Ver 2.0 2025-02
313-4920100-003
Uric Acid Control Solution
User Manual
Intended Use
FORA Uric Acid Control Solution is intended to be used to ensure that your meter and test strips are working together properly. Use with Multi-functional Monitoring Systems from ForaCare.
This system is intended for use outside the body (in vitro diagnostic use) by people with diabetes at home and by health-care professionals in clinical settings as an aid to monitor if the system is functioning correctly.
About Control Solution
FORA Uric Acid Control Solution contains a known amount of uric acid that reacts with uric acid strips made for ForaCare.
Operation of the control solution test is similar to a uric acid test except that FORA Uric Acid Control Solution is used instead of a drop of blood. The control solutions should provide results within an expected range printed on the test strip vial label or foil pack.
IMPORTANT
The control solution range printed on the test strip label or foil pack may differ. Make sure to use the range printed on the current test strip packaging.
Do a control solution test when:
► you first receive the device.
► you begin using a new vial of test strips.
► you suspect the device or test strips are not working properly.
► your uric acid test results are not consistent with how you feel, or if you think the results are not accurate.
► you have dropped or think you may have damaged the device.
CAUTION
The control solution range printed on the test strip vial or foil pack is for control solution only. It is used to test meter and test strip performance. It is not a recommended range for your uric acid level.
Chemistry
FORA Uric Acid Control Solution is a red solution that contains less than 0.013% Uric Acid.
| Ingredient | Percent |
| Active ingredient: Uric Acid | 0.005 – 0.013% |
| Inert ingredients | > 99.987% |
Storage and Handling
► For in vitro diagnostic use only.
► Use only FORA Uric Acid Control Solution with Multi-functional Monitoring System from ForaCare.
► Check the expiration date on the vial. Do not use if expired.
► Tightly close the control solution and store at room temperature at 2°C to 30°C (36°F to 86°F). Do not freeze.
► Use only for 3 months after first opening. Record the date first opened on the control solution vial. Discard after 3 months.
Test Procedures
To do a test with FORA Uric Acid Control Solution, follow the instructions printed in the Owner’s Manual that came with the system.
Accurate results depend upon following instructions carefully. Follow the technical guidelines below whenever you test with FORA Uric Acid Control Solution:
- We recommend that you allow the meter, control solution, and test strips to reach room temperature (20°C-25°C / 68°F-77°F) before testing.
- Shake the control solution well before use.
- Discard the first drop of control solution you squeeze out.
- Squeeze out another drop onto the vial cap, and touch the drop with the absorbent hole of the test strip.
- Wipe off the dispenser tip to keep it clean.

Expected Results
At room temperature, FORA Uric Acid Control Solution test results should be within the expected range printed on the test strip vial label.
If test results are outside the expected range printed on the test strip vial or foil pack, please check the troubleshooting guide in the Owner’s Manual and repeat the test. Results that are outside the expected range may indicate:
► An error in the way you are doing the test.
► Expired test strips and control solutions.
► A problem with the meter.
► Control solution that is too warm or too cool.
► The meter code does not match the code printed on the test strip vial or foil pack.
► Failure to shake the control solution vial vigorously.
If the repeat result is still outside the range, the system may not be working properly.
Do not use the system to test your blood. Please contact the dealer for service.
Availability
FORA Control Solution is available from retail pharmacies.
OUR COMMITMENT TO YOU
Our goal is to provide you with quality healthcare products and dedicated customer service. If you are not fully satisfied with this product, or if you have questions about the use of any FORA product, please contact the dealer for service.
Symbol Information
| Symbol | Referent |
| In vitro diagnostic medical device | |
| Consult instructions for use | |
| Temperature limitation | |
| Use by / Expiry date | |
| Use within 3 months after opening | |
| Model no. | |
| Distributor | |
| Batch code | |
| Manufacturer | |
| CE mark | |
| Authorised representative in the European Union | |
| Unique Device Identification | |
| Humidity limitation | |
| Importer |
Use only with FORA 6 Plus / Connect.
ForaCare Suisse AG
Neugasse 55, 9000 St. Gallen, Switzerland
www.foracare.ch
MedNet EC-REP GmbH
Borkstraβe 10, 48163 Mϋnster, Germany
For self-testing. Not intended for human consumption.

Documents / Resources
![]() | Uric Acid Control Solution |
References
- ForaCare Suisse AG - Comprehensive Healthcare Solutionswww.foracare.ch
- User Manualmanual.tools