Treedix TRX5-0903-D

Treedix Mini Electronic Load Tester (TRX5-0903-D) User Manual

Model: TRX5-0903-D

1. Introduction

The Treedix Mini Electronic Load Tester (Model TRX5-0903-D) is a versatile and compact device designed for testing various power sources, including chargers, power banks, and batteries. It supports operation at up to 50V voltage, 15A current, and 120W power, providing comprehensive data monitoring for voltage, current, power, energy, capacity, discharge time, and temperature. This manual provides detailed instructions for the safe and effective use of your electronic load tester.

2. Safety Information

WARNING: ELECTRIC SHOCK HAZARD. Always exercise caution when working with electrical circuits.

  • Ensure proper ventilation during operation, especially when discharging at high power, as the device generates heat.
  • Do not exceed the maximum input voltage of 50V or current of 15A.
  • Always connect the load tester to the power source correctly, observing polarity.
  • Keep the device away from water, moisture, and flammable materials.
  • Do not attempt to disassemble or modify the device. Refer all servicing to qualified personnel.
  • The device features intelligent safety protection functions including over-voltage, under-voltage, over-current, and over-power protection. However, these are secondary safeguards and should not replace careful operation.

3. Product Overview

Familiarize yourself with the components of your Treedix Mini Electronic Load Tester.

Diagram showing the components of the Treedix Mini Electronic Load Tester, including load input, DC jack, screen, Type-C input, confirm button, rotary encoder, fan socket, heat sink, and cooling fan.

Figure 3.1: Product Construction and Description

  1. Load Input Positive: Terminal for positive connection of the device under test.
  2. Load Input Negative: Terminal for negative connection of the device under test.
  3. DC Jack: 12V DC power input for the load tester itself.
  4. Screen: 1.54-inch high-definition color screen displaying test parameters and settings.
  5. Type-C Input: USB Type-C port for 12V PD protocol power supply to the load tester.
  6. Confirm Button (OK): Used to confirm selections and start/stop tests.
  7. Rotary Encoder (SET): Used to navigate menus and adjust parameter values.
  8. Fan Socket: Connection for the cooling fan.
  9. Heat Sink Aluminum Block: Dissipates heat generated during discharge.
  10. Cooling Fan: Activates automatically to cool the heat sink when needed.

4. Specifications

The following table outlines the technical specifications of the Treedix Mini Electronic Load Tester.

FeatureSpecification
BrandTreedix
ModelTRX5-0903-D
Power Source (for tester)DC 12V or Type-C PD 12V
Max. Input Voltage (Device Under Test)50 Volts
Max. Operating Current15 Amps
Max. Power Dissipation120 Watts
Display1.54-inch High-Definition Color Screen
Protection FeaturesOver-voltage, Under-voltage, Over-current, Over-power, Over-temperature
Included ComponentsLoad Tester, Adapter Board
Dimensions (approx.)100mm x 80mm x 53mm (3.94in x 3.15in x 2.09in)
Upper Temperature Rating90 Degrees Celsius
Diagram showing the dimensions of the Treedix Mini Electronic Load Tester: 100mm (3.94in) length, 80mm (3.15in) width, and 53mm (2.09in) height.

Figure 4.1: Product Dimensions

5. Setup

5.1. Powering the Load Tester

The Treedix Mini Electronic Load Tester requires a 12V power supply for its own operation. It supports two methods:

  • DC 12V Power Supply: Connect a standard 12V DC power adapter to the DC jack on the left side of the main circuit board.
  • Type-C PD 12V Power Supply: Connect a USB Type-C to Type-C cable from a PD protocol compatible 12V power supply to the Type-C input on the back side of the main board, below the display.

Important: This load instrument cannot be powered via Type-C to Type-A cables. The input power supply must provide 12V. If the voltage is incorrect, a "Power Error!" message will be displayed.

Diagram illustrating two power supply methods for the load tester: 1. DC12V Power Supply via barrel connector, and 2. Type-C to Type-C Power Supply. A warning about Type-C to Type-A cables and incorrect voltage is also shown.

Figure 5.1: Power Supply Methods

5.2. Connecting the Device Under Test (DUT)

Use the provided adapter board and cables to connect the device you wish to test (e.g., battery, power bank, charger) to the load input terminals marked "Load +" and "Load -". Ensure correct polarity.

Image demonstrating the wiring method, showing the load tester connected to an external power supply unit with red and black wires, and the load tester itself powered via a USB-C cable.

Figure 5.2: Wiring Method Example

6. Operation

6.1. Display Parameters Overview

The 1.54-inch color screen displays various real-time and accumulated test parameters. Understanding these parameters is key to effective testing.

Screen display of the electronic load tester showing various load parameters: U (Load Real-time Voltage), I (Load Real-time Current), P (Load Real-time Power), E (Load Accumulated Energy), Q (Load Accumulated Capacity), T (Load Cumulative Operating Time), OWH (Discharge Energy Setting), OAH (Discharge Capacity Setting), OVT (Discharge Time Setting), Constant Current Mode, IOP (Load Operating Current Setting), TEMP (Temperature), Error Warning, and Load Switch.

Figure 6.1: Display Parameters

  1. U: Load Real-time Voltage
  2. I: Load Real-time Current
  3. P: Load Real-time Power
  4. E: Load Accumulated Energy (Wh)
  5. Q: Load Accumulated Capacity (Ah)
  6. T: Load Cumulative Operating Time
  7. OWH: Discharge Energy Setting (automatic cutoff when reached; 0 means disabled)
  8. OAH: Discharge Capacity Setting (automatic cutoff when reached; 0 means disabled)
  9. OVT: Discharge Time Setting (automatic cutoff when reached; 0 means disabled)
  10. MODE:CC: Constant Current Mode indicator
  11. IOP: Load Operating Current Setting
  12. TEMP: Temperature (automatic over-temperature protection)
  13. LVP: Low Voltage Protection (Error Warning)
  14. OFF/ON: Load Switch status

6.2. Setting Test Parameters

Use the rotary encoder (SET knob) to navigate and adjust settings:

  • Adjusting Current: Short press the SET knob to highlight the current (IOP) value. Rotate the knob to adjust the desired discharge current.
  • Setting Cutoff Limits: Press the SET knob again to move the cursor to other parameters like OWH (Watt-hour), OAH (Amp-hour), or OVT (time). Rotate the knob to set a specific cutoff value. A value of '0' disables the cutoff for that parameter.
  • Confirming Settings: Once parameters are set, press the OK button to confirm.

6.3. Starting and Stopping a Test

  • After setting the desired parameters, press the OK button (lower right corner) to start the discharge test. The 'OFF' indicator will change to 'ON'.
  • To stop the test manually, press the OK button again. The 'ON' indicator will change to 'OFF'.
  • The test will automatically stop if any of the set cutoff limits (OWH, OAH, OVT) are reached, or if a safety protection (e.g., over-voltage, under-voltage) is triggered.

6.4. Applications

The Treedix Mini Electronic Load Tester is suitable for a wide range of applications:

  • Discharge aging tests for chargers and power banks.
  • Capacity testing for lithium batteries and other battery types.
  • Testing current adapters and DC power supplies.
  • Educational experiments and electronic hobbyist projects.
Collage showing various applications of the electronic load tester, including testing DC power supplies, lithium batteries, chargers, and power banks, as well as for teaching experiments.

Figure 6.2: Typical Applications

Two images demonstrating intelligent detection: one showing the tester connected to a wall charger via a data cable, and another showing it connected to a power bank, both displaying test parameters.

Figure 6.3: Intelligent Detection Examples

7. Safety Features

The Treedix Mini Electronic Load Tester incorporates several intelligent protection mechanisms to ensure safe operation:

7.1. Over-voltage Protection (OVP) and Under-voltage Protection (LVP)

  • OVP: If the input voltage exceeds the set maximum value (default 50V), the system will automatically shut down to protect the device under test and the load tester.
  • LVP: If the input voltage falls below a user-defined minimum voltage, the system will automatically shut down to prevent over-discharge of batteries or abnormal operation.
Screen display showing 'Load Settings' with OVP (Overvoltage Protection) set to 50.00V and LVP (Undervoltage Protection) highlighted, indicating these protection features.

Figure 7.1: Overvoltage and Undervoltage Protection

7.2. Over-current Protection (OCP) and Over-power Protection (OPP)

  • OCP: The system will automatically shut down if the discharge current exceeds a maximum threshold (default 15A).
  • OPP: The system will automatically shut down if the discharge power exceeds a maximum threshold (default 120W).
Screen display showing 'Load Settings' with OCP (Overcurrent Protection) set to 15.00A and OPP (Overpower Protection) set to 120.00W, indicating these protection features.

Figure 7.2: Overcurrent and Overpower Protection

7.3. Over-temperature Protection

The device monitors its internal temperature. If the heat sink temperature rises above 40°C, the cooling fan will activate. If the temperature continues to rise to a critical level, the system will automatically shut down to prevent damage.

8. Maintenance

  • Cleaning: Use a soft, dry cloth to clean the device. Do not use liquid cleaners or solvents. Ensure no dust accumulates on the cooling fan or heat sink.
  • Storage: Store the device in a cool, dry place away from direct sunlight and extreme temperatures.
  • Inspection: Periodically inspect the cables and connections for any signs of wear or damage.

9. Troubleshooting

ProblemPossible CauseSolution
Device does not power on.Incorrect power supply voltage or connection.Ensure a stable 12V DC or Type-C PD 12V power supply is connected correctly. Check for "Power Error!" message.
Test stops unexpectedly.Safety protection triggered (OVP, LVP, OCP, OPP, Over-temperature) or cutoff limit reached.Check the display for error messages. Verify input voltage, current, power, and temperature. Adjust cutoff limits if necessary.
Readings appear inaccurate.Poor connection or faulty cables.Ensure all connections are secure and cables are in good condition.
Cooling fan not running.Fan only activates when heat sink temperature exceeds 40°C.This is normal operation. If the device is hot and the fan is still not running, contact support.

10. Customer Support

For further assistance, technical support, or warranty inquiries, please contact Treedix customer service through the retailer where you purchased the product or visit the official Treedix website.