Druck UPS4E Series Loop Calibrator Owner’s Manual

UPS4E Series Loop Calibrator

Product Information:

Product Name: UPS4E series Druck loop
calibrator

Features and Benefits:

  • Rugged and compact design for one-handed operation
  • Essential tool for loop testing/powering process control mA
    loops and devices
  • Easy-to-read display with time-saving features like step, span,
    and valve check
  • Utilizes advanced electrical calibration technology

Specifications:

Function Range Resolution 1 year accuracy

Usage Instructions:

1. Display:

The graphic display provides high contrast back-light for
daylight readability. It shows the main measurement with percentage
of FS reading.

2. Measure or Source 0 to 24 mA:

An internal 24V loop supply is available in both measure and
source modes. In Source mode, the current is actively controlled as
set by the UPS4E. In Measure mode, the current is controlled by the
external equipment being tested.

3. Simultaneous % Readout:

The display shows mA and the percentage value of 4 to 20 mA or 0
to 20 mA.

FAQ:

Q: Can the UPS4E calibrator measure voltage?

A: Yes, it can measure voltage with a resolution of 0.001.

Q: What is the internal memory capacity for datalogging?

A: The UPS4E has an internal memory capacity of 100,000 points
for datalogging.


“`

UPS4E series

Druck loop calibrator
The UPS4E is a rugged and extremely compact loop calibrator, designed to be operated with one hand. It is an essential tool for loop testing/powering process control mA loops and mA devices. Perfect for instrument maintenance with an easy-to-read display and simple-to-use time saving features including step, span and valve check.
The UPS4E utilises the most advanced electrical calibration technology to create a dependable, cost-effective loop calibrator.

Features and benefits
· Measure or source 0 to 24 mA · Accuracy 0.01% of reading+1 count · Dual mA and % readout, linear or flow · Step, span check, valve check, ramp. · 30 VDC measurement and continuity check · 100,000 datalog points · In built 250 resistor for supporting HART operation · 24 V DC loop power with mA measure or source mode · Auto shutdown to conserve battery life

Display
The graphic display provides a robust design with high contrast back-light for daylight readability. The unit shows the main measurement with percentage of FS reading.

Measure or source 0 to 24 mA
An internal 24V loop supply is available in both measure and source modes, essential during plant shutdowns.
In Source mode the current is actively controlled as set by the UPS4E.
In Measure mode the current is controlled by the external equipment being tested.

Simultaneous % readout
Displays mA and the percentage value of 4 to 20 mA or 0 to 20 mA.

Step, span check and valve check
Step quickly through linearity tests and zero/span adjustments. In valve mode, seating can be checked and the actuator set.

Step mode 4 to 20 mA linear 0 to 20 mA linear 4 to 20 mA flow 0 to 20 mA flow 4 to 20 mA span check 0 to 20 mA span check 4 to 20 mA valve 0 to 20 mA valve

0%

25%

4

8

0

5

4

5

0

1.25

4

0

3.8, 4, 4.2 –

0, 0.2

50% 12 10 8 5 12 10

75% 16 15 13 11.25 –

100% 20 20 20 20 20 20 19, 20, 21 19, 20, 21

UPS4E specifications

Function

Range

Resolution

1 year % rdg

+acccouurnatcsy(1)

Source mA with external loop power

20 mA 24 mA

0.001 0.001

0.01% Rdg + 1 count 0.01% Rdg + 2 counts

Measure mA with external loop power

±20 mA ±24 mA

0.001 0.001

0.01% Rdg + 1 count 0.01% Rdg + 2 counts

Source mA with internal loop power

20 mA 24 mA

0.001 0.001

0.01% Rdg + 1 count 0.01% Rdg + 2 counts

Measure mA with internal loop supply

20 mA 24 mA

0.001 0.001

0.01% Rdg + 1 count 0.01% Rdg + 2 counts

Measure V

±20 V ±30 V

0.001 0.001

0.01% Rdg + 1 count 0.01% Rdg + 2 counts

(1) Accuracy includes temperature effects 10°C to 30°C (50°F to 86°F). For use outside these limits add 0.003%/°C (0.015%/°F)

Auto step and ramp
Outputs have adjustable ramp/step rates (1 to 599 seconds) for singlehanded loop testing, valve set-up and slew rate checks.

“Fast Set” output
Allows values to be set to within 0.001 mA using the keys. Holding a key quickly ramps the output to the next set point value. This function is ideal for alarm trip tests.

30 VDC measurement
For loop diagnostics and maintenance of voltage output instruments.

Datalogging
Ability to record over 100,000 datalog points of data from the main reading channel. Users can manually activate via the settings menu and configure user set intervals via the display.
Data files can be transferred to a PC for further analysis.

Other features
· Industry standard AA batteries · Supports inrush current to a max of 40 mA for 500 ms · Battery status indicator · Information screen reports serial number, software version
and calibration date · PIN protected calibration and firmware updates.

General specifications

Internal memory

100,000 point data logging memory

Operating temperature

-10 to 50°C (14 to 122°F)

Storage temperature -20 to 70°C (-4 to 158°F)

Ingress protection

IP54. Protected against dust and splashing water from any direction

Humidity

0 to 90% RH non-condensing

Shock/vibration

MIL-PRF-28800F CLASS 2

EMC

EN 61326-1:2013

Electrical safety

EN 61010-1:2010 EN 61010-2-030:2023

Enclosure materials

High impact ABS

Loop resistance high Flashes primary reading to indicate error condition

Out of range reading Flashes “—” on primary reading

Approvals

CE marked

Display

Size: 72.7 x 42.5mm (2.86 x 1.67in). Custom segment display

Size (L x W x H)

Safe Area: 145mm X 73mm X 25mm

Weight

Safe Area: 300 g including batteries

Power supply

4x 1.5 V AA-size alkaline batteries

Battery life

Up to 72 hours in measure mode Up to 18 hours at 12 mA loop current

Auto power down

10 minutes after last key press or disabled

Battery status

Battery symbol displayed with 4 segments

Electrical Terminals

Safe Area: 2x Gold plated, 4mm sockets

HART resistor

250 Ohm optional loop resistor to support external HART communications

Connectivity

USB Type C connection to access data log and for firmware update. Safe Area variant: Device can be powered through any commercially available certified USB adapter.

Ordering information for UPS4E loop calibrator

The UPS4E is supplied with a quick start & safety guide, set of alkaline batteries, test leads, USB cable and calibration certificate.
Please use the following part numbers when ordering:

Model Type: UPS4E

Description: Safe Area loop calibrator

UKAS calibration
We also offer UKAS accredited calibrations, please advise at the time of order placement if required.

Accessories
Please state any accessories required as separate items when placing order. The UPS4E are compatible with the following accessories unless otherwise specified.

Part code IO-USB-C-CABLE IOUPS4E-CASE

Description USB A-C 2m Cable UPS4E Carry Case

Related products
Druck offer a comprehensive range of portable pressure, temperature and electrical field calibrators. To help you select the right process calibration tools we have categorised our products into three ranges. Druck’s Essential range provides rugged, accurate and easy-to-use test and calibration tools. Druck’s Expert range introduces higher accuracy and increased functionality. Druck’s Elite range provides a modular system offering state of-the-art multifunction communication and calibration.

Supporting services
Druck’s highly trained staff can support you, no matter where you are in the world. We can provide nationally accredited calibration (both initially and at periodic intervals), extended warranty terms, maintenance and even rental of portable or laboratory instruments. Please contact your local Druck customer care team for more information. Druck offers a portfolio of high accuracy, high performance test and calibration equipment in safe or hazardous area (IS) variants that are ideal for the calibration and adjustment of Druck pressure sensors.

Delivering world class pressure measurement and calibration technology

Copyright 2025 Baker Hughes Company. All rights reserved.

920-723A BHCS39573

(05/2025)

druck.com

Documents / Resources

Druck UPS4E Series Loop Calibrator [pdf] Owner's Manual
UPS4E Series, UPS4E Series Loop Calibrator, Loop Calibrator, Calibrator

References

Leave a comment

Your email address will not be published. Required fields are marked *