UV Ink Densitometer Instruction Manual
MODEL: BASIC / Measure Fluorescence

FLUO BASIC UV Ink DensitometerFLUO BASIC UV Ink Densitometer - fig1FLUO BASIC UV Ink Densitometer - fig2Invisible Ink Densitometer Fluorescence and
Phosphorescence Colorimeters
UV Ink Densitometer
Trademark & Counterfeit Protection
Control the Ink Film Thickness of Invisible Inks

FLUO BASIC UV Ink Densitometer - qr code1 FLUO BASIC UV Ink Densitometer - qr code2 FLUO BASIC UV Ink Densitometer - qr code3 FLUO BASIC UV Ink Densitometer - qr code4
http://betascreen.net/products/fluo-invisible-ink-densitometer http://betascreen.net/products/fluo-invisible-ink-densitometer http://betascreen.net/products/fluo-invisible-ink-densitometer http://betascreen.net/products/fluo-invisible-ink-densitometer

Measure
CIE, XYZ, xy, Lch, BiFL, Delta E Red Fluorescence Green Fluorescence Blue Fluorescence Fluorescence dot area %

FLUO BASIC UV Ink Densitometer - fig3

Technology
Near UV LED Tristimulus sensor Auto dynamic range

FLUO BASIC UV Ink Densitometer - fig4

FLUO Advanced (Fluorescence & Phosphorescence)

Improve Brand Security & Protection

  • Solid Ink Coverage
  • Fluorescent Tint Images
  • Overprint Control

FLUO BASIC UV Ink Densitometer - fig5

MODEL TYPES

FULL BASIC
Fluorescent Measuring UV LED for Near UV (365nm)
FULL ADVANCED
Fluorescent and Phosphorescence Measuring, Dual UV LED for Near (365nm) and Far (256nm)

Incorporate Protection with Clever Designs

FLUO BASIC UV Ink Densitometer - fig6

Serial Numbers, QR Codes Digital Signatures
Tapes and Labels Synthetic Paper Smartphone Apps
FLUO BASIC UV Ink Densitometer - fig7 Track & Trace Holographic Images RFID Printed Memory Tracer Particles DNA Markers

FLUO BASIC UV Ink Densitometer - fig8

FLUO Invisible Ink Densitometer Instruction Manual

FLUO Fluorescence Densitometer▲▲►

The FLUO BASIC Fluorescence measurement device is equipped with near UV (365 nm). The dominant wavelength of the near UV illumination is 365nm (nanometers).
On the receiver side, there are three channels to measure the fluorescent effect according to the sensitivity of the human eye, i.e. X (red sensitivity), Y (green sensitivity), and Z (blue sensitivity).

FLUO BASIC UV Ink Densitometer - fig9

The FLUO BASIC is supplied with a USB 2.0 cable to output data to a PC.

FLUO BASIC UV Ink Densitometer - fig10

Safety Instructions

For safety reasons, it is absolutely necessary to read and understand the user’s guide. If the safety recommendations and instructions in this User Guide are not complied with, measurement errors or data loss or physical injury or property damage may result.

  • FLUO is not intrinsically safe. Therefore, the device cannot be used in an environment with explosive vapors where there is a risk of spark ignition. It should be protected against strong electromagnetic fields, chemicals, corrosive vapors, strong mechanical vibrations, and impacts.
  • Use the FLUO in ambient temperatures between 15°C (59°F) and 40°C (104°F), and do not expose it to direct sunlight.
  • The FLUO should never be opened as there are no user-serviceable parts. Doing so voids the guarantee. Contact your authorized dealer if repairs are necessary.
  • To prevent the operator from looking directly into the UV light source the FLUO must be placed on a substrate while taking measurements. A sensor detects if it is not placed on a substrate and prevents the UV illumination from being activated. No measurement will take place under this condition.
  • To avoid incorrect handling, the FLUO should only be used by trained personnel
  • Use only original FLUO spare parts and accessories.
  • Use the original packaging exclusively when transporting.
  • The FLUO is equipped with UV LEDs. Never look directly into the optical system of the device during measurement, even though it is equipped with a safety interlock.

Important: This manual describes the current version of the FLUO hardware and software. Future enhancements or modifications are reserved.

How to use the FLUO
During a measurement, the FLUO BASIC will calculate three measurement values, one for X, one for Y, and one for Z.
The FLUO can be used in two modes: the color mode will show XYZ values, the intensity mode will show only the intensity of the near UV. The mode is selected in the Configuration shown below.
The LCD shows the actual measurement value. The active light source and the dominant channel are displayed by the flashing LEDs at the top of the FLUO device.
Example:

FLUO BASIC UV Ink Densitometer - fig11

Example 1: green color channel (Y), near UV illumination, Color mode

FLUO BASIC UV Ink Densitometer - fig12

Example 2: near UV illumination, Intensity mode

RESET the FLUO
Press the RESET button on the bottom side of the device to perform a RESET. Resetting will delete all internally stored references. The display will show the device type for a few seconds.

FLUO BASIC UV Ink Densitometer - fig13

FLUO equipped with near UV LED only

After a few seconds, the display will change and show the current firmware version number

FLUO BASIC UV Ink Densitometer - fig14

After initialization, the device will show ‘0’. The FLUO is now ready for measurements. FLUO BASIC UV Ink Densitometer - fig15

FLUO BASIC UV Ink Densitometer - icon2 Press the RESET button and keep the right key depressed during reset to perform a FACTORY RESET. The factory Reset will reset all settings and configurations to its defaults.
Configure the device to your needs
The FLUO can be configured to your needs in a special Configuration mode. Press and hold the right key for at least 5 seconds to enter Configuration mode

FLUO BASIC UV Ink Densitometer - icon2 The configuration mode is indicated by a ‘~’ in the left top corner of the LCD.

FLUO BASIC UV Ink Densitometer - fig16

Two settings can be configured:
a) Average Solid Reference
b) Color Mode or Intensity Mode

FLUO BASIC UV Ink Densitometer - icon2 Click the right key to adjust the Average Solid Reference settings and Color or Intensity mode selection.
a) Use an average solid reference
The FLUO can be configured to calculate the average of a series of up to 10 readings on your reference patch. The number of readings that need to be measured to calculate the reference can be set in the Configuration mode. Averaging multiple readings can be useful when there is unevenness or variability in the reference patch print quality.

FLUO BASIC UV Ink Densitometer - fig17

FLUO BASIC UV Ink Densitometer - icon3Click the up key on the device to increment the number of readings
FLUO BASIC UV Ink Densitometer - icon4 Click the down key to decrement the number of readings

b) Select Color display mode or Intensity display mode The FLUO can be used in two different modes.

In Color mode the dominant fluorescent color after any 100% reference reading will be displayed automatically. By clicking the right key after a measurement the next XYZ color channel will be displayed – followed by the XYZ color channels of the near UV readings.

FLUO BASIC UV Ink Densitometer - fig18

The Intensity mode will display one value, expressing the total brightness of the light emission within the visual spectrum.
FLUO BASIC UV Ink Densitometer - fig19FLUO BASIC UV Ink Densitometer - icon4 Toggle between the two modes by clicking the left key.

Exit the configuration mode

FLUO BASIC UV Ink Densitometer - icon2 Press and hold the right key for at least 5 seconds to exit the Configuration mode. The current configuration will be saved.

Zero references and target reference calibration
The FLUO measures relative fluorescence using a non-fluorescent reference (0% or substrate) and a target fluorescent reference (100% target fluorescent ink layer). Follow the steps outlined below to set the Zero and Reference values.

Set the Zero reference
FLUO BASIC UV Ink Densitometer - icon4Position the FLUO with the aperture on a non-fluorescent surface, for example on the substrate without fluorescent ink. Press and hold the ‘down’ key and push the device to the front until zero appears in the LCD.

FLUO BASIC UV Ink Densitometer - fig20

All measurement channels will be zeroed simultaneously.

Set a single solid reference [C 0]..[C 1]
If the average number in the settings is set to zero, the calculation of an average is disabled.

FLUO BASIC UV Ink Densitometer - icon3 Position the FLUO with the aperture on the fluorescent reference surface, for example on the substrate with fluorescent ink. Press and hold the ‘up’ key and push the device to the front until 100 appears in the LCD.FLUO BASIC UV Ink Densitometer - fig21

The maximum reference measurement measures all channels simultaneously and sets the maximum fluorescent signal for each illuminant separately to 100%. The channel with the highest fluorescent response will automatically be displayed. When the Intensity mode has been selected the intensity of the reference sample is set to 100% for near UV.FLUO BASIC UV Ink Densitometer - fig22

FLUO BASIC UV Ink Densitometer - icon2 In Color mode, you can check the relationship between fluorescent responses in red, green, and blue by stepping through the colors by pressing the ‘next’ button.
Measure an average solid reference [C 2]..[C 10]
If the reference measurement average counter is greater than 1, you must take as many solid reference readings as specified in the setting.
FLUO BASIC UV Ink Densitometer - icon3 Position the FLUO with the aperture on the fluorescent reference surface, for example onto the substrate with fluorescent ink. Press and hold the ‘up’ key and push the device to the front until [C 1] appears in the LCD.FLUO BASIC UV Ink Densitometer - fig23

Now measure the next reference again with the middle key ‘up’ depressed.FLUO BASIC UV Ink Densitometer - fig24

Continue to measure references until the number of references specified in the settings is reached. Once all reference measurements are taken, the device will calculate the average and set this value to 100%
FLUO BASIC UV Ink Densitometer - fig25Measure samples in Color mode [CoL ]
Position the FLUO with the aperture on a fluorescent surface. Push the device to the front and hold it there until the measurement result appears in the LCD. The measurement value of the currently selected color channel is displayed.
FLUO BASIC UV Ink Densitometer - fig26The LED of the dominant visual color will blink.
FLUO BASIC UV Ink Densitometer - fig27 The number displayed (79%) tells you that the sample emits only 79% of the light in the ‘Y’ channel (green) spectrum compared to your reference. This indicates to the operator that insufficient ink is being applied to the substrate.
FLUO BASIC UV Ink Densitometer - icon2 Click the right key to select the next color channel. The color channel selection will step through all color channels for near UV.

Measure a sample in Intensity mode [Int ]
If your device has been configured to display the intensity only, the difference between the reference intensity and the sample intensity will be calculated and displayed in terms of a percentage value.
Position the FLUO with the aperture on a fluorescent surface. Push the device to the front until the measurement result appears on the LCD. The measurement value of the currently selected near UV is displayed.
FLUO BASIC UV Ink Densitometer - fig28The number -21 tells you that the fluorescent intensity of your sample is 21% lower than the fluorescent intensity of your reference. This indicates to the operator that insufficient ink is being applied to the substrate.
The near UV intensity measurement is displayed with RGB LED flashing. The far intensity is displayed with FUV LED flashing.
FLUO BASIC UV Ink Densitometer - fig29Underflow and Overflow measurements
The FLUO automatically checks the validity of measurement results and displays a warning in case of invalid measurement:

Underflow

An underflow occurs when the measurement resolution is too low. This can occur when the fluorescence of the sample is extremely low. In this case, the FLUO displays uF in the proper measurement channel.
FLUO BASIC UV Ink Densitometer - fig30Overflow
The FLUO has a very large dynamic range with Automatic Gain Control. An overflow error is very rare, occurring only when the fluorescent signal in one channel exceeds the range of the AGC. The overflow for the affected channel is then displayed as oF.
FLUO BASIC UV Ink Densitometer - fig31Open aperture
For safety reasons, the UV LED will not switch on if the device is not positioned on a substrate. In this case, the display will show ‘ooo’.
FLUO BASIC UV Ink Densitometer - fig32Measurement time too short
Keep the FLUO in the front position until the measurement result appears in the LCD. If you move the FLUO back to its parking position too early, the ‘—‘ in the LCD tells you that the measurement time was too short.
FLUO BASIC UV Ink Densitometer - fig33

Replacing the batteries

The FLUO is powered by two 1.5V AA alkaline batteries. No charger is needed as they will typically last many months in normal service
FLUO BASIC UV Ink Densitometer - fig34

  • Remove the cover from the battery compartment by pressing down on the center near the front of the battery cover and sliding it towards the rear
  • Remove the old batteries and dispose of them in accordance with local regulations
  • Insert the new batteries with the polarity as shown on the label inside the compartment
  • Replace the battery cover
  • Always replace always both batteries at the same time
  • If you do not use the devices for an extended period of time it is recommended to remove the batteries from the battery compartment.

FLUO logo2Contact for Technical Support
Beta Industries – The Quality Control Company
Telephone: Toll-free: 800-272-7336, 201-939-2400 (U.S.)
Email: [email protected]
Website: www.betascreen.net
FLUO BASIC UV Ink Densitometer - qr code5 https://betascreen.net/collections/beta-fluo-invisible-uv-ink-densitometer-colorimeter-measure-uv-ink-density-other-print-production-parameters-for-better-process-control

Documents / Resources

FLUO BASIC UV Ink Densitometer [pdf] Instruction Manual
BASIC, UV Ink Densitometer, BASIC UV Ink Densitometer, Ink Densitometer, Densitometer

References

Leave a comment

Your email address will not be published.