TSI OmniTrak Solution Core Module

User Calibration Guide
- Safety & Handling Notes
- Use calibration gases in a well-ventilated area.
- Do not exceed recommended gas concentrations or pressures.
- Secure gas cylinders upright and use appropriate regulators.
- Always wear PPE when handling compressed gases.
- Kanus-a ang Pag-calibrate
- Factory calibration – ensures initial accuracy
- Matag tuig
- Field calibration – compensates for real-world influences and aging over time
- At least daily or before each study
- More frequently if:
- Readings appear inaccurate or drift beyond expected range.
- It’s exposed to high VOC or CO2 environments, high humidity, dust, or pollutants.
- Regulatory or certification programs require it.
- Factory calibration – ensures initial accuracy
- Gikinahanglan nga Kagamitan
- OmniTrak™ Core module calibration shroud (7590-84)

- Zero air source and/or span gas (see Recommended Gases and Concentrations table)
- Gas regulator for ≤ 0.3 L/min
- FEP lined tubing and regulator
- OmniTrak™ Smart Station
- OmniTrak™ Core module calibration shroud (7590-84)
- Mga Instruksyon sa Pag-setup
- Start with the Core module paired to the Smart Station
- Attach the calibration shroud to the Core Module next to the probe
- Connect tubing from gas source to the calibration shroud
- On the Smart Station, tap the Connection menu (network icon in the top right corner). Find the Core module to calibrate and tap the kebab (three vertical dots) menu next to the module. Select ‘Calibrate’ and follow the instructions.
- Pag-calibrate
For the VOC-LOW sensor, Zero Calibration and Span Calibration are available. If both are performed, Zero Calibration should be completed before Span Calibration.
VOC Zero Calibration
- On the Smart Station Core Module Calibration screen:
- Select ‘VOC LOW’
- Tap ‘Calibrate’
- Tap ‘Zero’
- Apply the zero gas bottle. Tap ‘Continue.’
- Flow zero air into the sensor at ≤ 0.3 L/min
- Wait at least 2 minutes for reading to stabilize, then press ‘Apply’
- Wait for the confirmation message that the zero calibration was successful.

VOC Span Calibration
- On the Smart Station Core Module Calibration screen:
- Select ‘VOC LOW’
- Tap ‘Calibrate’
- Tap ‘Span’
- Flow VOC span gas (i.e., 10,000 ppb isobutylene) into the sensor at ≤ 0.3 L/min
- Wait at least 2 minutes for reading to stabilize, then press ‘Apply’
- Enter Span Gas Concentration (ppb)
- Wait for confirmation message that the span calibration was successful.

CO2 Span Calibration
- On the Smart Station Core Module Calibration screen:
- Select ‘CO2’
- Tap ‘Calibrate’
- Tap ‘Span’
- Flow CO2 span gas (i.e., 400 ppb CO2) into the sensor at ≤ 0.3 L/min
- Wait at least 2 minutes for reading to stabilize, then press ‘Apply’
- Enter Span Gas Concentration (ppm)
- Wait for confirmation message that the span calibration was successful.
Post-Calibration Verification
- Flow gas again and confirm sensor is reading within +/- 10% of target
- Return to ambient air and confirm baseline VOC and CO2 return to expected values.
Pag-troubleshoot
| Isyu | Posibleng Hinungdan | Solusyon |
| Napakyas ang pagkakalibrate | Dili lig-on nga mga pagbasa | Let readings stabilize before retrying |
| VOC reading too high post-zero | Contamination in zero air source | Replace or verify zero air source |
| CO2 span fails | Incorrect gas concentration | Verify gas cylinder and regulator |
Apendise
Recommended Gases and Concentrations
| Type sa Sensor | Uri sa Pag-calibrate | Gas Used | Katuyoan |
|
VOC |
Zero | Air with zero VOCs/reference zero air, ≤ 0.3 L/min | Establish baseline free of ambient VOCs |
|
Span |
The known VOC of interest, isobutylene in air is recommended. <10,000 ppb (10 ppm) recommended, ≤ 0.3 L/min | Ensure correct sensitivity to target VOCs | |
| CO2 | Span | CO2 in air, 400 to 2000 ppm recommended,
≤ 0.3 L/min |
Ensure accurate readings at expected levels |
TSI Technical Support
Please contact TSI or your local channel partner to make service arrangements and to receive a Service Request number. To fill out an online Service Request form, visit TSI Incorporated’s website sa tsi.com/support.
Gisakup ang TSI
500 nga Dalan sa Cardigan
baybayonview, MN 55126
Tel:……….+1 800-680-1220
……….+1 651-490-2860
Fax:………+1 651-490-3824
TSI Incorporated - Bisitaha ang among website www.tsi.com para sa dugang impormasyon.

TSI and the TSI logo are registered trademarks of TSI Incorporated in the United States and may be protected under other countries’ trademark registrations.
- USA Tel: +1 800 680 1220
- UK Tel: +44 149 4 459200
- France Tel: +33 1 41 19 21 99
- Alemanya Tel: +49 241 523030
- India Tel: +91 80 67877200
- China Tel: +86 10 8219 7688
- Telepono sa Singapore: +65 6595 6388
P/N 6018700 Rev. A US (10/7/2025)