9769HPFS Heated In Line Filter
Pump Assembly User Manual
Part # 9769HPFS
User Manual
Heated In-line Filter Pump Assembly
Questions? Contact us at 800-223-3977 or online at
http://www.cleanair.com/equipment/Express/main.html
9769HPFS Heated In Line Filter Pump Assembly
IMPORTANT!!!
BEFORE YOU BEGIN READ THIS!!!
READ AND UNDERSTAND ALL INSTRUCTIONS BEFORE USING THIS SYSTEM!
SAVE THESE INSTRUCTIONS!!!
- To Avoid Accidents…
Keep your work area clean and well lit.
Keep bystanders away.
Exercise common sense.
- Electrical Safety…
Do not operate in combustible environments.
DO NOT operate these products when wet or in water.
ALWAYS be sure that the components of this system are running with the correct voltage (120V or 240V).
Never remove a grounding prong or modify a plug.
Do not abuse the power cord or plug.
- Personal Safety…
The Heated In-line Filter and Pump Assembly will most likely be hot during and after use. Use caution when handling equipment during or after testing.
Stay alert and watch what you are doing.
Dress appropriately. Wear the appropriate personal safety devices.
- Equipment Maintenance…
Clean Air Express can not ensure that the Heated In-line Filter and Pump Assembly is compatible with any other systems. It is highly recommended that the assembly be used in conjunction with Clean Air approved analyzers and Clean Air Express heated sample lines. See http://www.cleanair.com or call 800-223-3977 for more information.
All internal maintenance and repairs should be performed by one of Clean
Air Express’s trained technicians.
Customer Feedback
Clean Air Engineering takes pride in our quality products and services. We strive to provide the highest quality products and services in the industry. We realize the importance of end user input in the continual improvement of our products and services. Customer feedback is of paramount importance. We encourage your feedback with any suggestions or problems that can help us improve our performance. A customer feedback form is available online at http://www.cleanair.com/About/feedback.html. To emphasize our commitment to quality products and complete customer satisfaction, Clean Air Engineering’s manufacturing division, Clean Air Express, offers what we feel is the best and most comprehensive warranty in the environmental industry.
Safety
1.1 Weight and Bulk
The Heated In-line Filter Pump is relatively heavy (around 38 pounds). Use proper heavy lifting technique to avoid injury.
1.2 Temperature
The filter and pump of the Heated In-line Filter and Pump are both temperature controlled, and the possibility for burns does exist. Unplug the Heated In-line Filter Pump Assembly and allow the heated head of the pump and the heated filter to cool before handling or performing any maintenance on it. One could also use insulating gloves when allowing the pump or filter to cool would be impractical. Exercise caution when moving the assembly before cooling and be sure to use the handle to avoid burns.
1.3 Electrical Shock
The Heated In-line Filter and Pump Assembly is powered by a standard 120 VAC line, meaning potentially fatal shocks are possible. The assembly is no more dangerous than many household appliances in this regard; however, care must
be taken to avoid shock. Before performing any maintenance on the assembly, turn off and unplug the apparatus from the 120 VAC line.
1.4 Operation in Ignition Prone Environments
Extreme caution must be taken when using the Heated In-line Filter and Pump Assembly in combustible or electrically charged environments, as the assembly is not intrinsically safe. The apparatus can potentially ignite flammable gasses, starting a fire or causing an explosion. Use extreme care and thoroughly evaluate the operating environment prior to use.
Principles of Operation
Maintaining a steady temperature around 250°F is important to ensure that moisture, acid gas, and other constituents of the gas stream do not condense before the analyzing equipment. This can result in the damage of equipment as well as the loss of important data. The Heated In-line Filter and Pump Assembly helps to ensure the accuracy of gas samples by providing a heated route for the gas before analysis.
2.1 Heated In-line Filter and Pump Assembly Schematic
2.2 System Components
Figure 1 shows the Heated In-line Filter and Pump Assembly. The assembly was designed for manual operation, requiring manual set up before use.
2.2.1 Temperature Controllers
The Heated In-line Filter and Pump Assembly utilizes two temperature controllers, shown in Figure 1. The temperature controllers are capable of maintaining the user-specified temperature of the filter and pump head.
Additionally, a heated blanket maintains the temperature of the airflow path between the filter and pump at 250° F. Maintaining a steady temperature around 250°F is important so that moisture, acid gas, and other constituents of the gas stream do not condense out, introducing bias and corroding equipment.
2.2.2 Sample Filters and Heated Filter Holder
The filter holder and a glass/PTFE fluorocarbon microfiber filter scrub the sample before it enters the pump. These type of filters have a 99.9% efficiency and are effective up to 500° F. The large size of the filter (1” diameter, 7” long) ensures that the heated in-line filter and pump assembly can be used in many different applications, over long periods of time. However, the filter should be periodically changed to ensure accurate and consistent sampling.
2.2.3 Heated Head Pump
The heated head pump drives the fluid flow of the sample from the extraction point to the analyzer. The pump is functional in ambient temperatures of up to 140°F and the heated head is able to heat the sample contained up to 400° F.
2.2.4 Needle Valve
The needle valve controls the flow of sample air through the filter and the supply flow rate to your analyzer. It is recommended that the user sets the flow rate of the sample air through the assembly before sampling begins. This can be done by connecting a flow meter to the inlet of the filter and adjusting to the desired flow rate using the needle valve. Be sure to disconnect the flow meter after setting your sample flow rate and before sampling, as most flow meters cannot withstand temperatures above 250° F.
2.2.5 Power Requirements
The Heated In-line Filter and Pump Assembly requires a 120 VAC line capable of supplying at least 10 Amps. The whole assembly is protected by a 15 Amp breaker.
2.3 System Operation
2.3.1 Pre-sampling
Unscrew the filter handle and visually inspect the filter element. If particulate has built up on the filter element, extract it and install a new clean filter. Connect the outlet of a flow meter to the inlet of the heated filter. Plug in the assembly and adjust the needle valve so that the pump is pulling at the desired flow rate. This flow rate will probably be determined by the analyzer being used and can be found in the analyzer’s manual. When the desired flow rate is reached, turn off the pump and disconnect the flow meter.
2.3.2 Assembly Setup
Connect the inlet of the heated filter to the sample line. Connect the outlet of the pump to the sample analyzer. Ensure that your sample collector is not currently in the sample stream and run a system leak check. Turn on the heater controller displays and set your target temperatures for the heated filter and the heated head of the pump. A video that shows this process can be found here: http://express.cleanair.com/HeatedSampleLine/temperaturecontroller.html.
When temperatures of the heated filter and heated head have reached your target temperatures, you are ready to sample, and can locate your probe in the gas sample stream.
2.3.3 Sampling
Locate your sampler in the gas stream and begin sampling. The flow rate should stay relatively constant unless you are sampling from a gas stream extremely high in particulate.
2.3.4 High Particulate Applications
If you are sampling in an extremely high particulate application, you must plan to routinely pause sampling to replace the filter if necessary so that your flow rate does not deteriorate.
Routine Maintenance and Inspection

- Always pull clean air through the assembly after completed testing to cleanse the sample gas path
- Check the electrical connections for bent, frayed, or damaged wires
- Leak-check the assembly from inlet of the filter to the outlet of the pump, ensuring that it holds both positive and negative pressure
- Inspect the filter housing gaskets for cracks or tears, and replace if necessary
- Verify that the filter and pump head can heat to approximately 250 °F or desired temperature up to 275 °F
- Lastly, general cleanliness promotes good equipment health as well as accurate test results
Troubleshooting
The Heated In-line Filter and Pump Assembly is very reliable. When something is not working, these points often can provide a solution.
- Unable to pull sufficient sample flow rate
If you cannot achieve your desired flow rate, ensure that the needle valve is in the fully open position. Then check to ensure that the filter is not clogged. If it is not, check the sample line; make sure the line is not crimped or clogged. - System cannot pass a leak check
First ensure that all of your connections to and from the assembly are tight, and connected correctly (not cross-threaded). If one of these connections is not the source of the leak, check the filter gasket to ensure that it is not cracked or broken. If no other source of leakage is found, the diaphragm to the pump may be ripped or cracked and if so, should be replaced. - Temperature controller is heating filter and pump head above temperature set points
If this is happening, check all wires and ensure that they are connected properly and not damaged. If they are connected properly and are not damaged, the relay inside of the temperature controller may be faulty, and must be replaced.
Wiring Diagram
Parts List
HEATED IN-LINE FILTER AND PUMP ASSEMBLY
Parent Part #: 9769HPFS
Type: F
Component | Description | Qty Required | UM | T |
0035C | CEM CATECO TEMP. CONTROLLER TFE COATED THERMOCOUPLE | 1 | EA | F |
0315T | WIRE | 4 | FT | R |
0316M | STD MALE THERMOCOUPLE PLUG | 2 | EA | R |
0419SNV | NEEDLE VALVE 3/8″T S.S. HOT POCKET-MET 80 JUNCTION | 1 | EA | R |
1233 | BOX | 1 | EA | R |
1233F | HEATED FILTER 1″X 7″,TYP K TCS | 1 | EA | R |
8701 | FOAM GRIP 1″X1″ 5″ LENGTH | 1 | EA | R |
9332 | SHRINK TUBING 1/4″ | 4 | FT | R |
9334 | SHRINK TUBING 1/2″ BLACK MATTE | 6 | EA | R |
9334C | SHRINK TUBING 1/2 CLEAR | 3 | FT | R |
940310A | 3 PIN CONNECTOR ASSY 3 PIN BOX MOUNT FLANGED | 1 | EA | R |
940310FL | RECEP. | 1 | EA | R |
9409 | 4 PIN CABLE CLAMP | 1 | EA | R |
9410 | 4 PIN CABLE RECEPTACLE | 1 | EA | R |
94110 | CABLE CLAMP SIZE 10 | 1 | EA | R |
9417 | 4 PIN MALE INSERT BRACKET FOR HEATED FILTER | 1 | EA | R |
9769HFB | FOR | 1 | EA | F |
9769S2015 | END CAP BLACK 80/20 MATERIAL | 4 | EA | R |
9769S3059 | BUTTON HEAD SOCKET CAP SCREW BUTTON HEAD SOCKET CAP | 4 | EA | R |
9769S3061 | SCREW | 14 | EA | R |
9769S3280 | 1/4-20 DOUBLE NUT SLIDE IN | 23 | EA | R |
9769S3287 | 1/4-20 TRIPLE NUT | 1 | EA | R |
9769S3342 | 1/4-20 FLANGED SCREW | 59 | EA | R |
9769S3382 | FASTENER T NUT | 21 | EA | R |
9769S4107 | 2 HOLE JOINING INSERT 80/20 | 1 | EA | R |
9769S4112 | 7 HOLE TEE JOINING PLATE | 1 | EA | R |
9769S4118 | 3 HOLE JOINING STRIP | 1 | EA | R |
9769S4132 | JOINING CONNECTOR 2 HOLE | 1 | EA | R |
9769S4150 | JOINING PLATE 4 HOLE 90 DEGREE | 1 | EA | R |
9769S4155 | 8 HOLE TEE JOINING PLATE | 1 | EA | R |
9769S4509 | 2 HOLE TRANSITION INSIDE CORNR | 1 | EA | R |
9769SBAR | 1″ X 1″ SLOTTED 80/20 BAR RAIL | 107 | FT | R |
9769T2 | PUMP ADI SS HEATED HEAD | 1 | EA | R |
99511 | BUMBER 1-1/2″ OD X 1/4″ SCREW | 4 | EA | R |
Our Guarantee
Clean Air Engineering warrants products to be free from defects and workmanship for a period of one year after delivery date. The sole and exclusive remedy for defective goods shall be repair or replacement of defective parts or payment price of the goods for which damages are claimed, at Clean Air Engineering’s option.
© Clean Air Engineering – June 2008