ESBE logoCRS200 Series Setpoint Controls Sensor
Instruction Manual

ESBE CRS200 Series Setpoint Controls SensorESBE CRS200 Series Setpoint Controls Sensor - icon

CRS200 Series Setpoint Controls Sensor

Series CRS210
LVD 2014/35/EU
EMC 2014/30/EU
RoHS 2011/65/EU
EN 300220-2

ESBE CRS200 Series Setpoint Controls Sensor - icon2ESBE CRS200 Series Setpoint Controls Sensor - QR code1DECLARATION OF CONFORMITY
https://www.esbe.eu/group/support (Contorllers/Certificates)

ESBE CRS200 Series Setpoint Controls Sensor - 13

ESBE CRS200 Series Setpoint Controls Sensor - icon4 Series CRS210

ESBE CRS200 Series Setpoint Controls Sensor - 1

Sensor

  1. Install the supplied sensor adapter in a T-piece or equivalent.
  2. Install the flow temperature sensor in the sensor adapter.
  3. Pressurize the system, if necessary, tighten adapter or flow temperature sensor.

ESBE CRS200 Series Setpoint Controls Sensor - icon4 Working Direction

ESBE CRS200 Series Setpoint Controls Sensor - 2

Working Direction

  1. Set the working direction of the controller by pressing the joystick to right for clockwise opening or left for counter clockwise opening (A).
  2. Keep the joystick in the desired direction and connect power (B).
  3. Wait 2 seconds until joystick is released (C), the correct working direction is now set.

ESBE CRS200 Series Setpoint Controls Sensor - icon5 Change of target temperature

ESBE CRS200 Series Setpoint Controls Sensor - 3

Change of target temperature

  1. In order to change the target flow temperature, press the joystick to the right or to the left (A).
  2. Press the joystick down to confirm new target temperature (B).

ESBE CRS200 Series Setpoint Controls Sensor - icon3 Advanced Settings

ESBE CRS200 Series Setpoint Controls Sensor - 15

Advanced Settings

  1. Press the joystick for 5 seconds to reach advanced settings (A).
  2. To move between menus press joystick to left or right (B).
  3. Press down the joystick to enter desired menu (C).

ESBE CRS200 Series Setpoint Controls Sensor - icon6 Maximum Flow Temperature

ESBE CRS200 Series Setpoint Controls Sensor - 5

Maximum Flow Temperature

  1. To change the maximum flow temperature press down the joystick for 5 seconds to reach advanced settings (A).
  2. Choose menu “Max” by pressing the joystick (B) OK (C).
  3. Decide the maximum flow temperature by pressing the joystick (D) OK (C).
  4. Press down the joystick for 5 seconds to return to main menu (A).

ESBE CRS200 Series Setpoint Controls Sensor - 6

ESBE CRS200 Series Setpoint Controls Sensor - 7

Minimum Flow Temperature

  1. To change the minimum flow temperature press down the joystick for 5 seconds to reach advanced settings (A).
  2. Choose menu “Min” by pressing the joystick (B) OK (C).
  3. Decide the minimum flow temperature by pressing the joystick (D) OK (C).
  4. Press down the joystick for 5 seconds to return to main menu (A).

ESBE CRS200 Series Setpoint Controls Sensor - 8

ESBE CRS200 Series Setpoint Controls Sensor - icon7 Alternative temperature

ESBE CRS200 Series Setpoint Controls Sensor - 9

Alternative temperature
An alternative target flow temperature can be activated for example during legionella flushing by an external signal (A).
To set the normal target flow temperature the external signal shall not be activated. Set the target temperature as described in menu 3 (B)ESBE CRS200 Series Setpoint Controls Sensor - icon9
To set the alternative target flow temperature the external signal must be activated. Set the target temperature as described in menu 3 (B)ESBE CRS200 Series Setpoint Controls Sensor - icon8

ESBE CRS200 Series Setpoint Controls Sensor - 10

ESBE CRS200 Series Setpoint Controls Sensor - icon7 Activate T2 mode

ESBE CRS200 Series Setpoint Controls Sensor - 11

Activate Alternative temperature
T2 mode is activated by an external signal.

  1. Disconnect power to the controller.
  2. Unscrew the cover (A) and connect two conductors to the green connector (B).
  3. Place the connector on the PCB (C) and assembly the cover back (D).
  4. Connect power to the controller (E).

When the two conductors are interconnected the alternative temperature is activated and theESBE CRS200 Series Setpoint Controls Sensor - icon8  symbol is shown in the display. The connection shall be without any voltage or current and with a maximum resistance of 100Ω. (F)

ESBE CRS200 Series Setpoint Controls Sensor - 12

ESBE logowww.esbe.eu
ESBE CRS200 Series Setpoint Controls Sensor - QR code
https://esbe.eu/group/
Mtrl.nr. 98141102 Ritn.nr. 9334 utg. A
Rev. 2021-07-07ESBE CRS200 Series Setpoint Controls Sensor - icon1

Documents / Resources

ESBE CRS200 Series Setpoint Controls Sensor [pdf] Instruction Manual
CRS200 Series, Setpoint Controls Sensor, CRS200 Series Setpoint Controls Sensor, Controls Sensor, Sensor

References

Leave a comment

Your email address will not be published. Required fields are marked *