heliX +
User Manual
AMINE COUPLING KIT 2
Coupling of molecules with primary amines to the ligand strand – spin column purification
Dynamic Biosensors GmbH
HK-NHS-2 v6.1
Key Features
- Allows for coupling of biomolecules with primary amines (e.g. NH2-terminus, lysines) to the Ligand strand in a single reaction tube.
- Convenient standard chemistry (NHS chemistry).
- Compatible with heliX® Adapter Chip.
- Compatible with spin column purification of the ligand-DNA conjugate (> 7 kDa).
- Coupling of multiple ligands can be performed simultaneously.
- Yields > 95 % pure ligand-DNA conjugate.
- Includes reagents for five individual conjugation reactions (approx. 10-50 regenerations each; up to max. 500).
- Compatible with automated standard regeneration process.
Workflow Overview
3-Step Conjugation Workflow
- DNA Modification
The Ligand strand is functionalized with a primary amine reactive NHS. - Ligand Conjugation
The biomolecule (ligand) is added to the functionalized Ligand strand and incubated for at least 1 h. - Purification
The Ligand strand conjugate is purified using the provided anion exchange spin columns.
After buffer exchange the conjugates are aliquoted and stored. - Ready-to-use
The conjugate stock solutions are ready to use in the heliX® biosensor.
Timeline: Hands on time < 1 h | Incubation ~ 2 h | Total ~ 3 h
Product Description
Order Number: HK-NHS-2
Table 1. Contents and Storage Information
Material | Cap | Amount | Storage |
Ligand strand NHS | Blue | 5 x | -20°C |
Buffer A [1] | Transparent | 1 x 1.8 mL | -20°C |
Buffer C(2) | Transparent | 5 x 1.8 mL | -20°C |
Buffer PE40131 | Transparent | 5 x 1.5 mL | -20°C |
Buffer E48 [31 | Transparent | 1.5 mL | -20°C |
ddH2O | Transparent | 1.5 mL | -20°C |
Crosslinker | Brown | 5 x | -20°C |
Purification spin column | Red | 10 x | 2-8°C |
2.0 mL reaction tubes for purification spin column | 10 x | RT | |
Anion exchange (AEX) spin column | 5 x | RT | |
Collection tube for AEX spin column | 10 x | RT | |
Centrifugal filter unit (3 kDa MWC0)14] | 5 x | RT | |
Centrifugation collection tube | 10 x | RT |
For research use only.
This product has a limited shelf life, please see expiry date on label.
IMPORTANT
Products may be shipped at different temperatures, but storage should adhere to the guidelines outlined in the Table.
The kit contains reagents sufficient for five conjugations of approximately 50-200 µg of biomolecule each.
The resin slurry in the purification spin column contains 0.02 % sodium azide.
Additional Materials Required
Table 2. Additional Materials
Material | Comments |
Benchtop microcentrifuge | Required speed range of between 1,000 x g to 13,000 x g |
Vortex | |
1.5 mL reaction tubes | |
UV-Vis Spectrophotometer (e.g. Nanodrop) | For determination of the Ligand strand conjugate’s concentration |
All necessary solutions and buffers are included in the kit.
Important Notes
a. The lyophilized Ligand strand may not always be found at the bottom of the tube; it could remain on the tube wall or in the tube cap. Please always check for the presence of the lyophilized Ligand strand, identifiable by its clear pellet appearance (you may need to remove the tube label to see it). If it is not at the bottom, please centrifuge the tube at high speed for a couple of minutes before dissolving the DNA in buffer. Alternatively, place the tip of your pipette near the DNA pellet and dispense the buffer directly onto it; the DNA will quickly dissolve.
b. The crosslinker will be linked to the primary amine groups (-NH2) of the ligand. Primary amines exist at the N-terminus of each polypeptide chain and in the side-chain of lysine amino acid residues.
c. Avoid using any buffers containing primary amines (i.e. Tris, Glycine) during the conjugation process (Please check Compatibility Sheet section).
d. Up to 1 mM of Dithiothreitol (DTT) can be used during the conjugation process. Avoid using 2-Mercaptoethanol or any other thiol-based reducing agents during the conjugation process. If a reducing agent is necessary, TCEP is recommended up to 1 mM.
e. Avoid using partially purified protein samples or protein samples containing carriers (e.g. BSA).
f. To ensure the highest reaction yields, the ligand should be dissolved in Buffer C. Buffer exchange is recommended prior to the conjugation process.
g. Before starting, briefly centrifuge all tubes with blue, brown and transparent caps to ensure that all material is at the bottom of the tubes.
h. If the pI of the protein is < 6, a low pH kit for conjugation (Order No: HK-NHS-3) is recommended. For more information, please email support@dynamic-biosensors.com.
3-Step Conjugation of a Biomolecule to a Ligand strand
Please read the entire protocol before starting and perform all steps without interruption.
TIP
This protocol can be performed simultaneously for multiple coupling reactions.
Avoid using partially purified protein samples or protein samples containing carriers (e.g., BSA).
Before starting allow the crosslinker to reach room temperature before use.
I. Nanolever Modification
- Dissolve Ligand strand NHS in 40 µL Buffer A prior to use, vortex until all solids are completely dissolved and briefly spin down.
- Dissolve the crosslinker (brown cap) by adding 100 µL ddH2O, vortex until all solids are completely dissolved and briefly spin down. IMPORTANT: Always use fresh compound.
- Add 10 µL of the freshly prepared linker solution to one Ligand strand aliquot. Discard the remaining linker solution from step 2.
- Vortex the reactants for 10 sec, spin down and incubate for 20 minutes at room temperature.
IMPORTANT
Do not exceed incubation time or the reaction yield will decrease. - In the meantime, equilibrate two purification spin columns (red cap) for one coupling reaction:
a. Remove the column’s bottom seal and loosen cap (do not remove cap).
b. Place the column in a 2.0 mL reaction tube.
c. Centrifuge at 1,500 × g for 1 minute to remove the storage solution.
d. Add 400 µL of Buffer C to the column’s resin bed. Centrifuge at 1,500 × g for 1 minute to remove buffer.
e. Repeat step d and discard the resulting buffer from the reaction tube. The purification spin column should now be in a dry state. - Sample loading
a. Place the columns from step 5 in new 1.5 mL reaction tubes.
b. Remove the cap of spin column number 1 and apply the sample from step 4 to the top of the resin bed.
c. Centrifuge at 1,500 × g for 2 minutes to collect the sample (flow-through). Discard the Purification spin column after use.
d. Remove the cap of spin column number 2 and apply the sample from step c to the resin bed.
e. Centrifuge at 1,500 × g for 2 minutes to collect the sample (flow-through). Discard the Purification spin column after use.
II. Ligand Conjugation
- Add approx. 100 µg (up to a maximum of 200 µg) of the ligand (concentration approx. 0.5 – 50 mg/mL) to the sample from step 6. For optimal conditions use a volume of approx. 50 µL.
EXAMPLE: Adjust protein concentration to 2 mg/mL and use 50 µL for conjugation.
IMPORTANT
Ensure the storage buffer of the ligand does not contain any primary amines, e.g. Tris buffers, glycine (please check Important Notes). - Mix the reaction by pipetting up and down and let it react at room temperature for at least 1 hour.
IMPORTANT
Do not vortex. If necessary, the reaction can be carried out at 4 °C with a longer reaction time (e.g. overnight).
III. Spin column Purification and Buffer Exchange
- Perform a purification reaction using anion exchange spin columns.
a. To equilibrate the spin column add 400 µL Buffer A and put the spin column in the provided collection tube.
Centrifuge at 2,000 x g for 5 minutes and discard the flow-through.
Note: To achieve even liquid flow-through the membrane using a fixed-angle rotor, align the printed letter (Q) toward the center of the rotor for all chromatography steps.
b. Add the complete sample to the spin column, incubate for 1 minute and centrifuge for 2 minutes at 2,000 x g.
Discard the flow-through.
c. Place the column in a new collection tube and add 100 µL of the Elution Buffer E48 to the spin column, incubate for 5 minutes and centrifuge for 2 minutes at 2,000 x g. Repeat the elution step and combine the two flowthroughs.
d. Add 200 µL of PE40 (or TE40, HE40) buffer to the eluted and combined sample from step c.
IMPORTANT
If the protein is not stable in PE40 (or TE40, HE40), please check buffer compatibility with the switchSENSE® compatibility sheet.
e. Apply the sample to the centrifugal filter unit and centrifuge at 13,000 x g (up to 14,000 x g) for 10 minutes and discard flow-through. (Please check Additional information: Buffer Exchange and Concentration with Centrifugal Filter Units). - Add 350 µL of PE40 (or TE40, HE40) buffer and centrifuge at 13,000 x g for 10 minutes. Discard the flow-through.
- Add 350 µL of PE40 (or TE40, HE40) buffer and centrifuge at 13,000 x g for 15 minutes. Discard the flow-through.
- To recover the Ligand strand conjugate, place the centrifugal filter unit upside down in a new centrifugal collection tube (provided in the kit).
Spin at 1,000 x g for 2 minutes to transfer the sample to the tube.
IV. Aliquots and Storage
- Measure the absorbance of the Ligand strand conjugate at 260 nm (= A260nm) on a UV-Vis Spectrophotometer (e.g. Nanodrop).
- Determine the concentration of the Ligand strand conjugate (c₁) by inserting (A260nm) into the following equation:
where d is the path length (usually equal to 1 cm; however, please check the UV-Vis Spectrophotometer user manual)
- For a ready to use solution for a biochip functionalization, please adjust the concentration to 500 nM (or up to 1 µM) with PE40 (or TE40, HE40) buffer (including up to 10 % glycerol, if needed) and prepare 20 µL aliquots.
- Store between -86 °C and 8 °C, as desired.
Stability of the solution is related to the stability of the ligand molecule.
IMPORTANT
Before a switchSENSE® interaction measurement, please add the appropriate adapter strand to the conjugate solution.
Additional Information
Buffer Exchange and Concentration with Centrifugal Filter Units
- Take one centrifugal filter unit, add the appropriate volume of buffer in the filter device, and cap it.
- Place capped filter device into the centrifuge rotor, aligning the cap strap toward the center of the rotor; counterbalance with a similar device.
- Spin the device at 13,000 x g (or 14,000 x g) for the given time.
- Remove the flow through and repeat steps 1-3.
- Remove the assembled device from the centrifuge and separate the filter device from the microcentrifuge tube.
- To recover the conjugate, place the filter device upside down in a clean centrifugal tube, aligning open cap towards the center of the rotor; counterbalance with a similar device. Spin for 2 minutes at 1,000 x g to transfer the sample from the device to the tube.
Compatibility Sheet
Buffer additives
The conjugation of ligands with all available coupling kits can be performed with many different additives. The following list shows all tested ones, but please note that others not listed here may also be successfully used.
Additive | Up to | Suitability Amine Coupling | Suitability Thiol Coupling |
EDTA | 1 mM | ![]() |
![]() |
DTT* | 1 mM | ![]() |
![]() |
TCEP | 1 mM | ![]() |
![]() |
Tris** | 1 mM | ![]() |
![]() |
DMSO | 2% | ![]() |
![]() |
ATP | 0.5 mM | ![]() |
![]() |
MgCl2 | 2.5 mM | ![]() |
![]() |
Glycine** | — | ![]() |
![]() |
Mannitol | 8% | ![]() |
![]() |
Glycerol | 10% | ![]() |
![]() |
Trehalose | 8% | ![]() |
![]() |
Histidin** | 30 mM | ![]() |
![]() |
Acetonitrile*** | 50% | ![]() |
![]() |
Trifluoroacetic acid | 0.10% | ![]() |
![]() |
* thiol-based reducing agents
** contains primary amines
*** caution, may harm the ligand
pH/pI
The pH value for the conjugation buffer may range from pH 5.0 to pH 8.0, depending on the ligand characteristics. When performing a conjugation of proteins with a pI of < 6, please note that using a buffer with lower pH may result in a better yield of conjugate.
Buffer | pH | Order No | Composition |
Phosphate-Citrate Buffer | pH 5 | – | 50 mM buffer salt, 150 mM NaCl |
Buffer M | pH 6.5 | BU-M-150-1 | 50 mM MES, 150 mM NaCl |
Buffer A | pH 7.2 | BU-P-150-10 | 50 mM Na2HPO4/NaH2PO4, 150 mM NaCl |
Buffer C | pH 8.0 | BU-C-150-1 | 50 mM Na2HPO4/NaH2PO4, 150 mM NaCl |
Salt concentration
For standard conjugations, 50 mM buffer salt and 150 mM NaCl (monovalent salt) are used.
When performing conjugation of strongly charged ligands, make sure that the concentration of NaCl is sufficiently high (up to 400 mM NaCl is recommended). Otherwise, precipitation of DNA may occur.
The shielding effect of monovalent sodium cations leads to DNA stabilization through neutralization of the negative charge on the sugar phosphate backbone.
Useful Order Numbers
Table 3. Order Numbers
Product Name | Amount | Order No |
heliX® Amine coupling kit 1 (proFIRE® purification) | 5 conjugations | HK-NHS-1 |
heliX® Amine coupling kit 3 (low pI biomolecules) | 5 conjugations | HK-NHS-3 |
heliX® Thiol coupling kit 1 | 5 conjugations | HK-MAL-1 |
Centrifugal filter unit (3 kDa MWCO) | 5 pcs. | CF-003-5 |
Centrifugal filter unit (10 kDa MWCO) | 5 pcs. | CF-010-5 |
10x Buffer A[1] | 50 mL (yielding 500 mL) | BU-P-150-10 |
5x Buffer B[5] | 50 mL (yielding 250 mL) | BU-P-1000-5 |
1x Buffer C[2] | 12 mL | BU-C-150-1 |
Contact
Dynamic Biosensors GmbH
Perchtinger Str. 8/10
81379 Munich
Germany
Bruker Scientific LLC
40 Manning Road, Manning Park
Billerica, MA 01821
USA
Order Information order@dynamic-biosensors.com
Technical Support support@dynamic-biosensors.com
www.dynamic-biosensors.com
Instruments and chips are engineered and manufactured in Germany.
©2025 Dynamic Biosensors GmbH
For Research Use Only. Not for use in clinical diagnostic procedures.
[2] Buffer C: 50 mM Na2HPO4/NaH2PO4, 150 mM NaCl, pH 8.0
[3] Buffer PE40: 10 mM Na2HPO4/NaH2PO4, 40 mM NaCl, pH 7.4, 0.05 % Tween, 50 µM EDTA, 50 µM EGTA
[4] For conjugation of proteins with a molecular weight higher than 20 kDa: Centrifugal filter units with a MWCO of 10 kDa can be ordered for a faster concentration process (Order No: CF-010-5).
[5] Buffer B: 50 mM Na2HPO4/NaH2PO4, 1 M NaCl, pH 7.2
Documents / Resources
![]() |
BRUKER heliX plus Amine Coupling Kit 2 [pdf] User Manual HK-NHS-2, heliX plus Amine Coupling Kit 2, heliX plus, Amine Coupling Kit 2, Coupling Kit 2, Kit 2 |