INSTALLATION

  1. When possible, connect the Chassis Ground on the back of your potentiostat to known, good earth ground.GAMRY-INSTRUMENTS-Reference 600+620-USB-Potentiostat-Calibration-FIG-1
  2. Connect the cell cable to the 200 Ohm Calibration Cell included with your instrument.GAMRY-INSTRUMENTS-Reference 600+620-USB-Potentiostat-Calibration-FIG-2
  3. Place the dummy cell inside the Calibration Shield, close the lid, and connect the black floating-ground lead of your cell cable to the Shield’s grounding post.GAMRY-INSTRUMENTS-Reference 600+620-USB-Potentiostat-Calibration-FIG-3
  4. Open Gamry Framework™. Select Experiment > Utilities > Calibrate InstrumentGAMRY-INSTRUMENTS-Reference 600+620-USB-Potentiostat-Calibration-FIG-4
  5. Choose your potentiostat. Select the radio button for calibration type Both, and click the OK button.GAMRY-INSTRUMENTS-Reference 600+620-USB-Potentiostat-Calibration-FIG-5

After you confirm several instructional messages, the calibration proceeds automatically, and you are notified if the calibration is successful.

TROUBLESHOOTING A FAILED CALIBRATION

Calibration is used to check the potentiostat’s health, and to “zero” many of the measurement circuits to your laboratory environment. A warning does not necessarily indicate a critical failure, and Gamry can use calibration information to determine the source of the warning.GAMRY-INSTRUMENTS-Reference 600+620-USB-Potentiostat-Calibration-FIG-6

  1. Double-check the following:
    • Connections on the 200 Ohm Calibration Cell.
    • Floating-ground cable is connected to a calibration shield.
    • If possible, the Chassis Ground is attached to known, good earth ground.
  2. Click the Retry button and the rest of the calibration restarts. Click the Ignore button for any other calibration warnings that may appear, and continue to Step 3.
  3. After a failed calibration attempt:
    • Find Calibration Results PC5-#####.txt in your My Gamry Data folder.*
    • E-mail the file, along with complete contact information, to techsupport@gamry.com

WHAT DOES GAMRY SOFTWARE DO?

GAMRY-INSTRUMENTS-Reference 600+620-USB-Potentiostat-Calibration-FIG-7

Documents / Resources

GAMRY INSTRUMENTS Reference 600+/620 USB Potentiostat Calibration [pdf] User Guide
Reference 600, Reference 620, USB Potentiostat Calibration, Reference 600 620 USB Potentiostat Calibration

References

Leave a comment

Your email address will not be published. Required fields are marked *