REF CDC-34063-HPK1A
Pressure Injectable Three-Lumen PICC
Contents:
1: Three Lumen TaperFree® Catheter: 6 Fr. (2.14 mm OD) x 40 cm, Pressure Injectable, T-Port Connector, Blue FlexTip®, Contamination Guard and Placement Wire
1: GlideThrum Peel-Away Sheath: 6 Fr. x 2-3/4″ (7 cm) Radiopaque over 6 Fr. Dilator
1: Spring-Wire Guide, Nitinol, Marked: .018″ (0.46 mm) dia. x 17-3/4″ (45 cm) (Straight Soft Tip on One End – Straight Stiff Tip on Other) with Arrow Advancer
1: Injection Needle: SafetyGlideTM1 25 Ga. x 5/8″ (1.60 cm) RW
1: Safety Introducer Needle: Echogenic 21 Ga. x 2-3/4″ (7 cm) TW
1: Blunt Fill Needle: 18 Ga. x 1-1/2″ (3.81 cm)
1: Introducer Needle: Echogenic 21 Ga. x 2-3/4″ (7 cm) TW
1: Syringe: 10 mL Luer-Lock 1: Syringe: 3 mL Luer-Lock 2: 10 mL Luer-Lock Pre-Filled Saline Syringe 1: Skin Protectant Prep
1: 5 mL 1% Lidocaine HCI Solution and Alcohol Prep
1: 3 mL Applicator 2% CHG and 70% IPA ChloraPrep®2 One-Step Solution
1: 3 mL Applicator 2% CHG and 70% IPA ChloraPrep®2 One-Step Solution with Hi-Lite OrangeTM Tint
3: Dust Cap: Non-Vented
1: SecondSiteTM Adjustable Hub: Fastener 1: SecondSiteTM Adjustable Hub: Catheter Clamp
1: SharpsAway® Disposal Cup
1: SharpsAway® II Locking Disposal Cup
1: Catheter Trimmer 1: Towel: 24″ x 36″ (61 cm x 91 cm)
1: Maximal Barrier Drape TM with two 13 cm (5″) x 15 cm (6″) fenestrations
1: Towel
1: Safety Scalpel: #11
1: Sterile Procedure Sign
1: Patient ID Card
1: Chart Sticker
1: Patient Information Booklet
1: Medication Label: 1% Lidocaine
1: Checklist/CLIP Sheet
2: Paper Tape Measure
1: Tourniquet 5: Gauze Pad: 2″ x 2″ (5 cm x 5 cm)
10: Gauze Pad: 4″ x 4″ (10 cm x 10 cm)
1: Dressing: Tegadermes 15.5 cm x 10 cm
1: Surgical Apparel: Impervious Gown
1: Dressing: BIOPATCH®4
1: Dressing: STATLOCKe5 Catheter Stabilization Device
1: Surgical Apparel: Mask with Eye Shield
1: Surgical Apparel: Bouffant Cap
1: Surgical Apparel: Mask
1: Tape: Steri-Strip®3
1: HemoHopper® Fluid Receptacle
1A trademark of Becton, Dickinson and Company.
2A registered trademark of CareFusion or one of its subsidiaries.
3A registered trademark of 3M Company.
4A registered trademark of Johnson & Johnson Corporation.
5A registered trademark of C. R. Bard, Inc.
Not made with natural rubber latex.
Store between 20 — 25°C (68 — 77°F).
Arrow International provides the enclosed medication label(s) for your convenience. Please ensure that these labels are applied to the correct syringe and corresponding medication.
Contraindications: The Pressure Injectable P1CC is contraindicated wherever there is the presence of device-related infections or the presence of thrombosis in the intended insertion vessel or catheter pathway. A clinical assessment of the patient must be completed to ensure no contraindications exist.
![]() Lumen |
Priming Volume* (mL) |
Gravity Flow Rated (mL/hr) |
Pump Flow Rate (mL/hr) |
MAX Pressure Injection Flow Rate** (mL/sec) |
Distal (17 Ga.) Medial (18 Ga.) Proximal (18 Ga.) |
0.55 0.44 0.42 |
1187 343 363 |
6800 2820 2920 |
6 not rated _ not rated |
* Priming volumes are approximate and are done without accessories. t Flow rate values are approximate and are determined using deionized water at 100 cm head height.
++Pump flow rates are determined at a maximum pump pressure of 10 psig and represent
approximate flow capabilities.
**Pressure injection flow rates are determined at the injector pressure setting of 300 psi maximum using media of 11.8 centipoise viscosity, with 152 cm pressure tubing.
LBL056585 R01 (2022-04)