GOSSEN METRAWATT logoShort-form Operating Instructions
5/12.21
METRAHIT ISO AERO
TRMS Multimeter with Insulation Measurement

Metrahit ISO Aero TRMS Multimeter

Please make sure to read the detailed operating instructions at www.gossenmetrawatt.com.
The short-form instructions are no substitute for the detailed instructions!
Symbol GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Symbol 1 indicates parameter settings which are only described in the detailed operating instructions.

Standard Equipment

  • 1 Isolation multimeter
  • 1 Protective rubber cover
  • 1 Cable set KS17-2
  • 1 DAkkS calibration certificate
  • 2 Batteries 1.5 V, type AA inserted in unit
  • 1 Power pack NA X-TRA
  • 1 Short-form operating instructions*
  • * Detailed operating instructions for download on the Internet at www.gossenmetrawatt.com

Application of measuring cable set KS17-2

Maximum Rated Voltage 600 V 1000V
Measuring Category CAT III CAT I I
Maximum Rated Current 1 A 16 A
with safety cap applied
without safety cap applied

Safety Instructions

In order to maintain the flawless condition of the instrument, and to ensure its safe operation, it is imperative that you read the operating instructions thoroughly and carefully before placing your instrument into service, and that you follow all instructions contained therein.

Observe the following safety precautions:
The multimeter may not be used in potentially explosive atmospheres.
The multimeter may only be operated by persons who are able to recognize contact hazards and take the appropriate safety precautions. Contact hazards exist wherever voltages of more than 33 V (RMS value) and/or 70 V DC occur.
The maximum voltage allowable according to standard between the voltage inputs or all inputs towards earth respectively is equal to 1000 V, category II / 600 V, category III.
Attention: Unexpected voltages may occur at defective devices, capacitors,…!
The insulation of the measurement cables may not be damaged, cables and plugs may not be interrupted!
No measurements may be made in electrical circuits with corona discharge (high-voltage)!
Special care is required when measurements are made in HF electrical circuits where dangerous pulsating voltages may be present.
Measurements under moist ambient conditions are not permissible.
Do not overload the measuring ranges beyond their allowable capacities!
The input of the current measuring ranges is fitted with a fuse.
Use original fuses only, see label on the housing or technical data section!
Only operate the instrument with batteries inserted. Otherwise dangerous currents or voltages will not be indicated and your instrument may be damaged.
The device may not be operated with the fuse or battery compartment cover removed.

Switching on / off / Light on

GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Switching on offBattery TestGOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Battery Test

Battery Replacement

2 Batteries: IEC LR6 / AA – AM3 – Mignon
GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Symbol 2 Disconnect the instrument from the measuring circuit before opening the battery compartment lid! Turn the slotted head screw counter-clockwise for this purpose. Observe the correct polarity of the batteries!
When refitting the battery compartment lid the side with the guide hooks must be inserted first. Then turn the slotted head srew clockwise.

Select Resolution GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Select ResolutionSelect Measuring Function GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Select Measuring Function

Measuring Range Selection

GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Measuring Range SelectionMeasuring Inputs GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Measuring InputsDATA-Hold/-Compare GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - CompareMIN/MAX memory GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - MAX memoryZERO
Zero BalancingGOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Zero BalancingΩ Resistance Measurement
GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Symbol 3 Continuity Testing
GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Resistance Measurement

GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Symbol 4 Diode Testing

GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Diode TestingTemp RTD
Temp TC
Temperature Measurement GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Temperature MeasurementV/ V GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Symbol 5
Direct Voltage / Pulsating Voltage MeasurementGOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Voltage MeasurementV~/ Hz
AC Voltage – Frequency, without/with Low-pass FilterGOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Low-pass FilterA / A GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Symbol 5
DC / Pulsating Current MeasurementGOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Pulsating Current MeasurementA~ / Hz
Alternating Current / Frequency MeasurementGOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Frequency Measurement GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Symbol 6 V / Hz
Measurement with Clip-on Current SensorGOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Measurement with Clip-on Current Sensor

Transformation ratio
CLIP
Measuring Ranges Current sensor
300
mV
3 V 30 V
1:1
1mV/1mA
0,3 A 3 A 30 A WZ12C
1:10
1mV/10mA
3 A 30 A 300 A WZ12B
Z201A/B
VETRAFLEX
1:100
1mV/100mA
30 A 300 A 3000 A Z202A/B
VETRAFLEX
1:1000
1 mV/1 A
300
A
3000 A (30kA) WZ12C
Z202A/B
Z203A/B
VETRAFLEX

GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Measurement with Clip-on Current Sensor 2GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Current Transformer 1

Transformation ratio
CLIP
Measuring Ranges Zange Current transformer
30
mA
300
mA
3 A
1:1
1mA/1mA
30
mA
300
mA
3 A
1:10
1mA/10mA
300
mA
3 A 30 A
1:100
1mA/1 00mA
3A 30A 300A
1:1000
1 mA/1 A
30 A 300 A 3000 A WZ12A
WZ12D
WZ11A
73511
73512
73514

GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Current Transformer 2GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Symbol 7 CapacitanceGOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - CapacitanceV 1MΩ (Ri = 1 MΩ)
Interference Voltage

MΩ@UISO GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Interference VoltageDevice and Measuring Parameters
SETUP GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Symbol 1
GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Device and Measuring Parameters 1Main Menus →… SEnd … StorE … GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Symbol 1GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Device and Measuring Parameters 2Sub-menus / Parameters ↓GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Device and Measuring Parameters 3

Technical Data

GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Symbol 1

Measuring Range Measuring Uncertainty Overload capacity
GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Symbol 8
300 mV ±(0,2% + 3 D)
2)
±1/1,5% + 3 D (> 100 D) 1000 V
DC
AC
sine
continuous
3 V ±(0,15% + 2 D) ±1/1,5% + 3 D (> 100 D)
30 V ±(0,15% + 2 D)
300 V ±(0,15% + 2 D)
1000 V ±(0,2% + 2 D)
1 Hz … 300,0 kHz ±(0,1% + 2 D) 1000 V 3) max. 10 s
GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Symbol 8
300 μA ±(0,5% + 5 D) ±1,5% + 5 D (> 100 D) 0,3 A continuous
3 mA ±(0,2% + 3 D)
30 mA ±(0,5% + 3 D)
300 mA ±(0,2% + 3 D)
3 A ±(1% + 5 D) 10 A 5 min
10 A ±(1% + 5 D)
1 Hz … 30,00 kHz ±(0,1% + 2 D) 4)
±(… % v. MW/rdg. + … D)
300 Ω 0,5 + 3 2) 1000 V
DC
AC
sine
max. 10 s
3 kΩ 0,5 + 2
30 kΩ
300 kΩ
3 MΩ
30 MΩ 2 + 5
GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Symbol 3 300 Ω 3 + 5
GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Symbol 4 5,100 V 2 + 5
±(… % v. MW/rdg. + … D)
30 nF ±(1% + 6 D) 2) 1000 V
DC
AC
sine
max. 10 s
300 nF ±(1% + 6 D)
3 μF
30 μF
300 μF ±(5% + 6 D)
  1. bei / at 0 ºC … + 40 ºC
  2. bei / with function ZERO active
  3. Power limiting: 3 · 10⁶ V · Hz @ U > 100 V
  4. max. current values see current measuring range

Insulation Measurement 1)

Measuring Range nominal voltage UISO [V] Measuring Uncertainty
±(… % v. MW/rdg. + … D)
0,3 V … 1000 V GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Symbol 8 2) Ri=1MΩ 3 + 30 > 100 Digit
5 … 310,0 kΩ 10//50/100/250/500 V 5 + 30 // 3 + 5
0,280 … 3,100 MΩ 10//50/100/250/500 V 5 + 30 // 3 + 5
02,80 … 31,00 MΩ 10//50/100/250/500 V 5 + 30 // 5 + 5
028,0 … 310,0 MΩ 10//50/100/250/500 V 5 + 30 // 5 + 5
0280 … 3100 MΩ 500 5 + 5
  1. During insulation resistance measurement (MΩ@UISO): If „Error“ is displayed >> limits: Uinterference > 10 … 20 V and Uinterference ≠ Uiso,
    Ri < 10 kΩ@ Uiso 10 V, Ri < 50 kΩ@ Uiso 50 V,
    Ri < 100 kΩ @ Uiso 100 V, Ri < 250 kΩ@ Uiso 250 V,
    Ri < 500 kΩ@ Uiso 500 V
  2. Interference voltage measurement TRMS (V AC + DC) with 1 MΩ input resistance, Bandwidth 15 Hz … 500 Hz, measuring error 3% + 30 Digit

Electrical Safety

Protection class II
– nach / per IEC 61010-1:2010/DIN EN 61010-1:2011/VDE 0411-1:2011
Measuring Category CAT II / CAT III
Nominal Voltage 1000 V / 600 V
Pollution degree 2
Test Voltage (nach/per IEC 61010-1) 5,2 kV~
Protection Housing: IP54 (pressure equalization by means of the housing)

Electromagnetic Compatibility EMC

Interference Emission
EN 61326-1:2013 class B
Interference Immunity
EN 61326-1:2013, EN 61326-2-1:2013

Ambient Conditions

Accuracy range 0°C … + 40°C
Operating temperature –10 °C … + 50°C
Storage temperature without battery – 25 °C … + 70°C
relative humidity 40 … 75 %
no condensation allowed
Elevation up to 2000 m maximum
GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Symbol 9
Fuse
FF(ultrarapid) 10 A / 1000 V AC DC
10 mm x 38 mm
Electric Warning Icon breaking capacity: 30 kA (Min 10 kA)

If you use other fuses than the one indicated above you forfeit your product guarantee.

Internal Fuse Test

GOSSEN METRAWATT Metrahit ISO Aero TRMS Multimeter - Internal Fuse TestFuse Replacement

warning 2 Disconnect the instrument from the measuring circuit before opening the fuse compartment lid! Turn the (captive) slotted head screw counter-clockwise for this purpose. Remove the fuse with the flat end of the fuse compartment lid.
When refitting the fuse compartment lid the side with the guide hooks must be inserted first. Then turn the slotted head srew clockwise.

Repair and Replacement Parts Service
Calibration Lab and Rental Instrument Service
When you need service, please contact:

GMC-I Service GmbH
Service-Center
Beuthener Straße 41
90471 Nürnberg • Germany
Phone +49 911 817718-0
Fax +49 911 817718-253
E-Mail service@gossenmetrawatt.com
www.gmci-service.com

Product Support
When you need support, please contact:

Gossen Metrawatt GmbH
Product Support Hotline
Telefon D 0900 1 8602-00
A/CH +49 911 8602-0
Phone +49 911 8602-0
Fax +49 911 8602-709
E-Mail support@gossenmetrawatt.com

© Gossen Metrawatt GmbH
Edited in Germany • Subject to change without notice / errors excepted • A pdf version is available on the Internet
All trademarks, registered trademarks, logos, product names, and company names are the property of their respective owners.

GOSSEN METRAWATT logoGossen Metrawatt GmbH
Südwestpark 15
90449 Nürnberg • Germany
Telefon +49 911 8602-0
Fax +49 911 8602-669
E-Mail info@gossenmetrawatt.com
www.gossenmetrawatt.com