ZIBOO ET1013

ZIBOO ET1013 Digital Manometer User Manual

Model: ET1013

1. Introduction

The ZIBOO ET1013 Digital Manometer is a precision instrument designed for measuring gauge and differential pressure in various applications. It offers 11 selectable pressure units and features data hold, recording, and zero functions for accurate and reliable measurements. This manual provides essential information for the safe and effective use of your manometer.

2. Safety Information

Always observe the following safety precautions to prevent injury and damage to the instrument:

3. Package Contents

Verify that all items are present in the package:

ZIBOO ET1013 Manometer package contents including device, bag, adapters, tubes, and manual.

Figure 3.1: Contents included in the ZIBOO ET1013 package.

4. Product Overview

4.1. Device Layout and Functions

ZIBOO ET1013 Digital Manometer showing buttons for UNIT, HOLD, DIF, REC, Backlight, and Power.

Figure 4.1: Front view of the ZIBOO ET1013 Digital Manometer with key functions labeled.

4.2. LCD Display Interface

ZIBOO ET1013 Digital Manometer LCD display showing differential pressure mode, hold data, pressure measurement, Max/Min mode, and pressure units.

Figure 4.2: Detailed view of the LCD display showing various indicators and readings.

The large LCD display provides clear readings, including:

4.3. Dual Channel Inputs

Close-up of the ZIBOO ET1013 Manometer showing the dual channel input ports.

Figure 4.3: The dual channel input ports for pressure measurement.

The manometer features two input ports for measuring differential pressure. These ports are clearly marked with '+' and '-' for positive and negative pressure connections.

4.4. Dimensions

Dimensions of the ZIBOO ET1013 Digital Manometer: 5.7 inches long, 2 inches wide, 1.2 inches thick.

Figure 4.4: Physical dimensions of the ZIBOO ET1013 Digital Manometer.

The compact and slim design (approximately 5.79" x 2.16" x 1.26") allows for easy one-handed operation and portability.

5. Setup

5.1. Battery Installation

The ZIBOO ET1013 Manometer requires a 9V battery (not included). To install:

  1. Locate the battery compartment cover on the back of the device.
  2. Use a small screwdriver to open the battery compartment.
  3. Insert a new 9V battery, ensuring correct polarity.
  4. Close the battery compartment cover securely.

Video 5.1: Unboxing and battery installation for a similar digital manometer. This demonstrates the general process for installing the 9V battery.

5.2. Connecting Pressure Tubes

Connect the provided soft tubes to the manometer's input ports and to the pressure source you wish to measure.

  1. Attach one end of a soft tube to the '+' port on the manometer.
  2. Attach the other end of the tube to the high-pressure side of your system.
  3. For differential pressure, attach a second soft tube to the '-' port and to the low-pressure side of your system.
  4. Use the RMA316 tube adapters if required for specific connections.
Close-up of a pagoda pressure connector with a soft tube attached, illustrating how to connect the tubes to the manometer.

Figure 5.2: Example of connecting a soft tube to a pagoda pressure connector.

6. Operating Instructions

6.1. Power On/Off

Press the Power button to turn the manometer on. The device will perform a self-test and display readings. To turn off, press the Power button again. The device features an auto power-off function after 20 minutes of inactivity to conserve battery life.

6.2. Unit Selection

Press the UNIT button to cycle through the 11 available pressure units (e.g., psi, kPa, bar, inHg, mmHg, inH₂O, ftH₂O, cmH₂O, ozin², kgfcm², mbar). The selected unit will be displayed on the LCD.

Video 6.1: Demonstration of unit selection and other basic operations on a digital manometer.

6.3. Data Hold

Press the HOLD button to freeze the current reading on the display. The 'HOLD' indicator will appear. Press the button again to release the hold function and resume live readings.

6.4. Recording (MIN/MAX/AVG)

Press the REC button to enter recording mode. The display will show the maximum (MAX), minimum (MIN), and average (AVG) pressure values recorded since entering this mode. Short press the REC button to cycle through MAX, MIN, and AVG values. Long press the REC button for 5 seconds to exit recording mode.

Video 6.2: Overview of features including data recording and unit options on a digital manometer.

6.5. Differential Pressure Measurement

To measure differential pressure, connect the high-pressure source to the '+' port and the low-pressure source to the '-' port. Press the DIF button. The display will show the pressure difference between the two inputs. Pressing the DIF button again will reset the differential measurement to zero, allowing for quick offset correction.

6.6. Backlight

Press the Backlight button (often indicated by a light bulb icon) to turn the LCD backlight on or off. This feature enhances readability in dimly lit environments.

7. Applications

The ZIBOO ET1013 Digital Manometer is suitable for a wide range of professional and industrial applications, including:

Images depicting various application scenarios for a digital manometer, including chemical fiber industry, pressure parts, HVAC systems, and clean rooms.

Figure 7.1: Diverse applications of the digital manometer in industrial and technical settings.

8. Specifications

The following table outlines the technical specifications of the ZIBOO ET1013 Digital Manometer:

FeatureDescription
DisplayDual LCD
Accuracy±0.3%FSO (25°C)
Repeatability±0.2% (Max +/-0.5%FSO)
Linearity/Hysteresis±0.29%FSO
Maximum Pressure10psi (HT-1890)
Response Time0.5 Seconds typical
Low Battery IndicatorYes
Over Range IndicatorErr.1
Under Range IndicatorErr.2
Operating Conditions32°F to 122°F (0 to 50°C)
Storage Conditions14°F to 140°F (-1 to 60°C)
Power Supply1x 9V Battery (not included) or External 9VDC
Standard AccessoriesPouch & Tubing
Item Weight300 g
Parcel Dimensions20.6 x 10.9 x 5.69 cm
Table showing parametric indicators for the ZIBOO ET1013 Digital Manometer, including display type, accuracy, repeatability, maximum pressure, and operating conditions.

Figure 8.1: Parametric indicators of the ZIBOO ET1013 Digital Manometer.

9. Maintenance

To ensure the longevity and accuracy of your ZIBOO ET1013 Digital Manometer:

10. Troubleshooting

If you encounter issues with your ZIBOO ET1013 Digital Manometer, refer to the following common problems and solutions:

For persistent issues not covered here, please contact customer support.

11. Warranty and Support

Specific warranty details and support contact information may vary by region and retailer. Please refer to the documentation provided with your purchase or contact the manufacturer directly for assistance.

Related Documents - ET1013

Preview User Manual: Earth Clamp Tester (Model UT278A)
Comprehensive user manual for the ZIBOO Earth Clamp Tester, detailing safety precautions, technical specifications, operation methods, and field applications for measuring grounding resistance. Includes instructions for multi-point, limited-point, and single-point grounding systems.
Preview ZIBOO DC Resistance Tester UT620C User Manual
User manual for the ZIBOO DC Resistance Tester (Model UT620C), detailing safety precautions, technical specifications, operation, and accessories for measuring low DC resistance.
Preview ZIBOO FT-1000W/FT-2000W Solar MPPT Power Meter User Manual & Specifications
User manual and technical specifications for ZIBOO FT-1000W and FT-2000W Solar MPPT Power Meters, covering safety precautions, general specifications, operation, and maintenance.