1. Introduction
The ANENG SZ302 is a versatile digital multimeter designed for measuring various electrical properties. This device is suitable for a wide range of applications, including automotive repair, home electrical work, educational purposes, and professional use in engineering and electrical fields. It accurately measures AC/DC voltage, AC/DC current, resistance, continuity, diodes, capacitance, and includes a Non-Contact Voltage (NCV) detection feature and a 1.5V battery test function.
This manual provides essential information for the safe and effective use of your ANENG SZ302 Digital Multimeter. Please read it thoroughly before operation.

Figure 1: ANENG SZ302 Digital Multimeter with included test leads and AA batteries.
2. Safety Information
Always adhere to safety precautions when using electrical testing equipment. Failure to do so may result in electric shock, injury, or damage to the meter or equipment under test.
General Safety Guidelines:
- Do not exceed the maximum input values for any function.
- Ensure the test leads are properly connected and the function switch is set to the correct range before making measurements.
- Exercise extreme caution when working with voltages above 36V DC or 25V AC RMS, as these can pose a shock hazard.
- Inspect test leads for damaged insulation or exposed metal before use. Replace if damaged.
- Do not operate the meter if it appears damaged or if the case is open.
- Remove test leads from the circuit before changing functions or ranges.
- Always disconnect power to the circuit and discharge all high-voltage capacitors before measuring resistance, continuity, or diodes.
- The meter is designed with fuses for anti-burn and overload protection. However, proper usage is essential for safety.
3. Components and Controls
Familiarize yourself with the parts of your ANENG SZ302 Multimeter:

Figure 2: Labeled diagram of the ANENG SZ302 Digital Multimeter.
Key Components:
- LCD Backlit Screen: Displays measurement values, units, and function indicators.
- Rotary Function Switch: Selects the desired measurement function (e.g., V~ for AC Voltage, V- for DC Voltage, Ω for Resistance).
- SEL Button: Selects between AC/DC modes or different sub-functions within a range (e.g., AC/DC current, Diode/Continuity).
- HOLD Button: Freezes the current reading on the display. Press again to release.
- Flashlight / Backlight Button: Activates the built-in flashlight and/or screen backlight.
- NCV Sensor: Detects non-contact AC voltage.
- Input Jacks:
- COM (Common) Jack: For the black (negative) test lead.
- VΩmA™+ (Positive) Jack: For the red (positive) test lead for voltage, resistance, continuity, diode, capacitance, and small current measurements.
- 10A MAX (Positive) Jack: For the red (positive) test lead for large current measurements (up to 10A).
- Protective Silicone Case: Provides durability and protection against drops.
- Fold-out Stand: Allows for hands-free operation.
4. Setup
Battery Installation:
- Ensure the multimeter is turned off and test leads are disconnected.
- Locate the battery compartment cover on the back of the unit.
- Use a screwdriver to open the battery compartment.
- Insert two AA 1.5V batteries, observing the correct polarity (+/-) as indicated inside the compartment.
- Replace the battery compartment cover and secure it with the screw.
Connecting Test Leads:
- For most measurements (voltage, resistance, continuity, diode, capacitance, and small current), insert the black test lead into the COM jack and the red test lead into the VΩmA™+ jack.
- For large current measurements (up to 10A), insert the black test lead into the COM jack and the red test lead into the 10A MAX jack.
5. Operating Instructions
The ANENG SZ302 offers various measurement functions. Always select the appropriate function and range using the rotary switch before connecting the test leads to the circuit.

Figure 3: Overview of measurement functions and connections.
5.1. DC Voltage Measurement (V-)
- Set the rotary switch to the desired DC Voltage range (e.g., 2V, 20V, 200V, 600V).
- Connect the black test lead to the COM jack and the red test lead to the VΩmA™+ jack.
- Connect the test probes in parallel across the component or circuit to be measured.
- Read the voltage value on the LCD.
5.2. AC Voltage Measurement (V~)
- Set the rotary switch to the desired AC Voltage range (e.g., 2V, 20V, 200V, 600V).
- Connect the black test lead to the COM jack and the red test lead to the VΩmA™+ jack.
- Connect the test probes in parallel across the AC voltage source.
- Read the voltage value on the LCD.
5.3. DC Current Measurement (A-)
- Important: Ensure the circuit is de-energized before connecting the meter in series.
- Set the rotary switch to the desired DC Current range (e.g., 20mA, 200mA, 10A).
- For currents up to 200mA, connect the black test lead to COM and the red test lead to VΩmA™+. For currents up to 10A, connect the red test lead to 10A MAX.
- Break the circuit and connect the meter in series with the load.
- Apply power to the circuit and read the current value on the LCD.
5.4. AC Current Measurement (A~)
- Important: Ensure the circuit is de-energized before connecting the meter in series.
- Set the rotary switch to the desired AC Current range (e.g., 20mA, 200mA, 10A).
- For currents up to 200mA, connect the black test lead to COM and the red test lead to VΩmA™+. For currents up to 10A, connect the red test lead to 10A MAX.
- Break the circuit and connect the meter in series with the load.
- Apply power to the circuit and read the current value on the LCD.
5.5. Resistance Measurement (Ω)
- Important: Ensure the circuit is de-energized and all capacitors are discharged before measuring resistance.
- Set the rotary switch to the desired Resistance range (e.g., 200Ω, 2kΩ, 20kΩ, 200kΩ, 2MΩ, 200MΩ).
- Connect the black test lead to COM and the red test lead to VΩmA™+.
- Connect the test probes across the component to be measured.
- Read the resistance value on the LCD.
5.6. Continuity Test (
)
- Set the rotary switch to the Continuity/Diode position. Press SEL if necessary to select continuity mode.
- Connect the black test lead to COM and the red test lead to VΩmA™+.
- Connect the test probes across the circuit or component.
- If the resistance is below approximately 50Ω, the buzzer will sound, indicating continuity.
5.7. Diode Test (
)
- Set the rotary switch to the Continuity/Diode position. Press SEL if necessary to select diode mode.
- Connect the black test lead to COM and the red test lead to VΩmA™+.
- Connect the red probe to the anode and the black probe to the cathode of the diode.
- The forward voltage drop will be displayed. Reverse the probes; an open circuit (OL) reading indicates a good diode.
5.8. Capacitance Measurement (F)
- Important: Ensure the capacitor is fully discharged before measurement to prevent damage to the meter.
- Set the rotary switch to the Capacitance range (e.g., 20nF, 200nF, 2µF, 20µF, 200µF, 2000µF).
- Connect the black test lead to COM and the red test lead to VΩmA™+.
- Connect the test probes across the capacitor.
- Read the capacitance value on the LCD.
5.9. Non-Contact Voltage (NCV) Detection
- Set the rotary switch to the NCV position.
- Move the NCV sensor (located at the top of the meter) close to the AC voltage source.
- The meter will emit an audible beep and the NCV indicator will light up, with increasing frequency as it gets closer to the voltage source.

Figure 4: Using the NCV function to detect AC voltage.
5.10. 1.5V Battery Test (BAT 1.5V)
- Set the rotary switch to the BAT 1.5V position.
- Connect the black test lead to COM and the red test lead to VΩmA™+.
- Connect the red probe to the positive terminal and the black probe to the negative terminal of the 1.5V battery.
- Read the battery voltage on the LCD.
5.11. Data Hold Function

Figure 5: The Data Hold function freezes the current reading on the display.
Press the HOLD button to freeze the current reading on the LCD. This is useful for recording measurements in difficult-to-view situations. Press the HOLD button again to release the reading and return to live measurement.
5.12. Backlight and Flashlight

Figure 6: The backlit screen and flashlight function enhance visibility in low-light conditions.
Press the Flashlight/Backlight button to turn on the screen backlight. Press and hold the button to activate the built-in flashlight. Press again to turn off.
5.13. Automatic Shut-down

Figure 7: The multimeter features an automatic shut-down function to conserve battery life.
The multimeter will automatically power off after approximately 15 minutes of inactivity to conserve battery life. To reactivate, turn the rotary switch to OFF and then back to the desired function.
6. Maintenance
Cleaning:
- Wipe the meter with a damp cloth and mild detergent. Do not use abrasives or solvents.
- Ensure the meter is completely dry before use.
Battery Replacement:
When the battery indicator appears on the LCD, replace the batteries as described in the "Battery Installation" section (Section 4).
Fuse Replacement:
If the current measurement function fails, the fuse may need replacement. Refer to the specifications for the correct fuse type and rating. Fuse replacement should only be performed by qualified personnel.
7. Troubleshooting
| Problem | Possible Cause | Solution |
|---|---|---|
| Meter does not power on. | Dead or incorrectly installed batteries. | Check battery polarity or replace batteries. |
| No reading or "OL" displayed. | Incorrect range selected, open circuit, or damaged test leads. | Select a higher range, check circuit connections, or inspect/replace test leads. |
| Inaccurate readings. | Incorrect function/range, poor test lead contact, or low battery. | Verify function/range, ensure good contact, or replace batteries. |
| Current measurement fails. | Blown fuse. | Replace the fuse (refer to specifications for type). |
8. Specifications
| Parameter | Value |
|---|---|
| Model | SZ302 |
| DC Voltage | 200mV / 2V / 20V / 200V / 600V |
| AC Voltage | 2V / 20V / 200V / 600V |
| DC Current | 20mA / 200mA / 10A |
| AC Current | 20mA / 200mA / 10A |
| Resistance | 200Ω / 2kΩ / 20kΩ / 200kΩ / 2MΩ / 200MΩ |
| Capacitance | 20nF / 200nF / 2µF / 20µF / 200µF / 2000µF |
| Diode Test | Yes |
| Continuity Test | Yes (Buzzer) |
| NCV Detection | Yes |
| Battery Test | 1.5V |
| Display | Backlit LCD |
| Power Supply | 2 x AA 1.5V Batteries |
| Automatic Shut-down | Approx. 15 minutes |
| Product Dimensions | 1.97 x 1.97 x 0.79 inches (approx. 5 x 5 x 2 cm) |
| Item Weight | 0.02 ounces (approx. 0.6 grams) |
| Manufacturer | ANENG |
9. Warranty and Support
ANENG provides customer service, after-sales support, and technical assistance for the SZ302 Digital Multimeter. For any inquiries, technical issues, or warranty claims, please contact ANENG customer support through the official channels or your point of purchase.
Please retain your purchase receipt as proof of purchase for warranty purposes.