PeakTech 8010 Sound Level Calibrator

Bedienungsanleitung / Operation manual

Schallpegelkalibrator / Sound Level Calibrator

1. Safety Precautions

This product complies with the requirements of the following directives of the European Union for CE conformity: 2014/30/EU (electromagnetic compatibility), 2014/35/EU (low voltage), 2011/65/EU (RoHS). Pollution degree 2.

To ensure safe operation of the equipment, the following safety precautions must be observed. Damages resulting from failure to observe these safety precautions are exempt from any legal claims whatsoever.

2. Cleaning the cabinet

Clean the equipment only with a damp, soft cloth and a commercially available household cleanser. Ensure that no water gets inside the equipment to prevent possible shorts and damage to the equipment.

3. Introduction

The sound level calibrator PeakTech® 8010 meets the requirements of EN 60942: 2003 Class 2, which is used to check the accuracy of sound level meters. The calibrator generates a defined sound level signal with a sound level of 94 dB (decibels) or 114 dB at a sound frequency of 1000 Hz. Sound level meters equipped with a microphone with a diameter of 1/2" (12.7mm) can be inserted into the calibration port of the calibrator. The specified accuracies are given at an atmospheric pressure of 1013 hPa, an ambient temperature of 23°C, and a humidity of 65% RH.

4. Front and Panel Description

The device features the following controls and indicators:

[Description of the device image: The front view shows a blue and white cylindrical device with the PeakTech logo, model number 8010, and the selection switch. The back view shows warning labels, CE marking, battery compartment details, and manufacturer information.]

5. Measurement Operation

  1. Insert the microphone of your sound level meter into the calibration port of the PeakTech 8010. Ensure the microphone is fully inserted.
  2. Select the required sound level for testing (94 dB or 114 dB) using the selection switch (2).
  3. Read the measured value on your sound level meter's display.
  4. If the measured value does not match the calibration signal, readjust your sound level meter. Refer to your sound level meter's manual or consult your dealer for assistance.
  5. After each measurement, turn off the sound level calibrator by selecting the "OFF" position of the switch.

6. Technical Specifications

ParameterSpecification
Output signal94 dB or 114 dB sound pressure level
Output frequency1000 Hz; +/-4%
Reference conditionsTemperature: 23°C (73°F)
Pressure: 1013 mbar
Humidity: 65% RH
Accuracy+/- 0.5 dB
Low Battery testThe calibrator stops the output of a sound pressure level when the battery voltage falls below the minimum requirement.
Standards complianceEN 60942:2003 Class 2
CE: EN 61326-1, EN 61326-2-1
Dimensions (WxHxD)51 x 120 x 43mm
Weight130 g

7. Replacing the battery

If no sound pressure level is output, the battery needs to be replaced. Follow these steps:

  1. Loosen the screw on the back of the battery compartment cover and remove the cover.
  2. Disconnect the old battery from the battery cable and remove it from the battery compartment.
  3. Connect the new 9V battery to the battery cable and place the battery in the compartment.
  4. Replace the battery compartment cover and secure it with the screw.

8. Notification about the Battery Regulation

The delivery of many devices includes batteries, which serve, for example, to operate remote controls. Batteries or accumulators may also be built into the devices themselves. In connection with the sale of these batteries or accumulators, PeakTech is obliged under the Battery Regulations to notify its customers of the following:

Please dispose of old batteries as required by law. Disposal in household waste is expressly forbidden by the Battery Regulations. Dispose of them at a municipal collection point or return them to a local shop free of charge. Batteries obtained from PeakTech can be returned free of charge after use at the address provided on the last page, or by posting them back with sufficient postage.

Batteries containing hazardous substances are marked with a symbol consisting of a crossed-out waste bin and the chemical symbol (Cd, Hg, or Pb) of the heavy metal responsible for the classification as a pollutant:

9. Copyright and Disclaimer

All rights, including translation, reprinting, and reproduction of this manual or parts thereof, are reserved. Reproduction of any kind (photocopying, microfilm, or other processes) is only permitted with written authorization from the publisher. This manual is based on the latest technical knowledge. Technical changes in the interest of progress are reserved. PeakTech herewith confirms that all devices meet the specifications mentioned in our documentation and are delivered factory-calibrated. A recalibration after 1 year is recommended.

Copyright © PeakTech® 07/2021/Po./Mi./Ehr.

PeakTech Prüf- und Messtechnik GmbH, Gerstenstieg 4, DE-22926 Ahrensburg / Germany. Phone: +49-(0) 4102-97398 80, Fax: +49-(0) 4102-97398 99, Email: info@peaktech.de, Website: www.peaktech.de

Models: 8010, Sound Level Calibrator, Level Calibrator, Sound Calibrator, Calibrator

File Info : application/pdf, 16 Pages, 625.23KB

PDF preview unavailable. Download the PDF instead.

9b19d568021ffa189de18b446f9a9f6d871d

References

Microsoft Word 2016 Microsoft Word 2016

Related Documents

Preview PeakTech 5175 Digital Sound Level Meter Operation Manual
Comprehensive operation manual for the PeakTech 5175 Digital Sound Level Meter, detailing safety instructions, features, technical specifications, operating procedures, and battery information.
Preview PeakTech 5175 Digital Sound Level Meter Operation Manual
Comprehensive operation manual for the PeakTech 5175 digital sound level meter, covering features, specifications, safety precautions, and usage instructions for accurate sound level measurements.
Preview PeakTech 8005 Sound Level Meter User Manual
Comprehensive user manual for the PeakTech 8005 Sound Level Meter, covering safety precautions, operation, PC software integration, technical specifications, and battery replacement.
Preview PeakTech 8005 Digital Sound Level Meter Operation Manual
Comprehensive operation manual for the PeakTech 8005 Digital Sound Level Meter, covering safety precautions, features, measurement procedures, PC connectivity, software installation, calibration, and technical specifications. Learn how to accurately measure sound levels with A/C weighting, fast/slow response, and data logging capabilities.
Preview PeakTech 5175 Digital Sound Level Meter Operation Manual
Comprehensive operation manual for the PeakTech 5175 Digital Sound Level Meter, covering safety precautions, features, technical specifications, operating instructions, and battery replacement.
Preview PeakTech 8005 Digital Sound Level Meter Operation Manual
The PeakTech 8005 Digital Sound Level Meter Operation Manual provides comprehensive instructions for operating, calibrating, and maintaining the device. Learn about its features, including A/C weighting, fast/slow response, data logging, and PC connectivity for real-time measurements and data analysis. This guide covers safety, technical specifications, and software installation for Windows 98/2000/XP/VISTA/WIN 7.
Preview PeakTech P 8005 Professional Sound Level Meter with Datalogger
Discover the PeakTech P 8005 Professional Sound Level Meter with Datalogger, a versatile digital device for precise noise measurement in various environments. Featuring a 4-digit LCD, USB interface, and internal memory for up to 32,000 readings, it's ideal for long-term measurements and quality control. Compliant with IEC-61672-1 class 2 standards, it offers 30-130 dB range, A/C weighting, and fast/slow response.
Preview Manuale Operativo Fonometro PeakTech 8005 - Misuratore di Livello Sonoro Professionale
Scopri il manuale operativo completo per il fonometro digitale PeakTech 8005. Guida all'uso, installazione software, calibrazione e specifiche tecniche per misurazioni accurate del livello sonoro da 30 a 130 dB.