MH-Z14 Intelligent Infrared Gas Module

MH-Z14-Intelligent-circut-board

Profile

Main functions and features:

  • High sensitivity, High resolution
  • Low power consumption
  • Output modes: UART, analog voltage signal, PWM wave
  • Quick response
  • Temperature compensation, excellent linear output
  • Good stability
  • Long lifespan
  • Anti-water vapor interference
  • No poisoning
    MH-Z14-Intelligent-circut-board

Main technical parameters

Working voltage 4.5 V ~ 5.5V DC
Average current < 85 mA
Interface level 3.3 V
Measuring range 0~5%VOL optional
 

Output signal

PWM
UART
0.4-2V DC
Preheat time 3min
Reponse Time T90    < 90s
Working temperature 0℃ ~ 50℃
Working humidity 0~95%RH
Weight 15 g
Lifespan >5 year
Dimension 57.5×34.7×16mm(L×W×H)

 

Target Gas Measuring

Range

Accuracy Mark
 

 

 

 

Carbon Dioxide (CO2)

0~2000ppm  

 

 

±(50ppm

+5%readin g value)

Temperature

compensation

0~5000ppm Temperature

compensation

0~1%VOL Temperature

compensation

0~3%VOL Temperature

compensation

0~5%VOL Temperature

compensation

Structure

MH-Z14-Intelligent-Structure

Definition for pins

MH-Z14-Intelligent-Structure

PIN

Description

Pad1、 Pad15、Pad17

Vin(input voltage 4.5V~5.5V)

Pad2、Pad3、 Pad12、Pad16

GND

Pad4

Vout2     (0.4~2V)

Pad5

Vout1     (0~2.5V)

Pad6

PWM

Pad8

HD

Pad7、 Pad9

NC

Pad11、Pad14、Pad18

UART(RXD)     0~3.3V input digital

Pad10、Pad13、Pad19

UART(TXD)    0~3.3V output digital

Circuit

MH-Z14-Intelligent-Circut-Structure

MH-Z14-Intelligent-Circut-Structure

Operating instruction

Analog output connections

The output value of Vout1 is 0-2.5V, which stands for 0 to full range.
The output value of Vout2 is 0.4-2V, which stands for 0 to full range.

Vin –5V
GND- Power Ground
Vout2-ADC input
After preheating, the value of output voltage from Vout2 represents gas concentration.

Voltage-output-graph

PWM output (taking PWM output from 2000ppm as example):

CO2 output range: 0ppm-2000ppm
Cycle: 1004ms±5%

High level output for beginning: 2ms (in name)
Middle of cycle: 1000ms±5%
Low level output for ending: 2ms(in name)
Account formula for CO2 concentration which gets through PWM:
PWM-output-example

Voltage-output-graph

Digital connects

Vin-5V power
GND- Power Ground
RXD connect sensor TXD
TXD connect sensor RXD
You can read gas concentration via Uart, no need to calculate.

Communication protocol

  1. General Settings
    Baud rate 9600
    Date byte 8 byte
    Stop byte 1byte
    Calibrate byte no
  2. Command
    Each command or return:
    Contains 9 bytes (byte 0 ~ 8) starting byte fixed 0 XFF
    command contains sensor number (factory default to 0 x01) to check and endCommand List:
    0x86 Gas concentration
    0x87 Calibrate zero point(ZERO)
    0x88 Calibrate span point(SPAN)

    Read gas concentration

    Send command

    Byte0

    Byte1 Byte2 Byte3 Byte4 Byte5

    Byte6

    Byte7 Byte8

    Starting byte

    Sensor No. command

    Check value

    0XFF

    0x01 0x86 0x00 0x00 0x00

    0x00

    0x00 0x79

    Return value

    Return

    Byte0

    Byte1 Byte2 Byte3 Byte4 Byte5

    Byte6

    Byte7 Byte8

    Starting

    byte

    command High level concentration Low level concentration

    Chec k

    value

    0XFF 0x86 0x02 0x60 0x47 0x00

    0x00

    0x00 0xD1

    Gas concentration= high level *256+low level

    Calibrate zero point

    Send command

    Byte0

    Byte1 Byte2 Byte3 Byte4 Byte5

    Byte6

    Byte7 Byte8

    Starting

    byte

    Sensor No. command

    Check value

    0XFF

    0x01 0x87 0x00 0x00 0x00 0x00 0x00

    0x78

    No return value

    Calibrate span point

    Send command

    Byte0

    Byte1 Byte2 Byte3 Byte4 Byte5 Byte6

    Byte7

    Byte8

    Starting

    byte

    Sensor No. command High level span

    point

    Low level span

    point

    Check value
    0XFF 0x01 0x88 0x07 0xD0 0x00 0x00

    0x00

    0xA0

    No return value

     

  3. Calibration and calculation
    The checksum = (invert (byte 1 +… + 7)) + 1
    Reading gas concentration:

    Send command

    Byte0

    Byte1 Byte2 Byte3 Byte4 Byte5 Byte6

    Byte7

    Byte8

    Starting
    byte

    Sensor No. command

    Check value

    0XFF

    0x01 0x86 0x00 0x00 0x00 0x00

    0x00

    0x79

    Except byte 0 ,add the other bytes together
    0x1 + 0x86 + 0 + 0 + 0 + 0 + 0 = 0x87
    Get the value from the first step, then invert it.
    0xff – 0x87 = 0x78
    The second value plus one
    0x78 + 0x01 = 0x79
    Program :C language

    char getCheckSum(char *packet)
    {char i, checksum; for( i = 1; i < 8; i++)
    {
    checksum += packet[i];

    }
    checksum = 0xff – checksum;
    checksum += 1;
    return checksum;
    }

Notes for maintenance

7.1 The sensor should be calibrated regularly. The cycle time is better to be no more than 6 months.
7.2 Do not use the sensor in the high dusty environment for long time.
7.3 Please use the sensor with correct power supply.
7.4 Forbidden to cut the sensor pin.

 