be required by the case length. To adjust the automatic shutoff timer go to: Settings -> Automatic Shutoff Timer -> Use – or + to adjust the time.
CLEANING THE CONTROLLER
Warnings
- DO NOT use a dripping wet cloth, and DO NOT immerse the Controller in liquid. Moisture will damage the components, and thermal injury may result.
Precautions
- DO NOT use pure harsh solvents (e.g., MEK, acetone, etc.) or cleaners containing hydrogen peroxide to clean the Controller.
Frequency
As needed
Tools/Equipment
- Sponge or soft cloth
- Mild detergent or anti-microbial spray
- Dry soft cloth
Method
- Disconnect the Controller from the power source before cleaning.
- Wipe the Controller with a moistened sponge or soft cloth; avoid pushing fluids into any openings.
- Dry with a separate soft cloth.
CLEANING WARMING BLANKETS AND INTEGRATED WARMING PADS
Intro:
Warming Blankets and Integrated Warming Pads are non-sterile electric warming devices for use in the operating room and in pre- and post-operative areas in healthcare facilities. Clean and disinfect Integrated Warming Pads and Blankets between patient uses if they appear visibly soiled. If Integrated Warming Pads and Blankets are not visibly soiled, disinfection at the end of the operating day is recommended.
Cautions:
Do not immerse Warming Blankets or Integrated Warming pads in water. Do not use high-level disinfectants to clean Warming Blankets or Integrated Warming Pads. The CDC (USA) recommends against the use of high-level disinfectants for cleaning environmental surfaces that may contact the patient since the chemicals are highly toxic. Do not spray cleaning solutions into the electrical connector.
Do not use hydrogen peroxide-based cleaning solutions as these can adversely affect the internal heater.
Do not use sodium hypochlorite (diluted bleach) to clean the Warming Pads.
Do not place the Warming Devices in an autoclave, sterilizer, automatic washer – disinfector, or any other high- temperature system as this may damage the Devices.
Cleaning Steps
Blankets and Mattresses should be cleaned following protocols for non-critical medical devices that may contact intact skin. Examples of similar devices are blood pressure cuffs, exam table covers, operating room table pads and surgical supports. The cleaning steps are general recommendations and are not meant to replace hospital-specific cleaning protocols.
- Avoid getting cleaning fluids into the electrical connector.
- If visible body fluids or soiling are present, these must be removed before applying a disinfectant. Scrub the areas using detergent and a soft brush or sponge to remove any organic matter. Wipe the surface of the Warming Device with water using a dampened cloth. Do not immerse blankets in liquids.
- Apply a low- or intermediate-level disinfectant to the entire Device by spraying or wiping. Follow the disinfectant manufacturer’s application instructions to ensure adequate disinfection.
- Following cleaning, ensure that the Device is dry before using again.
Frequently Asked Questions
- How does Hillrom Patient Warming work?
Hillrom Warming Blankets and Integrated Warming Pads use a conductive polymer fabric called ThermAssure. A low voltage DC current flows over this light, flexible fabric and the resistance generates even warmth. Hillrom Warming Blankets and Integrated Warming Pads do not use carbon fiber or ink, which could break and create hot spots. - Why is Hillrom Patient Warming safe?
The Hillrom Temperature Management Controller MP is really a microprocessor with many built-in safety features. It monitors connected warming devices at the patient and will automatically stop operation if readings are out of safe parameters. The blankets and mattresses use a low-voltage floating isolated DC current to warm. The flexible conductive polymer fabric generates uniform heat with no hot spots. - Are the Warming Blankets and Integrated Warming Pads difficult to clean?
The cleaning process takes 30 seconds or less. Blankets can be cleaned in the OR by wiping with a low- to intermediate-level disinfectant. Do not use with cleaners that contain hydrogen peroxide or sodium hypochlorite. Blankets are designed for easy cleaning. The edges are heat-sealed to eliminate crevices. The CDC says noncritical items are safe and present virtually no risk of cross-contamination. - Is there a greater risk of cross-contamination with reusable Hillrom Warming Blankets than with disposable FAW?
No. Hillrom warming products are considered “non-critical items,” meaning they only come in contact with intact skin. According to the CDC, “Virtually no risk has been documented for transmission of infectious agents to patients through noncritical items…” Another thing to consider is that the risk of contamination may actually be greater with FAW. FAW is only partially disposable. The blower and hose are used with thousands of different patients, sometimes moving from –one OR to another. One published study showed that 92% of FAW blowers are contaminated with bacteria, and 58% internally generated and emitted germ-sized particles (Albrecht, AJIC, 2011). The contamination is significant because high-velocity air blows across the germ colonies. The contaminated hot air vents under the drapes, mixes with “dirty” floor air and rises into the sterile field. - How is Hillrom Patient Warming safer for orthopedic surgeries?
Air-free Hillrom Patient Warming is safer for surgeries involving implanted foreign materials—such as orthopedic and cardiac surgery—because there is no waste heat disrupting the sterile field with contaminants. Rising waste heat from forced-air warming contaminates the sterile surgical field above the table with dirty air from the floor by generating convection currents. There is a large body of peer-reviewed evidence published on this issue. - How do I get the BEST warming results?
BODY SURFACE AREA: Warm as much surface area as possible. Warming the core is more effective than the periphery. EARLY START: Start warming as soon as the patient arrives in the OR. SENSOR CONTACT: Ensure that the sensor is in contact with the patient. THIN BARRIER: Use the thinnest barrier possible between the patient and the warming blanket. - Can Hillrom Patient Warming be used in x-rays?
Yes. The heater fabric is completely radiolucent. However, each warming blanket and mattress has parallel buss bars that run along the long edge of the warming device. These can be seen on x-ray. In addition, the area around the sensor is also radio-opaque. - Does Hillrom offer any warming options for cases using steep Trendelenburg?
Yes. Procedures where the patient is positioned in steep Trendelenburg historically result in high rates of hypothermia due to the small surface area available for warming. The WaffleGrip Trendelenburg Positioning System effectively prevents the patient from sliding on the table while still providing warmth under the patient. - Why do Hillrom Warming Blankets and Integrated Warming Pads expire?
Over time the electric current flowing over the conductive polymer fabric oxidizes it, changing its resistance and the time it needs to reach temperature. When new it only takes a few minutes to reach the set temperature. After the expiry date, which is found on the power entry cable, the device will take closer to 10 minutes to reach the set temperature. We have no data to support the use of the Devices beyond expiry.
SYSTEM COMPONENTS AND ACCESSORIES

DEFINITION OF PRODUCT SYMBOLS



TECHNICAL MANUAL
Read Before Servicing Equipment
Repair, preventive maintenance, safety testing, and servicing of the System require the skill of qualified medical- equipment service technicians, who are familiar with good practice for medical device repair. Do not open the Controller. There are no user-serviceable parts. If service is required, contact Technical Support. Perform all maintenance activities in accordance with the instructions in this technical manual. Approved personnel: unplug the Warming Device before servicing internal components.
MAINTENANCE & TESTING
Electrical Safety Checks and Functional Testing
Frequency
These tests should be completed once every 24 months (or more frequently if required by hospital guidelines).
Tools/Equipment
- Warming Device Cable: the yellow Warming Blanket Cable (2083508) or the blue Integrated Warming Pad Cable (2083511)
- Ground continuity tester
- Leakage current tester
- Calibrated, fast-reacting thermocouple and meter
- Warming Blanket or Integrated Warming Pad (optional)
Method
- Insert the Controller power plug into a properly grounded hospital-grade electrical outlet and confirm that no cables or devices are connected to any of the ports.
WARNING: To avoid risk of electric shock, this equipment must only be connected to a supply mains with protective earth ground.
Note: The Controller is grounded and should not be attached to un-grounded tables intended for use with a hyfrecator or equivalent devices. - Perform the following tests on the Controller per standard institutional protocol:
A. Ground continuity
B. Connect a Warming Blanket to the Controller and test leakage current to ensure the maximum leakage current does not exceed the requirements in Table 3.

1. Perform “Functional Testing” described on the following pages.
Functional Testing Method for Controller
Place the Controller in Diagnostic Test mode by navigating to the Service Menu (Main Menu>Service>Diagnostic Test). To run the Diagnostic Test, click the green checkmark button. The test will not begin until all Warming Devices have been disconnected from the Controller. If a failure is observed during any of these steps, call Customer Service.
To verify functionality of Alarms, Alarm should sound near the end of the test.
Per IEC 2-35: Test verifies that the independent thermal cut-out (i.e. secondary overtemperature sensor) is operational
When the test is successfully completed, “Passed” will display on the screen. If the test is unsuccessful, “Failed” will display.
Functional Testing for Blanket and Controller
Use a Warming Blanket to perform the steps outlined below. If a failure is observed during any of these steps, repeat testing using a different Warming Blanket. If failure is observed with the second Warming Blanket contact Technical Support.
- Tape a calibrated, fast-reacting temperature sensor to the patient-facing surface of the Warming Blanket or Mattress directly over the sensor marking.
- Fold the Warming Blanket back on itself (black face to black face in the case of using a Warming Blanket) so that the temperature sensor is inside the folded area. Place a 750 to 1000g weight (such as a small book or notebook) over the sensor location to ensure that the Device remains folded and that there is good contact between the sensor and the folded Warming Blanket.
Figure 3: Warming Blanket Test Configuration

- Turn the Mains Power Switch on the Controller to the ON position. Connect the Warming Blanket cable to the Controller. The Controller will emit an audible tone when the Warming Blanket is connected.
- Set the Controller to the temperature that is to be verified. If checking all set-points, start with the low temperature.
- After the Device reaches the set point (indicated when the set point readout is no longer flashing), allow the temperature to stabilize for an additional 10 minutes. NOTE: A temperature overshoot will be noted when testing this way, which is normal.
- After 10 minutes, the reading on the temperature sensor should be within +/-1°C of the set-point temperature. When measuring temperature, the accuracy and tolerance of the sensor must be taken into account. This will depend on the type of sensor being used and can range from +/- 0.2°C to +/- 2.0°C. The measurement tolerance of the sensor must be added to the +/- 1.0°C tolerance for the System to determine the pass/fail criteria for this test. For example: If the Controller is set to 41°C, and the measurement is being made with a temperature sensor that has a +/- 1.0°C measurement tolerance, the acceptable range of measured temperatures will be 39 to 43°C. (i.e. 41 +/- 2°C).
- Repeat steps 4-6 for the next temperature setting, if required.
Functional Testing for Integrated Warming Pad and Controller
Use an Integrated Warming Pad, a controller, and a blanket or sheet to perform the steps outlined below. If a failure is observed with the Integrated Warming Pad, contact Technical Support.
- Tape a calibrated, fast-reacting temperature sensor to the patient-facing surface of the Integrated Warming Pad directly over the sensor marking.
- Place blanket, sheet or warming blanket over the integrated warming pad fully covering the temperature sensor. Place a small book or notebook over the location of the temperature sensor, and place a weight of approximately 10 lbs. on the notebook to ensure there is good contact between the two temperature sensors (one inside the warming pad, and the second being used for verification).
• Figure 4: Integrated Warming Pad Test Configuration

- Turn the Mains Power Switch on the Controller to the ON position. Connect the Warming Pad cable to the Controller. The Controller will emit an audible tone when the Warming Pad is connected.
- Set the Controller to the temperature that is to be verified. If checking all set-points, start with the low temperature.
- After the Warming Pad reaches set point (indicated when the set point readout is no longer flashing), allow the temperature to stabilize for an additional 10 minutes.
- After 20 minutes, the reading on the temperature sensor should be within +/-1°C of the set-point temperature. When measuring temperature, the accuracy and tolerance of the sensor must be considered. This will depend on the type of sensor being used and can range from +/- 0.2°C to +/- 2.0°C. The measurement tolerance of the sensor must be added to the +/- 1.0°C tolerance for the System to determine the pass/fail criteria for this test. For example: If the Controller is set to 40°C, and the measurement is being made with a temperature sensor that has a +/- 1.0°C measurement tolerance, the acceptable range of measured temperatures will be 35 to 40°C. (i.e. 40 +/- 2°C).
- Repeat steps 4-6 for the next temperature setting, if required.
Error Codes: Alerts & Alarms
When an Alarm or Alert condition occurs, the associated error code will remain on the display until the condition is resolved.
If multiple Alarm or Alert conditions occur sequentially, the code associated with the highest priority Alert condition will be displayed first, followed by the next highest priority alarm or alert condition until all Alarm or Alert conditions have been displayed to the user. Once all Alarm or Alert conditions have been displayed, the display will return to the main operating screen where the error codes will still be present on the screen in place of the temperature set point.
If a new Alarm or Alert condition occurs, all active Alarm or Alerts will be displayed to the user again sorted by priority as described above.
To resolve an Alarm, follow the on-screen instructions. Devices will not be active when an Alarm is occurring.
Alarm “audio paused” duration is 10 minutes, after which time audio resumes.

SPECIFICATIONS







ELECTROMAGNETIC COMPATIBILITY (EMC)
The System requires special precautions regarding EMC and must be installed and put into service according to the EMC information provided in this User Manual.
Warning
- Warning per IEC 1-2 Standard:
- Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
- Use of accessories, transducers, and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
- Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the Temperature Management Controller MP, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Technical Support Phone: 800-445-3720, Option 2 Email: technical.support@hillrom.com
General Questions or re-ordering: 800-433-5774, Option 1 Email: orderssurgical@hillrom.com
Devices are protected by some or all of the following patents: (US Patents 7,543,344; 7,714,255; 7,851,729; 7,786,408; 8,062,343; 8,283,602; 8,604,391; 8,624,164; 8,772,676; 8,986,359; 9,962,122; 9,668,303; 10,154,543; 10,201,935; 10,206,248; 10,506,668; PCT Patent EP 2,062,460).
Other patents are pending.
© 2024 Augustine Surgical, Inc. All rights reserved.
Documents / Resources
![]() | MP Temperature Management Controller |
References
- mailto:orderssurgical@hillrom.comhillrom.com
- mailto:technical.support@hillrom.comhillrom.com
- p65warnings.ca.govwww.p65warnings.ca.gov
- p65warnings.ca.gov/www.p65warnings.ca.gov
- User Manualmanual.tools
