Braun IRT6525 ThermoScan 7+ Ear Thermometer Instruction Manual

Product description

The Braun ThermoScan® Ear thermometer is intended for home use for intermittent measurement of body temperature in individuals from newborn to adult. It is not intended for clinical use in a professional environment. The Age Precision feature is not intended for pre-term babies or small-for-gestational age babies. This thermometer is not intended to interpret hypothermic temperatures. Children under 12 should not take their temperatures unattended.

Warnings and Precautions:

Age Precision® color-coded display

Clinical research shows that the definition of fever changes with age. Braun's Age Precision feature helps interpret temperature readings by using a color-coded display (green for normal, yellow for fever, red for high fever) based on the selected age group.

How to use your Braun ThermoScan®

  1. Remove the thermometer from the protective case.
  2. Press the Power button. The display will show all segments during an internal self-check, then the last temperature taken for 5 seconds.
  3. The Hygiene cap indicator will blink. Attach a new, clean Hygiene cap by pushing the probe straight into the cap and pulling out. The thermometer will not work without a Hygiene cap.
  4. Select the age group using the Age Precision button. Press the button to cycle through age groups (0-3 months, 3-36 months, 36+ months). An age must be selected to take a measurement.
  5. Fit the probe snugly into the ear canal, pointing towards the opposite temple. Push and release the Start button.
  6. The ExacTemp® light will pulse during measurement and remain solid for 3 seconds to confirm a successful reading. A long beep signals completion if the probe is correctly inserted. Short beeps and an error message indicate incorrect positioning.
  7. A confirmation beep indicates an accurate measurement. The displayed temperature is equivalent to oral temperature.
  8. For the next measurement, press the Eject button to remove the used Hygiene cap and attach a new one.

Note: The thermometer turns off automatically after 60 seconds of inactivity or when the Power button is pressed.

Temperature-taking hints

Night mode

Press the night mode button to enter night mode. Pressing it again turns night mode off. In night mode, audible beeps are muted, and a Night mode icon is displayed. The flashlight turns on and flashes twice upon measurement completion. Night mode remains active until the button is pressed again or the device is turned off.

How to illuminate the display

The thermometer has a night light that illuminates the display in dim environments. The light turns on when any button is pressed and stays on for up to 15 seconds after the thermometer becomes inactive.

Last reading recall

The previous measurement is displayed during startup. Subsequently, it is shown in the upper right corner of the display with an 'M'. The memory updates with the new reading after the first button press following measurement completion.

How to set a default age group

You can set a default Age Precision age group for the device to power on with. To do this:

  1. With the device off, hold both the Power and Age Precision buttons for 4 seconds until the backlight turns yellow. Release the buttons.
  2. All Age Icons will flash. Press the Age Precision button to cycle through the age groups (0-3 months, 3-36 months, 36+ months). A full beep sounds with each press.
  3. After 3 seconds of no button press, the selected icon remains, and a full beep sounds. The device turns off. The default is now set and will remain until batteries are removed and reinserted.

How to change the temperature between °C and °F

Your Braun ThermoScan is shipped with Celsius (°C) activated. To switch to Fahrenheit (°F) or back:

  1. Ensure the thermometer is turned off.
  2. Press and hold the Power button. After approximately 3 seconds, the display will show the sequence: °C / SET / °F / SET.
  3. Release the Power button when the desired temperature scale is shown. A short beep confirms the setting, and the thermometer turns off automatically.

Care and cleaning

The probe tip is delicate and must be clean and intact for accurate readings. If used without a Hygiene cap:

Replacing the batteries

  1. Insert new batteries when the battery symbol appears.
  2. Open the battery compartment, remove old batteries, and insert new ones with correct polarity.
  3. Snap the battery door shut.

Errors and troubleshooting

Error Message Situation Solution
No Hygiene cap attached. Attach new, clean Hygiene cap.
The thermometer probe was not positioned securely in the ear. Ensure correct probe positioning and stability. Replace Hygiene cap and reposition.
An accurate measurement was not possible. (ERROR + position error animated icon) Press Start button to begin a new measurement.
Ambient temperature is not within the allowed operating range (10–40 °C or 50–104 °F). Allow the thermometer to remain for 30 minutes in a room where the temperature is between 10 and 40 °C (50 and 104 °F).
Temperature taken is not within typical human temperature range (34–42.2 °C or 93.2–108 °F). (HI = too high, LO = too low) Ensure the probe tip and lens are clean, a new Hygiene cap is attached, and the thermometer is properly inserted. Then, take a new temperature.
System error - self-check display flashes continuously and will not be followed by the ready beep and the ready symbol. Wait 1 minute until the thermometer turns off automatically, then turn on again. If error persists, remove and reinsert batteries. If still persistent, contact the Service Centre.
Battery is low, but thermometer will still operate correctly. Insert new batteries.
Battery is too low to take correct temperature measurement. Insert new batteries.
System error without messages or completely dead batteries. Insert new batteries. If the error persists, call customer service.

Product specifications

Subject to change without notice.

This medical device carries the CE mark and conforms to RoHS Directive 2011/65/EU and other applicable directives/regulations. Kaz Europe Sarl, Place Chauderon 18, CH-1003 Lausanne, Switzerland, assumes responsibility for product conformance. ASTM laboratory accuracy requirements for IR thermometers in the range of 37 to 39 °C (98 to 102 °F) is ±0.2 °C (±0.4 °F), compared to ±0.1 °C (±0.2 °F) for mercury-in-glass and electronic thermometers.

MEDICAL ELECTRICAL EQUIPMENT requires special precautions regarding EMC. Do not use this device in the presence of electromagnetic or other interference outside the normal range specified in EN 60601-1-2. For EMC requirements, visit the website (Support / Document center) or contact the service center.

Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL EQUIPMENT.

This product contains batteries and recyclable electronic waste. Dispose of it at appropriate local collection points, not in household waste.

Calibration

The product was calibrated at manufacture. Periodic calibration is not required if used according to instructions. This device is not a substitute for regular doctor check-ups. Consult your doctor if you have any doubts about temperature readings.

Inquiry

In Europe, report any serious incidents (e.g., death, life-threatening injury, surgical intervention) related to this thermometer to Kaz Europe (Helen of Troy Authorized Service Centers) and the competent authority of your member state. Provide REF, LOT, and Serial numbers for inquiries or claims. The LOT number indicates the day and year of manufacture (e.g., LOT No.: 12313tav means made on day 123, year 2013, manufacturer code tav).

Guarantee

Retain your receipt as proof of purchase for any guarantee claims. The device is guaranteed for Two Years (2 Years) from the date of purchase against defects in materials or workmanship under normal use. The guarantee does not cover defects or damage from abuse or failure to follow user instructions. The guarantee is void if the device is opened, tampered with, or used with non-Braun branded parts/accessories, or if repairs are undertaken by unauthorized persons. Accessories and consumables are excluded from warranty.

For support requests, visit www.BraunHealthcare.com/uk_en/ or find service contact information in the manual. This guarantee applies to Europe, Russia, Middle East, and Africa. UK consumers' statutory rights are not affected.

Documents / Resources

BRAUN IRT6525 ThermoScan 7+ Ear Thermometer [pdf] Instruction Manual

References

Models: IRT6525, IRT6525 ThermoScan 7 Plus Ear Thermometer, ThermoScan 7 Plus Ear Thermometer, Ear Thermometer, Thermometer

File Info : application/pdf, 12 Pages, 478.48KB

PDF preview unavailable. Download the PDF instead.

original

References

wkhtmltopdf 0.12.6 Qt 4.8.7

Related Documents

Preview Braun ThermoScan 7+ Ear Thermometer User Manual
Comprehensive user manual for the Braun ThermoScan 7+ Ear Thermometer (Model IRT6525). Provides instructions on product description, intended use, warnings, how to use, temperature taking tips, troubleshooting, product specifications, and calibration.
Preview Braun IRT6525 ThermoScan 7+ Ear Thermometer Instruction Manual
Comprehensive instruction manual for the Braun IRT6525 ThermoScan 7+ Ear Thermometer, detailing product description, warnings, usage instructions, temperature settings, care, troubleshooting, and specifications.
Preview Braun ThermoScan 7 FAQs: Your Guide to Accurate Temperature Readings
Find answers to common questions about the Braun ThermoScan 7 digital ear thermometer, including error messages, display indicators, proper storage, and fever guidance for accurate health monitoring.
Preview Braun ThermoScan 7+ Ear Thermometer: Age-Adjustable Fever Guidance
Learn about the Braun ThermoScan 7+ ear thermometer with AgeSmart technology for accurate, age-specific fever readings. Features ExacTemp stability indicator, night mode, and hygienic lens filters for family health.
Preview Braun ThermoScan 6 IRT6515 Ear Thermometer User Manual
Comprehensive user manual for the Braun ThermoScan 6 IRT6515 ear thermometer, covering features, usage instructions, warnings, troubleshooting, and specifications for accurate temperature measurement.
Preview Braun ThermoScan PRO 3000 Ear Thermometer User Manual & Instructions
Comprehensive user manual for the Braun ThermoScan PRO 3000 ear thermometer, providing detailed instructions on usage, care, troubleshooting, and temperature measurement guidelines for accurate health monitoring.
Preview Braun ThermoScan PRO 6000 Ear Thermometer: Directions for Use and User Manual
This document provides comprehensive directions for use and operational guidance for the Braun ThermoScan PRO 6000 ear thermometer, a clinical-grade device for professional medical environments. Learn about its features, setup, usage, maintenance, and troubleshooting.
Preview Braun ThermoScan 7 IRT6520 Ear Thermometer User Manual and Guide
Comprehensive guide for the Braun ThermoScan 7 IRT6520 ear thermometer, covering setup, usage, features like Age Precision™ technology, troubleshooting, and specifications.